Method:
Please follow guidance as below strictly.
1) First of all please power on the instrument and close the operation software, then enter “\\Dirui\CS-400
2) Please input “DiRuiWB” as password in the pop-out interface, then click “Login”
3) After login the interface as picture 2-1 will pop out, first please click “Connect” to connect upgrading
software (if “Connect” button is gray, please check whether the instrument is open or the serial port cable
is connected well), if the upgrading software is on-line successfully there will be a hint “connect ok!” as
picture 2-2 showed.
Picture 2-3
6) After upgrading complete in order to accomplish upgrading process please power off the instrument and
reboot it after 10 seconds for every upgraded firmware.