HT 10
Table of Contents
1. General Information 6
1.1 Read me.............................................................................................6
1.2 Glossary .............................................................................................6
2. Safety 7
2.1 Basic Safety Rules .............................................................................7
2.1.1 General Regulations ................................................................7
2.2 Used Symbols in this Operating Instruction .......................................7
2.3 Used Safety and Warning Signs ........................................................8
2.4 Specific Risks .....................................................................................8
2.5 Residual Risks ...................................................................................8
2.6 Organizational / Personnel.................................................................9
2.7 Training, Calibration, Qualification, Maintenance ..............................9
3. Product Description 10
3.1 General ............................................................................................10
3.1.1 Dedicated Use .......................................................................10
3.1.2 Conformity..............................................................................10
3.1.3 Unit Identification Label .........................................................10
3.1.4 Requirements for the Place of Use ........................................11
3.2 Technical Data .................................................................................12
3.2.1 Overview ................................................................................12
3.2.2 Electrical Characteristics .......................................................12
3.2.3 Level of Noise Emission ........................................................12
3.2.4 Dimensions and Weight .........................................................12
4. Assembly and Function 13
4.1 Mechanical Assembly of HT 10 .......................................................13
5. Function Description 14
6. Commissioning 15
6.1 Installation ........................................................................................15
6.1.1 Scope of Delivery (Standard Accessories) ............................15
6.1.2 Unpacking and Checking .......................................................15
6.2 Functional Components and Display ...............................................16
6.2.1 Diagram of the Control Panel ................................................16
6.2.2 Diagram of the Interfaces ......................................................17
6.3 Initial Startup ....................................................................................17
6.4 Configuration HT 10 .........................................................................18
6.4.1 Configuration EPROM ...........................................................27
6.4.2 Sensor....................................................................................27
6.5 Password Administration .................................................................28
6.5.1 General ..................................................................................28
6.5.2 Password Levels ....................................................................28
6.5.3 Access Authorization .............................................................30
1. General Information
1.1 Read me
Please read the operating instruction before starting up the unit. Please take
special note of all instructions and warnings regarding danger that could arise
when using the unit.
The operating instruction forms an integral part of the product and contains
basic instructions to be followed during installation, operation and mainte-
nance. For this reason, the operating instruction must be kept at the disposal
of all persons dealing with the system.
1.2 Glossary
Abbreviation / Term Description
CE Conformité Européenne
EP European Pharmacopeia
FDA Food and Drug Administration
HT 10 SOTAX Tablet Hardness Tester
JP Japanese Pharmacopeia
IQ Installation Qualification
SOTAX Mechanical Tablet (simple verification of HT
MT
functionality between two qualifications)
OQ Operational Qualification
PL Plausibility –Limit
PQ Performance Qualification
T1, T2, T3 Tolerance Limits
The qualification activities should produce documented
evidence that provides a high degree of assurance that
Qualification
the unit consistently performs as intended and meets
specifications.
USP US Pharmacopeia
Validation is the process of checking that a product, ser-
Validation vice or system meets specifications and that it fulfills its
intended purpose.
2. Safety
The symbols used in this operating instruction have the following meaning:
Danger:
This symbol indicates a high risk of injury to personnel. The information
given must be observed.
Warning:
This symbol indicates information which, if ignored, can lead to substantial
material damage. The safety instructions must be observed.
Important:
This symbol indicates information containing important instructions regard-
ing use. Failure to observe the instructions can lead to errors.
To prevent the risk of crushing body parts, the following notice must be fol-
lowed:
An operation of this unit shall only take place when the control box
and the cover are closed and the splinter protection cover is properly
set up.
Switch off the unit or unplug it from the power supply to cut off the electricity
completely, if e.g. the unit is not used for some time.
Place the power cable so that no one can stand on it or trip over it.
A risk management file has been created. It is only for internal use and not
intended for release to the customer. In the course of an internal audit it can
be presented to the customer for inspection.
Furthermore, they must have read and understood the chapter “safety”.
Never bypass existing security and monitoring devices. They serve to protect
people from injury. Every change, ignoring or deactivation of security and sur-
veillance devices must be previously approved by SOTAX in written form.
Please also ask SOTAX or the local representative about the range of services
(calibration, qualification IQ/OQ, and maintenance).
3. Product Description
3.1 General
The tablet hardness tester SOTAX HT 10 is designed to determine automati-
cally the physical properties of tablets, coated tablets, oblongs and similar ad-
ministration forms. The unit enables hardness, thickness, diameter/length and
weight of samples to be checked.
The dedicate use includes also the observation of these operating instruc-
tions, the adherence of personal qualifications, the safety regulations as well
as the maintenance and overhaul conditions.
3.1.2 Conformity
The HT 10 Hardness Tester meets the essential health and safety require-
ments of all relevant European Directives (see chapter 13 „Declaration of Con-
formity“).
Address of manufacturer, the unit type, the serial number and the year of man-
ufacture, the art. no. of unit, the hardness range as well as details concerning
mains voltage, frequency, power and fuse protection.
Also on the type plate shown are the CE – mark and the symbol of the directive
2002/96/EC on waste electrical and electronic equipment (crossed-out
wheeled bin symbol with a black bar underneath it) for setting collection, recy-
cling and recovery targets for all types of electrical goods.
Temperature
The ambient temperature should be between 15°C and 30°C. Avoid larger
temperature fluctuations. Keep the unit sufficiently far away from radiators.
Humidity
The relative humidity should be between 10% and 75% without condensa-
tion.
Environment
The environment should be free of dust, smoke and corrosive fumes. The
unit shall not be operated in a potentially explosive atmosphere.
Exposure to light
The unit must be protected from direct sunlight. Diffuse illumination provides
the ideal lightning.
Space requirements
We advise to have at least twice the width and depth of the unit available
(see details in chapter 3.2.4 „Dimensions and Weight“).
Vibrations
Place the unit firmly to a suitable vibration-free and stable base.
H 10 355 mm
250 mm 540 mm
4
1
2 3
5
No. Description
1 Magazine with disc
2 Cover
3 Splinter protection cover
4 Control box
5 Collecting tray
2
3 1
H 10
No. Description
Magazine with opening for feeding
1
the test samples
2 Control panel
Disc with permanent magnet and
3 numbering (turning counter-clock-
wise)
5. Function Description
The tablet hardness tester SOTAX HT 10 is designed to determine automati-
cally the physical properties of tablets, coated tablets, oblongs and similar ad-
ministration forms. The unit enables hardness, thickness, diameter/length and
weight of samples to be checked.
The hardness of a sample can be tested with the HT 10 unit according to two
different measurement principles "Constant Speed" (the tablet is broken at a
constant speed) or "Constant Force" (linear force increase up to the breakage
of the tablet). The speed and the force increase can be freely selected in order
to simulate exactly the measurement principle of various hardness testers.
During the design of HT 10, a high value has been set on a precise mecha-
nism. In addition, this testing unit, very easy to operate, is sturdy and nearly
maintenance free.
6. Commissioning
6.1 Installation
6.1.1 Scope of Delivery (Standard Accessories)
Tablet Hardness Tester
Model: SOTAX HT 10
Art. No.: 08900
Accessories not mentioned here are listed on the delivery note accompanying
the consignment.
Report any damage to the transport company immediately and inform the local
representative.
11 10
2 3 4 5 6 7 8
No. Description
1 Display
2 Function keys
3 Alignment rollers in/out
4 Alignment rollers on/off
5 SET key
6 Help
7 Menu
8 Input confirmation (ENTER)
9 Delete (Clear)
10 Input keys
11 Movement keys
5
6
3 4
Should any problems arise during installation or initial start up, switch the
unit off, unplug the power cable and contact the SOTAX service department
or your local representative.
If passwords are assigned, the following display is shown after the initiali-
zation:
After entering the password and confirmation with the ENTER-key, the unit
automatically switches over to the main menu.
If no password is assigned, the main menu display is shown right after the
initialization.
Press the F1-key (print) to print the configuration data (see chapter 7.8.4
“Configuration Print-Out”).
Defaults of units
Set the corresponding parameters using the SET-key.
Default Option
Weights mg g or mg
Dimensions mm mm or inch
Thickness, width,
diameter/length
Hardness N N, Kp, Sc oder UD
1N = 0.102 kp
1 kp = 9.81 N
1 Sc = 6.874 N
1 kp = 1.428 Sc
1 Sc = 0.7 kp
Backofft The “Backoff” distance is the distance which is added to the di-
ameter/length measured. The breaker jaw is moving at maximal
speed up to this distance and is then switched over to the meas-
urement speed.
Default 0.50 mm
Range 0.10 - 2.00 mm
Dia. Offset While using a special breaker jaw, a “Dia. Offset” can be set.
The diameter offset can be positive or negative. It is used for the
compensation calculation while making a diameter/length
measurement.
Default 00.00 mm
Range 00.00 - 30.00 mm
UD Factor The “UD Factor” (user defined hardness unit of measure) will be
User defined multiplied by the determined hardness value to calculate a user
factor defined value.
Default 10.000
Range 65.000 - 0.001
Mode Set the relevant “Mode” using the SET-key.
Const. Speed The breaker jaw is moving towards the sam-
ples at a constant speed.
Const. Force The speed of the breaker jaw is adjusted in
order to press against the sample at a con-
stant force increase.
Default: Const. Speed
Speed/Force “Speed” or “Force” is shown according to the setting in “Mode”.
At Const. Speed Speed = 1.00 mm/sec (Default)
At Const. Force Force = 10 N/sec (Default)
Range : 0.10 mm -2.50 (Const. Speed)
Range: 1-100 N/sec (Const. Force)
Sensitivity The “Sensitivity” defines the sensitivity of force measurement,
at which the tablet is detected as broken.
Default 25 %
Range 1 % - 50 %
(Notice: The smaller the percentage, the more sensitive is the
break detection )
Acceptance Range
Hardness Maximum admissible deviation to the nominal
weight at the calibration of the unit
Default 0.10 kg
Range 0.00 - 9.99 kg
Dimensions Maximum admissible deviation to the nominal
Thickness, diameter/length length at the calibration of the unit
Default 0.05 mm
Range 0.00 - 9.99 mm
Verif. Int. Set the requested “Verification Interval” for the
Calibration intervals next calibration using the SET-key.
None No request and also no entry in
the test protocol (print out)
1 day
1 week
1 month
3 months
6 months
1 year (Default)
Print out Specify which printer format is connected
A4
Letter
Label Printer (Default)
Jaw Delay After starting a test, the breaker jaw starts not until
the “jaw delay” is expired to have more orientation
time for the sample if needed.
Default 0 sec
Range 0 - 99 sec
T123 Print Set the requested test protocol parameter using the
SET-key (see chapter 7.8.2 “Method Print-Out”)
Inlimit see example below
Outlimit (Default) see example below
Lock Op. Set to “Yes”, the menu areas “Read” and “Setup” are
locked for the operator when starting a test without
method.
The access to the menu areas is possible for all user,
but changes are not possible (Default No)
Print view The “print view” can be selected either using SOTAX,
FreeWeigh (Mettler) or InfinityQS. See also the flow
chart below (Default SOTAX).
Selfcheck See description below.
Pre-settings:
Tablet
T3 = 130 %
Third tolerance range
T2 = 120 %
Second tolerance range
T1 = 110 %
T1 = 90 %
Second tolerance range
T2 = 80 %
Third tolerance range
T3 = 70 %
Result:
Pre-settings:
Tablet
T3 = 130 %
Third tolerance range
T2 = 120 %
Second tolerance range
T1 = 110 %
T1 = 90 %
Second tolerance range
T2 = 80 %
Third tolerance range
T3 = 70 %
Result:
5 tablets inside the first tolerance range (with no tolerance indication for the
values in the printout).
3 tablets outside the first tolerance range (in printout value T1)
2 tablets outside the second tolerance range > plausibility-limit (in printout
value T2).
Configuration HT
with 1, 5, C
Selfcheck
INI
Yes/No
Force
Diameter
Thickness
6.4.2 Sensor
Press the F1-key (Test) to execute the relevant steps for the hardness- and
magazine sensor.
6.5.1 General
Passwords can be assigned with a maximum of eight alphanumeric char-
acters and are only viewed with „**** ”characters.
Password entries must be confirmed with „ENTER“.
The password assignment is set after entering the characters in the first
row and with the confirmation in the second row. If the characters in the
second row do not match the first row, the password must be assigned
again.
If a password is assigned, it is required when switching on the unit and is
valid until shut-off.
It is possible to assign the same password for different levels.
Without the "Master" password, it is not possible to assign lower levels
passwords.
Deleting a password can only carry out the legitimate level by pressing the
"Clear" and "ENTER" and must be made at the first and the second line.
There is no traceability for changed passwords.
Enter Password
Operator With the "Operator" password you can start a method
and print protocols.
Method With the “Method” password, you can create, modify
and delete methods, but not changing passwords.
If an "Operator" password is assigned, the unit may be
also started with the "Method" password.
If the "Method" password is not defined, the operator
can change methods.
Enter Password
Config. With the “Config” password you can change not only
the configuration of the unit but also the passwords
“Operator”, “Method” and “Config”.
If an "Operator" password is assigned, the unit may be
also started with the "Config" password.
If a "Config" password is not defined, the "Operator" or
"Method" user can change the unit configurations.
If a “Config" password is defined, it is always needed
when accessing the menus "New", "Edit", "Delete" or
"Device". The menus will be left automatically after 5
minutes
Master With the "Master" password you can execute all the
described activities and change all passwords.
The main reason of the "Master" password assign-
ment is to have a backup in case one or more of the
above described password has been forgotten.
With no "Operator", "Method" or "Config" password,
the "Master" password is not relevant.
To go back to the main menu, switch of the unit and make a re-start after
approx. 5 sec.
To measure the weight of a sample, you can connect a balance using the RS-
232 port on the back of the unit (see chapter 6.2.2 “Diagram of the Interfaces”.
In order to communicate between HT 10 and the balance, it must be ensured
that the relevant parameters are set. The procedure is shown in the respective
balance manual.
The following manufacturers can be selected:
6.6.1 Mettler
Configuration:
9600 baud
8 bit
1 stop bit
No parity
The connecting cable is included in the chapter 8.2.2 “Accessories”.
6.6.2 Sartorius
Configuration:
mg units 3 1 13 (not possible with all unit types)
9600 baud 517
Space 522
1 stop bit 5 3 1 (standard)
software handshake 541
ASCII 5 5 1 (not selectable with all unit types)
for print 611
lineformat 7 2 1 (not selectable with all unit types)
The connecting cable is included in the scope of supply of manufacturer.
6.6.3 Precisa
Configuration:
9600 baud
8 bit
1 stop bit
No parity
No handshake
The connecting cable is included in the scope of supply of manufacturer.
7. Operation
If a password protection is assigned, you will see then the following display:
After entering the password and confirmation with the ENTER key (see chap-
ter 6.5 „Password Administration“) the unit switches automatically to the main
menu.
If no password protection is assigned, the main menu appears directly.
Hardness Yes/No
Diameter Yes/No
Thickness Yes/No
Weight Yes/No
Size of Pill Requested in order to position the sample exactly
onto the rollers. (Default 8 mm)
Spinning Speed Speed of the carousel to transport the sample.
10 % - 100 % (Default 100 %).
Press the F1-key (Ini) to initialize the unit. All data from previous measure-
ments will be deleted. The counters are set to 0 (starting point).
Press the F2-key (Read) to perform the first measurement. The counter states
how many measurements already have been performed. Up to 99 samples
can be measured.
After the first measurement (counter bigger than 0), „Back“ is shown at the F3-
key instead of „Start“. Press the F3-key (Back) in order to delete the last value
and to set the counter back by 1.
After the first measurement, the statistics may be viewed using the F4-key
(Data) (see chapter 7.4.4 „Data“). At this display, a report can be generated
(see chapter 7.8.3 „Test Protocol without Method “)
By pressing the F4-key (OK), it will be re-checked again whether the balance
is tarred. If found correct, the following display will be shown:
Once the sample has been laid on the balance and the weight stays constant,
the value is automatically taken over. After a short delay time, the next display
will be shown.
Feed the magazine manually with the sample and press the F2-key (Read) to
start the measurements.
The positioning of the tablet onto the rollers (to align the tablet) is described
in chapter 7.1.2 „Selection of Tests to be performed“).
At the first measurement, the breaker jaw moves forward with reduced speed.
The size of the sample must match with the nominal value.
Alphanumeric inputs can be entered using the control panel. Figures may be
entered directly on the corresponding keys at the control panel. For letters,
dots and spacing press the "ABC" key. The display below is shown. Use the
movement-keys (see chapter 6.2.1 „Diagram of the Control Panel“) to select
the characters and press the ENTER-key to confirm.
The main menu should be shown. If not, press the menu-key on the control
panel.
Press the F2-key (New Method). The method editing window will be shown.
With the movement-keys you can select individual rows and change the pa-
rameters. Use the ENTER-key to confirm settings.
To move between the displays, use the F4-key (Next) or F3-key (Prev).
Press the F1-key (Test) to check the “Size of Pill” and “Spinning Speed” pa-
rameters.
Press the F4-key (Next) to access the specification displays, for which a test
has to be performed.
Enter the nominal values of the parameters to be tested and confirm with the
ENTER-key.
The unit set in the configuration dialog is displayed and can be changed by
pressing the SET-key (see chapter 6.4 „Configuration HT 10 “).
The plausibility limit must be set to differ between realistic measurements and
faulty measurements.
If “Nominal” is selected the tolerance limits are based on the nominal value.
Are faulty measurements existing, pull along as many tablets as PL deviations
have taken place during measurement.
If “Average” is selected the tolerance limits are based on the average value
of measured results.
If “No Limit” is selected no tolerance limits are considered. The test report
printout is not showing any tolerance values.
If you go with the cursor to the “%”, using the SET-key you can select the
tolerance limit between relative “%” and absolute (N, mm, mg, etc.) (Default
%).
Input values T1 > T2 and T2 > T3 will be not accepted. The tolerance limits must
not be outside of the measuring range (see chapter 3.2.1 “Overview”.
Now you need to save the method. 30 storage locations are available. 10 stor-
age locations are shown. Move the cursor to one of the storage locations and
enter a name with maximum 8 characters (using the “ABC” key) or a number
under which the method has to be saved. Press the F4-key (Save) to save the
method.
By pressing the "downward" or "upward" movement-keys you can switch be-
tween 10 storage locations.
Press the F3-key (Prev) to change back to the specification display.
Press the F2-key (Exit) and Menu-key to return to the main menu.
The selected method will be not deleted until the confirmation F3-key (Yes) is
pressed.
Press the F3-key (Yes) or F4-key (No) to switch back to the “Select Method”
display.
The selected method will be not printed out until the confirmation F3-key (Yes)
is pressed.
Press the F3-key (Yes) or F4-key (No) switch back to the “Select Method”
display.
Enter now the needed information like “Batch No.”, “Press” and “Comment”.
Confirm each entry with the ENTER-key.
Press the F3-key (Yes) and the balance will be tarred. The following display
appears:
Deposit the sample on the balance. The stable value is displayed briefly and
then taken over. The following display appears:
Take the sample from the balance and place it into the magazine position 1.
Tare the balance now manually. The following display will be shown:
Deposit the next sample on the balance, make a weight measurement and
place the sample into the magazine position 2.
C D
A
B
A Status display. Shows always the status of the unit. Requests for man-
ual inputs will be shown here.
B Displays the measured values.
C Number of tests performed and remaining.
D Displays the date and time in the relevant configured format.
E Name or number of the current test. If the test has not been started
through a method, no number/name is shown under the line method.
If you want to finish the test, press the F2-key (Abort Test). The values meas-
ured up to this moment will be not stored.
Press the F4-key (Data) to view and print the statistics of the tests already
performed.
7.4.4 Data
The statistics of the already performed tests can be viewed.
Use the F3-key (Prev) and F4-key (Next) to switch among the tests (e.g. Hard-
ness and Diameter.
To switch between the Dialog Menu and the Main Menu, press the Menu-key.
Press the F1-key (View) to view all available methods, the F2-key (Edit) to edit
a method, the F3-key (Delete) to delete a method and the F4-key (Print) to
print a method (see chapter 7.8.2 „Method Print-Out“).
As a basic principle, all SOTAX units will be adjusted and calibrated prior to de-
livery.
To assure, that the unit is still suitable for the intended use, SOTAX recommends
a Re-Qualification interval of half a year to be observed.
With every start up of the unit, a check is performed to make sure that the cali-
bration date for thickness, hardness and diameter has not expired. When the
calibration interval set under chapter 6.4 „Configuration HT 10 “ has passed, the
message “Calibration date expired” will be shown.
Changes to Calibration acceptance ranges can be done in setting configuration
mode.
Select with pressing the F1-key (Hardness) or the F2-key (Diameter) the de-
sired parameter, you want to calibrate and confirm this with the ENTER-key.
Assembly
Open the control box (5) and tip it to the side. Unscrew and remove the two
screws on the load cell. Screw the calibration tower (4) to the unit with the two
short screws (Art.V100-3754) from the calibration set (Art.9276). Fix the load
cell in a vertical position and tighten with the two long screws (V100-4809)
from the calibration set (Art.9276).
Put the calibration disk (2) into the guidance (3). Position the weight (1) care-
fully in the middle of the calibration plate.
1
5 2
ENTER
3
C
9
3
8
?
7
0
SET
4
c
F4
F3
F2
F1
No. Description
1 Calibration weight
2 Calibration disk
3 Guidance
4 Calibration tower
5 Control box
Select the corresponding fields with the movement-keys. Enter the value of a
weight and confirm it with the ENTER-key. To delete a weight value, press the
“C” key and confirm it with the ENTER-key.
Calibration
Position the weight (1) entered at the first calibration point onto the calibration
disk (2) (input field no.2). Move the cursor to the field "Read". Press the EN-
TER-key to perform the measurement. The deviation (Dev.) to the nominal
weight is shown on the right side.
Once all measurements are completed, press the F4-key (Next) to enter the
next display where the “Acceptance Range” (see chapter 6.4 „Configuration
HT 10 “) and the max. determined deviation of calibration is shown.
Use the F3-key (Yes) to save the protocol. With this saving it will overwrite the
eldest protocol and set the new calibration date. If the test has to be repeated,
press the F4-key (No).
Press the F3-key (Yes) to print the protocol (see chapter 7.8.1 „Calibration “).
When the calibration work is finished, remove the calibration tower and install
the load cell. Make sure that the load cell is mounted free movable and paral-
lel.
Push the load cell towards the back of unit before tightening the screws.
Make sure, that the internal Plexiglas cover is in place and prevent a falling
down of the control box while carefully closing using both hands (see
also information sign “open+close carefully sensitive electronics inside!).
Execution of Calibration
Remove the carousel for the calibration. Position the relevant gauge block
between the breaker jaws. Move the cursor to the field "Read". Press the EN-
TER-key to perform the measurement. The deviation (Dev.) to the nominal
value is shown on the right side.
If you have entered several nominal values, proceed as described above.
Once all measurements are completed, press the F4-key (Next) to enter the
next display where the “Acceptance Range” (see chapter 6.4 „Configuration
HT 10 “) and the max. determined deviation of calibration is shown.
Use the F3-key (Yes) to save the protocol. With this saving it will overwrite the
eldest protocol and set the new calibration date. If the test has to be repeated,
press the F4-key (No).
Press the F3-key (Yes) to print the protocol (see chapter 7.8.1 „Calibration “)
Execution of Calibration
Remove the carousel for a more easy calibration. Position the first gauge block
entered at the first verification point below the thickness measuring station. If
you have entered thickness 0 as nominal value, do not insert any gauge block.
In this case, the plunger moves directly to the support. Move the cursor to the
field "Read". Press the ENTER-key to perform the measurement. The devia-
tion (Dev.) to the nominal thickness is shown on the right side.
If you have entered several thickness values, proceed as described above.
Once all measurements are completed, press the F4-key (Next) to enter the
next display where the “Acceptance Range” (see chapter 6.4 „Configuration
HT 10 “) and the max. deviation determined during the calibration is shown.
Use the F3-key (Yes) to save the protocol. With this saving it will overwrite the
eldest protocol and set the new calibration date. If the test has to be repeated,
press the F4-key (No).
As soon as the protocol is saved the next dialog appears:
Press the F3-key (Yes) to print the protocol (see chapter 7.8.1 „Calibration “ )
Mount the carousel and the splinter cover protection. Close the cover of unit.
Perform an adjustment only when the equipment no longer meets the ac-
ceptance criteria of the corresponding parameter.
ENTER
3
C
9
3
8
?
7
0
SET
4
c
F4
F3
F2
F1
No. Description
1 Calibration weight
2 Calibration disk
3 Guidance
4 Calibration tower
5 Control box
Select the corresponding fields with the movement-keys. Enter the value of
the weight and confirm with the ENTER-key. To delete a weight value, press
the ‘C’ key and confirm it with the ENTER-key.
0-value measurement
The calibration disk remains in place for the 0 value measurement. Move the
cursor to the field "rd" (read) of the first line. Press the ENTER-key to perform
the measurement.
Move the cursor to the field "st" (store). Press the ENTER-key to overwrite the
current value with the new one.
Adjustment
Position the weight (1) entered at the first calibration point onto the calibration
disk (2). Move the cursor to the field "rd" (read) of the first line. Press the
ENTER-key to perform the measurement. As described above, move the cur-
sor to the field "Set". Press the ENTER-key to overwrite the current value with
the new measured value. You may perform as much measurements as you
want before confirming a value with the “set” key.
Once all adjustments are completed, press the F4-key (Next). The new ad-
justment date will be set and you enter automatically the “calibration menu”
display.
Execution of Adjustment
Remove the carousel for the adjustment. Position the relevant gauge block
between the breaker jaws. If you have entered length 0 as nominal value, do
not insert any gauge block. In this case, the breaker jaw moves directly to the
very end. Move the cursor to the field "rd" (read). Press the ENTER-key to
perform the measurement.
You see now both “Counts”: the currently saved one and the new measured
one. Move the cursor to the field "st" (store). Press the ENTER-key to over-
write the current value with the new one. No “Counts” are shown in the second
column. The test “st” (store) is no longer shown.
Position the gauge block entered at the second adjustment between the
breaker jaws. Move the cursor to the field "rd" (read) of the second line. Press
the ENTER-key to perform the measurement. As described above, move the
cursor to the field "st" (store). Press the ENTER-key to overwrite the current
value with the new measured value.
Once all adjustments are completed, press the F4-key (Next). The new ad-
justment date will be set and you enter automatically the “calibration menu”
display.
Execution of Adjustment
Remove the carousel for the adjustment. Position the gauge block entered at
the first adjustment point below the thickness measurement station. If you
have entered thickness 0 as nominal value, do not insert any gauge block. In
this case, the plunger moves directly to the support. Move the cursor to the
field "rd" (read) of the first line. Press the ENTER-key to perform the measure-
ment.
You see now both “Counts”: the currently saved one and the new measured
one. Move the cursor to the field "st" (store). Press the ENTER-key to over-
write the current value with the new one. No “Counts” are shown in the second
column. The text “st” (store) is no longer shown.
Position the gauge block entered at the second adjustment point below the
thickness measurement station. Move the cursor to the field "rd" (read) of the
second line. Press the ENTER-key to perform the measurement. As described
above, move the cursor to the field "st" (store). Press the ENTER-key to over-
write the current value with the new measured value.
Once all adjustments are completed, press the F4-key (Next). The new ad-
justment date will be set and you enter automatically the “calibration menu”
display.
Hardness
last adjustment date: : 02.08.10
used weights: : 0.00kg
: 10.00kg
: 20.00kg
Calibration date: : 02.08.10 18:17
Nominal Actual Dev.
Weight 1 [kg]: 0.00 0.00 0.00
Weight 2 [kg]: 10.00 10.03 0.03
Weight 3 [kg]: 20.00 20.05 0.05
Acceptance Range: 0.10 Max. Dev. 0.05
Calibration: within Limit
Diameter
last adjustment date: : 02.08.10
used gauge block : 0.00 mm
: 30.00 mm
Calibration date : 02.08.10 18:20
Nominal Actual Dev.
Gauge 1 [mm]: 0.00 0.00 0.00
Gauge 2 [mm]: 4.00 4.01 0.01
Gauge 3 [mm]: 16.00 15.99 -0.01
Acceptance Range: 0.05 Max. Dev. 0.01
Calibration: within Limit
Thickness
last adjustment date: 02.08.10
used gauge block: : 0.00mm
: 4.00mm
: 20.00mm
Calibration date: : 02.08.10 18:23
Nominal Actual Dev.
Gauge 1 [mm]: 0.00 0.01 0.01
Gauge 2 [mm]: 2.00 2.01 0.01
Gauge 3 [mm]: 4.00 4.00 0.00
Gauge 4 [mm]: 16.00 15.98 -0.02
Acceptance Range: 0.05 Max. Dev. -0.02
Calibration: within Limit
Operator: ………………….. Signature: ……………………
Method No.: 4
Product No.: 2
Batch No.:
Product Name: Test Tablet
Press:
Form: Tablet
Comment:
Method Parameters
Hardness [N] Diameter [mm] Thickness [mm]
Tests 6 6 6
T3 min 150 7.84
T2 min 155 8.96
T1 min 165 10.08
Nom/Avg 171 11.20
T1 max 188 12.32
T2 max 205 13.44
T3 max 222 14.56
Measuring Parameters
Hardness [N] Diameter [mm] Thickness [mm]
1 157 T2 11.20 T1 5.42
2 169 T1 11.20 T1 5.42
3 132 >PL 11.19 T1 5.42
4 163 T2 11.20 T1 5.46
5 182 T1 11.21 T1 5.45
6 179 T1 11.20 T1 5.45
7 168 T1 11.18 T1 5.42
Statistic
Number 6 6 6
min. 157 11.18 5.42
max. 182 11.21 5.46
max.-min. 25 0.03 0.04
Avg. 169.7 11.198 5.437
Std. Dev. 9.5 0.010 0.019
RSD [%] 5.6 0.1 0.3
Method No.: 4
Product No.: 2
Batch No.:
Product Name: Test Tablet
Press:
Form: Tablet
Comment:
Method Parameters
Hardness [N] Diameter [mm] Thickness [mm]
Tests 6 6 6
T3 min 150 7.84
T2 min 155 8.96
T1 min 165 10.08
Nom/Avg 171 11.20
T1 max 188 12.32
T2 max 205 13.44
T3 max 222 14.56
Measuring Parameters
Hardness [N] Diameter [mm] Thickness [mm]
1 159 T1 11.19 5.44
2 157 T1 11.20 5.43
3 155 T1 11.20 5.46
4 177 11.19 5.45
5 165 11.21 5.43
6 171 11.22 5.44
Statistic
Number 6 6 6
min. 155 11.19 5.43
max. 177 11.22 5.46
max.-min. 22 0.03 0.03
Avg. 164.0 11.202 5.442
Std. Dev. 8.6 0.012 0.012
RSD [%] 5.3 0.1 0.2
Method No.: 4
Product No.: 2
Batch No.:
Product Name: Test Tablet
Press:
Form: Tablet
Comment:
Method Parameters
Hardness [N] Diameter [mm] Thickness [mm]
Tests 6 6 6
T3 min 12.3 30.0 30.0
T2 min 9.4 20.0 20.0
T1 min 3.5 10.0 10.0
Nom/Avg average average average
T1 max 10.0 10.0 10.0
T2 max 20.0 20.0 20.0
T3 max 30.0 30.0 30.0
Statistic
Number 6 6 6
min. 130 11.19 5.42
max. 172 11.22 5.45
max.-min. 42 0.03 0.03
Avg. 157.7 11.207 5.438
Std. Dev. 14.4 0.010 0.012
RSD [%] 9.2 0.1 0.2
Method No.: 4
Product No.: 2
Batch No.:
Product Name: Test Tablet
Press:
Form: Tablet
Comment:
Method Parameters
Hardness [N] Diameter [mm] Thickness [mm]
Tests 6 6 6
T3 min 12.3 30.0 30.0
T2 min 9.4 20.0 20.0
T1 min 3.5 10.0 10.0
Nom/Avg average average average
T1 max 10.0 10.0 10.0
T2 max 20.0 20.0 20.0
T3 max 30.0 30.0 30.0
Statistic
Number 6 6 6
min. 123 11.20 5.40
max. 177 11.22 5.44
max.-min. 54 0.02 0.04
Avg. 156.0 11.207 5.428
Std. Dev. 20.3 0.008 0.016
RSD [%] 13.0 0.1 0.3
Method Parameters
Hardness [N] Diameter [mm] Thickness [mm]
Tests 20 20 20
Test Protocol HT 10
Measuring Parameters
Statistic
Number 2 2 2
min. 110 11.22 5.43
max. 123 11.31 5.45
max.-min. 13 0.09 0.02
Avg. 116.5 11.265 5.440
Std. Dev. 9.2 0.064 0.014
% RSD 7.9 0.6 0.3
When performing a test with method and a measured value is outside the
plausibility limit (>PL) the whole packet of this tablet is not sent and an addi-
tional tablet for testing is requested.
Only the hardness measurement will be sent using the SOTAX standard.
E.g. if the hardness unit is configured with N, the lines look like this:
@ 1 210 N<cr><lf>
@ 2 217 N<cr><lf>
@ 3 219 N<cr><lf>
@ 4 216 N<cr><lf>
@ 5 212 N<cr><lf>
@ 6 218 N<cr><lf>
@ 7 219 N<cr><lf>
@ 8 213 N<cr><lf>
@ 9 221 N<cr><lf>
@ 10 212 N<cr><lf>
@ 11 214 N<cr><lf>
@ 12 222 N<cr><lf>
E.g. if the hardness unit is configured with Kp, the lines look like this:
@ 1 22.3 Kp<cr><lf>
@ 2 22.5 Kp<cr><lf>
@ 3 22.8 Kp<cr><lf>
@ 4 22.6 Kp<cr><lf>
@ 5 23.1 Kp<cr><lf>
@ 6 23.1 Kp<cr><lf>
@ 7 22.9 Kp<cr><lf>
@ 8 22.7 Kp<cr><lf>
@ 9 22.2 Kp<cr><lf>
@ 10 22.6 Kp<cr><lf>
When performing a test with method and a measured value is outside the
plausibility limit (>PL) the whole packet of this tablet is not sent and an addi-
tional tablet for testing is requested.
With a test without method, units and precision is according settings in config-
uration mode (see chapter 6.4 “Configuration HT 10 ”).
Units and precision always corresponds to units and precision shown on the
display.
A data line starts with one of the tokens : 'WGT', ' DIA', 'HRD', 'THK' according
to weight-, diameter-, hardness- and thickness-measurement. Following a '='
character is the measurement value with the corresponding unit and unit des-
ignator.
Interface configuration:
Baud rate: 9600
Data bits: 8
Stop bits: 1
Parity: none
Handshake: none
String terminator: CR / LF
Example of a strings:
HRD = 123.1VN
WGT = 567.8Vmg
DIA = 9.87Vmm
THK = 13.45Vmm
Parser definition:
3x Reply: Individual value:
Header HRD WGT DIA THK
Header position 1 1 1 1
# of strings to ignore 0 0 0 0
# of separators value 1 1 1 1
Separator = = = =
Unit N mg mm mm
Variable type Hardness Weight Thickness Length
With a “test with method” (HT 10 only) first all weight measurements are sent,
then all thickness, all diameter and finally all hardness measurements (see
example).
When performing a test with method and a measured value is outside the
plausibility limit (>PL) the whole packet of this tablet is not sent and an addi-
tional tablet for testing is requested.
Line format:
@<tablet number>, <abbreviation of measurement>, <measurement value>,
<unit>, <cr> <lf>
8. Maintenance
Maintenance and repair work should only be carried out by certified SOTAX
service department technicians or by certified SOTAX distributors.
Service Contract
In the service contract the following will be defined:
Invoice- and delivery address, service contract number, start of contract, ser-
vice interval, serial numbers of main system / auxiliary components, reserva-
tions, description of work to performed with required hours and cost (mainte-
nance, eventually qualification, travel time, board and lodging).
Phone: +91 22 42 95 01 91
E-mail: services_asia@sotax.com
8937 Cover
8686 Face
8680 Block
230585 MT holder
9. Overhauling
9.1 Warranty
The manufacturer grants a 12-month warranty from date of delivery in the
event of material or manufacturing faults. Any claims extending there be-
yond are expressly declined.
The warranty does not cover damage due to incorrect use or handling to
normal wear and tear.
If the user carries out his own repairs (without approval by SOTAX) during
the warranty period, these are not covered by the SOTAX warranty.
Liability for consequential damage or plant standstill is expressly excluded.
Products that are defective or objected should be returned to SOTAX, free
factory. Any travel and transport costs for delivery and return will be
charged to the customer.
The manufacturer grants a 6-month warranty for spare parts installed dur-
ing regular repair/maintenance work.
9.2 Trouble-Shooting
9.2.1 A Power Failure of the Unit
could be caused by the following:
The fuses at the power switch are damaged.
Always switch the unit off and unplug the power cable before replacing the fuses.
HT 10
Voltage 230 V 115 V
Fuses 2 x 6.3 AF 2 x 6.3 AF
1
No. Description
1 Voltage selector with fuses
I 2 2 Power switch
0 3 Power plug
If this display appears, the backup battery must be replaced as soon as pos-
sible. Another indication for a weak backup battery is a black point in the
main menu.
Print out the configuration of the unit and all programmed methods and con-
tact your local representative.
Attention:
When making a “Factory Reset”, all customer specific adjustments will be
overwritten with the factory settings.
10.2 Storage
SOTAX recommends using the original shipment wrapping of the unit for stor-
age.
Please contact SOTAX or your local representative, if you have any questions.
12. Disposal
SOTAX recommends contacting the local representative regarding decom-
missioning and disposal of the unit and the packaging material.