Intended Use
ME and SE models are high-resolution ME and SE balances save work and Symbols
balances of special accuracy for speed up both simple and complex The following symbols are used in
extremely precise measurement of mass. routine applications through: these instructions:
These series cover a range from
0.001 mg to 610 g. – Ultrafast response times G indicates steps you must perform
A broad range of special performance Built-in application programs; $ indicates steps you must perform
features makes the ME and SE balances application level 1: only under certain conditions
ideal for use as measuring and test – Second weight unit
equipment in ISO or GLP quality man- – Counting > describes what happens after you
agement systems. – Weighing in percent have performed a certain step
These features include: – Animal weighing
– Recalculation – indicates an item in a list
– The fully automatic self-calibrating – Calculation
and adjustment function, isoCAL – Density determination ! indicates a hazard
(time- and temperature-dependent) – Differential weighing
– Air buoyancy correction
– reproTEST for quick determination of – Air density determination Conventions Used in These
the standard deviation to check the for ME5, SE2 Operating Instructions:
repeatability of results – Diameter determination – The pictures in these Operating
Instructions are based on the
– ISO/GLP-compliant recording capability Application level 2: ME215S model. On other models,
for printouts – Checkweighing some display readouts and printouts
– Time-controlled functions may differ slightly from the ones
– Password-protected menu lock shown. This will be explained in
Application level 3: cases where this is important for
– Display of maintenance | – Totalizing operation of the balance.
service intervals when due – Formulation
– Statistics Hotline:
ME and SE balances meet the highest For advice on the use of applica-
requirements placed on the accuracy with the following additional functions: tions, just call or fax your local
and reliability of weighing results – Second tare memory Sartorius office. For the address,
through the following features: – Identification codes please visit our Internet website at:
– Product data memory www.sartorius.com
– Efficient filtering-out of vibration – SQmin function
– Manual data storage in
– Fully automatic draft shield with three application level 3
motorized, self-teaching draft shield – DKD uncertainty of measurement
elements and soft-touch technology
– Automatic initialization when you
– Stable and repeatable results switch on the balance
ME215/235/254/414/614:
– Integrated static electricity eliminator
feature to neutralize interfering electro-
static charges (ionizer)
2
Contents
3
Warning and Safety Instructions
This balance complies with the – The only way to switch the power off
European Council Directives as well as completely is to disconnect
international regulations and the AC adapter
standards for electrical equipment,
electromagnetic compatibility, and – The balance housing is protected
the stipulated safety requirements. against the penetration of solid objects
Improper use or handling, however, with a diameter of more than 2.5 mm
can result in damage and/or injury. (such as accumulated dust) and
dripping water falling vertically (IP32) –
Read these operating instructions the housing is not completely dust-
thoroughly before using your and leak-tight, however
balance to prevent damage to the
equipment. Keep these instructions in – Protect the AC adapter from contact
a safe place. with liquid
4
General Views of the Balance
Models ME215/235/254/414/614
26
1
25
2
24
23
3
22
4
21
5
20
20
6
19 7
8 8
9 9
18 10
11
17 12
16 13
15 14
Pos. Designation Spare Part Order No. Pos. Designation Spare Part Order No.
1 Draft shield door grips 17 On/off key
2 Weighing pan 69 ME0001 18 Toggle key for alphabetic input
3 Shield disk (ME235S/P only) 19 Function keys
4 Shield plate 69 ME0002 20 Leveling foot 69MA0091
5 Level indicator 21 Lug for attaching an antitheft
6 Operating panel locking device
7 Display 22 Menu access switch
8 Tare key 23 Serial printer port (PRINTER)
9 Key for opening/closing 69ME0007 24 Serial communications port
draft shield (set of small parts) (PERIPHERALS)
10 Decimal point key 25 Terminal for connecting an
11 Print key equipotential bonding conductor
12 Ionizer on/off key 26 DC jack
13 CF key (clear function)
14 Numeric keys Not shown:
15 Toggle key for changing the application program Set of dust covers 6960ME01
16 Setup key for configuring the balance Set of small parts (operating panel) 69ME0007
Set of caps 69ME0008
5
General Views of the Balance
ME36S
28
27 1
2
3
4
26
5
26
6
25
24
22
23
7
8
23
22 9
10
11
21
12
20
MES_06.cdr
19
18 13
17 14
16 15
6
General Views of the Balance
3 4
1 5
6
7
7
8
10
11
12
30
29
28 13
27
26 14
15
15
16
25
17
24
23 18
22 19
21 20
7
Operating Design
c Clear function
Deletes keypad input
Interrupts a calibration and
adjustment routine in progress
Quits application programs
8
Operating Design
1
> Switch left:
Standard/not legal for trade
Switch right:
Legal for trade
> Note:
Do not move switch 2 * = Verification scale interval “e” =
scale interval “d”
9
Tare Memory, Calculated Values: Display for Menu Parameter To set a parameter:
The symbols displayed here indicate when Settings (Setup) G Press the Q or q soft key repeatedly
there is a value in one of the tare This display is divided into three sections. until the desired setting is selected
memories or when the value shown is a (displayed inversely)
result of calculation rather than direct Line for Operating State
measurement. G Confirm your selection:
Input and Output Window
press the l soft key
These symbols are as follows:
Soft Key Labels
To change the numeric value of
a Calculated value a parameter:
Status Line: G Press the Q or q soft key repeatedly,
s1 Net value | tare memory The status line of shows the function if necessary, until the desired setting is
of the display screen page. In the Setup selected (displayed inversely)
s2 used by an application program menu, the current menu “path” is
(e.g., formulation, second tare shown here. G Enter a new value or character:
memory) use the 0 1 … 9 . keys
Setup Menu Example: “Balance/scale or the a key and enter
Application Pictograms: functions”: the desired letters
The pictograms displayed here indicate
the application(s) selected. The pictogram G Confirm your selection:
SETUP BAL.FUNC.
is displayed inversely (white on a black press the l soft key
background) when the corresponding
application is active. To exit Setup: press the oo soft key
10
Input Data Output Error Codes
If you press a key that has no function,
Bar Code Scanner or Keyboard Input The balance provides two interface or which is blocked at a certain point in
ports for outputting weights, calculated an application program, this error is
You can use a bar code scanner or an values and parameter settings: indicated as follows:
external keyboard to input alphanumeric – Serial communications port
values. These inputs are processed in (PERIPHERALS – Serial I/O) – a double beep is sounded as an acoustic
the same manner as keypad inputs on signal if the key has no function
the display and control unit of the bal- – Serial printer port
ance. Bar code and keyboard inputs are (PRINTER – Serial Out) – a double beep is sounded and the
only displayed; they cannot activate any message “No function” is displayed in
function. the text line if the key function is not
Serial Printer Port available at that time
To assign a bar code scanner or In addition to Sartorius printers (such as
keyboard input to a function, press one the YDO03-0CE), you also have the The response to an operator error is
of the following soft keys: choice of connecting a remote display identical in all operating modes. See the
– Lot or an external checkweighing display to chapter entitled “Error Codes” for a
– Samples the printer port. detailed description.
– Measured values
– Sample number You can configure the data output
– Tare value functions in the Setup menu to meet Storing Settings
– Initial weight your various requirements, including Saving Parameter Settings
– Backweighed value ISO | GLP requirements. The settings configured remain stored
– Sample ID in the balance’s non-volatile memory. In
ISO: International Organization addition, you can reload the factory set-
for Standardization tings.
Foot or Hand Switch Input
You can connect a foot switch or GLP: Good Laboratory Practice
a hand switch to the balance to have Saving Settings
this device perform a keypad You can have printouts generated Under “Setup > Device parameters >
function (such as c or w). automatically, or by pressing r; password” you can assign passwords in
generation can be dependent on or order to block access to:
independent of the stability or time – Balance | scale functions
PC Input parameters.
You can use a computer to control the – Device parameters
functions of the balance and display See the section on “Data Output
and control unit via the communica- Functions” in the chapter entitled – Application parameters
tions port (see the “Data Output “Operating the Balance” for a detailed
Function” section in the chapter description. – Printout
entitled “Operating the Balance”).
– Factory settings
Serial Communications Port
You can connect a PC, a remote display,
an external checkweighing display
or a standard (non-verifiable) printer
to this port.
11
Getting Started
§ Lift the inner package containing the balance out of the outer packaging by the strap.
12
§ Remove the following parts from the recessed spaces at the top of the
inner packaging:
– AC adapter (in cardboard packaging)
– Weighing pan
– Shield plate
§ Remove the two padding blocks that make up the inner packaging by pulling outward.
§ Remove the shield disk (included with ME235S/P models only) and dust cover.
§ Place one hand under the display and control unit and the other under the back of the
balance and lift the balance out of the lower packaging.
13
! Do not lift the balance by the draft shield or the front panel, as this can result
in damage.
! Save the box and all parts of the packaging in case it should become necessary to
transport the balance over a long distance. Only the original packaging provides the
best protection for shipment (see also “Transporting the Balance on page 20“).
Before packing the balance, unplug all connected cables to prevent damage.
Equipment Supplied
The following individual components
are supplied:
ME215/235/254/414/614 ME36S
– Balance – Weigh cell
– AC adapter with power cord – Electronic evaluation unit
– Weighing pan with hanger for below- – AC adapter with power cord
balance weighing – Weighing pan
– Shield disk – Shield disk
– Dust cover for the balance housing
– Dust cover for the display and control ME5-F, SE2-F
unit – Weigh cell
– Instruction manual – Draft shield cover
– Shield ring
ME5, SE2 – Electronic evaluation unit
– Weigh cell – Connecting cable
– Draft shield – AC adapter with power cord
– Electronic evaluation unit – Accessories kit
– Connecting cable
– AC adapter with power cord The accessories kit includes:
– Accessories kit – Filter pan, 50 mm d
– Weighing pan
The accessories kit includes: – Shield disk
– Weighing pan – Interior draft shield (only for SE2)
– Shield disk – Brush
– Interior draft shield (only for SE2) – Forceps
– Brush – Cloth
– Forceps
– Cloth
14
Installation Instructions
The ME/SE balances are designed to provide reliable weighing results under normal ambient
conditions in the laboratory and in industry. Choose the right location to set up your balance
by observing the following so that you will be able to work with added speed and accuracy:
– Set up the balance on a completely even surface on a low-vibration balance table or
wall console
– Avoid placing the balance in close proximity to a heater or otherwise exposing the
balance to heat or direct sunlight, as this can considerably increase the temperature
inside the draft shield (greenhouse effect), resulting in incorrect readouts due
to convection currents, turbulence and buoyancy effects.
– Protect the balance from drafts that come from open windows or doors
– Avoid brief fluctuations in room temperature
– Protect the balance from aggressive chemical vapors
– Do not expose the balance to extreme moisture
Moisture in the air can condense on the surfaces of a cold balance whenever it is brought
into a substantially warmer place. If you transfer the balance to a warmer area, make
sure to condition it for about 2 hours at room temperature, leaving it unplugged from
AC power. Afterwards, if you keep the balance connected to AC power, the continuous
positive difference in temperature between the inside of the balance and the outside
will practically rule out the effects of moisture condensation.
ME215/235/254/414/614:
Setting Up the Balance ME215/235/254/414/614
§ Place the components listed below inside the weighing chamber in the order given:
– 1) Shield plate
– 2) Position the weighing pan and turn to the left or right until it snaps into place
– 3) Shield disk (ME235S/P only)
3
$ Note: the shield disk minimizes the effects of drafts within the weighing chamber
2
15
ME36S: Setting Up the ME36S Balance
§ Place the components below in the weighing chamber in the order given:
– Shield plate
– Weighing pan
ME5, SE2:
Setting Up the ME5 or SE2 Balance
§ Place the components below on the weigh cell base in the order given:
– Shield disk
– Weighing pan
Please note: after placing the weighing pan on the base, turn the pan slightly towards
the left then the right while pressing down on it to lock.
– Draft shield: center the hole over the pan (see arrows)
ME5-F, SE2-F:
Setting Up the ME5-F or SE2-F Balance
§ Place the components below on the weigh cell base in the order given:
– Shield disk
– 50 mm dia. filter pan or weighing pan (or, optionally, the 75 or 90 mm dia. filter pan)
Please note: after placing the pan on the base, turn the pan slightly towards the left
then the right while pressing down on it to lock
! If the weighing pan is removed during operation, turn the balance off and then on again
after you return it to the balance.
16
§ Turn draft shield assembly by approx. 90 degrees toward the left
(loosen knurled knob to turn)
Use only
– Original Sartorius AC adapters and power cords
$ To use a main feeder cable from the ceiling or to mount a CEE plug,
have a certified electrician install it
§ Insert the AC adapter plug with the angle facing downward into the jack on the balance
17
§ Plug power cord into the AC adapter
§ To power the balance with AC current, plug the power cord into
a wall outlet (mains supply)
Safety Precautions
The AC adapter rated to Class 2 can be plugged into any wall outlet
without requiring any additional safety precautions. The ground or earth terminal is
connected to the scale housing, which can be additionally grounded, if required.
The data interface is also electrically connected to the balance housing (ground).
Note:
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in
a commercial environment. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense. Changes or
modifications not expressly approved by Sartorius AG could void the user‘s authority
to operate the equipment.
Warmup Time
Each time you move your balance to another location, you must condition it for at least
12 hours to the new location. To deliver exact results, the balance must warm up for at
least 12 hours after initial connection to AC power. Only after this time will the balance
have reached the required operating temperature.
18
Antitheft Locking Device
To fasten an antitheft locking device, use the lug located on the rear panel
of the balance.
Always level the balance again any time it has been moved.
§ Turn the leveling feet as shown in the diagram until the air bubble is centered exactly
within the circle of the level indicator
19
Shipping the Balance
$ The balance must be disconnected from power before preparing it for shipment.
Disconnect the AC adapter and all interface cables from the balance.
! Do not lift the balance by the draft shield or the front panel, as this can result
in damage.
§ Open the draft shield doors and carefully position the balance in the lower packaging.
§ Press the foam padding against the housing.
20
§ Attach the retainers to secure the front draft shield panel and the display and
control unit.
§ Position the two padding blocks that make up the inner packaging and press inward.
§ Place the following parts in the recessed spaces at the top of the inner packaging:
– Shield plate
– Weighing pan
– AC adapter (in cardboard packaging)
21
§ Place the strap around the package and tighten it.
§ Lift the packaged balance by the strap and place it in the shipping box, with the
bottommost padding already in the box.
§ Close the shipping box and seal it appropriately for the intended transport
or shipment.
22
Configuring the Balance
23
Navigating in the Setup Menu (Examples):
24
Example: Entering the time and date
25
Setting the Balance Functions (BAL.FUNC.)
26
Balance Functions (Overview)
ο factory setting
√ user-defined setting(s)
g g
tin tin
y set y set
tor tor
Fac Fac
Setup Calibration | Cal | isoTST External calibration | adjustment** with
Balance Adjustment key function factory-defined calibration weight
functions External calibration | adjustment**;
user-defined weight
Internal calibration | adjustment
Internal linearization
Cal key blocked
reproTEST
ο Selection mode for cal. | adjustment
27
g
tin
y set
tor
Fac
Setup – Stability delay Without delay
Balance ο Short delay
functions Average delay
Long delay
Auto zero ο On
Off
Factory settings: No
only weighing Yes
parameters
(= balance | scale
functions)
28
Setting the Device Parameters (Device)
29
§ New password: Enter the numbers Extra Functions
and/or letters for the new password
(8 characters max.) § Exit the Setup menu:
If “none” is displayed, this means press the oo soft key
no password has been assigned
To delete the user password: > Restart the application
Press . or c and confirm
§ Print the parameter settings:
§ To confirm: – If the device parameters
press the l soft key are displayed: press r
Interfaces
Serial communicati
SBI
Baudrate
1200 baud
Number of data b
7 data bits
Parity
Odd
Number of stop b
1 stop bit
Handshake mode
Hardware handshake
after 1 char
Serial printer (PR
YDP03
Baudrate
1200 baud
Parity
Odd
Handshake mode
Hardware handshake
after 1 char
Function: external
Print key
Function: control
Output
Display
Contrast
2
etc.
30
Device Parameters (Overview)
ο factory setting
√ user-defined setting(s)
g g g g
Setup – tin tin tin tin
y set y set y set y set
Device tor tor tor tor
parameters Fac Fac Fac Fac
Ionizer* Off
ο On Auto-off time: 10 s 1 .. 99 seconds
Parity Space 2)
o Odd
Even
None3)
For the display, keys and extra functions, see next pages
31
Setup – ing ing
sett sett
Device y y
tor tor
parameters Fac Fac
Parity Space
ο Odd
Even
YDP01IS Label
(label printer)
YDP02IS
YDP02IS Label
(label printer)
Parity Odd
Even
ο None 2)
YDP04IS Label
For the display, keys and extra functions, see next page
32
Setup – ing ing
sett sett
Device y y
tor tor
parameters Fac Fac
Background ο White
Black
Application Off
symbols ο On
Power-on ο Off/on/standby
mode Auto on
Factory settings: No
only Yes
device parameters
33
Setting the Application Parameters (Application)
Factory settings:
only application parameters
34
Application Parameters (Overview)
ο factory settings
√ user-defined setting(s)
g g
tin tin
y set y set
tor tor
Fac Fac
For counting through air buoyancy correction, see the following pages
35
g
tin
y set
tor
Fac
Setup – Application 1 Counting Accuracy – Average ο Display accuracy
Application (Basic settings) piece weight calculation + 1 decimal place
parameters (resolution) + 2 decimal places
Printout None
ο Average weight only
Average and calculated value
Recalculation*
* = How to run this application is described in detail in our Masterpro “LA….” Installation and Operating Instructions.
Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see “download”)
36
g
tin
y set
tor
Fac
Setup – Application 1 Calculation Decimal places in None
Application (Basic settings) calculated result 1 decimal place
parameters ο 2 decimal places
3 decimal places
4 decimal places
5 decimal places
6 decimal places
7 decimal places
Printout ο None
All data
Tare weighing No
ο Yes
Autosave ο Off
values On, first value at stability
On, last value at stability
On, value bet.
70 – 130% at stabil.
Save statistics ο No
Yes
* = Setting can only be changed when the program is initially run and when the Wg.seq. key option is set to “No”
37
g g
tin tin
y set y set
tor tor
Fac Fac
Setup – Application 1 Differential weighing Generate printout None
Application (Basic settings) ο Automatic after backweighing
parameters Auto. after initial weigh.
and backweighing
Automatic after tare-, initial-
and backweigh.
Wg.seq key No
ο Yes
Change steel/aluminum ο No
references Yes
Diameter determination
Application 2 ο Off
(Control functions) Checkweighing* Activation of port lines ο Within checkweighing range
Always on
Stability and
checkweighing range
At stability
Stability + checkweigh.
range on
* = How to run this application is described in detail in our Masterpro “LA….” Installation and Operating Instructions.
Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see “download”)
38
g g
tin tin
y set y set
tor tor
Fac Fac
Setup – Application 3 ο Off
Application (Data records)
parameters Totalizing* Automatic storage ο Off
On, first value at stability
On, last value at stability
On, value bet. 70% – 130%
at stability
Printout of individual No
components ο Yes
Printout of individual No
components ο Yes
* = How to run this application is described in detail in our Masterpro “LA….” Installation and Operating Instructions.
Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see “download”)
39
g g
tin tin
y set y set
tor tor
Fac Fac
Setup – Application 3 Statistics Automatic storage o Off
Application (Data records) On, first value at stability
parameters On, last value at stability
On, value 70% – 130%
at stability
Printout of individual No
components o Yes
40
g
tin
y set
tor
Fac
Setup – Extra Functions ο Off
Application F4 key
parameters Extra Functions 2nd tare memory Container tare weight ο No
F5 key Yes
Manual storage
in app. 3
memory M+ (totalizing,
formulation, statistics)
Print in o Off
GLP Header On
Print1) U (absolute)
U* (relative)
PA (process accuracy)
Auto-start On
application when ο Off
power goes on
Factory settings No
only for Yes
application-
parameters
1) an asterisk (*) indicates an activated menu item. You can select up to 3 items.
41
Selecting the Printout Function (Printout)
42
Printout Parameters (Overview)
ο factory setting
√ user-defined setting(s)
g g
tin tin
y set y set
tor tor
Fac Fac
SETUP Application- Stability parameter Without stability 1)
Printout defined output o With stability
ISO/GLP o Off
printout Only for calibration/adjustment
Always on
Factory settings: No
printout only Yes
43
Printout Configuration
– Total printout: For totalizing, – Print item “Form feed” for footer:
formulation or statistics applications, Advance to beginning of next label in
press MR the “YDP01IS-Label” and “YDP02IS-
Label” [printer] interface mode
– Backweighing printouts or records:
automatically generated after back-
weighing or manually by pressing the Extra Functions
r key when the result is displayed § Exit printout configuration:
at the end of backweighing press oo soft key
Printouts for Differential Weighing: – LIST: print the currently selected list
These printouts can be generated as SELECT: printout items that can still
standard or configured (user-defined) be selected
reports.
§ When the select bar is on
You can configure the following LIST or SELECT:
printouts: press the r key
– automatically, if configured
in the Setup menu [Application
parameters: Application 1: Differential
weighing: Generate printout: Auto]
44
Example:
Configure an Individual Printout for Counting Application to Include Dotted Line, Date/Time, Piece Count and Net Weight
45
Device Information
46
Operating the Balance
Basic Weighing Function Therefore, we recommend that before Weighing Magnetic or Magnetizable
you begin the actual weighing series you Samples
Purpose open and close the weighing chamber at
The basic weighing function is always the same rate as you will be doing during It is technically impossible to avoid the use
accessible and can be used alone weighing. After the weighing chamber of magnetizable materials in the manufac-
or in combination with an application has been closed, the weight readout will ture of balances. Ultimately, the operating
program (Toggle between Weight Units, usually stabilize after about 8 seconds. principle of high-resolution balances is
Counting, Weighing in Percent, etc.). The accuracy of the weight readouts will based on electromagnetic force compensa-
increase as you continue weighing with tion of the load placed on the weighing
Features greater consistency. pan.
– Taring the balance
– Assigning IDs to weights When magnetic or magnetizable samples
– Printing weights Weighing Electrostatically Charged or containers (e.g., a beaker with a stirrer)
– Printing ID codes for weights Samples and Containers are weighed, interactions among the
Major measuring errors can occur when above-mentioned components of the
Soft Key Functions electrostatically charged samples and balance may occur, distorting weight
Cal Start calibration/adjustment containers are weighed. This problem readouts.
isoCAL Press when necessary to particularly involves samples that have
start calibration and extremely poor conductivity (glass, plastic, Unlike deviations caused by electrostatic
adjustment filters) since they can discharge electro- charges, magnetic interference is usually
S ID Stored ID entered static – i.e., friction-induced – charges only constant over time. However, it is sensitive
over a relatively long period of time. The to and dependent on the position of the
result is an interaction of forces among the sample container on the weighing pan and
charges adhering to the sample and the is also characterized by poor repeatability.
General Instructions stationary components of the balance
for “Analytical Weighing” (weighing chamber base plate, draft shield To reduce the effect described above,
construction, balance housing). This is we recommend increasing the distance
Handling Samples and Containers noticeable when the weight readout drifts. between the sample and the weighing pan
The sample should be conditioned At high humidity, this effect is not by inserting a non-magnetizable material
to the temperature inside the balance. very pronounced or may not occur at all, between them (the reduction in force is
This is the only way to avoid errors due to the thin layer of water that proportional to the quadrate of the distance).
caused by air buoyancy and deviations condenses on the sample and, through In special cases, soft-magnetic plates
caused by convection currents at the conductive discharge, counteracts should be used to shield against interfering
surface of the sample. interfering static electricity. magnetic effects.
Since these effects increase proportion- In addition to taking purely mechanical In the presence of extremely strong
ally to the volume and surface of the counteractive measures (protecting the magnetic fields – for instance, when
sample, make sure that the size of the sample using a special antistatic weighing measuring the susceptibility of a sample
tare vessel selected is appropriate for pan – see the “Accessories”), you can in an electromagnet – you should use
the initial sample. neutralize the surface charges by the below-balance weighing port, which
“bombarding” them with ions of the comes standard on your balance.
Never use your bare hands to touch opposite polarity (see page 53 for
samples to be weighed. In addition to instructions on activating the ionizer).
the effect of the temperature, the This is a highly effective procedure
extremely hygroscopic behavior for eliminating static electricity.
of fingerprints left on the sample will
otherwise cause considerable inter- The balance’s environment, including the
ference during weight measurement. operator, can considerably interfere with
weighing results, due to static electricity.
Use forceps or a similar utensil that is The balances of the ME/SE series have
appropriate to place your sample care- been designed to counteract this phenom-
fully on the pan. Working with your enon: the glass surfaces of the draft shield
balance requires a steady hand and a have a special metallic coating.
smooth, uninterrupted technique.
The rear panel of the balance has a
If the weighing chamber has not been terminal for connecting an equipotential
opened for a relatively long period, grounding conductor. It is used for
it may have a temperature different additional grounding of a peripheral
from that of the balance’s surrounding device (for example, a vibrating spatula).
environment. When you open the This terminal is designed for single
weighing chamber, a change in temp- grounding wires up to .25” standard
erature will inevitably occur, due to the gauge or 6 mm2 stranded wires and for .18”
laws of physics, and may show up as standard gauge or 4mm2 stranded wires.
a change in the weight readout.
47
Operating the Balance
Below-Balance Weighing
A below-balance weighing hanger is located on the bottom of the balance.
§ Attach a wire to the sample, if necessary, and hang it on the notched hook
48
For Service: Additional Functions Calibration
Using Verified Balances as Legal G Press the Cal soft key
Measuring Instruments in the EU*: In addition to the functions:
This balance is not allowed to be used – alphanumeric input > See the section on “Calibration/Adjust-
for weighing goods intended for direct – taring (not during ment” for further instructions
sale to the public. The type-approval alphanumeric input)
certificate for verification applies only – printing Setup
to non-automatic weighing instru- G Press o
ments; for operation with or without you can also access the following func-
auxiliary measuring devices, you must tions from the weighing application: > See the chapter entitled “Configuring
comply with the regulations of your – calibration/adjustment (not during the Balance” for further instructions
country applicable to the place of alphanumeric input)
installation of your balance. – setup Turning Off the Balance
– turning off the balance G Press e
G Before using the balance as a legal
measuring instrument, calibrate and > The balance shuts off
adjust it at the place of use: see the sec-
tion on “Calibration, Adjustment” in > The display goes blank
this chapter
Examples
Example W1: Simple weighing
49
Example W2
Note:
– The sample ID generally applies to one weighing operation only
– The ID is deleted after data output
2. Select the required letter group “A” ABCDEF soft key Max 210 g d=0.01mg
0% 100% A
F aH
S
A
000000 gOOO
B C D E
L
sS
F
T
4. Select the letter group and ABCDEF soft key Max 210 g d=0.01mg
enter “B” B soft key 0% 100% A
F aH
S AB L
sS
ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-? :#*"&U
T
5. Select the letter group and ABCDEF soft key Max 210 g d=0.01mg
enter “C” C soft key 0% 100% A
(If you enter only letters, F aH
conclude input a) S ABC L
sS
ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-? :#*"&U
T
50
Device Parameters
Opening and Closing This function can be deactivated. – A lower weight resolution is possible
If not, the draft shield doors will close when the draft shield doors are open.
the Draft Shield automatically if the balance has not
been used for 2 minutes (protects – The left and right K keys for operating
Purpose chamber from dust). the draft shield doors can
The ME/SE balance is a high-resolution – have the same function
weighing instrument. It has a draft – The function “Close draft shield auto- – have separate functions
shield so that convection currents can- matically when function is activated” – be switched off.
not affect the weighing result. To load a can be combined with functions and
sample on and remove it from the applications that require the “with Draft Shield on
weighing pan, a draft shield door must stability” parameter in order for weights ME215|235|414|614|36S:
be opened and closed. You can do this to be accepted: – You can define which draft shield
in various ways, depending on the – Turn on the balance (tare when the door(s) will open and close when you
menu setting you select. power goes on) press the left or right K key (the Genius
– Tare after stability balance has a self-teaching capability)
Features – Print on request after stability
– The draft shield doors can be opened – Start all adjustment functions – If a door encounters an obstacle while
and closed at any time, regardless of the – 2nd tare memory moving, the following will happen:
application used – Manual weight storage mode for the – While opening: the door will stop
following functions: – While closing: the door will re-open
– The draft shield doors can be opened – Counting, weighing in percent,
and closed by pressing the respective recalculation, density determination, Factory Settings of the Parameters
keys, by activating an external switch or differential weighing Left/right draft shield keys:
by sending a command to the balance’s – Checkweighing, time-controlled Same function
interface port functions with storage mode
– Totalizing, formulation, statistics Automatic mode: Off
– The draft shield doors can be set so that
they operate automatically in one of the Weight resolution when door is open:
following ways when a specific balance Show all decimal
function is performed, such as taring places
(zeroing):
– Close automatically
– Close automatically, then open
51
Preparation
G Turn on the balance: press e
G Select “Device parameters”: press the q soft key, then the O soft key
ο = factory setting
G Store settings and exit the Setup menu: press the oo soft key
$ If open, close all draft shield doors G Apply moderate pressure to the right-
hand draft shield door (3) to slide it
G Apply moderate pressure to both door toward the back so that the door is
1 2 3 grips for top and right-hand draft opened by motor.
shield doors (2 and 3) to move them G Press the left-hand K key to save this
simultaneously towards the back door-opening mode; the door now
closes. If you press the left-hand K key
G Press the right-hand K key to save this afterwards, the right-hand draft shield
door-opening mode; the doors now door will open and close.
close. If you press the right-hand K
key afterwards, the top and right-hand G Apply moderate pressure to the left-
doors will open and close. hand draft shield door (1) to slide it
toward the back so that the door
is opened by motor.
52
ME215/235/414/614 Models: Password Interfaces
Static Electricity Eliminator You can enter a password to block access
(Ionizer) J to the menu parameter settings and to ID Purpose
code and exact calibration weight inputs. This item enables you to set the
Purpose See the detailed description in the chapter parameters for the following interfaces:
Ionization of air (i.e., charging the air on “Configuring the Balance: Setting the – Serial communications port
with ions of both positive and negative Device Parameters.” – Serial printer port
charges) causes the air surrounding a – External switch function
sample to become electrically conductive. – Control port function
Charges are neutralized in the air stream User ID
or grounded. You can enter your own personal password Serial Communications Port
(20 characters max.) You can set the serial communications
Be especially careful when weighing port to use for the following modes:
electrostatically charged samples. – SBI
Clock – XBPI
Features ISO/GLP printouts in particular must be – Sartonet
– The ionizer can be turned on and off generated with the date and time stamp
independently of the position of the of the specific measurement. This date and Serial Printer Port
draft shield doors time stamp is optional on other printouts. You can set the serial printer port to use
for the following printers:
– When the draft shield doors are open, See the chapter on “Configuring the – YDPO1IS
the static electricity eliminator (ionizer) Balance: Setting the Device Parameters.” – YDP02
will shut off after the preset time has – YDP03
elapsed. However, the ionizing time will – YDPO1IS Label
start over if the static electricity – YDPO2IS
eliminator is still running when a door – YDP02IS Label
closes. – Universal
– YDP04IS
– The static electricity eliminator (ionizer) – YDP04IS Label
can be configured in the Setup menu so
that it will stay on only for a certain External Universal Remote Switch
time. You can connect an external universal
remote switch (foot switch or bar code
– When the ionizer is active, this is scanner or an external keyboard) to one
indicated by the flashing symbol J of the two serial ports. Then you can
(flashes from the outside toward the assign one of the following functions to
inside and vice versa) be performed when
the switch is activated:
Factory Setting for the Parameters – Print key
53
Control Port Function Display Keys
You can connect either a checkweighing You can assign different functions to
display or an external universal switch to You can configure the display for your the c key for deleting input and
the serial communications port on the individual needs. applications.
balance (factory setting).
The contrast can be adjusted in 5 levels: When you delete applications, you can
To do so, you need to configure the Contrast delete either the data stored for all
interface for input or output. applications or just selected data.
Characters can be displayed in black on CF function
Pin Assignment Chart of the Female white or vice versa: Background in application
Interface Connector
Max 210 g d=0.01mg When you delete input, you can either
0% 100% A
PinFunction: Input F aL
delete all the data input in a field, or
15 r key; see
“Universal switch”
S 2148576 g
COUNTING: nRef = 10 pcs
Cal
S
sT
Start
only the last character entered.
CF function for inputs
16 Left K key
17 Soft key 6 (Cal) You can blank out either the bar graph or You can block key functions; you can
fgh Q the text line or both choose whether to block all keys (except
18 Soft key 1 (F) e, o, draft shield left/right and
19 w key Digit size ionizer) or just the alphanumeric keys.
Max 210 g d=0.01mg
Block key functions
Pin Function: Output 0% 100% A
F aL
15 “External switch”
(see above)
S 2148576 g
COUNTING: nRef = 10 pcs
Cal
S
sT
Start
Extra Functions
54
Calibration, Adjustment, Linearization
* including the Signatories of the You can have the balance automatically
Agreement on the European Economic display an adjustment prompt after a
Area certain time interval has elapsed since
the last calibration/adjustment or when
the ambient temperature changes by
a defined amount.
55
Preparation Releasing Access to
G Select the balance function for “calibration/adjustment”: press o External Calibration
G Remove the cover plate from the back
G To select the Balance/scale functions: press the O soft key of the balance housing
G Select Calibration/adjustment: press the O soft key G Move switch 1 in the direction
of the arrow
ο = factory setting
G Save settings and exit Setup menu: press the oo soft key
56
Preparation
Example: Set the parameters for calibration and adjustment; e.g., with manual calibration/adjustment, isoCAL off
S
Cal
000000 g
2. Select the Setup menu o
57
Step Press key(s) (or follow instructions) Display/Output
S
Cal
000000 g
58
Selecting the Calibration/Adjustment – Internal calibration/adjustment: – External calibration/adjustment with a
Parameter Internal cal./ calibration weight determined by the
The setting Selection mode adjustment user:
must be selected in the Setup menu Ext. cal./adj.;
(factory setting). After pressing the – Internal linearization: user-defined wt.
Cal soft key, you can choose Internal lin.
from among the following settings by G Start the desired routine:
pressing the Select soft key: – Repeatability test Press the Start soft key
reproTEST
In the selection mode: Perform external calibration followed by automatic adjustment with the factory-set weight
Configuration:
factory settings
4. Place the weight on the Place std. weight on balance Max4200 g d= 0.01g
balance (e.g., 200.00000 g) 0% 100% A
F aH
Minus sign –: Weight too low
Plus sign + Weight too high
no plus/minus sign: Weight o.k
D 20000000 g
CAL: Extern. adj. factory-def. wt. T
Start Wahl
L
sS
59
Internal Calibration/Adjustment Internal Linearization
First set either Internal In the Setup menu (Balance/scale func-
cal./adjustment or tions: Calibration/adjustment: CAL key
Selection mode (factory function), you need to have set
setting) in the Setup menu Internal linearization
(Balance/scale functions: or select this using the Selection
Calibration/adjustment: CAL key mode (factory setting).
function).
Inside the balance housing are built-in
Inside the balance housing are built-in motorized calibration/adjustment
motorized calibration/adjustment weights for linearization.
weights.
The internal linearization sequence
The internal calibration/adjustment is as follows:
sequence is as follows:
G Unload the balance
G Unload the balance
G Select the calibration function: press G Select the linearization function: press
the Cal soft key and then the the Cal soft key, the Select soft
Start soft key key, and then the Start soft key
> The internal calibration weight > The internal linearization weights are
is applied automatically
applied automatically.
> The balance is calibrated
> The balance is linearized
> If the setting Calibrate,
then auto adjust (factory
> The internal linearization weights are
setting) is selected in the Setup menu,
the balance is now automatically removed
adjusted
> The balance is adjusted automatically
> If the setting Calibrate, following internal linearization
then manual adjust
is selected in the Setup menu,
the internal calibration routine is now > ISO/GLP printout: see page 139
ended without adjusting the balance
(otherwise, see “Calibration and
Adjustment Sequence,” on next page)
60
Calibration and Adjustment Sequence External Calibration/Adjustment with
In the Setup menu, you can configure a User-Defined Calibration Weight
the balance so that: In the Set up menu (Balance/scale
– calibration is always followed functions: Calibration/adjustment:
automatically by adjustment CAL key function), you need to set
Ext. cal./adj.;
Calibrate, then auto user-defined wt. or select
adjust (factory setting) or this using the Selection mode
(factory setting).
– you have the choice of ending the
sequence or starting adjustment after You can define a weight for adjustment.
calibration Calibrate, then External adjustment must be performed
manual adjust with weights that are traceable to a
national standard and that have error
If no deviation is determined in limits that are no greater than 1/3 of
calibration, or the deviation is within the required tolerance of the display
the tolerance limits dictated by the accuracy.
degree of accuracy you require, it is not
necessary to adjust the balance. In this Sequence of external calibration
case, you can end the calibration/ (adjustment: see left column).
adjustment sequence after calibration. First select External adjustment:
There are 2 soft keys active at this User-def. wt.
point:
– Start to start adjustment The balance has a factory-set weight
value (see “Specifications”).
– End to end the sequence
To reset a user-defined calibration
weight to the original factory setting:
$ Enter the factory-defined value
manually (see “Specifications”)
61
Define the Calibration Weight
S
Cal
000000 g
62
isoCAL: Automatic Calibration and Adjustment In the Setup menu, you can configure
Automatic Calibration, Adjustment at Set Times * the balance so that after calibration and
and Linearization adjustment
In the Setup menu, you can now enter – the application program
First set either up to three different times of day for is restarted On and reset
On and reset the automatic calibration/adjustment (see the application
application, isoCAL menu tree on page 55). When one
and linearization on of these times is reached, the balance – isoCAL and linearization are activated
or On without resetting displays the flashing calibration prompt and the application program must be
the app. (factory setting) in (“isoCAL”). Calibration/adjustment is restarted
the “Setup: Balance/scale functions” not performed if the balance is isoCAL and
menu. – off (standby) or linearization on
– in the Setup mode at the time set for
The “isoCAL” display automatically calibration. – the application program remains at its
begins flashing if the ambient previous status
temperature changes in relation to the If the balance is being operated at On without resetting
temperature at the time of the last the time set for automatic calibration/ the app.
calibration/adjustment, or after a adjustment, the calibration/adjustment
defined time interval has elapsed. sequence will not be prompted or Also in Setup, you can configure the
The balance is telling you that it wants performed afterwards. balance so that it displays a calibration
to adjust itself. prompt, but does not perform the
Automatic calibration/adjustment calibration/adjustment functions
This automatic calibration and adjust- is prompted at set times when: automatically:
ment prompt is activated when: – The set time is reached Only adjustment prompt
– The change in temperature is greater
than1.5 Kelvin or the elapsed time – The balance Setup mode is not active
interval is greater than 4 hours
– No alphanumeric input is active
– The balance Setup mode is not active (e.g., equation for calculation)
– No number or letter input is active – The load has not been changed within
the last 2 minutes
– The load has not been changed within
the last 2 minutes – The balance has not been operated
within the last 2 minutes
– The balance has not been operated
within the last 2 minutes – The load on the balance does not
exceed 2% of the maximum capacity
– The load on the balance does not
exceed 2% of the maximum capacity.
63
Calibration/Adjustment Printout -------------------- GLP header
Data Block Printout 13.03.2000 09:17
You can have the results of a calibration/ SARTORIUS
adjustment procedure printed out. You Model ME215S
can configure whether the printout is Ser. no. 60419914
generated as soon as the procedure is Ver. no. 01-41-02
completed, or whether a number of ID
calibration/adjustment procedures -------------------- List of Calibration/
(up to 50) are printed as a data block 24.04.2000 12:03 Adjustment Procedures:
printout. Start: manual Example 1:
Diff. + 0.00001 g Internal calibration
External calibration
Data Block Printout of Calibration/ completed
Adjustment Data
With the following Setup menu 25.04.2000 12:10 Example 2:
configuration, you can store the data Start: isoCAL/temp isoCAL activated by difference
from up to 50 calibration/adjustment Diff. + 0.00001 g in temperature
procedures and have them printed Internal adjustment
on request: completed
– isoCAL printout Diff. + 0.00000 g
On request, from
record memory 25.04.2000 18:30 Example 3:
Start: Adj. time isoCAL at defined time
When the memory contains 50 data Diff. + 0.00001 g
records: Internal adjustment
– additional records are output completed
automatically Diff. + 0.00000 g
64
Purpose Preparation
Repeatability Test (reproTEST) The “reproTEST” function automatically G Turn on the balance: press e
determines the repeatability (also called
Definition “reproducibility”) of results (based on > The Sartorius logo is displayed
Repeatability is the ability of the balance 6 individual measurements). In this way, the
to display identical readouts when it is balance determines one of the most impor- G Select reproTEST in the Setup menu:
loaded several times with the same weight tant quantities in relation to the place of press o
under constant ambient conditions. The installation. The results are displayed with
standard deviation for a given number of the balance’s accuracy. G Select “Balance/scale functions:
measurements is used to quantify the Calibration/adjustment: CAL key
repeatability. function: use the O q soft keys
CAL: reproTEST
Start
OO6 g Wahl
Start
000001 g
CAL: reproTEST
End
5. End reproTEST End soft key
or restart reproTEST Start soft key
65
Application Programs
– Calculated values = o, G
66
Toggle between Two Weight Units, ( )
Purpose Preparation
With this application, you can switch Balances used as legal measuring instruments in the EU*: grams, carats and
the display of a weight value back and milligrams are the only weight units available
forth between two weight units
by pressing a soft key. Standard balances: All weight units listed below:
Display/ Line for
You can use this application in Unit Conversion factor Printout metrological data
combination with any program chosen Grams 1.00000000000 g g
from Application 2 (checkweighing, Kilograms # 0.00100000000 kg kg
time-controlled functions) and one from Carats 5.00000000000 ct ct
Application 3 (totalizing, formulation, Pounds 0.00220462260 lb lb
statistics), as well as with the extra Ounces 0.03527396200 oz oz
functions. Troy ounces 0.03215074700 ozt ozt
Hong Kong taels 0.02671725000 tlh tlh
Singapore taels 0.02645544638 tls tls
Features Taiwanese taels 0.02666666000 tlt tlt
– Toggling the displayed weight Grains 15,43235835000 GN GN
Pennyweights 0.64301493100 dwt dwt
– Setting the display accuracy Milligrams 1000.00000000000 mg mg
Parts per pound 1.12876677120 /lb lb
– Other features as for the basic weighing Chinese taels 0.02645547175 tlc tlc
function Momme 0.26670000000 mom M
Austrian carats 5.00000000000 K K
Tola 0.08573333810 tol tol
Factory Settings of the Parameters Baht 0.06578947437 bat bat
Mesghal 0.21700000000 MS MS
ME215/235/254/414/614:
Weight unit 1: Grams /g G Select the Toggle Weight Units application in the Setup menu: press o
ME36S, ME5, ME2: G Select Application parameters: press the q soft key 2 x,
Weight unit 1: then the O soft key once
Milligrams/mg
G Application 1 (basic settings): press the O soft key
Display accuracy 1: All digits
G Select Toggle wt. units: (repeatedly) press the Q or q soft key
Weight unit 2: Milligrams /mg
G Confirm Toggle wt. units: press the O soft key
Display accuracy 2: All digits
Toggle wt. units Weight unit 1 see above
* including the Signatories of the
Agreement on the European Economic Display accuracy 1 ο All digits
Area Fewer for weight change
Last digit off*
ο = factory setting
* = not for verified balances used as legal measuring instruments
# = not on ME36S, ME5, SE2
G Save settings and exit the Setup menu: press the oo soft key
67
Additional Functions Toggling to the Next Application
In addition to the functions for:
G Press d
– alphanumeric input,
> See the section on the
– taring (not during corresponding application program for
alphanumeric input), further instructions
G Press e
Example
Toggle the Display from Grams [g] (1st Unit) to Carats [ct] (2nd Unit)
Setup: Application parameters: Application 1: Toggle wt. units: Weight unit 2: Carats/ct
68
Counting A
– Setting the accuracy when the Weigh. Toggle to the Weighing mode
reference sample weight “wRef” is
stored for calculating a piece count Start Store current weight value for
preselected piece count
– Automatic output of the quantity and
sample weight via the data interface
port after initialization or reference
sample updating while running the
Counting program (Printout:
Application-defined printout: Auto
print upon initialization: All values)
69
Preparation Additional Functions
To calculate a piece count, the average weight of one piece must be known. In addition to the functions for:
This average piece weight can be entered into the Counting program in one of three ways: – alphanumeric input,
– Enter the average piece weight using the numeric keys and store it;
– taring (not during
– The last reference sample quantity entered is loaded and displayed when you turn on alphanumeric input),
the balance. Place the same number of parts on the balance and initialize the Counting
program; – printing,
– When the automatic initialization parameter (see previous page), is on (Setup: > Printout: you can also access the following
Application-defined output: Autoprint upon initialization: All values; see page 39), the functions from this application:
balance goes into the “counting” mode when you turn it on and loads the last average
piece weight and corresponding reference sample quantity that were entered or calculated.
Calibration/Adjustment
Reference Sample Updating G Press the Cal soft key
You can have the average piece weight updated during counting (with the piece
count displayed) if “AWP update” is set to “manual” or “automatic” in the Setup menu. > See “Calibration/Adjustment”
Manual updating can only be performed when the Update soft key is displayed. for further instructions
Reference sample updating must be completed before using an application program from
Application 3.
Toggling to the Next Application
The Update soft key is displayed when: G Press d
– the balance has stabilized (stability symbol displayed)
> See the section on the
– the current piece count is not more or less than double the original piece count corresponding application program
for further instructions
– the current piece count is more than 100
– the internally calculated piece count (e.g., 17.24) differs from the nearest whole number Setup (Setting Parameters)
(in this case: 17) by less than ±0.3 G Press o
Reference updating can be repeated several times with an approximately > See “Configuring the Balance” for
doubled piece count. further instructions
> The Sartorius logo is displayed > The balance shuts off
G Select the Counting application in the Setup menu: press o > The display goes blank, then
standby or off is displayed with
G Select the Application parameters: press the q key 2x, backlighting
then the O soft key once
ο = factory setting
G Save settings and exit the Setup menu: press the oo soft key
70
Example
Determining an Unknown Piece Count; Weighing-In the Preset Reference Sample Quantity
Setup: Application parameters: Application 1: Counting: Average piece weight updating: Manual
2. Prepare a container for the Place the empty container Max 210 g d=0.01mg
parts to be counted on the balance 0% 100% A
F aL
S 5006012 g
COUNTING: nRef = 10 pcs
Cal
S
sT
Start
4. Place the reference sample Place the displayed number Max 210 g d=0.01mg
quantity on the balance of parts in the container 0% 100% A
F aL
(example: nRef = 10 pcs)
S 2148031 g
COUNTING: nRef = 10 pcs
Cal
S
sT
Start
5. Determine the average piece Start soft key Max 210 g d=0.01mg
weight (number of decimal 0% 100% A
F aL
places displayed depends on
the balance model) S 10 pcs
COUNTING: wRef = 2.14800 g
Cal Weigh.
S
sT
6. If necessary, increase the number Place additional parts Max 210 g d=0.01mg
of parts and perform reference in the container 0% 100% A
F aL
sample updating (example: Update soft key
7 additional parts) S 17 pcs
COUNTING: wRef = 2.14800 g
Cal
S
sT
Weigh.Update
nRef + 17 pcs
wRef +2.1480300 g
71
Weighing in Percent B
72
Preparation Additional Functions
To calculate a value in percent, the reference percentage must be known. In addition to the functions for:
This value can be entered into the weighing-in-percent program in one of three ways: – alphanumeric input,
– The last reference percentage entered is loaded and displayed when you turn on the bal-
ance. Place the corresponding weight on the weighing platform and initialize the weigh- – taring (not during alphanumeric input),
ing-in-percent program;
– printing,
– With automatic initialization switched on (see previous page), the balance goes into the
“weighing in percent” mode when you turn on the power and loads the last reference you can also access the following
percentage entered as well as the corresponding reference weight (Setup: Printout: Appli- functions from this application:
cation-defined output: Auto print upon initialization: All values);
– Enter the reference weight using the numeric keys and store it (Wxx% soft key). Calibration/Adjustment
§ Press the Cal soft key
G Turn on the balance: press e
> See “Calibration/Adjustment”
> The Sartorius logo is displayed for further instructions
G Select the Weighing in Percent application (“Percent weigh”) in the Setup menu:
press o Toggling to the Next Application
§ Press d
G Select Application parameters: press the q soft key 2 x,
then the O soft key once > See the section on the corresponding
application program for further instruc-
G Select Application 1 (basic settings): press the O soft key tions
Percent weigh. Weight storage ο Display accuracy > See “Configuring the Balance”
accuracy +1 decimal place for further instructions
+2 decimal places
ο = factory setting
G Save settings and exit the Setup menu: press the oo soft key
Equations
The following equations are used for the corresponding calculations:
Ratio1 (percent-Ratio 1:) = (100% weight – current weight) / current weight x 100%
73
Examples
4. Place the reference weight Place weight equal Max 210 g d=0.01mg
on the balance (example: to reference weight in 0% 100% B
aL
(18.21480 g equals 100%) the container
S 1821480 g
PERCENT WEIG.: nRef = 100
Cal
B
S
sT
Start
pRef + 100 %
Wxx% + 18.21480 g
74
Calculation D
75
Preparation Additional Functions
G Turn on the balance: press e In addition to the functions for:
– alphanumeric input,
> The Sartorius logo is displayed
– taring (not during alphanumeric input),
G Select the Calculation application program in the Setup menu: press o
– printing,
G Select the Application parameters: press the q soft key 2 x,
then the O soft key once you can also access the following
functions from this application:
G Select Application 1 (basic settings): press the O soft key
Calibration/Adjustment
G Select Calculation: press the Q or q soft key, repeatedly, if necessary § Press the Cal soft key
76
Example
Calculate the weight per unit area (grammage or gsm weight) of paper: determine the gsm of a sheet of A4 paper
with the dimensions 0.210 m x 0.297 m = 0.06237 m 2. The gsm weight is a product of the division of the weight by the surface area.
6. Turn on the calculated result display Start soft key Max 210 g d=0.01mg
0% 100% D
F aL
S
EQUAT.=W/0.06237
Cal
000 og S
sT
Equat.Weigh.
77
Density Determination E
You can use this application in combination No. of decimals for display of vol. density:
with any program chosen from Application 2 decimals
2 (e.g., checkweighing, timer functions) Wa
and one from Application 3 (totalizing, Printout: None
formulation, statistics) as well as with the
extra functions.
Wfl Soft Key Functions
Wa Store weight of sample in air
– displacement.
Wfl With liquid density, buoyancy
and displacement methods:
– Store weight of sample
in liquid
78
Equations Used to Determine Density
The equation considers the following variables: the number of bars or wires,
the bar/wire diameter of the sample holder, and the inner diameter of the
vessel used.
The equation considers the following variables: the number of wires or bars,
the wire/bar diameter of the sample holder, and the inner diameter of the
vessel used.
79
Preparation Additional Functions
G Turn on the balance: press e In addition to the functions for:
– alphanumeric input,
> The Sartorius logo is displayed; the self-test is performed
– taring (not during alphanumeric input),
G Select the Density application in the Setup menu: press o
– printing,
G Select the Application parameters: press the q soft key 2 x,
then the O soft key once you can also access the following
functions from this application:
G Select Application 1 (basic settings): press the O soft key
Calibration/Adjustment
G Select Density: press the Q or q soft key, repeatedly, if necessary G Press the CAL soft key
see also the “Application Parameters (Overview)” in the chapter entitled > The display goes blank, then
“Configuring the Balance” OFF or Standby is displayed with
backlighting
G Save settings and exit the Setup menu: press the oo soft key
80
Example
Buoyancy: Determine the Density of Samples of a Solid Using the Buoyancy Method. Reference Liquid: Water
oo Q q l
8. Store weight value in liquid Wfl soft key Max 210 g d=0.01mg
0% 100% C
Density of sample is displayed E
H
F aL
S 838 g/
gcm
DENSITY: calculated density
Cal Param. Vol.
3 N
S
sT
Start
81
Step Press key(s) (or follow instructions) Display/Output
82
Differential Weighing
83
Differential Weighing: Defining the 3. Combined Weighing Factory Settings of the Parameters
Weighing Sequence The tare and initial weight, in that Weighing sequence:
You can choose from among four order, of each sample is measured first, Group weighing
sequences for measuring tare then the backweight of each sample is
weights, initial sample weights and measured. Tare weighing: Yes
backweighed residue (“backweight”)
during differential weighing: Measured weights/values Result with decimal point:
2 decimal places
Samples
1. Individual Weighing
Tare weight, initial weight and back- Tare Init.wt. Backwt.
Autosave values: No
weight are measured in that order. 1 1 1
Minimum load for autosave:
Measured weights/values 10 digits
Sample
84
Printout for Differential Weighing The following printout is generated:
85
Preparation 1
) Setting can only be changed when
G Turn on the balance: press e the application is first run and when
the Wg. seq key option is set to
> The Sartorius logo is displayed; a self-test is performed “No”
Tare weighing No
ο Yes
Autosave ο Off
values On; first value at stability
On; last value at stability 2)
On; value bet. 70 – 130% at stabil. 3)
Save statistics ο No
Yes
Include sample ID ο No
in text line Yes
86 ο = factory setting
Equations Direct Selection of Lot/Sample/Value
Backweight in %: backweight / initial weight · 100% When the measured values are
displayed, you can enter numbers and
Loss in weight: backweight – initial weight letters to:
– change the lot and sample directly (dis-
Loss in %: (backweight – initial weight) / initial weight · 100% played in the text line)
Calculated loss: (backweight – initial weight) · factor – directly access the display pages for
samples and values
Ratio 1 in %: (initial weight – backweight) / backweight · 100%
G Enter lot/sample/value ID
Ratio 2 in %: initial weight / backweight · 100%
87
List Function for Differential Weighing
The list function has 4 display pages: one each for lots, samples, values and results.
LOTS: 792 Smpl.avail. The display page for lots shows all of the lots that have already been created, as well as the
1 1 Sample T
122 1 Sample T,N number of samples in each lot and the processing status of the selected sample (tare, initial
AB05 20 Samples T,N,R1
CH0001 10 Samples T,N
and backweight). On this display page you can create, rename, delete and print lots.
CH01234
oo DeleteCreate
2
Q
Samples T,N,R1
q Sample
You can also define a factor for calculation of loss; for instance, to have weight per unit area
calculated (such as grams per square meter). You can also enter a lot ID alphanumerically
to access a lot directly.
SMPL: avail.792 Lot: CH0001 This display page shows the samples contained in a selected lot, as well as the processing
Sample 1: T,N,R(1) CX87
Sample 2: T,N,R(1) CX88 status of the samples (tare, initial and backweight) and the sample IDs. You can also enter
Sample 3: T,N
Sample 4: T,N a sample ID alphanumerically to access a sample directly.
Sample 5: T,N
oo Delete o Q q Values
VALUES: Lot: CH0001 Smpl:2 This display page shows the date and time of sampling, as well as the sample ID and the
Date,time: 16.11.1998 15:11:17
Name: ID CX88 values measured, for a selected sample.
Tare: T1 + 24.72654 g
Net initial wt: N1 + 14.45432 g
BackwghId res: R (1)+ 93.55678 g
oo Result o Q q
RESULT: Lot: CH0001 Smpl:2 This display page shows the calculated values for a selected sample. These include back-
Residue: R + 20.74321 g
dResidue: R + 80.48 % weighed residue, loss, loss calculated using a factor, and the ratio values. The d symbol
Loss: D - 5.03565 g
Loss: D - 19.52 % indicates the value that is selected for display immediately following a backweighing
Ratio1: DR + 24.25 %
oo Values o Q q l
procedure. To change this setting, use the q and Q soft keys to move the highlight bar to
the desired value, and press l to confirm.
STATISTICS: Lot:CH6789 This page shows the characteristic data for a lot (date; time; statistics on, for example, the
Statistics on: R (1) 5 Spls
Statistics on: R (2) 3 Spls backweighed residue; number of samples) as well as the calculated values (mean value,
Statistics on: R (*) 8 Spls
standard deviation).
oo Messw. o Q q l
To select a set of statistics from a lot with different numbers of backweighing procedures:
STATISTICS: Lot:CH6789
Date,time: 04.02.1999 14:31:30
Statistics on: R (1) OResidueo Press the l soft key to display the selected set of statistics:
No.of values: n 2
Mean value: Mean + 93.28 %
Std. deviation: s 0.01 %
oo Messw. o Q q l
88
Selecting Display Pages in the List Function for Differential Weighing
Lot
Samples
o
Delete
Values
o o
[numeric] Result
oo Values
Result oo
89
View and Print Display Pages You can manually print the display page Omit or Include Sample
You can use the manual mode to for statistics when it is shown on the G Activate the display page for samples
print display pages (for lots, samples, balance
values and results). G Select the desired (or omitted) sample
To view the display page for statistics:
To view and print a display page G Select statistics: press the Stat. G Delete: Press the Delete key
for values: soft key
G Show the display page for lots: press the G Omit: Press the Omit key
Lot soft key $ For samples each with a different
number of backweights:
G Show the display page for samples: Select the kind of statistics:
press the Sample soft key press the q or Q soft key
G Show the display page for values: press $ Confirm selection: press the l soft key Example: Sample 3 has been omitted
the Values soft key
G Print the display page for values: Deleting or Omitting a Lot or Sample Additional Functions
press r Lots can be deleted; samples can be In addition to the functions for:
deleted or omitted. – alphanumeric input,
PRINT: Lot: CH0001 Smpl:1
Current sample
All samples (5) You can choose between – taring (not during alphanumeric
– deleting the current lot and input), and
oo o Q q l – deleting all lots.
– printing,
$ Select amount of data to be included on You can choose whether
the printout: press the q or Q soft key – the active sample is deleted entirely, or you can also access the following
– only the values from the active sample functions from this application:
G Confirm print command: are deleted, or
press the l soft key – all samples are deleted completely, or
– only the values from all samples are Calibration/Adjustment
The display pages for lots and samples deleted, or G Press the Cal soft key
can be printed when they are shown on – a sample is omitted
the balance display. > See the section entitled “Calibration/
Deleting a Lot/Sample Adjustment” for further instructions
View the Display Page for Results: G Activate the display page for
G Show the display page for lots: press the lots/samples
Lot soft key Setup (Parameter Settings)
G Select the desired lot/sample G Press the o key
G Show the display page for samples:
press the Sample soft key G Select the “Delete” function: > See the chapter entitled “Configuring
Press the Delete key the Balance” for further instructions
G Show the display page for values:
press the Values soft key G Define the lot(s)/sample(s) to be deleted
and confirm Turning Off the Balance
G Show the display page for results: G Press the e key
press the Result soft key G Select “Yes” to complete the delete
function or “No” to cancel it > The balance shuts off
G Print the display page for results:
see instructions for printing the display > The display goes blank, then OFF or
page for values Standby is displayed with backlighting
90
Example
Differential weighing: Consecutive individual weighing; select lot, determine the difference in weight between initial weights and backweights
of two samples; generate and printout statistics; do not generate any individual and backweighing printouts
Setup: Application parameters: Application 1: Differential weighing: Weighing sequence: Consecutive individual weighing
Setup: Application parameters: Application 1: Differential weighing: Save statistics: Yes
Setup: Application parameters: Application 1: Differential weighing: Generate printout: No
3. Create or select lot Lot soft key LOTS: 985 Smpl avail.
CH0001 4 Samples T,N,R(1)
– select: q or Q soft key CH0002 8 Samples T,N,R(1)
continue with step 7.
– create: see step 4.
oo DeleteCreate Q q Sample
6. Confirm lot name input l soft key LOTS: 985 Smpl avail.
CH0001 4 Samples T,N,R(1)
CH0002 8 Samples T,N,R(1)
CH1234 0 Samples
oo DeleteCreate Q q Sample
91
Step Press key(s) (or follow instructions) Display/Output
8. Measure 1st tare weight Place 1st empty container Max 210 g d=0.01mg
on balance/scale 0% 100% A
F
S 7202512 g
CONS.IND.W: CH1234 #1 avail.
Cal Lot Wg.seq Wvgen
H
L
S
ZZT
Tare
10. Unload the balance Remove the empty container Max 210 g d=0.01mg
0% 100% A
F aH
S 000000 g
CONS.IND.W: CH1234 #1 T
Cal
L
dS
zzT
Lot Wg.seq Wvgen Ini.wt
11. Measure the initial weight Fill the 1st container Max 210 g d=0.01mg
0% 100% A
(in this case: 24.51 g) Place filled container
aH
on balance F
S 2451213 g
CONS.IND.W: CH1234 #1 T
Cal
L
s1 S
ZZT
Lot Wg.seq Wvgen Ini.wt
12. Save initial weight value Ini.wt soft key Max 210 g d=0.01mg
Remove the filled container 0% 100% A
F aH
S 000000 g
CONS.IND.W: CH1234 #1 T,N
Cal
L
dS
zzT
Lot Wg.seq Wvgen Backw.
13. Treat sample in 1st container
(for example, by drying)
15. Save backweight (the value to Backw. soft key Max 210 g d=0.01mg
be displayed is defined on the 0% 100% A
aH
display page for results; in this case:
backweighed residue in %)
F
S 7927 %
CONS.IND.W: CH1234 #1 OResidueo
Cal Stat.
L
s1 S
ZZT
Lot Wg.seq OTare Result
16. Unload the balance Remove the 1st container Max 210 g d=0.01mg
0% 100% A
F aH
S
Cal
000000 g
CONS.IND.W: CH1234 #2 avail.
Stat. Lot Wg.seq Wvgen
L
dS
zzT
Tare
17. Measure the second tare weight Place 2nd empty container Max 210 g d=0.01mg
0% 100% A
on balance
aH
F
S 7219117 g
CONS.IND.W: CH1234 #2 avail
Cal Stat. Lot Wg.seq Wvgen
L
dS
ZZT
Tare
92
Step Press key(s) (or follow instructions) Display/Output
18. Save tare weight Tare soft key Max 210 g d=0.01mg
0% 100% A
F aH
S
Cal
000000 g
CONS.IND.W: CH1234 #2 T
Stat. Lot Wg.seq Wvgen
L
s1 S
zzT
Tare
19. Unload the balance Remove the empty container Max 210 g d=0.01mg
0% 100% A
F aH
S 000000 g
CONS.IND.W: CH1234 #2 T
Cal Stat.
L
dS
zzT
Lot Wg.seq Wvgen Ini.wt
20. Measure the initial weight Fill the 2nd container Max 210 g d=0.01mg
(in this case: 25.77 g) Place filled container 0% 100% A
aH
on balance F
S 2577201 g
CONS.IND.W: CH1234 #2 T
Cal Stat.
L
s1 S
ZZT
Lot Wg.seq Wvgen Ini.wt
21. Save initial sample weight Ini.wt soft key Max 210 g d=0.01mg
0% 100% A
Remove the filled container
F aH
S 000000 g
CONS.IND.W: CH1234 #2 T,N
Cal Stat.
L
dS
zzT
Lot Wg.seq Wvgen Backw.
22. Treat sample in 2nd container
(for example, by drying)
26. Start second backweighing Place 2nd container on the Max 210 g d=0.01mg
of Sample 2 balance again; press 2, 0% 100% A
aH
then Spl# soft key F
S 1992174 g
CONS.IND.W: CH1234 #2 R(1)
Cal Stat.
L
s1 S
ZZT
Lot Wg.seqOResul Backw.
93
Step Press key(s) (or follow instructions) Display/Output
94
Air Buoyancy Correction G
95
Factory Settings of the Parameters Air Buoyancy Correction
Air density determination: Off
Diagram for Air Buoyancy Correction
Change steel/aluminum references: Off
Correction
of the mass ∆m g
ρw = 0.8 ——–
Soft Key Functions (milligrams) cm3
Diff. Display the difference
between the value measured
and the specification for g
steel/aluminum ρw = 1.0 ——–
cm3
to be weighed
for Air Buoyancy Correction”
on this page)
g
Corr. Start air buoyancy correction ρw = 2.0 ——–
cm3
with numeric input of the
density of a sample
cm3
density, specifications for
Weight displayed
steel, aluminum, etc.)
Subtract the mass error from
RhoA Activate air density determina- To determine the mass of a sample, its weight is multiplied by the following factor K:
tion or display RhoA for 2
seconds in the text line if K = (1 – RhoA / RhoST) / (1 – RhoA / RhoW)
the parameter “Air density
determination Off” is set where:
96
Preparation Additional Functions
G Turn on the balance: press e In addition to the functions for:
– alphanumeric input,
> The Sartorius logo is displayed; a self-test is performed
– taring (not during alphanumeric input),
G Configure the Air Buoyancy Correction application in the Setup menu: press o
G Select Application parameters: press the q soft key 2 x, – printing,
then the O soft key once
you can also access the following
G Select Application 1 (basic settings): functions from this application:
press the O soft key
Calibration/Adjustment
G Select Air buoyancy correction: press the Q or q soft key, repeatedly, § Press the Cal soft key
if necessary
> See “Calibration/Adjustment” for
G Confirm Air buoyancy correction: press the O soft key further instructions
97
Example
To determine the correct weight of a sample, enter the density of this sample. In this example, use the air density saved in the balance.
Setup: Application parameters: Application 1: Air buoyancy correction: Air buoyancy determination: On
5. Place the sample on the balance Load balance Max 210 g d=0.01mg
0% 100% G
(for example, 42.12648 g)
F aL
S 4212648 g
CORR.: RhoW = 1.0120000 g/cm3
Cal PT1/T1 RhoA Weigh. Fact.
S-ID
S
dT
6. Display the saved air density, RhoA soft key Max 210 g d=0.01mg
0% 100% G
if desired L
F
S 4212648 g
AIR D: RhoA = 1.2000000 kg/m3 (Def)
oo PT1/T1Param. Wvgen Start
S-ID
S
dT
98
Air Density Determination
Equations for Air Density Determination
The balance uses the equation below to calculate the air density for steel and aluminum
based on the reference weights supplied:
where:
RhoA = air density [kg/m3]
RhoAL = density of aluminum [kg/m3] RhoST = density of steel [kg/m3]
mA = mass of aluminum mST = mass of steel
WAL = weight value of aluminum WST = weight value of steel
where:
mAL = mass of aluminum (specification)
MAL = conventional mass value of aluminum
RhoAL = density of aluminum [kg/m3]
where:
mST = mass of steel (specification)
MST = conventional mass value of steel
RhoST = density of steel [kg/m3]
You can obtain the air density value in one of two ways:
G Turn on the balance and select the “Application parameters” as described for
Air Buoyancy Correction
G Use the numeric keys to enter the air density (1.0 – 1.4 kg/m3): 1 . 2 ... 0
G Save value for air density: press the RhoA soft key
G Exit the application for determining the air density: press the oo soft key
2. Weighing and Saving the Reference Weights for Steel and Aluminum
99
Example
Determination of the Air Density by Weighing, Using the Reference Weights Supplied for Steel and Aluminum
Setup: Application parameters: Application 1: Air buoyancy correction: Air buoyancy determination: Yes
Air buoyancy correction: Change steel/aluminum references: Yes
100
Step Press key(s) (or follow instructions) Display/Output
12. Place steel ref. weight on Load balance Max 210 g d=0.01mg
0% 100% G
the balance
F aL
S 20002142 g
AIR D: St.ref
oo
:Save ZZ
S
dT
PT1/T1Param. Diff. Store
13. Save steel reference weight Store soft key Max 210 g d=0.01mg
0% 100% G
F aL
S 20002142 g
AIR D: St.ref :Remove
oo
ZZ
S
dT
PT1/T1Param. Diff. Store
14. Remove steel reference weight Unload balance Max 210 g d=0.01mg
0% 100% G
F aL
D
oo
000000 g S
dT
AIR D: Al.ref 200.00348 g :Load zz
PT1/T1Param. Diff. S-ID
16. Save aluminum reference Store soft key Max 210 g d=0.01mg
weight 0% 100% G
F aL
S 20001082 g
AIR D: Al.ref
oo
:Remove
PT1/T1Param. Diff.
S
dT
ZZ
17. Remove aluminum reference Unload the balance Max 210 g d=0.01mg
0% 100% G
weight (the calculated air density
F aL
is displayed; in this example,
1.3195259) D 000000 g
AIR D: RhoA = 1.3195259 kg/m3
oo PT1/T1Param.
S
dT
Start
S-ID
101
Diameter Determination
102
Example
Determining the diameter of metal wires and filaments (such as a filament used in a light bulb)
103
Step Press key(s) (or follow instructions) Display/Output
104
Time-Controlled Functions K
105
Preparation Additional Functions
G Turn on the balance: press e In addition to the functions for:
– alphanumeric input,
> The Sartorius logo is displayed
– taring (not during alphanumeric input),
G Select the Time-Controlled Functions application in the Setup menu: press o
– printing,
G Select the Application parameters: press the q key 2 x,
then the O soft key you can also access the following
functions from this application:
G Select Application 2 (control functions):
press the q soft key, then the O soft key Calibration/Adjustment
§ Press the Cal soft key
G Select Time-controlled functions: press the Q or q soft key
> See “Calibration/Adjustment” for
G Confirm Time-controlled functions: press the O soft key further instructions
Print then tare Off > See “Configuring the Balance” for
ο On further instructions
ο = factory setting
Turning Off the Balance
see also the “Application Parameters (Overview)” in the chapter entitled § Press e
“Configuring the Balance”
> The balance shuts off
G Save settings and exit the Setup menu: press the oo soft key
> The display goes blank, then OFF or
Standby is displayed with backlighting
Print Net Values without Printout of Time
106
Example
Document the evaporated amount of a sample with defined surface, temperature and air pressure at preset intervals of 1 minute, 30 seconds.
107
Statistics N
108
Factory Settings of the Parameters Soft Key Functions Printout of Statistics
Automatic storage: Off M+ Add weight values or application The transaction or component counter is
values to the total in the totaliz- printed in front of each measured value
Minimum load for automatic storage: ing memory. The component (weight). When an intermediate or final
10 digits or transaction counter value evaluation is printed out, all results up to
Source of data for auto storage: increases by one each time you this point are included.
Application 1 press this key.
n 5
Evaluated values: Net M- Delete the last value added to Total +151.67321 g
memory. The transaction Avg. + 33.0 pcs
Evaluation mode, MR key function: counter value decreases by 1. You s + 3.2 pcs
Intermediate cannot delete previous values by srel + 9.70 %
evaluation, print repeatedly pressing this key. Total + 165 pcs
Min + 29 pcs
M+/M– function, then tare: Off MR Print or display an intermediate or Max + 37 pcs
final evaluation Diff + 8 pcs
Printout of individual components: Yes
nDef Store the input number of n: Transaction counter
Stability range: 2 digits components
Total: Sum of all values
Printout: Application-defined output: Print
on request then tare: Off Mean: Average
s: Standard deviation
Min: Minimum
Max: Maximum
109
Preparation Additional Functions
G Turn on the balance: press e In addition to the functions for:
– alphanumeric input,
> The Sartorius logo is displayed
– taring (not during alphanumeric input),
G Select the Statistics application in the Setup menu: press o
– printing,
G Select Application parameters: press the q soft key 2 x,
then the O soft key once you can also access the following
functions from this application:
G Select Application 3 (data records):
press the q soft key 2 x, then the O soft key once Calibration/Adjustment
G Press the Cal soft key
G Select Statistics: press the Q or the q soft key
> See “Calibration/Adjustment”
G Select Statistics: press the O soft key for further instructions
Evaluation mode, ο Intermediate evaluation, print > The display goes blank, then OFF or
MR function Final evaluation, print Standby is displayed with backlighting
Intermediate evaluation, display+print
Final evaluation, display+print
Printout of individual No
components ο Yes
ο = factory setting
G Save settings and exit the Setup menu: press the oo soft key
110
Example: Animal Weighing with Statistics (Averaging)
Let’s suppose that you need to determine each weight of 7 very small animals and statistically evaluate and print them.
5. Weigh 1st animal Place 1st animal in container Weight fluctuates because of animal activity
Max 210 g d=0.01mg
0% 100% C
F aL
S
ANIMAL WEIG.: xNet
Cal M+
g
xRes
s
New
N
T
6. Start automatic animal weighing Start soft key Max 210 g d=0.01mg
0% 100% C
F aL
S
ANIMAL WEIG.: mDef = 2
Cal M+
g s
N
----T
Start
111
Step Press key(s) (or follow instructions) Display/Output
The balance delays starting the When this criterion is met, the Max 210 g d=0.01mg
subweighing operation until subweighing series begins 0% 100% C
three successive subweights lie F aL
within the range defined for
an “active” animal
S
Cal
8888000g
ANIMAL WEIG.: m = 20
M+ 19
s
Start
N
T
18
...
1
112
Extra Functions
113
Example
Setup: Application parameters: Extra function (F4): 2nd tare memory: Automatic printout: Tare/preset tare
PT1 + 50.00 g
4. Determine net weight of bottles Place filled bottles Max 210g d=0.01mg
(in this case: net contents = 125 g) on the balance 0% 100% C
F L
S 12500000 g ST
100
TARA1: Ubernahme PT1
Cal PT1/T1Rekal. Comp.
a
s1
S-ID
114
Individual Identification Codes (ID)
115
Preparation
G Turn on the balance: press e
> The Sartorius logo is displayed
G Select Extra function (F4) or Extra function (F5) in the Setup menu: press o
G Enter ID name: Select “Printout”: press the q soft key, then the O soft key
G Select “Identification #”: press the q soft key 5 x, then the O soft key once
G Select ID1
G Enter name for ID1 and confirm: use the numeric keys for numbers and/or the
soft keys to enter letters
G Save settings and exit the Setup menu: press the oo soft key
Example
See next page
116
Example
Include company address and sample lot number on the printout. Each ID line begins with the name.
Print this ID for each net value.
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Extra function (F4): Identification codes
Setup: Input: ID1: Company
Setup: Input: ID2: Location
Setup: Input: ID3: Street
Setup: Input: ID4: Lot
1. If necessary,
turn on the balance e
117
Step Press key(s) (or follow instructions) Display/Output
12. Place the first sample on the Place load on balance Max 210 g d=0.01mg
balance (ex.: weight of 0% 100% C
F aL
110.53214 g)
S 11053214 g100
TARA1: Ubernahme PT1
Cal ID Rekal. Comp.
S
dT
S-ID
118
Saving Values Manually in M+
Purpose Preparation
This extra function enables you to load G Turn on the balance: press e
weight values and calculated results
directly from Application 1 (such as > The Sartorius logo is displayed
counting, weighing in percent) or
Application 2 (checkweighing, time- G Select Extra function (F4) or Extra function (F5) in the Setup menu: press o
controlled functions) into Application 3
(totalizing, formulation, statistics). G Select Application parameters: press the q soft key 2 x,
then the O soft key
G Save settings and exit the Setup menu: press the oo soft key
Factory Settings of the Parameters
No user-definable parameters
119
Changing the Resolution
Purpose Preparation
To change the resolution of the § Turn on the balance: press e
weighing result. This enables quicker
weighing at a lower resolution. > The Sartorius logo is displayed
Features § Select Extra function (F4) or Extra function (F5) in the Setup menu: press o
– Weights can be displayed with a
reduced resolution. § Select Application parameters: press the q soft key twice,
then O soft key once
– Once this function is selected, the dis-
play will appear as usual, e.g., § To select Extra function (F4) or Extra function (F5):
» ... d=0.01 mg« will be shown. To tog- press the q soft key repeatedly, then O soft key once
gle to the 4-digit range:
press the d*10 key. § Select Change Resolution
In the metrological line of the display,
the readout will toggle accordingly to § Confirm the Change Resolution function.
» ... d=0.1 mg«. Afterwards, the soft key
should be labeled with d/10. See also: “Application Parameters (Overview)” in the chapter entitled “Configuring the
Balance”
– In general, the balance is tared each
time the number of decimal digits dis- § Store the settings and exit the setup menu: press the oo soft key.
played is changed. This ensures that the
unit is precisely tared in accordance
with the regulations governing legal
metrology each time the readout is
toggled.
120
Example
Quick determination of the weights of a few consecutive samples with reduced resolution.
121
Product Data Memory
Preparation
G Turn on the balance: press e
> The Sartorius logo is displayed
G Select Extra function (F4) or Extra function (F5) in the Setup menu: press o
G Save settings and exit the Setup menu: press the oo soft key
122
Example
Create a new product base data record for initializing the checkweighing program, including: target value, minimum, maximum
oo q l
oo Q l
4. Toggle to display of product data ProDat soft key PROD.DATA: PERCENT WGH
PERCENT WGH40 Wxx% 168.75432 g
(existing data records are displayed; CALCULATION8 pRef 100 %
in this example, 3 data records have COUNTING13
been stored)
oo Delete Load q Q Change
123
SQmin Function
Purpose Preparation
To display the allowable minimum sam- § Turn on the balance: press e
ple quantity “SQmin” in accordance
with the United States Pharmacopeia > The Sartorius logo is displayed
(USP). According to USP guidelines, the
uncertainty of measurement may not § Select Extra function (F4) or Extra function (F5) in the Setup menu: press o
exeed 0.1 % of the sample quantity
when substances are weighed with the § Select Application parameters: press the q soft key twice,
highest degree of accuracy for volume then press the O soft key once
determination. This additional function
ensures that weight results lie within § Select Extra function (F4) or Extra function (F5):
defined tolerance limits corresponding press the q soft key repeatedly, then press the O soft key
to the requirements of your quality
assurance system. § Select SQmin.
Factory-set parameters
Display: Text display
124
Example
Determining the weights of samples while monitoring the minimum sample quantity (here: SQmin: 30 mg)
125
DKD Uncertainty of Measurement
– Display of the uncertainty of measure- § In the Setup menu, select “Extra functions (F4)” or “Extra functions (F5)”:
ment, for example: press o
Absolute uncertainty of measurement:
U = 0.000292 g § Select the Application parameters: press the q soft key 2 times,
Relative uncertainty of measurement: then the O soft key
U* = 0.00029 %
Process accuracy: § Select Extra function (F4) or Extra function (F5):
PA = 0.00087 % press the q soft key repreatedly, then press the O soft key
126
Example
Perform a weighing procedure with the “DKD uncertainty of measurement” application
Settings:
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Display: PA (process accuracy)
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Print: PA (process accuracy)
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Display: Input: 3.00000 (factory setting)
127
Combining Applications
The following table summarizes the possibilities for combination of the application programs described here. Each line stands for one
combination. The weighing function is generally available, and does not have to be combined with a calculating function.
128
Practical Combination of Several Applications
Density determination of a solid sample using the displacement method with water and statistical evaluation
of 10 measurements
129
Step Press key(s) (or follow instructions) Display/Output
8. Store
Max 210 g weight in liquid; Wfl soft
d=0.01mg
0% 100% C
the density of the sample H
E
F g/ aL
is displayed (toggle if nec.,
density/volume/weight) S
DENSITY: calculated density
Cal M+
128 g
Param. Vol.
cm3 s STN
Start
130
Data Output Functions
– Text line
Line for Metrological Data (on balances verified for legal metrology)
This line shows:
Max 210 g – Maximum weighing capacity (such as 210 g)
Min 1 g – Minimum weighing capacity; the weight must not go below this limit
in Germany (such as 1 g)
e= 0.1mg – Verification scale interval; irrelevant if the balance is not used in legal metrology
(such as 0.1 mg)
d=0.01mg – Readability/actual scale interval: indicates the balance’s display increment in digits
(such as 0.01 mg)
You can turn off (blank) the bar graph display (Setup: Device parameters:
Display: Digit size: 13 mm + text display or 13 mm)
– “Busy” symbol
J
– Plus or minus sign
FH
– Zero symbol (indicating the scale has been zeroed)
S
131
Line for Measured Values
This line shows:
1250é – The current weight value
Application Symbols
This column shows:
(ABCD – Symbol for Application 1 (toggling between weight units, counting, weighing in percent,
animal weighing, calculation, etc.)
Text Line
This line contains:
COUNTING: nRef = 10 pcs – Explanatory text about the application program (for example, about “Counting”)
Balance Information
In the Setup menu, you can select Setup: Info: Device information
for a display of balance information. The display includes:
SETUP INFO
Version no:
Device
01-41-05
– Software version number
Wgh.sys. ver. #: 00-21-09 – Balance version number
Draft sh. ver.#: 05-01-03
Model: ME215S – Draft shield version number
Serial no: 91205355
oo o q – Balance model
– Balance serial number
– Date: next maintenance
– Service phone
– Minimum sample quantity SQmin
132
Interfaces
Purpose – In addition, the following equipment – Printouts generated from the applica-
The ME/SE series balances have two can be connected to the printer port: tion programs or by the configurable
interfaces that allow weights and other – Remote display print function can be output to the
measured values, calculated values and – Hand switch serial printer port or to the serial
parameter settings to be output to – Foot switch communications port or to both.
a printer, PC or checkweighing display, – External checkweighing display
etc. Control commands (for foot switch – Bar code scanner* – If you have selected the automatic print
functions) and alphanumeric inputs – Keyboard* mode, data will be output to the serial
(such as those from an online bar code communications port; printouts
scanner) can also be input in the * using the YCC01-0024M01 adapter generated by the application programs
balance via the two interfaces. (see “Accessories”) will then only be output to the serial
printer port.
– The serial communications port has a
Features 25-contact D-Sub female connector as – In the xBPI mode, the serial
– ME/SE series balances have two serial a standard feature. This connector can communications port can operate
interface ports: be exchanged for either of the two independently of the serial printer port
– Serial printer port (PRINTER – female connectors below: (this means you can transfer data from
Serial Out) – 12-contact round connector (RS-485 the balance to a PC and use this PC to
– Serial communications port for xBPI; RS-232 for SBI, xBPI) control your balance while generating
(PERIPHERALS – Serial I/O) – 9-contact D-Sub connector for direct printouts via the serial printer port).
connection of a PC
– The serial printer port has a – In the SBI mode, you can use ESC
permanently installed 25-contact D-Sub – Both the 12-contact and the commands from your PC to control the
female connector (RS-232) 9-contact female interface connectors balance via the serial communications
are additionally equipped with a 5-pin port.
– The following printers can be connected male connector to directly interface an
to this printer port: external bar code scanner or a keyboard. For printing an individual value on
– YDP02 request, either by pressing the print key
– YDP03 – The serial communications port can be r or by sending an ESC P print
– YDP01IS used in the following modes: command, the particular menu setting
– YDP01IS Label – SBI determines which data port will be
– YDP02IS – xBPI (BPI) selected for data output.
– YDP02IS Label
– Universal – The following equipment can be
– YDP04IS connected to this serial communications
– YDP04IS Label port:
– Printer not verifiable for legal
! You may need to use an metrology in the EU
external power supply to operate – PC
peripheral devices. – Remote display
– Hand switch
– Foot switch
– External checkweighing display
– T-connector
– Bar code scanner*
– Keyboard*
133
Factory Settings of the Parameters Preparation
Device parameters: Interfaces:
Serial communication: SBI Configuring the Interfaces
G Turn on the balance: press e
Serial printer: YDP03
> The Sartorius logo is displayed; a self-test is performed
Printout: Output to interface ports: serial
communicaton (PERIPHERALS): G Configure the interfaces : press o
Application-defined output
G Select Device parameters: press the q soft key, then the O
Printout: Output to interface ports:
Serial printer (PRINTER): G Select Interfaces: press the q soft key 5 x, then the O soft key once
Application-
defined output Interfaces Serial ο SBI
communication XBPI
(PERIPHERALS)
ο = factory setting
G Save settings and exit the Setup menu: press the oo soft key
134
Configuring a Printout
G Turn on the balance: press o
G Select Printout: press the q soft key 3 x, then the O soft key once
Serial Off
printer ο Application-defined output
(PRINTER)
ISO/GLP/GMP ο Off
printout Only for calibration/adjustment
Always on
Factory settings: No
only printout Yes
ο = factory setting
* = not applicable to verified balances
1
) = Information on use in legal metrology: Only permitted for control purposes; printouts are not allowed
2
) = auto print when load change is > 10d and stability is reached; no printout until residual difference in load value is < 5d
135
Printouts
136
Configuring Printout Formats Generating an ISO/GLP Printout
For a number of application programs, you In the Setup menu, you have a choice of
need to set initialization values. All values three settings:
upon initialization or only the main values – No ISO/GLP printout generated (Off)
can be automatically printed as soon as
you have configured this in the Setup – ISO/GLP printout generated only for
menu: Auto print upon ini- calibration/adjustment (Only for
tialization calibration/adjustment)
– YDP02
– YDP03
– YDP01IS
– YDP01IS Label
– YDP02IS
– YDP02IS Label
– YDP04IS
– YDP04IS Label
137
r (Print) Key Examples
Pressing this key causes the current value +153.00000 g Weight in grams
shown on the display to be printed out + 58.56234 ozt Weight in Troy ounces
(weight with unit, calculated value, + 253 pcs Piece count
alphanumeric readout) + 88.23 % Percentage
+ 105.78 o Calculated value
Setting:
Printout: Application-defined output or
Automatic output of displayed value
Line Format
The current value displayed can be printed
with a data ID code of up to 6 characters
at the beginning of the line. You can use ID ABC123DEF456GH Identification number*
this data ID code, to designate a weight L ID ABC123DEF456GH Lot number (weighing series)*
readout as a net weight (N) or a calculated W ID ABC123DEF456GH Weight set number*
value as a piece count (QNT) N +153.00000 g Net value
Qnt + 253 pcs Quantity
Setting: Prc + 88.23 % Percentage
Setup: Printout: Line format: For other
apps./GLP (22 characters)
* = only for ISO/GLP-compliant
records/printouts
Sample ID
You can have each weighed or calculated
value that you print preceded by a line of
text containing numbers and/or letters.
You can either print this ID immediately as
alphanumeric input (press r) or store it S ID ABC123DEF456GH Sample ID
as the sample ID (S ID soft key) to be (with less than 14 characters)
included on the next printout, if the “For ABC123DEF456GHI789JK Sample ID
other apps./ GLP (22 characters)” setting (with up to 20 characters)
is configured. NUM 12345678 Numeric key output when r pressed
Setting:
1
Setup: Printout: Automatic output of displayed value
2
Setup: Printout: Automatic output of displayed value: Time-dependent auto print
3
Setup: Printout: Automatic output of displayed value: Stability parameter
138
ISO/GLP Printout -------------------- Dotted line
You can have the parameters pertaining 17.01.2000 16:12 Date/time
to weighing conditions printed before SARTORIUS Balance manufacturer
(GLP header) and after (GLP footer) the Model ME215S Balance model
values from the weighing series. Ser. no. 91205355 Balance serial number
These parameters include: Ver. no. 01-41-05 Software version
(display and control unit)
GLP header: ID 12345678901234 Balance ID no.
– Date -------------------- Dotted line
– Time at the beginning of a weighing L ID 12345678901234 Weighing series no. (lot)
series nRef 10 pcs Application initialization value
– Balance manufacturer wRef 1.35274 g Application initialization value
– Balance model Qnt + 235 pcs Counting result
– Model serial number Qnt + 4721 pcs Counting result
– Software version S ID 12345678901234 ID for counting result
– ID number Qnt + 567 pcs Counting result
-------------------- Dotted line
GLP footer: 17.01.2000 16:13 Date/time
– Date Name: Field for operator signature
– Time at the end of the weighing series Blank line
– Field for operator signature -------------------- Dotted line
G Press c
> Text line: CF selected: clear application
139
Serial Communications Port
Purpose Features
The ME/SE balance has a serial
communications port (labeled Type of interface: Serial port
“PERIPHERALS”) to which you can
connect a computer, a remote display Operating mode: Full duplex
or an external checkweighing display. Standard: RS-232 (RS-485 optional)
Interface connector: D-SUB female connector, 25-contact
You can use an on-line computer to Optional: round female connector, 12-contact
change, start and/or monitor the func- Optional: D-SUB female connector, 9-contact
tions of the balance and the application (Each of the optional connectors comes with a DIN 5-contact
programs. female connector)
The communications port also provides Transmission rates: 150; 300; 600; 1200; 2400; 4800; 9600 and 19,200 baud
data output port lines for the over/under Parity: Odd, even, none
checkweighing program. This port Character transmission: Start bit, 7/8-bit ASCII, parity, 1 or 2 stop bits
can also be used to connect a hand or Handshake: For 2-wire interface: software (XON/XOFF);
foot switch. for 4-wire interface: hardware (CTS/DTR)
! Warning When Using Pre-wired Communication mode: SBI, xBPI*
RS-232 Connecting Cables Network address**: 1, 2, ..., 31, 32
Manual print mode Without stability, after stability
RS-232 cables purchased from other
manufacturers often have incorrect pin Automatic print mode Without stability, at stability, when weight changes
assignments for use with Sartorius bal- Data output format
ances. Be sure to check the pin assign- of the balance: 16 characters, 22 characters
ment against the chart before connect-
ing the cable, and disconnect any lines * xBPI communication mode always with 9600 baud, 8-bit ASCII, uneven parity,
marked “Internally Connected” (e.g., pin 1 stop bit
6). Failure to do so may damage or even ** Network address is only for the xBPI communication mode
completely ruin your balance and/or
peripheral device.
Factory Settings of the Parameters:
Preparation
G For the pin assignment charts, see the description starting on page 145.
140
Output Format (Line Format) The following characters can be output, depending on the characters displayed on the balance:
You can output the values displayed in the
line for measured values and the weight Normal Operation
unit with or without a data ID code
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Example: Without data ID code + D D D D D D D D D * U U U CR LF
+ 253 pcs
or – . . . . . . . . . * * *
Example: With data ID code or * * * * * * * * * *
Qnt + 253 pcs
*: Space
Configure this parameter in the Setup D: Digit or letter
menu (Setup: Printout: Line format). U: Unit symbol
CR: Carriage return
The output with a data ID code has LF: Line feed
16 characters; without a data ID code,
22 characters. Special Codes
Output Format With 16 Characters Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Display segments that are not activated * * * * * * – – * * * * * * CR LF
are output as spaces. Characters without or * * * * * * A * * * * * * * CR LF
a decimal point are output without a
decimal point. or * * * * * * A B * * * * * * CR LF
and only upon request with ESC w0 (no print command):
* * * * * * S * X X X Y Y Y CR LF
or * * * * * * I * X X X * * * CR LF
*: Space
AB = – –: Final readout mode A = C: Cal/adjustment
A = H: Overload S: Draft shield status
AB = H H: Overload in checkweighing I: Ionizer
A = L: Underload Y,Y,Y = Draft shield doors
AB = L L: Underload in checkweighing
XXX = Decimal value calculated from individual status bits:
Control information:
0 ¡ Bit 6 – Motorized draft shield operation
+ 0 ¡ Bit 4 – Draft shield open
+ 8 ¡ Bit 3 – Learning function on
+ 0 ¡ Bit 1 – Draft shield motor off
+ 0 ¡ Bit 0 – No error
= 08
141
Example: output weight of + 111.25507 mg ID code characters I 1)
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Stat Status
+ 1 1 1 . 2 5 5 0 7 * m g * CR LF ID Identification (identifier)
L ID Weighing series no. (lot)
Position 1: Plus + or minus sign – or space W ID Weight set number
Position 2: Space or weight value digit
Positions 3–10: Weight with decimal point; leading zeros are output as spaces Nom Exact calibration weight
Position 11: Space (value)
Positions 12–14: Characters for unit of measure or space S ID Sample ID
Position 15: Carriage return NUM Numeric input
Position 16: Line feed T1 Application tare memory 1
N Net weight (T1 = 0)
Data Output Format with 22 Characters N1 Net weight (T1 0)
When data with an ID code is output, the ID code consisting of 6 characters precedes the Qnt Piece count
data with the 16-character format. These 6 characters identify the subsequent value. Prc Percent
nRef Reference sample quantity
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
pRef Reference percentage
I I I I I I + D D D D D D D D D * U U U CR LF
wRef Reference piece weight
* * * * * – . . . . . . . . . * * *
Wxx% Reference percentage weight
* * * * * * * * * *
mDef Target value for animal
weighing
I: ID code character1) U: Unit symbol1) see “Toggle Mul Calculated result
*: Space between Weight Units” in animal weighing
D: Digit or letter CR: Carriage return
LF: Line feed x-Net Result in animal weighing
x-Res Calculated result in animal
1
) depends on balance type; e.g., not all units and characters are weighing
available on balances verified for use in legal metrology Res Result using equation
(calculation)
Setp Target value for
Special Codes checkweighing
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
Min Lower limit for checkweighing
S t a t * * * * * * * * – – * * * * * * CR LF
Max Upper limit for checkweighing
H H
Time Time that a value was stored
L L Compxx Component no. xx in formula-
C tion
Tot.cp Total weight in formulation
*: Space L: Underload n Transaction counter
– –: Final readout mode
Total Sum of all values
(unstable weight) L L: Underload in checkweighing
H: Overload C: Calibration/adjustment Avg. Average in statistics
H H: Overload in checkweighing Draft shield and ionizer status; similar s Standard deviation
to data output format with 16 characters srel Variation coefficient
Diff Difference between maximum
Error Codes and minimum
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * E r r * # # # * * * * CR LF
142
Data Input Format Esc: Escape
You can connect a computer to your balance to send commands via the balance interface !: Command character
port to control balance functions and applications. #: Number
&: Number or letter
The commands sent are control commands and may have different formats; e.g., control _: Underline (ASCII: 95)
commands can have up to 26 characters. Each character must be transmitted according to CR: Carriage return (optional)
the settings configured in the Setup menu for data transmission. LF: Line feed (optional)
max: depends on command character;
i.e., parameter: once the max. length
Format for Control Commands is reached, input received
is truncated, rather than discarded
Format 1: Esc ! CR LF as with keyboard input
Format 2: Esc ! # _ CR LF
Format 3: Esc ! # & (max. 20 &) & _ CR LF
Format 4: Esc ! # & (max. 20 &) & _ CR LF
Format 5: Esc ! # # # _ CR LF Exc ! # _CR LF
Format 1 (e.g., ESC K) Format 2 (e.g., ESC f3_) Control Commands for the Draft Shield
on the ME5 and SE2:
! Meaning !# Meaning w0 Draft shield status
I Weighing mode 1 f3 Zero w1 Open draft shield 100° towards the
L Weighing mode 2 f4 Tare (without zeroing) left (stored position is deleted)
M Weighing mode 3 f5 Left draft shield key (closes and w2 Close draft shield
N Weighing mode 4 opens as “learned” or standard)
w3 Open draft shield up to position
O Block keys f6 Right draft shield key (closes and saved
opens as “learned” or standard)
P Print w4 Open draft shield 100° towards the
f9 J function key right (stored position is deleted)
Q Beep (acoustic signal)
kF1 Soft key 1 * Function depends on
R Unblock keys
setting in application
S Restart kF6 Soft key 6* program Format 3
T Tare and zero kF7 o function key (not allowed in the Setup menu
Z Internal calibration/adjustment kF8 d function key For example: ESC z5 1234567_)
m0 Ionizer status
!# Meaning
m1 Ionizer on
z5 Input (GLP) balance ID no.
m2 Ionizer off
z6 Input weight set “inventory” no.
s3 c function key
z7 Print weighing series no.
x0 Perform internal calibration
x1 Print balance model
x2 Print serial no. of weighing platform Format 4
x3 Software version of weighing platform ! Meaning
x4 Software version of display & control unit t Text input in display
x5 Print (GLP) balance ID no.
x6 Print weight set “inventory” no.
Format 5
x7 Print weighing series no.
(only for ME5, SE2: for example:
ESC t120 _ f5_)
Control Commands for the Draft Shield on
Models ME215/235/254/414/614/36S: ESC txxx _ CR LF ESC f5 _ CR LF:
Save left-opening position xxx in degrees
w0 Draft shield status
ESC txxx _ CR LF ESC f6 _ CR LF:
w1 Left draft shield door open Save right-opening position xxx in degrees
w2 Close all draft shield doors
w3 Open top draft shield door * counted from right to left
w4 Open right-hand draft shield door
w5 Open left & top draft shield doors
w6 Open left and right draft shield doors
w7 Open right-hand and top draft shield doors
w8 Open all draft shield doors
143
“External Keyboard” Functions Handshake Transmitting Device:
(PC-Keyboard) The balance interface (Sartorius Balance Once XOFF has been received, it prevents
Setting Interface = SBI) has transmit and receive further transmission of characters.
Setup: Interfaces: Function “external buffers. You can define the handshake When XON is received, it re-enables the
switch”: extra keyboard parameter in the Setup menu: transmitting device to send data.
The key codes implemented here are spe- – Hardware handshake (CTS/DTR) Receiving Device:
cific to the German keyboard layout. The – Software handshake (XON, XOFF) XOFF is transmitted after the 26th
following alphanumeric characters are character has been stored. To prevent too
implemented (some require “Shift” key): many control commands from being
a–z, A–Z, 0–9, <space>, Hardware Handshake received at one time, XON is not trans-
„,.\+'<>/"$@%/();=:_?* With a 4-wire interface, 1 more mitted until the buffer is almost empty.
character can be transmitted after CTS
Function keys: (Clear to Send). Activating Data Output
You can define the data output parameter
PC keyboard ME/SE Software Handshake so that output is activated either when a
F1 w key The software handshake is controlled print command is received or automatical-
F2 o key via XON and XOFF. When a device is ly and synchronously with the balance dis-
F3 Softkey F6 (far left) switched on, XON must be transmitted play or at defined intervals (see application
Cal key to enable any connected device to program descriptions and auto print set-
F4 F5 soft key (2nd from left) communicate. ting).
F5 F4 soft key (3rd from left)
F6 F3 soft key When the software handshake is Data Output by Print Command
(3rd from right) configured in the Setup menu, the The print command can be transmitted
F7 F2 soft key hardware handshake becomes active by pressing r or by a software
(3rd from right) after the software handshake. command (Esc P).
F8 F1 soft key (far right)
F9 d key Automatic Data Output
F10 – The data transmission sequence is In the “auto print” operating mode, data is
F11 r key as follows: output to the interface port without an
F12 – extra print command. You can choose to
Print r key Balance ––– byte –––> Computer have data output automatically at defined
Return L (Enter) F1 soft key (far right) (trans- ––– byte –––> (receiving print intervals with or without the stability
Cursor Up F3 soft key mitting ––– byte –––> device) parameter. Whichever parameter you select,
(3rd from right) device) ––– byte –––> the data will be output as the
Cursor Left F4 soft key (3rd from left) <––– XOFF ––– readouts appear on the balance display.
Cursor Down F2 soft key ––– byte –––> The display update frequency depends on
(2nd from right) ––– byte –––> both the model of the balance and the
Cursor Right F1 soft key (far right) … current operating status.
POS 1 (Home) F6 soft key (far left) (Pause)
Cal key … If you select the auto print setting, data
Backspace c key <––– XON ––– will be transmitted immediately the
ESC c key ––– byte –––> moment you turn on the balance. In the
––– byte –––> Setup menu, you can configure whether
Synchronization ––– byte –––> this automatic output can be stopped and
During data communication between the ––– byte –––> started by pressing r.
balance and an on-line device (computer),
messages consisting of ASCII characters are
transmitted via the interface. For error-free
data communication, the parameters for
baud rate, parity, handshake mode and
character format must be the same for
both units.
144
Pin Assignment Charts
5 2 4
Pin Assignment for the 5-Contact DIN Female Connector:
Pin1: Keyboard Clock
Pin2: Keyboard Data
Pin3: Not Connected
Pin4: Signal GND
Pin5: +5 V
! The YRB02FC bar code scanner requires an external power source if you have
connected a printer and a second display to the balance.
The PC keyboard also requires an external source of power.
Cabling Diagram
Diagram for interfacing a computer or different peripheral device to the balance using the
RS-232/V24 standard and cables up to 15 m (~50 ft.) long
TxD 2 2 RxD
RxD 3 3 TxD
CTS 5 4 DTR
DTR 20 8 CTS
GND 4 6 DSR
GND 14 5 GND
TxD 2 3 RxD
RxD 3 2 TxD
CTS 5 20 DTR
DTR 20 5 CTS
GND 4 6 DSR
GND 14 7 GND
146
Error Codes and Messages
Error codes are displayed in the main display or text line for 2 seconds.
The program then returns automatically to the previous status.
147
Error Code/Message Displayed Cause Solution
Ref.wt. too light Error in storing reference weight Weight too light or there is
(with the counting or weighing- no sample on the balance
in-percent application)
Cannot update Reference updating not possible See “Counting” in “Operating
(with the counting application) the Balance” for reference updating criteria
Not a number Input wrong (with any Follow the instructions for the
xxxxx Too low application program), e.g., application programs
xxxxx Too high alphabetic input not allowed
Too many char. Input text too long Allowable text lengths, incl. decimal point:
– S ID, NUM, L ID, ID: max. 20 characters
– W ID: max. 14 characters
Wrong line format Configured printout, printout memory Select the 22-character format
and 16-character format selected
Limits unequal Unit entered for tolerance limits Adjust tolerance limits
for unit in checkweighing different
from the application used
Equation too long Equation exceeds 28 characters Limit equation to 28 characters
informulation
Cancel, enter No reference parameters entered Enter missing reference parameters
ref. parameters for air density determination
Function active Function is being performed –
Fewer than 999 samples can be Product memory is full Delete some of the data
saved in up to 100 lots in the product memory
Err 10 x Key is stuck Release key
or
Key pressed when switching on the balance: contact your local Sartorius Service Center
x= 1: n (F1, F2, F5, F6), c
x= 2: d, n (F3), 0, 3, 4, 9
x= 3: 2, 5, 6, ., r, w-right
x= 4: 1, 7, 8, n (F4), a,
w-left
”Checkerboard” pattern displayed o key was pressed when turning on
continuously the balance, or is stuck
Err 320 Operating program memory defective Contact your local Sartorius Service Center
Err 340 Operating parameter (EEPROM) is wrong Turn the balance off, then back on again.
RAM lost data If this error remains displayed,
Factory settings deleted please contact your local Sartorius Service Center
Err 341 Battery needs to be recharged Leave the balance power on for at least 10 hours
No WP Weighing cell is defective Contact your local Sartorius Service Center
blocked Function blocked None
The special code J None of the keys has been pressed Press a key
remains displayed since the balance was turned on
The weight readout changes constantly Unstable ambient conditions Set up the balance in another area
Too much vibration, or the balance Change Setup configurations to adapt
is exposed to a draft the balance to the ambient conditions
A foreign object is caught between Remove the foreign object
the pan and the balance housing
The weight readout is obviously wrong The balance has not been
calibrated/adjusted Calibrate/adjust the balance
The balance was not tared before weighing Tare before weighing
The balance is not level Level the balance
If any other errors occur, contact your local Sartorius Service Center!
148
Error Messages Displayed Cause Solution/Remarks
during Differential Weighing
149
Care and Maintenance
Repairs
Repair work must be performed by
trained service technicians. Any attempt
by untrained persons to perform repairs To remove the weighing pan and clean:
may lead to hazards for the user. § Reach under the shield plate and lift it
together with the weighing pan off the
! Warning! base to avoid damaging the weighing
Particles drawn in by the fan can build system.
up inside the balance housing. If you
are using this balance in the chemical ! Make sure that no dust or liquid enters
industry, be sure to handle or treat the balance housing
defective parts of the balance or those
to be cleaned according to your country’s
current rules and regulations.
Cleaning
! Make sure that no dust or liquid enters
the balance housing
150
Cleaning the Weighing Chamber on Safety Inspection
ME5 | SE2 If there is any indication that safe
§ Use a small hand-held car vacuum operation of the balance with the AC
cleaner with a mini-hose attached to adapter is no longer warranted:
carefully remove spilled powder beneath
the shield disk § Turn off the power and disconnect the
power cord from an electrical outlet
§ Use blotting paper to remove spilled (mains supply) immediately
liquid
> Lock the AC adapter and power cord
in a secure place to ensure that the
equipment cannot be used during this
time
151
Recycling
Sartorius
Service Center
Weender Landstrasse 94-108
37075 Goettingen, Germany
152
Specifications
Standard Models
Model ME235S ME215S ME235P ME215P
Readability mg 0.01 0.01 0.01/0.02/0.05 0.01/0.02/0.05
Weighing capacity g 60/230 60/210 60/110/230 60/110/210
Tare range (subtractive) g –230 –210 –230 –210
Repeatability <±mg 0.015/0.025 0.015/0.025 0.015/0.04/ 0.015/0.04/
0.04 0.04
Linearity <±mg 0.1 0.1 0.15 0.15
Off-center loading error mg 0.15 – 0.2 –
at one-half max. cap.
(positions acc. to OIML R76)
Sensitivity drift within
+10 to +30 °C (50 to 86 °F) <±/K 1·10–6 1·10–6 1·10–6 1·10–6
Response time (average) s <8 <8 <8 <8
External calibration weight g 200 (E2) 200 (E2) 200 (E2) 200 (E2)
(of at least accuracy class...)
Allowable ambient operating temperature +5 to +40 °C (41 to 104 °F)
Operating temperature range +10 to +30 °C (50 to 86 °F)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update rate (depends on filter level) 0.2 – 0.4
Pan size mm d 90
Balance dimensions (W x D x H) mm 252 x 533 x 292
Weighing chamber height mm 239
Net weight, approx. kg 11.1
Dust and water protection rating of the
balance housing according to EN 60529 IP32
Power connection Using wide-range AC adapter for voltage ratings of 100 V to 240 V
Nominal frequency 50 – 60 Hz
Power consumption max. 35 VA
Hours of operation with fully charged
YRB05Z external battery pack, approx. 10 h
DC nominal supply voltage 10.5 to 25 Vdc
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams, parts per
pound, Chinese taels, momme, Austrian carats, tola, baht, mesghal
Selectable application programs Toggle weight units, counting, weighing in percent, animal weighing,
recalculation, calculation, density determination, differential weighing,
air buoyancy correction, air density determination, checkweighing,
time-controlled functions, totalizing, formulation, statistics,
2nd tare memory, identification codes, product data memory
153
Specifications
Standard Models
Model ME614S ME414S ME254S
Readability mg 0.1 0.1 0.1
Weighing capacity g 610 410 250
Tare range (subtractive) g –610 –410 –250
Repeatability <±mg 0.1 0.1 0.07
Linearity <±mg 0.4 0.3 0.15
Off-center loading error mg 0.6 0.4 0.3
at one-half max. cap.
(positions acc. to OIML R76)
Sensitivity drift within
+10 to +30 °C (50 to 86 °F) <±/K 1·10–6
Response time (average) s <3 < 2.5 < 2.5
External calibration weight g 500 (E2) 2+ 200 (E2) 200 (E2)
(of at least accuracy class...)
Allowable ambient operating temperature +5 ... +40 °C
Operating temperature range +10 ... +30 °C
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update rate (depends on filter level) 0.2 – 0.4
Pan size mm d 90
Balance dimensions (W + D + H) mm 252 + 533 + 292
Weighing chamber height mm 239
Net weight, approx. kg 11.1
Dust and water protection rating of the
balance housing according to EN 60529 IP32
Power connection Using wide-range AC adapter for voltage ratings of 100 V to 240 V
Nominal frequency 50 – 60 Hz
Power consumption max. 35 VA
DC nominal supply voltage 10.5 to 25 Vdc
Hours of operation with fully charged
YRB05Z external battery pack, approx. 10 h
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams, parts per
pound, Chinese taels, momme, Austrian carats, tola, baht, mesghal
Selectable application programs Toggle weight units, counting, weighing in percent, animal weighing,
recalculation, calculation, density determination, differential weighing,
air buoyancy correction, air density determination, checkweighing,
time-controlled functions, totalizing, formulation, statistics,
2nd tare memory, identification codes, product data memory
154
Specifications
Standard Models
Model ME36S ME5 SE2 ME5-F SE2-F
Readability μg 1 1 0.1 1 0.1
Weighing capacity g 31 5.1 2.1 5.1 2.1
Tare range (subtractive) g 31 5.1 2.1 5.1 2.1
Repeatability <±μg 2 1 0.25 11) 0.252)
Linearity <±μg 10 4 0.9 41) 0.92)
Sensitivity drift within
+10 to +30 °C (50 to 86 °F) <±/K 1·10–6 1·10–6 1·10–6 1·10–6 1·10–6
Response time (average) s 14 – 18 < 10 < 10 < 101) <10 2)
External calibration weight g 30 (E2) 5 (E2) 2 (E2) 5 (E2) 2 (E2)
(of at least accuracy class...)
Allowable ambient operating temperature +5 ... +40 °C
Operating temperature range +10 ... +30 °C
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update rate (depends on filter level) 0.2 – 0.4 0.1 – 0.4 0.2 – 0.4 0.1 – 0.4 0.2 – 0.4
Pan size mm d 30 d 30 d 20 d 50 d 50
Balance dimensions (W + D + H)
– Weigh cell mm 223+410+232 122 +316+122 122 +316+122 122 +316+122 122 +316+122
– Display and control unit mm 254+320+106 254 +320+106 254 +320+106 254 +320+106 254 +320+ 106
Net weight, approx.
– Weigh cell kg 9.8 3.3 3.3 4.3 4.3
– Display and control unit kg 3.5 3.5 3.5 3.5 3.5
Dust and water protection rating of the
balance housing according to EN 60529 IP32
Power connection (AC adapter) Using wide-range AC adapter for voltage ratings of 100 V to 240 V
Nominal frequency 50 – 60 Hz
Power consumption max. 70 VA (average 21 VA)
Power consumption, weigh cell and approx. 7 W (typical)
display and control unit only
Hours of operation with fully charged
YRB05Z external battery pack, approx. 10 h
DC nominal supply voltage 10.5 to 25 Vdc
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams, parts per pound,
Chinese taels, momme, Austrian carats, tola, baht, mesghal
Selectable application programs Toggle weight units, counting, weighing in percent, animal weighing,
recalculation, calculation, density determination, differential weighing,
air buoyancy correction, air density determination, checkweighing,
time-controlled functions, totalizing, formulation, statistics,
2nd tare memory, identification codes, product data memory
1
) with standard weighing pan: d 30 mm
2
) with standard weighing pan: d 20 mm
155
Specifications
* = according to Council Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
156
Specifications
* = according to Council Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
157
Specifications
* = according to Council Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
158
Accessories (Options)
159
Dimensions (Scale Drawings)
ME215/235/254/414/614:
160
Dimensions (Scale Drawings)
ME36S
182
233
206,3 35
115
419
An electronic
Zu der Wägezelle cell also belongs to
ME36S...
gehört auch jeweils
each eine weigh
ME36S… Elektronikzelle.
cell. /
An electronic cell also belongs to each of
the ME36S... weigh cells.
106
189
254
106 215
174.6
320
323
14
112
13°
60
161
Dimensions (Scale Drawings)
ME5, SE2:
Model
5 . .
SE2 / SE2-OCE . .
.
.
162
Declarations of Conformity
163
C
164
165
166
167
168
Plates and Markings
169
170
Index
Operating design 8
Operating the balance 47
Options 158
171
Entering the General Password
General Password:
40414243
Sartorius Weighing Technology GmbH
Weender Landstrasse 94–108
37075 Goettingen, Germany
Phone +49.551.308.0
Fax +49.551.308.32 89
www.sartorius.com
Copyright by Sartorius,
Goettingen, Germany.
All rights reserved. No part
of this publication may
be reprinted or translated in
any form or by any means
without the prior written
permission of Sartorius.
The status of the information,
specifications and illustrations
in this manual is indicated
by the date given below.
Sartorius reserves the right to
make changes to the technology,
features, specifications and design
of the equipment without notice.
Status:
April 2013,
Sartorius Weighing Technology GmbH,
Goettingen, Germany