METROMAT®
LABORATOR DE ETALONARI
Client:/customer:
Adresal Adress:
Comanda /Order:
Obiect etalonat:\Caltvetednstument
Denumire /Desinaton:
Serie Ise No
Model / vost
Producétor/ onactwer
Interval de masurare /veasxing rea
ExactitatelAcsvay
Rezolutia/Rosotion
Etalon/ stander instument
Denumire etalon/ Desnaton’
Serie / senat
C.Eetalon Nr/ emis de:/ la data
Caltrain cerita Nob: at date:
Denumire etalon/ Cesgnaton
Serie / sera
C.E.etalon Nr emis de: la data:
Celtation caiiete No by a dae:
CERTIFICAT DE ETALONARE
Calibration certificate
Nr. H = 09 - 18 - 003
METROLOGY SOLUTIONS CONSULTING
Str. Liiacului, BI 13, Sc. B, Ap.12, Codlea, Jud. Brasov
L 1996 / 18.06.2018
TTermohigrometru electronic
387671781711
Testo 606-2
TESTO
( -10...50)°C /(0...100)%6rHt
20,5°C; 42,5 Sort (6...95)%RH
0,1°C ;0,1%0H
Termometru electronic
2217
6677-09.17/Metromav09.2017
Generator etalon umiditate si temperatura
33296
181241/KESSID-K-18476-01-00105.2018
Pag.ti2
Acest cetficat de elalonare
cocumenieazs —_vrasabilatea
reulaelor prezentate la Sistomul
Intenatjonal de Unit (8.
Tresabitateamasurslor este
tealzeti si menfnul& pin
talon periodice, in
‘conformitate cu SR EN ISOICE!
17025 2006.
Etalonarea a fost electat’ in
conformitate cu un sistem de
management a calla ceitiat
SR EN 180 9001:2008.Este
recomandat ca utlzatrul 88
‘sigue reetaonarea la intervale
e timp rezonabie
This calation —_cotifcato
documents the traceabily of ho
presented results 10 the
Intemational System of Unis (S)
‘The messurement traceability is
realized and mentsined by
peridicaly calibrations, in
conformiy with SR EN (SOICE!
170252006.
The calibration was performed in
compliance with an SR EN ISO
9001:2008 carted qualty
system
Its recommended to the user fo
‘assure the recaliration at
resonable tne intervals
[Data etalonari: Data elibersriicertificatului:
Cavtraton cate: GD ease cet Ce
‘Stampila: Responsabil tehnic: Etalonarea a fost efectuaté de:
Teeical manager Person in charge
Ing. Nafrédi Jené Ing.Cretu Constantin
Notd Cerfcaul de etlonae fra sermatura gi stampllnu este vali Note: Caltatonceticates without siete and seal re ot vals
METROMAT SRL
Tel Fax:0368 409 008
Str-Ady Endre, nr 44
e-mail labor@metromat.ro
0.505600 SACELE Brasov
iF: RO 4980861LABORATOR DE ETALONARI|
METROMAT®
Cerificat de etalonare’Nr. H- 09 - 18-003 Pag 22
Conditii de etalonare / Caltraton conditions:
Wetoda de etalonare /Watrodolay - PE-15 bazata pe FOX 15-120
Etalonarea a fost efectuata in camera de clima (a temperatura) si camera generatorului de umicitate (Ia umiditate), prin
compara cu termometrul si higrometrul etalon. Valorile masurate sunt medi ale valoilor indicate,
Locul etalonari/ Locaton ofcataton Laborator Temperaturi Metromat
Condit de mediu / Antien conitions Temperatura ambianta: (22+ 3)°C.
Presiunea atmosferica: (944 +15) hPa
Umiditatea relativa: (45 +25) %erH
Rezultatele etalonarii | Cattyation sults
‘Inainte de ajustare | As found:
Vvabateadereerns | Vatorea mswls oes Inserting ce sure
wo) ‘mas("0) tras (0) rc
700 100) oo oa
250 250 00 O4
200 38 02 OA
Vacateadereens |] Vaearea asus coca oatidhea de maurre
cn ‘rae sens) ut
20 Ed 27 2
500 HO 10 20
700 702 22 22
Dupa ajustare | as et
Nu au fost efectuate reglaje de ajustare.
Incertudinea arbuité este incerttudnea extins® obit prin mulipicarea incotudin standard ou factorl de acoperie k=2,Iceritusinea a fost
expr in onforitale ou EA 4-02. Valoarea masurandull seal n interval de valor desernat cu 0 probabil de 95%,
The uncertainty the expanded uncaanty obtained by muting te standard uncertanty by the cores factor K=2. has een dtemined in scordance wih EA 4
02.The va af he messurand Ses minh assigned inter of vais wih a probatilyo $5%.
Incheierea certficatului de etalonare
End of ealoation cerifieate
Nota: METROWAT@ este marca registra la OSIM pentu servic metroogice. Reproducerea paral acetfcauui ese interzs
Note: NETROMAT @ i tade mak by OSIM for mera] sores. This caltaton coil may na bereqrodued parity.
METROMAT SRL StrAdy Endre, nr.44 RO-505600 SACELE Brasov
Tol Fax:0968 409 008 e-mail: labor@metromat.ro CIF: RO 4980864BIROUL ROMAN DE METROLOGIE LEGALA
INSTITUTUL NATIONAL DE METROLOGIE
NATIONAL INSTITUTE OF METROLOGY
INSTITUT_NATIONAL DE METROLOGIE
CERTIFICAT DE ETALONARE
Nr. 05.01 — 218/2018
Obiect: Umidimetru pentru lemn
Producator: TESTO SE &Co.KGaA Germania
Tip: testo 606-2
Serie / numar: 38767178/711
pentir. METROLOGY Qo ONS = CoNSUGITNG
Beneficiar: §.C. METROMAT §.R.L
P-ta Libertatii nr. 2, Sacele - Brasov
Comanda nr: 2521/21.06.2018
Numéar de pagini: 2
Data etalonarii: 25.06.2018
‘Stampila INM Data: 25.06.2018 7 Sef de laborator
CMUMMOn
Dr. ing. loan Cirmeanu
Fara aprobarea INM, acest ceriiical de etalonare nu poate fireprodus decat integral
Cerificatele de etalonare far semnaturi si stampild originale sunt nule
INM Sos. Vitan-Barzesti 11 + Sector4 + Bucuresti 042122 + Romania
Tel: (+4021) 334 50.60; 334.48 30 + Fax: (+4021) 334 63 45; 334 65 33 + E-mail: ofice@inm.roINM Certificat de etalonare Nr. 05.01 — 218/2018
Pagina 2/2
Metoda de etalonare: Comparare directa, conform PS-11-05.01-INM, cu materiale de
referin{& certificate pentru lemin cu CE nr, 05.01 - 012/2018.
Locul etalonarii: INM — Laboratorul Materiale de Referinta
Conditii de masurare: Au fost efectuate 7 masurari in conditii de repetabilitate,
de referinta: Temperatura ambianta: (23 + 2) °C
Umiditatea relativa a aerulul: (45 + 18) %
Nr] Specie | Vaioarea conv. | Valoarea medie | Inceritudinea |
crt. | lemn | adevarata a a indicatilor | de masurare,
| | “umiaitti, umidiayi, | U
| (4) a or)
7,00 867
1. | Brad 11100 10,57 0,32
1400 1343,
Incertitudinea de masurare: conform tabel
Nota 1: Incertitudinea atribuitd este incertitudinea extins4 obtinuta prin multiplicarea incertitudinii standard cu
factorul de extindere k = 2 sia fost estimata in conformitate cu GUM. Standardul roman echivalent cu
‘GUM este SR Ghid ISOICE! 98-3:2010: Incertitudine de masurare - Partea 3: Ghid pentru exprimarea
incertitudinil de masurare.
\Valoarea masurandului se afla in intervalul de valor indicat, cu un nivel de Incredere de 95 %.
Nota 2: Rezultatele masurarilor sunt trasabile la valorile materialelor de referinta utilizate la etalonare.
Romania este membra a Conventiei Metrulut iar INM este co-semnatar al MRA.
Etalonarea a fost efectuata de: dr. ing. lonescu George-Vietor
Incheierea certificatului de etalonare