Professional Documents
Culture Documents
Research Group for Technical Analytical Chemistry of the Hungarian Academy of Sciences,
Institute for General and Analytical Chemistry, Technical University of Budapest, GellCrt Tkr 4.,
H-l 11I Budapest, Hungary
Received February 10, 1992; accepted February 19, 1992
The performance characteristics (selectivity, slope of the electrode response, response
time) of polyvinyl chloride and modified polyvinyl chloride based solvent polymeric, po-
tassium selective membrane electrodes were evaluated in relevance to the analysis of ionic
constituents in biological samples. A flow-injection system incorporating a macro wall-jet
potentiometric detector was designed and its utility for the potassium assay in undiluted
urine samples was evaluated. Analytical data obtained by flow methods for the potassium
content of patients’ blood serum and urine samples had been correlated with flame atomic
emission spectrometric data. 0 1992 Academic press, IN.
INTRODUCTION
Flow analysis with on-line detection offers a modern approach to automation of
analytical procedures that find utility in clinical, pharmaceutical, and environmen-
tal analyses, as well as in process monitoring. Electrochemical detection in flow
analysis has become important in two aspects; namely the rapidly developing flow
analytical techniques need selective and sensitive detectors in order to reduce the
complexity of the analysis system, while electrochemical sensors in general ex-
hibit improved performance characteristics under hydrodynamically controlled
conditions (Z-3).
The special features of solvent polymeric membrane electrode-based ion-
selective potentiometry (e.g., the sensor size and geometry can easily be adapted
to the analytical task, the interfering effect of biochemicals on ionophore-based
plasticized polyvinyl chloride membranes can be minimized by membrane tech-
nology) made its use in clinical chemistry attractive and competitive by the ex-
tensively used flame photometry (4-6).
If ionophore-based solvent polymeric membrane electrodes are to be used for
the assay of ionic constituents in biological fluids it should be considered that the
analysis will be performed in extremely complex but more or less well-defined
matrices of naturally occurring and/or therapeutic samples. The conditions in
blood and urine samples are defined in certain respect by distinct physiological
concentration ranges. Thus, due to the special nature of the sample matrix the
requirements for the performance characteristics of the sensors can be predicted
(4).
232
0026-265x/92 $1.50
Copyright 0 1992 by Academic Press. Inc.
All rights of reproduction in any form reserved.
FLOW-INJECTION ASSAY OF F’OTASSIUM 233
The demand for the analysis of ionic constituents of electrolytes of biological
origin on a large scale prompted the automation of analysis in clinical chemistry.
Since the appearance of ionophore-based solvent polymeric membrane elec-
trodes, continuous flow methods with flow-through ion-selective detection have
become more and more popular and relatively less work has been devoted to the
use of the dynamic flow-injection technique for the assay of biological fluids.
The purpose of this contribution is to demonstrate the importance of ion-
selective membrane technological optimization for clinical analysis as well as to
explore the analytical value of a flow-injection potentiometry with a macro wall-
jet detector cell for the assay of potassium in biological samples.
N4
W-COO-c-ooc-H
BME-44
KTpClPE
-O-Or
0
DOS
FIG. 1. Membrane constituents.
(SSM) at the 0.1 M concentration level of the relevant chloride salts (II), by
considering theoretical slope values.
The mean activity coefficients of the electrolytes were calculated with the
Debye-Htickel equation (6) while the liquid junction potentials were estimated
with the help of the Henderson formation (6).
The double junction reference electrode used, Ag; AgCl, KC1 (1 iU) 1
CHJOOLi (1 M), was a Radelkis model (Budapest, Hungary). The EMF mea-
surements were carried out at room temperature with a six-channel electrode
switch (Orion Model 605, Orion Research Inc., Cambridge, MA) coupled to a
Radiometer pHM 64 (Radiometer Copenhagen, Denmark) digital pH-mV meter.
Reproducibility Test
The stability of the sensor is characterized in terms of the reproducibility of the
potential data measured between two alternating standards samples of different
concentration and/or composition (4). For the reproducibility test the electrodes
were immersed alternately in 10e2 M KC1 in BSE, then in low3 M KC1 in BSE,
or 10e3 M KC1 in BSE + 7 wt% BSA, at least 10 times, and the potential readings
were taken 120 s after each immersion. Between the measurements the electrodes
were washed with distilled water and wiped with tissue paper.
FLOW-INJECTION ASSAY OF POTASSIUM 235
Response Time Measurements
Dynamic response curves were recorded in a computer-controlled, switched
wall-jet measuring arrangement (12) with an activity perturbation. The switching
time was 7 ms. Eight preset scan rates were available, ranging from 0.661 to 147.6
ms time resolution, corresponding to total data acquisition times of 1 s to 4 min.
FTIR-ATR Measurements
The effect of albumin on ex situ conditioned ion-selective membranes of differ-
ent compositions was followed by FTIR spectroscopy using the ATR technique.
All membranes were treated in 1 M KC1 + 7 wt% BSA solution for 30 min and
rinsed for an instant with distilled water. The details and the capability of the
FTIR-ATR technique are described in (1.3).
Flow-Injection Measurements
An LKB pump (12000 Varioperpex, LKB-Produkter AB, Bomma, Sweden)
with a variable flow rate option was used to adjust the carrier stream. Samples
were introduced into the flow system by means of a six-port manual injector
(Rheodyne 7125, Cotati, CA) provided with loops of different volumes. A macro
wall-jet cell was located as a detector at the end of the analysis channel.
The injector and the macro wall-jet detector cell are connected to the flow-
system with Teflon ferrules. All connections and the analysis channel were made
of Teflon tubing (i.d., 0.5 mm; o.d., 1.8 mm).
Depending on the analytical task, the geometry of the analysis channel could
easily be altered by the incorporation of a dispersion coil, mixing chamber, or
straight-line tubes of different volumes (14). Coil reactors of different lengths were
constructed by wrapping the teflon tubing around a chart paper cardboard tube
(o.d., 17 mm).
Potential measurements were carried out with a Radelkis OP 208/l pH-mV
meter (Budapest, Hungary), and recorded with an Omniscribe DSOOO(Huston
Instrument Div., Gistel, Belgium) strip chart recorder.
Wall-Jet Cell
After the introduction of the voltammetric wall-jet detector (e.g. (15, 26)) into
liquid chromatography this cell arrangement became more widely used with other
types of detectors as well as in different fields of analytical chemistry (17-19). The
characterization of an ion-selective electrode wall-jet detector for flow-injection
analysis has been reported recently (20).
In our work a potentiometric wall-jet cell arrangement with small effective
volume and a macro ion-selective electrode was also designed and elaborated for
flow-injection analysis (Fig. 2). A Philips IS-560 electrode body (ISE, 1) and a
Radelkis OP-830 calomel electrode (REF, 3) with a bottom-ceramic frit were fitted
into the cell body made of plexi (poly(methylmethacrylate)). The reference elec-
trolyte (4) flowed continuously around the reference electrode frit, forming a thin
solution film around the ISE then entered a reservoir (6). The streaming carrier
solution was directed directly onto the ion-selective membrane surface (2) through
236 JENEY ET AL.
3 6
the variable diameter jet (5). The jet-nozzle-electrode membrane distance could
be varied by adjusting the nozzle. The wall-jet cell can be positioned both hori-
zontally and vertically. Since air bubbles occasionally exist in the carrier solution
and/or in the sample solution, positioning the wall-jet arrangement horizontally is
more advantageous.
millisecond range and only slightly affected by the nature of the ionophores or the
plasticizers with relevant lipophilicity values above log P 2 5. In this experimental
work the effect of mobile or fixed ion-exchanger sites incorporated in the ion-
selective membranes on the dynamic response characteristics was evaluated with
the aim of selecting the best membrane system for assay of biological fluids in the
flow-injection operation mode (Fig. 3).
The comparison of the response time curves relevant to different membrane
compositions, including the use of carboxylated PVC, demonstrates in accor-
dance with earlier findings (25, 26) that the increase in the site concentration of
ion-selective membranes reduces the response time (twS) of the sensor. More-
over, in contrast to the use of nonderivatized polyvinyl chloride, the conditioning
time in aqueous electrolyte solution increases significantly the response time of
ion-selective membranes incorporating carboxylated PVC, probably due to the
water uptake and the development of a hydrated surface gel-layer. This fact can
seriously limit the application of carboxylated PVC as a membrane support.
In respect to the application of plasticized polyvinyl chloride-based ion-
selective membrane electrodes in clinical chemistry, the problem associated with
the influence of proteins on the electrode response function has been disputed in
several publications (5, 23). The drift of the potentiometric cell incorporating
liquid membrane electrodes has been ascribed to protein adsorption, and it was
claimed that its effect can be overcome by membrane technology, e.g., by the use
of vinyl chloride-vinyl alcohol copolymers.
In the present work the nature of the protein-liquid membrane interaction was
studied in the function of PVC-types used for membrane preparation by the mo-
lecular specific FTIR spectroscopy using the ATR technique (1.9, in parallel with
electrochemical studies.
In contrast to expectation (23), the spectroscopic data relevant to potassium-
selective ionophore (BME-44)-based liquid membranes of different nature and
composition, being conditioned in 7 wt% BSA containing 1 M KCl, showed that
the protein deposition at the membrane surface was pronounced even at mem-
238 JENEY ET AL.
TABLE 2
Selectivity Properties of the BME-44 Ionophore-Based Potassium Sensors
lIj
u -- K -K -__ K
__ NH4
-3 1 __ Na
~ No -H
-H
- Ll
-H
Li
-4
- co
1
- co CO
-5 ’
Time (msec)
50.--
-30-
-500 I I I I I
“t om “N
4 8 w
lime (msec)
FIG. 3. Dynamic response characteristics of potassium sensors based on PVC and PVC-COOH in
two diierent time domains. (-) BME-44, PVC, DOS; (+ + +) BME-44, PVC, DOS, KTpClPB; (---)
BME-44, PVC-COOH, DOS. (A) Recorded with unconditioned membranes, (B) recorded with mem-
branes conditioned in 10m3M KC1 for 5 days. Concentration step: lo-* M KC1 -+ 10e3 M KC1 in 0.1
M NaCl background electrolyte.
240 JENEY ET AL.
0.25
DATE 21/01/91
TIME 19:41:47
ATR KRSS
TRANSMISSION APT=I.2
FLS=JKRSIOS
0.0 ‘1 1 FLR=JKRSlOI
2000 1500 I( 3oc
)
WAVENUMBER CM-I NSS=IOO
RES=4.0
VEL=IO
SMF=O
FIG. 4. FIIR-ATR spectrum of BME-44, PVC-COOH, DOS membrane treated in 7 wt% BSA
containing 1 M KC1 solution.
BME44,PVC,DOS,KTpCIPB BME44,PVC,DOS,KTpCIPB
E (mV)
*M)
wx2o
BME44,PVC-COOH,DOS BME44,PVC-COOH,DOS
E (mV) E hV)
260 210 -
*-0.23 s-o.17
2w- 100
a-O.26
110- , r-0.24
l 160. + +
100 2 ,w
0 2 lo u 0 1 10 I?
Nb. of m&u,.me”f Nb. Of mc:w,.,n:.
A B
FIG. 5. Reproducibility study corresponding to BME-44based ion-selective membranes of diierent
compositions. Measurements were carried out alternately in (A) lo-* M KC1 (0) and 10m3M KC1 (+)
containing BSE (B) 10 -* M KC1 (0) and 10e3 M KC1 + 7 wt% BSA (+) containing BSE.
Since almost all continuous analyzers require the predilution of urine samples,
the flow-injection analysis approach with controlled dispersion has advantages,
like short contact time with the interfering sample solution, self cleaning proper-
ties due to the alternating measuring-conditioning nature of the technique, in situ
dilution, etc., for the analysis of undiluted urine samples.
242 JENEY ET AL.
I--
50 - A
30’-* x
E +
S
10 -
6
J
5
+
cf -10 -
xt+++++
x ++++
2 I I x K x x I .:+++
xx.::: ;:++++++++++
xxxxxxxxxx
-30 -
Time (msec)
Time (msec)
FIG. 6. Dynamic response curves recorded with potassium membranes with fixed and mobile ion
exchanger sites in two different time domains. (+) BME-44, PVC-COOH, DOS; (x) BME-44, PVC,
KTpClPB, DOS. Concentration step: 10W2M KCI + lo-’ M KC1 + 7 wt% BSA in 0.1 M NaCl
background electrolyte.
- SME 44 (KC11
250 - + BME 44 (KCIIBSE)
+ BME 44 (KCMJLE)
200 -
150 -
100 -
50
I I I I I I I
-50’
-6 -7 -6 -5 -4 -3 -2 -1 0
log a
FIG. 7. EMF vs log uk response in different electrolyte solutions. Membrane composition: 2 wt%
BME-44, 34 wt% PVC, 63.3 wt% DOS, 0.7 wt% KTpClPB.
and the length of the analysis channel, respectively, the variation of the signal
corresponding to 10 injections is shown in Fig. 9. For selecting the most favorable
parameters, a fast, reagent saving method with desired dispersion was also con-
sidered.
5.
“L Y - A + Ex / I
Y 3.
5.5
K CFlamd mmol/l
FIG. 8. Correlation study for the potassium concentrations of 50 blood serum samples.
244 JENEY ET AL.
*s (mV)
- 20 ul - 60 ul -1ooul -2oou1
*) _.,...._...... I.
1 2 8 4 6 0
flow rate (mlhnid
04
0
0 1 2 a 4 6 0 7
flow rate (ml/mid
2.~1..1~
-0 1 2 3 4 8 0 7
flow ‘rate (ml/mid
FIG. 9. Optimization of the parameters of the FIA manifold.
FLOW-INJECTION ASSAY OF POTASSIUM 245
batch mode
wC(potroaium) mM
r-0.34
4S-
10 12 14 16 18 20
No: of mearurement8
flow-through mode
,~~
2 4 6 8 10 12 14 18 18 20
No. of mearurementr
flow-injection mode
;;,
2 4 6 8 10 12 14 16 18 20
No. of merrurementr
FIG. 10. The within run precision of the analytical data (n = 15).
246 JENEY ET AL.
150
Y - A + Bx
A - 1.2 125 ! Y - A + 9x
B - 0.95 A - -2.96
rsd.- 5. 6
0.95
y 100 t-l - 0.927 100i rsd.- 4.9
- 0.975
2
I
P
75
6
Y
2% 50
0
SO 100 1 0
B
A
FIG. 11. Correlation study for the potassium ion concentrations in urine samples. (A) HORIBA
SERA-210 Ion analyzer vs FAES. (B) Flow-injection analysis vs HORIBA SERA-210 Ion analyzer.
The within run precision of the analytical data relevant to urine samples, per-
formed by the optimized flow-injection system, is exhibited in Fig. 10 in compar-
ison with data obtained by the batch operation mode and by the commercial
SERA-210 Ion analyzer, respectively.
The potassium concentration data of undiluted urine samples of individual pa-
tients determined by the flow-injection analysis technique were correlated to
those obtained with the continuous flow method (SERA-210) as well as with
FAES (Fig. 11).
The correlation presented proves the value of the FIA method as well as the
macro wall-jet cell arrangement with a commercially available potassium ion-
selective membrane electrode of optimized composition, for the analysis of bio-
logical samples of microliter volumes with a rate of 100 samples per hour.
ACKNOWLEDGMENTS
The authors thank Prof. Dr. R. Kellner for the helpful discussion on FTIR-ATR investigations and
the Hungarian Academy of Sciences for providing financial support for this research (OTKA 421/88).
REFERENCES
1. Pungor, E.; Fehbr, Zs.; Nagy, G.; T&h, K. CRC Wt. Rev. Anal. Chem., 1983, 14(3), 175-230.
2. Pungor, E.; Fehbr, Zs.; Nagy, G.; T&h, K.; Horvai, G.; Gratzl, M. Anal. Chim. Acta, 1979,109,
l-24.
3. T&h, K.; Fucsk6, J.; Lindner, E.; Feher, Zs.; Pungor, E. Anal. Chim. Acta, 1986, 179, 359-370.
4. Oesch, U.; Anker, P.; Amman, D.; Simon, W. Membrane technological optimization of ion-
selective electrodes based on solvent polymeric membranes for clinical applications. In Zon-
Selective Electrodes (E. Pungor, Ed.), Vol. 4, pp. 81-112. Akademiai Kiad6, Budapest, 1985.
5. Kiilpmann, W. R. J. Clin. Chem. Clin. Biochem., 1989, 27, 815-824.
6. Meier, P. C.; Amman, D.; Morf, W. E.; Simon, W,. Liquid-membrane ion-selective electrodes and
their biomedical applications. In Medical and Biological Applications of Electrochemical De-
vices (J. Koryta, Ed.), pp. 13-91. Wiley, New York, 1980.
7. TBke, L.; Bitter, I.; Agai, B.; Csongor, E.; T6th, K.; Lindner, E.; Horvath, M.; Harfouch, S.;
Pungor, E. Liebigs Ann. Chem., 1988, 349-353.
FLOW-INJECTION ASSAY OF POTASSIUM 247
8. Craggs, A.; Moody, Cl. J.; Thomas, J. D. R. J. Chem. Educ., 1974, 51, 541-544.
9. Meier, P. C.; Amman, D.; Osswald, H. F.; Simon, W. Med. Progr. Technol., 1977, 5, l-12.
10. Burr, R. G. Clin. C/tern., 1982, 28(7), 1710-1711.
II. Guilbault, G. G.; Durst, R. A.; Frant, M. S.; Freiser, H.; Hansen, E. H.; Light, T.; Pungor, E.;
Rechnitz, G. A.; Rice, N. M.; Rohn, T. J.; Simon, W.; Thomas, J. D. R. Pure Appl. Chem.,
1976, 48, 127-132.
12. Lindner, E.; T&h, K.; Pungor, E.; Berube, T. R.; Buck, R. P. Anal. Chem., 1987,59,2213-2216.
13. Kellner, R.; G&zinger, G. Microchim. Acta, 1984, II, 61-74.
14. T&h, K.; Lindner, E.; Horvath, M.; Jeney, J.; Bitter, I.; Agai, B.; Meisel, T.; Tiike, L. Anal.
Lert., 1989, 22(S), 1185-1207.
15. Horvai, G.; Szepesvary, T.; Pungor, E. Hungarian Patent 185451,tiled August 23, 1981.
16. Gunasingham, H.; Fleet, B. Anal. Chem., 1983, 55, 1409-1414.
17. Frenzel, W. Analyst, 1988, 113, 1039-1046.
18. Trojanowicz, M. Some aspects of application of ion-selective electrode detectors in flow analysis.
In Modern Trends in Analytical Chemistry (E. Pungor, G. Veress, and I. Buzas, Ed%), pp.
189-215. Akadtmiai Kiad6, Budapest, 1984.
19. Lindner, E.; T&h, K.; Jeney, J.; Horvath, M.; Pungor, E.; Bitter, I.; Agai, B.; Tiike, L. Micro-
chim. Acta, 1990, I, 157-168.
20. Douglas, J. G. Anal. Chem., 1989, 61, 922-924.
21. Tarcali, J.; Nagy, G.; Tbth, K.; Pungor, E. Anal. Chim. Acta, 1985, 178, 231-237.
22. Ma, S. C.; Chaniotakis, N. A.; Meyerhoff, M. E. Anal. Chem., 1988, 60, 2293-2299.
23. Diirselen, L. F. J.; Wegmann, D.; May, K.; Oesch, U.; Simon, W. Anal. Chem., 1988, 60,
1455-1458.
24. T6th, K.; Lindner, E.; Jeney, J.; Graf, E.; Horvath, M.; Pungor, E.; Bitter, I.; Meisel, T.; Agai,
B.; TBke, L.; Buck, R. P.; Kellner, R. New directions in the research of solvent-polymeric
membranes. In Zon-Selective Electrodes (E. Pungor, Ed.), Vol. 5, pp. 181-206. Akademiai
Kiadb, Budapest, 1989.
25. Lindner, E.; T6th, K.; Pungor, E.; Morf, W. E.; Simon, W. Anal. Chem., 1978, 50, 1627-1631.
26. Huser, M.; Gehrig, P. M.; Morf, W. E.; Simon, W.; Lindner, E.; Jeney, J.; T6th, K.; Pungor, E.
Anal. Chem., 1991, 63, 1380-1386.
27. Van der Linden, W. E.; Gestervink, R. Anal. Chim. Acta, 1978, 101, 419-422.
28. FehCr, Zs.; Nagy, G.; Tbth, K.; Pungor, E. Anal. Chim. Acta, 1978, 98, 193-203.