800 Series Part Numbers & Product Descriptions

HOW DOES AN 800 SERIES CALL SYSTEM WORK? Two methods of reset are available - button or magnetic
key. Reset buttons are fine for non-secure applications DISABLED PERSONS TOILET ALARM KITS OVERDOOR LIGHTS AND SOUNDERS

800 SERIES
INFRARED TRANSMITTERS
Depending on the type of calling devices used, the Can be used to increase call indication levels
system can be set up to operate as a standard system – on but reset keys are usually preferred in care NC951 Standard disabled WC alarm kit For use with infrared call points & ceiling receivers
establishments as they help prevent patients c/w NC942B call controller; NC807C ceiling pull,; NC806C Overdoor light QT412RXA Infrared transmitter
which standard (constant tone) calls are generated – or NC806CS overdoor light & sounder; NC809DBB NC806CS Overdoor light & sounder
inadvertently cancelling their own calls. Push/pull for emergency call
an emergency system – on which standard (constant tone) button reset point c/w sounder; WC sticker
NC806C/SS Overdoor light, s/steel (requires a QT424/1 or QT424/10 for
and emergency (intermittent tone) calls are generated. Call points without an on-board reset facility, such as NC951/SS Standard disabled WC alarm kit, s/steel NC806CS/SS Overdoor light & sounder, s/steel charging)

call system
ceiling pulls, must be reset via a reset point or call point c/w NC941/SS call controller; NC807C ceiling pull,
NC887D Switchable sounder QT412RXCA Infrared transmitter
On standard systems: NC806CS/SS overdoor light & sounder; NC809DB/SS
with an on-board reset facility. NC888D Remote sounder Push for standard call / pull for emergency
The user calls for assistance by activating a ceiling pull, button reset point; WC sticker
(requires a QT424/1 or QT424/10 for charging)
standard call point, monitoring point, water resistant A full list of standard and emergency call points and the
SINGLE ZONE EMERGENCY INDICATOR PANELS STANDARD CALL POINTS QT424/1 Single-way charging unit
alert point or infrared ceiling receiver (as fitted). A red facilities they offer can be found at the back of this leaflet. Indicate standard and emergency calls for use with QT412 range transmitters
Generate standard calls only
confidence light on the calling device illuminates to Ancillary calling devices NC942 Single zone call controller NC807C Ceiling pull unit QT424/10 10-way charging unit
reassure the user that the call has been registered. A tail call lead or hand/foot operated pneumatic pad can c/w 12V 140mA PSU, reset button & relay No on-board reset, no remote socket for use with QT412 range transmitters
At the system’s indicator panel(s), an indicator light be connected to a call point’s remote socket (if fitted) to NC942B Single zone call controller NC802DB Standard call point QT432 Patient neck pendant
generate a standard call. For safety reasons, a standard c/w 12V 140mA PSU, reset button, relay and Button reset, with remote socket
illuminates and a buzzer sounds to inform staff that onboard rechargeable battery backup supply NC802DM Standard call point
c/w anti-ligature lanyard, push for standard

someone is in need of assistance. Relevant overdoor call will also be made if the ancillary calling device is call, battery operated, requires 12V A23/23A
Ideal for use in:
NC941 Single zone call controller Magnetic reset, with remote socket (8LR50) alkaline battery (supplied)
lights, remote sounders and area indicators also operate inadvertently pulled from its socket. c/w 12V 250mA PSU & reset button NC917L Standard call push with protruding button
(as fitted) to provide additional audible and visual Monitoring points NC941/SS Single zone call controller, s/steel No on-board reset, no remote socket RESET POINTS Hotels
indication of the call. Monitoring points can be used to trigger a standard call if c/w 12V 250mA PSU & reset button NC802DEWM Master water resistant alert point NC809DB Button reset point
Leisure centres
NC941B 0.3AHr battery backup kit magnetic reset, no remote socket
A member of staff visits the calling room to assist the a fire exit, drug cupboard or storeroom is opened and can NC809DBB Button reset point c/w sounder
for use with NC941 & NC941/SS call controller NC802DEWS Slave water resistant alert point
user and resets the calling device to return the system to be isolated via an on-board keyswitch if required. They Up to 10 per master
Label also includes Braille text Nursing homes
normal. can also be connected to any device with a normally open 4-12 ZONE EMERGENCY INDICATOR PANELS NC802DESS IP69K sauna alert point NC809DB/SS Button reset point, s/steel
or normally closed switch, such as a pressure mat, allowing Indicate standard and emergency calls Must be used as a slave to an NC802DEWM NC809DM Magnetic reset point Care homes
On emergency systems: it to operate as a standard call point when activated. NC944 Four zone call controller NC802DEM/SS Stainless steel call point NC899IRR Keyswitch reset point
In addition to doing everything a standard system does, c/w mute/reset button, requires 12V PSU Magnetic reset, no remote socket Hospitals
a more urgent / extra level of assistance can be Call latch modules NC944/SS Four zone call controller, s/steel MAGNETIC RESET KEYS
summoned by activating an ‘emergency call’ via an Call latch modules operate in a similar way to monitoring c/w mute/reset button, requires 12V PSU STANDARD MONITORING POINTS For use with all magnetic call points, monitoring Shops
emergency call point, water resistant alert point or points but are smaller (for discreet mounting) and do not NC930 12V 250mA PSU Generate standard calls only points and reset points
include an onboard reset or isolating facility. powers up to 2 x four zone call controllers NC894DKB Isolatable monitoring Point Lavatories
infrared ceiling receiver (as fitted). NC803M/10 10 x magnetic reset keys
NC930/SS 12V 250mA PSU, s/steel Button reset, no remote socket
Once activated, the red confidence light on the calling Infrared call points, ceiling receivers and transmitters powers up to 2 x four zone call controllers NC894DKM Isolatable Monitoring Point NC803M/50 50 x magnetic reset keys Guest houses
device will flash to confirm the emergency call has been To help protect staff against aggressive visitors, distressed NC941B 0.3AHr battery backup kit Magnetic reset, no remote socket
patients or intruders, infrared ceiling receivers and call for use with 12V 250mA PSU ANCILLARY CALLING DEVICES Gymnasiums
registered and the relevant indicator panel light and its NC925B1 12V 1A white boxed PSU EMERGENCY CALL POINTS Generate standard call via a call point’s remote socket
buzzer will flash and sound a more insistent tone to points can be used in conjunction with infrared Fitness suites
powers up to 3 x four zone call controllers Generate standard and/or emergency calls (as shown) NC805C/6 1.8m (6ft) tail call lead
attract attention. Relevant overdoor lights, remote transmitters to remotely trigger standard and/or
AC284 2.1AHr backup battery NC802DEB-1/2 Emergency call point NC805C/14 4.2m (14ft) tail call lead
sounders and area indicators will also pulse (as fitted) to emergency calls. Up to three slave ceiling receivers can for use with 12V 1A boxed PSU Button reset, with remote socket, Solariums
be connected to one master ceiling receiver or infrared can generate standard & emergency calls NC805D 1.2-3.6m (4-12ft) tail call lead
provide additional audible and visual indication of the 10-30 ZONE EMERGENCY INDICATOR PANELS Police cells
emergency call. call point to provide additional coverage in large or L NC802DEM Emergency call point NC805P Hand/foot pneumatic pad
shaped rooms. Infrared transmitters typically provide a Indicate standard and emergency calls Magnetic reset, with remote socket requires 1 x NC805AS
10m line-of-sight transmitting range. NC812KE 10 Zone Emergency master panel can generate standard & emergency calls NC805AS Remote air switch for NC805P Prisons
WHICH SYSTEM SHOULD YOU CHOOSE? c/w 12V PSU, backup battery & mute button NC802DEWM Master water resistant alert point NC805MD Portable movement detector
Overdoor lights NC811KE 10 Zone Emergency repeater panel magnetic reset, no remote socket, can generate Factories
The type of system required will depend upon the Overdoor lights (if fitted) are normally installed outside standard or emergency calls (not both) NC805DSP Double remote socket
NC822KE 20 Zone Emergency master panel
application and the needs of the user so always check with rooms to increase call indication levels. If required, up to c/w 12V PSU, backup battery & mute button
allows two ancillary calling devices to Warehouses
NC802DEWS Slave water resistant alert point be used at the same time
the approving authority/client before work commences. four zones can be connected to one overdoor light to NC821KE 20 Zone Emergency repeater panel Max. 10 per master
MISCELLANEOUS 800 SERIES DEVICES Offices
In residential care homes, a standard system is normally provide area indication (in corridors, etc) with additional NC832KE 30 Zone Emergency master panel NC802DESS IP69K sauna alert point
acceptable as the main requirement is for the reset zones connectable via one or more input expanders. c/w 12V PSU, backup battery & mute button Must be used as a slave to an NC802DEWM NC899KD Area isolator Reception areas
can be used to isolate areas/zones
function to be carried out at the source of the call. In NC831KE 30 Zone Emergency repeater panel
Remote sounders NC802DEM/SS Stainless steel call point
nursing homes, however, some authorities insist on Up to three switchable or remote sounders can be Larger panels available to order magnetic reset, no remote socket, can generate
NC884DE Emergency 5:1 input expander GP surgeries
can be used to provide area/zonal indication
standard and emergency call facilities so an emergency standard or emergency calls (not both)
connected to an indicator panel’s sounder output to 10-30 ZONE STANDARD INDICATOR PANELS NC885 Standard 5:1 input expander Government buildings
system may be required. In most other applications, such NC804DE Emergency call / reset point
increase call levels. Indicate standard calls ONLY Button reset, no remote socket, can be used to provide area/zonal indication
as shops, leisure centres, hotels and public sector Relay units NC810K 10 Zone Standard master panel can generate emergency calls only NC889 Call latch module Interview rooms
buildings, the type of system and devices required will External strobes, sounders, beacons, etc, can be connected c/w 12V PSU & reset button NC809DEM Emergency call / reset point can trigger a standard call from a normally
depend almost entirely on what the user wants to NC811K 10 Zone Standard repeater panel Magnetic reset, no remote socket, open or normally closed switch Housing offices
to any zone or group of zones via a 12V relay unit. can generate emergency calls only
NC889ERM Multi-purpose call latch module
achieve. In most instances, referring to the typical NC820K 20 Zone Standard master panel Restaurants
applications diagram overleaf will help all interested Tone only paging c/w 12V PSU & reset button
INFRARED CALL POINTS & CEILING RECEIVERS
can trigger a standard and/or emergency call
from a normally open switch (includes
parties come to a decision on what components should A simple ‘tone-only’ paging option is available (requires NC821K 20 Zone Standard repeater panel confidence LED and onboard magnetic reset) Changing rooms
Generate standard and/or emergency calls when
be purchased. a DP874B/4 UHF paging transmitter). When a call is NC830K 30 Zone Standard master panel triggered by a compatible infrared transmitter NC883D 12V relay unit
made, tone only pagers will beep, prompting staff to c/w 12V PSU & reset button
NC302RXC Infrared master ceiling receiver can be used to switch sirens, strobes, etc. and much, much more
visit the nearest call indicator panel to determine the NC831K 30 Zone Standard repeater panel No on-board reset, no remote socket
PLANNING A SYSTEM exact source and nature of the call. Larger panels available to order NC302RXCS Infrared slave ceiling receiver
NC848/50 50 x Anti-tamper screws
can be used to secure stainless steel devices
The following points should be considered when Up to 3 per master. in high risk environments.
Master reset buttons NC802DERB Infrared call point
designing or planning an 800 Series Call System. Some indicator panels have an optional master reset Button reset, with remote socket
DP874B/4 Tone only UHF paging transmitter
supplied with one tone-only pager
Indicator panels button. If enabled, this facility allows standard calls to NC802DERM Infrared call point
Magnetic reset, with remote socket DP875/4 Extra tone-only pager
Indicator panels should be sited internally in a clean dry be globally cancelled without the source of the call
area which is readily accessible to staff. Various types being investigated. As there is a definite requirement Distributed by:
and sizes are available covering 1 to 90 zones. Repeater in most applications for the caller to be visited before
panels are also available. the call is reset, this facility is normally only utilised in
carefully managed staff-to-staff call systems. In
Call Points
most other instances we recommend this facility is
Any number of call points (of any kind), reset points and disabled.
overdoor lights can be wired onto one zone.
Mute buttons C-TEC
Standard call points (including ceiling pulls and All multi-zone emergency indicator panels are supplied
monitoring points) will work on 800 Series standard and with an optional mute button. If enabled, this allows Stephens Way, Wigan,
emergency systems but emergency call points will work any standard calls on the system to be silenced but WN3 6PH. United Kingdom
on emergency systems only. leaves relevant indicator light(s) lit. Although a new call Tel: +44 (0)1942 322744
Some emergency call points can generate standard and from a different call point will re-activate the sounders, Fax: +44 (0)1942 829867
emergency calls whereas others, such as water resistant to ensure callers are visited we recommend this facility is
alert points and stainless steel call points, can be set up used only in very carefully managed systems. Emergency Email: sales@c-tec.co.uk Quality System Certificate No. 176
to generate standard or emergency calls (not both). calls cannot be muted. Web: www.c-tec.co.uk Assessed to ISO9001 : 2008

Approved Document. No. DNS8002255 Rev 4. The manufacturer operates a policy of continuous improvement and reserves the right to alter product specifications at its discretion and without prior notice. Errors and omissions excepted.
No responsibility can be accepted by the manufacturer or distributor for any misinterpretation of an instruction or guidance note, or for the compliance of the system as a whole.
800 SERIES 800 SERIES
call system
Typical Applications & Wiring Details call system
Tried, trusted and respected by users and CALL RESET

installers alike, C-TEC’s 800 Series Call System


nursing home
ALERT

is ideal for use in all types of private, NC807C NC806C


NC802DERM MAGNETIC
NC807C
NC806C
OVERDOOR
NC807C
CEILING
NC802DEB-1/2
BUTTON RESET
CEILING PULL OVERDOOR LIGHT CEILING PULL LIGHT PULL EMERGENCY
commercial, residential and public sector RESET EMERGENCY

& hospital
Extra ceiling CALL POINT
CALL RESET INFRA-RED CALL POINT RESET

pulls can
buildings. Incorporating the UK’s market- ALERT CALL RESET be fitted as
required
leading disabled persons toilet alarm kit, a
Call
Call

call systems


NC802DEB-1/2 NC806CS NC809DBB
wide range of indicator panels and an extensive
Call

BUTTON RESET OVERDOOR BUTTON NC807C


EMERGENCY NC803M LIGHT & RESET CEILING
QT412RXCA SOUNDER POINT C/W
array of system components, it is without
CALL POINT MAGNETIC PULL
INFRARED TRANSMITTER RESET KEY SOUNDER
TAIL CALL LEAD (PUSH FOR CALL / PULL IMPORTANT: There is a definite requirement in NC894DKB
doubt one of the most versatile and reliable FOR EMERGENCY) nursecall systems for the reset function to be PRESSURE PAD ON
RESET
ISOLATABLE
BUTTON
carried out at the room from which the call is made
conventional call systems on the market. RESET
ISOLATED

Call
MONITORING

hotel &
NC802DB POINT
NC806C NC302RXC NC302RXCS
BUTTON RESET OVERDOOR INFRA-RED MASTER INFRA-RED SLAVE TAIL CALL LEAD OR
CALL POINT LIGHT CEILING RECEIVER CEILING RECEIVER

guest house
CALL RESET
ON F Switch
U
S Fused
Hundreds of applications - suitable for use in hotels, Two levels of call - standard (constant tone) and
E

Small localised system


MUTE RESET MUTE RESET
Spur
Call Call Call
Call

may be linked to indicate


leisure centres, nursing homes, shops, offices, etc. emergency (intermittent tone). on larger master panels
NC944 FOUR NC944 FOUR NC930 12V 250mA

call systems
ZONE CALL ZONE CALL POWER SUPPLY DOOR P.I.R


CONTROLLER CONTROLLER RESET
NC803M CONTACTS SENSOR
Optional button or magnetic key reset facilities
MAGNETIC
Hard wired for total reliability - 4 to 6 core QT412RXA
INFRARED
RESET KEY BED MOVEMENT
DETECTOR
to detect patients
wandering about
security cable is ideal for most installations. (dependent on items purchased). TRANSMITTER
(PUSH / PULL FOR NC809DM
1 core per zone EMERGENCY) MAGNETIC RESET POINT
Wide range of indicator panels covering 1 to 90 zones. No limit to system size - components can be mixed & common -Ve

and matched to meet the layout and call Tone-only paging


Extensive range of ancillaries including ceiling pulls, facilities also available
communication requirements of virtually any CALL RESET
NC803M
MAGNETIC
wall-mounting call points, monitoring points, NC302RXCS NC302RXC NC899IRR RESET KEY
building - see typical applications diagram for details. INFRA-RED SLAVE INFRA-RED MASTER KEYSWITCH
water resistant alert points, stainless steel call points, CEILING RECEIVER CEILING RECEIVER RESET POINT
NC812KE / 822KE / 832KE NC811KE / 821KE / 831KE NC806C
Call

NC802DM
overdoor lights, remote sounders, relays and Optional paging and infrared people protection 10, 20 OR 30 ZONE
EMERGENCY MASTER PANEL
10, 20 OR 30 ZONE
EMERGENCY REPEATER PANEL
OVERDOOR LIGHT MAGNETIC RESET
CALL POINT
call latch modules. facilities.
ON

ISOLATED

Call

12V d.c. system - simple to install, easy to operate FREE plan marking up and technical advice Up to three slave
ceiling receivers
Switch
Fused
F
U
1 core per zone
NC944/SS
STAINLESS STEEL NC930/SS
QT412RXA S
NC941B 12V BATTERY
and highly reliable. service available for larger systems. can be used per FOUR ZONE CALL STAINLESS STEEL
E

Spur + 3 core control 1 core per zone


INFRARED TRANSMITTER CONTROLLER 12V 250mA PSU BACK-UP KIT
master receiver + 2 core power & common -Ve
(PUSH / PULL FOR EMERGENCY)
(use 1mm2 for +Ve & -Ve)

Most wall-mounting plastic components fit standard Extremely cost-effective - all components CALLING
ON

25mm single or double gang back boxes. Ceiling normally available ex-stock. Small localised system Call MUTE RESET
Call

NC802DEWS NC802DEWM
pulls and water resistant alert points are surface may be linked to indicate Optional 12V 0.3AHr

retail, sports
SLAVE WATER MASTER WATER
Manufactured in the UK by an ISO 9001 accredited on larger master panels battery back-up kit
F


RESISTANT RESISTANT U

mountable.
S
E
ALERT POINT ALERT POINT
company with over 25 years’ experience of making ON OFF
Switch
Attractive compact design blends perfectly quality life safety and call system products. Fused Spur NC803M

& leisure
NC803M

general surgery
CALL
MAGNETIC
into any sort of decor. MAGNETIC RESET KEY
Call Call
Call Call

Can be used to help building managers and service RESET KEY NC888D
REMOTE
NC887D
SWITCHABLE Call

call systems
All power supplies capable of 12V sealed lead acid providers comply with BS8300, the Disability s SOUNDER SOUNDER NC806C/SS S/STEEL NC802DEM/SS

& interview room


OVERDOOR LIGHT S/STEEL CALL POINT
battery charging. Discrimination Act and the Care Standards Act. NC806CS
Switchable and Remote Sounders are
available to increase sound levels
NC802DEM/SS
OVERDOOR
S/STEEL

call systems
LIGHT &
CALL POINT


CALL
SOUNDER CALLING

NC917L CALL PUSH WITH

WHAT IS AN 800 SERIES CALL SYSTEM?


PUSH TO CALL
Call NC803M MAG. RESET KEY PROTRUDING BUTTON
NC915L desk pushes & NC920 wall sockets are also
RESET
External CALL RESET NC807C available for use as discreet panic alarm buttons.

In its simplest form, an 800 Series Call System: - bell push


at delivery
CEILING PULL
gate or
• Allows the general public, a member of staff or a patient to call for assistance; NC899IRR
Call
NC894DKM ISOLATABLE
Call

NC807C door
MAGNETIC RESET CEILING PULL NC809DB BUTTON NC889 CALL NC802DB BUTTON KEYSWITCH ON

ISOLATED
MONITORING POINT RESET POINT LATCH MODULE RESET CALL POINT RESET POINT
• Confirms that the call has got through;
Call

Standard RESET
ALERT RESET
NC302RXC
burglar ON Relay Unit
INFRA-RED
MASTER
• Makes sure the caller is visited; and, if required;
Relay
alarm TO OTHER EQUIPMENT Activated RESET normally open QT412RXCA
ISOLATED
NC803M CEILING
contacts Call
or normally INFRARED TRANSMITTER
Call
MAGNETIC e.g. strobes, sirens, ON
RECEIVER
NC804DE NC806CS closed voltage (PUSH FOR STANDARD CALL
RESET KEY ISOLATED

• Allows extra or more urgent assistance to be summoned using an EMERGENCY OVERDOOR


beacons, door releases, Call Call free contacts PULL FOR EMERGENCY)
a CALL / BUTTON LIGHT & etc. (some of which may
optional ‘emergency’ call facility. RESET POINT SOUNDER require an external PSU)
NC883D
12V RELAY
NC894DKB ISOLATABLE
BUTTON RESET
EQUIPMENT TO
BE MONITORED
UNIT MONITORING POINT
b le

However, the unrivalled flexibility and versatility of the 800 Series Call System NC809DBB
means it can also be used to: -
NC807C BUTTON RESET NC806CS OVERDOOR CABLE & WIRING DETAILS
il a

CEILING PULL POINT C/W SOUNDER LIGHT & SOUNDER


va

a
warehouse
■ STANDARD CALL POINTS (including ceiling pulls, monitoring points
• Inform staff that someone is being attacked via an optional infrared NC899IRR NC302RXC NC302RXCS EXTRA LOW VOLTAGE (7/0.2) SECURITY CABLE
RESET

ow
KEYSWITCH INFRA-RED MASTER INFRA-RED SLAVE and NC809DB and NC809DM reset points): Run 1 signal wire per zone
n ‘staff attack’ facility.
RESET

RESET POINT CEILING RECEIVER CEILING RECEIVER ■ Typically 4 or 6 core dependent on layout and devices plus a common negative to the nearest panel
ies
ON
used (see right).
ar
Call

■ EMERGENCY CALL POINTS (including infrared ceiling receivers and


c il l
& factory
ISOLATED

Wid • Monitor storage cupboards, cash offices and stock rooms for unauthorised
Call ■ Multi-core cable can be used to reduce wiring runs
an
NC899IRR and NC809DBB reset points): Run 1 signal wire per zone
e ra n
s t e el
ON Door NC894DKB NC806C but more joints will result.
plus a common negative and a common positive to the nearest panel
ge o f s ta i nl e s s access. NC951 STANDARD KIT Switch
Fused
F
U
ON CALLING
Call
ISOLATED
contacts ISOLATABLE
BUTTON RESET
OVERDOOR
LIGHT
■ We recommend you leave about 20-25% spare cores.
■ OVERDOOR LIGHTS: Connect to the signal wire of the zone being
S

(NC951B ELITE KIT WITH BATTERY E


■ DO NOT use solid core cable as it breaks easily.

call systems
Spur QT412RXA INFRARED MONITORING POINT
• Monitor doorbells, telephones and machinery for activation or failure.
RESET

BACK-UP FACILITY ALSO AVAILABLE) indicated and to common negative and common positive


Call
TRANSMITTER (PUSH /
NC942 SINGLE ZONE CALL PULL FOR EMERGENCY) 220-240V MAINS FEED, e.g. 1mm T&E ■ REMOTE SOUNDERS: Usually connected directly to an indicator panel
CONTROLLER / 12V PSU
• Send notification of calls to hand-held paging equipment. using two cores
© C-TEC 1999-2006. Errors and omissions excepted. No responsibility can be accepted by C-TEC for any misinterpretation of the above or for the compliance of the system as a whole. We reserve the right to alter product specifications at our discretion and without prior notice. OTHER (as specified) ■ INDICATOR PANELS / REPEATERS: Wire as specified in diagram
800 SERIES 800 SERIES
call system
Typical Applications & Wiring Details call system
Tried, trusted and respected by users and CALL RESET

installers alike, C-TEC’s 800 Series Call System


nursing home
ALERT

is ideal for use in all types of private, NC807C NC806C


NC802DERM MAGNETIC
NC807C
NC806C
OVERDOOR
NC807C
CEILING
NC802DEB-1/2
BUTTON RESET
CEILING PULL OVERDOOR LIGHT CEILING PULL LIGHT PULL EMERGENCY
commercial, residential and public sector RESET EMERGENCY

& hospital
Extra ceiling CALL POINT
CALL RESET INFRA-RED CALL POINT RESET

pulls can
buildings. Incorporating the UK’s market- ALERT CALL RESET be fitted as
required
leading disabled persons toilet alarm kit, a
Call
Call

call systems


NC802DEB-1/2 NC806CS NC809DBB
wide range of indicator panels and an extensive
Call

BUTTON RESET OVERDOOR BUTTON NC807C


EMERGENCY NC803M LIGHT & RESET CEILING
QT412RXCA SOUNDER POINT C/W
array of system components, it is without
CALL POINT MAGNETIC PULL
INFRARED TRANSMITTER RESET KEY SOUNDER
TAIL CALL LEAD (PUSH FOR CALL / PULL IMPORTANT: There is a definite requirement in NC894DKB
doubt one of the most versatile and reliable FOR EMERGENCY) nursecall systems for the reset function to be PRESSURE PAD ON
RESET
ISOLATABLE
BUTTON
carried out at the room from which the call is made
conventional call systems on the market. RESET
ISOLATED

Call
MONITORING

hotel &
NC802DB POINT
NC806C NC302RXC NC302RXCS
BUTTON RESET OVERDOOR INFRA-RED MASTER INFRA-RED SLAVE TAIL CALL LEAD OR
CALL POINT LIGHT CEILING RECEIVER CEILING RECEIVER

guest house
CALL RESET
ON F Switch
U
S Fused
Hundreds of applications - suitable for use in hotels, Two levels of call - standard (constant tone) and
E

Small localised system


MUTE RESET MUTE RESET
Spur
Call Call Call
Call

may be linked to indicate


leisure centres, nursing homes, shops, offices, etc. emergency (intermittent tone). on larger master panels
NC944 FOUR NC944 FOUR NC930 12V 250mA

call systems
ZONE CALL ZONE CALL POWER SUPPLY DOOR P.I.R


CONTROLLER CONTROLLER RESET
NC803M CONTACTS SENSOR
Optional button or magnetic key reset facilities
MAGNETIC
Hard wired for total reliability - 4 to 6 core QT412RXA
INFRARED
RESET KEY BED MOVEMENT
DETECTOR
to detect patients
wandering about
security cable is ideal for most installations. (dependent on items purchased). TRANSMITTER
(PUSH / PULL FOR NC809DM
1 core per zone EMERGENCY) MAGNETIC RESET POINT
Wide range of indicator panels covering 1 to 90 zones. No limit to system size - components can be mixed & common -Ve

and matched to meet the layout and call Tone-only paging


Extensive range of ancillaries including ceiling pulls, facilities also available
communication requirements of virtually any CALL RESET
NC803M
MAGNETIC
wall-mounting call points, monitoring points, NC302RXCS NC302RXC NC899IRR RESET KEY
building - see typical applications diagram for details. INFRA-RED SLAVE INFRA-RED MASTER KEYSWITCH
water resistant alert points, stainless steel call points, CEILING RECEIVER CEILING RECEIVER RESET POINT
NC812KE / 822KE / 832KE NC811KE / 821KE / 831KE NC806C
Call

NC802DM
overdoor lights, remote sounders, relays and Optional paging and infrared people protection 10, 20 OR 30 ZONE
EMERGENCY MASTER PANEL
10, 20 OR 30 ZONE
EMERGENCY REPEATER PANEL
OVERDOOR LIGHT MAGNETIC RESET
CALL POINT
call latch modules. facilities.
ON

ISOLATED

Call

12V d.c. system - simple to install, easy to operate FREE plan marking up and technical advice Up to three slave
ceiling receivers
Switch
Fused
F
U
1 core per zone
NC944/SS
STAINLESS STEEL NC930/SS
QT412RXA S
NC941B 12V BATTERY
and highly reliable. service available for larger systems. can be used per FOUR ZONE CALL STAINLESS STEEL
E

Spur + 3 core control 1 core per zone


INFRARED TRANSMITTER CONTROLLER 12V 250mA PSU BACK-UP KIT
master receiver + 2 core power & common -Ve
(PUSH / PULL FOR EMERGENCY)
(use 1mm2 for +Ve & -Ve)

Most wall-mounting plastic components fit standard Extremely cost-effective - all components CALLING
ON

25mm single or double gang back boxes. Ceiling normally available ex-stock. Small localised system Call MUTE RESET
Call

NC802DEWS NC802DEWM
pulls and water resistant alert points are surface may be linked to indicate Optional 12V 0.3AHr

retail, sports
SLAVE WATER MASTER WATER
Manufactured in the UK by an ISO 9001 accredited on larger master panels battery back-up kit
F


RESISTANT RESISTANT U

mountable.
S
E
ALERT POINT ALERT POINT
company with over 25 years’ experience of making ON OFF
Switch
Attractive compact design blends perfectly quality life safety and call system products. Fused Spur NC803M

& leisure
NC803M

general surgery
CALL
MAGNETIC
into any sort of decor. MAGNETIC RESET KEY
Call Call
Call Call

Can be used to help building managers and service RESET KEY NC888D
REMOTE
NC887D
SWITCHABLE Call

call systems
All power supplies capable of 12V sealed lead acid providers comply with BS8300, the Disability s SOUNDER SOUNDER NC806C/SS S/STEEL NC802DEM/SS

& interview room


OVERDOOR LIGHT S/STEEL CALL POINT
battery charging. Discrimination Act and the Care Standards Act. NC806CS
Switchable and Remote Sounders are
available to increase sound levels
NC802DEM/SS
OVERDOOR
S/STEEL

call systems
LIGHT &
CALL POINT


CALL
SOUNDER CALLING

NC917L CALL PUSH WITH

WHAT IS AN 800 SERIES CALL SYSTEM?


PUSH TO CALL
Call NC803M MAG. RESET KEY PROTRUDING BUTTON
NC915L desk pushes & NC920 wall sockets are also
RESET
External CALL RESET NC807C available for use as discreet panic alarm buttons.

In its simplest form, an 800 Series Call System: - bell push


at delivery
CEILING PULL
gate or
• Allows the general public, a member of staff or a patient to call for assistance; NC899IRR
Call
NC894DKM ISOLATABLE
Call

NC807C door
MAGNETIC RESET CEILING PULL NC809DB BUTTON NC889 CALL NC802DB BUTTON KEYSWITCH ON

ISOLATED
MONITORING POINT RESET POINT LATCH MODULE RESET CALL POINT RESET POINT
• Confirms that the call has got through;
Call

Standard RESET
ALERT RESET
NC302RXC
burglar ON Relay Unit
INFRA-RED
MASTER
• Makes sure the caller is visited; and, if required;
Relay
alarm TO OTHER EQUIPMENT Activated RESET normally open QT412RXCA
ISOLATED
NC803M CEILING
contacts Call
or normally INFRARED TRANSMITTER
Call
MAGNETIC e.g. strobes, sirens, ON
RECEIVER
NC804DE NC806CS closed voltage (PUSH FOR STANDARD CALL
RESET KEY ISOLATED

• Allows extra or more urgent assistance to be summoned using an EMERGENCY OVERDOOR


beacons, door releases, Call Call free contacts PULL FOR EMERGENCY)
a CALL / BUTTON LIGHT & etc. (some of which may
optional ‘emergency’ call facility. RESET POINT SOUNDER require an external PSU)
NC883D
12V RELAY
NC894DKB ISOLATABLE
BUTTON RESET
EQUIPMENT TO
BE MONITORED
UNIT MONITORING POINT
b le

However, the unrivalled flexibility and versatility of the 800 Series Call System NC809DBB
means it can also be used to: -
NC807C BUTTON RESET NC806CS OVERDOOR CABLE & WIRING DETAILS
il a

CEILING PULL POINT C/W SOUNDER LIGHT & SOUNDER


va

a
warehouse
■ STANDARD CALL POINTS (including ceiling pulls, monitoring points
• Inform staff that someone is being attacked via an optional infrared NC899IRR NC302RXC NC302RXCS EXTRA LOW VOLTAGE (7/0.2) SECURITY CABLE
RESET

ow
KEYSWITCH INFRA-RED MASTER INFRA-RED SLAVE and NC809DB and NC809DM reset points): Run 1 signal wire per zone
n ‘staff attack’ facility.
RESET

RESET POINT CEILING RECEIVER CEILING RECEIVER ■ Typically 4 or 6 core dependent on layout and devices plus a common negative to the nearest panel
ies
ON
used (see right).
ar
Call

■ EMERGENCY CALL POINTS (including infrared ceiling receivers and


c il l
& factory
ISOLATED

Wid • Monitor storage cupboards, cash offices and stock rooms for unauthorised
Call ■ Multi-core cable can be used to reduce wiring runs
an
NC899IRR and NC809DBB reset points): Run 1 signal wire per zone
e ra n
s t e el
ON Door NC894DKB NC806C but more joints will result.
plus a common negative and a common positive to the nearest panel
ge o f s ta i nl e s s access. NC951 STANDARD KIT Switch
Fused
F
U
ON CALLING
Call
ISOLATED
contacts ISOLATABLE
BUTTON RESET
OVERDOOR
LIGHT
■ We recommend you leave about 20-25% spare cores.
■ OVERDOOR LIGHTS: Connect to the signal wire of the zone being
S

(NC951B ELITE KIT WITH BATTERY E


■ DO NOT use solid core cable as it breaks easily.

call systems
Spur QT412RXA INFRARED MONITORING POINT
• Monitor doorbells, telephones and machinery for activation or failure.
RESET

BACK-UP FACILITY ALSO AVAILABLE) indicated and to common negative and common positive


Call
TRANSMITTER (PUSH /
NC942 SINGLE ZONE CALL PULL FOR EMERGENCY) 220-240V MAINS FEED, e.g. 1mm T&E ■ REMOTE SOUNDERS: Usually connected directly to an indicator panel
CONTROLLER / 12V PSU
• Send notification of calls to hand-held paging equipment. using two cores
© C-TEC 1999-2006. Errors and omissions excepted. No responsibility can be accepted by C-TEC for any misinterpretation of the above or for the compliance of the system as a whole. We reserve the right to alter product specifications at our discretion and without prior notice. OTHER (as specified) ■ INDICATOR PANELS / REPEATERS: Wire as specified in diagram
800 Series Part Numbers & Product Descriptions
HOW DOES AN 800 SERIES CALL SYSTEM WORK? Two methods of reset are available - button or magnetic
key. Reset buttons are fine for non-secure applications DISABLED PERSONS TOILET ALARM KITS OVERDOOR LIGHTS AND SOUNDERS

800 SERIES
INFRARED TRANSMITTERS
Depending on the type of calling devices used, the Can be used to increase call indication levels
system can be set up to operate as a standard system – on but reset keys are usually preferred in care NC951 Standard disabled WC alarm kit For use with infrared call points & ceiling receivers
establishments as they help prevent patients c/w NC942B call controller; NC807C ceiling pull,; NC806C Overdoor light QT412RXA Infrared transmitter
which standard (constant tone) calls are generated – or NC806CS overdoor light & sounder; NC809DBB NC806CS Overdoor light & sounder
inadvertently cancelling their own calls. Push/pull for emergency call
an emergency system – on which standard (constant tone) button reset point c/w sounder; WC sticker
NC806C/SS Overdoor light, s/steel (requires a QT424/1 or QT424/10 for
and emergency (intermittent tone) calls are generated. Call points without an on-board reset facility, such as NC951/SS Standard disabled WC alarm kit, s/steel NC806CS/SS Overdoor light & sounder, s/steel charging)

call system
ceiling pulls, must be reset via a reset point or call point c/w NC941/SS call controller; NC807C ceiling pull,
NC887D Switchable sounder QT412RXCA Infrared transmitter
On standard systems: NC806CS/SS overdoor light & sounder; NC809DB/SS
with an on-board reset facility. NC888D Remote sounder Push for standard call / pull for emergency
The user calls for assistance by activating a ceiling pull, button reset point; WC sticker
(requires a QT424/1 or QT424/10 for charging)
standard call point, monitoring point, water resistant A full list of standard and emergency call points and the
SINGLE ZONE EMERGENCY INDICATOR PANELS STANDARD CALL POINTS QT424/1 Single-way charging unit
alert point or infrared ceiling receiver (as fitted). A red facilities they offer can be found at the back of this leaflet. Indicate standard and emergency calls for use with QT412 range transmitters
Generate standard calls only
confidence light on the calling device illuminates to Ancillary calling devices NC942 Single zone call controller NC807C Ceiling pull unit QT424/10 10-way charging unit
reassure the user that the call has been registered. A tail call lead or hand/foot operated pneumatic pad can c/w 12V 140mA PSU, reset button & relay No on-board reset, no remote socket for use with QT412 range transmitters
At the system’s indicator panel(s), an indicator light be connected to a call point’s remote socket (if fitted) to NC942B Single zone call controller NC802DB Standard call point QT432 Patient neck pendant
generate a standard call. For safety reasons, a standard c/w 12V 140mA PSU, reset button, relay and Button reset, with remote socket
illuminates and a buzzer sounds to inform staff that onboard rechargeable battery backup supply NC802DM Standard call point
c/w anti-ligature lanyard, push for standard

someone is in need of assistance. Relevant overdoor call will also be made if the ancillary calling device is call, battery operated, requires 12V A23/23A
Ideal for use in:
NC941 Single zone call controller Magnetic reset, with remote socket (8LR50) alkaline battery (supplied)
lights, remote sounders and area indicators also operate inadvertently pulled from its socket. c/w 12V 250mA PSU & reset button NC917L Standard call push with protruding button
(as fitted) to provide additional audible and visual Monitoring points NC941/SS Single zone call controller, s/steel No on-board reset, no remote socket RESET POINTS Hotels
indication of the call. Monitoring points can be used to trigger a standard call if c/w 12V 250mA PSU & reset button NC802DEWM Master water resistant alert point NC809DB Button reset point
Leisure centres
NC941B 0.3AHr battery backup kit magnetic reset, no remote socket
A member of staff visits the calling room to assist the a fire exit, drug cupboard or storeroom is opened and can NC809DBB Button reset point c/w sounder
for use with NC941 & NC941/SS call controller NC802DEWS Slave water resistant alert point
user and resets the calling device to return the system to be isolated via an on-board keyswitch if required. They Up to 10 per master
Label also includes Braille text Nursing homes
normal. can also be connected to any device with a normally open 4-12 ZONE EMERGENCY INDICATOR PANELS NC802DESS IP69K sauna alert point NC809DB/SS Button reset point, s/steel
or normally closed switch, such as a pressure mat, allowing Indicate standard and emergency calls Must be used as a slave to an NC802DEWM NC809DM Magnetic reset point Care homes
On emergency systems: it to operate as a standard call point when activated. NC944 Four zone call controller NC802DEM/SS Stainless steel call point NC899IRR Keyswitch reset point
In addition to doing everything a standard system does, c/w mute/reset button, requires 12V PSU Magnetic reset, no remote socket Hospitals
a more urgent / extra level of assistance can be Call latch modules NC944/SS Four zone call controller, s/steel MAGNETIC RESET KEYS
summoned by activating an ‘emergency call’ via an Call latch modules operate in a similar way to monitoring c/w mute/reset button, requires 12V PSU STANDARD MONITORING POINTS For use with all magnetic call points, monitoring Shops
emergency call point, water resistant alert point or points but are smaller (for discreet mounting) and do not NC930 12V 250mA PSU Generate standard calls only points and reset points
include an onboard reset or isolating facility. powers up to 2 x four zone call controllers NC894DKB Isolatable monitoring Point Lavatories
infrared ceiling receiver (as fitted). NC803M/10 10 x magnetic reset keys
NC930/SS 12V 250mA PSU, s/steel Button reset, no remote socket
Once activated, the red confidence light on the calling Infrared call points, ceiling receivers and transmitters powers up to 2 x four zone call controllers NC894DKM Isolatable Monitoring Point NC803M/50 50 x magnetic reset keys Guest houses
device will flash to confirm the emergency call has been To help protect staff against aggressive visitors, distressed NC941B 0.3AHr battery backup kit Magnetic reset, no remote socket
patients or intruders, infrared ceiling receivers and call for use with 12V 250mA PSU ANCILLARY CALLING DEVICES Gymnasiums
registered and the relevant indicator panel light and its NC925B1 12V 1A white boxed PSU EMERGENCY CALL POINTS Generate standard call via a call point’s remote socket
buzzer will flash and sound a more insistent tone to points can be used in conjunction with infrared Fitness suites
powers up to 3 x four zone call controllers Generate standard and/or emergency calls (as shown) NC805C/6 1.8m (6ft) tail call lead
attract attention. Relevant overdoor lights, remote transmitters to remotely trigger standard and/or
AC284 2.1AHr backup battery NC802DEB-1/2 Emergency call point NC805C/14 4.2m (14ft) tail call lead
sounders and area indicators will also pulse (as fitted) to emergency calls. Up to three slave ceiling receivers can for use with 12V 1A boxed PSU Button reset, with remote socket, Solariums
be connected to one master ceiling receiver or infrared can generate standard & emergency calls NC805D 1.2-3.6m (4-12ft) tail call lead
provide additional audible and visual indication of the 10-30 ZONE EMERGENCY INDICATOR PANELS Police cells
emergency call. call point to provide additional coverage in large or L NC802DEM Emergency call point NC805P Hand/foot pneumatic pad
shaped rooms. Infrared transmitters typically provide a Indicate standard and emergency calls Magnetic reset, with remote socket requires 1 x NC805AS
10m line-of-sight transmitting range. NC812KE 10 Zone Emergency master panel can generate standard & emergency calls NC805AS Remote air switch for NC805P Prisons
WHICH SYSTEM SHOULD YOU CHOOSE? c/w 12V PSU, backup battery & mute button NC802DEWM Master water resistant alert point NC805MD Portable movement detector
Overdoor lights NC811KE 10 Zone Emergency repeater panel magnetic reset, no remote socket, can generate Factories
The type of system required will depend upon the Overdoor lights (if fitted) are normally installed outside standard or emergency calls (not both) NC805DSP Double remote socket
NC822KE 20 Zone Emergency master panel
application and the needs of the user so always check with rooms to increase call indication levels. If required, up to c/w 12V PSU, backup battery & mute button
allows two ancillary calling devices to Warehouses
NC802DEWS Slave water resistant alert point be used at the same time
the approving authority/client before work commences. four zones can be connected to one overdoor light to NC821KE 20 Zone Emergency repeater panel Max. 10 per master
MISCELLANEOUS 800 SERIES DEVICES Offices
In residential care homes, a standard system is normally provide area indication (in corridors, etc) with additional NC832KE 30 Zone Emergency master panel NC802DESS IP69K sauna alert point
acceptable as the main requirement is for the reset zones connectable via one or more input expanders. c/w 12V PSU, backup battery & mute button Must be used as a slave to an NC802DEWM NC899KD Area isolator Reception areas
can be used to isolate areas/zones
function to be carried out at the source of the call. In NC831KE 30 Zone Emergency repeater panel
Remote sounders NC802DEM/SS Stainless steel call point
nursing homes, however, some authorities insist on Up to three switchable or remote sounders can be Larger panels available to order magnetic reset, no remote socket, can generate
NC884DE Emergency 5:1 input expander GP surgeries
can be used to provide area/zonal indication
standard and emergency call facilities so an emergency standard or emergency calls (not both)
connected to an indicator panel’s sounder output to 10-30 ZONE STANDARD INDICATOR PANELS NC885 Standard 5:1 input expander Government buildings
system may be required. In most other applications, such NC804DE Emergency call / reset point
increase call levels. Indicate standard calls ONLY Button reset, no remote socket, can be used to provide area/zonal indication
as shops, leisure centres, hotels and public sector Relay units NC810K 10 Zone Standard master panel can generate emergency calls only NC889 Call latch module Interview rooms
buildings, the type of system and devices required will External strobes, sounders, beacons, etc, can be connected c/w 12V PSU & reset button NC809DEM Emergency call / reset point can trigger a standard call from a normally
depend almost entirely on what the user wants to NC811K 10 Zone Standard repeater panel Magnetic reset, no remote socket, open or normally closed switch Housing offices
to any zone or group of zones via a 12V relay unit. can generate emergency calls only
NC889ERM Multi-purpose call latch module
achieve. In most instances, referring to the typical NC820K 20 Zone Standard master panel Restaurants
applications diagram overleaf will help all interested Tone only paging c/w 12V PSU & reset button
INFRARED CALL POINTS & CEILING RECEIVERS
can trigger a standard and/or emergency call
from a normally open switch (includes
parties come to a decision on what components should A simple ‘tone-only’ paging option is available (requires NC821K 20 Zone Standard repeater panel confidence LED and onboard magnetic reset) Changing rooms
Generate standard and/or emergency calls when
be purchased. a DP874B/4 UHF paging transmitter). When a call is NC830K 30 Zone Standard master panel triggered by a compatible infrared transmitter NC883D 12V relay unit
made, tone only pagers will beep, prompting staff to c/w 12V PSU & reset button
NC302RXC Infrared master ceiling receiver can be used to switch sirens, strobes, etc. and much, much more
visit the nearest call indicator panel to determine the NC831K 30 Zone Standard repeater panel No on-board reset, no remote socket
PLANNING A SYSTEM exact source and nature of the call. Larger panels available to order NC302RXCS Infrared slave ceiling receiver
NC848/50 50 x Anti-tamper screws
can be used to secure stainless steel devices
The following points should be considered when Up to 3 per master. in high risk environments.
Master reset buttons NC802DERB Infrared call point
designing or planning an 800 Series Call System. Some indicator panels have an optional master reset Button reset, with remote socket
DP874B/4 Tone only UHF paging transmitter
supplied with one tone-only pager
Indicator panels button. If enabled, this facility allows standard calls to NC802DERM Infrared call point
Magnetic reset, with remote socket DP875/4 Extra tone-only pager
Indicator panels should be sited internally in a clean dry be globally cancelled without the source of the call
area which is readily accessible to staff. Various types being investigated. As there is a definite requirement Distributed by:
and sizes are available covering 1 to 90 zones. Repeater in most applications for the caller to be visited before
panels are also available. the call is reset, this facility is normally only utilised in
carefully managed staff-to-staff call systems. In
Call Points
most other instances we recommend this facility is
Any number of call points (of any kind), reset points and disabled.
overdoor lights can be wired onto one zone.
Mute buttons C-TEC
Standard call points (including ceiling pulls and All multi-zone emergency indicator panels are supplied
monitoring points) will work on 800 Series standard and with an optional mute button. If enabled, this allows Stephens Way, Wigan,
emergency systems but emergency call points will work any standard calls on the system to be silenced but WN3 6PH. United Kingdom
on emergency systems only. leaves relevant indicator light(s) lit. Although a new call Tel: +44 (0)1942 322744
Some emergency call points can generate standard and from a different call point will re-activate the sounders, Fax: +44 (0)1942 829867
emergency calls whereas others, such as water resistant to ensure callers are visited we recommend this facility is
alert points and stainless steel call points, can be set up used only in very carefully managed systems. Emergency Email: sales@c-tec.co.uk Quality System Certificate No. 176
to generate standard or emergency calls (not both). calls cannot be muted. Web: www.c-tec.co.uk Assessed to ISO9001 : 2008

Approved Document. No. DNS8002255 Rev 4. The manufacturer operates a policy of continuous improvement and reserves the right to alter product specifications at its discretion and without prior notice. Errors and omissions excepted.
No responsibility can be accepted by the manufacturer or distributor for any misinterpretation of an instruction or guidance note, or for the compliance of the system as a whole.
800 Series Part Numbers & Product Descriptions
HOW DOES AN 800 SERIES CALL SYSTEM WORK? Two methods of reset are available - button or magnetic
key. Reset buttons are fine for non-secure applications DISABLED PERSONS TOILET ALARM KITS OVERDOOR LIGHTS AND SOUNDERS

800 SERIES
INFRARED TRANSMITTERS
Depending on the type of calling devices used, the Can be used to increase call indication levels
system can be set up to operate as a standard system – on but reset keys are usually preferred in care NC951 Standard disabled WC alarm kit For use with infrared call points & ceiling receivers
establishments as they help prevent patients c/w NC942B call controller; NC807C ceiling pull,; NC806C Overdoor light QT412RXA Infrared transmitter
which standard (constant tone) calls are generated – or NC806CS overdoor light & sounder; NC809DBB NC806CS Overdoor light & sounder
inadvertently cancelling their own calls. Push/pull for emergency call
an emergency system – on which standard (constant tone) button reset point c/w sounder; WC sticker
NC806C/SS Overdoor light, s/steel (requires a QT424/1 or QT424/10 for
and emergency (intermittent tone) calls are generated. Call points without an on-board reset facility, such as NC951/SS Standard disabled WC alarm kit, s/steel NC806CS/SS Overdoor light & sounder, s/steel charging)

call system
ceiling pulls, must be reset via a reset point or call point c/w NC941/SS call controller; NC807C ceiling pull,
NC887D Switchable sounder QT412RXCA Infrared transmitter
On standard systems: NC806CS/SS overdoor light & sounder; NC809DB/SS
with an on-board reset facility. NC888D Remote sounder Push for standard call / pull for emergency
The user calls for assistance by activating a ceiling pull, button reset point; WC sticker
(requires a QT424/1 or QT424/10 for charging)
standard call point, monitoring point, water resistant A full list of standard and emergency call points and the
SINGLE ZONE EMERGENCY INDICATOR PANELS STANDARD CALL POINTS QT424/1 Single-way charging unit
alert point or infrared ceiling receiver (as fitted). A red facilities they offer can be found at the back of this leaflet. Indicate standard and emergency calls for use with QT412 range transmitters
Generate standard calls only
confidence light on the calling device illuminates to Ancillary calling devices NC942 Single zone call controller NC807C Ceiling pull unit QT424/10 10-way charging unit
reassure the user that the call has been registered. A tail call lead or hand/foot operated pneumatic pad can c/w 12V 140mA PSU, reset button & relay No on-board reset, no remote socket for use with QT412 range transmitters
At the system’s indicator panel(s), an indicator light be connected to a call point’s remote socket (if fitted) to NC942B Single zone call controller NC802DB Standard call point QT432 Patient neck pendant
generate a standard call. For safety reasons, a standard c/w 12V 140mA PSU, reset button, relay and Button reset, with remote socket
illuminates and a buzzer sounds to inform staff that onboard rechargeable battery backup supply NC802DM Standard call point
c/w anti-ligature lanyard, push for standard

someone is in need of assistance. Relevant overdoor call will also be made if the ancillary calling device is call, battery operated, requires 12V A23/23A
Ideal for use in:
NC941 Single zone call controller Magnetic reset, with remote socket (8LR50) alkaline battery (supplied)
lights, remote sounders and area indicators also operate inadvertently pulled from its socket. c/w 12V 250mA PSU & reset button NC917L Standard call push with protruding button
(as fitted) to provide additional audible and visual Monitoring points NC941/SS Single zone call controller, s/steel No on-board reset, no remote socket RESET POINTS Hotels
indication of the call. Monitoring points can be used to trigger a standard call if c/w 12V 250mA PSU & reset button NC802DEWM Master water resistant alert point NC809DB Button reset point
Leisure centres
NC941B 0.3AHr battery backup kit magnetic reset, no remote socket
A member of staff visits the calling room to assist the a fire exit, drug cupboard or storeroom is opened and can NC809DBB Button reset point c/w sounder
for use with NC941 & NC941/SS call controller NC802DEWS Slave water resistant alert point
user and resets the calling device to return the system to be isolated via an on-board keyswitch if required. They Up to 10 per master
Label also includes Braille text Nursing homes
normal. can also be connected to any device with a normally open 4-12 ZONE EMERGENCY INDICATOR PANELS NC802DESS IP69K sauna alert point NC809DB/SS Button reset point, s/steel
or normally closed switch, such as a pressure mat, allowing Indicate standard and emergency calls Must be used as a slave to an NC802DEWM NC809DM Magnetic reset point Care homes
On emergency systems: it to operate as a standard call point when activated. NC944 Four zone call controller NC802DEM/SS Stainless steel call point NC899IRR Keyswitch reset point
In addition to doing everything a standard system does, c/w mute/reset button, requires 12V PSU Magnetic reset, no remote socket Hospitals
a more urgent / extra level of assistance can be Call latch modules NC944/SS Four zone call controller, s/steel MAGNETIC RESET KEYS
summoned by activating an ‘emergency call’ via an Call latch modules operate in a similar way to monitoring c/w mute/reset button, requires 12V PSU STANDARD MONITORING POINTS For use with all magnetic call points, monitoring Shops
emergency call point, water resistant alert point or points but are smaller (for discreet mounting) and do not NC930 12V 250mA PSU Generate standard calls only points and reset points
include an onboard reset or isolating facility. powers up to 2 x four zone call controllers NC894DKB Isolatable monitoring Point Lavatories
infrared ceiling receiver (as fitted). NC803M/10 10 x magnetic reset keys
NC930/SS 12V 250mA PSU, s/steel Button reset, no remote socket
Once activated, the red confidence light on the calling Infrared call points, ceiling receivers and transmitters powers up to 2 x four zone call controllers NC894DKM Isolatable Monitoring Point NC803M/50 50 x magnetic reset keys Guest houses
device will flash to confirm the emergency call has been To help protect staff against aggressive visitors, distressed NC941B 0.3AHr battery backup kit Magnetic reset, no remote socket
patients or intruders, infrared ceiling receivers and call for use with 12V 250mA PSU ANCILLARY CALLING DEVICES Gymnasiums
registered and the relevant indicator panel light and its NC925B1 12V 1A white boxed PSU EMERGENCY CALL POINTS Generate standard call via a call point’s remote socket
buzzer will flash and sound a more insistent tone to points can be used in conjunction with infrared Fitness suites
powers up to 3 x four zone call controllers Generate standard and/or emergency calls (as shown) NC805C/6 1.8m (6ft) tail call lead
attract attention. Relevant overdoor lights, remote transmitters to remotely trigger standard and/or
AC284 2.1AHr backup battery NC802DEB-1/2 Emergency call point NC805C/14 4.2m (14ft) tail call lead
sounders and area indicators will also pulse (as fitted) to emergency calls. Up to three slave ceiling receivers can for use with 12V 1A boxed PSU Button reset, with remote socket, Solariums
be connected to one master ceiling receiver or infrared can generate standard & emergency calls NC805D 1.2-3.6m (4-12ft) tail call lead
provide additional audible and visual indication of the 10-30 ZONE EMERGENCY INDICATOR PANELS Police cells
emergency call. call point to provide additional coverage in large or L NC802DEM Emergency call point NC805P Hand/foot pneumatic pad
shaped rooms. Infrared transmitters typically provide a Indicate standard and emergency calls Magnetic reset, with remote socket requires 1 x NC805AS
10m line-of-sight transmitting range. NC812KE 10 Zone Emergency master panel can generate standard & emergency calls NC805AS Remote air switch for NC805P Prisons
WHICH SYSTEM SHOULD YOU CHOOSE? c/w 12V PSU, backup battery & mute button NC802DEWM Master water resistant alert point NC805MD Portable movement detector
Overdoor lights NC811KE 10 Zone Emergency repeater panel magnetic reset, no remote socket, can generate Factories
The type of system required will depend upon the Overdoor lights (if fitted) are normally installed outside standard or emergency calls (not both) NC805DSP Double remote socket
NC822KE 20 Zone Emergency master panel
application and the needs of the user so always check with rooms to increase call indication levels. If required, up to c/w 12V PSU, backup battery & mute button
allows two ancillary calling devices to Warehouses
NC802DEWS Slave water resistant alert point be used at the same time
the approving authority/client before work commences. four zones can be connected to one overdoor light to NC821KE 20 Zone Emergency repeater panel Max. 10 per master
MISCELLANEOUS 800 SERIES DEVICES Offices
In residential care homes, a standard system is normally provide area indication (in corridors, etc) with additional NC832KE 30 Zone Emergency master panel NC802DESS IP69K sauna alert point
acceptable as the main requirement is for the reset zones connectable via one or more input expanders. c/w 12V PSU, backup battery & mute button Must be used as a slave to an NC802DEWM NC899KD Area isolator Reception areas
can be used to isolate areas/zones
function to be carried out at the source of the call. In NC831KE 30 Zone Emergency repeater panel
Remote sounders NC802DEM/SS Stainless steel call point
nursing homes, however, some authorities insist on Up to three switchable or remote sounders can be Larger panels available to order magnetic reset, no remote socket, can generate
NC884DE Emergency 5:1 input expander GP surgeries
can be used to provide area/zonal indication
standard and emergency call facilities so an emergency standard or emergency calls (not both)
connected to an indicator panel’s sounder output to 10-30 ZONE STANDARD INDICATOR PANELS NC885 Standard 5:1 input expander Government buildings
system may be required. In most other applications, such NC804DE Emergency call / reset point
increase call levels. Indicate standard calls ONLY Button reset, no remote socket, can be used to provide area/zonal indication
as shops, leisure centres, hotels and public sector Relay units NC810K 10 Zone Standard master panel can generate emergency calls only NC889 Call latch module Interview rooms
buildings, the type of system and devices required will External strobes, sounders, beacons, etc, can be connected c/w 12V PSU & reset button NC809DEM Emergency call / reset point can trigger a standard call from a normally
depend almost entirely on what the user wants to NC811K 10 Zone Standard repeater panel Magnetic reset, no remote socket, open or normally closed switch Housing offices
to any zone or group of zones via a 12V relay unit. can generate emergency calls only
NC889ERM Multi-purpose call latch module
achieve. In most instances, referring to the typical NC820K 20 Zone Standard master panel Restaurants
applications diagram overleaf will help all interested Tone only paging c/w 12V PSU & reset button
INFRARED CALL POINTS & CEILING RECEIVERS
can trigger a standard and/or emergency call
from a normally open switch (includes
parties come to a decision on what components should A simple ‘tone-only’ paging option is available (requires NC821K 20 Zone Standard repeater panel confidence LED and onboard magnetic reset) Changing rooms
Generate standard and/or emergency calls when
be purchased. a DP874B/4 UHF paging transmitter). When a call is NC830K 30 Zone Standard master panel triggered by a compatible infrared transmitter NC883D 12V relay unit
made, tone only pagers will beep, prompting staff to c/w 12V PSU & reset button
NC302RXC Infrared master ceiling receiver can be used to switch sirens, strobes, etc. and much, much more
visit the nearest call indicator panel to determine the NC831K 30 Zone Standard repeater panel No on-board reset, no remote socket
PLANNING A SYSTEM exact source and nature of the call. Larger panels available to order NC302RXCS Infrared slave ceiling receiver
NC848/50 50 x Anti-tamper screws
can be used to secure stainless steel devices
The following points should be considered when Up to 3 per master. in high risk environments.
Master reset buttons NC802DERB Infrared call point
designing or planning an 800 Series Call System. Some indicator panels have an optional master reset Button reset, with remote socket
DP874B/4 Tone only UHF paging transmitter
supplied with one tone-only pager
Indicator panels button. If enabled, this facility allows standard calls to NC802DERM Infrared call point
Magnetic reset, with remote socket DP875/4 Extra tone-only pager
Indicator panels should be sited internally in a clean dry be globally cancelled without the source of the call
area which is readily accessible to staff. Various types being investigated. As there is a definite requirement Distributed by:
and sizes are available covering 1 to 90 zones. Repeater in most applications for the caller to be visited before
panels are also available. the call is reset, this facility is normally only utilised in
carefully managed staff-to-staff call systems. In
Call Points
most other instances we recommend this facility is
Any number of call points (of any kind), reset points and disabled.
overdoor lights can be wired onto one zone.
Mute buttons C-TEC
Standard call points (including ceiling pulls and All multi-zone emergency indicator panels are supplied
monitoring points) will work on 800 Series standard and with an optional mute button. If enabled, this allows Stephens Way, Wigan,
emergency systems but emergency call points will work any standard calls on the system to be silenced but WN3 6PH. United Kingdom
on emergency systems only. leaves relevant indicator light(s) lit. Although a new call Tel: +44 (0)1942 322744
Some emergency call points can generate standard and from a different call point will re-activate the sounders, Fax: +44 (0)1942 829867
emergency calls whereas others, such as water resistant to ensure callers are visited we recommend this facility is
alert points and stainless steel call points, can be set up used only in very carefully managed systems. Emergency Email: sales@c-tec.co.uk Quality System Certificate No. 176
to generate standard or emergency calls (not both). calls cannot be muted. Web: www.c-tec.co.uk Assessed to ISO9001 : 2008

Approved Document. No. DNS8002255 Rev 4. The manufacturer operates a policy of continuous improvement and reserves the right to alter product specifications at its discretion and without prior notice. Errors and omissions excepted.
No responsibility can be accepted by the manufacturer or distributor for any misinterpretation of an instruction or guidance note, or for the compliance of the system as a whole.