Service Manual
Table of Contents
Service Manual for RESmart BPAP V1.0
1. Introduction .............................................................................................................................................................1
1.1 Intended Use ...................................................................................................................................................1
IMPORTANT! ......................................................................................................................................................1
I.2 Implements ......................................................................................................................................................1
2. Structure ..................................................................................................................................................................1
2.1 System Features ..............................................................................................................................................1
2.2 Internal Structure ............................................................................................................................................2
2.3 Spare Parts ......................................................................................................................................................2
3. Software ..................................................................................................................................................................2
3.1 Upgrading Device Firmware ..........................................................................................................................2
3.2 Hidden Function .............................................................................................................................................3
4. Malfunction and Countermeasure .......................................................................................................................4
4.1 Error Code List for RESmart ........................................................................................................................4
4.2 Malfunction Determination.............................................................................................................................6
Appendices .................................................................................................................................................................7
Appendix A: ..........................................................................................................................................................7
Device Configuration .................................................................................................................................7
1. Main device.......................................................................................................................................7
2. Humidifier.........................................................................................................................................9
Components Illumination ........................................................................................................................ 11
1. Device Inside .................................................................................................................................. 11
2. Device Outside (All BPAP machine is grey, CPAP is white) .......................................................13
3. Humidifier.......................................................................................................................................14
Service Manual for RESmart BPAP V1.0
IMPORTANT
All data in this manual have been verified correctly. It is sufficient for servicing and repairing
this device. If applying this manual on other purpose, the validation should be confirmed by
BMC Medical Co., Ltd. Otherwise, BMC Medical Co., Ltd. has no responsibility for the result.
All information in this manual is protected by law and regulation. All copyrights belong to
BMC Medical Co., Ltd.
IMPORTANT, CAUTIONS and WARNINGS in this manual are to emphasize dangers to service
people.
Safety Notice
Electric Shock
There is high voltage over 100VAC inside this device, please be sure to repair the device after power
off.
Chemical Safety
There may be risk of harmful bacteria and viruses after using by patient, please clean the device or
wear protective glove before servicing and repairing.
Please wash your hands with disinfectant after operating the device.
Please deal with the waste according to your local regulation.
Service Manual for RESmart BPAP V1.0
1. Introduction
1.1 Intended Use
The RESmart BPAP 20T/25T/30T system is a Bi-level Positive Airway Pressure device, which is
intended to provide non-invasive ventilation for patients with respiratory insufficiency or obstructive
sleep apnea (OSA), either in the hospital or at home.
The RESmart BPAP 20T/25T/30T is to be used only on the instruction of a licensed health care
professional. Your home care provider will make the correct pressure settings according to your
health care professional’s prescription.
Several accessories are available to make your treatment with the RESmart BPAP 20T/25T/30T
as convenient and comfortable as possible. To ensure that you receive the safe, effective therapy
prescribed for you, use only BMC accessories.
The RESmart BPAP 20T/25T/30T is only used for single patient and must not be re-used on another
person. This is to avoid the risk of cross-infection.
IMPORTANT!
Read and understand the entire user manual before operating this system. If you have any
questions concerning the use of this system, contact your home care provider or health care
professional.
I.2 Implements
1. Implements for repairing
Type ‘+’ screwdriver
Pincers
2. Implements for measuring
Pressure meter
2. Structure
2.1 System Features
Handle
Display Screen
User Buttons
Humidifier Controller
Air Outlet
Humidifier Power
1/14
Service Manual for RESmart BPAP V1.0
Medical Product
Note (bottom)
Indicator
Outlet Port Light
Heater
Plate
Fill Line
Water Chamber
2.2 Internal Structure
See Appendix A
3. Software
3.1 Upgrading Device Firmware
Note: Once the RESmart BPAP device is powered on, the display screen will show the current
firmware version, the below picture for example, in the red frame is the firmware version.
2/14
Service Manual for RESmart BPAP V1.0
6. Press and hold + buttons together as below, in the meantime, power on (connect the
power) the device. Note: Please do not release the buttons until the upgrade process finished.
3/14
Service Manual for RESmart BPAP V1.0
NOTE
If the device experiences the error code, it is basically caused by the malfunction of the blower unit
or the Main PCBA, so the repair principle is to first replace the most possible malfunction part, such
as the Main PCBA,. However, if the problem is still there, place the Main PCBA, back, and then replace
the blower unit, the problem should be solved. If you still cannot solve it, please feel free to contact
us.
E-mail: zhangjianwei@bmc-medical.com
Error 01: Overtemperature: Main PCBA temperature is too high, or the Main PCBA temperature
sensor is damaged.
If the roomtemperature is not above 35°C, please replace the Main PCBA.
Error 02: The blower does not run: Malfunction of the blower, Main PCBA or pressure sensor results
in closed-loop control error.
Please replace the blower unit.
Error 03: The blower runs abnormally slow: Malfunction of the blower, Main PCBA or pressure sensor
results in closed-loop control error.
Please replace the blower unit first. if the problem cannot be solved,replace the blower unit back, and
then replace the Main PCBA.
Error 07: The drift of zero point for the pressure sensor is out of limitation: It may be the parameters
error or the pressure sensor is damaged.
Please replace the Main PCBA.
Error 08: Firmware version matching error: Used a wrong (non-matched) firmware version when
upgrading the device.
Please select the right upgrade firmware to conduct the upgrade again, such as 1.90-TN.upd
1.9E.04-TN.upd. The suffix of the firmware version is the same.
Error 09: Device parameters error: The device has no production process parameters or CPU error.
Please replace the Main PCBA.
Error 12: Overtemperature: Flow sensor temperature is too high, or the flow sensor is damaged.
If the room temperature is not above 35°C, please replace the Main PCBA.
4/14
Service Manual for RESmart BPAP V1.0
Error 13: Device temperature is lower than 0°C, or the digital pressure sensor fails.
If the room temperature is not below 0°C, please replace the Main PCBA.
Error 14: Overtemperature: Blower temperature is too high, or the blower temperature sensor is
damaged.
Please replace the blower unit.
Error 16: Lack of hardware parameters: The production process was not finished; the production
firmware should be replaced.
Please replace the Main PCBA.
Error 17: Firmware error for BPAP: Wrongly wrote firmware of BPAP 25T/30T into BPAP 25/25A.
Please select the right upgrade firmware, and conduct the upgrade again.
Error 18: Firmware error for BPAP: Wrongly wrote firmware of BPAP 25/25A into BPAP 25T/30T.
Please select the right upgrade firmware, and conduct the upgrade again.
Error 19, Error 20: Blower drive circuit experiences overtemperature, or blower temperature sensor
is damaged.
Please replace the blower unit first. If the problem cannot be solved,place the blower unit back, and
then replace the Main PCBA.
Error 21: Auto CPAP, BPAP: The Main PCBA temperature sensor and flow sensor both experience
over temperature or are damaged.
CPAP: The Main PCBA temperature sensor and power tube thermistor both experience
overtemperature.
Please replace the Main PCBA first. If the problem cannot be solved,place the Main PCBA back, and
then replace the blower unit.
Error 74: Validation error for the blower data returned from the serial port.
Please replace the Main PCBA.
5/14
Service Manual for RESmart BPAP V1.0
Malfunction
Can Start Up
Power Module
Power Supply
Humidifier Self-Check
Others
No Heat
E E E E E E E E E
Replace Main board
Low E
Temperture r r r r r r r r r
r
r r r r r r r r r
r
N o o o r o r o o o
Check AC
o
Power
r r r r r r r r r Stop
r
1 2 3 4 7 8 9 13 14 After
Humidifier PCBA
12
Water Tank
Start Up
Replace
Replace
Replace
Replace
E
r
r
Refer error o
Repair Replace code table r
15
End
6/14
Service Manual for RESmart BPAP V1.0
Appendices
Appendix A:
Device Configuration
1. Main device
Fig. A - Main board (backside) Fig. B – Outlet (left side)
3 2
Hole #1 in Fig. B connect to sensor YL in Fig. A via silicon rubber canal; Hole #2 in Fig. B connect to
sensor LL 1 in Fig. A via silicon rubber canal; Hole #3 in Fig. B connect to sensor LL 2 in Fig. A via
silicon rubber canal;
Specification of canal
Fig. B Fig. A Applicable Model
φ2mm, Φ4mm, L12cm
Hole #1 Sensor YL CPAP/Auto CPAP
φ2mm, Φ4mm, L10cm
Hole #2 Sensor LL 1 Auto CPAP
φ2mm, Φ4mm, L10cm
Hole #3 Sensor LL 2 Auto CPAP
Table 1
7/14
Service Manual for RESmart BPAP V1.0
8/14
Service Manual for RESmart BPAP V1.0
Connecting relationship:
Part 1 2 3 4
Table 2
2. Humidifier
Fig. G – Humidifier PCBA (back side)
P1
P2
9/14
Service Manual for RESmart BPAP V1.0
Left Right
Black Orange
3-pin Plug
Blue
Blue
Gray
Gray
Orange
Orange
Black
Black
Heater PCBA
Connecting relationship:
Part
1 2 3 4
Fig.
I Fig. G: P2 Fig. G: P1 Fig. H: Left Fig. H: Right
Table 3
10/14
Service Manual for RESmart BPAP V1.0
Components Illumination
1. Device Inside (All BPAP machine is grey shell, CPAP is white shell,Auto CPAP is black
shell)
11/14
Service Manual for RESmart BPAP V1.0
12/14
Service Manual for RESmart BPAP V1.0
2. Device Outside (All BPAP machine is grey shell, CPAP is white shell,Auto CPAP is black
shell)
13/14
Service Manual for RESmart BPAP V1.0
3. Humidifier
S/N Assembly Figure
14/14