You are on page 1of 7

ARTICLE IN PRESS

Current Problems in Diagnostic Radiology 000 (2019) 1 7

Current Problems in Diagnostic Radiology


journal homepage: www.cpdrjournal.com

Common and Uncommon Artifacts in T1 FLAIR SAG Sequences


of MRI Brain
Eleftherios Lavdas, PhDa,b, Maria Papaioannou, BScb, Nadia Boci, BSca,
Efthimios Dardiotis, MDc, Violeta Roka, MD, PhDd, Georgios Sakkas, Phde,
Glykeria Apostolopoulou, BScb, Lida Gogou, Phda, Panayiotis Mavroidis, PhD, DABRf,*
a
Department of Medical Radiological Technologists, Technological Education Institute of Athens, Athens, Greece
b
Department of Medical Imaging, Animus Kyanoys Larisas Hospital, Larissa, Greece
c
Department of Neurology, University Hospital of Larissa, Larissa, Greece
d
Health Center of Farkadona, Trikala, Greece
e
Department of Sport Science, University of Thessaly, Trikala, Greece
f
Department of Radiation Oncology, University of North Carolina, Chapel Hill, NC

A B S T R A C T

Objective: This study aims at identifying, classifying, and measuring the frequency the different artifacts that show up in the images of the Sagittal T1 Fluid Atten-
uated Inversion Recovery (FLAIR) sequence.
Materials and Methods: A total of 101 subjects underwent brain magnetic resonance imaging examination with the following sequences: Axial T1 FLAIR, Axial
T2-weighted imaging, Diffusion Weighted Imaging, 2D Multiple Echo Recombined Gradient Echo, Sagittal T1 FLAIR, Coronal T2 Turbo Spin Echo, Spin Echo T1-
weighted imaging, and 3D Fast Spoiled Gradient-echo. In these images, we observed the following categories of artifacts: (a) ghost artifacts, (b) aliasing behind
the occipital bone, (c) aliasing inside the sphenoid cavity, (d) susceptibility artifacts, and (e) pulsation artifacts. In order to recognize and verify the artifacts, we
used not only the Sagittal T1 FLAIR sequence, but also Sagittal reconstructions from the 3-dimensional Fast Spoiled Gradient-echo sequence and the other rou-
tine sequences.
Results: Aliasing artifacts and especially aliasing of nose are present in 41% of the cases. In 45% of these cases the uncommon aliasing artifacts, which took place
into the brain parenchyma (sphenoid cavity, subarachnoid bay, or pituitary) originated from nose. In 33% of the subjects, ghost artifacts are presented, which
stem from the nose, the orbits, or other pulsating structures (pulsation artifacts) or even from fat tissue. Moreover, susceptibility artifacts comprise 8% of all the
artifacts. Finally, 19% of brains were presented without artifact.
Conclusions: We suggest in addition to T1 FLAIR, the application of Sagittal SE or TSE sequences in magnetic resonance imaging examination of brain, trying to
include the nose in the square of FOV.
© 2019 Elsevier Inc. All rights reserved.

Introduction The fluid attenuated inversion recovery (FLAIR) technique obtains


T2-weighted images and also T1-weighted images with nulling of the
Magnetic Resonance Imaging (MRI) of the brain is widely used in a cerebrospinal fluid (CSF) signal.3-9 FLAIR is a sequence that attains
variety of medical applications. Those applications involve research, heavy T2 weighting with the use of a long TE. Specifically, T1 FLAIR is
diagnosis, treatment, surgical planning, and image guided surgeries. a sequence that attains T1 weighting with the use of a long TE.
The magnetic resonance brain images are often corrupted with vari- FLAIR imaging has proved to have an important clinical utility in
ous artifacts and may affect the image of the brain and lead to a several intracranial conditions.2-8,10-14 It has proved to be useful in
default of the diagnosis. head injury, multiple sclerosis, acute subarachnoid hemorrhage, car-
Specifically, an artifact is a feature appearing in an image that is bon monoxide poisoning, tuberous sclerosis, and mesial temporal
not present in the original subject. Those artifacts may be very notice- sclerosis.4,6,7,9-14 Moreover, the use of FLAIR sequences has been
able or just a few pixels out of balance but can lead to confusing arte- reported in the diagnosis of intracranial infectious diseases.
factual appearances with pathology that may be misdiagnosed. Many In addition, T1 FLAIR sequence presents some important advan-
different kind of artifacts may occur during brain imaging, some tages like great nulling of CSF, reduced pulsation and flow artifacts,
affecting the diagnostic quality, while others may be interpreted as very good contrast between white and gray matter and generally bet-
pathology.1,2 ter CNR, definite imaging of hemorrhages and dysplasia, and ability to
detect high lesions of brain.2,8,9
However, FLAIR’s usefulness has been reported to be limited in
*Reprint requests: Panayiotis Mavroidis PhD, DABR, Department of Radiation Oncol-
ogy, University of North Carolina, 101 Manning Dr, Chapel Hill, NC 27599-7512.
the diagnosis of intracranial tumors.10 Aliasing or Wrap Around arti-
E-mail address: panayiotis_mavroidis@med.unc.edu (P. Mavroidis). facts occur when the field of view (FOV) is smaller. This is an artifact

https://doi.org/10.1067/j.cpradiol.2019.08.001
0363-0188/© 2019 Elsevier Inc. All rights reserved.
ARTICLE IN PRESS
2 E. Lavdas et al. / Current Problems in Diagnostic Radiology 00 (2019) 1 7

that occurs when the region beyond the body part being imaged proj-
ects on the other side of the image.11 On the other hand, discrete
''ghost'' artifacts may occur along the phase-encode direction when-
ever the position or signal intensity of imaged structures within the
FOV vary or move in a regular (periodic) fashion. Pulsatile flow of
blood or CSF, cardiac motion, and respiratory motion are the most
important subject-related causes of ghost artifacts in clinical MRI.
The intensity of these ghost artifacts increases with the amplitude of
the periodic motion, as well as with the signal intensity of the moving
tissue. Susceptibility artifacts share distortions or local signal change
due to local magnetic field inhomogeneities from a variety of com-
pounds. They are especially encountered while imaging near metallic
orthopedic hardware or dental work.
The purpose of this study is to retrospectively evaluate the com-
mon and uncommon artifacts in T1 SAG sequence in order to avoid
pitfalls in medical findings in relation with MRI brain examinations.
Additionally, we compared T1 SAG FLAIR, 3-dimensional Fast Spoiled
Gradient Echo (3D FSPGR), and all the other routine sequences in
order to verify the presence of the artifacts and avoid misdiagnosis.

Methods

This is retrospective study, which included 101 consecutively


examined subjects (50 male, 51 female; mean age = 55 years; age
range = 17-81 years), who were referred for brain MRI. Of those sub-
jects, 9 were inpatients and 92 outpatients and only a few subjects
were excluded because their images were not diagnostic due to
motion and we could not evaluate the artifacts. For this cohort, we
recorded and classified the technical artifacts which were present in
the MRI scans and specifically in the T1 SAG FLAIR sequences during
those MRI brain examinations. This study was approved by the local
institutional review board and written informed consent was
obtained from all the subjects participating in the study protocol.
The examinations were performed on all the subjects using a 1.5 T
MRI scanner (GE Signa).The following sequences were performed in
every brain examination (Axial T1 FLAIR, Axial T2-weighted imaging,
diffusion-weighted imaging , 2D Multiple Echo Recombined Gradient
Echo, SAG T1 FLAIR, Coronal T2 Turbo Spin Echo, Spin Echo T1-
weighted imaging, and 3D FSPGR. Generally, we used only 1
sequence in sagittal plane in order to find another way to verify the
results.
For the artifacts’ verification we used the FSPGR sequence with 3D
reconstructions in the sagittal plane. In that way, we evaluated
FIG 1. (A) OBL 3D FSPGR; (B) SAG T1 FLAIR; (C) SAG T1 FLAIR. The images show aliasing
whether the artifacts in T1 SAG FLAIR sequence were nominal or not.
artifacts in the sphenoid sinus (arrows). Specifically, the image (A) proves that the
At the same time, we recorded the number and type of artifacts sphenoid cavity is clear without aliasing artifacts due to different dimension of phase
(ghost, aliasing behind occipital bone, aliasing inside the sphenoid encoding. The images (B) and (C) show aliasing artifacts. More specifically, the nose is
cavity, susceptibility, pulsation artifacts). The scan parameters that larger than 15 cm and the artifacts show up in the sphenoid cavity because there was
we used are presented in Table 1. not enough space to be presented behind the occipital bone.
Also, we recorded the anterior-posterior (A/P) dimension of every
brain because the aliasing artifact of the nose is presented in that
dimension. For that reason, we took the larger A/P dimension in the artifact of nose based on whether this was present behind the occipi-
level of nose (nose included) in order to modify the FOV to eliminate tal bone or inside the sphenoid cavity.
that artifact. Additionally, we grouped the subjects with the aliasing
Results

TABLE 1
The aliasing artifacts are presented in the 41% of the cases and in
Summary of the examined sequences that were applied during the MR examinations
the 80% of these cases the dimension of brain (A/P) was over 22 cm.
SAG T1 FLAIR* 3D FSPGR More specifically, it was observed that in cases that the dimension of
Thickness 5.0 1.2 the head is larger than 22 cm this artifact appears in the sphenoid
FOV 24 £ 24 24 £ 18 cavity and when it is smaller than 22 cm and there is enough space
TR 2278 7.8 the artifact shows up behind of occipital bone.
TE 25.9 3 In total, 33% of the subjects presented ghost artifacts from the
NEX 2.0 1.0
Matrix 320 £ 256 256 £ 256
nose, orbits, or other pulsating structures or even from fat tissue.
Phase A/P S/I Also, the frequency of the susceptibility artifacts was around 8% of all
rFOV 100% 75% the artifacts, and they were related to metallic matter like dental
* The inversion time that was used in T1 FLAIR was 750 ms. implants in the jawbone.
ARTICLE IN PRESS
E. Lavdas et al. / Current Problems in Diagnostic Radiology 00 (2019) 1 7 3

FIG 2. (A, B) SAG 3D FSPGR sequences; (C-E) Different series of T1 SAG FLAIR; (F,G) Different series of T2 Coronal Fast Spin Echo. The figures show aliasing in pituitary (arrows).

In addition, the aliasing artifact of nose was present in 41% of the increasing number of phase encoding steps but this would increase
cases and in all these subjects the A/P dimension of brain was over image acquisition time. For correction, a larger FOV may be cho-
22 cm. As a result, the nose did not appear behind the head but sen.2,15 Particularly, in our study, we used a larger A/P dimension in
showed up in the sphenoid cavity. Additionally, around 15% of the the level of nose with the nose included in order to modify the FOV
cases presented pulsation artifacts from the vascular structures par- and achieve elimination of the aliasing artifact (Fig 1).
ticularly in the middle-sagittal slices. As mentioned “ghost” artifacts (Fig. 3, 4, and 7) may occur along
Figures 1-8 show examples of the artifacts examined as well as the phase-encoding direction16 whenever the position or signal
images verifying that those findings are artifacts. intensity of imaged structures within the field-of-view vary or move
in a regular (periodic) fashion. These artifacts may be caused by arte-
Discussion rial pulsations, swallowing, breathing, peristalsis, and physical move-
ment of the subject. When projected over anatomy it can mimic
The aliasing artifacts shown in Figures 1, 2, 5, and 6 are a result of pathology, and needs to be recognized (Fig. 3, 4, and 7). In clinical
mismapping of anatomy that lies outside the FOV but within the slice brain MRI, pulsatile flow of blood or CSF or 1 of the above aforemen-
volume. The selected FOV is smaller than the size of the imaged tioned causes are the most important and usual subject-related
object. The anatomy is usually displaced to the opposite side of the causes of ghost artifacts. The intensity of these ghost artifacts
image. It can be caused by nonlinear gradients or by under sampling increases with the amplitude of the periodic motion such as vascular
of the frequencies contained within the return signal.14 The sampling pulsation, as well as with the signal intensity of the moving tissue.
rate must be twice the maximal frequency that occurs in the object The spacing between these ghosts is related to the repetition time
(Nyquist sampling limit). If not, the Fourier transform will assign very and the frequency of the motion. During our efforts to eliminate
low values to the frequency signals greater than the Nyquist limit. ghost artifacts which are caused by motion artifacts we used T2-fluid
These frequencies will then ‘wrap around’ to the opposite side of the FLAIR recovery sequences with different k-space trajectories (BLADE).
image (Fig. 1 and 2) showing up as low-frequency signals. In the fre- This process and results have been previously repoted.17
quency encoding direction a filter can be applied to the acquired sig- FLAIR is a pulse sequence with an inversion time that nulls the
nal to eliminate frequencies greater than the Nyquist frequency. In signal from CSF, providing heavy T2 weighting, due to its very long
the phase encoding direction, artifacts can be reduced by an TE.18-20 However, there is a limitation of FLAIR MRI is the ventricular
ARTICLE IN PRESS
4 E. Lavdas et al. / Current Problems in Diagnostic Radiology 00 (2019) 1 7

FIG 3. (A, B) SAG T1 FLAIR sequences. The figures show ghost artifacts because of the existence of fat (arrows).

FIG 4. (A) SAG 3D FSPGR sequence; (B) SAG T1 FLAIR sequence. The figures show ghost artifacts in medulla (arrows).

FIG 5. (A, B) Different series of SAG T1 FLAIR. The figures show calcifications from Venus sinuses (red arrows). Crista galli (white arrows). (Color version of figure is available online.)
ARTICLE IN PRESS
E. Lavdas et al. / Current Problems in Diagnostic Radiology 00 (2019) 1 7 5

FIG 6. (A-F) Different series of SAG T1 FLAIR. The aliasing of nose could be presented even in consecutive images (arrows).

CSF pulsation artifact (VCSFA). This artifact might compromise the multiparametric. Specifically, in the locations where CSF goes into
study of ventricular abnormalities by leading to false-negative or the brain sections, between the inversion pulse and the commence-
false-positive interpretations.21,22 There are many studies which try ment of signal sampling, it is not appropriately nulled, and remains
to explain the VCSFA. It seems that the causes of VCSFA are hyperintense on FLAIR images.18,23
ARTICLE IN PRESS
6 E. Lavdas et al. / Current Problems in Diagnostic Radiology 00 (2019) 1 7

FIG 7. (A) Images of SAG T1 FLAIR showing ghost artifacts in the subarachnoid bay (arrow). (B, C) Two consecutive images of SAG 3D FSPGR verifying that the subarachnoid bay is
clear.

Vascular pulsation artifacts are recognized by their alignment


with the responsible vessel along the phase encoding direction of the
image. These artifacts reproduce the cross-sectional size and shape of
the responsible vessel, but not necessarily its signal intensity.24-26
There are many studies, which have proposed technical variations to
reduce this artifact, such as, using a wider slice-selective inversion to
increase the inversion of CSF outside the imaging section, to use other
section selective inversion pulses.27 We may try to eliminate this
kind of artifacts with BLADE sequences.28
Susceptibility artifacts (Fig 8) occur as a result of the microscopic
gradients or variations in the magnetic field strength, which occur
near the interfaces of substances of different magnetic susceptibility.
Large susceptibility artifacts are commonly seen surrounding ferro-
magnetic objects inside diamagnetic materials (such as the human
body). These gradients cause dephasing of spins and frequency shifts
of the surrounding tissues. The net results are bright and dark areas
with spatial distortion of surrounding anatomy. Those artifacts can
be made less prominent by performing imaging at low magnetic field
strength, using smaller voxels, decreasing TE, and increasing receiver
bandwidth. We may eliminate those artifacts by decreasing time of
echo and increase receiver bandwidth.29 It should be mentioned that
Gradient-echo and echo-planar sequences should be avoided,
because they accentuate susceptibility artifacts. The use of spin-echo
and particularly fast spin-echo sequences should be considered. FIG 8. SAG T1 FLAIR sequence. Susceptibility artifacts in the oral cavity from possible
The magnetic resonance artifacts in brain cause image degrada- dental metal implants (white arrow) and ghost artifact in the same line. (red arrow).
tion, and can occasionally mimic pathological lesions. Particularly, (Color version of figure is available online.)
ARTICLE IN PRESS
E. Lavdas et al. / Current Problems in Diagnostic Radiology 00 (2019) 1 7 7

we consider that they are more distinct in the sagittal plane. The pur- 8. Sundarakumar D, Smith C, Hwang W, et al. Evaluation of focal cervical spinal cord
pose of this study is to learn how to recognize the artifacts (common lesions in multiple sclerosis: Comparison of white matter—suppressed T1 inver-
sion recovery sequence versus conventional STIR and proton density—weighted
and uncommon) and especially the aliasing that nose may cause, turbo spin-echo sequences. AJNR Am J Neuroradiol 2016;37:1561–6.
which are rarely described in the literature. At the beginning of this 9. Ganesan K, Bydder M. A prospective comparison study of fast T1 weighted fluid
study, a lot of time was consumed to verify the cause of the observed attenuation inversion recovery and T1 weighted turbo spin echo sequence at 3T in
degenerative disease of the cervical spine. Br J Radiol 2014;87:20140091.
artifacts. We believe that it would be very useful to inform about the 10. Tsuchiya K, Mizutani Y, Hachiya J. Preliminary evaluation of fluid-attenuated
existence and the reason of the appearance of those artifacts in order inversion-recovery MR in the diagnosis of intracranial tumors. AJNR Am J Neurora-
to avoid diagnostic pitfalls. We suggest comparing the images of SAG diol 1996;17:1081–6.
11. Ashikaga R, Araki Y, Ishida O. MR imaging of traumatic head injuries using FLAIR
FLAIR T1 with 3D reconstructions from FSPGR sequences which is technique [in Japanese]. Nippon Acta Radiol 1994;54:1184–6.
more reliable in relation with such artifacts (eg, small SNR, different 12. Hashemi RH, Bradley WG, Chen DY, et al. Suspected multiple sclerosis: MR imag-
dimension of phase encoding, superior/inferior direction). Three- ing with a thin-section fast FLAIR pulse sequence. Radiology 1995;196:505–10.
13. Noguchi K, Ogawa T, Inugami A, et al. Acute subarachnoid hemorrhage: MR imag-
Tesla (3.0 T) MRI gives the potential of improved image quality and
ing with fluid-attenuated inversion recovery pulse sequences. Radiology
decreased scan times.30-32 However, in 3 T different types of artifacts 1995;196:773–7.
are presented, too.30 Because in FLAIR sequences in 3 T the artifacts 14. Jack CR, Rydberg CH, Krecke KN, et al. Mesial temporal sclerosis: Diagnosis with
are more than the 1.5 T30 they may be studied retrospectively, fluid-attenuated inversion-recovery versus spin-echo MR imaging. Radiology
1996;199:367–73.
because we suppose that they may be more obvious and more often 15. Somasundaram K, Kalavathi P. Analysis of imaging artifacts in MR brain images.
appearing. In cases that we are not sure for the result, it could be use- Orient J Comp Sci Technol 2012;5:135–41.
ful to increase the FOV with parallel cost of resolution. Alternatively, 16. Krupa K, Bekiesin  ska-Figatowska M. Artifacts in magnetic resonance imaging. Pol J
Radiol 2015;80:93–106.
we could apply the technique phase over wrap with cost in time, 17. Lavdas E, Mavroidis P, Kostopoulos S. al.Improvement of image quality using
because the FOV is not square. BLADE sequences in brain MR imaging. Magn Reson Imaging 2013;31:189–200.
18. Tanaka N, Abe T, Kojima K, et al. Applicability and advantages of flow artifact-
insensitive fluid-attenuated inversion-recovery MR sequences for imaging the
Conclusions posterior fossa. AJNR Am J Neuroradiol 2000;21:1095–8.
19. Alibek S, Adamietz B, Cavallaro A, et al. Contrast-enhanced T1-weighted fluid-
attenuated inversion-recovery BLADE magnetic resonance imaging of the brain:
We suggest the application of SAG Spin Echo or Turbo Spin Echo An alternative to spin-echo technique for detection of brain lesions in the unse-
sequences in MRI examinations of brain in addition to the use of the dated pediatric patient? Acad Radiol 2008;15:986–95.
T1 FLAIR, which is the standard sequence in most practices. In those 20. Alkan O, Kizilkiliç O, Yildirim T, et al. Comparison of contrast-enhanced T1-
weighted FLAIR with BLADE, and spin echo T1-weighted sequences in intracranial
sequences, we should try to include the nose in the square of FOV. If
MRI. Diagn Interv Radiol 2009;15:75–80.
we do not like to change the dimension of phase encoding, we have 21. Bakshi R, Caruthers SD, Janardhan V, et al. Intraventricular CSF pulsation artifact on
to check very carefully the region of the sphenoid cavity and the par- fast fluid-attenuated inversion recovery MR images: Analysis of 100 consecutive
allel axis in order to insure that there are no aliasing artifacts, which normal studies. AJNR Am J Neuroradiol 2000;21:503–8.
22. Bakshi R, Kamran S, Kinkel PR, et al. Fluid-attenuated inversion recovery MR imag-
are projected in the FOV. ing in acute and subacute cerebral intraventricular hemorrhage. AJNR Am J Neuro-
radiol 1999;20:629–36.
23. Oatridge A, Curati WL, Herlihy AH, et al. Evaluation of a FLAIR sequence designed
Declaration of Competing Interest toreduce CSF and blood flow artifacts by useof k-space reordered by inversion
time at each slice position (KRISP) in high grade gliomas of the brain. J Comput
The authors have declared that no competing interests exist. Assist Tomogr 2001;25:251–6.
24. Barish MA, Jara H. Motion artifact control in body MR imaging. Magn Reson Imag-
ing Clin N Am 1999;7:289–301.
References 25. Hedley M, Yan H. Motion artifact suppression: A review of post-processing techni-
ques. Magn Reson Imaging 1992;10:627–35.
26. Huber ME, Hengesbach D, Botnar RM. Motion artifact reduction and vessel
1. Hamilton J, Franson D, Seiberlich N. Recent advances in parallel imaging for MRI. enhancement for free-breathing navigator-gated coronary MRA using 3D k-space
Prog Nucl Magn Reson Spectrosc 2017;101:71–95. reordering. Magn Reson Med 2001;45:645–52.
2. Mavroidis P, Giankou E, Tsikrika A, et al. Brain imaging: Comparison of T1W FLAIR 27. Tha KK, Terae S, Kudo K, et al. Differential diagnosis of hyperintensecerebrospinal
BLADE with conventional T1W SE. Magn Reson Imaging 2017;37:234–42. fluid on fluid-attenuated inversion recovery images of the brain. Part II: Non-path-
3. Erasmus LJ, Hurter D, Naude  M, et al. A short overview of MRI artifacts. S Afr J
ological conditions. Br J Radiol 2009;82:610–4.
Radiol 2004;8:13–7. 28. Lavdas E, Mavroidis P, Hatzigeorgiou V, et al. Elimination of motion and pulsation
4. Neema M, Guss Z, Stankiewicz M, et al. Normal findings on brain FLAIR MRI scans artifacts using BLADE sequences in knee MR imaging. Magn Reson Imaging
at 3 T. AJNR Am J Neuroradiol 2009;30:911–6. 2012;30:1099–110.
5. Saranathan M, Worters P, Rettmann D, et al. Physics for Clinicians: Fluid-attenu- 29. Mavroidis P, Boci N, Lavdas E, et al. Investigation of the possibility to reduce sus-
ated inversion recovery (FLAIR) and double inversion recovery (DIR) imaging. J ceptibility artifacts in MRI knee examination. J Med Biol Eng 2017;37:644–52.
Magn Reson Imaging 2017;46:1590–600. 30. Lavdas E, Tsougos I, Kogia S, et al. T2 FLAIR artifacts at 3-T brain magnetic reso-
6. Gabr R, Hasan K, Haque M, et al. Optimal combination of FLAIR and T2-weighted nance imaging. Clin Imaging 2014;38:85–90.
MRI for improved lesion contrast in multiple sclerosis. J Magn Reson Imaging 31. Phalke VV, Gujar S, Quint DJ. Comparison of 3.0 T versus 1.5 T MR: Imaging of the
2016;44:1293–300. spine. Neuroimaging Clin N Am 2006;16:241–8.
7. Wiggerrmann V, Hernadez Torres E, Traboulsee A, et al. FLAIR: A combination of 32. Ramnath RR. 3 T MR imaging of the musculoskeletal system (Part II): Clinical
FLAIR and T2 for improved MS lesion detection. AJNR Am J Neuroradiol applications. Magn Reson Imaging Clin N Am 2006;14:27–40.
2016;37:259–65.

You might also like