SAND2018-4454PE
E
u.s.N
oyireOF AIL M Alnk031 Sandia National Laboratories is a multimission laboratory managed and operated by National Technology and Engineering Solutions of Sandia, LLC, a wholly owned
Y Lk subsidiary of Honeywell International, inc., for the U.S. Department of Energys National Nuclear Security Administration under contract DE-NA0003525.
Sandia
Outline National
Laboratories
5. Future development
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 2
Sandia
• •
*KEY94GHT
+ 00000 'y' •
7 8 9 E
1441.
2 4t L
4 - I 5 L 6 !-
I 2 3
7- a
-7
3458A digital multimeters are used throughout the Sandia PSL and DOE
metrology labs to make precision electrical measurements.
• Calibrated measurement functions:
• Resistance: <1 viD to 1 GQ
• DC voltage: <1 pV to 1000 V
• DC current: <100 pA to 1 A
• AC voltaget: <10 pV to 1000 V, 1 Hz to 10 MHz
• AC currentt: <50 vtA to 1 A, 1 Hz to 100 kHz
• Frequency: 1 Hz to 10 MHz
• Manufactured by Hewlett Packard, Agilent, Keysight
• Measurements can be manual or automated (with LabView, Python, MetCal, etc.)
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 3
Sandia
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 4
Sandia
MMS 2300
• Physical equipment contained in one rack
• Measurements automated with MetCal procedure
Standards
• Resistors (decades from 100 mQ to 1 GQ)
• Zener voltage standard
• Digital multimeter
• Calibrator
• AC measurement standard
• Signal generator
Other equipment
• Reference voltage divider (self-calibrating)
• Computer running MetCal
• Low thermal scanner for resistors
• Thermometer
,,,,,,,,,,
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 5
Sandia
2 2 2
System uncertainty (F X std dev(x) +(1 UUT_RES)
Uexp
C nfidence interval + )
Measurement
Traceability to SI
Function
Resistance Calibrated resistors
DC Voltage Zener voltage standard; NVLAP-accredited calibrated 3458A
DC Current Calibrated resistors; NVLAP-accredited calibrated 3458A
AC Voltage NVLAP-accredited calibrated Fluke 5730A calibrator & 5790B AC mmt standard
AC Current NVLAP-accredited calibrated Fluke 5730A calibrator
Frequency Calibrated signal generator
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 6
Sande
Resistance laboratwies
MI 4220
3458A DUT
low thermaJ
(R range) Traceability to SI:
scanner
Resistors calibrated at PSL against
L&N 4010 primary 1 ohm resistors
,t441itrw-, '14
1 1Gcl
Calibrated at NIST against
Quantized Hall Resistance
(intrinsic standard)
r Zen
I
reference
Traceability to SI:
752A 732B Zener calibrated at PSL against
input standard
Reference Zener voltage Josephson Voltage Array
divider standard (intrinsic standard)
Fluke 5730
null meter
calibrator
3458A Std NVLAP-accredited CCL or
(null meter) PSL calibration nhf
V=
2e
Equipment Diagram of Calibrator Output Voltage
Characterization.
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 8
Sandia
DC Current National
Laboratories
NVLAP-accredited CCL or
PSL calibration
Time-of-test voltage
characterization 3458A DLIT Traceability to SI:
(DCI range) Resistor calibrated at PSL against
Fluke 5730 Rstd L&N 4010 primary 1 ohm resistors
calibrator
‘1,.....
h
RHall =
ne 2
System uncertainty: 1-yr specs for standard 3458A measuring voltage drop across standard resistor
RSS (Resistor's certified ue,p, 3458A's 1-year specs for voltage and range used (100 mV or 1 V))
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 9
Sandia
AC Voltage National
Laboratories
___ NVLAP-accredited
5790B AC CCL calibration
measurement
standard
Fluke 5730
1
NVLAP-accredited
3458A DUT
(ACV range)
calibrator
CCL calibration
System uncertainty: 5790B measurement standard specs OR 5730 calibrator's wideband specs
• Up to 1 MHz: 5790E3's 1-year specs for ppm of reading + floor
• Over 1 MHz: 5730A's 1-year specs for ppm of voltage output + floor
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 10
Sandia
AC Current National
Laboratories
3458A DUT
(AO Ira nge)
Fluke 5730 NVLAP-accredited
calibrator CCL calibration
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 11
Sandia
Frequency National
Laboratories
PSL calibration
33250
345BA DUT
signa1
(frequency range)
generator
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 12
Sandia
MMS 2300 has been used to calibrate DC Lab's 3458A's. The DC Lab will begin
to calibrate 3458A's belonging to other PSL Labs and the wider DOE complex.
THANKS:
• Gary Smith and Jim Novak in DC Lab (design, development, software)
• Dave Hofmann and Selina Pedroza (electrical hazard analysis / mitigation)
DOE Metrology, May 2nd, 2018 Qualification of 3458A DMM Calibration System 13