Bante510 型电导表简易操作步骤
Training project

Bante510 Type conducitvity meter operation procedures

培训日期 2019.02.28 培训老师


Training Date Trainer
巡回检查的内容和标准(Contents and standards of patrol inspection):

Bante510 型电导表简易操作步骤
Bante510 Type conducitvity meter operation procedures
1. 插入电源,按仪表 Meas 键,预热 20min。
1. Plug in the power, press “Meas”, warming up for 20min.
2. 按 Mode 键,选择 Cond 模式。
2. Press “Mode”, select “Cond” mode.
3. 按住 Cal 键,屏幕显示默认的电极常数 Cell 1,按▲或▼键选择数字与所
连接的电极上标注的常数一致,按 Enter 键,仪表自动进入校准点设置模
式。仪表显示 Cal 1(1 点校准),按▲或▼键选择数字 2 或 3(这里选择 2
点校准),选择好后,按 Enter 键。再按 Meas 键回到测量模式。
3. Press “Cal”, the screen will show the default electrode constant “Cell 1”, press
▲ or ▼ to select the electrode constant to make sure that it is
same with the constant on the connecting electrode. After that,
press “Enter”, the instrument will automatically enter the
calibration point setting mode. Instrument reads “Cal 1” (point
1 calibration). Press ▲ or ▼ to select the number 2 or 3 (in
here, select 2). After that, press “Enter”. Press “Meas” again
to back to the measuring mode.
4. 按 Cal 键,仪表显示 Cal 184.0μS,用除盐水清洗电极,甩去电极上残留的
水,将电极浸入到第一点 84.0μS/cm 的标准溶液中(若电极常数为 1 时,两
印尼庞卡兰苏苏火电项目部
PANGK ALAN SUSU COAL FIRED STEA M POWER PLANT PROJECT UNIT
3&4 2X200MW

点校准选用 84.0μS/cm 和 1413μS/cm 的标准液,若电极常数为 10 时,两点


校准选用 1413μS/cm 和 12.88mS/cm 的标准液)。按 Enter 键,第一个数字闪
动,可以用▲或▼键修改数字,如果不需要修改数字时,按一次 Enter 键数
字向右移动一位,直至仪表上方 Calibration 闪动,等待数秒后,显示 Cal 2
84.0μS
, 提起电极,甩去电极上残留的溶液,然后把电极浸入到第二点
1413μS/cm 的标准液中,按 Enter 键,第一个数字闪动,可以用▲或▼键修
改数字,,如果不需要修改数字时,按一次 Enter 键数字向右移动一位,直
至仪表上方 Calibration 闪动,等待数秒后,显示 End,校准结束,仪表自动
回到测量模式。
4. Press “Cal”, instrument reads Cal 184.0μS. Use demineralized water to flush the
electrode. Shake off the water, and immerse it into the standard solution of point 1
84.0μS/cm (if the electrode constant is 1, point 2 calibration shall select the
standard solutions of 84.0μS/cm and 1413μS/cm; if the electrode constant is 10,
point 2 calibration shall select the standard solutions of 1413μS/cm and
12.88mS/cm). Press “Enter”, the first number will be flashing, and the number can
be modified by ▲ or ▼. If there is no need to modify the number, press “Enter”
to move to the right once, until “Calibration” on the top of screen is flashing, wait
84.0μS
for several seconds, screen shows Cal 2 . Take out the electrode and shake
off the residual solution on the electrode, and then immerse it into the standard
solution of point 2 1413μS/cm. Press “Enter”, the first number will be flashing,
and the number can be modified by ▲ or ▼. If there is no need to modify the
number, press “Enter” to move to the right once, until “Calibration” on the top of
screen is flashing, wait for several seconds, screen shows “End”, calibration is
finished, and the instrument will back to measuring mode automatically.