You are on page 1of 17

VIASKINER® MesoLED LASER

User Manual

CERTIFICATE

PROF.COSMETICA Sp. z o.o.


Ul.Sierakowska 29
05-092 Łomianki, Warszawa
tel./fax. 22 610-44-40, 22 610-43-10
www.chantarelle.pl e-mail: biuro@chantarelle.pl

VIASKINER® MesoLED LASER


DESCRIPTION OF VIASKINER® MesoLED Laser
An easy-to -use advanced multifunctional 3-head device, designed for non-invasive
procedures applied in derma esthetics and aesthetic medicine, offering high efficiency of
modern photo and electrotherapy methods. VIASKINER® features an exclusive
combination of 4 cutting-edge technologies, delivering excellent treatment results:

SONOPHORESIS Head SONO1 Active sonophoresis treatment (without passive electrodes);


ultrasound head frequency: 1 MHz. 5 power options: 20%, 40%,
60%, 80% and 100%. The SONO1 head is designed to use by
continuous motion; direct contact surface of 15.2cm2.
MESOPORATION Head MESO2 No-needle mesotherapy treatment by active electroporation
(without passive electrodes). Electroporation utilized by
VIASKINER® devices is based on the phenomenon of increased
permeability of cell membranes by alternating their resting
potential (rheobase) in response to an electric pulse of specific,
precisely set parameters – a phenomenon the discovery of which
has given its researchers the 2003 Nobel Prize in chemistry. The
method enables the introduction of any kind of active ingredients
(lipo-, hydrophilic molecules of different sizes) directly into the
target cell without the limitations typical for other electrical
treatment approaches. The MESO2 head is designed to use by
continuous motion, direct contact surface of 10.2cm2.
LED Chromotherapy treatment based on photobiomodulation of
CHROMOTHERAPY cellular processes in the skin using 36 LEDs in four colors: UV-
ultraviolet 395nm, blue 470nm, green 525nm, red 650nm and
three laser diodes 635nm (R) used to obtain an additional
laserporation effect. The CHROMO3 head is designed to use by
continuous motion; direct contact surface of 12.6cm2.
PDT LED LASER R PDT photodynamic treatment combines the beneficial interaction
PHOTODYNAMIC of photosensitive chromophores and light energy absorbed by the
Head CHROMO3
THERAPY mitochondria in the skin, causing an increase in energy within cells
and a stimulation of the fibroblast activity. 9 head options
CHROMO3 enable to apply the treatment using only LED light or
together with 635nm (R) laser, allowing to apply the treatment in
patients with contraindications against laser treatment (light
allergy is the only contraindication against LED therapy).The
CHROMO3 head is designed to use by continuous motion; direct
contact surface of 12.6cm2.

VIASKINER® - comprehensive treatment for beautiful and healthy skin

Effective advanced methods for skin rejuvenation and reduction of skin problems in one
device
VIASKINER® allows to use painless application methods for the following procedures:
mesotherapy, LED phototherapy, PDT photodynamic therapy and sonophoresis, delivering
spectacular therapeutic effects.

Alternative to injectable procedures


Non-invasive VIASKINER® technologies allow to apply 58 different procedures using 17
special operation modes and CHANTARELLE dermocosmetics. All procedures are painless.
They do not cause any complications, delivering immediate and clearly visible effects.

Versatility and convenience of use


VIASKINER® is both easy-to-use and highly efficient, offering substantial effectiveness of
new electrotherapy methods. 17 different operation modes, with carefully designed and
precisely programmed parameters, allow to apply 58 procedures. The modes include 3
levels of mesotherapy, 9 LED laser procedures and 5 sonophoresis power options.

Head ergonomic design


3 treatment heads SONO1, MESO2 and CHROMO3 of VIASKINER® allow for both smooth
and safe application of the procedures even onto very sensitive parts of the face, neck and
bust.

Small, portable and easy-to-carry


VIASKINER® is a multifunctional device the size of a laptop. It is light and easy-to-carry,
both around the treatment room and outside.

Patient comfort and satisfaction


VIASKINER® delivers immediate therapeutic effects. Treatment is non-invasive and
painless. After treatment application, patients can come back to their normal activities.

VIASKINER ® LED light effects


The VIASKINER® technology uses a combination of LED light with different wavelengths:

Red light 650 nm –  stimulates tissue metabolism  stimulates healing processes 


stimulates fibroblasts  reduces wrinkles increases absorption of active ingredients
contained in dermocosmetics into the skin;

Green light 525 nm –  inhibits melanogenesis brightens the skin and reduces skin
discoloration spots improves blood circulation

Blue light 470 nm –  delivers anti-inflammatory and anti-bacterial effects


(Propionibacterium Acnes)  reduces acne and skin inflammation  reduces pores.

Ultraviolet light 395 nm -  reduces pigmentation  delivers anti-bacterial and anti-


inflammatory effects

VIASKINER® HEADS

VIASKINER® has THREE heads: SONO1, MESO2 and CHROMO3 (depending on a version):

SONO1 HEAD
– this head is designed for active sonophoresis treatment (without passive electrodes) to
use by continuous motion
– ultrasound head frequency 1 MHz.
– five operation modes for sonophoresis treatment corresponds to five power options:
20%, 40%, 60%, 80% and 100%
– the selection of a given mode depends on a kind of product applied by sonophoresis and
on skin condition/type.
– direct contact surface of 15.2cm2
RECOMMENDATION:
To achieve optimum therapeutic results of derma esthetic procedures, for sonohoresis
treatment applied with the SONO1 head use products recommended for sonophoresis
treatment provided by CHANTARELLE Laboratory Derm Aesthetics, including:
→ 14 treatment concentrates in ampoules,
→ acids in ampoules – almond acid Mandelic Peel, AHA Anti-Photo-Ageing Peel acids
→ 18 active serum - BLACK ESSENTIAL concentrates
→ SPECIAL AESTHETICS Hydro-Gel Ultrasound.

MESO2 HEAD
– this head is designed for mesoporation / no-needle mesotherapy treatment by active
electroporation (without passive electrodes) method.
– the principle of electroporation used in VIASKINER® is based on the phenomenon of
increased permeability of cell membranes by altering their resting potential (rheobase) in
response to an electric pulse with specific, precisely set parameters
– the discovery of this phenomenon has given its researchers the 2003 Nobel Prize in
chemistry.
– the method enables the introduction of any kind of active ingredients (lipo-, hydrophilic
molecules of different sizes) directly into the target cell without the limitations typical for
other electrical treatment approaches.
– the MESO2 head has 3 operation modes allowing to select appropriate parameters for a
specific skin type.
– direct contact surface of 10.2cm2.

RECOMMENDATION:
To achieve optimum dermo aesthetic results use meso-serum products for no-needle
mesotherapy treatment, applied with the MESO2 head VIASKINER®, provided by
CHANTARELLE Laboratory Derm Aesthetics → the offer includes 13 products recommended
for no-needle mesotherapy.

CHROMO3 HEAD
– this head is designed for color therapy using 36 LEDs in 4 colors:
UV-ultraviolet 395nm, blue 470nm, green 525nm, red 650nm - operating in appropriately
optimized systems. The head also has 3 laser diodes 635nm (R) delivering additional effects
of laser treatment.
– The CHROMO3 head has 9 operation modes allowing to apply several procedures based
on color therapy.
– The CHROMO3 head is designed to use by continuous motion; direct contact surface of
12.6cm2.

RECOMMENDATION:
To achieve optimum therapeutic results of derma esthetic procedures, for color therapy
treatment provided by VIASKINERTM use products containing chromophores – chromoactive
ingredients delivered by CHANTARELLE Laboratory Derm Aesthetics. → the offer includes
28 products containing chroma sensitive active ingredients.
The CHROMO3 head can be used to apply effective color therapy procedures, also without
application of special products.

CONTRAINDICATIONS AGAINST THE USE OF THE DEVICE-SPECIFIC HEADS


For specific heads, you must comply to all relevant contraindications against
physiotherapeutical procedures. Each head and its use must be viewed as separate
electrical devices. Before applying any procedure, you must interview the patient and
obtain the patient’s consent to the procedure.

Absolute contraindications:
- pregnancy
- active peptic ulcer disease
- bleeding, in particular gastrointestinal bleeding
- acute viral, bacterial and fungal infections
- active pulmonary tuberculosis
- active cancer disease
- electronic implants (e.g. cardiac pacemaker, cardiac stimulator, etc.)
- organ transplants
- some autoimmune diseases
- body temperature above 38 C
- mental disorders, emotional instability, unreal expectations.

Relative contraindications, requiring medical consultation :


- abnormal secretion of endocrine glands, in particular hyper- and hypothyroidism
- unstable diabetes
- arrhythmia
- acute heart and blood circulation failure syndromes
- light sensitivity and taking photosensitizing medications
- ischaemic heart disease.

GENERAL INSTRUCTIONS FOR USE

VIASKINER® is a professional device and must be used only by a professional who has received the
company’s training.
The use of the device is very intuitive. All subsequent steps are displayed on a very clear control
panel (Step byStep system).

General, practical instructions for application of the procedures:


– the procedure is applied directly onto the skin. The flat part of the head is pressed against
the skin, moved within the area to be treated WITHOUT losing contact with the skin, and
depending on a given head and procedure - applied together with a suitable product
recommended for this specific procedure;
- in majority of cases effects are visible immediately - skin condition is being improved
during the treatment procedure;
– in some patients, procedures may cause a bit stronger reaction than in others (resulting
in skin redness and increased anxiety); in such cases, you may shorten the time of the
procedure and, at the next visit, reduce its power;
– all parameters of laser treatment and LED phototherapy - for all operation modes - have
been designed to respond to a specific skin problem and these settings cannot be changed
- you can only change the time of the procedure.

Frequency of procedures:
Best long-term effects and the highest effectiveness of treatment are achieved by applying
procedures in series of 4-6 procedures, depending on a specific skin problem -once a week,
or as a single procedure.

Hygiene and safety:


VIASKINER® stands out with a modern design and a user-friendly size of a small easy-to-
carry laptop computer; the raised display installed in the device makes the use of the
device much easier as it allows to adjust the view.
At the bottom of the case, there are elastic stabilizers, preventing the device from shaking
during operation. The ergonomic head holder protects the heads from accidental dropping
on the floor.

The design of the heads allows for easy disinfection and keeping the heads clean:
WARNING: To clean the heads, you must wipe the working surface with a disinfecting
preparation– DO NOT DIP the heads into any fluids!

RECOMMENDATION:
To clean the working surface of the head, we recommend to use IDEAL PURE Anti-Bacterial
Spray Lotion offered by CHANTARELLE Laboratory Derm Aesthetics.

TECHNICAL DATA OF VIASKINER® MesoLED LASER


Power supply: 100-240 V, 50/60 Hz, 20 VA
Power supply – safety: - class 2 (grounding not required)
Three heads: SONO1, MESO2 and CHROMO3
Lasers / nominal power: 3pcs x 5 mW (ONLY for P03 and P07 modes of the CHROMO3 head)
Laser safety class: 3R (required eye protection for the operator)
Application area: direct contact method
Contact surface: depending on the head: 6cm 2 – 12.6cm2
Maximum continuous operating time (per a single procedure): 25 minutes
Allowable transportation temperature +5 to +35°C
Allowable humidity (rel.) for transportation 15 to 93%, non-condensing
Allowable atm. pressure for transportation 700 to 1060 hPa
Allowable ambient temperature for storage/operation -40 to +70°C
Allowable ambient humidity (rel.) for storage/operation 10 to 100%
Allowable ambient atm. pressure for storage/operation 500 to 1060 hPa
Dimensions (width x depth x height): 280 x 285 x 25mm
Weight (body, without accessories) approx.1.5 kg

VIASKINER® MesoLED LASER PACKAGE CONTENTS


Device (body) VIASKINER®MESO LED-Laser – 1 pc.
Head SONO1 (for SONOphoresis; metal truncated cone front part of the head) – 1 pc.
Head MESO2 (for no-needle MESOtherapy -electroporation; flat metal-ribbed front part of the head) – 1 pc.
Head CHROMO3 (for LEDtherapy; transparent front of the head) – 1 pc.
Power cord -1 pc.
Head holder – 1 pc.
Safety glasses for the operator, delivered in a case, with the CE certificate (to use with the CHROMO3 head
for P03 and P07 modes) – 1 pair (depending on a version)
User Manual – 1 pc.

THE DEVICE HAS A 24-MONTH WARRANTY IN ACCORDANCE WITH THE ‘TERMS AND CONDITIONS OF
WARRANTY AND GUARANTEE’ CONSTITUTING AN INTEGRAL PART OF THE DEVICE.
THE DEVICE HAS THE CE CERTIFICATE

STARTING YOUR VIASKINER® MesoLED LASER


INSTALLATION OF THE HEAD HOLDER:
Remove the covers of adhesive stripes from the metal holder and attach the holder on the
underside of the casing (first put your device upside-down); for stabilization, use the hole
in the holder and the pin on the underside of the device.

CONNECTING THE HEADS:


Select the heads, place the heads in the holder, one head on each side of the holder,
connect the cords of the heads to appropriate outlets provided on the front side of the
device; plug to the end - so that the plug flange almost touches the front side of the casing:
- to OUT.1, you can connect interchangeably two heads: SONO1 or MESO2; connect the
head you want to use at a given time;
- to OUT.2, you must connect the CHROMO3 head.

CONNECTING THE POWER CORD:


WARNING: Check the position of the power switch at the back of your VIASKINER® - the
position of the switch MUST be ‘0’.

Then plug the power cord into the socket and connect it to an external power outlet with
ground fault protection – the device can be powered by AC (alternating current), voltage
100 to 240 V. Make sure you use only the original power cords (!).

TURNING ON THE DEVICE:


Turn on the device by setting the power switch (at the back of the device) to ‘1’.
The head operating status is indicated by a green diode next to POWER on the side
of the currently operating head. The head connection status is indicated by the RUN
light on the display.

VERY IMPORTANT INFORMATION:


DO NOT turn on your device without first connecting the heads.
DO NOT connect/disconnect the heads cords when the heads are still operating!
The head operating status is indicated by the green diode next to ‘POWER’ on the
side of the currently operating head. The head connection status is indicated by
the RUN light on the display.
If you change the heads when they are still operating, you may damage the
electronic system of the device – the heads can be connected/disconnected ONLY
when the head operation mode has stopped to operate.

TURNING OFF THE DEVICE:


If not used for a long time, the device goes to the silent mode. To use the device again, you
must press any button on the device.
Turning off the device:
Turn off your device by pressing the button at the back of the device, setting the switch
from ‘1‘ to ‘0‘ and unplug the power cord.

Communication with  VIASKINER® MesoLED LASER - SELECTION OF HEADS


The use of the device is very intuitive – at all subsequent levels of the operation, on the
display, you can see clear commands to follow.
To make sure you can see all commands on the display, you must raise the part of the
device with the display (mechanically, with your fingers) to the optimum viewing angle
preferred by the operator; the position you have selected is automatically blocked to
prevent the display from falling during the operation of the device.

After turning on your device, on the display you will see a welcome message (see Image 1.)
and a short auto-test will be performed – all letter boxes will be on. After the test, the
device is ready to use.
Next to the POWER diode on the panel beside the display, the green diode will be on.

VIASKINER VIASKINER
MesoLED Laser MesoLED Laser
Ver. 1.2 Ver. 1.2
PRESS ANY KEY TO CONTINUE Przyciśnij dowolny przycisk aby kontynuować

Image 1.

After pressing any button, on the display you will see the information about connected
heads (Image 2.).
Using ◄► buttons, you must select the head you want to use at a given time: when
selecting a head, the currently selected head is highlighted. If you want to select the head
connected to OUT1, confirm the head currently connected to OUT.1 – by pressing
CONTROL1, for the head connected to OUT.2 – press CONTROL2.
(CONTROL1 and CONTROL2 are on the same side as the output of a given head).
If you want to select the head connected to OUT.1 – confirm with CONTROL1. The light
next to OUT.1 will be on.

SONOphoresis: Head H1 in OUT.1 SONOphoresis: Head H1 in OUT.1


MESOtherapy: Head H2 in OUT.1 MESOtherapy: Head H2 in OUT.1
CHROMOtherapy: Head H3 in OUT.2 CHROMOtherapy: Head H3 in OUT.2
◄►- SELECT A HEAD, CTRL1 – CONFIRM ◄►- WYBIERZ GŁOWICĘ , CTRL1 – AKCEPTUJ

Image 2.

If you select to use the CHROMOtherapy head connected to OUT.2 - confirm with CONTROL2
(Image 3). If you confirm the head you want to select with CONTROL2, the diode next to OUT.2 will
be on.

SONOphoresis: Head H1 in OUT.1 SONOphoresis: Head H1 in OUT.1


MESOtherapy: Head H2 in OUT.1 MESOtherapy: Head H2 in OUT.1
CHROMOtherapy: Head H3 in OUT.2 CHROMOtherapy: Head H3 in OUT.2
◄►- SELECT A HEAD, CTRL2 – CONFIRM ◄►- WYBIERZ GŁOWICĘ , CTRL2 – AKCEPTUJ

Image 3.

Communication with VIASKINER® MesoLED LASER - SELECTION OF OPERATION MODES

1. Head SONOphoresis: Head H1 in OUT.1


Head SONO1 (H1) is connected to OUT.1 and controlled by CONTROL1.

After confirming the selection of the SONOphoresis (H1) head, on the display you will see the list of
operation modes which can be used with this head (one mode – one screen). Each of the 5
operation modes programmed for this head in VIASKINER® has its own flag on the display as
showed below (Image 4.1-4.5).

All operation modes can be viewed using ◄►buttons.

You must confirm the mode you want to select by pressing CONTROL suitable for the head you
intend to use (CONTROL1 for the head connected to OUT.1, CONTROL2 for the head connected to
OUT.2).

The SONO1 (H1) head offers 5 operation modes of different operation power options from 20% to
100% every 20%, four modes offer pulse operation (BURST MODE) and one – 100% intensity -
continuous (CONTINUOUS).

PROG.P01 PROG.P01
OUT.1 : H1-SONO TIME:10min INTENSITY: 20% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 20%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 20% TRYB IMPULSOWY – MOC 20%

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START

Image 4.1

If you want to select PROG.P01, you must press CONTROL1.


On the display, you can see the option of passing time of the procedure (minutes: seconds) – you
can also see how much time is left to complete the procedure (Image 5). On the panel, next to the
display, the orange diode lights up, next to the RUN button.
At the same time, at the bottom row, there appears a bar of moving squares indicating the
continuing operation of the head.

You can pause the procedure at any time by pressing CONTROL1; countdown is stopped, on the
panel next to the display the green diode next to PAUSE lights up and at the bottom row the text
appears saying about the option to unpause the procedure (Image 6.).
If you press CONTROL1 again, the procedure and countdown will be continued.

PROG.P01 9:25 PROG.P01 9:25


OUT.1 : H1-SONO TIME:10min INTENSITY: 20% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 20%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 20% TRYB IMPULSOWY – MOC 20%

¤ ■ ■■■■■■■■■■■■■■■■■■ ¤ ¤ ■ ■■■■■■■■■■■■■■■■■■ ¤
Image 5.

To change the time of the procedure of the selected operation mode, you must press and hold
(for a while!) M button – the function of the time of the procedure is highlighted, use ◄► buttons
to change the time to 5min, 10min, 15min, 20min, or 25min.
Confirm the time of the procedure you have selected by pressing (short click!) M button.
The change of the time of the procedure is confirmed by the text at the bottom of the display –
Image 6.
The standard time set for all modes available using the SONO1 head is 10 minutes.

WARNING: you cannot change the time of the procedure once it has been started! The time of the
procedure must be set before you start applying the procedure.
PROG.P01 PROG.P01
OUT.1 : H1-SONO TIME:10min INTENSITY: 20% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 20%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 20% TRYB IMPULSOWY – MOC 20%

¤ CHANGE VALUE BY ◄ OR ► KEY & PRESS OK ¤ ¤ ZMIEŃ WARTOŚĆ ◄ LUB ► PRZYCIŚNIJ OK ¤


Image 6.

During the procedure, you can, at any time, go to the VIASKINER® main menu (with the list of all
heads to choose from) using CONTROL1 and M buttons.

To view all operation modes available for the SONO1 head, press ◄►buttons; confirm the option
you want to use by pressing CONTROL1

PROG.P02 PROG.P02
OUT.1 : H1-SONO TIME:10min INTENSITY: 40% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 40%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 40% TRYB IMPULSOWY – MOC 40%

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START

Image 4.2

PROG.P03 PROG.P03
OUT.1 : H1-SONO TIME:10min INTENSITY: 60% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 60%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 60% TRYB IMPULSOWY – MOC 60%

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START

Image 4.3

PROG.P04 PROG.P04
OUT.1 : H1-SONO TIME:10min INTENSITY: 80% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 80%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 80% TRYB IMPULSOWY – MOC 80%

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 4.4

PROG.P05 PROG.P05
OUT.1 : H1-SONO TIME:10min INTENSITY: 100% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 100%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
CONTINUOUS MODE – POWER 100% TRYB CIĄGŁY – MOC 100%

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START

Image 4.5

2. Head MESOtherapy: Head H2 in OUT.1

JUST AS A REMINDER: Before changing the heads connected to the same outlet – this refers to
OUT.1 and SONO1 and MESO2 heads – you MUST turn off your VIASKINER® using the ‘0/1‘ switch at
the back of the device (when changing the heads, the switch must be set to 0). You must not
change the heads when the device is on as this may cause damage to the electrical circuit of both
the device and the head.

Head MESO2 (H2) is connected to OUT.1 and controlled by CONTROL1.


After confirming the selection of the MESOtherapy (H2 – no-needle mesotherapy) head - at the
beginning (Image 2) - on the display you will see the list of operation modes which are available
using this head (one mode – one screen). Each of the 3 operation modes programmed for this head
in VIASKINER® has its own flag on the display as showed below (Image 7.1-7.3).

The list of all MESO2 (H2) operation modes can be viewed using ◄►buttons.

You must confirm the mode you want to select by pressing CONTROL1 when the flag of the option
you intend to use appears on the display.

For the MESO2 (H2) head, there are 3 operation modes available, offering different AC frequency
depending on the patient’s skin type and skin problems.
The device has different parameters set for three groups:
- P01 frequency of 1.0 kHz – skin type/skin problems: thin, sagging, inelastic, stretch marks;
programmed procedure time: 10 minutes;
- P02 frequency of 1.5 kHz – skin type/skin problems: mature, dry, sagging, wrinkled, spots/
/discoloration; programmed procedure time: 15 minutes;
- P03 frequency of 2.0 kHz – skin type/skin problems: oily, irritated, with furrows, scars;
programmed procedure time: 20 minutes.

PROG.P01 PROG.P01
OUT.1 : H2-MESO TIME:10min FREQ: 1.0 kHz OUT.1 : H2-MESO CZAS:10min CZĘSTOTLIWOŚĆ: 1.0 kHz
MESO THERAPY NO NEEDLE – ELECTROPORATION 1 MEZOTERAPIA BEZIGŁOWA – ELEKTROPORACJA 1
SKIN: THIN, SAGGING, INELASTIC, STRETCH MARKS SKÓRA: CIENKA, WIOTKA, NIEELASTYCZNA, ROZSTĘPY

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START

Image 7.1

PROG.P02 PROG.P02
OUT.1 : H2-MESO TIME:15min FREQ: 1.5 kHz OUT.1 : H2-MESO CZAS:15min CZĘSTOTLIWOŚĆ: 1.5 kHz
MESO THERAPY NO NEEDLE – ELECTROPORATION 2 MEZOTERAPIA BEZIGŁOWA – ELEKTROPORACJA 2
SKIN: MATURE, DRY, SAGGING, WRINKLED, SPOTS SKÓRA: DOJRZAŁA, SUCHA, WIOTKA, ZMARSZCZKI, PLAMY

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START

Image 7.2

PROG.P03 PROG.P03
OUT.1 : H2-MESO TIME:20min FREQ: 2.0 kHz OUT.1 : H2-MESO CZAS:20min CZĘSTOTLIWOŚĆ: 2.0 kHz
MESO THERAPY NO NEEDLE – ELECTROPORATION 2 MEZOTERAPIA BEZIGŁOWA – ELEKTROPORACJA 2
SKIN: OILY, IRRITATED, WITH FURROWS, SCARS SKÓRA: ŁUSTA, PODRAŻNIONA, BRUZDY, BLIZNY

¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START

Image 7.3

To start the program you have selected, PROG.P01, PROG.P02 or PROG.P03, press CONTROL1.
On the display, you can see the option of passing time of the procedure (minutes: seconds) – you
can also see how much time is left to complete the procedure, as in the case of the SONO1 head
(Image 5). On the panel, next to the display, the orange diode lights up, next to the RUN button. At
the same time, at the bottom row, there appears a bar of moving squares indicating the continuing
operation of the head.

You can pause the procedure at any time by pressing CONTROL1; countdown is stopped, on the
panel next to the display the green diode next to PAUSE lights up and at the bottom row the text
appears saying about the option to unpause the procedure, as in the case of the SONO1 head
(Image 6.).
If you press CONTROL1 again, the procedure is unpaused and the countdown continues.
To change the time of the procedure of the selected operation mode, you must press and hold (for
a while!) M – the function of the time of the procedure is highlighted. Use ◄► buttons to change
the time to 5min, 10min, 15min, 20min, or 25min. Confirm the time of the procedure you have
selected by pressing (short click!) M.

The change of the time of the procedure is confirmed by the text at the bottom of the display. .

WARNING: you cannot change the time of the procedure once it has been started! The time of the
procedure must be set before you start applying the procedure.

During the procedure, you can, at any time, go to the VIASKINER® main menu (with the list of all
heads to choose from) using CONTROL1 and M buttons.

3. Head CHROMOtherapy: Head H3 in OUT.2

Head CHROMO3 (H3) is connected to OUT.2 and controlled by CONTROL2.

After confirming the selection of the CHROMOtherapy (H3 – laser led light therapy) head - at the
beginning (Image 2) - on the display you will see the list of operation modes which are available
using this head (one mode – one screen).

The CHROMO3 head has 36 LEDs producing pulses of light with wavelengths corresponding to the
following colors: R (650nm Red), G (525nm Green), B (470nm Blue), UV (395nm Ultraviolet), and 3
low-power laser diodes producing red light 635nm.
The lasers have been used for two programs carried out by using the CHROMO3 head: P03 and P07,
All remaining programs use only the combinations of LED diodes corresponding to the patient’s skin
type and the patient’s skin problems.

For the CHROMO3 (H3) head used in VIASKINER®, there are 9 operation modes available, offering
different parameters suitable for different skin types and skin problems.

The list of all CHROMO3 (H3) operation modes can be viewed using ◄►buttons.

You must confirm the mode you want to select by pressing CONTROL2 when the flag of the option
you intend to use appears on the display.

The basic time of the procedures applied with the CHROMO3 head is 15 minutes. The time of the
procedure can be changed in accordance with the instructions described for the SONO1 and
MESO2 heads.
Other functions such as pausing and unpausing the procedure, or the change of the procedure
before its completion, are standard for this device and applied according to the instructions
described for the SONO1 and MESO2 heads. Please note , however, that the button used to control
the CHROMO3 head is CONTROL2.

P01. LED:R (red LED diodes 650nm)

-USED BEFORE OTHER PROCEDURES - this is due to a wide range of special properties of red light
which delivers effects that can be compared with that of pre-massage applied to increase the
bioavailability of the skin for active ingredients and to improve the biomechanical properties of the
skin,
-IMPROVING (BLOOD) CIRCULATION – this is due to the fact that red light dramatically improves
blood circulation and lymph flow (when combined with gentle massage applied during the
procedure), as well as boosts metabolism, thus, allowing skincare products and cosmetic
procedures to deliver better effects,
-BLACK ESSENTIAL PDT – used intentionally, PDT (Photo Dynamic Therapy) increases the
effectiveness of the procedure as dermocosmetics contain special chromophores responding to
specific wavelengths of light. Dermocosmetics BLACK ESSENTIAL CHANTARELLE Laboratory Derm
Aesthetics contain 650nm red monochromatic light photosensitive chromophores and have been
designed for 24 different cosmetic procedures (Appendix no. 1 to this User Manual).

PROG.P01 PROG.P01
OUT.2 : H3-CHROMO TIME:15min LED: R OUT.2 : H3-CHROMO CZAS:15min LED: R
BEFORE TREATMENT CIRCULATION IMPROVEMENT PRZED ZABIEGAMI POPRAWA CYRKULACJI
& BLACK ESSENTIAL PDT & BLACK ESSENTIAL i metoda PDT

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.4

P02. LED: GREEN+R (green LEDs 525nm and red LEDs 650nm)

-LIFTING - this is due to the special properties of red and green light, which stimulate skin’s
metabolic processes – increase the number of metabolic reactions and enhance enzymatic channels
by improving skin hydration. When the two types of light are combined, they deliver perfect skin
lifting effects.
-MOISTURIZING – this is particularly true for green light, as one of the main characteristics of green
light (apart from reducing spots, discoloration, and visibility of capillaries) is the stimulation of
various processes helping improve skin hydration.
-REDUCTION OF WRINKLES – this is due to the fact that green and red light work in synergy,
delivering spectacular effects on the skin.
-IMPROVES SKIN TONE – both red and green light, regulating and stimulating skin regeneration
processes, help dramatically improve the appearance of the skin by reducing wrinkles, uneven skin
tone, dullness and signs of fatigue. The skin becomes smooth, looks healthy and radiant, without
visible signs of degradation.

PROG.P02 PROG.P02
OUT.2 : H3-CHROMO TIME:15min LED: G+R OUT.2 : H3-CHROMO CZAS:15min LED: G+R
LIFTING & HYDRATING EFFECT & SMOOTHING LIFTING & EFEKT NAWILŻENIA & WYGŁADZENIE
FINE LINES & SKIN TONING ZMARSZCZKI & TONUS SKÓRY

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.5

P03. LASER:R LED: G+R (red laser 635nm, green LEDs 525nm and red LEDs 650nm)

-ANTI-AGEING & ANTI-STRETCH & ANTI-INFLAMMATORY – this is an intensive skin rejuvenation


program recommended for patients WITHOUT contraindications against laser biostimulation.
WARNING: When performing procedures using this program, the operator MUST wear laser safety
glasses (delivered together with VIASKINER®).
This program is recommended due to the powerful properties of red and green light, helping
directly improve skin texture (when combined together, these types of light significantly stimulate
the production of elastic fibers) and improve functioning of the skin’s metabolic pathways, and help
restore basic parameters of epidermal homeostasis.

PROG.P03 PROG.P03
OUT.2 : H3-CHROMO TIME:15min LASER:R LED: G+R OUT.2 : H3-CHROMO CZAS:15min LASER:R LED: G+R
ANTI-AGEING & ANTI-STRETCH PRZECIW STARZENIOWY & PRZECIW ROZSTĘPOM
& ANTI-INFLAMMATORY & PRZECIW OBJAWOM STANÓW ZAPALNYCH

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.6

P04. LED: BLUE +GREEN +RED (blue LEDs 470nm, green LEDs 525nm and red LEDs 650nm)

-ANTI-REDNESS – this combination is recommended to use to treat and PREVENT SKIN REDNESS as
blue light has strong soothing properties whereas green light reduces the visibility of skin redness
and pigmentation.
-ANTI-COUPEROSE & ANTI-ROSACEA – the ANTI-COUPEROSE & ANTI-ROSACEA effects of the
combination of these types of light are possible as green light helps moisturize and soothe the skin,
reduce the visibility of broken capillaries, discoloration and other similar imperfections, whereas
red light improves blood circulation and deeply regenerates the skin.

PROG.P04 PROG.P04
OUT.2 : H3-CHROMO TIME:15min LED: B+G+R OUT.2 : H3-CHROMO CZAS:15min LED: B+G+R
ANTI-REDNESS & ANTI-COUPEROSE PRZECIW ZACZERWIENIENIOM & PRZECIW NACZYNKOM
& ANTI-ROSACEA & PRZECIW TRĄDZIKOWI RÓŻOWATEMU

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.7

P05. LED: UV +BLUE+RED (ultraviolet LEDs 395nm, blue LEDs 470nm and red LEDs 650nm)

-MIXED SKIN & PORE REDUCTION & MATIFFYING & ANTI-INFLAMMATORY – Blue light soothes the
skin, reduces signs of skin irritation. UV light delivers anti-bacterial and balancing effects on the
skin. Researchers also found other powerful effects of the combination of ultraviolet light, blue light
and red light (!) Strong anti-inflammatory effects, helping treat acne of mature skin (Acne Tarda).
-HYDRATING – the direct indication for red light.

PROG.P05 PROG.P05
OUT.2 : H3-CHROMO TIME:15min LED: UV+B+R OUT.2 : H3-CHROMO CZAS:15min LED: UV+B+R
MIXED SKIN & PORE REDUCTION & ANTI- SKÓRA MIESZANA & REDUKCJA ROZSZERZONYCH PORÓW
INFLAMMATORY & MATTIFYING HYDRATING & PRZECIW ZAPALNE & MATUJĄCE NAWILŻAJĄCE

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.8

P06. LED: UV+BLUE (ultraviolet LEDs 395nm and blue LEDs 470nm)

-OILY SKIN/ACNE PRONE & SEBUM REDUCTION & ANTI-INFLAMMATORY & CALMING – Blue light
soothes the skin and reduces signs of skin irritation while UV light gives anti-inflammatory and
balancing effects. Blue and ultraviolet light help considerably reduce sebum production and soothe
existing inflammation, also in the form of active lesions in clogged sebaceous ducts. As a result of
mixed UV and blue light, the two types of light help reduce the causes and symptoms of
inflammation, delivering calming and soothing effects.

PROG.P06 PROG.P06
OUT.2 : H3-CHROMO TIME:15min LED: UV+B OUT.2 : H3-CHROMO CZAS:15min LED: UV+B
OILY SKIN/ACNE PRONE & SEBUM REDUCTION SKÓRA TŁUSTA/TRĄDZIKOWA & REDUKCJA SEBUM
& ANTI-INFLAMMATORY & CALMING & PRZECIW ZAPALNE & USPOKAJAJĄCE

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.9

P07. LASER:R LED: BLUE (red laser 635nm blue LEDs 470nm)
This program is recommended for patients WITHOUT contraindications against laser biostimulation.
WARNING: When performing procedures using this program, the operator MUST wear laser safety
glasses (delivered together with VIASKINER®).

-SKIN CONDITIONING – Red light helps replenish and restore impaired skin at several levels, both at
a chemical level (stimulating metabolism) and directly, at a physical level – improving blood
circulation and stimulating skin elasticity.
-ACNE-PRONE ADVANCED – this program helps treat severe acne. Blue light calms and soothes the
skin when combined with red light stimulates metabolism and detoxification of skin cells, and thus
brings the inflammation soothing effects much faster, as well as reduces skin redness.
-ANTI-SCARS – the combination of red light and blue light reduces the visibility of scars by softening
scar tissue, helping smooth the scars and even skin tone, improves blood circulation in the scar area
and deeply soothes the skin.

PROG.P07 PROG.P07
OUT.2 : H3-CHROMO TIME:15min LASER:R LED: B OUT.2 : H3-CHROMO CZAS:15min LASER:R LED: B
SKIN CONDITIONING & ACNE-PRONE ADVANCED POPRAWA JAKOŚCI SKÓRY & ZAAWANSOWANY TRĄDZIK
& ANTI-SCARS & PRZECIW BLIZNOM

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.10

P08. LED: UV+BLUE+GREEN+ R (ultraviolet LEDs 395nm, blue LEDs 470nm, green LEDs 525nm and
red LEDs 650nm)

-DEPIGMENTATION & ANTI-SPOT & ANTI-PHOTOAGEING – the combination of UV, blue, green and
red light has been used to provide cumulated powerful effects to reduce spots and skin
pigmentation caused by different factors ranging from skin aging, caused by excessive sun tanning
and photoaging but also these resulting from oxidative changes due to some external and internal
factors.
This program is also recommended to use to reduce the visibility of discoloration appearing in areas
difficult to treat e.g. eye area and eyelids. In these cases, it is recommended to shorten the time of
the procedure to 10 minutes, 5 minutes per eyelid/eye. After the procedure, the patient should
stay in the chair for 10 more minutes WITHOUT OPENING the eyes. During this time, the operator
may put on the eyes pads moistened with appropriate tonic or apply appropriate serum.

PROG.P08 PROG.P08
OUT.2 : H3-CHROMO TIME:15min LED: UV+B+G+R OUT.2 : H3-CHROMO CZAS:15min LED: UV+B+G+R
DEPIGMENTATION & ANTI-SPOT DEPIGMENTACJA & PRZECIW PLAMOM
& ANTI-PHOTOAGEING & PRZECIW OBJAWOM FOTOSTARZENIA

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.11

P09. LED: UV+BLUE +GREEN (ultraviolet LEDs 395nm, blue LEDs 470nm and green LEDs 525nm)

-AFTER TREATMENT – P09 is a program which is a combination of ultraviolet LEDS delivering anti-
bacterial effects, blue LEDs which soothe and calm the skin, and green LEDs helping reduce skin
redness. This program is recommended to apply at the end of each procedure – performed before
applying the last cream. This extra procedure makes the skin look healthy, radiant and relaxed, and
increases antibacterial protection levels of the skin which is particularly important in treatments
with a high level of antibacterial uncertainty e.g. procedures damaging the epidermis and after
deep exfoliation.
-CALMING & ANTI-INFLAMMATORY & REDNESS REDUCTION – Ultraviolet, blue and green light offer
powerful soothing, anti-irritation, anti-inflammatory and anti-redness effects on the skin. The
procedure delivers practically immediate results. P09 can be performed at any stage of any
procedure, if necessary.

PROG.P09 PROG.P09
OUT.2 : H3-CHROMO TIME:15min LED: UV+B+G OUT.2 : H3-CHROMO CZAS:15min LED: UV+B+G
AFTER TREATMENT CALMING & ANTI- NA ZAKOŃCZENIE ZABIEGU USPOKOJENIE &
INFLAMMATORY & REDNESS REDUCTION PRZECIW PODRAŻNIENIOM & REDUKCJA ZACZERWIENIEŃ

¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START

Image 7.12

Warranty period for VIASKINER® MesoLED LASER: 24 months. The warranty does not cover any
mechanical damage to the heads and wiring/outlets and electronics damaged as a result of
changing the heads during the operation of the device. Any attempt to connect other devices to
any of the outlets of the heads will damage the control electronics!.

It is prohibited to copy or distribute this User Manual VIASKINER® MesoLED LASER or any part of it.

 The device should be discarded as electronic waste. The device does not contain any hazardous materials
that can threaten human health.

PRECAUTIONS FOR USE:


The device must be kept away from moisture! It can be used only in dry places.

If it is necessary to replace the power cord – it must be replaced only with a cord of the same class and
with the same parameters.

Never disconnect the heads during the operation of VIASKINER® MesoLED LASER! This may damage the
electronic system of the device. The heads can be connected/disconnected ONLY when the device is off.
To power VIASKINER® MesoLED LASER, It is recommended to use a cord with a safety switch.
WARNING: The lasers used in the CHROMO3 head of VIASKINER® MesoLED LASER have a 3R safety class
– you should never look directly into the light/ beam of the laser neither with your eyes nor using any optical
device. The operator, to perform procedures using P03 and P07 programs with the CHROMO3 head, must put
on special certified safety glasses designed for this class of lasers. The patient must also wear special safety
pads.

To prevent fire, electric shock or device damage, the device must not be exposed to rain and
moisture, It must be protected from water and other liquids. Do not put any object filled with liquids on the
device, e.g. vases.
Use only the accessories recommended in this User Manual.
Do not repair the device yourself.

Warning! Do not put the device in a confined space e.g. bookshelf or a built-in cabinet. Make sure
the device has appropriate ventilation. To prevent electric shock or fire which may result from the
overheating of the device, make sure that no materials, e.g. curtains, cover the ventilation holes.
Do not cover the ventilation holes of the device with newspapers, tablecloths, curtains, etc.
Do not put any naked flame sources e.g. candles on the device.
On the VIASKINER® MesoLED LASER casing and on the CHROMO3 head power cord, there are the following
warning and information signs:

Visible laser radiation LASER APERTURE


AVOID DIRECT EYE EXPOSURE Laser radiation is
Class 3R laser device emitted at the tip
Diode laser 635nm of the applicator
Max output power <5mW
EN 60825-1:2007