Professional Documents
Culture Documents
User Manual
CERTIFICATE
Effective advanced methods for skin rejuvenation and reduction of skin problems in one
device
VIASKINER® allows to use painless application methods for the following procedures:
mesotherapy, LED phototherapy, PDT photodynamic therapy and sonophoresis, delivering
spectacular therapeutic effects.
Green light 525 nm – inhibits melanogenesis brightens the skin and reduces skin
discoloration spots improves blood circulation
VIASKINER® HEADS
VIASKINER® has THREE heads: SONO1, MESO2 and CHROMO3 (depending on a version):
SONO1 HEAD
– this head is designed for active sonophoresis treatment (without passive electrodes) to
use by continuous motion
– ultrasound head frequency 1 MHz.
– five operation modes for sonophoresis treatment corresponds to five power options:
20%, 40%, 60%, 80% and 100%
– the selection of a given mode depends on a kind of product applied by sonophoresis and
on skin condition/type.
– direct contact surface of 15.2cm2
RECOMMENDATION:
To achieve optimum therapeutic results of derma esthetic procedures, for sonohoresis
treatment applied with the SONO1 head use products recommended for sonophoresis
treatment provided by CHANTARELLE Laboratory Derm Aesthetics, including:
→ 14 treatment concentrates in ampoules,
→ acids in ampoules – almond acid Mandelic Peel, AHA Anti-Photo-Ageing Peel acids
→ 18 active serum - BLACK ESSENTIAL concentrates
→ SPECIAL AESTHETICS Hydro-Gel Ultrasound.
MESO2 HEAD
– this head is designed for mesoporation / no-needle mesotherapy treatment by active
electroporation (without passive electrodes) method.
– the principle of electroporation used in VIASKINER® is based on the phenomenon of
increased permeability of cell membranes by altering their resting potential (rheobase) in
response to an electric pulse with specific, precisely set parameters
– the discovery of this phenomenon has given its researchers the 2003 Nobel Prize in
chemistry.
– the method enables the introduction of any kind of active ingredients (lipo-, hydrophilic
molecules of different sizes) directly into the target cell without the limitations typical for
other electrical treatment approaches.
– the MESO2 head has 3 operation modes allowing to select appropriate parameters for a
specific skin type.
– direct contact surface of 10.2cm2.
RECOMMENDATION:
To achieve optimum dermo aesthetic results use meso-serum products for no-needle
mesotherapy treatment, applied with the MESO2 head VIASKINER®, provided by
CHANTARELLE Laboratory Derm Aesthetics → the offer includes 13 products recommended
for no-needle mesotherapy.
CHROMO3 HEAD
– this head is designed for color therapy using 36 LEDs in 4 colors:
UV-ultraviolet 395nm, blue 470nm, green 525nm, red 650nm - operating in appropriately
optimized systems. The head also has 3 laser diodes 635nm (R) delivering additional effects
of laser treatment.
– The CHROMO3 head has 9 operation modes allowing to apply several procedures based
on color therapy.
– The CHROMO3 head is designed to use by continuous motion; direct contact surface of
12.6cm2.
RECOMMENDATION:
To achieve optimum therapeutic results of derma esthetic procedures, for color therapy
treatment provided by VIASKINERTM use products containing chromophores – chromoactive
ingredients delivered by CHANTARELLE Laboratory Derm Aesthetics. → the offer includes
28 products containing chroma sensitive active ingredients.
The CHROMO3 head can be used to apply effective color therapy procedures, also without
application of special products.
Absolute contraindications:
- pregnancy
- active peptic ulcer disease
- bleeding, in particular gastrointestinal bleeding
- acute viral, bacterial and fungal infections
- active pulmonary tuberculosis
- active cancer disease
- electronic implants (e.g. cardiac pacemaker, cardiac stimulator, etc.)
- organ transplants
- some autoimmune diseases
- body temperature above 38 C
- mental disorders, emotional instability, unreal expectations.
VIASKINER® is a professional device and must be used only by a professional who has received the
company’s training.
The use of the device is very intuitive. All subsequent steps are displayed on a very clear control
panel (Step byStep system).
Frequency of procedures:
Best long-term effects and the highest effectiveness of treatment are achieved by applying
procedures in series of 4-6 procedures, depending on a specific skin problem -once a week,
or as a single procedure.
The design of the heads allows for easy disinfection and keeping the heads clean:
WARNING: To clean the heads, you must wipe the working surface with a disinfecting
preparation– DO NOT DIP the heads into any fluids!
RECOMMENDATION:
To clean the working surface of the head, we recommend to use IDEAL PURE Anti-Bacterial
Spray Lotion offered by CHANTARELLE Laboratory Derm Aesthetics.
THE DEVICE HAS A 24-MONTH WARRANTY IN ACCORDANCE WITH THE ‘TERMS AND CONDITIONS OF
WARRANTY AND GUARANTEE’ CONSTITUTING AN INTEGRAL PART OF THE DEVICE.
THE DEVICE HAS THE CE CERTIFICATE
Then plug the power cord into the socket and connect it to an external power outlet with
ground fault protection – the device can be powered by AC (alternating current), voltage
100 to 240 V. Make sure you use only the original power cords (!).
After turning on your device, on the display you will see a welcome message (see Image 1.)
and a short auto-test will be performed – all letter boxes will be on. After the test, the
device is ready to use.
Next to the POWER diode on the panel beside the display, the green diode will be on.
VIASKINER VIASKINER
MesoLED Laser MesoLED Laser
Ver. 1.2 Ver. 1.2
PRESS ANY KEY TO CONTINUE Przyciśnij dowolny przycisk aby kontynuować
Image 1.
After pressing any button, on the display you will see the information about connected
heads (Image 2.).
Using ◄► buttons, you must select the head you want to use at a given time: when
selecting a head, the currently selected head is highlighted. If you want to select the head
connected to OUT1, confirm the head currently connected to OUT.1 – by pressing
CONTROL1, for the head connected to OUT.2 – press CONTROL2.
(CONTROL1 and CONTROL2 are on the same side as the output of a given head).
If you want to select the head connected to OUT.1 – confirm with CONTROL1. The light
next to OUT.1 will be on.
Image 2.
If you select to use the CHROMOtherapy head connected to OUT.2 - confirm with CONTROL2
(Image 3). If you confirm the head you want to select with CONTROL2, the diode next to OUT.2 will
be on.
Image 3.
After confirming the selection of the SONOphoresis (H1) head, on the display you will see the list of
operation modes which can be used with this head (one mode – one screen). Each of the 5
operation modes programmed for this head in VIASKINER® has its own flag on the display as
showed below (Image 4.1-4.5).
You must confirm the mode you want to select by pressing CONTROL suitable for the head you
intend to use (CONTROL1 for the head connected to OUT.1, CONTROL2 for the head connected to
OUT.2).
The SONO1 (H1) head offers 5 operation modes of different operation power options from 20% to
100% every 20%, four modes offer pulse operation (BURST MODE) and one – 100% intensity -
continuous (CONTINUOUS).
PROG.P01 PROG.P01
OUT.1 : H1-SONO TIME:10min INTENSITY: 20% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 20%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 20% TRYB IMPULSOWY – MOC 20%
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 4.1
You can pause the procedure at any time by pressing CONTROL1; countdown is stopped, on the
panel next to the display the green diode next to PAUSE lights up and at the bottom row the text
appears saying about the option to unpause the procedure (Image 6.).
If you press CONTROL1 again, the procedure and countdown will be continued.
¤ ■ ■■■■■■■■■■■■■■■■■■ ¤ ¤ ■ ■■■■■■■■■■■■■■■■■■ ¤
Image 5.
To change the time of the procedure of the selected operation mode, you must press and hold
(for a while!) M button – the function of the time of the procedure is highlighted, use ◄► buttons
to change the time to 5min, 10min, 15min, 20min, or 25min.
Confirm the time of the procedure you have selected by pressing (short click!) M button.
The change of the time of the procedure is confirmed by the text at the bottom of the display –
Image 6.
The standard time set for all modes available using the SONO1 head is 10 minutes.
WARNING: you cannot change the time of the procedure once it has been started! The time of the
procedure must be set before you start applying the procedure.
PROG.P01 PROG.P01
OUT.1 : H1-SONO TIME:10min INTENSITY: 20% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 20%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 20% TRYB IMPULSOWY – MOC 20%
During the procedure, you can, at any time, go to the VIASKINER® main menu (with the list of all
heads to choose from) using CONTROL1 and M buttons.
To view all operation modes available for the SONO1 head, press ◄►buttons; confirm the option
you want to use by pressing CONTROL1
PROG.P02 PROG.P02
OUT.1 : H1-SONO TIME:10min INTENSITY: 40% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 40%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 40% TRYB IMPULSOWY – MOC 40%
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 4.2
PROG.P03 PROG.P03
OUT.1 : H1-SONO TIME:10min INTENSITY: 60% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 60%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 60% TRYB IMPULSOWY – MOC 60%
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 4.3
PROG.P04 PROG.P04
OUT.1 : H1-SONO TIME:10min INTENSITY: 80% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 80%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
BURST MODE – POWER 80% TRYB IMPULSOWY – MOC 80%
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 4.4
PROG.P05 PROG.P05
OUT.1 : H1-SONO TIME:10min INTENSITY: 100% OUT.1 : H1-SONO CZAS:10min INTENSYWNOŚĆ: 100%
SONOPHORESIS 1 MHz SONOFOREZA 1 MHz
CONTINUOUS MODE – POWER 100% TRYB CIĄGŁY – MOC 100%
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 4.5
JUST AS A REMINDER: Before changing the heads connected to the same outlet – this refers to
OUT.1 and SONO1 and MESO2 heads – you MUST turn off your VIASKINER® using the ‘0/1‘ switch at
the back of the device (when changing the heads, the switch must be set to 0). You must not
change the heads when the device is on as this may cause damage to the electrical circuit of both
the device and the head.
The list of all MESO2 (H2) operation modes can be viewed using ◄►buttons.
You must confirm the mode you want to select by pressing CONTROL1 when the flag of the option
you intend to use appears on the display.
For the MESO2 (H2) head, there are 3 operation modes available, offering different AC frequency
depending on the patient’s skin type and skin problems.
The device has different parameters set for three groups:
- P01 frequency of 1.0 kHz – skin type/skin problems: thin, sagging, inelastic, stretch marks;
programmed procedure time: 10 minutes;
- P02 frequency of 1.5 kHz – skin type/skin problems: mature, dry, sagging, wrinkled, spots/
/discoloration; programmed procedure time: 15 minutes;
- P03 frequency of 2.0 kHz – skin type/skin problems: oily, irritated, with furrows, scars;
programmed procedure time: 20 minutes.
PROG.P01 PROG.P01
OUT.1 : H2-MESO TIME:10min FREQ: 1.0 kHz OUT.1 : H2-MESO CZAS:10min CZĘSTOTLIWOŚĆ: 1.0 kHz
MESO THERAPY NO NEEDLE – ELECTROPORATION 1 MEZOTERAPIA BEZIGŁOWA – ELEKTROPORACJA 1
SKIN: THIN, SAGGING, INELASTIC, STRETCH MARKS SKÓRA: CIENKA, WIOTKA, NIEELASTYCZNA, ROZSTĘPY
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 7.1
PROG.P02 PROG.P02
OUT.1 : H2-MESO TIME:15min FREQ: 1.5 kHz OUT.1 : H2-MESO CZAS:15min CZĘSTOTLIWOŚĆ: 1.5 kHz
MESO THERAPY NO NEEDLE – ELECTROPORATION 2 MEZOTERAPIA BEZIGŁOWA – ELEKTROPORACJA 2
SKIN: MATURE, DRY, SAGGING, WRINKLED, SPOTS SKÓRA: DOJRZAŁA, SUCHA, WIOTKA, ZMARSZCZKI, PLAMY
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 7.2
PROG.P03 PROG.P03
OUT.1 : H2-MESO TIME:20min FREQ: 2.0 kHz OUT.1 : H2-MESO CZAS:20min CZĘSTOTLIWOŚĆ: 2.0 kHz
MESO THERAPY NO NEEDLE – ELECTROPORATION 2 MEZOTERAPIA BEZIGŁOWA – ELEKTROPORACJA 2
SKIN: OILY, IRRITATED, WITH FURROWS, SCARS SKÓRA: ŁUSTA, PODRAŻNIONA, BRUZDY, BLIZNY
¤ ◄►- PROGRAM, M- HEAD, CTRL1 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL1 – START
Image 7.3
To start the program you have selected, PROG.P01, PROG.P02 or PROG.P03, press CONTROL1.
On the display, you can see the option of passing time of the procedure (minutes: seconds) – you
can also see how much time is left to complete the procedure, as in the case of the SONO1 head
(Image 5). On the panel, next to the display, the orange diode lights up, next to the RUN button. At
the same time, at the bottom row, there appears a bar of moving squares indicating the continuing
operation of the head.
You can pause the procedure at any time by pressing CONTROL1; countdown is stopped, on the
panel next to the display the green diode next to PAUSE lights up and at the bottom row the text
appears saying about the option to unpause the procedure, as in the case of the SONO1 head
(Image 6.).
If you press CONTROL1 again, the procedure is unpaused and the countdown continues.
To change the time of the procedure of the selected operation mode, you must press and hold (for
a while!) M – the function of the time of the procedure is highlighted. Use ◄► buttons to change
the time to 5min, 10min, 15min, 20min, or 25min. Confirm the time of the procedure you have
selected by pressing (short click!) M.
The change of the time of the procedure is confirmed by the text at the bottom of the display. .
WARNING: you cannot change the time of the procedure once it has been started! The time of the
procedure must be set before you start applying the procedure.
During the procedure, you can, at any time, go to the VIASKINER® main menu (with the list of all
heads to choose from) using CONTROL1 and M buttons.
After confirming the selection of the CHROMOtherapy (H3 – laser led light therapy) head - at the
beginning (Image 2) - on the display you will see the list of operation modes which are available
using this head (one mode – one screen).
The CHROMO3 head has 36 LEDs producing pulses of light with wavelengths corresponding to the
following colors: R (650nm Red), G (525nm Green), B (470nm Blue), UV (395nm Ultraviolet), and 3
low-power laser diodes producing red light 635nm.
The lasers have been used for two programs carried out by using the CHROMO3 head: P03 and P07,
All remaining programs use only the combinations of LED diodes corresponding to the patient’s skin
type and the patient’s skin problems.
For the CHROMO3 (H3) head used in VIASKINER®, there are 9 operation modes available, offering
different parameters suitable for different skin types and skin problems.
The list of all CHROMO3 (H3) operation modes can be viewed using ◄►buttons.
You must confirm the mode you want to select by pressing CONTROL2 when the flag of the option
you intend to use appears on the display.
The basic time of the procedures applied with the CHROMO3 head is 15 minutes. The time of the
procedure can be changed in accordance with the instructions described for the SONO1 and
MESO2 heads.
Other functions such as pausing and unpausing the procedure, or the change of the procedure
before its completion, are standard for this device and applied according to the instructions
described for the SONO1 and MESO2 heads. Please note , however, that the button used to control
the CHROMO3 head is CONTROL2.
-USED BEFORE OTHER PROCEDURES - this is due to a wide range of special properties of red light
which delivers effects that can be compared with that of pre-massage applied to increase the
bioavailability of the skin for active ingredients and to improve the biomechanical properties of the
skin,
-IMPROVING (BLOOD) CIRCULATION – this is due to the fact that red light dramatically improves
blood circulation and lymph flow (when combined with gentle massage applied during the
procedure), as well as boosts metabolism, thus, allowing skincare products and cosmetic
procedures to deliver better effects,
-BLACK ESSENTIAL PDT – used intentionally, PDT (Photo Dynamic Therapy) increases the
effectiveness of the procedure as dermocosmetics contain special chromophores responding to
specific wavelengths of light. Dermocosmetics BLACK ESSENTIAL CHANTARELLE Laboratory Derm
Aesthetics contain 650nm red monochromatic light photosensitive chromophores and have been
designed for 24 different cosmetic procedures (Appendix no. 1 to this User Manual).
PROG.P01 PROG.P01
OUT.2 : H3-CHROMO TIME:15min LED: R OUT.2 : H3-CHROMO CZAS:15min LED: R
BEFORE TREATMENT CIRCULATION IMPROVEMENT PRZED ZABIEGAMI POPRAWA CYRKULACJI
& BLACK ESSENTIAL PDT & BLACK ESSENTIAL i metoda PDT
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.4
P02. LED: GREEN+R (green LEDs 525nm and red LEDs 650nm)
-LIFTING - this is due to the special properties of red and green light, which stimulate skin’s
metabolic processes – increase the number of metabolic reactions and enhance enzymatic channels
by improving skin hydration. When the two types of light are combined, they deliver perfect skin
lifting effects.
-MOISTURIZING – this is particularly true for green light, as one of the main characteristics of green
light (apart from reducing spots, discoloration, and visibility of capillaries) is the stimulation of
various processes helping improve skin hydration.
-REDUCTION OF WRINKLES – this is due to the fact that green and red light work in synergy,
delivering spectacular effects on the skin.
-IMPROVES SKIN TONE – both red and green light, regulating and stimulating skin regeneration
processes, help dramatically improve the appearance of the skin by reducing wrinkles, uneven skin
tone, dullness and signs of fatigue. The skin becomes smooth, looks healthy and radiant, without
visible signs of degradation.
PROG.P02 PROG.P02
OUT.2 : H3-CHROMO TIME:15min LED: G+R OUT.2 : H3-CHROMO CZAS:15min LED: G+R
LIFTING & HYDRATING EFFECT & SMOOTHING LIFTING & EFEKT NAWILŻENIA & WYGŁADZENIE
FINE LINES & SKIN TONING ZMARSZCZKI & TONUS SKÓRY
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.5
P03. LASER:R LED: G+R (red laser 635nm, green LEDs 525nm and red LEDs 650nm)
PROG.P03 PROG.P03
OUT.2 : H3-CHROMO TIME:15min LASER:R LED: G+R OUT.2 : H3-CHROMO CZAS:15min LASER:R LED: G+R
ANTI-AGEING & ANTI-STRETCH PRZECIW STARZENIOWY & PRZECIW ROZSTĘPOM
& ANTI-INFLAMMATORY & PRZECIW OBJAWOM STANÓW ZAPALNYCH
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.6
P04. LED: BLUE +GREEN +RED (blue LEDs 470nm, green LEDs 525nm and red LEDs 650nm)
-ANTI-REDNESS – this combination is recommended to use to treat and PREVENT SKIN REDNESS as
blue light has strong soothing properties whereas green light reduces the visibility of skin redness
and pigmentation.
-ANTI-COUPEROSE & ANTI-ROSACEA – the ANTI-COUPEROSE & ANTI-ROSACEA effects of the
combination of these types of light are possible as green light helps moisturize and soothe the skin,
reduce the visibility of broken capillaries, discoloration and other similar imperfections, whereas
red light improves blood circulation and deeply regenerates the skin.
PROG.P04 PROG.P04
OUT.2 : H3-CHROMO TIME:15min LED: B+G+R OUT.2 : H3-CHROMO CZAS:15min LED: B+G+R
ANTI-REDNESS & ANTI-COUPEROSE PRZECIW ZACZERWIENIENIOM & PRZECIW NACZYNKOM
& ANTI-ROSACEA & PRZECIW TRĄDZIKOWI RÓŻOWATEMU
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.7
P05. LED: UV +BLUE+RED (ultraviolet LEDs 395nm, blue LEDs 470nm and red LEDs 650nm)
-MIXED SKIN & PORE REDUCTION & MATIFFYING & ANTI-INFLAMMATORY – Blue light soothes the
skin, reduces signs of skin irritation. UV light delivers anti-bacterial and balancing effects on the
skin. Researchers also found other powerful effects of the combination of ultraviolet light, blue light
and red light (!) Strong anti-inflammatory effects, helping treat acne of mature skin (Acne Tarda).
-HYDRATING – the direct indication for red light.
PROG.P05 PROG.P05
OUT.2 : H3-CHROMO TIME:15min LED: UV+B+R OUT.2 : H3-CHROMO CZAS:15min LED: UV+B+R
MIXED SKIN & PORE REDUCTION & ANTI- SKÓRA MIESZANA & REDUKCJA ROZSZERZONYCH PORÓW
INFLAMMATORY & MATTIFYING HYDRATING & PRZECIW ZAPALNE & MATUJĄCE NAWILŻAJĄCE
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.8
P06. LED: UV+BLUE (ultraviolet LEDs 395nm and blue LEDs 470nm)
-OILY SKIN/ACNE PRONE & SEBUM REDUCTION & ANTI-INFLAMMATORY & CALMING – Blue light
soothes the skin and reduces signs of skin irritation while UV light gives anti-inflammatory and
balancing effects. Blue and ultraviolet light help considerably reduce sebum production and soothe
existing inflammation, also in the form of active lesions in clogged sebaceous ducts. As a result of
mixed UV and blue light, the two types of light help reduce the causes and symptoms of
inflammation, delivering calming and soothing effects.
PROG.P06 PROG.P06
OUT.2 : H3-CHROMO TIME:15min LED: UV+B OUT.2 : H3-CHROMO CZAS:15min LED: UV+B
OILY SKIN/ACNE PRONE & SEBUM REDUCTION SKÓRA TŁUSTA/TRĄDZIKOWA & REDUKCJA SEBUM
& ANTI-INFLAMMATORY & CALMING & PRZECIW ZAPALNE & USPOKAJAJĄCE
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.9
P07. LASER:R LED: BLUE (red laser 635nm blue LEDs 470nm)
This program is recommended for patients WITHOUT contraindications against laser biostimulation.
WARNING: When performing procedures using this program, the operator MUST wear laser safety
glasses (delivered together with VIASKINER®).
-SKIN CONDITIONING – Red light helps replenish and restore impaired skin at several levels, both at
a chemical level (stimulating metabolism) and directly, at a physical level – improving blood
circulation and stimulating skin elasticity.
-ACNE-PRONE ADVANCED – this program helps treat severe acne. Blue light calms and soothes the
skin when combined with red light stimulates metabolism and detoxification of skin cells, and thus
brings the inflammation soothing effects much faster, as well as reduces skin redness.
-ANTI-SCARS – the combination of red light and blue light reduces the visibility of scars by softening
scar tissue, helping smooth the scars and even skin tone, improves blood circulation in the scar area
and deeply soothes the skin.
PROG.P07 PROG.P07
OUT.2 : H3-CHROMO TIME:15min LASER:R LED: B OUT.2 : H3-CHROMO CZAS:15min LASER:R LED: B
SKIN CONDITIONING & ACNE-PRONE ADVANCED POPRAWA JAKOŚCI SKÓRY & ZAAWANSOWANY TRĄDZIK
& ANTI-SCARS & PRZECIW BLIZNOM
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.10
P08. LED: UV+BLUE+GREEN+ R (ultraviolet LEDs 395nm, blue LEDs 470nm, green LEDs 525nm and
red LEDs 650nm)
-DEPIGMENTATION & ANTI-SPOT & ANTI-PHOTOAGEING – the combination of UV, blue, green and
red light has been used to provide cumulated powerful effects to reduce spots and skin
pigmentation caused by different factors ranging from skin aging, caused by excessive sun tanning
and photoaging but also these resulting from oxidative changes due to some external and internal
factors.
This program is also recommended to use to reduce the visibility of discoloration appearing in areas
difficult to treat e.g. eye area and eyelids. In these cases, it is recommended to shorten the time of
the procedure to 10 minutes, 5 minutes per eyelid/eye. After the procedure, the patient should
stay in the chair for 10 more minutes WITHOUT OPENING the eyes. During this time, the operator
may put on the eyes pads moistened with appropriate tonic or apply appropriate serum.
PROG.P08 PROG.P08
OUT.2 : H3-CHROMO TIME:15min LED: UV+B+G+R OUT.2 : H3-CHROMO CZAS:15min LED: UV+B+G+R
DEPIGMENTATION & ANTI-SPOT DEPIGMENTACJA & PRZECIW PLAMOM
& ANTI-PHOTOAGEING & PRZECIW OBJAWOM FOTOSTARZENIA
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.11
P09. LED: UV+BLUE +GREEN (ultraviolet LEDs 395nm, blue LEDs 470nm and green LEDs 525nm)
-AFTER TREATMENT – P09 is a program which is a combination of ultraviolet LEDS delivering anti-
bacterial effects, blue LEDs which soothe and calm the skin, and green LEDs helping reduce skin
redness. This program is recommended to apply at the end of each procedure – performed before
applying the last cream. This extra procedure makes the skin look healthy, radiant and relaxed, and
increases antibacterial protection levels of the skin which is particularly important in treatments
with a high level of antibacterial uncertainty e.g. procedures damaging the epidermis and after
deep exfoliation.
-CALMING & ANTI-INFLAMMATORY & REDNESS REDUCTION – Ultraviolet, blue and green light offer
powerful soothing, anti-irritation, anti-inflammatory and anti-redness effects on the skin. The
procedure delivers practically immediate results. P09 can be performed at any stage of any
procedure, if necessary.
PROG.P09 PROG.P09
OUT.2 : H3-CHROMO TIME:15min LED: UV+B+G OUT.2 : H3-CHROMO CZAS:15min LED: UV+B+G
AFTER TREATMENT CALMING & ANTI- NA ZAKOŃCZENIE ZABIEGU USPOKOJENIE &
INFLAMMATORY & REDNESS REDUCTION PRZECIW PODRAŻNIENIOM & REDUKCJA ZACZERWIENIEŃ
¤ ◄►- PROGRAM, M- HEAD, CTRL2 – START ¤ ¤ ◄►- PROGRAM, M- GŁOWICA, CTRL2 – START
Image 7.12
Warranty period for VIASKINER® MesoLED LASER: 24 months. The warranty does not cover any
mechanical damage to the heads and wiring/outlets and electronics damaged as a result of
changing the heads during the operation of the device. Any attempt to connect other devices to
any of the outlets of the heads will damage the control electronics!.
It is prohibited to copy or distribute this User Manual VIASKINER® MesoLED LASER or any part of it.
The device should be discarded as electronic waste. The device does not contain any hazardous materials
that can threaten human health.
If it is necessary to replace the power cord – it must be replaced only with a cord of the same class and
with the same parameters.
Never disconnect the heads during the operation of VIASKINER® MesoLED LASER! This may damage the
electronic system of the device. The heads can be connected/disconnected ONLY when the device is off.
To power VIASKINER® MesoLED LASER, It is recommended to use a cord with a safety switch.
WARNING: The lasers used in the CHROMO3 head of VIASKINER® MesoLED LASER have a 3R safety class
– you should never look directly into the light/ beam of the laser neither with your eyes nor using any optical
device. The operator, to perform procedures using P03 and P07 programs with the CHROMO3 head, must put
on special certified safety glasses designed for this class of lasers. The patient must also wear special safety
pads.
To prevent fire, electric shock or device damage, the device must not be exposed to rain and
moisture, It must be protected from water and other liquids. Do not put any object filled with liquids on the
device, e.g. vases.
Use only the accessories recommended in this User Manual.
Do not repair the device yourself.
Warning! Do not put the device in a confined space e.g. bookshelf or a built-in cabinet. Make sure
the device has appropriate ventilation. To prevent electric shock or fire which may result from the
overheating of the device, make sure that no materials, e.g. curtains, cover the ventilation holes.
Do not cover the ventilation holes of the device with newspapers, tablecloths, curtains, etc.
Do not put any naked flame sources e.g. candles on the device.
On the VIASKINER® MesoLED LASER casing and on the CHROMO3 head power cord, there are the following
warning and information signs: