Basics of Calibrating Pressure Transmitters

Pressure transmitters need to be calibrated on a regular basis for


maximum performance. When do you do it? How do you do it? And
who does it?

By Keith Riley, Ehren Kiker and Duane Muir, Endress+Hauser

• Are there any local, national, safety


or environmental regulations that
must be observed?
• What is your reason for requiring
calibration: quality, safety or
standard maintenance?

Process conditions:
• Is there a homogeneous process
fluid with a stable pressure/
temperature?
• Will the process conditions
fluctuate significantly?
• Is there risk of buildup, corrosion
or abrasion to the pressure
transmitter?
• Will heavy vibration be present?

Pressure transmitters used in the As with most things in life, there is Ambient conditions:
process industries are very durable no “one size fits all” answer. However, • Will the pressure transmitter
and reliable instruments. Even so, there are simple best-practice be installed in a well-controlled
they still require periodic maintenance guidelines, which can be modified to environment with low humidity,
and calibration to ensure optimal fit specific applications. This article normal/stable temperatures, and
performance. This is an area of helps answer the basic questions few contaminants such as dust or
confusion for many, with these and facing process plant personnel with dirt?
other questions typical: regard to calibration. • Is an outdoor transmitter exposed to
widely varying weather conditions
• Are we calibrating our transmitters How Often? or high humidity?
too often, resulting in excessive
downtime and unnecessary mainte- Each process plant has to determine If you have no significant history or
nance expense? correct calibration intervals based regulatory requirements to guide
• Are we calibrating our transmitters upon historical performance and you in developing your calibration
too infrequently, resulting in quality process-related requirements. Factors procedures, a good place to start is
issues and possible loss of product? you need to consider that may with the following general guidelines.
• Are we calibrating our transmitters influence this decision are:
correctly?
• Direct mounted pressure Many make the mistake of adopting The test equipment you intend to use
transmitters installed inside in a the manufacturer’s reference should be traceable to the National
controlled environment on a process accuracy as their calibration target. Institute of Standards and Technology.
with stable conditions should be Unfortunately, this means they will As a general recommendation, your
calibrated every four to six years. have a MPE that is too tight, with a reference equipment should be at least
• Direct mounted pressure high rate of non-conformance in their three times more accurate than the
transmitters installed outside on calibration process. In the worst case pressure transmitter being calibrated.
a process with stable conditions with a very tight tolerance MPE, it
should be calibrated every one to may not be possible for their field or Performing the Calibration
four years, depending upon ambient lab test equipment to calibrate some
conditions. of their transmitters. Once your calibration interval and
MPE have been established, you
If a remote diaphragm seal is A manufacturer’s reference accuracy are ready to perform the actual
employed on a pressure transmitter, is based upon tightly controlled calibration procedure on your
the calibration interval should be environmental conditions seldom pressure transmitter. The best
reduced by a factor of two; i.e., a if ever duplicated in a plant practice recommendation from
four to six year interval is reduced to environment. Using that reference Endress+Hauser is:
two to three years. This is because a accuracy for a calibration target also
remote diaphragm seal will employ fails to take into account the long term 1. Mount the transmitter in a stable
more fill fluid than a direct mounted stability of the instrument. fixture free from vibration or
configuration. Consequently it will movement
experience more mechanical stress Over time, all instruments will 2. Exercise the sensor/membrane
from process or ambient temperature experience slight accuracy degradation before performing the calibration.
fluctuations. Most remote diaphragms due to aging and simple wear and This means applying pressure and
are flush faced where the diaphragm/ tear on mechanical components. raising the level to approximately
membrane is susceptible to physical This needs to be considered when 90% of the maximum range. For
damage (dents or abrasions) that can establishing the MPE. In general, a 150 psi cell that would mean
cause offset or linearity issues. unless there are mitigating pressurizing it to 130-135 psig.
circumstances, it is better to set a Hold this pressure for 30 seconds
If the process regularly experiences reasonable MPE achievable with then vent. Your overall results
significant pressure swings or over standard field and lab test equipment. will be much better than if you
pressurization events, reducing the calibrate “cold.”
calibration interval by a factor of two Test equipment starts with an accurate 3. Perform a Position Zero
is a good rule of thumb. pressure source to simulate the Adjustment (zero the
transmitter input. The corresponding transmitter). This is important
How Accurate? output is measured with a multimeter because the fixture used for
for a 4-20mA transmitter, or with calibration may be different than
How good is good enough? In a specialized device for smart how the transmitter is mounted
other words, what is the Maximum transmitters with digital outputs in the process. Failing to correct
Permissible Error (MPE) for your such as HART, Foundation Fieldbus, for this by skipping this step can
calibration? Profibus or EtherNet/IP. result in non-conformance.

Recommended calibration intervals (years)

2
4. Begin the calibration procedure. transmitter—even on new units from maintaining the documentation.
Typically this means three points the manufacturer. A sensor trim Repeatability is the key and in the
up (0% / 50% / 100%) and then corrects the digital reading from the
three points down. The 4-20mA sensor after the A/D conversion. world of calibration, if it isn’t properly
output should be 4mA, 12mA Performing a sensor trim on a new documented, it didn’t happen.
and 20mA at the three points transmitter is essentially a single
(or the correct digital values for point calibration under current plant Occasionally there are some
a smart transmitter). Each test environment conditions, as opposed
point should be held and allowed to sticking with the original factory calibrations that cannot be performed
to stabilize before proceeding to calibration. in a standard maintenance shop
the next. Normally that should by maintenance technicians. For
take no more than 30 seconds. Factory calibrations of pressure these cases, an ISO17025 accredited
More points can be used if you transmitters are performed in a
require a higher confidence in the tightly controlled environment and organization is required. Not only can
performance of the instrument. incorporate up to as many as 100 an ISO17025 accredited organization
5. Compare the results of your test points. Performing a sensor trim perform more stringent calibrations,
pressure transmitter to your on a new pressure transmitter under they provide other value as well:
reference device. field conditions will result in a unit
6. Document the results for your that operates at less than optimal
records. capacity. A sensor trim should only be • Accredited labs can simplify the
performed by a qualified technician calibration audit process.
The calibration should be performed under the manufacturer’s guidance. • The process and methodology used
in as stable an environment as
possible because temperature and Who Should Perform the by an accredited lab is extremely
humidity can influence the pressure Calibrations? repeatable, thus producing a high
transmitter being tested as well as the level of confidence in the results
pressure reference. Even with the sophisticated calibration from an auditor’s perspective.
and reference equipment currently
If the results of your calibration are available, there is no substitute for • Annual audits of the accredited
within the MPE, do not attempt to a properly trained technician when lab ensure they are consistently
improve the performance of the it comes to calibrating pressure performing at a high level for their
transmitter. transmitters. Not only does the registered scope of work.
technician need to be trained on
One mistake many end users make the mechanics of the calibration
is to regularly perform a sensor process, he or she also needs to be
trim adjustment of their pressure equally qualified in completing and

3
Summary

The “correct” calibration cycle for a pressure transmitter


will depend on the purpose of the calibration and the
application. The same pressure transmitters employed in
different operating units or processes at the same plant
may require different calibration intervals.

Even more important than the calibration interval of the


instrument are:

• Establishing correct and realistic MPEs


• Following correct calibration procedures
• The training of the person performing the calibration
• Proper documentation of calibration results.

Following these guidelines and using judgment based on


actual plant operation conditions will help establish proper
calibration practices, saving money while maintaining
acceptable performance.

About the Authors

Keith Riley joined Endress+Hauser in 2008 as a level product manager. He was responsible for the business
development and technical direction of level products. After four years, he became the national product team leader
for pressure and temperature products. In this role, Riley markets the temperature and pressure products and
oversees the strategic direction for the U.S. temperature and pressure product business team.

Ehren Kiker has spent the past 20 years consulting on process instrumentation and control as well as electrical
control and distribution equipment. He received his bachelor’s degree in industrial and systems engineering from
the University of Florida and his MBA from the University of Houston. Kiker is the national product marketing
manager for Endress+Hauser U.S. He is responsible for technology application, marketing, and business
development for a wide range of pressure and temperature products.

Duane Muir is currently a technical product specialist for level and pressure products at Endress+Hauser with 20
years of experience in application, installation, calibration, and troubleshooting instrumentation. He received his
BS at Indiana State University in electronics technology. In his career, Muir has specialized in five different areas of
instrumentation technology and has multiple certifications from Endress+Hauser.