2.
1.5
Intensity (a.u.)
1.0
0.5
0.0
-0.5
0.00 0.05 0.10 0.15 0.20 0.25
Concentration (mg/L)
Zero: Ex 350.00 ; Em 450.00 ( 0.896 )
Zero: Ex 350.00 ; Em 450.00 ( 0.987 )
Concentration Report
Report time: 8/28/2019 [Link] PM
Software version: 1.1(132)
Operator:
Instrument Parameters
Instrument Cary Eclipse
Instrument Serial Number FL1108M003
Data mode Fluorescence
Excitation wavelength (nm) 350.00
Emission wavelength (nm) 450.00
Ex. Slit (nm) 5
Em. Slit (nm) 10
Excitation filter Auto
Emission filter Open
Ave time (sec) 0.10000
PMT voltage (V) Medium
Standard replicates 3
Standard averaging OFF
Sample replicates 1
Sample averaging OFF
Weight and volume corrections OFF
Fit type Linear
Min R² 0.95000
Concentration units mg/L
Comments:
Calibration
Collection time 8/28/2019 [Link] PM
Standard Concentration F Mean SD %RSD Readings
mg/L
________________________________________________________________________
Std 1 -0.112
-0.001
0.050 -0.013 0.0934 >100 0.073
Std 2 0.642
0.537
0.100 0.573 0.0604 10.5 0.539
Std 3 0.732
0.979
0.150 0.826 0.1335 16.1 0.767
Std 4 1.213
1.112
0.200 1.241 0.1452 11.7 1.398
Std 5 1.592
1.639
0.250 1.612 0.0241 1.49 1.604
Calibration eqn Int = 7.83542*Conc -0.32774
Correlation coefficient 0.98684
Calibration time 8/28/2019 [Link] PM
Analysis
Collection time 8/28/2019 [Link] PM
Sample Concentration F Readings
mg/L
_________________________________________________
Sample 1 101.204 O 792.646
Sample 2 0.358 O 2.477
Sample 3 0.154 0.881
Sample 4 0.110 0.536
Sample 5 0.120 0.610
Read sequence cancelled
Results Flags Legend
U = Uncalibrated O = Over-range concentration
N = Not used in calibration R = Repeat reading
@ = Over-range signal