LIGHT METER

Model: LX-IOIA

-
1\

LIGHT
B C

Your purchase of this UGHT METER marks a step forward for you
into the field of precision measurement. Although this UGHT
METERis a complex and delicate instrument, its durable structure
developed. Please read the following instructions carefully and
always keep this manual within easy reach.

OPERATION MANUAL
___ --'I~

TABLE OF CONTENTS
1. FEATURES
1. FEATURES 1
• Sensor used the exclusive photo diode & color
2. SPECiFiCATIONS 1 correction filter, spectrum meet C.LE. photopic.
2-1 General Specifications 1 • High accuracy in measuring.
2-2 Electrical Specifications 2 • Wide measurement, 3 ranges: 2,000 LUX, 20,000 LUX,
& 50,000 LUX .
3. FRONT PANEL DESCRiPTION 3 * Separate LIGHT SENSOR allows user to measure the
3-1 Display 3 light at an optimum position.
3-2 Off/On Switch 3 • 151 circuit provides high reliability and durability.
3-3 Range Switch 3 • LCD display allows clear read -out even at high
3-4 Battery Compartment/Cover 3 ambient light level.
3-5 light Sensor 3 • Pocket size, easy to carry out & operation.
3-6 Zero Adjust VR 3 • Compact, light weight and excellent operation.
• Built-in low battery indicator.
4. MEASURING PROCEDURE 4

5. MEASURING CONSiDERATION 5 2. SPECIFICATIONS


6. BA TIERY REPLACEMENT.. 5 2-1 General SL ecifications
Display 13mm (0.5') LCD, 3 1/2 diqits,
Max. indication 1999.
Measurement o to 50,000 Lux, 3 ranges
Sensor The exclusive photo diode & color
correction filter.
Over Input Indication of "1".
Displav
Power Supply 006P DC 9V battery.
Power Current Approx. DC 2.7 mA.
Zero adjustment External VR on front panel
Operatinq Temp. o °C to 50°C (32°F to 122°Fl. 3. FRONT PANEL DESCRIPTION
Operating Less thanb 80% RH.
Humidity
WeiQht 195 Q I 0.43 LB (includinq battery).
Dimension Main instrument:
131 x 70 x 25 mm ( 5.2x2.8xl.0 inch)
/"
Sensor probe
82 x 55 x 7 mm ( 3.2 x 2.2 x 0.3 inch l. II u=d;~-3-5
Standard Instruction ManuaL ..................... 1 PC
.. 3- 1
••• 1000
Accessories Sensor orobe ................................ 1 PC
I
2 2 Electrical Specifications (23 + 5°C) Inn..
Resolution Accuracy 3-2
RanQe
0- 1,999 Lux 1 Lux ~ 0- - 3-6
10 Lux + (5% + 2d) 3 J
2,000 19,990 Lux
20,000 50,000 Lux 100 Lux

~
r_._ 3 4

Fig. 1

3-1 Display
3 - 2 Off/On Switch
3-3 Range Switch
3-4 Battery Compartment/Cover
3-5 Light Sensor
3-6 Zero Adjust VR

2 3
5. MEASURING CONSIDERATION
4. MEASURING PROCEDURE
~ TIlt:" R.lIJge B "isdesigl1cd to mcesurc 2.()()() to 19,990 LlH' only:
'# The" Rauge C"isdesigucd to rueasure 20,(){)()/050,000 Lux only.
1) Slide the" Off/On Switch" ( 3 - 2, Fig. 1 ) to the" On "
.• If the measured value < 2.000 Lux. it should select (0 " Range A ':
position.
'# If the measured value lVitJu"n2.(){)() (0 19.990Lux. it should select to" Rrmge B ":
2) Select the max. range (2,000 Lux, 20,000 Lux or 50,000
'# lfthe measured value 2 ZO.OOOLux; it should select the Range C ":
It

Lux) on the" Range Switch" (3-3, Fig.I ).


3) Hold the" Light Sensor" (3-5, Fig. 1 ) by hand & face
6. REPLACEMENT OF BATTERY
the measuring light opposite. then the Display ( 3 -1,
Fig. 1 ) will show light values directly.
1) When the left corner of LCD display show" LO BAT ",
4) Zero adjustment:
it is necessary to replace the battery. However, in -spec
Due to drift of environment temperature value, battery
measurement may still be made for several hours after
power change or, meter used for a long time or other
low battery indicator appears before the instrument
reasons ... The display value may exist not zero value (
become inaccurate.
few digits) after blanking the" Light Sensor" (3-6, Fig. 1 ).
2) Slide the Battery Cover (3-5, Fig. 1) away from the
General speaking those not zero value will not effect
instrument and remove the battery.
the measurement typically. However if intend to make
Replace with 9V battery and reinstate the cover.
the precision measurement, the following zero
Make sure the battery cover is secured after change the
adjustment procedures should be executed as :
battery.
Select the • Range Switch' ( 3-3, Fig. 1 )
to the' 2000 Lux • range. Blank the
• Light Sensor' (3-5, Fig. 1 ). Use "-
convenient screw driver to adjust • Zero
adjust VR' (3-6, Fig. 1 ) until the
disnlav reach the' ZERO' value.

5
0202-LX-101A-ZERO
4
,........ I _~_ •••
.J~

ANEMOMETER

Your purchase of this ANEMOMETER


marks a step forward for you into the field
of precision measurement. Although this
ANEMOMETER is a complex and delicate
instrument, its durable structure will allow
many years of use if proper operating
techniques are developed. Please read the
following instructions carefully and always
keep this manual within easy reach.
TABLE OF CONTENTS 1. FEATURES

1. FEATURES 1 * The portable anemometer provides fast, accurate


readings, with digital readability and the
2. SPECIFICATIONS 2 convenience of a remote sensor separately.
2-1 General Specifications 2 * Multi-functions for air flow measurement: mis, krn/h,
2-2 Electrical Specifications 3 ft/rnin, knots.
* Low-friction ball-bearing design allows free vane
3. FRONT PANEL DESCRIPTION 4 movement, resulting in accuracy at both high & low
3-1 Display 4 velocities.
3-2 Off/On/Hold Switch 4 * A sensitive balanced vane wheel rotates freely in
3-3 Function ( mis, krn/h, ft/rnin, knots) Switch. 4 response to air flow.
3-4 Battery Compartment/Cover 4 * Conventional twisted vane arms, always a source of
3-5 Vane Probe handle .4 unreliability have been eliminated.
3-6 Vane Probe head 4 * DATA HOLD function for stored the desired value on
display. Large LCD display, easy to read.
4. MEASURING PROCEDURE s * LCD display for low power consumption & clear
read-out even in bright ambient light condition.
5. REPLACEMENT of BATTERY 5 * Used the durable, long-lasting components,
including a strong, light weight ABS-plastic housing
case.
* Compact housing cabinet, easy to carry out.
* Built-in low battery indicator.
* Wide applications: use this anemometer to check air
conditioning & heating systems, measure air
velocities, wind speeds, temperature ... etc.

..~
.'.
2. SPECIFICATIONS 2-2 Electrical Specifications ( 23 % 5 't )

CMeasure- iRange TResolutio Accuracy


r'-I m:..::e::.:n.:.:t'---~-tl~~_ ----1
2-1 General SpeCifications ~ 10.4- 30.0 m/s 10.1 m/s ± (2%+0.2m/s)
Display 18 mm ( 0.7" ) LCD ( Liquid Crystal
1.4 - 108.0 km/h 0.1 km/h ± (2%+0.8km/h)
Dis lay~/,-,,2,-,d~i,,-,i::::ts,,-.
,.,..--~~~~_
m/s (meters per second), 0.8 - 58.3 knots 0.1 knots ± 2%+0.4knots
krn/h (kilometers per hour), ft/min 80 - 5910 ft/min 10 ft/rnin ± @o+40 ft/rnin)
ft/rnin (feet/per minute), m/s - meters per second krn/h - kilometers per hour
knots (nautical miles per hour), ft/min - feet/per minute knots - nautical miles per hour
Data hold. mph - miles per hou_r~~_ ~national knot)
Operating -0 to 50 ·C (32 to 122 OF). Remark,'
Te~ratur~ Above specification are tested under the environment RF Field
Operating -tess than 80 % RH. Strength less than 3 VIM & frequency less than the 30 MHz only.
Humidity __ _--,-~,--~~~~..,--~_

!Air Velocity
Sensor
Structure
Conventional twisted vane arms and
low-friction ball-bearing design.

PoOwWeerr
SupplOo6P DC 9V battery ( heavy duty type ).
Approx. DC 9 mA.
Consumption
tP J
Wei~ 325 g/O.72 Ib ~ including batteiY -).~- ~
Dimension Instrument 168 x 80 x 35 mm
f--
6.6 x 3.2 x 1.2 inch
Sensor Head Round 72 mm Dia.
Standard Instruction Manual 1 Pc.

[Accessories ~",sc=e,-,ns-,-o"-r",p,-,r""o=be.=..","",""-""-""-"",",,,",,,",,,",,,,",,-"
ICarrying case ,.., -,l,-P,-C=c.,--~_
1 Pc.

2 3

".'.!,:
3. FRONT PANEL DESCRIPTION 4. MEASURING PROCEDURE
1) Select the" Off/On/Hold Switch" ( 3-2, Fig. 1 ) to the
" On " position.
2) Select the" Function Switch" ( 3-3, Fig. 1 ) to the
" m/s ". "km/h ", " ft/min " or " knots" position
according to the measuring requirement.
3)Hold the" Vane Probe Handle" ( 3-5, fig. 1) by
hand and let the "Vane Probe Head" ( 3-6, Fig. 1 ) is
opposite to the measuring air flow source, then the
Display ( 3-1, Fig. 1) will show air velocities directly .
.3- 2-++---iFi
.3-3-++--L~ Measuring Consideration:
The yellow dot mark on the sensor head indicates
3-5
the "yellow dot mark " need to face against the
direction of air flow.

4)During the measurement, it will hold the display


values if select the" Off/On/Hold Switch" ( 3-2,
Fig. 1 ) to the" Hold" position.
L-----3-4
S. REPLACEMENT OF BATTERY
Fig. 1 l)When the left corner of LCD display show" BAT ".
it is necessary to replace the battery. However,
in-spec. measurement may still be made for
3-1 Display several hours after low battery indicator appears
3-2 Off/On/Hold Switch before the instrument become inaccurate.
3-3 Range ( mis, km/h, ft/min, 2) Loose the" Battery Cover Screw" ( 3-4, Fig. 1 ),
knots/Temp ) Switch slide the battery cover away from the instrument
3-4 Battery Compartment/Cover and remove the battery.
3-5 Vane Probe handle 3) Replace with 9V battery ( heavy duty type) and
3-6 Vane Probe head reinstate the cover.
4) Make sure the battery cover is secured with the
screw after changing battery.

4
5
1407-AM4201
SOUND LEVEL
METER

Your purchase of this DIGITAL SOUND LEVEL


METER marks a step forward for you into the field of
precision measurement. Although this SOUND LEVEL
METER is a complex and delicate instrument, its
ruggedness will allow many years of use if proper
operating techniques are developed. Please read the
following instructions carefully and always keep this
manual within easy reach.
TABLE OF CONTENTS 1. FEATURES

1. FEATURES 1 * Large LCD display, easy to read.


• Frequency weighting networks are designed to meet the
2. SPECIFICATIONS 1 lEC 61672 Class 2.
• A & C weighting networks are conformity to standards.
3. FRONT PANEL DESCRIPTION .4 • Time weighting (FAST & SLOW) dynamic characteristic
3-1 Electric condenser microphone 4 modes.
3-2 Display 4 * ACIDC output for system expansion.
3-3 Power switch & Output type selector 4 * Build in CALmRATION VR on front panel.
3-4 A/C weighting & Calibration selector .4 • Condenser microphone for high accuracy & long-term
3-5 Time Weighting ( Fast/Slow )/Max. hold stability.
selector 4 • Max. Hold function for stored the maximum value on
3-6 Range selector .4 display.
3-7 Signal output terminal. .4 * Warning indicator for over and under load.
3-8 Battery Compartment/Cover .4 • LCD display for low power consumption & clear
3-9 Range upper/lower indicator 4 read -out even in bright ambient light condition.
3 -1 0 Calibration VR ( Accuracy adjust VR ) .4 • Used the durable, long-lasting components, including
a strong, light weight ABS-plastic housing case.
4. MEASURING PROCEDURE 5 • Small and light weight design allow one hand operation.
+ Low battery indicator.
5. MEASURING CONSiDERATION 6
2. SPECIFICATIONS
6. SIGNAL OUTPUT 6
Display 18mm (0.7") LCD (Liquid Crystal Display),
7. REPLACEMENT of BATTERy 7 3 1/2 dlqits.
Function dB (A & C frequency weighting),
8. CALIBRATION 7 Time weighting(Fast, SlOW),
Max. hold, AC & DC output.
9. FREQUENCY WEIGHTING CHARACTERISTICS Measurement 35 to 130 dB, input signal only,
OF A & C NETWORKS 9 Ranae 3 ranqes. -
Resolution 0.1 dB.
10. TIME WEIGHTING (FAST & SLOW)
CHARACTERiSTiCS 9
Accuracy Frequency weighting meet IEC 61672 Time Fast - t= 200 ms
(23 ± 5°C) Class 2, calibrating input signal on 94dB weighting * "Fast" range is simulated the human ear
( 31.5 Hz to 8 kHz), then the accuracy (Fast & Slow) response time weiahtino.
of A weighting is specified as following: Slow - t = 500 ms
31.5 Hz - ± 3.5 dB, 63 Hz - ± 2.5 dB, * "Slow" range is easy to get the average
125 Hz - ± 2.0dB, 250Hz - ± 1.9 dB, values of vibration sound level.
500 Hz - ± 1.9dB, 1 kHz - ± 1.4 dB, Calibration * Build in external calibration VR on front
2 kHz - ± 2.6 dB, 4 kHz - ± 3.6 dB, panel.
8 kHz - ± 5.6 dB * Calibrated via external SOUND
Characteristics of A & C. CALIBRATOR.
A weighting : * Calibrated via internal 94 dB/1 KHz
The characteristic is simulated as "Human aenerator.
Frequency Ear Listing" response. Typical, if making Output Signal AC output:
Weighting the environmentalsound level measurement, AC 0.5 Vrms corresponding to each range
Network always select to A weiahtina. step.
C weighting DC output:
The characteristic is near the "FLAT" DC 0.3 to 1.3 VDC, 10 mV per dB.
response. Typical, it is suitable for Out put impedance - 600 ohm.
checking the noise of machinery (Q.C. Output 3.5 mm Dia. phone output terminal is
check) & knowing the sound pressure level of terminal provided for connection with analyzer,
the tested eauioment. level recorder, tape recorder.
Freauencv 31.5 Hz to 8,000 Hz. Operating O°Cto 50°C ( 32°F to 122°F).
Calibrator B & K (Bruel & Kjaer) : Temperature
Multi-fuction acoustic calibrator, Max. 90% RH (0° to 35°C).
Operatirp
model : 4226. Humlctl
Microphone Electric condenser microphone. Power Suoolv 006P DC 9V battery ( heaw dutVtVPe 1.
Size of 1/2 inch standard size. Power Approx. DC 6 mA.
microohone Consumption
Range 30 to 80 dB, 50 to 100 dB, 80 to 130 dB, Dimension 245 x 80 x 35 mm ( 9.6 x 3.2 x 1.4 inch ).
selector 50 dB on each step, with over & under Weiaht 300 g/0.66 LB ( includina battery ).
range indicating. Standard Instruction Manual ............... 1 PC.
* 30to 80 dB range are operated from 35 Accessories Calibration screw driver .........1 PC.
80 dB ottica/lv. Optional 94 dB Sound Calibrator, model :SC-941
Calibrator 94 dB Sound Calibrator, model: SC-940A
2 3
3. FRONT PANEL DESCRIPTION 4. MEASURING PROCEDURE
-I--~-3-1
1) Slide the" NC weighting selector" ( 3 -4, Fig. 1 ) to
" A " or" C" position for sound level measuring.

NOTE:
a. The characteristic table of A, C weighting,
please ref page 9.
b. The characteristic of A weighting is simulated
as the "Human Ear Listening" response.
Typically, if making the environmental sound
3-2
level measurement, always select the A
weighting typically.
~~~~~3-3
3-4
c. The C weighting characteristic is near the
r>-~---l-I- 3··10 "FL4 T" response. Typically, it is suitable for
3-5
checking the noise of machinery ( Q. C. check)
'-~~~++- 3-6 & knowing the real sound level of the tested
3-7 eauioment.

l':::====c=~_3-8 2) Determine proper measuring range by selecting


the" Range selector" ( 3 -6, Fig. 1 ) to minimize the
tolerance of readout. When left corner of LCD
show" ..•. "indicator, (3-9, Fig. 1 ), it shows the
3-1 Electric condenser microphone dB range selection is upper or lower setting. Slide
3-2 Display range selector to other range for measuring.
3-3 Power switch & Output type selector 3) According on various measuring sound source,
3-4
3-5
A/C weighting & Calibration selector
Time Weighting ( Fast/Slow) /Max. hold selector
t select the" Time Weighting selector" (3-5, Fig. 1 ) to
" Fast" or " Slow" position.
3-6
3-7
Range selector
Signal output terminal
\ 4) Hold the instrument in hand and point the
microphone at measured noise source, the sound
3-8 Battery Compartment/Cover level will be displayed on " dB " ( decibel) unit.
3-9 Range upper/lower indicator
3-10 Calibration VR (Accuracy adjust VR)

4 5
5) Max. hold - 7. REPLACEMENT OF BATTERY
During the sound level measurement, if intend to
store the maximum ( peak) value on display, please 1) When the left corner of LCD display show "BAT", it
slide the" Time weighting! Max. hold selector" ( 3 - 5, indicate a normal battery output of less than 6.5 V
Fig. 1 ) to the "Max. hold" position. - 7.5 V. It is necessary to replace the battery.
However, in - spec. measurement may still be
NOTE: made for several hours after LOW BATTERY
a. When measure long-term stability under INDICATOR appears before the instrument
slowly varying noise environment, please use the become inaccurate.
Max. hold function to read the maximum values. 2) Slide the" Battery Cover" (3-8, Fig. 1) & take the
b. Slide the selector to "Fast" or "Slow" position will battery away from the battery compartment.
cancel the max. hold values. 3) Replace with 9V battery ( heavy duty type) and
reinstate the cover.
5. MEASURING CONSIDERATION
8. CALIBRATION
1) Please should select the proper weighting
selector ( A weighting or C weighting ). The sound level meter is built in the" Internal calibration VR/
The A weighting selector will be engaged normally. Accuracy Calibration VR " ( 3 -10, Fig. 1 ) on the front
2) Please select proper measurements range to panel. Please according the following procedures to
minimize the tolerance of readout. calibrate the instrument accurately, if it is necessary.
3) Please don't keep or operate the instrument at
high temperature & humidity environment for a 8-1 Calibrated via external SOUND CALIBRATOR
long period. 1) Prepare the" SOUND CALIBRATOR, model:
u SC-941 " ( optional) or model: SC-940A ( optional ).
4) Keep microphone dry & avoid serious vibration.
Power on the Sound Calibrator & plug calibrator
6. SIGNAL OUTPUT output into the Electric condenser micro - phone
( 3 -1, Fig. 1 ) of the Sound Level Meter.
The instrument is provided an " 3.5 mm Dia. phone 2) Slide the Range selector (3-6, Fig. 1) to" 50 - 100
output" ( 3 -7, Fig. 1 ) terminal for connection with dB u position.
analyzer, level recorder, tape recorder, controller 3) Slide" Time Weighting selector" (3-5, Fig. 1 ) to
...etc. Slide" Power switch & Output type selector" " SLOW" position .
( 3-3, Fig. 1 ) to AC output or DC output according
the user requirement.

6 7
4) Slide the" NC weighting & Calibration 9. FREQUENCY WEIGHTING
selector" (3-4, Fig. 1 ) to " A weighting" position. CHARACTERISTICS OF A & C
5) Carefully adjust the" Calibration VR " (3-10, Fig. 1 ) NETWORKS
with" - " screw driver, until the display read within
" 94.0 ± 0.2 " dB. Frequency A Weighting C Weighting Tolerance
Hz Charac. Charac. (IEC 61672
8-2 Calibrated via internal 94 dB generator class 21
The sound level meter is built in the internal standard 31.5 -39.4 dB -3 dB + 3.5 dB
94 dB/1 KHz square wave generator for the purpose to 63 -26.2 dB -0.8 dB + 2.5 dB
calibrate some certain amplifier circuit. Please 125 -16.1 dB -0.2 dB + 2.0 dB
250 -8.6 dB OdB + 1.9 dB
according the following procedures to calibrate the
500 -3.2 dB OdB + 1.9 dB
instrument before making operation, if the instrument 1k OdB OdB + 1.4 dB
not in use for a long time. 2k +1.2 dB -0.2 dB + 2.6 dB
4k +1 dB -0.8 dB + 3.6 dB
1) Slide" Range selector" (3-7, Fig. 1) to" 50 - 100" 8k -1.1 dB -3 dB + 5.6 dB
position.
2) Slide" Fast/Slow selector" (3-6, Fig. 1 ) to "SLOW" 10. TIME WEIGHTING (FAST & SLOW)
position. CHARACTERISTICS
3) Slide" NC weighting & Calibration selector" (3-4,
Fig. 1 ) to " 94 dB CAL" position. Time Weighting Max. response Tolerance
4) Carefully adjust the" Calibration VR " ( 3 -10, Fig. 1 ) Charac. ref. continuos (IEC 61672class 2)
with" - " screw driver, until the display read within signal
" 94.0 ± 0.2 " dB. F (Fast) - 1.0 dB + 1 dB,
- 2 dB
S-( Slow I - 4.1 dB + 2dB
NOTE:
If intend to make calibration accurately, it is
recommended to use the calibrating procedures
of aboe 8-1 "Calibrated via external SOUND
CALIBRATOR n

8 9 0005-SL-4001
-
STATIC CHARGE METER (Digital) Eltech-03

USER MANUAL

The Static Charge Meter is a non-contact, hand held, light weight static monitoring and
locating instrument. It utilises a precision LCD display to indicate the magnitude and
polarity of the electrostatic charge.

The instrument has good directional sensitivity and is ideal for measuring electric charges
encountered in critical electronic assembly and storage areas as well as Antistatic Work
Stations. It can also be used for Plastics, Paper, Textiles, Food, Explosives and
Pharmaceutical Industries.

The unit is gives direct readings of electrostatic potentials on any surface, when heid at a
distance of 50 mm. However, virtually any electrostatic potential can be measured by
proportionately increasing the distance from the meter to the charged object.

The use of very low current circuitry results in a long battery life of over 6 months under
normal use.

The static meter incorporates state-of-the-art circuitry, to enable measurements to be made


for as long as 30 seconds without significant drift. In practice, measurement time is usually
less than 10 seconds.

OPERATING INSTRUCTIONS ARE GIVEN ON THE BACK OF THE INSTRUMENT

It is important to note that the accuracy of measurement depends on the following points.

1. The size and shape of the object.

2. The magnitude of the average charge distribution over the object's surface.

3. The predominating polarity of the charge since both polarities can be present at the
same time.

. 2
.. 2 ..

MAINTENANCE:
If used with care, this static charge instrument will provide many years of trouble-free
service. If the instrument fails to operate, first check the battery. If it still does not operate
return it to the factory for service. It does not have any user serviceable parts. It can be
cleaned with a damp cloth. Never use solvents as they may damage the plastic case.

CAUTION:
Avoid contact between sensor plate and charged object or personnel. Voltages in excess of
30V placed directly on the sensor may permanently damage the sensor amplifier.

SPECIFICATIONS

Indicating range : ± 10 KV at 5 cms from object.


Resolution : LSB 10 V
Display : 3.5 Digit LCD display.
Accuracy : 10% of reading
Test Actuation : Push On - Push Off Switch
Power Supply : 9 V battery.
Low Battery : Indication on Display at < 7.5V
Dimensions in mm : 125 x 75 x 29
Weight : 125 gms.
Calibration : Recommended once every 12 months.
Warranty : One year.
Accessories : Carrying case
Eltech Industries Unit No. A-417, Diamond Ind. Est.,
Ketki Pada Rd., Nr. Toll Naka,
Off. W.E. Highway, Dahisar (El,
Mumbai - 400 068. India
Tel: 022-64402868
Email: eltind@gmail.com
CALIBRATION CERTIFICATE
1. Customer Name : MIS. QUALITY QUEST ENG.

2. Product : STATIC CHARGE METER - DIGITAL

3. Model : eltech -03

4. Serial No. : 1349

5. Measurement Range : 0 to ± 10KV DC

6. Tolerance : ± 10% of reading

7. Calibration Conducted on : 19.04.2016

8. Calibration Due On : 18.04.2017

9. Result : Pass

10. Test Conditions

Temperature C : 26

Relative Humidity % : 54

11. New I Old : NEW

The above instrument is calibrated using test and measuring equipment with valid traceability to
National Standards.

Equipment used for Calibration:

1. Microdyne High Voltage AC generator Model 510 SI,No.960733


2..: High Voltage precision Rectifier 81 Nc;GS0733

Traceability: Microdyne Calibration Report NO.36026 dt.01.01.2015 further traceable to


Transcal Cal Report No. TSC/14-15 I 1197001 dt.29.09.2014 & Kusum-Meco ~/~'\~. ~~
133300200 dt. 13.01.2014. , u~-e.
.X- r:
iJJ ~
I-
\ ..J\

( /
IUM fl-v /
Test ~d By
Quality Control Department