Professional Documents
Culture Documents
ORIGINAL ARTICLE
Bicuspid Aortic Valve Anatomy and Relationship
With Devices: The BAVARD Multicenter Registry
A European Picture of Contemporary Multidetector Computed
Tomography Sizing for Bicuspid Valves
BACKGROUND: Sizing for transcatheter aortic valve implantation in Didier Tchetche, MD*
bicuspid aortic valves (BAV) remains controversial. Chiara de Biase, MD*
et al
METHODS AND RESULTS: The aim of the BAVARD (Bicuspid Aortic
Valve Anatomy and Relationship With Devices) retrospective registry is to
capture the sizing ratios used for transcatheter aortic valve implantation
in BAV and analyze the second-generation prostheses geometry
postimplantation. About 101 patients with BAV along with available pre-
and post-transcatheter aortic valve implantation multidetector computed
tomography were compared with 88 tricuspid aortic valves (TAV) patients.
Preprocedural multidetector computed tomography diagnosed type 0
and type 1 BAV in, respectively, 12.9% and 86.1 % of BAV. At baseline,
the ellipticity index was similar between BAV and TAV patients: 1.2±0.1
Downloaded from http://ahajournals.org by on December 9, 2019
https://www.ahajournals.org/journal/
circinterventions
T
patients were compared with 88 TAV patients, enrolled at 2
ranscatheter aortic valve implantation (TAVI) is
centers, in the same period, with available pre- and post-TAVI
an established treatment for intermediate to
MDCT. All patients provided consent for anonymized data ac-
high risk patients with symptomatic aortic ste- quisition and analysis. The registry was approved by each local
nosis (AS).1–3 Bicuspid aortic valves (BAV) are the most
Downloaded from http://ahajournals.org by on December 9, 2019
diameter divided by the perimeter-derived annulus diameter. device comparison and integration of the supra-annular posi-
The prosthesis-ICD ratio was defined as the labeled prosthesis tion of the leaflets in the Medtronic Evolut R prosthesis. The
diameter divided by ICD measured 4 mm above the annulus. prosthetic ellipticity index (Dmax/Dmin) was also measured at
the annulus level and 4, 8, 12 mm above (Figure 2; Video II in
the Data Supplement).
Post-TAVI MDCT Analysis
The depth of implantation was defined as the maximum pro-
trusion of THV into the left ventricular outflow tract; level 0 Statistical Analysis
being the native aortic annulus. The distance from the left For statistical analysis, quantitative parameters were described
main coronary artery to the annulus at baseline was used to using the following descriptive statistics: mean, SD, median,
locate the annular plane postimplant. The mean perimeter- Q1, Q3, extreme values (min and max). In the same way,
derived diameter of THV was measured at the aortic an- qualitative parameters were described using the number
nulus (level 0) and 4, 8, and 12 mm above it. In that region and percentage of each modality. Continuous variables were
of interest, representing the 12 distal mm of a given device, plotted using boxplots and scatter plots. Comparisons be-
all prostheses have a cylindrical configuration, allowing for tween groups were computed using the parametric Student
Downloaded from http://ahajournals.org by on December 9, 2019
Table 1. Baseline Clinical Characteristics and Aortic Root Dimensions only of exploratory nature, and no adjustment to the α-risk
of the Study Populations
despite multiple variables have been done.
Tricuspid Bicuspid The individual ratio between the prosthesis diameter and
(n=88) (n=101) P Value the mean aortic annular diameter provides information about
Age, y 83.1±5.7 78.2±10.1 <0.01 appropriate sizing. A ratio >1 indicates oversizing while a
ratio <1 relates to undersizing.
Men, n (%) 41 (46) 66 (65) <0.01
Statistical analyzes have been computed using SAS Version
Hypertension, n (%) 79 (90) 64 (63) <0.01
9.4 TS Level 1M4 Copyright ©2002–2012 by SAS Institute
Dyslipidaemia, n (%) 50 (57) 37 (37) <0.01 Inc, Cary, NC. A 2-sided P<0.05 was considered statistically
Diabetes mellitus, n (%) 13 (15) 17 (17) 0.69 significant.
Dialysis, n (%) 2 (2) 1 (1) 0.48
Current smoker, n (%) 10 (11) 7 (7) 0.16
RESULTS
Previous MI, n (%) 8 (9) 5 (5) 0.26
Baseline Clinical and MDCT Data
Previous PCI, n (%) 8 (9) 32 (32) <0.01
Previous CABG, n (%) 26 (29) 2 (2) <0.01
The baseline clinical characteristics of patients with tri-
cuspid and bicuspid AS are reported in Table 1. Several
Atrial fibrillation, n (%) 27 (31) 24 (24) 0.3
differences were identified between both cohorts: BAV
Previous stroke/TIA, n (%) 11 (12) 10 (10) 0.57
patients were younger (78.2±10.1 versus 83.1±5.7
Previous pacemaker, n (%) 10 (11) 7 (7) 0.29 years; P<0.01), more frequently males (65% versus
COPD, n (%) 23 (26) 26 (26) 0.91 46%; P<0.01), with higher risk profiles as expressed
BMI, kg/m2 28.1±18.4 28.7±34.3 0.47 by the Society for Thoracic Surgery score for mortality
BSA, m 2
1.7±0.2 2.2±0.1 0.05 (11.3±8.5 versus 7.6±4.4; P<0.01). Several elements
LVEF, % 53.4±13.3 54.3±14.9 0.48
like chronic immunosuppressive drugs, higher cre-
atinin level, concomitant mitral valvular disease, and
NYHA class I–II, n (%) 33 (37) 38 (38) 0.98
atrial arrhythmias partly explain this difference in the
NYHA class III–IV, n (%) 50 (57) 52 (51) 0.46
Society for Thoracic Surgery score. We did not perform
STS score, % 7.6±4.4 11.3±8.5 <0.01 any matching of both cohorts because of the rela-
Annulus maximal diameter, 26.1±2.3 29.9±3.8 <0.01 tively small sizes and the focus of this study on MDCT
mm
outcomes.
Downloaded from http://ahajournals.org by on December 9, 2019
Annulus minimal diameter, 20.7±1.9 23.3±3.3 <0.01 Preprocedural MDCT diagnosed type 0 and type
mm
1 BAV in, respectively, 12.9% and 86.1 % of the bi-
Annulus mean perimeter- 23.6±1.9 26.8±3.1 <0.01
cuspid patients. Only 1 patient (1%) had a type 2
derived diameter, mm
BAV (Figure 1). Baseline MDCT analysis, identified a
Annulus mean area derived 23.2±1.9 26.3±3.0 <0.01
diameter, mm
larger annulus and ascending aorta for the BAV group
(26.8±3.1 versus 23.6±1.9; P<0.01; 36.7±5.4 versus
Ellipticity index 1.2±0.1 1.2±0.1 <0.01
30.4±3.4; P<0.01), with a significantly higher calcium
Calcium score, cm3/ 3038.6±3671.1 5143.1±4730.2 0.09
threshold 450 HU
volume (5143±4730 versus 3038±3671 mm3; P=0.02;
Table 1). The ellipticity index was not significantly dif-
ICD at 4 mm above the na 27±3.1 0.02
annulus, mm ferent between BAV and TAV patients: 1.2±0.1 versus
Ascending Aorta, 3 cm 30.4±3.4 36.7±5.4 na
1.2±0.1, P=0.09. Of note, while the mean perimeter-
above the annulus, mm derived diameter of the aortic annulus of BAV patients
Left coronary height, mm 13.4±3.1 15±3.7 <0.01 was 26.8±3.1 mm, the ICD (measured at +4 mm) was
27.0±3.1 mm, P=ns. Thirty-four patients (33.7%) with
Right coronary height, mm 17.1±3.1 18.3±4.2 0.01
BAV had a concordance between the annulus mean
BMI indicates body mass index; BSA, body surface area; CABG, coronary
perimeter-derived diameter and the ICD (ratio 0.9–
artery by-pass grafting; COPD, chronic obstructive pulmonary disease; HU:
Hounsfield Unit; ICD, intercommissural distance; LVEF, left ventricular ejection 1.1), 53 (52.5%) had a flared configuration (annulus
fraction; MI, myocardial infarction; NYHA, New York Heart Association; PCI, smaller than the ICD; ratio <0.9) and only 14 (13.8%)
percutaneous coronary intervention; STS, Society for Thoracic Surgery; and TIA,
had a tapered configuration (annulus greater than the
transient ischemic attack.
ICD; ratio >1.1).
statistics in case of normal distribution and using the non-
parametric Wilcoxon statistics otherwise. Qualitative variables Procedural Details
were described and compared using the χ2 or Fisher statistics
depending on variables distribution. TAVI procedural details are depicted in Table 2.
All statistical analyses were performed at the 0.05 global Overall the femoral route was the predominant
significance level (type I error rate), using 2-sided tests. As this access site. In TAV patients, the Edwards Sapien 3,
is an exploratory study, CIs and statistical tests provided are Medtronic Evolut R, and Boston Lotus devices were
*Comparison between tricuspid and bicuspid cohorts. ICD indicates intercommissural distance.
Figure 3. Sizing ratios utilized for different transcatheter aortic valve implantation devices in the bicuspid cohort.
A, Prosthesis labeled diameter/aortic annulus mean perimeter-derived diameter ratio; (B) prosthesis labeled size/intercommissural distance ratio.
BAVARD was not focused on clinical outcomes. elliptical. These findings echo the conclusions from Son
However, we observed similar the Valve Academic et al.22 Watanabe et al23 compared the outcomes in TAV
Research Consortium-2 criteria outcomes in TAV and and BAV patients post-TAVI with first-generation pros-
BAV patients. In a recent report, Yoon et al21 already theses: patients with BAV had higher gradients, larger
demonstrated the improved safety of TAVI with sec- annulus perimeters, and more calcified valves. Higher
ond-generation prostheses in bicuspid anatomies as postprocedural gradient and valve underexpansion
compared with first-generation devices. were frequently observed.
MDCT analysis was the main focus of our registry. In TAV patients, we observed that the aortic annulus
Given the final prosthesis depth of implantation in TAV influenced the final diameter of the TAVI devices. In the
and BAV patients, the MDCT region of interest ex- region of interest, the most frequently used second-gen-
tended from 4 mm below the annulus to 8 mm above. eration devices (S3, ER, and Lotus) matched the aortic
At baseline, as compared with TAV patients, the annulus mean diameter, remained cylindrical, with con-
aortic annulus in BAV patients was larger but not more stant diameters and ellipticity indexes. With BAV, the
Table 3. VARC-2 Clinical and Echocardiographic Outcomes at 30 Days Table 4. Geometry of THV as Assessed in Post-TAVI MDCT
Vascular complications, 24 (27) 16 (16) 0.05 Prosthesis ellipticity index at 4 mm 1.16±0.1 1.17±0.13 0.94
n (%)
Prosthesis diameter at 4 mm/ICD na 0.88±0.11 na
Major vascular 0 5 (5) 0.06 ratio
complications, n (%)
Mean Prosthesis perimeter 23.4±1.9 23.4±2.8 0.78
Minor vascular 22 (26) 11 (11) <0.01 diameter at 8 mm, mm
complications, n (%)
Prosthesis ellipticity index at 8 mm 1.17±0.12 1.18±0.13 0.75
Pacemaker implantation, 12 (14) 13 (13) 0.87
Mean Prosthesis perimeter 23.7±2.2 23.9±2.8 0.55
n (%)
diameter at 12 mm, mm
Echographic outcomes
Prosthesis ellipticity index at 12 mm 1.15±0.1 1.17±0.12 0.23
Effective orifice area, cm2 2.1±0.5 1.9±0.6 0.07
ICD indicates intercommissural distance; MDCT, multidetector computed
Indexed effective orifice 1.33±0.37 1.17±0.4 <0.01 tomography; TAVI, transcatheter aortic valve implantation; and THV,
area, cm2/m2 transcatheter heart valve.
Mean gradient, mm Hg 9.4±4.9 10.7±4.9 0.15
Mild-moderate aortic 11(12.5) 21(20.8) 0.11
conserve stable diameters and ellipticity when they
regurgitation, n (%) meet a point of high resistance, with similar patterns
in both TAV and BAV. In TAV patients that point of re-
Downloaded from http://ahajournals.org by on December 9, 2019
ellipticity in BAV patients as compared with TAVI. One annular-based sizing would result in selecting a device
possible explanation for that discrepancy with our find- potentially too large for the patient, with inherent risks
ings, could be difference in sizing, with bigger prosthe- of aortic root rupture or greater device underexpansion.
ses used in their series and devices potentially failing The calcium burden is likely a major player in the final
to achieve their maximal diameter and circularity in a expansion of TAVI prostheses. It should be quantified
relatively too constrained landing zone. In our series, and integrated in the sizing process.
minimal oversizing was applied (3%–4%), when using From our registry, we can hypothetize that in type
the mean perimeter-derived aortic annulus diameter, 0 and type 1 BAV, (1) annulus-based sizing is valid, (2)
for sizing in BAV. minimal oversizing (3%–4%) ensures good clinical out-
In patients with TAV, the landing zone usually inte- comes, (3) in gray zones and when the ICD is smaller
grates the aortic annulus and the left ventricular out- that the mean annular diameter (taper), selecting a
flow tract, 4 mm below it. As an analogy, given the smaller THV size could avoid excessive oversizing and its
location of constraint points, the landing zone in BAV consequences, and (4) landing THVs 3 to 4 mm below
patients could run from the aortic annulus to 4 mm the aortic annulus could be safe and associated with
above it. That explains, in an effort of simplification, good clinical outcomes. Our findings and simplified siz-
our proposal of integration of the ICD at 4 mm above ing algorithm need to be validated in larger prospective
the annulus for sizing in BAV patients, at least type 0 registries, ideally evaluating the different types of pros-
and type 1 variations. Several configurations can be theses separately. This sizing algorithm will be part of
identified. In a tubular configuration, the mean aortic the BIVOLUT X prospective registry.
annulus diameter matches the ICD and can be used
for sizing with an average oversizing of 3% in our
cohort. In a flared configuration, in which the mean Limitations
aortic annulus diameter is smaller than the ICD, it could Our study has several limitations. It is a retrospective
also be used as the reference for sizing. In a tapered registry with selection bias in THV type and sizing from
configuration (mean perimeter-derived diameter of one center to the other. As not all consecutive patients
the annulus greater than ICD), the ICD could be inte- had pre- and post-TAVI MDCT assessment, our find-
grated, with a 0.9–1/1 ratio because prostheses were ings are exposed to interpretation bias but the number
systematically smaller than the ICD in our BAV patients of patients with interpretable data may strengthen our
(Figure 4). Importantly, annulus-based sizing was appli- conclusions. The distribution of the different types of
Downloaded from http://ahajournals.org by on December 9, 2019
cable to 88% of our BAV patients (Table I in the Data second-generation prostheses was not homogeneous
Supplement). Even though rare, it remains important to between BAV and TAV patients. Various sizing algo-
identify a tapered configuration. In such anatomy, an rithms were used and only a prospective registry could
Figure 4. Various configuration of the landing zone in bicuspid patients and simplified sizing algorithm.
ICD indicates intercommissural distance.
validate our proposed simplified sizing methodology for and Boston Scientific. Dr Van Mieghem has received research grant support
from Abbott vascular, Edwards LifeSciences, Medtronic, Boston Scientific,
BAV. Differences in baseline clinical characteristics be- Claret Medical. Dr de Biase has been supported by a research grant provided by
tween groups also represent a limitation of our registry. the Cardiopath PhD program. The other authors report no conflicts.
Given the relatively small size of the study population, we
liberally did not perform any comparison of devices nor
paired analysis, so, our findings have to be analysed with REFERENCES
caution. The small size of our cohort may explain the 1. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, Tuzcu
EM, Webb JG, Fontana GP, Makkar RR, Brown DL, Block PC, Guyton RA,
lack of mortality observed at 30 days in bicuspid patients. Pichard AD, Bavaria JE, Herrmann HC, Douglas PS, Petersen JL, Akin JJ,
Larger prospective registries are required, exploring the Anderson WN, Wang D, Pocock S, Investigators PT. Transcatheter aortic-
appropriate sizing algorithm for each prosthesis type. valve implantation for aortic stenosis in patients who cannot undergo sur-
gery. N Engl J Med. 2010;363:1597–607. doi: 10.1056/NEJMoa1008232
2. Adams DH, Popma JJ, Reardon MJ. Transcatheter aortic-valve replacement
with a self-expanding prosthesis. N Engl J Med. 2014;371:967–968. doi:
Conclusions 10.1056/NEJMoa1400590
3. Leon MB, Smith CR, Mack MJ, Makkar RR, Svensson LG, Kodali SK,
Second-generation TAVI prostheses similarly reshape the
Thourani VH, Tuzcu EM, Miller DC, Herrmann HC, Doshi D, Cohen DJ,
aortic annulus in tricuspid and bicuspid aortic valves. On Pichard AD, Kapadia S, Dewey T, Babaliaros V, Szeto WY, Williams MR,
average prostheses were deployed 3.4 mm below the Kereiakes D, Zajarias A, Greason KL, Whisenant BK, Hodson RW, Moses
aortic annulus, in the BAVARD registry, with excellent JW, Trento A, Brown DL, Fearon WF, Pibarot P, Hahn RT, Jaber WA,
Anderson WN, Alu MC, Webb JG; PARTNER 2 Investigators. Transcatheter
clinical outcomes. Devices keep consistent diameters or surgical aortic-valve replacement in intermediate-risk patients. N Engl J
from distal edge to 12 mm above. Prosthesis underex- Med. 2016;374:1609–1620. doi: 10.1056/NEJMoa1514616
pansion is constantly observed in BAV. Annular-based siz- 4. Siu SC, Silversides CK. Bicuspid aortic valve disease. J Am Coll Cardiol.
2010;55:2789–2800. doi: 10.1016/j.jacc.2009.12.068
ing is accurate in type 0 and type 1 bicuspid valves with 5. Rodríguez-Caulo EA, Araji OA, Barquero JM. Transapical aortic valve im-
minimal oversizing. The ICD, 4 mm above the annulus, plantation in bicuspid aortic valves: must be an absolute contraindication?
could be integrated in the sizing process for gray zones. Res Cardiovasc Med. 2012;1:37–39. doi: 10.5812/cardiovascmed.4498
6. Philip F, Faza NN, Schoenhagen P, Desai MY, Tuzcu EM, Svensson LG,
Kapadia SR. Aortic annulus and root characteristics in severe aortic ste-
nosis due to bicuspid aortic valve and tricuspid aortic valves: implica-
ARTICLE INFORMATION tions for transcatheter aortic valve therapies. Catheter Cardiovasc Interv.
Received June 23, 2018; accepted November 29, 2018. 2015;86:E88–E98. doi: 10.1002/ccd.25948
The Data Supplement is available at https://www.ahajournals.org/doi/ 7. Sabet HY, Edwards WD, Tazelaar HD, Daly RC. Congenitally bicuspid aortic
suppl/10.1161/CIRCINTERVENTIONS.118.007107. valves: a surgical pathology study of 542 cases (1991 through 1996) and a
literature review of 2,715 additional cases. Mayo Clin Proc. 1999;74:14–
Downloaded from http://ahajournals.org by on December 9, 2019
Transcatheter aortic valve replacement with early- and new-generation the Valve Academic Research Consortium-2 consensus document. Eur
devices in bicuspid aortic valve stenosis. J Am Coll Cardiol. 2016;68:1195– Heart J. 2012;33:2403–2418. doi: 10.1093/eurheartj/ehs255
1205. doi: 10.1016/j.jacc.2016.06.041 20. Sievers HH, Schmidtke C. A classification system for the bicuspid
14. Zhao ZG, Jilaihawi H, Feng Y, Chen M. Transcatheter aortic valve im- aortic valve from 304 surgical specimens. J Thorac Cardiovasc Surg.
plantation in bicuspid anatomy. Nat Rev Cardiol. 2015;12:123–128. doi: 2007;133:1226–1233. doi: 10.1016/j.jtcvs.2007.01.039
10.1038/nrcardio.2014.161 21. Yoon SH, Bleiziffer S, De Backer O, Delgado V, Arai T, Ziegelmueller J,
15. Tanaka R, Yoshioka K, Niinuma H, Ohsawa S, Okabayashi H, Ehara S. Barbanti M, Sharma R, Perlman GY, Khalique OK, Holy EW, Saraf S,
Diagnostic value of cardiac CT in the evaluation of bicuspid aortic ste- Deuschl F, Fujita B, Ruile P, Neumann FJ, Pache G, Takahashi M, Kaneko H,
nosis: comparison with echocardiography and operative findings. AJR Am Schmidt T, Ohno Y, Schofer N, Kong WKF, Tay E, Sugiyama D, Kawamori H,
J Roentgenol. 2010;195:895–899. doi: 10.2214/AJR.09.3164 Maeno Y, Abramowitz Y, Chakravarty T, Nakamura M, Kuwata S, Yong G,
16. Mylotte D, Dorfmeister M, Elhmidi Y, Mazzitelli D, Bleiziffer S, Wagner A, Kao HL, Lee M, Kim HS, Modine T, Wong SC, Bedgoni F, Testa L, Teiger E,
Noterdaeme T, Lange R, Piazza N. Erroneous measurement of the aortic Butter C, Ensminger SM, Schaefer U, Dvir D, Blanke P, Leipsic J, Nietlispach
annular diameter using 2-dimensional echocardiography resulting in inap- F, Abdel-Wahab M, Chevalier B, Tamburino C, Hildick-Smith D, Whisenant
propriate CoreValve size selection: a retrospective comparison with mul- BK, Park SJ, Colombo A, Latib A, Kodali SK, Bax JJ, Søndergaard L, Webb
tislice computed tomography. JACC Cardiovasc Interv. 2014;7:652–661. JG, Lefèvre T, Leon MB, Makkar R. Outcomes in transcatheter aortic valve
doi: 10.1016/j.jcin.2014.02.010 replacement for bicuspid versus tricuspid aortic valve stenosis. J Am Coll
17. Binder RK, Webb JG, Willson AB, Urena M, Hansson NC, Norgaard BL, Cardiol. 2017;69:2579–2589. doi: 10.1016/j.jacc.2017.03.017
Pibarot P, Barbanti M, Larose E, Freeman M, Dumont E, Thompson C, 22. Son JY, Ko SM, Choi JW, Song MG, Hwang HK, Lee SJ, Kang JW.
Wheeler M, Moss RR, Yang TH, Pasian S, Hague CJ, Nguyen G, Raju R, Measurement of the ascending aorta diameter in patients with severe bi-
Toggweiler S, Min JK, Wood DA, Rodés-Cabau J, Leipsic J. The impact of cuspid and tricuspid aortic valve stenosis using dual-source computed to-
integration of a multidetector computed tomography annulus area siz- mography coronary angiography. Int J Cardiovasc Imaging. 2011;27(suppl
ing algorithm on outcomes of transcatheter aortic valve replacement: a 1):61–71. doi: 10.1007/s10554-011-9956-5
prospective, multicenter, controlled trial. J Am Coll Cardiol. 2013;62:431– 23. Watanabe Y, Chevalier B, Hayashida K, Leong T, Bouvier E, Arai T, Farge
438. doi: 10.1016/j.jacc.2013.04.036 A, Hovasse T, Garot P, Cormier B, Morice MC, Lefèvre T. Comparison of
18. Hayashida K, Bouvier E, Lefèvre T, Chevalier B, Hovasse T, Romano M, multislice computed tomography findings between bicuspid and tricuspid
Garot P, Watanabe Y, Farge A, Donzeau-Gouge P, Cormier B, Morice aortic valves before and after transcatheter aortic valve implantation.
MC. Transcatheter aortic valve implantation for patients with severe bi- Catheter Cardiovasc Interv. 2015;86:323–330. doi: 10.1002/ccd.25830
cuspid aortic valve stenosis. Circ Cardiovasc Interv. 2013;6:284–291. doi: 24. Kawamori H, Yoon SH, Chakravarty T, Maeno Y, Kashif M, Israr S,
10.1161/CIRCINTERVENTIONS.112.000084 Abramowitz Y, Mangat G, Miyasaka M, Rami T, Kazuno Y, Takahashi N,
19. Kappetein AP, Head SJ, Généreux P, Piazza N, van Mieghem NM, Jilaihawi H, Nakamura M, Cheng W, Friedman J, Berman D, Sharma R,
Blackstone EH, Brott TG, Cohen DJ, Cutlip DE, van Es GA, Hahn RT, Makkar RR. Computed tomography characteristics of the aortic valve
Kirtane AJ, Krucoff MW, Kodali S, Mack MJ, Mehran R, Rodés-Cabau J, and the geometry of SAPIEN 3 transcatheter heart valve in patients
Vranckx P, Webb JG, Windecker S, Serruys PW, Leon MB. Updated stan- with bicuspid aortic valve disease. Eur Heart J Cardiovasc Imaging.
dardized endpoint definitions for transcatheter aortic valve implantation: 2018;19:1408–1418. doi: 10.1093/ehjci/jex333
Downloaded from http://ahajournals.org by on December 9, 2019