C62 Flow Calibration
{A flow calibration can be used to reduce errors resulting from inaccuracies in path length, path
angle, pipe diameter and path location. The mean ratio between the output signal and the flow
velocity (sometimes called “meter factor”) can be determined by a flow calibration. The
calibration should be performed in a way to ensure that the test rig does not influence the test
results. This incorporates a steady, fully developed velocity flow profile, free from swirl and
pulsation. Generally these requirements can be achieved by using sufficient straight lengths of
pipe upstream and downstream from the meter tube. If necessary, an upstream flow conditioner
may be installed. As a minimum, the manufacturer's reference to the installation conditions must
be observed.
The calibration should be performed as closely as possible to the Reynolds number of the actual
application. During the calibration, the meter output, which may be via serial communications or
a frequency signal, is compared with one or more reference instruments. If using analog output,
the resolution of the conversion to a digital signal needs to be considered.
To improve accuracy, the calibration should be conducted according to good laboratory practice
and in accordance with methods recognized by intemational standards (c.g, 150 4185, 1S0 8316,
1S0 9300). Any flow calibration has a degree of uncertainty, depending on the methods of
calibration and the facility. It is determined by the random and systematic errors in measurement
of the flow velocity and by the random and systematic errors of the laboratory. -
‘The uncertainty of the laboratory can be as low as 0.2% to 0.
‘The calibration should be made over a statistically significant number of runs and over a range of
flow velocities. A common practice is to calibrate at least 6 or 10 velocities, logarithmically
spaced over the meter range, taking the mean of at least three measurements, of 100 seconds, at
tach velocity. In the lower part of the range, the number of measurements may be increased to $
or 10. In determining the number of repeat points at each flow condition, the variability of the
instrument under test should be considered, so that the random errors are sufficiently averaged
and the remaining difference with respect to the calibration standard is predominantly the bias of
the instrument under test
63 Calibration Facilities
jficant problem is the current lack of facilities capable of calibrating these meters, which
lly need to be calibrated at very high flow rates and pipeline pressures. Cost of the
calibration may be a significant portion of the cost of the meter, and the test may only cover the
ower end of the meter capacity
Facilities to calibrate meters larger than 12" in diameter are virtually nonexistent.
C64 Transducer Replacement
Ultrasonic meters are fitted with pairs of electrical ultrasonic transducers. In the event that
transducer must be removed due to malfunction or damage, single transducers or pairs of
transducers are replaced, depending on the recommendations of the specific manufacturer. The
2