Professional Documents
Culture Documents
Vista 120: Instructions For Use
Vista 120: Instructions For Use
Vista 120
WARNING
To properly use this medical device, read
Patient Monitor
and comply with these Instructions for Software 1.n
Use. Release date April 25 2011
Screen images
Schematic renderings of screen images are used, which may differ in appearance or in
configuration from the actual screen images.
WARNING
A WARNING statement provides important information about a potentially
hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially hazardous
situation which, if not avoided, may result in minor or moderate injury to the user or
patient or in damage to the medical device or other property.
NOTE:
A NOTE provides additional information intended to avoid inconvenience during
operation.
Application......................................................................................................................... 6
Intended use...............................................................................................................................6
Restrictions for use ...................................................................................................................6
Overview ............................................................................................................................ 7
Front view ...................................................................................................................................8
Side view...................................................................................................................................10
Back view.................................................................................................................................. 11
Abbreviations ...........................................................................................................................12
Symbols ....................................................................................................................................14
Vista 120 configuration ...........................................................................................................15
Getting started................................................................................................................. 16
Initial inspection .......................................................................................................................16
Installing the monitor...............................................................................................................16
Installing the monitor on a flat surface .........................................................................16
Installing the monitor on a wall ......................................................................................16
Installing the monitor on a roll stand/trolley .................................................................16
Connecting the power cord ....................................................................................................16
Checking the monitor ..............................................................................................................16
Checking the strip recorder....................................................................................................17
Setting the date and time .......................................................................................................17
System Configuration..................................................................................................... 18
Opening User Maintain Menu................................................................................................18
Entering Demo Mode ..............................................................................................................18
Selecting Lead Placement .....................................................................................................18
I
Operation ......................................................................................................................... 19
Using keys ................................................................................................................................20
Changing monitor settings .....................................................................................................21
Adjusting screen brightness...........................................................................................21
Changing date and time .................................................................................................22
Adjusting volume .....................................................................................................................22
Adjusting key volume ......................................................................................................22
Adjusting alarm volume ..................................................................................................22
Adjusting beat volume ....................................................................................................22
Checking the monitor version ................................................................................................22
Setting languages....................................................................................................................22
Understanding screens...........................................................................................................23
Calibrating screens .................................................................................................................23
Alarms.............................................................................................................................. 24
Alarm category.........................................................................................................................24
Physiological alarms .......................................................................................................24
Technical alarms ..............................................................................................................24
Prompts.............................................................................................................................24
Alarm levels ..............................................................................................................................24
Controlling alarm .....................................................................................................................25
Switching the individual alarm off..................................................................................25
Temporary alarm mute....................................................................................................25
Alarm mute .......................................................................................................................25
Setting alarm limits ..........................................................................................................25
Latching alarms........................................................................................................................26
Testing alarms ..........................................................................................................................26
Managing patients........................................................................................................... 39
Admitting a patient...................................................................................................................39
Patient category and paced status ...............................................................................39
Quick admit...............................................................................................................................40
Editing patient information......................................................................................................40
Update a patient ......................................................................................................................40
Monitoring RESP............................................................................................................. 56
Overview ...................................................................................................................................56
RESP safety information ........................................................................................................56
Resp display.............................................................................................................................56
Electrode placement for monitoring resp.............................................................................56
Cardiac overlay ........................................................................................................................57
Chest expansion ......................................................................................................................57
Abdominal breathing ...............................................................................................................57
Selecting resp lead..................................................................................................................57
Changing hold type .................................................................................................................58
Changing the size of the respiration waveform ..................................................................58
Using Resp alarms ..................................................................................................................58
Changing the apnea time .......................................................................................................58
III
Monitoring SpO2 .............................................................................................................. 59
Overview ...................................................................................................................................59
SpO2 safety information..........................................................................................................59
Measuring SpO2 .......................................................................................................................60
Measurement procedure ........................................................................................................60
Understanding SpO2 alarms ..................................................................................................61
Adjusting alarm limits ..............................................................................................................61
Setting SpO2 as pulse source................................................................................................61
Setting pitch tone.....................................................................................................................61
Setting sensitivity.....................................................................................................................61
Monitoring PR.................................................................................................................. 62
Overview ...................................................................................................................................62
Setting PR volume...................................................................................................................62
Using pulse alarms..................................................................................................................62
Selecting the active alarm source.........................................................................................62
Monitoring TEMP............................................................................................................. 68
Overview ...................................................................................................................................68
TEMP safety information ........................................................................................................68
TEMP monitoring setup ..........................................................................................................68
Calculating temp difference ...................................................................................................68
IV
Monitoring CO2 (optional) ............................................................................................... 74
Overview ...................................................................................................................................74
CO2 safety information............................................................................................................74
Monitoring procedures ............................................................................................................75
Zeroing the sensor ..........................................................................................................75
C5 CO2 module................................................................................................................75
Setting CO2 waveform setup .................................................................................................77
Setting CO2 corrections ..........................................................................................................77
Changing CO2 alarms.............................................................................................................77
Changing apnea alarm ...........................................................................................................78
Freeze............................................................................................................................... 79
Overview ...................................................................................................................................79
Entering/exiting freeze status ................................................................................................79
Entering freeze status.....................................................................................................79
Exiting freeze status........................................................................................................79
Reviewing frozen waveform...................................................................................................80
Review.............................................................................................................................. 81
Trend graph review .................................................................................................................81
Selecting trend graph of specific parameter................................................................81
Setting resolution.............................................................................................................81
Scrolling the screen.........................................................................................................81
Switching to the trend table............................................................................................82
Record...............................................................................................................................82
Trend table review ...................................................................................................................82
Setting resolution.............................................................................................................82
Scrolling the screen.........................................................................................................82
Switching to trend graph.................................................................................................82
Recording .........................................................................................................................83
NIBP review..............................................................................................................................83
Scrolling the screen.........................................................................................................83
Recording .........................................................................................................................83
Alarm review.............................................................................................................................83
Scrolling the screen.........................................................................................................83
Selecting alarm event of specific parameter ...............................................................83
Setting time index ............................................................................................................83
Recording .........................................................................................................................84
Arr review..................................................................................................................................84
Selecting specific arrhythmia event ..............................................................................84
Scrolling the screen.........................................................................................................84
Deleting arrhythmia events ............................................................................................84
Recording .........................................................................................................................84
Maintenance .................................................................................................................... 99
Inspecting .................................................................................................................................99
Maintenance task and test schedule ....................................................................................99
Replacing fuse .......................................................................................................................100
EMC Declaration.............................................................................................................116
General information............................................................................................................... 116
Electromagnetic emissions .................................................................................................. 116
Electromagnetic environment .............................................................................................. 116
Electromagnetic immunity .................................................................................................... 117
Recommended separation distances to portable and mobile RF telecommunication
devices ....................................................................................................................................120
VII
Instructions for Use Vista 120 Patient Monitor
Maintenance
WARNING
The medical device must be inspected and serviced regularly by professionals
who possess the required qualifications due to their training and their experience.
Repair of the medical device must also be performed by trained personnel with
additional product-specific DrägerService training.
Dräger recommends that a service contract is obtained with DrägerService and
that all repairs are performed by DrägerService. Dräger further recommends that
only authentic Dräger repair parts are used for maintenance.
If the above are not complied with, the correct functioning of the medical device
may be compromised.
See chapter "Maintenance".
Accessories
WARNING
Only the accessories indicated on the chapter “Accessories” have been tested and
approved for use with the medical device.
Therefore, it is strongly recommended that only these accessories are used in
conjunction with the medical device. Otherwise, the correct functioning of the
medical device may be compromised.
-1-
Instructions for Use Vista 120 Patient Monitor
Connected devices
WARNING
Risk of electric shock and of device malfunction.
Any connected devices or device combinations not complying with the
requirements mentioned in these Instructions for Use may compromise the correct
functioning of the medical device. Before operating any combination of devices,
refer to and strictly comply with the Instructions for Use for all connected devices
and device combinations.
Patient safety
The design of the medical device, the accompanying documentation, and the labeling on
the medical device are based on the assumption that the purchase and the use of the
medical device are restricted to professionals, and that certain inherent characteristics of
the medical device are known to the user. Instructions and WARNING and CAUTION
statements are therefore largely limited to the specifics of the Dräger medical device.
These Instructions for Use do not contain references to various hazards which are
obvious to professionals who operate this medical device as well as references to the
consequences of medical device misuse, and to potentially adverse effects in patients
with different underlying diseases. Medical device modification or misuse can be
dangerous.
CAUTION
Risk of patient injury
Do not make therapeutic decisions based solely on individual measured values and
monitoring parameters.
-2-
Instructions for Use Vista 120 Patient Monitor
Sterile accessories
CAUTION
Do not use sterile-packaged accessories if the packaging has been opened, is damaged
or there are other signs of non-sterility. Disposable articles must not be reprocessed and
resterilized. Reuse, reprocessing, or resterilization can lead to a failure of the medical
device and cause injury to the patient.
Installing accessories
CAUTION
Install accessories to the basic device in accordance with the Instructions for Use of the
basic device. Make sure that there is a safe connection to the basic device system.
Strictly observe Assembly Instructions and Instructions for Use.
-3-
Instructions for Use Vista 120 Patient Monitor
WARNING
Route all cables away from patient’s throat to avoid possible strangulation.
WARNING
Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off during patient monitoring may
result in a hazard to the patient. Remember that the most reliable method of patient
monitoring combines close personal surveillance with correct operation of
monitoring equipment.
WARNING
When interfacing with other equipment, a test for leakage current must be
performed by qualified hospital technical personnel before using with patients.
WARNING
During monitoring, if the power supply is off and there is no battery for standby,
the monitor will be off. After reconnecting the power supply, the user should turn
on the monitor for monitoring.
WARNING
Keep away from fire immediately when leakage or foul odor is detected.
WARNING
The device and accessories are to be disposed of according to local regulations
after their useful lives. Alternatively, they can be returned to the dealer or the
manufacturer for recycling or proper disposal. Batteries are hazardous waste. Do
NOT dispose them together with house-hold garbage. At the end of their life hand
the batteries over to the applicable collection points for the recycling of waste
batteries. For more detailed information about recycling of this product or battery,
contact your local Dräger sales representative.
WARNING
This equipment is not intended for home usage.
WARNING
Devices connected to the equipment must meet the requirements of the applicable
IEC standards (e.g., IEC 60950 “Safety standards for information technology
equipment” and IEC 60601-1 “Safety standards for medical electrical equipment”)
The system configuration must meet the requirements of the IEC 60601-1-1
“Medical electrical systems” standard. Any personnel who connect devices to the
equipment’s signal input/output port is responsible for providing evidence that the
safety certification of the devices has been performed in accordance to the IEC
60601-1-1.
CAUTION
Do not immerse transducers in liquid. When using solutions, use sterile wipes to avoid
pouring fluids directly on the transducer.
-4-
Instructions for Use Vista 120 Patient Monitor
CAUTION
Do not use autoclave or gas to sterilize the monitor, recorder or any accessories.
CAUTION
Disposable devices are intended for single use only. They should not be reused as
performance could degrade or contamination could occur.
CAUTION
Remove a battery whose life cycle has expired from the monitor immediately.
CAUTION
Avoid liquid splash on the device. The temperature must be kept between 5 and 40 °C
while working. And it should be kept between -20 and +55°C during transportation and
storage.
NOTE
• Position the device in a location where the operator can easily see the screen and
access the operating controls.
• The monitor can only be used on one patient at a time.
• If the monitor gets damp, put it in dry circumstance to dry it until it can work normally.
If liquid pours on the monitor, contact DrägerService.
• Do not use this monitor for diagnostic purposes.
• When the monitor is shut off, only the patient information and alarm settings can be
saved.
• The pictures and dialog boxes in these Instructions for Use are for reference only.
• Regular preventive maintenance should be carried every two years. The user is
responsible for any requirements specific to their country.
-5-
Instructions for Use Vista 120 Patient Monitor
Application
Intended use
The Vista 120 monitors parameters such as ECG (3-lead or 5-lead selectable), respiration (RESP),
functional arterial oxygen saturation (SpO2), invasive or non-invasive blood pressure (dual-IBP,
NIBP), temperature (dual-TEMP), and expired CO2.
Vista 120 is intended to be used only under regular supervision of clinical personnel. It is
applicable to adult, pediatric, and neonatal usage in a hospital environment and during patient
transport inside a healthcare facility.
-6-
Instructions for Use Vista 120 Patient Monitor
Overview
Vista 120 is optimized for surgical, cardiac, medical and neonatal care environments, and can
store data for both trends and events. The user can also view and record graphical and tabular
trends (vital signs).
Vista 120 has a 15-inch TFT color flat panel display. Up to 11 waveforms can be displayed on
the screen.
-7-
Instructions for Use Vista 120 Patient Monitor
Front view
A
B
C
D
E
F
G
I
A Alarm When an alarm occurs, the alarm indicator will light or flash. The alarm
indicator level is color coded.
B Alarm Mute Press this button to pause the alarm. All audio alarms stop and the
Temporary Alarm Mute **s and the symbol are displayed in the
information area. If the user presses the button again or if the pause time
ends, the system will resume normal monitoring status and the Temporary
Alarm Pause **s and icon will clear.
The symbol is displayed in the information area. Pressing or holding
the button again resumes the alarm.
C NIBP Press to inflate the cuff and perform NIBP measurement. Press again to
measurement stop the measurement and deflate the cuff.
D Trend Press this button to enter trend table review dialog box.
E Freeze In normal mode, press this button to freeze all the waveforms on the screen.
In Freeze mode, press this button to restore the waveform refreshing.
F Recording Press this button to start a real-time recording. Press again to stop
recording.
G Menu Press to return to the main dialog box.
H Rotary Knob The user can turn the rotary knob clockwise or counter-clockwise to
-8-
Instructions for Use Vista 120 Patient Monitor
highlight the desired item. Press the rotary knob to select the item.
I Mains/Battery Refer to chapter “Battery power indicator” for details.
indicator
J On/Off When connected to the AC power supply, press the key to turn the monitor
on. Press the key again to turn the monitor off.
Further Alarm Mute information can be found in the chapter “Audio Off”.
-9-
Instructions for Use Vista 120 Patient Monitor
Side view
A
B
C
D I
E
F
G
H
A NIBP port
B ECG port
C etCO2 port
D SpO2 port
E T1 port
F T2 port
G IBP 1 port
H IBP 2 port
I Recorder
J Battery compartment door
- 10 -
Instructions for Use Vista 120 Patient Monitor
Back view
A
L
K
J B
I
C
H G F E D
A Fan
B Speaker
C Equipotential grounding terminal. If the monitor is used with other devices, connect this
terminal to eliminate potential ground differences between devices.
D Security lock. Used to prevent the power cord from detaching.
E Ventilation grid
F Anti-theft lock
G USB interface. This port is used to connect the USB device.
H RJ45 network interface
I RS232 interface
J VGA output
K Analog output. Vista 120 outputs the waveform through the auxiliary output port.
L Nurse Call port. This port may be connected to an external nurse call system, not provided
by Dräger.
- 11 -
Instructions for Use Vista 120 Patient Monitor
Abbreviations
Abbr. Explanation Abbr. Explanation
AC Alternating current Adu Adult
AHA American Heart Association Art Arterial
aVF Left foot augmented lead aVL Left arm augmented lead
aVR Right arm augmented lead awRR Airway respiration rate
BP Blood pressure BTPS Body temperature and pressure,
saturated
CISPR International Special Committee CO2 Carbon dioxide
on Radio Interference
COHb Carboxyhemoglobin CVP Central venous pressure
DC Direct current Dia Diastolic
ECG Electrocardiogram EEC European Economic Community
EMC Electromagnetic compatibility EMI Electromagnetic interference
ESU Electrosurgical unit et End-tidal
etCO2 End-tidal carbon dioxide FCC Federal Communication
Commission
FiCO2 Fraction of inspired carbon dioxide Hb Hemoglobin
Hb-CO Carboxyhemoglobin HR Heart rate
IBP Invasive blood pressure ICP Intracranial pressure
ICU Intensive care unit ID Identification
IEC International Electrotechnical IEEE Institute of Electrical and
Commission Electronic Engineers
LA Left arm LAP Left arterial pressure
LCD Liquid crystal display LED Light emitting diode
LL Left leg MAP Mean arterial pressure
MDD Medical Device Directive MetHb Methemoglobin
N/A Not applicable Neo Neonate
NIBP Non-invasive blood pressure O2 Oxygen
oxyCRG Oxygen cardio-respirogram PA Pulmonary artery
PAWP Pulmonary artery wedge pressure Ped Pediatric
Pleth Plethysmogram PR Pulse rate
PVC Premature ventricular complex R Right
- 12 -
Instructions for Use Vista 120 Patient Monitor
- 13 -
Instructions for Use Vista 120 Patient Monitor
Symbols
Alternating current
Mains indicator
Serial number
Power supply switch
Caution Recycle
- 14 -
Instructions for Use Vista 120 Patient Monitor
- 15 -
Instructions for Use Vista 120 Patient Monitor
Getting started
Initial inspection
Before unpacking, check the packaging and check that there are no signs of mishandling or
damage. If the shipping cartons are damaged, contact your local supplier for assistance.
Open the package carefully and remove the monitor and accessories. Check that all correct
options and accessories are included.
If you have any question, contact your local supplier.
WARNING
Do not use the device if you detect any damage or if the monitor displays error
messages. Contact your hospital technical personnel or Customer Service Center
immediately.
- 16 -
Instructions for Use Vista 120 Patient Monitor
NOTE
• Check that all monitor functions are operating correctly.
• If rechargeable batteries were provided, charge them each time before using the
device to ensure adequate power.
• If you power the monitor off, wait at least 60 seconds before powering it on again.
- 17 -
Instructions for Use Vista 120 Patient Monitor
System Configuration
The user can not change the system configuration of the monitor. After the monitor is installed
and checked properly, a service engineer needs to change the system configuration.
3. Type the password into the displayed dialog box and press OK to enter the User Maintain
menu.
- 18 -
Instructions for Use Vista 120 Patient Monitor
Operation
Everything the user needs to operate the monitor is displayed on its screen. Nearly every element
on the screen is interactive. Screen elements include measurement numerics, waveforms, screen
keys, information fields, alarms fields and menus. The monitor can be configured in a number of
different ways. For example, the user can access an item through the on-screen setup menu, via a
hard key, or via a shortcut key. This Instructions for Use describes how to access items via an
on-screen menu.
1 2 3 4 5
16
15 14 13 12 11 10 9 8
1 Department 10 Date and time
2 Bed number 11 Battery status symbol
3 Patient name 12 AC power supply symbol
4 Patient type 13 Shortcut key area
5 Alarm status area 14 Scroll left to display more shortcut keys
6 Alarm off 15 Mute key
7 Measurement value 16 Parameter waveform
8 Menu
9 Scroll right to display more shortcut keys
- 19 -
Instructions for Use Vista 120 Patient Monitor
Using keys
The monitor has four different types of keys:
Permanent keys
A permanent key is a graphical key that is permanently located at the bottom of the main screen
allowing fast, direct access to functions.
Shortcut keys
A shortcut key is a configurable graphical key located at the bottom of the main screen. It gives
the user direct access to functions. The selection of shortcut keys available on the monitor
depends on the monitor configuration and options purchased. The following functions are
available:
- 20 -
Instructions for Use Vista 120 Patient Monitor
Hardkeys
Hardkeys are the physical keys on the front panel of the monitor. The Vista 120 has the following
hardkeys: Mute, NIBP, Trend, Freeze, Recording, and Menu. Refer also to the chapter “Front
view”.
Pop-up keys
Pop-up keys are task-related graphical keys that appear automatically on the screen when
required. For example, the confirmation pop-up key appears only when the user need to confirm a
change.
- 21 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Changing date and time will affect the storage of trend data.
Adjusting volume
Adjusting key volume
The key volume is audible when the user selects any field on the monitor screen or when the user
turns the knob. To adjust the key volume:
1. Select the Key Volume key on the screen directly or
2. Select Menu > System Setup > Key Volume, then select the appropriate setting: Five is the
loudest and Zero is the quietest.
Setting languages
To change the language:
1. Select Menu > Maintenance > User Maintain, then type the maintenance password into the
displayed dialog box.
2. Select the Language option on the pop-up dialog box to display the language list.
3. Select the desired language from the list. To make the change permanent, restart the monitor.
- 22 -
Instructions for Use Vista 120 Patient Monitor
Understanding screens
The monitor has a set of preconfigured screens, which are optimized for common monitoring
scenarios such as OR adult or ICU neonatal. A screen defines the overall selection, size and
position of waveforms, numerics and shortcut keys displayed when the user starts the monitor.
The user can switch between different screens during monitoring. Screens do not affect alarm
settings, patient category etc. If the user changes from a complex to a less complex screen layout,
some measurements may not be visible but are still monitored in the background. For detailed
information, refer to chapter “User Interface”.
Calibrating screens
To calibrate the screen:
1. Select the Touch Calib shortcut key on the screen directly or select Menu > Maintenance >
User Maintain, then type the maintenance password into the displayed dialog box, then
select TouchScr Calibration from the pop-up dialog box.
- 23 -
Instructions for Use Vista 120 Patient Monitor
Alarms
The following alarm information is applicable for all measurements. Refer to chapters on specific
alarms for more information.
WARNING
A potential hazard can exist if different alarm presets are used for the same or
similar equipment in different areas, e.g., an intensive care unit or an cardiac
operating room.
Alarm category
The monitor has physiological, technical, and prompt alarms.
Physiological alarms
If one or more physiological parameters monitoring the patient exceed the predefined alarm limit
(for example: APNEA and SpO2 value exceed the alarm limit), the monitor alarms. For more
information, refer to chapter “Physiological alarm information”.
Technical alarms
If, for example, the device has a low battery or defective lead the monitor will initiate a technical
alarm. Technical alarms cannot be disabled. For more information, refer to chapter “Technical
Alarm Information”.
Prompts
The device monitors processes and other functions, such as: ARR Relearning. For more
information, refer to chapter “Prompts”.
Alarm levels
There are three alarm levels consisting of high, medium, and low.
1. High level alarms
A life threatening situation exists and emergency treatment is required.
2. Medium level alarms
The patient’s vital signs or system status is abnormal. Response is required.
3. Low level alarms
The patient’s vital signs or system status is abnormal. Attention is required.
- 24 -
Instructions for Use Vista 120 Patient Monitor
Alarm sound
The high/medium/low-level alarms have the following audible characteristics:
Alarm level Prompt
Mode is “DO-DO-DO------DO-DO, DO-DO-DO------DO-DO”, which
High is triggered once every 5 seconds. The alarm indicator flashes in red,
with a frequency of 1.4Hz~2.8Hz.
Mode is “DO-DO-DO”, which is triggered once every 25 seconds. The
Medium
alarm indicator flashes in yellow, with a frequency of 0.4Hz~0.8Hz.
Low Mode is “DO-”, which is triggered once every 30 seconds.
The sound pressure range for audible alarm signals is from 45 dB to 85 dB.
Controlling alarm
Switching the individual alarm off
To switch the alarm off, select XX Setup > Alarm Setup (XX stands for parameter name) and
set the alarm from the pop-up list.
The temporary alarm mute symbol is displayed and the remaining pause time is displayed on
a red colored background.
Alarm mute
To mute the alarm, select Menu > Maintenance > User Maintain > Alarm Setup, and set Mute
to On.
Press the key on the front panel for more than three seconds or the shortcut key Mute on the
screen to mute the alarm.
WARNING
Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
- 25 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Setting alarm limits to extreme values may cause the alarm system to become
ineffective.
High
Alarm Limit
Upper arrow or
lower arrow to
increase or decrease
the alarm limit
Setting value of
high alarm limit
Latching alarms
To set the alarm latch function, select Menu > Maintenance > User Maintain > Alarm Setup
and set Alarm Latch from the pull-down list. If set to On, when an alarm occurs, the monitor
will display the alarm message of the parameter in the alarm status area. If several parameters
latch alarms, each message is displayed in two second intervals.
To deselect the alarm latch, set Alarm Latch to Off. When Alarm Latch is set to Off, the latch
function is invalid.
Testing alarms
When the user switches the monitor on, a self test is started. Verify that the alarm indicator lights
and emits a single tone. The self test verifies that the visible and audible alarm indicators are
functioning correctly. For further testing of individual measurement alarms, perform the
measurement on a subject or use a simulator. If necessary, adjust the alarm limits and check that
the system is operating correctly.
- 26 -
Instructions for Use Vista 120 Patient Monitor
Alarm information
Physiological alarm information
Message Cause Alarm level
ECG Too Weak Cannot detect the signal in designated time period. High
ST-X high ST value is above the upper alarm limit. User-selectable
(X stands for I, II, III, aVR, aVL, aVF, V)
ST-X low ST value is below the lower alarm limit. User-selectable
(X stands for I, II, III, aVR, aVL, aVF, V)
ASYSTOLE No QRS is detected for 4 consecutive seconds User-selectable
VFIB/VTAC Ventricular tachycardia: The fibrillation waveform lasts User-selectable
for 4 consecutive seconds; or the number of continuous
Vent beats is larger than the upper limit of cluster Vent
beats (>5). The RR interval is less than 600 ms.
VT>2 3< the number of cluster PVCs < 5 User-selectable
COUPLET 2 consecutive PVCs User-selectable
BIGEMINY Vent bigeminy User-selectable
TRIGEMINY Vent trigeminy User-selectable
R ON T A type of single PVC under the condition that HR<100, User-selectable
R-R interval is less than 1/3 the average interval, followed
by a compensating pause of 1.25X the average R-R
interval (the next R wave advances onto the previous T
wave).
PVC Single PVCs not belonging to the type of above User-selectable
mentioned PVCs.
TACHY 5 consecutive QRS complex, User-selectable
RR interval is less than 0.5 s.
BRADY 5 consecutive QRS complex, User-selectable
RR interval is longer than 1.5 s.
MISSED When HR is less than 100 beats/min., no heart beat is User-selectable
BEATS detected during the period 1.75 times of the average RR
interval; or
When HR is higher than 100beat/min, no beat is detected
within 1 second.
IRR IRREGULAR RHYTHM: The patient has irregular heart User-selectable
rate, check patient’s condition, electrodes, cables and
leads.
- 27 -
Instructions for Use Vista 120 Patient Monitor
- 28 -
Instructions for Use Vista 120 Patient Monitor
- 29 -
Instructions for Use Vista 120 Patient Monitor
- 30 -
Instructions for Use Vista 120 Patient Monitor
ECG Comm Fail ECG module failure or High Stop monitoring ECG
communication failure and notify hospital
technical personnel or
customer support.
ECG Noise ECG signal is too noisy. Low Check lead connection
and patient condition
ECG Signal The amplitude of ECG is too Low Modify the ECG gain.
Overflow tall.
RESP Comm Fail RESP module failure or High Stop monitoring RESP,
communication failure and notify hospital
technical personnel or
customer support.
SpO2 Sensor Off SpO2 sensor may be Low Ensure the sensor is
disconnected from the properly connected to
patient or the monitor. the patient’s body.
SpO2 No Sensor SpO2 sensor was not Low Ensure the monitor and
connected well or connected sensor are properly
to the monitor, or the connected. Reconnect
connection is loose. the sensor.
SpO2 Low Perfusion The pulse signal is too weak Low Reconnect the SpO2
or the perfusion of the sensor and change the
measurement site is too low. measurement site. If
the problem continues,
notify hospital
technical personnel or
customer support.
SpO2 Comm Fail SpO2 module failure or High Stop monitoring SpO2
communication failure and notify hospital
technical personnel or
customer support
NIBP Comm Fail NIBP module failure or High Stop monitoring NIBP
communication failure and notify hospital
technical personnel or
customer support
- 31 -
Instructions for Use Vista 120 Patient Monitor
NIBP Invalid Reset The hardware pressure is too Low Measure again, if
high failure persists, stop
monitoring NIBP and
notify hospital
technical personnel or
customer support.
NIBP Time Out Measuring time has Low Measure again or use
exceeded 120 seconds other measuring
(adult/pediatric) or 90 method.
seconds (neonatal).
NIBP Tube Leak NIBP cuff or pump has a Low Check the NIBP cuff
leakage and pump for leakages
NIBP Cuff type The cuff type used is Low Confirm the patient type
Error inconsistent with the patient and change the cuff
type
- 32 -
Instructions for Use Vista 120 Patient Monitor
NIBP Pneumatic NIBP cuff or pump has a Low Check the NIBP cuff
Leak leakge and pump for leakages
NIBP Cuff Leak Cuff, pumnp or airway has a Low NIBP cuff is not
leakage properly connected, or
there is a leak in the
airway.
NIBP Leak Test Hardware abnormal High Check whether the
Error airway is occluded or
pressure sensor works
properly in pressure
meter mode. If the
problem still exists,
contact your service
personnel.
NIBP Range The patient’s blood pressure Low The patient’s blood
Exceeded is beyond the measurement pressure is beyond the
range measurement range
NIBP Air Leak The cuff pressure cannot Low Check the connections
reach the set value within 60 and the wrapped cuff to
sec. or 20 sec. Neo mode see whether they are all
well prepared.
- 33 -
Instructions for Use Vista 120 Patient Monitor
TEMP T2 Sensor Temperature cable of TEMP Low Make sure that the cable
Off channel 2 may be is properly connected.
disconnected.
- 34 -
Instructions for Use Vista 120 Patient Monitor
Prompts
Message Cause
ECG Arr Learning The QRS template building required for Arr. Analysis is in
process.
SpO2 Search Pulse When the sensor is connected to the patient, the SpO2 is
analyzing the patient signal and searching for the pulse to
compute the saturation.
Manual Measuring In manual measuring mode
Continual Measuring In continuous measuring mode
Auto Measuring In automatic measuring mode
- 35 -
Instructions for Use Vista 120 Patient Monitor
Message Cause
Measure Abort Measurement over
Calibrating During calibrating
Calibrate Abort Calibration over
Leak Testing During pneumatic test
Leak Test Ok NIBP module has passed leak test
Leak Test Abort Pneumatic test over
Resetting NIBP module in resetting
Please Start NIBP module is in idle status
Done NIBP measurement successfully done
Continual Measuring NIBP module performs continuous measuring
Stat Measuring NIBP module performs STAT measurement
Please Switch To Maintain NIBP module is in normal mode, the user can not start leak
Mode test and pressure calibration. Enter User Maintain > NIBP
Maintain and switch to Maintain Mode to perform leak test
or pressure calibration.
Please Switch To Normal NIBP module is in maintain mode, the user can not start blood
Mode pressure measurement. Enter User Maintain > NIBP
Maintain and switch to Normal Mode to perform blood
pressure measurement.
CO2 Standby Changing from measuring to standby mode places the module
in energy-saving status.
CO2 Sensor Warm Up The CO2 module is in warm-up state.
- 36 -
Instructions for Use Vista 120 Patient Monitor
- 37 -
Instructions for Use Vista 120 Patient Monitor
MAP 235 20
PED SYS 200 40
DIA 150 10
MAP 165 20
NEO SYS 135 40
DIA 100 10
MAP 110 20
- 38 -
Instructions for Use Vista 120 Patient Monitor
Managing patients
Admitting a patient
The monitor displays patient physiological data and records the information in trends. This allows
the user to monitor unadmitted patients. It is recommended, however, that the user admit patients
to record their information in reports.
During admission, the patient category setting determines the algorithm that the monitor uses to
process and calculate measurements. These include safety limits that are applied for specific
measurements and alarm limit ranges.
To admit a patient:
1. Select the Admission shortcut key on the screen or
2. Select Menu > Patient Setup > New Patient, then a message is displayed to ask the user to
confirm to update patient.
3. Click No to cancel this operation; click Yes, the Patient Info window is displayed.
4. Enter the patient information:
• Serial No: Enter the patient’s medical record number (MRN), for example 12345678.
• Last name: Enter the patient’s last name (family name), for example Smith.
• First name: Enter the patient’s first name, for example Joseph.
• Gender: Choose Male or Female.
• Type: Choose the patient type, either Adult, Pediat, or Neonat.
• BloodType: Choose the patient type N/A, A, B, AB, or O.
• Pace: Choose On or Off (The user must use On if the patient has a pacemaker).
• Date of Birth: Enter the patient’s date of birth.
• Date of Admission: Enter the patient’s date of admission.
• Height: Enter the patient’s height.
• Weight: Enter the patient’s weight.
• Doctor: Enter any extra information about the patient or treatment.
5. Select Exit to save the changes and exit this screen.
- 39 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Changing the patient category may change the arrhythmia and NIBP alarm limits.
Always check alarm limits to make sure that they are appropriate for your patient.
WARNING
For paced patients, you must set Pace to On. If it is incorrectly set to Off, the
monitor could mistake a pace pulse for a QRS and fail to alarm during asystole.
Quick admit
If the user does not have the time or information to fully admit a patient, fill in the rest of the
patient information later.
The user can quickly admit a patient as follows:
1. Select Menu > Patient Setup > Quick Admit, a message is displayed requesting that the
user confirm the patient update.
2. Click No to cancel this operation; click Yes to continue and the Patient Info window is
displayed, choose Type and Pace and set them to the correct mode.
3. Select Exit to save the changes and exit this screen.
Update a patient
The user should always perform an update before starting monitoring for a new patient. To
update a patient:
• Select Menu > Patient Setup > New Patient or Menu > Patient Setup > Quick Admit, a
menu is displayed.
• If the user selects Yes, the monitor will update the patient information.
• If the user selects No, the monitor won’t update the patient information and returns to
patient setup dialog box.
NOTE
Discharging a patient will clear all associated data in the monitor.
- 40 -
Instructions for Use Vista 120 Patient Monitor
User interface
Setting interface style
The user can set the interface style. The following options are available:
• Waveform sweep
• Monitored parameters
Only authorized personnel should set the interface style.
- 41 -
Instructions for Use Vista 120 Patient Monitor
This is a NICU only feature. Monitoring of SpO2, HR and Resp is different for neonates than for
adults.
- 42 -
Instructions for Use Vista 120 Patient Monitor
Monitoring ECG
Overview
The electrocardiogram (ECG) measures electrical activity in the heart and displays it both
numerically and as a waveform. This chapter also describes arrhythmia and ST segment
monitoring.
- 43 -
Instructions for Use Vista 120 Patient Monitor
CAUTION
If patients may be defibrillated, be sure to use defibrillation-proof ECG cables, otherwise,
the ECG cables will be damaged.
NOTE
• Interference from a non-grounded instrument near the patient and ESU interference
can cause inaccuracy of the waveform.
• IEC/EN60601-1-2 (protection against radiation is 3 V/m) specifies that the electrical
field density exceeding 1V/m may cause measurement error in various frequencies. It
is accordingly suggested to not use equipment generating electrical radiation near
ECG/RESP monitoring devices.
• In the default settings of the monitor, the ECG waveforms are the first two waveforms
from the top in the waveform area.
• For measurements in or near the heart connect the monitor to the potential
equalization system.
• For protecting the environment, the used electrodes must be recycled or disposed of
properly.
ECG display
The figure below is for reference only.
The symbol “①” indicates the lead name of the displayed waveform; refer to chapter “Selecting
calculation lead”.
The symbol “②” indicates waveform gain; refer to chapter “Changing the size of the ECG
waveform”.
The symbol “③” indicates Filter setting; refer to chapter “Changing the ECG filter setting”.
- 44 -
Instructions for Use Vista 120 Patient Monitor
- 45 -
Instructions for Use Vista 120 Patient Monitor
Monitoring procedure
Preparation
Skin is a poor conductor of electricity. Prepare the patient's skin to facilitate good electrode
conductivity.
• Select sites with intact skin, without impairment of any kind.
• Shave hair from sites, if necessary.
• Wash sites thoroughly with soap and water. (Never use ether or pure alcohol, because this
increases skin impedance).
• Rub the skin briskly to increase capillary blood flow in the tissues and remove skin scurf and
grease.
Installing electrodes
NOTE
The following table gives the corresponding lead names according to IEC (used in
Europe) and the AHA (USA) respectively.
AHA (American Standard) IEC (Europe Standard)
Electrode
Color Electrode Labels Color
Labels
RA White R Red
LA Black L Yellow
LL Red F Green
RL Green N Black
V Brown C White
- 46 -
Instructions for Use Vista 120 Patient Monitor
- 47 -
Instructions for Use Vista 120 Patient Monitor
- 48 -
Instructions for Use Vista 120 Patient Monitor
ECG leads are used mainly for monitoring patient vital signs. When using the patient monitor
with other HF surgical equipment, use defibrillator-proof ECG leads.
The placement of the ECG leads depends on the type of surgery. For example, during open heart
surgery the electrodes can be placed laterally on the chest or back. In the operating room, artifacts
may affect the ECG waveform due to the use of ES (electrosurgery) equipment. To help reduce
this interference, the electrodes can be placed on the right and left shoulders, the right and left
sides near the abdomen and the chest lead on the left side at mid-chest. Avoid placing the
electrodes on the upper arms, which might result in a small waveform.
NOTE
• If an ECG waveform is not accurate (with electrodes are tightly attached), change the
lead.
• Interference from a non-grounded instrument near the patient can cause inaccurate
waveforms.
- 49 -
Instructions for Use Vista 120 Patient Monitor
ECG display
The ECG display varies with the Lead Type. When Lead Type is set to 3 Leads, Display can be
set to Normal, and displays one ECG waveform on the main screen.
When Lead Type is set to 5 Leads, Display can be set to Normal, Full-Scr and Half-Scr. Select
Normal to display two ECG waveforms on the main screen. Select Full-Scr to display seven
ECG waveforms. Half-Scr displays seven ECG waveforms on the screen in an area of four
waveforms.
NOTE
If 3 Leads is selected in the ECG Setup menu, only Normal can be selected for Display
in the sub-menu.
- 50 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Some pace pulses can be difficult to reject. When this happens, the pulses are
counted as a QRS complex, and could result in an incorrect HR and failure to
detect cardiac arrest or some arrhythmias. Keep pacemaker patients under close
observation.
ECG calibration
Selecting this item from the ECG Setup menu inserts a square wave into the ECG waveform that
can be used to estimate the amplitude accuracy of that waveform. Select the item again to turn off
the square wave.
NOTE
The patient is not monitored while this square wave is displayed.
ST segment monitoring
The monitor performs ST segment analysis on normal and atrially paced beats and calculates ST
segment elevations and depressions. This information can be displayed in the form of ST
numerics and snippets on the monitor.
ST segment monitoring function is shut off by default. It can be switched to On when necessary.
When using the ST analysis function, the result is displayed on the main screen. Refer to the
following figure.
- 51 -
Instructions for Use Vista 120 Patient Monitor
Setting ST analysis
To change ST analysis, select ECG Setup > ST Setup > ST Analysis, then select On or Off
from the pop-up list.
ST display
The screen may look different from the illustrations.
NOTE
• ST Analysis can be used only in Adu mode.
• If ST Analysis is on, the monitor must be in Diagnos mode.
• ECG monitoring must be in Diagnos mode.
DEF POINT
- 52 -
Instructions for Use Vista 120 Patient Monitor
The ST and ISO measurement points need to be adjusted when monitoring is started, and if the
patient's heart rate or ECG morphology changes significantly. Always ensure that ST
measurement points are appropriate for the patient. Abnormal QRS complexes are not considered
in ST segment analysis.
Arrhythmia monitoring
Arrhythmia analysis
The arrhythmia algorithm is used to monitor clinical neonatal and adult ECG. It also detects heart
rate and ventricular rhythm changes, arrhythmia events, and generates alarm information. The
arrhythmia algorithm can monitor paced and non-paced patients. Qualified personnel can use
arrhythmia analysis to evaluate patient condition (such as heart rate, PVCs frequency, rhythm and
ectopic beat). The arrhythmia algorithm can also monitor patients and generate arrhythmia
specific alarms.
The monitor can support up to 16 different arrhythmia analyses.
ARR Types Occurring Condition
ASYSTOLE No QRS is detected for 4 seconds
Ventricular tachycardia: The fibrillation waveform lasts for 4 consecutive
seconds; or the number of continuous Vent beats is larger than the upper limit
VFIB/VTAC of cluster Vent beats (>5).
The RR interval is less than 600 ms.
VT>2 3< the number of cluster PVCs < 5
COUPLET 2 consecutive PVCs
BIGEMINY Vent bigeminy
TRIGEMINY Vent trigeminy
A type of single PVC when HR<100,R-R interval is less than 1/3 the average
R ON T interval, followed by a compensating pause of 1.25X the average R-R interval
(the next R wave encroaches on the previous T wave).
PVC All other single PVCs.
- 53 -
Instructions for Use Vista 120 Patient Monitor
- 54 -
Instructions for Use Vista 120 Patient Monitor
PVCs alarm
Select On in the menu to enable the prompt message when an alarm occurs. Select Off to disable
the alarm function. The symbol appears.
WARNING
When the PVCs Alarm is set to OFF, the monitor won’t give an alarm prompt even if
an alarm occurs. In order to avoid endangering the patient’s life, the user should
use this function cautiously.
ARR relearning
Select this item to start a learning procedure. ECG ARR LEARNING is displayed on the screen.
The ECG ARR LEARNING will start automatically the following:
• Changing lead type;
• Connecting leads;
• Updating the patient category;
• Starting ARR learning manually;
• After the ARR analysis is switched on;
• The module is set to on;
• Calibration mode is changed to normal measurement mode;
• Exiting the Demo mode;
• Exiting the standby mode;
ARR alarm
The user can switch on or off all arrhythmia alarms by selecting ECG Setup > ARR Analysis >
ARR Setup > ARR Alarm. Individual arrhythmia alarms can be turned on or off. These are:
ASYSTOLE, VFIB/VTAC, R-ON-T, VT>2, COUPLET, PVC, BIGEMINY, TRIGEMINY,
TACHY, BRADY, MISSED BEATS, IRR, PNC, PNP, VBRADY and VENT.
To switch individual alarms on or off, select ECG Setup > ARR Analysis > ARR Setup > ARR
Alarm.
- 55 -
Instructions for Use Vista 120 Patient Monitor
Monitoring RESP
Overview
The monitor measures respiration from the amount of thoracic impedance between two ECG
electrodes. The change of impedance between the two electrodes, (due to the thoracic movement),
produces a respiratory waveform on the screen.
Resp display
Resp 30
8
14
Electrode placement for monitoring resp
Correct patient skin preparation for electrode placement is important for Resp measurement.
Refer to the chapter on ECG for more information.
- 56 -
Instructions for Use Vista 120 Patient Monitor
The Resp signal is always measured between two of the ECG electrodes. There are two standard
ECG leads for selection: I lead (RA and LA) and II lead (RA and LL).
Cardiac overlay
Cardiac activity that affects the Resp waveform is called cardiac overlay. This occurs when Resp
electrodes detect impedance changes caused by rhythmic blood flow. Correct electrode placement
can reduce cardiac overlay.
Chest expansion
Some patients, especially neonates, expand their chests laterally. In this case, place the two
respiratory electrodes in the right midaxillary and left lateral chest areas at the maximum
expansion point. This placement optimizes respiratory waveform results.
Abdominal breathing
Some patients with restricted chest movement breathe mainly abdominally. In this case, place the
LL electrode on the left abdomen at the point of maximum abdominal expansion. This placement
optimizes respiratory waveform results.
NOTE
Place the red and green electrodes diagonally to optimize the respiration waveform. Do
not position electrodes over the liver or the heart’s ventricles. This is an especially
important safety procedure for neonates.
- 57 -
Instructions for Use Vista 120 Patient Monitor
- 58 -
Instructions for Use Vista 120 Patient Monitor
Monitoring SpO2
Overview
The SpO2 value is derived from the absorption of pulse blood oxygen by red and infrared light
using a finger sensor and a SpO2 measuring unit. SpO2 Plethysmogram measurement is employed
to determine the oxygen saturation of hemoglobin in the arterial blood. If, for example, 97 % of
the hemoglobin molecules in the red blood cells of the arterial blood combine with oxygen, then
the blood has a SpO2 oxygen saturation of 97 %. The SpO2 numeric on the monitor will read
97 %. The SpO2 numeric shows the percentage of hemoglobin molecules which have combined
with oxygen molecules to form oxyhemoglobin. The SpO2/PLETH parameter can also provide a
pulse rate signal and a plethysmogram waveform.
- 59 -
Instructions for Use Vista 120 Patient Monitor
Measuring SpO2
1. Select the correct patient category setting (adult/pediatric and neonatal). This is used to
optimize the calculation of the SpO2 and pulse numerics.
2. During measurement, ensure that the application site:
- has a pulsatile flow, ideally with a good circulation perfusion.
- has not changed in its thickness, causing an improper fit of the sensor.
Measurement procedure
1. Switch on the monitor.
2. Attach the sensor to the appropriate site of the patient finger.
3. Plug the sensor extension cable connector into the SpO2 socket on the SpO2 module.
WARNING
Inspect the application site every two to three hours to ensure skin quality and
correct optical alignment. If the skin quality changes, move the sensor to another
site. Change the application site at least every four hours.
NOTE
Injected dyes such as methylene blue or intravascular dyshemoglobins such as
methemoglobin and carboxyhemoglobin may lead to inaccurate measurements.
- 60 -
Instructions for Use Vista 120 Patient Monitor
WARNING
High oxygen levels may predispose a premature infant to retrolental fibroplasia. If
this is a consideration do not set the high alarm limit to 100 %, which is equivalent
to switching the alarm off.
Setting sensitivity
Sensitivity allows the user to set the refresh frequency. High is the highest SpO2 refresh
frequency value. To change the sensitivity:
1. Select the SpO2 Setup menu;
2. Select Sensitivity and set the desired sensitivity from the pop-up list.
- 61 -
Instructions for Use Vista 120 Patient Monitor
Monitoring PR
Overview
The pulse numeric counts the arterial pulsations that result from heart mechanical activity in beats
per minute (bpm) from the measured SpO2 signal.
Setting PR volume
Six selections are available: 0, 1, 2, 3, 4, and 5. 5 is the maximum volume. 0 is mute. PR Volume
can be changed in the PR Setup menu.
- 62 -
Instructions for Use Vista 120 Patient Monitor
Monitoring NIBP
Overview
This monitor uses the oscillometric method for measuring NIBP. It can be used for adult,
pediatric and neonatal patients.
Oscillometric devices measure the amplitude of pressure changes in the occluding cuff as the cuff
deflates from above systolic pressure. The amplitude suddenly increases as the pulse breaks
through the occlusion in the artery. As the cuff pressure decreases further, the pulsations increase
in amplitude, reach a maximum (which approximates to the mean pressure), and then diminish.
In adult and pediatric mode, the blood pressure measurements determined with this device comply
with the American National Standard for Electronic or Automated Sphygmomanometers
(ANSI/AAMI SP10-1992) in relation to mean error and standard deviation, when compared to
auscultatory measurements in a representative patient population. For the auscultatory reference,
the fifth Korotkoff sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements determined with this device comply with the
American National Standard for Electronic or Automated Sphygmomanometers (ANSI/AAMI
SP10-1992) in relation to mean error and standard deviation, when compared to intra-arterial
measurements in a representative patient population.
WARNING
Do not measure NIBP on patients with sickle-cell disease or any condition where
skin damage has occurred or is expected.
WARNING
Use clinical judgment to decide whether to perform frequent unattended blood
pressure measurements on patients with severe blood clotting disorders because
of the risk of hematoma in the limb fitted with the cuff.
WARNING
Before starting a measurement, verify that you have selected a setting appropriate
for your patient (adult, child, or neonate). An incorrect setting may be dangerous
for children, because of higher measurement pressures.
WARNING
Do not apply the cuff to a limb that has an intravenous infusion or catheter in place.
This could cause tissue damage around the catheter when infusion is slowed or
blocked during cuff inflation.
- 63 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Make sure that the air tubing connecting the blood pressure cuff and the monitor is
neither blocked nor tangled.
NOTE
• Do not start NIBP measurements during low battery conditions (the monitor may turn
off automatically).
• If an alarm occurs or measurement fails, discontinue the measurement.
• If you spill liquid onto the equipment or accessories, particularly if there is a chance
that it can get inside the tubing or the measurement device, contact your service
personnel.
• Continuous use of the automatic measuring mode for short intervals may lead to the
discomfort of the patient.
Measurement limitations
Measurements are impossible with heart rate extremes of less than 40 bpm or greater than 240
bpm, or if the patient is on a heart-lung machine.
The measurement may be inaccurate or impossible in the following situations:
• A regular arterial pressure pulse is hard to detect.
• Patients with cardiac arrhythmias.
• Patients with excessive and continuous movement such as shivering or convulsions.
• Patients with rapid blood pressure changes.
• Patients with severe shock or hypothermia that reduces blood flow to the periphery.
• Obese patients, where a thick layer of fat surrounding a limb dampens the oscillations coming
from the artery.
• Patients with an edematous extremity.
Measurement methods
There are three methods of measuring NIBP:
• Manual - measurement on demand.
• Auto - continually repeated measurements (between 1 and 480 minute adjustable interval).
• Continual - NIBP measurements run consecutively in five minutes, and then the monitor
switches to manual mode.
- 64 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Prolonged non-invasive blood pressure measurements in Auto mode may be
associated with purport, ischemia and neuropathy in the limb wearing the cuff.
When monitoring a patient, examine the extremities of the limb frequently for
normal color, warmth and sensitivity. If any abnormality is observed, stop the
blood pressure measurements.
Measurement procedures
1. Connect the air hose and switch on the monitor.
2. Attach the blood pressure cuff to the patient's arm or leg and follow the following
instructions. Ensure that the cuff is completely deflated.
Attach the appropriate size cuff to the patient (For information about cuff size selection, refer
to the chapter “NIBP accessories”), and ensure the symbol "Φ" is over the artery. Ensure that
the cuff is not wrapped too tightly around the limb. Excessive tightness can cause
discoloration and ischemia of the extremity.
NOTE
The width of the cuff should be either 40 % of the limb circumference (50 % for neonates)
or 2/3 of the upper arm length. The inflatable part of the cuff should be long enough to
encircle 50-80 % of the limb. The wrong size of cuff can cause erroneous readings. If you
are unsure what size to use, consult a physician or use a large size cuff.
Cuff Usage
3. Connect the cuff to the air tubing.
4. Check if the patient mode is appropriate. Access the Patient Setup menu from Menu and
select Type. Turn the knob to select the desired patient Type.
5. Select a measurement mode in the NIBP Setup menu. Refer to chapter “Operation Prompts”
for information.
- 65 -
Instructions for Use Vista 120 Patient Monitor
Operation prompts
1. Manual measurement
- Access the NIBP Setup menu and set the Measure Mode item to Manual.
Press the button on the front panel to start a manual measurement.
- During the idle period between AUTO measurements, press the button on the front
panel at any time to start a manual measurement. Press the button again to stop the
manual measurement. This allows the system to continue to take AUTO measurements
according to the selected time interval.
2. Automatic Measurement
Access the NIBP Setup menu and set the Measure Mode item to Auto.
Press the button on the front panel to start the AUTO measurement according to the
selected time interval.
3. Continuous measurement
Access the NIBP Setup menu and select the Continual item to start a continuous
measurement.
The continuous measurement will last 5 minutes.
4. Stopping continuous measurement
During continuous measurement, press the button on the front panel at any time to
stop.
NIBP alarm
When NIBP Alarm is On, the physological alarm occurs if any measurement value of Systolic
pressure, Mean pressure, Diastolic pressure is outside the alarm limits. The user can adjust the
alarm limit by accessing NIBP Setup > Alarm Setup > Sys Alarm/Mean Alarm/Dia Alarm.
Resetting NIBP
When NIBP does not work properly and the system does not generate an error message, select
Reset in the NIBP Setup menu to activate the self-test procedure. This function restores the
system.
- 66 -
Instructions for Use Vista 120 Patient Monitor
Calibrating NIBP
NIBP is not user-calibrated. Cuff-pressure transducers must be verified and calibrated, if
necessary, at least once every two years by a qualified service professional. See the Service
Guide for details.
Leak test
This item is used for the leak test. Turn the knob to select the Leak Test item in the User
Maintain > NIBP Maintain menu to start the air leakage test. When the item is selected, it will
change to Stop. If this item is selected again, the system will stop air leakage test and the item
returns to Leak Test.
WARNING
This pneumatic test, other than being specified in the EN 1060-1 standard, is to be
used by the user to simply determine whether there are air leaks in the NIBP airway.
If, at the end of the test, the system gives the prompt that the NIBP airway has air
leaks, contact the manufacturer for repair.
- 67 -
Instructions for Use Vista 120 Patient Monitor
Monitoring TEMP
Overview
Body temperature is measured using a thermistor probe (a semiconductor whose resistance
changes with temperature) placed appropriately on the patient.
Two TEMP probes can be used simultaneously to measure two TEMP values. The monitor
automatically calculates the temperature difference. The default configuration is axilla sensor for
adult.
- 68 -
Instructions for Use Vista 120 Patient Monitor
Monitoring procedures
Preparatory steps for IBP measurement:
1. Connect the pressure cable to the corresponding socket and switch on the monitor.
2. Flush the system with a normal saline solution. Ensure that the system is free of air bubbles.
3. Connect the patient’s catheter to the pressure line. Ensure there is no air present in the
catheter or pressure line.
4. Position the transducer so that it is at the same level as the patient’s heart, approximately
mid-axillary line.
- 69 -
Instructions for Use Vista 120 Patient Monitor
5. For the label name selection, refer to chapter “Selecting a Pressure for Monitoring”.
6. To zero the transducer, refer to chapter “Zeroing the Pressure Transducer”.
WARNING
If there are air bubbles in the pressure line or the transducer, flush the system with
the infused solution.
- 70 -
Instructions for Use Vista 120 Patient Monitor
If the user needs to perform this procedure himself, he will need the following equipment:
• standard sphygmomanometer
• 3-way stopcock
• tubing (approx. 25 cm long).
- 71 -
Instructions for Use Vista 120 Patient Monitor
7. Inflate to make the mercury bar rise to the setup pressure value.
8. Adjust repeatedly until the value in the menu is equal to the pressure value shown by the
mercury calibration.
9. Press the Start button, the device will begin calibrating.
10. Wait for the calibrated result. Corresponding measures should be taken based on the prompt
information.
11. After calibration, disassemble the blood pressure tubing and the attached 3-way valve.
2 5
- 72 -
Instructions for Use Vista 120 Patient Monitor
IBP alarm
When Alarm Switch is On, the physological alarm occurs if any measurement value of Systolic
pressure, Mean pressure, or Diastolic pressure is outside the alarm limits. The alarm limit can be
adjusted by accessing XX Options (XX stands for the label name) > Setup > SYS Alarm >
MEAN Alarm > DIA Alarm.
- 73 -
Instructions for Use Vista 120 Patient Monitor
- 74 -
Instructions for Use Vista 120 Patient Monitor
NOTE
After the low battery alarm appears, do not start the CO2 measurement. The monitor may
turn off from lack of battery power.
Monitoring procedures
Zeroing the sensor
Zeroing must be performed according to the following procedure when using a new airway
adapter:
1. Expose the sensor to room air, but not to any source of CO2. This includes the ventilator,
patient, and operator.
2. Select CO2 Setup, and change Work Mode from Standby to Measure.
3. In the CO2 Setup menu, select Zero Calibration.
4. If the system briefly displays Zeroing, the process is successful. After the zeroing calibration
is finished, you can start CO2 Monitoring. If the system displays Breath detected or Zero
required, zeroing has failed. Zero calibration must be performed again.
C5 CO2 module
NOTE
You must perform a zero calibration as described in this procedure each time you use a
new airway adapter.
C5 CO2 module
Measurement steps
1. Attach the sensor connector to the CO2 connector on the monitor.
2. Wait two minutes, allowing the sensor to reach its operating temperature and a stable thermal
condition.
3. Choose the appropriate airway adapter and connect it to the sensor head. The airway adapter
clicks into place when seated correctly.
- 75 -
Instructions for Use Vista 120 Patient Monitor
Connecting Sensor
4. To zero the sensor, refer to zeroing the sensor;
5. Install the airway adapter at the proximal end of the circuit between the elbow and the
ventilator Y-section.
WARNING
Accuracy is affected by temperature and barometric pressure.
WARNING
Installing or removing the CO2 module when the monitor is working may make the
monitor unstable. If you do it inadvertently, turn off the module in menu
immediately. The module enters STANDBY mode if you reconnect it to monitor
which it is powered on. If the readings are inaccurate, you should do calibration.
NOTE
• Replace the airway adapter if excessive moisture or secretions are observed in the
tubing or if the CO2 waveform changes unexpectedly without a change in patient
status.
- 76 -
Instructions for Use Vista 120 Patient Monitor
- 77 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Safety and effectiveness of the respiration measurement method in the detection
of apnea, particularly the apnea of prematurity and apnea of infancy, has not been
established.
- 78 -
Instructions for Use Vista 120 Patient Monitor
Freeze
Overview
When monitoring a patient, the user can freeze waveforms and examine them. Generally, the user
should not view frozen waveforms for longer than 60 seconds. The Freeze function has the
following features:
• Freeze status can be activated on any operating screen.
• Upon freezing, the system exits all other operating menus. The system freezes all waveforms
in the Waveform area of the Basic Screen, the Full Lead ECG waveforms and extra
waveforms on the Full Lead ECG interface (if any). The Parameter area refreshes normally.
• The frozen waveforms can be reviewed and recorded.
Press the hardkey on the control panel, and the Freeze menu will appear on the bottom part
of the screen. At the same time, the system freezes the waveforms.
• REC WAVE: can be set to any waveform of 8 s, such as IBP1, CO2, PLETH etc. It can also
be set to OFF.
• Review: Used to review frozen waveforms.
• Exit: The system closes the Frozen menu and exits Freeze status.
NOTE
Pressing the hardkey repeatedly over a short period of time may result in
discontinuous waveforms on the screen.
- 79 -
Instructions for Use Vista 120 Patient Monitor
- 80 -
Instructions for Use Vista 120 Patient Monitor
Review
The monitor provides 120-hour trend data of all parameters, storage of 1200 NIBP measurement
results and 60 alarm events. This chapter gives detailed instruction for review of all data.
Setting resolution
To set an appropriate resolution, select Menu > Review > Trend Graph. Choose Resolution to
display the list and select a resolution ranging from 1 sec, 5 sec, 1 min, 5 min and 10 min.
- 81 -
Instructions for Use Vista 120 Patient Monitor
Record
The monitor can print the data in the current trend window. For more information, refer to
chapter “Strip recording”.
Setting resolution
To set an appropriate resolution, select Menu > Review > Trend Table. Choose Resolution to
display the list and select a resolution ranging from 1 sec, 5 sec, 1 min, 5 min and 10 min, 30
min and 60 min.
- 82 -
Instructions for Use Vista 120 Patient Monitor
Recording
The monitor can make a tabular trend recording of the data in the current trend graph window.
The report uses the current trend interval settings. For more information, refer to chapter
“Recording”.
NIBP review
To review the NIBP measurement data, select the NIBP Review key on the screen or select
Menu > Review > NIBP Review. The NIBP Review window is displayed.
Recording
The monitor can record the measurement data in the NIBP review window. For more information,
refer to chapter “Recording”.
Alarm review
The monitor can store up to 60 alarm events and displays the last 10 alarm events.
To review the alarm event, select the Alarm Review key on the screen or select Menu > Review
> Alarm Review. The Alarm Review dialog box is displayed.
- 83 -
Instructions for Use Vista 120 Patient Monitor
If the user selects User Define, he can define the review time by setting Year, Month, Day,
Hour, Min and Sec. The alarm events occur before the User Define option is displayed on the
alarm event review dialog box.
Recording
The monitor can record the alarm event in the alarm event review window. For more information,
refer to chapter “Recording”.
Arr review
Select ECG Setup > Arr Analysis> Arr Review or Menu > Review > Arr Review to display
the Arr Review dialog box. The dialog box displays the latest arrhythmia events.
Recording
The monitor can record the waveforms of the arrhythmia events displayed on the current screen.
For more information, refer to chapter “Recording”.
- 84 -
Instructions for Use Vista 120 Patient Monitor
NOTE
The drug calculation function acts only as a calculator. The patient weights in Drug
Calculation menu and in Patient Information menu are independent of each other.
Therefore changing the Weight in Drug Calculation menu will not change the weight in
the Patient Information menu.
Drug calculation
Calculation procedures
1. The drug calculation window is displayed by selecting Menu > Common Function > Drug
Calc.
2. Select the right pull-down Drug Name option and select the desired drug name among the 15
drugs which are listed as follows. The names of drug Drug A, Drug B, Drug C, Drug D and
Drug E can be defined by the user.
• Drug A, Drug B, Drug C, Drug D and Drug E
• AMINOPHYLLINE
• DOBUTAMINE
• DOPAMINE
• EPINEPHRINE
• HEPARIN
• ISUPREL
• LIDOCAINE
• NIPRIDE
• NITROGLYCERIN
• PITOCIN
3. The system generates default values that cannot be treated as calculation results. The user
must input the correct parameter values.
4. Input the patient’s weight.
5. Input the correct parameter value.
6. Confirm if the calculation result is correct.
- 85 -
Instructions for Use Vista 120 Patient Monitor
Calculation unit
Each drug has the fixed unit or unit series that is used for it. Within the same unit series, the
numeric value is automatically scaled up or down as needed.
The calculation units of the drugs are as follows:
Drug Unit
DRUG A, DRUG B, DRUG C, AMINOPHYLLINE, g, mg, mcg
DOBUTAMINE, DOPAMINE, EPINEPHRINE,
ISUPREL, LIDOCAINE, NIPRIDE, NITROGLYCERIN
DRUG D, PITOCIN, HEPARIN Ku, mu, Unit
DRUG E mEq
When defining a drug, select Drug A, Drug B, Drug C, Drug D, and Drug E based on the unit
series.
NOTE
• The drug calculation is displayed as invalid prior to data entry.
• Drip Rate and Drop Size are invalid in neonatal mode.
Titration table
After completing the drug calculation, the user can display the Titration on the Drug Calc dialog
box.
The user can change the following items in the titration table:
• Basic
• Step
• Dose Type
The data in the trend table will vary with the changes above. The user can perform the following:
• Scroll up and down the screen by selecting the symbol and displayed on the trend
graph.
• Record the data displayed in the current dialog box by selecting the button Record.
- 86 -
Instructions for Use Vista 120 Patient Monitor
Strip recording
General information
A thermal dot matrix recorder is used for the monitor and can support a number of recording
types and output patient information, measurement data, and review data waveforms.
4 1
Recording type
The monitor provides several types of strip recordings:
• Continuous real-time recording
- 87 -
Instructions for Use Vista 120 Patient Monitor
• Time recording
• Alarm recording
• Trend graph, trend table recording
• Arrhythmia review recording
• Drug calculation titration recording
• NIBP review recording
• Alarm review recording
- 88 -
Instructions for Use Vista 120 Patient Monitor
Proper operation
• When the strip recorder is operating, do not pull the paper.
• Do not operate the strip recorder without record paper.
Paper out
When the Recorder Out Of Paper alarm is displayed, the strip recorder will not start.
Installling paper
1. Pull out the upper arc of the strip recorder casing as shown in the following figure.
- 89 -
Instructions for Use Vista 120 Patient Monitor
2. Insert a new roll of paper into the paper cassette, printing side facing upwards.
- 90 -
Instructions for Use Vista 120 Patient Monitor
NOTE
• If the monitor is not installed with a strip recorder, it will indicate RECORDER SETUP
NEEDED after pressing the Record button in the dialog box.
• Do not touch the thermo-sensitive print head when performing continuous recording.
- 91 -
Instructions for Use Vista 120 Patient Monitor
Other Functions
Nurse call
The monitor provides a dedicated nurse call port which can be connected to the hospital’s nurse
call system.
- 92 -
Instructions for Use Vista 120 Patient Monitor
Using battery
Battery power indicator
The indicator labeled Battery on the front panel of the monitor illuminates green when the
monitor is battery powered and yellow when being charged. The indicator is not illuminated
when the monitor is not powered or when using AC power.
When the monitor is battery powered, the monitor switches off automatically if there is no power.
The performance of rechargeable batteries may deteriorate over time. The battery should be
conditioned regularly as follows:
1. Disconnect the patient from the monitor and stop all monitoring and measurement.
2. Switch the monitor power on and charge the battery for more than 6 hours continuously.
3. Disconnect the monitor and let it run until there is no battery power remaining and the
monitor shuts off.
4. Reconnect the monitor to mains.
This should improve the battery running time. If the running time is less than listed in the
specification, change the battery or contact the customer service.
WARNING
Before using the rechargeable lithium-ion battery (hereinafter called battery), be
sure to read these Instructions for Use and follow all safety precautions.
WARNING
Do not destroy, disassemble or modify the battery in any way. Do not short circuit
the battery, or its terminals, to a metal object.
- 93 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Do not unplug the battery while the monitor is monitoring a patient.
WARNING
Do not heat or throw battery into a fire.
WARNING
Do not store the battery at temperatures above 60 °C.
WARNING
Do not immerse, throw, or wet battery in water/seawater.
WARNING
Use the battery only in the monitor. Do not connect battery directly to an electric
outlet or cigarette lighter charger.
WARNING
If liquid leaking from the battery gets into your eyes, do not rub your eyes. Wash
them well with clean water and go to see a doctor immediately. If liquid leaks of the
battery splash onto your skin or clothes, wash well with fresh water immediately.
WARNING
Keep away from fire immediately when leakage or foul odor is detected.
WARNING
Stop using the battery if abnormal heat, odor, discoloration, damage, deformation,
or abnormal condition is detected during use, charge, or storage. Keep it away
from the monitor.
Battery
door
- 94 -
Instructions for Use Vista 120 Patient Monitor
NOTE
To prolong the life of rechargeable battery, charge it at least once every six months for a
charging time of about five hours.
- 95 -
Instructions for Use Vista 120 Patient Monitor
General points
WARNING
Before cleaning, sterilizing, or disinfecting the monitor or any accessory, make
sure that the equipment is switched off and disconnected from the power line.
Keep the monitor, cables, and accessories free of dust and dirt.
CAUTION
To prevent the device from damage, follow this procedure:
• Always dilute cleaning, sterilizing, or disinfecting agents according to the
manufacturer’s instructions or use the lowest possible concentration.
• Do not immerse any part of the equipment or any accessory in liquid or pour liquid
onto them.
• Never use bleach.
• Do not allow liquid to enter the case.
• Never use abrasive material (such as steel wool or silver polish).
CAUTION
If you spill liquid on the equipment, battery, or accessories, or they are accidentally
immersed in liquid, contact your service personnel or DrägerService.
Cleaning
Cleaning the monitor
Regular cleaning of the monitor shell and the screen are strongly recommended. Use only
non-caustic detergents such as soap and warm water (40 °C maximum) to clean the monitor shell.
Do not use strong solvents such as acetone or trichloroethylene.
Exercise caution when cleaning the monitor screen. Do not permit liquids to enter the monitor
case. Do not allow water or cleaning solution to enter the measurement connectors.
Examples of disinfectants that can be used on the instrument casing are as follows:
• Tenside;
• Diluted Ammonia Water < 3%;
- 96 -
Instructions for Use Vista 120 Patient Monitor
Cables and leadwires can be cleaned with a warm, damp cloth and mild soap, or isopropyl
alcohol wipes. Refer to the previous page for safe cleaning techniques.
- 97 -
Instructions for Use Vista 120 Patient Monitor
An adapter can be cleaned by rinsing in a warm soapy solution, followed by soaking in a liquid
disinfectant, pasteurized, or cold sterilized (glutaraldehyde). It should then be rinsed with sterile
water and dried. Before reusing any adapter, ensure the windows are clean and free of any
residue.
For other applied parts, the manufacturer must be consulted for cleaning, sterilization, or
disinfecting methods.
Other accessories
For cleaning other accessories, contact the manufacturers for details.
Sterilization
To avoid extended damage to the equipment, sterilization is only recommended when stipulated
as necessary in the hospital’s policy. Sterilization facilities should be cleaned first.
Recommended sterilization material: Ethylate and Acetaldehyde.
Appropriate sterilization materials for ECG lead and blood pressure cuff are described in those
chapters.
WARNING
Sterilize and disinfect as needed to prevent cross infection between patients.
CAUTION
Use a moistened cloth to wipe up any agent remaining on the monitor.
Disinfection
WARNING
Mixing bleach and ammonia (or other disinfecting solutions) may produce
hazardous gases.
To avoid extended damage to the equipment, disinfection is only recommended when stipulated
as necessary in the hospital’s policy. Disinfection facilities should be cleaned first.
Appropriate disinfection materials for ECG leads, SpO2 sensor, blood pressure cuff, TEMP probe,
IBP sensor are described in those chapters.
Recommended types of disinfecting agents are:
• Alcohol
• Aldehyde
CAUTION
Do not use etO gas or formaldehyde to disinfect the monitor.
- 98 -
Instructions for Use Vista 120 Patient Monitor
Maintenance
WARNING
Failure on the part of the responsible institution to implement a satisfactory
maintenance schedule may cause undue equipment failure and possible health
hazards.
WARNING
If the monitor is mechanically damaged, or if it is not working properly, do not use
it. Contact your technical personnel.
WARNING
If you discover a problem with any of the equipment, contact your service
personnel or your authorized supplier.
Inspecting
The overall check of the monitor, including the safety check, should be performed by qualified
personnel every 24 months, and following repair.
The following items should be checked:
• If the environmental condition and power supply meet requirements.
• If the power supply cord is damaged and the insulation is not cut.
• If the device and accessories are damaged.
• Specified accessories.
• If the alarm system works properly.
• If the recorder works properly and there is sufficient paper.
• Battery performance
• If all monitoring functions are in good condition.
• If grounding resistance and leakage function properly.
If any damage or abnormality is found, do not use the monitor and contact customer support.
- 99 -
Instructions for Use Vista 120 Patient Monitor
Replacing fuse
To replace the blown fuse:
1. Switch off the monitor and unplug the power cable.
2. Carefully place the monitor upside down on a flat surface covered with cloth or other
protecting pad.
3. Pull ot the fuse box from the groove and you will see the fuse.
4. Remove the old fuse. Take out a new fuse with the following specifications:
Ф 5 mm x 20 mm: Model: T1.6AL 250 V.
5. Place the new fuse in the fuse socket and push the box back into position.
- 100 -
Instructions for Use Vista 120 Patient Monitor
Disposal
CAUTION
The device and its components must be disinfected and cleaned before disposal!
- 101 -
Instructions for Use Vista 120 Patient Monitor
Service policy
Repair of the device may only be carried out by trained service personnel otherwise the correct
functioning of the device may be compromised.
Regular bi-annual maintenance (functional and safety test) according to IEC 62353 is
recommended, in addition to national regulations and laws (for example, accident prevention
regulations).
Connecting the medical device to other medical devices could result in additional maintenance
requirements.
Consult the documentation for these other devices to identify additional requirements.
Dräger recommends contracting with DrägerService for any repairs. Use only authentic Dräger
repair parts during maintenance. Using non-Dräger repair parts may adversely affect the
operation of the device.
If the product fails to function properly, or if assistance is needed, for service and for spare parts,
contact Dräger’s service center. A representative will assist in troubleshooting the problem and
will make every effort to solve it over the phone or e-mail, avoiding potential unnecessary
returns.
In case a return can not be avoided, the representative will record all necessary information and
will provide a Return Material Authorization (RMA) form that includes the appropriate return
address and instructions. An RMA form must be obtained prior to any return.
If you have any questions about maintenance, technical specifications, or malfunctions of devices,
contact your local distributor.
- 102 -
Instructions for Use Vista 120 Patient Monitor
Accessories
Accessories can be ordered from your local Dräger representative.
WARNING
Never reuse disposable transducers, sensors or other accessories that are
intended for single-use or single patient use only. Reuse may compromise device
functionality and system performance and cause a potential hazard.
WARNING
Only use accessories that are approved by Dräger. Using accessories that are not
approved by Dräger may compromise device functionality and system
performance and cause a potential hazard.
WARNING
Do not use a sterilized accessory if its packaging is damaged.
NOTE
Transducers and sensors have a limited shelf life. Refer to the package labeling.
The following cables may not all be available in all countries. Check availability with your local
Dräger supplier.
ECG accessories
The following table lists the optional configuration for the monitor:
- 103 -
Instructions for Use Vista 120 Patient Monitor
SpO2 accessories
Part Number Accessory
NIBP accessories
Part Number Accessory
- 104 -
Instructions for Use Vista 120 Patient Monitor
Temp accessories
Part Number Accessory
IBP accessories
Part Number Accessory
CO2 accessories
Part Number Accessory
Device accessories
Part Number Accessory
2606229 Vista 120 Rechargeable Lithium-Ion Battery
2606231 Vista 120 Printing paper, 4 rolls
2606232 Vista 120 Ground Cable
2606233 Vista 120 Mounting adaptor
Spare parts
Part Number Accessory
2606229 Vista 120 Rechargeable Lithium-Ion Battery
2606260 Fuse
- 105 -
Instructions for Use Vista 120 Patient Monitor
Technical Data
Classification
Protection class Class I equipment and internal powered equipment
EMC class Group I, Class A
Degree of protection against CF: ECG (RESP), TEMP, IBP
electric shock BF: SpO2, NIBP, CO2
Liquid ingress protection IPX1
Disinfection/sterilization method Refer to chapter “Care and Cleaning” for details.
Mode of operation Continuous
Device specifications
Size 408 mm (L) × 157 mm (W)× 316 mm (H)
Weight 7 kg
Ambient conditions
Temperature
Operating +5 to +40 °C
Transport and storage -20 to +55 °C
Relative humidity
Operating 25 to 80 % (non-condensing)
Leakage current
Applied part Normal condition Single fault condition
Earth leakage current <0.5 mA <1 mA
Enclosure leakage current <0.1 mA <0.5 mA
Patient leakage current CF AC: <0.01 mA AC: <0.05 mA
DC: <0.01 mA DC: <0.05 mA
BF AC: <0.1 mA AC: <0.5 mA
- 106 -
Instructions for Use Vista 120 Patient Monitor
Power supply
Power supply 100 to 240 V~, 50 Hz/60 Hz
Pmax = 110 VA
FUSE T 1.6 AL, 250 V
Battery (optional)
Quantity 1
Capacity 4.2 Ah
Battery life 120 min (At 25 °C, continuous SpO2 measuring mode and
NIBP automatic measuring mode)
Battery charge time 320 min (Monitor is on or in standby mode.)
Display
Display Messages
Display screen: 15” color TFT A maximum of 11 waveforms
Resolution: 1024 × 768 One power LED
Two alarm LEDs
One charge LED
Recorder
Record width 48 mm
Paper speed 25 mm/s, 50 mm/s
Trace Up to 3
- 107 -
Instructions for Use Vista 120 Patient Monitor
Review
Trend review
Short 1 hr, 1 sec. resolution
Long 120 hrs, 1 min. resolution
Review 1200 sets NIBP measurement data
ECG
3-, 5-lead monitoring
Lead mode 3-Lead: I, II, III
5-Leads: I, II, III, aVR, aVL, aVF, V
- 108 -
Instructions for Use Vista 120 Patient Monitor
- 109 -
Instructions for Use Vista 120 Patient Monitor
Resolution 0.01 mV
HR averaging method
Method 1 Normally, heart rate is computed by averaging the 12 most
recent RR intervals.
Method 2 If each of three consecutive RR intervals is greater than
1200 ms, then the four most recent RR intervals are
averaged to compute the HR.
Range of sinus and SV rhythm
- 110 -
Instructions for Use Vista 120 Patient Monitor
RESP
Method Impedance between RA-LL, RA-LA
Baseline impedance range 200 to 2500 Ω (cable resistance = 0 K)
2200 to 4500 Ω (leads cables 1KΩ resistance)
Measuring sensitivity 0.3 Ω (baseline impedance 200 to 4500 Ω)
Noise < 0.1 Ω (3, 5-lead monitoring)
Max. dynamic range 500 Ω resistance, 3 Ω variable resistance, no clipping
Waveform bandwidth 0.2 to 2.5 Hz (-3 dB)
- 111 -
Instructions for Use Vista 120 Patient Monitor
RR measuring range:
Adult 0 to 120 rpm
Neo/Ped 0 to 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Gain selection ×0.25, ×0.5, ×1, ×2, ×3, ×4, ×5
Respiration excitation < 400 μA, sinusoid, 62.8 kHz (± 10%)
waveform
NIBP
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring interval in AUTO 1/2/3/4/5/10/15/30/60/90/120/240/480 min
mode
Continuous 5 min, interval is 5 s
Measuring type Systolic Pressure, Diastolic Pressure, Mean Pressure
Alarm type SYS, DIA, MAP
Measuring and alarm rang
Adult mode SYS: 40 to 270 mmHg
DIA: 10 to 215 mmHg
MAP: 20 to 235 mmHg
Pediatric mode SYS: 40 to 200 mmHg
DIA: 10 to 150 mmHg
MAP: 20 to 165 mmHg
Neonatal mode SYS: 40 to 135 mmHg
DIA: 10 to 100 mmHg
MAP: 20 to 110 mmHg
Cuff pressure measuring 0 to 300 mmHg
range
Pressure resolution 1 mmHg
Maximum standard 8 mmHg
deviation
- 112 -
Instructions for Use Vista 120 Patient Monitor
SpO2
Measuring Range 0 to 100 %
Alarm Range 0 to 100 %
Resolution 1%
Accuracy
Adult (including Pediatric) ±2 % (70 to 100 % SpO2)
Undefined (0 to 69 % SpO2)
Neonate ±3 % (70 to 100 % SpO2)
Undefined (0 to 69 % SpO2)
Pulse Rate
Measuring range 25 to 300 bpm
Alarm range 30 to 300 bpm
Accuracy ±2 bpm
Data update period 1s
Wave length
Red light 660±3 nm
Infrared light 905±5 nm
- 113 -
Instructions for Use Vista 120 Patient Monitor
TEMP
Channel 2
Measuring range 0 to 50 °C
Alarm range 0 to 50 °C
Sensor type YSI-10K
Resolution +0.1 °C
Accuracy (without sensor) ±0.1 °C
Refresh time Every 1 to 2 s
IBP
Pressure measuring range -50 to +300 mmHg
Resolution 1 mmHg
Accuracy (without sensor) ± 2 % or ±1 mmHg, whichever is larger
Pressure sensor
Sensitivity 5 (μV/V/mmHg)
Impedance 300 to 3000 Ω
Frequency response d.c. to 12.5 or 40 Hz
Zero Range: ±200 mmHg
Accuracy: ±1 mmHg
Measuring and alarm range
Art 0 mmHg to 300 mmHg
PA -6 to +120mmHg
CVP/RAP/LAP/ICP -10 to +40 mmHg
P1/P2 -50 to +300 mmHg
Volume displacement of MSI 4.5 x 10-4 in3 / 100 mmHg
CO2
Method Infra-red Absorption Technique
Unit mmHg, %, kPa
Measuring range
EtCO2 0 to 150 mmHg
- 114 -
Instructions for Use Vista 120 Patient Monitor
FiCO2 3 to 50 mmHg
AwRR 0 to 150 rpm (MainStream)
Resolution
EtCO2 1 mmHg
FiCO2 1 mmHg
AwRR 1 rpm
EtCO2 Accuracy ± 2 mmHg, 0 to 40 mmHg
± 5 % of reading, 41 to 70 mmHg
± 8 % of reading, 71 to 100 mmHg
± 10 % of reading, 101 to 150 mmHg
AwRR Accuracy ± 1 rpm
Apnea Alarm Delay 10 s, 15 s, 20 s (default), 25 s, 30 s, 35 s, 40 s
Calculation Method BTPS (Body Temperature Pressure Saturated)
Sample Gas Flowrate 50 ml/min
Stability
Short Term Drift Drift over 4 hours < 0.8 mmHg
Long Term Drift 120 hour period
O2 Compensation
Range 0 to 100 %
Resolution 1%
Default 16 %
Response time 60 ms
- 115 -
Instructions for Use Vista 120 Patient Monitor
EMC Declaration
General information
The EMC compliance of the medical device is also applicable to the external cables, transducers,
and accessories specified in the list of accessories. In addition, accessories which do not affect
EMC compliance may be used if no other reasons forbid their use (see other sections of the
Instructions for Use). The use of non-compliant accessories may result in increased emissions or
decreased immunity of the medical device.
The medical device may only be used adjacent to or stacked with other devices when the
configuration is approved by Dräger. If adjacent or stacked use of non-approved configurations is
inevitable, the medical device must be observed to verify normal operation in the said
configuration. In any case, strictly observe the Instructions for Use of the other devices.
Electromagnetic emissions
When selecting wireless systems (wireless communication devices, pager systems, etc.) for use in
installations where wireless networking is used, ensure that operating frequencies are compatible.
Low-level signals such as ECG are particularly susceptible to interference from electromagnetic
energy. While the equipment meets the testing described below, it will not ensure perfect
operation. The ‘quieter’ the electrical environment the better. In general, increasing the distance
between electrical devices decreases the likelihood of interference.
Electromagnetic environment
The medical device is intended for use in an electromagnetic environment as specified below.
The user must ensure its use in such an environment.
- 116 -
Instructions for Use Vista 120 Patient Monitor
Electromagnetic immunity
The medical device is intended for use in an electromagnetic environment as specified below.
The user must ensure its use in such an environment.
- 117 -
Instructions for Use Vista 120 Patient Monitor
⎡ 3 .5 ⎤
d =⎢ ⎥ P
⎣ V1 ⎦
⎡ 3 .5 ⎤
d = ⎢ ⎥ P 80 MHz
⎣ E1 ⎦
to 800 MHz
⎡7⎤
d = ⎢ ⎥ P 800
⎣ E1 ⎦
MHz to 2.5 GHz
Where P is the
maximum output power
rating of the transmitter
in watts (W) according
to the transmitter
manufacturer and d is
the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey1, should be less
than the compliance
level in each frequency
range2.
Interference may occur
in the vicinity of
equipment marked with
the following symbol:
- 118 -
Instructions for Use Vista 120 Patient Monitor
- 119 -
Instructions for Use Vista 120 Patient Monitor
- 120 -
Instructions for Use Vista 120 Patient Monitor
Default settings
This appendix documents the most important default settings of the monitor as it is delivered
from the factory.
Note: If the monitor has been ordered preconfigured to your requirements, the settings at delivery
will be different from those listed here.
- 121 -
Instructions for Use Vista 120 Patient Monitor
- 122 -
Instructions for Use Vista 120 Patient Monitor
RESP
RESP settings ADU PED NEO
Alarm switch On
Alarm record Off
Alarm level Medium
Alarm high limit 30 30 100
Alarm low limit 8 8 30
Apnea time 20 s
Calculation type Auto
Resp hold type Auto
Sweep 12.5 mm/s
Amplitude 1
SpO2
SpO2 settings ADU PED NEO
Alarm switch On
Alarm record Off
Alarm level Medium
Alarm high limit 100 100 95
Alarm low limit 90 90 88
Pitch tone Off
Sweep 12.5 mm/s
PR
PR settings ADU PED NEO
PR source SpO2
Alarm switch On
Alarm record Off
Alarm level Medium
Alarm high limit 120 160 200
Alarm low limit 50 75 100
Pulse volume 3
- 123 -
Instructions for Use Vista 120 Patient Monitor
NIBP
NIBP settings ADU PED NEO
Alarm switch On
Alarm record Off
Alarm level Medium
Alarm high limit (SYS) 160 120 90
Alarm low limit (SYS) 90 75 40
Alarm high limit (Mean) 110 90 70
Alarm low limit (Mean) 60 50 30
Alarm high limit (Dia) 90 70 60
Alarm low limit (Dia) 50 40 20
Inflation value 160 140 100
Unit mmHg
Interval Manual
TEMP
TEMP settings ADU PED NEO
Alarm switch On
Alarm record Off
Alarm level Medium
Alarm high limit (T1) 39.0 39.0 39.0
Alarm low limit (T1) 36.0 36.0 36.0
Alarm high limit (T2) 39.0 39.0 39.0
Alarm low limit (T2) 36.0 36.0 36.0
Alarm high limit (TD) 2.0 2.0 2.0
Unit °C
- 124 -
Instructions for Use Vista 120 Patient Monitor
IBP
IBP settings ADU PED NEO
Alarm switch On
Alarm record Off
Alarm level Medium
Unit mmHg
Filter 12.5 Hz
SYS, DIA, SYS, DIA, MEAN SYS, DIA, MEAN
MEAN
Alarm high limit 160, 90, 110 160, 90, 110 160, 90, 110
(ART, P1, P2)
Alarm low limit 90, 50, 70 90, 50, 70 90, 50, 70
(ART, P1, P2)
Alarm high limit (PA) 35, 16, 20 35, 16, 20 35, 16, 20
Alarm low limit (PA) 10, 0, 0 10, 0, 0 10, 0, 0
MEAN MEAN MEAN
Alarm high limit 10 10 10
(CVP, RAP, LAP, ICP)
Alarm low limit 0 0 0
(CVP, RAP, LAP, ICP)
CO2
CO2 settings ADU PED NEO
Alarm switch On
Alarm record Off
Alarm level Medium
Work mode Measure
Unit mmHg
Apnea time 20 s
O2 compensate 16 %
Anes agent 0%
Alarm high limit (EtCO2) 50 50 45
Alarm low limit (EtCO2) 15 20 30
- 125 -
Instructions for Use Vista 120 Patient Monitor
- 126 -
Instructions for Use Vista 120 Patient Monitor
- 127 -
Instructions for Use Vista 120 Patient Monitor
Passwords
The following dialog boxes are protected by a password to prevent unauthorized changes:
• User Maintain
• Demo Mode
The password can be found on this page of these Instructions for Use. Cut out this section and
store it in a safe place to keep it from unauthorized persons.
If the section with the password has been removed, ask the person responsible for the device
about the possibility of making changes in the above mentioned dialog boxes.
Cut out this section and store it in a safe place to keep it from unauthorized persons.
The following dialog boxes are protected by a password to prevent unauthorized changes:
- 128 -
These instructions for use apply only to
Vista 120 SW 1.n
with the Serial No.:
If no Serial No. has been filled in by Dräger,
these Instructions for Use are provided for
general information only and are not intended
for use with any specific machine or unit.
This document is provided for customer
information only, and will not be updated or
exchanged without customer request.
Directive 93/42/EEC
concerning Medical Devices
Manufacturer: