Professional Documents
Culture Documents
0.06
Introduction 0.04
0.02
Calibration certificates for indicating instruments, such as 0.00
Correction
thermometers, pressure gauges, and digital voltmeters -0.02
(DVMs), typically report a table of corrections for a mod- -0.04
-0.06
est number of calibration points. Usually the number of
-0.08
points is just enough for the calibration laboratory to con- -0.10
fidently characterise the major features of the instrument’s -40 -20 0 20 40 60 80 100 120 140 160
performance. Table 1 shows an example of a table from a Instrument reading
certificate for a digital thermometer. Figure 1. A plot of the data in Table 1. The straight lines joining
neighbouring points indicate the corrections obtained by linear
Table 1. Calibration data from a certificate. interpolation.
Reading (°C) Correction (°C) Uncertainty (°C)
–19.88 0.05 0.02 The value of the correction can also be calculated from
0.07 0.01 0.02 the numbers given in the table. The equation for each line
segment in the graph is
20.16 –0.03 0.02
39.88 –0.06 0.03 ( x − xL ) ( x − x)
y = yH + yL H ,
60.00 –0.08 0.03 ( xH − xL ) ( x H − xL )
80.10 –0.08 0.03
where:
100.09 –0.07 0.03
y is the correction value to be calculated,
119.89 –0.03 0.04
x is the reading for which the correction is required,
139.87 0.03 0.04
xL is the instrument reading at the left of the segment,
yL is the correction corresponding to xL.
When the instrument is used, it is common to find that xH is the instrument reading at the right of the segment,
there is no value in the table close to the instrument read- yH is the correction corresponding to xH,
ing, and the question arises, “what value of the correction These definitions are explained graphically in Figure 2,
should I use?” For example, we might want to know the which shows an expanded region around 28.5 °C. If we ap-
value of the correction when the instrument is reading ply the equation to determine the correction value for
28.5 °C. 28.5 °C, using the closest two readings surrounding the
This guide describes two methods for calculating the value 28.5 °C, we get
values for corrections between the entries reported in the
calibration certificate.
0.06
Method 1: Linear Interpolation 0.04
0.02
Figure 1 shows a graph of the correction versus instrument 0.00
Correction
(xL, yL)
-0.02
reading. Each point plotted on the graph corresponds to the
-0.04 (xH, yH)
instrument reading and correction reported in one row of -0.06 y
the calibration certificate. Also shown on the graph are -0.08 x
straight lines joining neighbouring points. The straight -0.10
0 20 40 60
lines may be used directly to read values for corrections off Instrument reading
the graph. For example, if we look along the horizontal axis
of the graph, we can see that for an instrument reading near Figure 2. Expanded region of Figure 1 showing the
definitions of the variables in the interpolating equation.
30 °C, the correction is near −0.05 °C.
0.03
alog box with the option for adding a trendline. Select
‘Add Trendline’ and an extra line should appear on the 0.02
Contact Details
Postal address: Measurement Standards Laboratory, Callaghan Innovation, PO Box 31-310, Lower Hutt 5040,
New Zealand.
Website: www.measurement.govt.nz
E-mail: info@measurement.govt.nz
The Measurement Standards Laboratory is New Zealand’s national metrology institute, operating under the authority of
the New Zealand Measurement Standards Act 1992. The Measurement Standards Laboratory is part of Callaghan
Innovation, a New Zealand Crown entity.
This work is licensed under a Creative Commons Attribution-No Derivatives 3.0 Licence
(CC BY-ND 3.0 NZ, see https://creativecommons.org/licenses/by-nd/3.0/nz/).
This licence allows for redistribution, commercial and non-commercial, as long as the work is passed
along unchanged and in whole, with appropriate credit.