Professional Documents
Culture Documents
Selecting The
Correct Clamp-On
I1 I2
Current Probe
▼
Answering the following questions
will help you to select the appropriate Z
probe for your applications.
1. Determine if you are measuring B1 (N1) B2 (N2)
AC or DC (DC current probes Iron core
are categorized as AC/DC
because they measure both). Figure 1
2. What is the the maximum current
you will measure, and what is
the minimum current you will
Introduction AEMC offers the widest selection of
current probes available to measure
measure? Check that the accuracy Clamp-on current probes are designed AC or DC current. Several AEMC
at low levels is appropriate, or to extend the current measuring probes are patented for their unique
select a low current measurement capabilities of DMMs, power instru- circuitry and design and the majority
probe. Most probes perform with ments, oscilloscopes, hand-held are UL approved.
greater accuracy at the upper scopes, recorders or loggers and
end of their range. Several other diverse instruments. The probe AC Clamp-On
probes are designed to measure is “clamped” around the current
very low DC or AC. carrying conductor to perform non- Current Probes
3. What size conductor will you clamp contact current measurements without Theory of Operation
onto? This parameter determines interrupting the circuit under test. The An AC clamp-on current probe may
the probe jaw size needed. probe outputs a current or voltage be viewed as a variance of a simple
signal directly proportional to the current transformer.
4. What type of probe output do
measured current, thereby providing
you need or can you work with
current measuring and displaying A transformer (Figure 1) is essentially
(mA, mV, AC, DC, etc.)? Check
the maximum receiver imped- capabilities to instruments with low two coils wound on a common iron
ance to ensure that the probe current or voltage inputs. core. A current I1 is applied through
will perform to specifications. When making a measurement, the the coil B1, inducing through the
current carrying conductor is not common core a current I2 in the coil
Other factors you may want to consider: B2. The number of turns of each
broken and remains electrically isolated
What is the working voltage of the from the meter input terminals. As a coil and the current are related by
conductor to be measured? AEMC result, the meter’s low input terminal N1 x I1 = N2 x I2, where N1 and N2
probes must not be used above may be either floated or grounded. It are the number of turns in each coil.
600V (see specifications)! From this relationship:
is not necessary to interrupt the power
What type of output termination do supply when using a clamp-on current I2 = N1/N2 x I1 or I1 = N2/N1 x I1
you need: lead with BNC, lead with probe for taking measurements, so
4mm safety banana plugs or jacks This same principle is applied to a
costly down time can be eliminated.
to accept 4mm leads? clamp-on current probe (Figure 2).
True RMS measurements within The articulated magnetic core holds
Will the probe be used for harmonics the probe frequency
or power measurements? Look at response are possible by
the frequency specifications and using most AEMC current Conductor/cable
▼
→
(Id) is held constant, the magnetic
field (B) is directly proportional to the • Select the function and range.
→
current in a conductor. Thus, the Hall • Clamp the probe around a single
output voltage (Vh) is representative Hall Air conductor.
of that current. Such an arrangement generator gap
has two important benefits for universal • Read the conductor’s current value.
Conductor
current measurement. Example (Figure 8):
Figure 7
First, since the Hall voltage is not AC: Probe Model: MD303
dependent on a reversing magnetic Ratio: 1000:1
field, but only on its strength, the The probe outputs are in mV (mVDC Output: 1mAAC/AAC.
device can be used for DC measure- when measuring DC, and mVAC when DMM: Set to mAAC range
ment. Second, when the magnetic measuring AC) and may be connected DMM Reading: 125mAAC
field strength varies due to varying to most instruments with a voltage Current in Conductor:
current flow in the conductor, input, such as DMMs, loggers, 125mA x 1000 = 125AAC
response to change is instantaneous. oscilloscopes, hand held scopes,
DC: Probe Model: MR521
Thus, complex AC wave forms may recorders, etc.
1mVDC/ADC (Hall sensor)
be detected and measured with high AEMC also offers different technologies DMM: Set to mVDC range
accuracy and low phase shift. for DC measurements such as in DMM Reading: 160mVDC
The basic construction of a probe jaw the K100 and K110 designed to Current in Conductor: 160ADC
assembly is shown in Figure 7, (Note: measure very low DC currents and
AC: Probe Model: MR411
one or two Hall generators are used using saturated magnetic technology.
Output: 1mVAC/AAC (Hall sensor)
depending on the type of current probe). AC/DC probes also offer the opportunity DMM: Set to mVAC range
The many AEMC AC/DC Current to display or measure True RMS in DMM Reading: 120mVAC
Probes were developed based on the AC or AC + DC. Current in Conductor: 120AAC
above principle, together with patented
DC: Micro Probe Model K100
electronic circuitry incorporating signal
Output: 1mVmA
conditioning for linear output and a
DMM: Set to mVDC range
temperature compensation network.
DMM Reading: 7.4mVDC
These have a wide dynamic range
Current in Conductor: 7.4mADC
and frequency response with highly
accurate linear output, for application
in all areas of current measurement
up to 1500A. Direct currents can be
measured without the need of
expensive, power consuming shunts,
and alternating currents up to several Power Load
kHz can be measured with fidelity to source
respond to the requirements of complex
signals and RMS measurements.
Id
Vh
B
Figure 6 Figure 8
▼
Probe Model: SR604
▼
I+IL
Ratio: 1000:1
DMM: Set to mAAC range
Turns in Probe Jaw: 10
DMM Reading: 60mAAC
Current in Conductor:
60mA x 1000/10 = 6000mA = 6A
When the probe is clamped around
two conductors with different
polarities, the resulting reading will
be the difference between the two
currents. If the currents are the same,
the reading will be zero (Figure 11). Motor
When a reading other than zero is
obtained, the reading is the amount IL
of leakage current on the load. To
measure low currents or leakage, Figure 11
1 BE
Figure 12
© 2002 Chauvin Arnoux®, Inc. d.b.a. AEMC ® Instruments Selection Guide to Clamp-On Current Probes (pdf) 07/02
Repair and Calibration Service - for information on repair & calibration, obtaining a user manual:
repair@aemc.com
Technical and Product Application Support –for technical and application support:
techinfo@aemc.com
Chauvin Arnoux
190, rue Championnet
75876 Paris Cedex 18, France
33 1 44 85 45 28 • Fax 33 1 46 27 73 89
info@chauvin-arnoux.com