You are on page 1of 9

Talanta 56 (2002) 365– 373

www.elsevier.com/locate/talanta

Diffusion coefficient measurements in microfluidic devices


Christopher T. Culbertson, Stephen C. Jacobson, J. Michael Ramsey *
Oak Ridge National Laboratory, Chemical Sciences Di6ision, P.O. Box 2008, Oak Ridge, TN 37831 -6142, USA

Received 25 May 2001; received in revised form 7 September 2001; accepted 18 September 2001

Abstract

Four methods for measuring diffusion coefficients were compared on a microfabricated fluidic device using
rhodamine 6G as the analyte. The measurements were made using a static imaging method and three dynamic
methods—stopped flow, varying the applied potential (E-field method), and varying the detection length (length
method). Under conditions where analyte–wall interactions (adsorption) are minimized, e.g. in a 50/50 (v/v)
methanol/aqueous buffer, the stopped flow (2.71 9 0.09× 10 − 6 cm2 s − 1), E-field (2.684 90.005× 10 − 6 cm2 s − 1) and
the static imaging (2.69 9 0.02× 10 − 6 cm2 s − 1) measurements were all within experimental error of one another and
previously reported values. Under 100% aqueous conditions, however, the diffusion coefficient measured dynamically
was 11% larger than that measured statically. Diffusion coefficients for rhodamine B, fluorescein, 2%,7%dichloro-fluores-
cein (DCF), rhodamine 6G, tetramethylrhodamine labeled glutamic acid and isoleucine, and fluorescein conjugated
bovine serum albumin and ovalbumin were also measured using the static imaging method. © 2002 Elsevier Science
B.V. All rights reserved.

Keywords: Diffusion coefficient; Microfluidic device; Rhodamine; Fluorescein

1. Introduction theoretical plates generated (N), is inversely pro-


portional to the diffusion coefficient (D) for a
The ability to quickly and accurately measure given applied electric field strength (E) and migra-
diffusion coefficients is important to assessing and tion distance (l) as shown in Eq. (1),
interpreting the quality of experimental results vekEl
obtained from liquid chromatography and capil- N= (1)
2D
lary/microchannel electrophoresis (CE). For ex-
ample, in CE under ideal conditions, the only where vek is the electrokinetic mobility of an
significant source of band broadening is longitudi- analyte. This equation is generally used to assess
nal (molecular) diffusion. The separation effi- how well a particular separation method is func-
ciency, therefore, as measured by the number of tioning, but often requires an estimation of an
analyte’s diffusion coefficient which gives rise to
* Corresponding author. Tel.: +1-865-574-5662; fax: + 1-
errors and uncertainties in the results. The actual
865-574-8363. analyte diffusion coefficients are generally not
E-mail address: ramseyjm@ornl.gov (J. Michael Ramsey). measured because of the time and effort involved.

0039-9140/02/$ - see front matter © 2002 Elsevier Science B.V. All rights reserved.
PII: S 0 0 3 9 - 9 1 4 0 ( 0 1 ) 0 0 6 0 2 - 6
366 C.T. Culbertson et al. / Talanta 56 (2002) 365–373

Three methods, however, for measuring diffusion an electric field, i.e. statically, by actually imaging
coefficients in CE have been reported: ‘on-the-fly- the analyte band in the capillary column over time
by-electrophoresis’ [1,2], ‘stopped flow’ [1,2], and would remove the contribution of these four
the E-field method [2]. These methods can be sources to the peak variance assuming that the
considered dynamic methods because the analyte equilibration time for any adsorption–desorption
is moving under the influence of an electric field process is small compared to the diffusion coeffi-
while the measurement is being made. In the cient measurement time. One static diffusion co-
on-the-fly-by-electrophoresis method the peak efficient measurement method has been developed
width, measured as the temporal peak variance, is in a capillary [3]; however, given the spatial extent
first converted to a spatial peak variance (| 2) and of the typical initial plug width (determined, in
then using the Einstein–Smoluchowski equation large part, by the injection plug length), it is
(Eq. (2)), difficult to evenly illuminate and image a large
enough section of capillary to perform this type of
| 2 =2Dt (2)
measurement. Performing static diffusion coeffi-
and the run time (t), a diffusion coefficient is cient measurements in channels on microfabri-
calculated. If the peak variance due to the injec- cated devices, however, is easier as the injection
tion is significant, it can be calculated from the plug lengths are shorter, and therefore, the area
injection parameters and subtracted from the total on the chip that needs to be illuminated and
peak variance prior to calculating the diffusion imaged is smaller. In addition, the planar surfaces
coefficient [2]. For the stopped flow method, two on the chip also make the imaging optics simpler.
consecutive electrophoretic runs are made. In the The ability to perform rapid diffusion coeffi-
first run the analyte is injected and electromi- cient measurements on microchips is also crucial
grated through the detection window. In the next to continuing the development of microfluidics
run the analyte is injected and migrated part of technology. While microfabricated fluidic devices
the way to the detector at which point the electric have been successfully demonstrated for a wide
field is removed for a specified period of time. variety of electrokinetic separations [4–7], the
After this time the electric field is reapplied, and separations must perform at or near their theoret-
the analyte is electromigrated through the detec- ical limit to be successful given the short channel
tion window. The difference in the spatial peak lengths and run times. Being able to quickly mea-
variances between the two runs and the stopped sure analyte diffusion coefficients provides the
flow time are used to calculate the diffusion coeffi- rapid feedback necessary to make such
cient using the Einstein equation (Eq. (2)). In the assessments.
E-field method, a series of electrophoretic runs is In this paper, we compare three dynamic meth-
made between which the field strength is varied ods with a static imaging method for measuring
but kept low to minimize any Joule heating ef- diffusion coefficients on microfabricated devices.
fects. The analyte spatial peak variance for each These comparisons are made using the fluorescent
of these runs is then plotted against the migration dye rhodamine 6G under conditions where ana-
time. The slope of this plot is equal to 2D. lyte–wall interactions are minimized by using a
For all of the dynamic methods described methanol/aqueous buffer and under aqueous con-
above, the assumption is made that the only ditions where significant analyte–wall interactions
source of peak variance is longitudinal diffusion. occur. The precision and accuracy of the measure-
This is often not the case as the initial analyte ments are evaluated and show that the static
injection width, analyte– wall interactions (ad- diffusion coefficient measurement method is in-
sorption), Joule heating and electrodispersion are deed accurate, fast and robust. Seven other com-
also frequently significant contributors. These pounds— fluorescein, 2%,7%-dichlorofluorescein
methods, therefore, may not reliably determine (DCF), rhodamine B, tetramethylrhodamine
the actual longitudinal diffusion coefficient. Mea- derivatized isoleucine and glutamic acid, and
suring the diffusion coefficient in the absence of fluorescein conjugated ovalbumin and bovine
C.T. Culbertson et al. / Talanta 56 (2002) 365–373 367

serum albumin—were also measured using the 2.3. Microchip operation


static method under aqueous conditions and their
diffusion coefficients are reported. The electrokinetic runs and constant volume
injections [9] were performed using four indepen-
dent and remotely programmable high voltage
2. Experimental
(0–10 kV) power sources from a multiple source
2.1. Reagents supply (2866, Bertan, Hicksville, NY). The proper
potentials to apply at each reservoir were deter-
2%,7%-Dichlorofluorescein was obtained from mined using Kirchhoff’s rules and Ohm’s Law.
Aldrich (Milwaukee, WI). Bovine serum albumin The high voltage control and data acquisition
fluorescein conjugate (BSA; 5.5 mol dye mol − 1), software were written in-house using LABVIEW
ovalbumin fluorescein conjugate (Ova; 2.8 mol (National Instruments). For the dynamic diffu-
dye mol − 1) and tetramethylrhodamine isothio- sion coefficient measurements the analyte was
cyanate (TRITC) were obtained from Molecular simply electromigrated through the detection
Probes Inc. (Eugene, OR). Fluorescein (Fl), glu- zone. For the static diffusion coefficient measure-
tamic acid (Glu) and isoleucine (Ile) were ob- ments the analyte was electromigrated into a sec-
tained from Sigma (St. Louis, MO). Rhodamine B tion of the channel imaged by a CCD camera
(RhB) and rhodamine 6G (R6G) were obtained (Fig. 1), and when in the center of this region the
from Eastman Chemical Co. (Kingsport, TN). All potentials were removed until the static measure-
chemicals were used as received. The amino acids ments were complete.
were individually derivatized with TRITC as pre-
viously reported [8]. All solutions were made us-
ing distilled deionized water from a Barnstead
Nanopure System (Dubuque, IA) and then
filtered through 0.45 mm Acrodiscs (Gelman Sci-
ences, Ann Arbor, MI).

2.2. Microchip design

A simple cross chip design was used in both the


static and dynamic diffusion coefficient measure-
ments (Fig. 1). The chips were fabricated on
soda-lime glass (Telic Company, Santa Monica,
CA) using standard photolithographic, wet chemi-
cal etching, and cover plate bonding techniques
[9]. Access holes to the channels were ultrasoni-
cally drilled (Sonic-Mill, Albuquerque, NM) into
one of the slides prior to bonding. Small fluid
reservoirs (140 ml capacity) were attached with
Epo-tek 353ND epoxy (Epoxy Technologies, Inc.,
Billerica, MA) at points where the access holes
were drilled. Channel widths were measured using
a stylus-based surface profiler (P-10, Tencor,
Mountain View, CA) and are reported as the
width at half the channel depth. Seven chips were
Fig. 1. Schematic of microchip used for diffusion coefficient
used in the experiments reported below and had measurements. The detection points and the imaged area for
depths ranging from 10 to 15 mm and widths at the various methods are shown in the schematic. The reported
half-depth ranging from 30 to 46 mm. detection distances were measured from the cross intersection.
368 C.T. Culbertson et al. / Talanta 56 (2002) 365–373

2.4. Single point detection for dynamic diffusion 2.5.2. E-field method
coefficient measurements In the E-field method several consecutive runs
were made in which the field strength was varied
All of the dynamic diffusion coefficient experi- from 46 to 370 V cm − 1, and the detection dis-
ments were performed using a single point detec- tance was set at 0.5 cm (Fig. 1). The peak vari-
tion system similar to that previously described ance was again plotted against the peak migration
[10]. Laser induced fluorescence (LIF) detection of time to obtain the diffusion coefficient from the
the analytes was performed using either the 488 slope of the plot.
(Fl, BSA, Ova) or 514 nm (DCF, RhB, R6G,
TRITC labeled amino acids) line of an Innova 90
2.5.3. Length method
Argon ion laser (Coherent, Inc., Palo Alto, CA)
In the length method several consecutive runs
as the excitation source at a power of 8 mW. The
were made with detection lengths of 0.2, 0.4, 0.6
laser beam was focused through a 300 mm focal
and 0.8 cm (Fig. 1). The peak variance was plot-
distance planoconvex lens to create an estimated
ted against the peak migration time to obtain the
excitation spot size of 50 mm on the chip. The
diffusion coefficient from the slope of the plot.
fluorescence was collected using a 40× micro-
scope objective (CD-240-M40X; Creative Devices,
Neshanic Station, NJ), and the image was focused 2.6. Imaging for static diffusion coefficient
onto a 800 mm pinhole. The signal passing measurements
through the pinhole was then spectrally filtered
using a 488 or 514 nm notch filter (Kaiser Optical The static diffusion coefficient measurements
Systems, Inc., Ann Arbor, MI) and a 530 or 580 were made using a TE300 inverted microscope
nm bandpass filter (530df30 or 580df30; Omega outfitted with an epifluorescence attachment from
Optical, Brattleboro, VT) before being measured Nikon, a Micromax 512 BFT CCD camera from
by a photomultiplier tube (PMT, 77348; Oriel Princeton Instruments (Trenton, NJ), and a Unib-
Instruments, Inc., Stratford, CT). The signal from litz electronic shutter for the high pressure mer-
the PMT was amplified using an SR570 low noise cury arc lamp from Vincent Associates
current preamplifier (Stanford Research Systems, (Rochester, NY). A BV1 filter cube which con-
Inc., Sunnyvale, CA) with a 100 Hz lowpass filter. tained the excitation, dichroic, and emission filters
The signal from the amplifier was sampled at 100 (Nikon) was used for the RhB, R6G, and TRITC
Hz using a PCI-MIO-16XE50 multifunction I/O labeled amino acid experiments, and a BV2 filter
card (National Instruments, Inc., Austin, TX) in a cube (Nikon) was used for the Fl, DCF, and
G3-300 Power Macintosh. All data analysis was protein measurements. Twenty-five images of the
performed using Igor Pro from Wavemetrics, Inc. sample plug in the channel were taken one every 5
(Lake Oswego, OR). s for RhB, DCF, R6G, Fl, TRITC labeled amino
acids or one every 20 s for the proteins. The
2.5. Dynamic diffusion coefficient measurements
exposure time for the small molecules was 200 ms
2.5.1. Stopped flow measurements per image and for the proteins was 1000 ms per
For the stopped flow method two consecutive image. The sample was only illuminated during
runs were made. In the first run R6G was injected the time that the image was taken. The data was
and electromigrated past the detection window. acquired using IPLab Spectrum (Scanalytics,
Immediately, thereafter, a second run was made. Fairfax, VA) and analyzed using scripts created in
This time, however, the R6G was electromigrated IPLab and macros developed using IGOR Pro.
half way to the detection window at which point To obtain accurate temperature measurements,
the potential was removed for anywhere between a thermistor (ON-402-PP with an HH42 digital
4 and 8 min —the ‘stop time’. The ‘stop time’ was thermometer; Omega Stamford, CT) was used for
set such that the ratio of the first to second run both the static and dynamic methods. For the
variance was B0.1. dynamic methods, the microchip was contained in
C.T. Culbertson et al. / Talanta 56 (2002) 365–373 369

Fig. 2. (A) Fluorescence images of R6G diffusing over time in a microchip obtained for use with the static imaging method. (B)
Profiles obtained from part A by summing across the width of the channel. Each profile was fitted by a Gaussian equation (gray
dotted lines) to extract the peak variance.

a light tight box which also served to reduce air hydrodynamic potential was applied. Fig. 2A
circulation, and the thermistor was placed in the shows the static diffusion of rhodamine 6G over
box on the surface of the chip. For the static time in a series of images. These images were flat
method the thermistor was also placed on the field corrected for light intensity variations, and
surface of the chip. In addition, for both methods axial concentration profiles of the analyte were
the reservoirs on the microchip were covered with extracted by summing across the width of the
parafilm to reduce solvent evaporation. channel (Fig. 2B). The variance of each peak was
then determined by fitting a Gaussian function to
the peak profile. Finally, the peak variance (| 2)
3. Results and discussion was plotted against the diffusion time (t). The
slope of the linear regression through this data is
The static diffusion coefficients were measured equal to two times the diffusion coefficient (D),
in the absence of fluid flow, i.e. no electrical or assuming that diffusion obeys the Einstein–
370 C.T. Culbertson et al. / Talanta 56 (2002) 365–373

Smoluchowski relation (Eq. (2)) (Fig. 3A). Good shuttered when data were not being taken so no
linearity was obtained from the linear regression photobleaching was observed as seen by the con-
with correlation coefficients (r) ranging from stant value of the peak area over the time course
0.99993 to 0.99998. The diffusion coefficients of the measurement (Fig. 3A).
measured at temperature X were all corrected to The results for the three dynamic diffusion
25 °C using Eq. (3). coefficient measurements are shown in Table 1.
For the stopped flow measurements the ‘stop
T25°C pX °C time’ was set such that the ratio of the first to
D25°C = DX °C (3)
TX °C p25°C second run variance was B 0.1 to reduce the
which is based on the Stokes– Einstein relation effects of measurement errors. In the E-field
[11], where T is the absolute temperature and p is method low field strengths were used to minimize
the solution viscosity. The corrected values are any effects from Joule heating. At 370 V cm − 1,
reported in Table 1. For all of the static diffusion the highest field used, the power dissipation was
coefficient measurements the excitation light was only 0.29 W m − 1 which is well below the 28 W
m − 1 power dissipation threshold, where Joule
heating was detected under similar conditions
[12]. A representative plot of the peak variance vs
migration time is shown in Fig. 3B. Good linear-
ity was obtained from the regression analyses for
all of the trials with correlation coefficients rang-
ing from 0.999 to 0.99995. For the length method,
a field strength of only 183 V cm − 1 was used to
minimize any effects from Joule heating. A typical
plot of the peak variance vs migration time is
shown in Fig. 3C. Good linearity was observed
from the regression analyses for all of the trials
with correlation coefficients ranging from 0.9995
to 0.99995.
The values obtained from the three dynamic
diffusion coefficient measurements vary by 7.5%
with the E-field method giving the smallest mea-
sured values and the length method giving the
largest values. The reason for the discrepancy in
the two results is unknown. The most likely rea-
son would be a systematic error in adjusting the
detection length; however, the micrometers used
to move the chip were checked and found to be
properly calibrated. Another possibility is that the
channel width, height, or surface roughness varied
Fig. 3. Peak variance vs diffusion time for (A) static; (B) along the length of the channel. The result, how-
E-field; and (C) length methods. R6G was the sample. The
lines represent the best least-squares fit to the data which are
ever, was not device specific as it was repeated on
represented by their respective error bars. The error bars are several different microchips. The E-field method,
drawn at 9 1.35% which assumes a temperature measurement stopped flow, and static diffusion coefficient mea-
uncertainty of 9 0.5 °C. The errors reported in the diffusion surements were within experimental error of one
coefficient values on the plots are due to the error in the another. More importantly, however, all of the
least-squares fitting. The diffusion coefficient values reported
on the plots have not been corrected to 25 °C as have the
results were within the experimental error of two
values in the tables. In (A) the peak areas are represented by other results reported in the literature using spec-
the triangles. troscopic methods [13,14]. The variation in ap-
C.T. Culbertson et al. / Talanta 56 (2002) 365–373 371

Table 1
R6G diffusion coefficients measured in both a 50/50 (v/v) methanol/aqueous solution and an aqueous solution

Method 50/50 Methanol/aqueous Aqueous

Diffusion coefficient (×10−6 cm2 s−1) at % RSDa n b Diffusion coefficient (×10−6 cm2 s−1) at nb
25 °C 25 °C

Static 2.69 90.02 0.74 6 4.14 90.1 6


Dynamic
Stopped flow 2.71 90.09 3.3 11
E-field 2.684 90.005 0.19 6 4.59 90.06 6
Length 2.88 90.17 5.9 8
Fister [13] 2.7 90.1 3.7 NA
Hansen [14] 2.5 90.3 12 NA

NA, not available.


a
Percent relative standard deviation.
b
Number of trials.

plied field and the static measurements were the static method to ensure that only longitudinal
most reproducible; both gave relative standard diffusion was being measured.
deviations (RSDs) of B1%. These results show Additional static diffusion coefficient measure-
that, under ideal circumstances, electrophoretic ments were made of seven other fluorescent dyes
separations on microfluidic devices are diffusion or molecules conjugated to fluorescent dyes—
limited. fluorescein, rhodamine B, 2%,7%-dichlorofluores-
While the static and best dynamic diffusion cein, TRITC-Glu, TRITC-Ile, fluorescein
coefficient measurements for R6G in the conjugated ovalbumin and fluorescein conjugated
methanol/aqueous buffer were within experimen- bovine serum albumin (Table 2). For the rho-
tal error of each other, under many circumstances damine and fluorescein families of dyes and the
the longitudinal diffusion coefficient of an analyte proteins, the diffusion coefficient decreased with
cannot be obtained from dynamic measurements increasing molecular weight as expected. This,
as the analyte may interact with the wall. Static however, was not the case for the TRITC labeled
diffusion coefficients, however, can be obtained as amino acids as the diffusion coefficient for
long as the equilibration time for adsorption– des- TRITC-Glu (MW 589) was larger than for
orption processes is small compared to the diffu- TRITC-Ile (MW 573). The reason for this appar-
sion coefficient measurement time. To examine ent discrepancy is that the van der Waals volume
the effect of analyte– wall interaction on the mea- for Ile (124 A, 3) is larger than for Glu (109 A, 3)
surement of the diffusion coefficient, we measured [15]. Diffusion coefficient values for the two
the dynamic diffusion coefficient for R6G in a 20 proteins have been reported previously in the
mM boric acid, 100 mM Tris solution using the literature and are also shown on the table. The
E-field method and compared the results to those static diffusion coefficient measurements for oval-
obtained from static measurements (Table 1). The bumin (at an ionic strength of 0.015 M) are 13%
diffusion coefficient from the dynamic measure- less than the average reported literature value.
ment is 11% larger than the static measurement The diffusion coefficient measurements, however,
indicating that there is a significant amount of reported in the literature were performed in solu-
analyte –wall interaction as the R6G migrates tions of different ionic strength which tends to
through the channel. For the rest of our diffusion change the diffusion coefficient, in some cases
coefficient measurements, therefore, we used the more than 20%, by changing the magnitude of the
372 C.T. Culbertson et al. / Talanta 56 (2002) 365–373

Table 2
Diffusion coefficients measured by the static imaging method in an aqueous solution

Analyte MWa Diffusion coefficient (×10−6 cm2 s−1) at 25 °C % RSD n Literature values

Rhodamine 6G 471 4.14 9 0.01 0.24 6


Rhodamine B 444 4.279 0.04 0.94 13
Fluorescein 332 4.25 9 0.01 0.24 6
2%,7% Dichlorofluorescein 401 4.10 9 0.07 1.7 12
TRITC-Glu 589 3.609 0.03 0.83 6
TRITC-Ile 573 3.43 9 0.04 1.2 5
Ovalbumin 45 000 0.6759 0.006b 0.9 6 0.789 [16]
0.776 [17]
0.8139 0.009c 1.1 16 0.759 90.014 [18]
BSA 65 000 0.638 9 0.015 2.4 5 0.63 90.02 [19]
0.62 90.01 [20]

a
Molecular weight.
b
Ionic strength 0.015 M.
c
Ionic strength 0.040 M.

Coulombic repulsions among the protein changes. This is problematic, however, in terms of
molecules. For our static measurements we in- accurately measuring the temperature inside of the
creased the ionic strength of the 20 mM boric channels. Significantly improved accuracy and pre-
acid/100 mM Tris buffer by adding 25 mM NaCl cision should be realized, however, with better
to the solution. This increased the diffusion coeffi- thermostatting of the chip.
cient by 20% from 6.75×10 − 7 cm2 s − 1 at an ionic The static coefficient measurements reported
strength (I) of 0.015 M– 8.13 ×10 − 7 cm2 s − 1 at above were all performed on microscopy setups in
an ionic strength of  0.040 M. These values our lab which we generally use to observe fluid
bracket those reported in the literature. The mea- movement on all of the microfabricated devices
surements for BSA are within experimental error of that we develop. As such, these measurements can
those reported in the literature. Molecular weight be performed with the typical instrumentation
(  2%) and, therefore, molecular volume changes found in a lab equipped for microfluidics research.
due to dye conjugation to the proteins— Ova (2.8 In addition, the measurements themselves only take
mol dye mol − 1 protein) and BSA (5 mol dye mol − 1 2–8 min to perform depending upon the size of the
protein)—does not change the diffusion coefficient molecule, and the post-data acquisition processing
measured within the error of the experiment. is less than 2 min. This technique, therefore, can be
At present the precision and accuracy of the performed rapidly and accurately with minimal
static measurements is limited by our ability to equipment investment.
accurately measure and control the temperature of
the chip. The temperature at the surface of the chip
was observed to vary by as much as 0.5 °C over Acknowledgements
the time course of an experiment. Such fluctuations
indicate that diffusion coefficient measurement er- This research is sponsored by The National
rors in terms of both accuracy and precision of up Cancer Institute, The National Institutes of Health,
to 1.35% might be expected. This is considerably under grant number CA83238. Oak Ridge National
larger, however, than most of the RSDs (measure- Laboratory is managed and operated by UT-Bat-
ment precision) reported in Tables 1 and 2 above telle, LLC under contract DE-AC05-00OR22725
indicating that the temperature inside the channels with the U.S. Department of Energy. The authors
remains relatively constant over the time course of wish to thank Christopher D. Thomas for fabrica-
the experiment even as the ambient air temperature tion of the microchips.
C.T. Culbertson et al. / Talanta 56 (2002) 365–373 373

[10] S.C. Jacobson, J.M. Ramsey, Analytical Chemistry 68


References (1996) 720.
[11] P.W. Atkins, Physical Chemistry, 3rd ed., W.H. Freeman
[1] Y. Walbroehl, J.W. Jorgenson, Journal of Microcolumn and Company, New York, 1986.
Separations 1 (1989) 41. [12] S.C. Jacobson, C.T. Culbertson, J.E. Daler, J.M. Ramsey,
[2] Y.J. Yao, S.F.Y. Li, Journal of Chromatograhic Science 32 Analytical Chemistry 70 (1998) 3476.
(1994) 117. [13] J.C. Fister III, S.C. Jacobson, L.M. Davis, J.M. Ramsey,
[3] C.T. Culbertson, Ph.D. Dissertation thesis, University of Analytical Chemistry 70 (1998) 431.
North Carolina, Chapel Hill, 1996. [14] R.L. Hansen, X.R. Zhu, J.M. Harris, Analytical Chemistry
[4] M.U. Kopp, H.J. Crabtree, A. Manz, Current Opinion in
70 (1998) 1281.
Chemical Biology 1 (1997) 410.
[15] T.E. Creighton, Proteins: Structures and Molecular Prop-
[5] V. Dolnik, S. Liu, S. Jovanovich, Electrophoresis 21 (2000)
erties, 2nd ed., W.H. Freeman, New York, 1993.
41.
[16] S.J. Gibbs, A.S. Chu, E.S. Lightfoot, T. Root, Journal of
[6] C.S. Effenhauser, G.J.M. Bruin, A. Paulus, Electrophoresis
Physical Chemistry 95 (1991) 467.
18 (1997) 2203.
[7] S.C. Jacobson, J.M. Ramsey, in: M.G. Khaledi (Ed.), [17] H.R. Mahler, E.H. Cordes, Biological Chemistry, 2nd ed.,
High-Performance Capillary Electrophoresis, vol. 146, Harper and Rowe, New York, 1971.
John Wiley and Sons, Inc., New York, 1998, p. 613. [18] M.S. Bello, R. Rezzonico, P.G. Righetti, Science 266 (1994)
[8] R.P. Haugland, Handbook of Fluorescent Probes and 773.
Research Chemicals, 6th ed., Molecular Probes, Eugene, [19] M. Placidi, S. Cannistraro, Chemical Physics Letters 310
OR, 1996. (1999) 130.
[9] S.C. Jacobson, R. Hergenröder, L.B. Koutny, R.J. War- [20] N. Meechai, A.M. Jamieson, J. Blackwell, Journal of
mack, J.M. Ramsey, Analytical Chemistry 66 (1994) 1107. Colloid and Interface Science 218 (1999) 167.

You might also like