Calibrations 3. The multimeter should read between 2.46 and 2.54 V DC. If not, adjust R224 on the CPU and R-Wave board until the reading is correct. Refer to Fig, 3, pg. 13 for the location of R224. Adjusting the gain 1. Connect the patient cable to a calibrated 1-mV source (safety analyzer or defibrillator energy analyzer). 2. Select [SENSORS]. Then, press (") to display the leads menu. Select 4, Press ("%") to start a real time printout and inject a I-mV pulse. ) to quit printing. Check the printout to see if the pulse is .5 mm). Ifitis not, adjust R127 on the CPU and R-Wave board until the pulse is correct. Refer to Fig. 3, pg. 13 for the location of R127. Adjusting the energy output 1. Ensure a pair of adult external paddles are connected and properly attached to a calibrated joule meter / defibrillator energy analyzer. 2, Set the energy select switch to the 20 Joules position. 3. Press @ (charge) on the apex paddle. An intermittent tone sounds as the unit charges. When the tone becomes continuous, the defibrillator is fully charged. This should occur within 10 seconds. 4. Within 30 seconds (before the energy is automatically dumped), discharge the defibrillator into the joule meter by pressing e (discharge) on both paddles simultaneously. 5. Ensure that the joule meter reading is 20 joules (48%). Ifnot, adjust R72 on the Power Supply board until the meter reading is correct. Refer to Fig. 4, pg. 14 for the location of R72. 6. Next, check the energy delivered display to make sure it reads 20 joules. Ifnot, adjust R143 on the Power Supply board until the display reading is correct. Refer to Fig. 4, pg. 14 for the location of R143, ‘Madic 5 Deftrillator Service Maral 11 Section 2 | Service 7. Repeat this procedure for each energy level. Verify that the joule meter readings are in the following ranges: 5-J level = 3.4-6.6 J; 10-J level =8.4- 11.6 J; 154] level = 13.4166 J, all other levels = +8%. Figure 2 . Control board | adjustments | | . . 6 | UB | \ maa ares sonar | | | : - a CONTROL BOARD, 2 Medic 5 Defrilatr Seroce Manual Figure 3 CPU and R-Wave board adjustments 1 n Calibrations var uz uaa LL vat ert us7 30 vo GAIN ADJUSTMENT ~ use uss ust ist ZERO OFFSET ADJUSTMENT — Medic 5 Defiriltr Serice Manual 13 Section 2 | Service Figure 4 Power Supply board adjustments C ‘ 2 “ B ENERGY DELIVERED ADJUSTMENT), fir143 ENERGY ACCURACY ADJUSTMENT \ u 2 R72 cs ror muri 2 R28 . \— 24-VOLT SUPPLY ADJUSTMENT ci @ D poweER SUPPLY BOARD 14 Medic 5 Deibriato Serice Maral

You might also like