Calibrations
3. The multimeter should read between 2.46 and 2.54 V DC. If not, adjust
R224 on the CPU and R-Wave board until the reading is correct. Refer
to Fig, 3, pg. 13 for the location of R224.
Adjusting the gain
1. Connect the patient cable to a calibrated 1-mV source (safety analyzer
or defibrillator energy analyzer).
2. Select [SENSORS]. Then, press (") to display the leads menu. Select
4, Press ("%") to start a real time printout and inject a I-mV pulse.
) to quit printing. Check the printout to see if the pulse is
.5 mm). Ifitis not, adjust R127 on the CPU and R-Wave board
until the pulse is correct. Refer to Fig. 3, pg. 13 for the location of R127.
Adjusting the energy output
1. Ensure a pair of adult external paddles are connected and properly
attached to a calibrated joule meter / defibrillator energy analyzer.
2, Set the energy select switch to the 20 Joules position.
3. Press @ (charge) on the apex paddle. An intermittent tone sounds as
the unit charges. When the tone becomes continuous, the defibrillator
is fully charged. This should occur within 10 seconds.
4. Within 30 seconds (before the energy is automatically dumped),
discharge the defibrillator into the joule meter by pressing e
(discharge) on both paddles simultaneously.
5. Ensure that the joule meter reading is 20 joules (48%). Ifnot, adjust R72
on the Power Supply board until the meter reading is correct. Refer to
Fig. 4, pg. 14 for the location of R72.
6. Next, check the energy delivered display to make sure it reads 20 joules.
Ifnot, adjust R143 on the Power Supply board until the display reading
is correct. Refer to Fig. 4, pg. 14 for the location of R143,
‘Madic 5 Deftrillator Service Maral 11Section 2 | Service
7. Repeat this procedure for each energy level. Verify that the joule meter
readings are in the following ranges: 5-J level = 3.4-6.6 J; 10-J level =8.4-
11.6 J; 154] level = 13.4166 J, all other levels = +8%.
Figure 2 .
Control board |
adjustments |
|
.
. 6 |
UB |
\ maa ares sonar |
| |
:
-
a
CONTROL BOARD,
2 Medic 5 Defrilatr Seroce ManualFigure 3
CPU and R-Wave
board adjustments 1
n
Calibrations
var
uz uaa
LL
vat
ert
us7 30 vo
GAIN ADJUSTMENT ~
use uss ust
ist
ZERO OFFSET ADJUSTMENT —
Medic 5 Defiriltr Serice Manual 13Section 2 | Service
Figure 4
Power Supply
board adjustments
C
‘
2
“
B ENERGY DELIVERED ADJUSTMENT),
fir143
ENERGY ACCURACY ADJUSTMENT
\
u 2 R72
cs
ror muri
2 R28
.
\— 24-VOLT SUPPLY ADJUSTMENT
ci
@
D poweER SUPPLY BOARD
14 Medic 5 Deibriato Serice Maral