Installation

and
Operations Manual for the Engle 310™ Operatory System
(229-1046) Engle 320™ Operatory System
P.O. Box 725 · 4115 24th Avenue Suite A · Forest Grove, Oregon 97116, USA
www.engledental.com

INSTALLATION AND OPERATION MANUAL

Customer Service: Technical Support:


7:00 A.M. to 4:30 P.M. (Pacific Time) 7:00 A.M. to 4:00 P.M. (Pacific Time)
Phone: 800-547-1906 Phone: 800-547-1906
Fax: 503-357-7286 Fax: 503-357-7286
E-mail: info@engledental.com E-mail: support@engledental.com

Warranty
Products sold to and installed by authorized Engle Dental Systems dealers are
guaranteed against defects in materials and workmanship for a period of five (5)
years after date of purchase. Engle 300 dental chair Electronic Positioning
Components have a Limited Warranty of ten (10) years after date of purchase.
Upholstery is guaranteed for one (1) year after date of purchase. All hand-piece
illumination systems are guaranteed for six (6) months after date of purchase. This
warranty does not cover light bulbs of any kind.
Faulty parts as determined by dealer service representative, will be repaired or
replaced; no claim for labor or consequential damages will be allowed. Invoices
for replacement parts may be subject to adjustment after the manufacturer has
inspected the alleged defective parts.

The warranty does not cover normal wear, stains, cuts, or scratches of upholstery
surface finishes, or parts sold to O.E.M. customers.
This warranty does not cover damage resulting from the use of cleaning,
disinfecting, or sterilization chemicals and processes.
The warranty does not cover damage from improper installation or maintenance,
accident, misuse, or accidents incurred in shipping and handling.

In the event of shipping damage, the purchaser must initiate all claims against
the freight carrier in writing within twenty-four (24) hours after the damaged items
are received. The claim is the responsibility of the customer.

PRODUCT MODIFICATIONS

Engle Dental Systems continually works to better our products. We reserve


the right to make modifications without notice or obligation to modify
previously manufactured products.

1
CLEANING AND DISINFECTING ENGLE EQUIPMENT
Infection control in the dental office continues to be a high priority for our customers and end users. O.S.H.A.,
the A.D.A., and the C.D.C. are also involved in this complex issue.

We will not attempt to specify the required intervals for disinfection nor can we state what is the overall best
surface disinfectant. Please refer to the Infection Control Recommendations published by the American
Dental Association for further reference.

Barrier Technique
ENGLE strongly advocates the barrier technique be used whenever possible to preserve the finish and
appearance of the equipment. Wherever possible disposable barriers should be used and changed between
patients. The barrier technique will ensure maximum long term durability of the surfaces and finishes of
equipment.
Chemical Disinfection
Regardless of the chemical disinfectant used it is imperative that the equipment be thoroughly washed with
mild soap and warm water at least once per day. This wash down will minimize the harmful effects of
chemical disinfectant residues being allowed to accumulate on the equipment.

When using a chemical disinfectant, always pay strict attention to the manufacturer’s application directions of
the disinfectant. When using a concentrated disinfectant, always pay strict attention to the manufacturer’s
diluting directions of the disinfectant. “Disinfectant solutions that are relatively harmless to surfaces at their
recommended strengths can be corrosive at a higher/purer concentrate.”

Unacceptable Disinfectants
These disinfectants WILL HARM the surface finishes of ENGLE equipment and are NOT RECOMMENDED.
USE OF THESE PRODUCTS WILL VOID YOUR WARRANTY.
CHEMICAL COMPOSITION TRADE NAMES
Strong Phenols/Phenol-Alcohol Combinations Lysol, Lysol 2, Lysol Professional
Coe Foam, Coe Spray Pump,
Vitaphone, Omni II
Sodium Hypochlorite/Household Bleach Clorox, Ajax, Purex

Alcohol
Household Cleaners

Conditionally Acceptable Disinfectants*


These disinfectants have been found to be the LEAST HARMFUL to the equipment surfaces.
None of these products are recommended for chair upholstery. Use the barrier technique.
CHEMICAL COMPOSITION TRADE NAMES
Iodophors** Biocide, Aseptic-IDC
Wescodyne, SD5, Promedyne, Ido-Five

Mild Phenols Procide ES, Asepti-Steryl Aerosol

Glutaraldehyde/Phenol sprays Sterall Spray, Coldspor

Synergized Super Quat Dis-cide, Cavicide, Kleenaseptic

Phenol/Water Sprays Top-Cide, Sporicidin Pump Spray, Birex se

* ENGLE DENTAL SYSTEMS makes no representation as to the disinfectant efficacy of these products. We make no warranty,
expressed or implied, that these disinfectants will not damage the surface finishes. Damage and discoloration of the surface finishes
is not covered under the warranty.

** Iodophor based disinfectants will cause yellow staining on many surfaces. Regular washing with soap and water will minimize this
staining. Iodophor neutralizers such as Promedyne are also available.

2
Engle 310/320 Utility Center Package - Installation Instructions
IL 1296 Page 1 of 4
Pre-Setup: Top
Post Casting
Plate
• Using the provided bolts, install the pre-assembled
Chair Bracket and Post Casting on the left or right side Figure 1.
of the Top Plate on the chair, located in front, below
the Seat assembly and above the Base Linkage
assembly. See figure 1.
• Remove your Utility Center Package from the Chair
shipping box. Bracket
with
Hidden
2 Umbilical

Delivery
Arm

Support
Arm

1 For a non-hidden Umbilical package insert the


2" diameter post into the Post Casting. For the
hidden Umbilical package, remove the
Umbilical Cover mounted by “four” screws, then
route the Umbilical thru the Post Casting and
insert the 2-1/4" diameter post into the Post 1
Casting. Re-install the Umbilical Cover with
the Umbilical inside routed thru the opening. 2

2 Adjust the “eight” adjustment Set Screws on the Post Casting


Post Casting until the Support Arm is vertically Umbilical
plumb. Determine this by using a level, being Cover
sure to check that the Support Arm is vertically
plumb in all directions. Tighten the adjustment 3 The Delivery Arm is under tension due to the
screws evenly and re-check the Support Arm to Packing Band. While securing the Delivery
ensure that it is still plumb. Note: It is extremely Arm with one hand, cut the packing band.
important that the Support Arm is vertically Warning: If the band is removed without
plumb, as this determines whether or not the securing the Delivery Arm, the Delivery Arm
Cuspidor, Delivery Arm, Tray and the Light will will rise up unexpectedly and could cause
be plumb, and therefore function properly. damage or injury.

3
Engle 310/320 Utility Center Package - Installation Instructions
IL 1296 Page 2 of 4

Delivery Arm
Contact Point
Assembly
Water Stream
Bowl
Flush
Tube
Cup Fill Tube

Contact Point

Engle 310 Cuspidor Cup Fill Tube/Bowl Flush Tube Installation


! Rotate the Delivery Arm assembly to the position shown.
! Insert the Cup fill tube fully into it’s Tube Mount, at the
location shown.
! Insert the Bowl Flush Tube fully into it’s Tube Mount, at the
location shown, “and make sure that the bowl and water
stream contact point is precisely at one of two locations
shown, Choose the location closest to patient”.

Engle 320 Linear Telescoping Arm Installation Delivery Arm


Assembly

Linear Telescoping
Arm Assembly

C.V. Can ! Rotate the Delivery Arm assembly to the


Assembly position shown.
! Insert the Linear Telescoping Arm assembly
fully into the mount of the Utility Center next
to the C.V. Can assembly as shown.
! Place the hand-pieces into their positions as
shown.

4
Engle 310/320 Utility Center Package - Installation Instructions
IL 1296 Page 3 of 4

Engle 300 Auto Control Assembly and Tray Mount Assembly Installation

4 1

Engle 300
Auto Control

4 1

3 Using the “male” and “female” Quick


Unit Mount
Assembly Disconnects attached to the tube ends of
the Umbilical and Engle 300 Auto Control,
match the tubes regarding “size” and
“color” and SECURELY connect the “male”
1 Remove the Top Plate of the Engle 300 Auto and “female” Quick Disconnects by twisting
Control assembly by loosening the “two” them together.
Access Nuts attached to the underside of the
assembly. EXAMPLE: ¼" diameter gray tubing from the
Umbilical would connect with ¼" diameter
2 Route the Umbilical from the Unit Mount gray tubing from the Engle 300 Auto
assembly / Delivery Arm Assembly through the Control.
2” X 2” square hole, located at the underside of
the Engle 300 Auto Control. Install the Engle 4 Re-install the Top Plate of the Engle 300
300 Auto Control to the Unit Mount assembly by Auto Control.
locking and tightening the “four” mounting
screws into the “four” keyholes surrounding the 5 Insert the Tray Holder assembly fully into the
2” X 2” square hole. Unit Mount assembly as shown. The Tray
rests on top of the Tray Holder assembly.
5
Engle 310/320 Utility Center Package - Installation Instructions
IL 1296 Page 4 of 4

Light Delivery 2 3
Assembly Installation Light
Sub-assembly

24“ Light Post

Figure 2.
Light Post
Assembly
Figure 1.
Engle 300 Trim Cover
Auto Control
Delivery
Assembly
Water Auxiliary Quick
Chassis Cover Connect/Disconnect

Water Toggle
Switch

Chassis
2 Housing Shroud

1 Rotate the Engle 300 Auto Control Delivery


assembly so that it will not interfere with the
5 Place the Trim Cover around the 24“ Light
Light Delivery assembly installation.
Post and install the Light Delivery assembly
2 Partially move the Chassis Housing Shroud down fully into the mount of the Utility Center next
by removing the “four” Acorn Nuts located at it’s to the Engle 300 Control Delivery assembly,
underside. Caution: The tubes for the Water as shown on [Figure 1.]. Adjust the ”four” Light
Toggle Valve and Water Auxiliary Quick Post adjustment Set Screw see [Figure 2.],
Connect/Disconnect are attached to there making sure that the 24” Light Post is
source. Carefully slide the chassis housing vertically plumb.
downward until the “four” Light Post adjustment CAUTION: DO NOT OVER TIGHTEN SET
Set Screws on the Utility Center Post Mount are SCREWS.
accessible, see Figure 2.
6 Connect the routed power cord of the Light
3 Route the power cord from the Light Sub- Sub-assembly to the power cord from the
assembly, through the Light Post assembly using Umbilical inside the Utility Center and re-
the pre-assembled routing string and insert the install the Chassis Housing Shroud.
Light sub-assembly fully into the to the Light Post
assembly. See Figure 1. 7 Finally, for a Non-Hidden Umbilical, connect
the Umbilical to the junction box, see (IL 160
4 Route the power cord from the Light Post in this manual). For a Hidden Umbilical, see
assembly to the Utility Center. (IL 1600 in this manual).
6
Umbilical/Junction Box - Installation Instructions
IL 160 Page 1 of 1

Engle Dental Systems has simplified the connection of utilities within the Junction Box. Refer to (IL 250 “Large
Junction Box Installation Diagram”) of this manual throughout this installation process.
NOTE: Actual size (1:1 SCALE) of IL 250 “Large Junction Box Installation Diagram” to be used as a template are
available upon request when a layout is required.

! A 1/2" threaded water supply pipe and a 1/2" threaded air supply
Supply
pipe must be 1” tall maximum, above the finished floor. Using the Stop
proper thread sealant, install the “two” included Supply Stop Valves Valve
to the 1/2” threaded pipes, one each pipe. Once set, use these
valves to purge the water and air supply pipes to remove any 1/2“
Threaded
supply line debris. (See Figure 1.) Supply Pipe
! The supplied Air Regulator Filter assembly and Water Regulator Filter
assembly are PRE-ADJUSTED. Insert the nipple of each assembly
into the compression fitting of it’s designated Manual Shut-off Valve Figure 1.
(“A” for air) (”W” for water), and tighten. (See Figure 2.)
! Remove the Cover from the frame of the Junction Box assembly,
and place the frame on the floor as shown on the IL 250 “Large Air Regulator
Junction Box Installation Diagram”. Make sure the Umbilical mount Filter Assembly
of the Frame that fits your Umbilical is facing the correct direction.
Secure the Frame to the floor with the “eight” provided screws. (See
Figure 3.)
! Determine the length required for the Umbilical to be routed from
the Utility Center to the Junction Box, then cut off the extra length
of Umbilical and insert that Umbilical end into the correct Umbilical
mount of the Frame.
! Connect the 1/8" clear tubing from the Water Regulator Filter Figure 2.
assembly to the 1/8" clear tubing from the Air Regulator Filter
assembly. To do this SECURELY connect the “male” and “female”
Quick Disconnects by twisting them together.
Cover
! Connect the 3/4" diameter gray ribbed tubing from the Umbilical
to the office Central Vacuum within the Junction Box, and the 3/4"
diameter clear tubing from the Umbilical to the office Gravity Drain
within the Junction Box. See IL 250 “Large Junction Box Installation
Diagram”.
! Using the “male” and “female” Quick Disconnects attached to the
tube ends of the Umbilical and both Air and Water Regulator
assemblies, match the tubes regarding “size” and “color” and
SECURELY connect the “male” and “female” Quick Disconnects by
twisting them together.

EXAMPLE: 1/8“ diameter yellow tubing from the Umbilical would connect Frame Umbilical
With 1/8” diameter yellow tubing from the Water Regulator assembly. Mount
! Plug the Power Cord from the Umbilical into the Electrical Box. Figure 3.
! Re-install the Cover to the Frame of the Junction Box.
7
Large Junction Box Installation Diagram (34-005)
IL 250

8
Engle 300 Hidden Umbilical Path - Partial Section View
IL 1600 Page 1 of 1

Before routing the Umbilical, position the chair so that the base is all the way up as shown.
Turn the power switch off and un-plug the Power Cord from the power source.

FRONT CHAIR MOUNT


UMBILICAL PATH

TOP MOUNT
REAR CHAIR MOUNT
SHROUD
UMBILICAL PATH
(If Equipped)

UPPER ELEVATOR
SHROUD

POWER
SHROUD
LARGE RIBBED
UMBILICAL TUBE

HYDRAULIC
ASSEMBLY
BRACKET

LOWER ELEVATOR
SHROUD
TO ENGLE 300
JUNCTION BOX

1 Now that the chair’s power cord is un-pluged from it’s power source. Remove the Power
Shroud by un-installing the “two” mounting screws, one each side. Remove the Upper Elevator
Shroud by un-installing the “two” mounting screws, in front. Un-install the Lower Elevator Shroud
by slightly pulling one mounting bracket outwards and sliding it off the “two” mounting pins
then slide the opposite mounting bracket of the remaining “two” mounting pins.
2 Route the Umbilical as shown. Caution: For the Engle 360 System make sure the Auto Control
Delivery, Engle Light Delivery, and Rear Vacuum Group Delivery can “fully“ articulate without
any articulation stress to it’s Umbilical. From the front chair mount, make sure the Umbilical
goes in thru the front 1-1/2“ wide slot, and out the underside large rectangular opening, of the
Top Mount Shroud. Then, in thru the large Ribbed Umbilical Tube. Finally thru the center tunnel
of the Hydraulic Assembly Bracket. From the rear chair mount (IF EQUIPPED), make sure the
Umbilical goes thru the 2-1/8” thru hole, and thru the large Ribbed Umbilical Tube. Finally thru
the center tunnel of the Hydraulic Assembly Bracket.
3 Re-install the ”three” shrouds previously removed, using step one in reverse order. NOTE: The
Hidden Umbilical length is pre-measured, once installation to the Engle 300 Junction Box is
complete, adjust the Umbilical slack where needed to prevent it from articulation stress. See
(IL 1160 ”Hidden Umbilical/Junction Box-Installation Instructions”) in this manual and connect
the Umbilical to the Engle 300 Junction Box assembly.
9
Hidden Umbilical/Junction Box - Installation Instructions
IL 1160 Page 1 of 1
Engle Dental Systems has simplified the connection of utilities within the Junction Box, it is important that the pre-
instalation of utilities for the Hidden Umbilical Engle 300 chair package is at the service location of the chair as
shown in (IL 1250 “Engle 300 Junction Box Installation Diagram”) of this manual, refer to IL 1250 throughout this
installation page.

NOTE: Actual size (1:1 SCALE) of IL 1250 “Engle 300 Junction Box Installation Diagram” to be used as a template
are available upon request when a layout is required.
Supply
! A 1/2" threaded water supply pipe and a 1/2" threaded air supply Stop
pipe must be 1” tall maximum, above the finished floor. Using the Valve
proper thread sealant, install the “two” included Supply Stop Valves 1/2“
to the 1/2” threaded pipes, one each pipe. Once set, use these Threaded
valves to purge the water and air supply pipes to remove any Supply Pipe
supply line debris. (See Figure 1.)
! The supplied Air Regulator Filter assembly and Water Regulator Filter Figure 1.
assembly are PRE-ADJUSTED. Insert the nipple of each assembly
into the compression fitting of it’s designated Manual Shut-off Valve
(“A” for air) (”W” for water), and tighten. (See Figure 2.)
! Remove the Cover from the frame of the Engle 300 Junction Box Air Regulator
assembly, and place the frame on the floor, in front, and centered Filter Assembly
to the Engle 300 chair Base Plate as shown on the IL 1250 “Engle
300 Junction Box Installation Diagram”. Secure the Frame to the
finished floor with the “four” provided screws. (See Figure 3.)
! Insert and secure the Umbilical from the Engle 300 chair into the
Umbilical mount of the Frame.
! Connect the 1/8" clear tubing from the Water Regulator Filter
assembly to the 1/8" clear tubing from the Air Regulator Filter
assembly. To do this SECURELY connect the “male” and “female” Figure 2.
Quick Disconnects by twisting them together.
! Connect the 3/4" diameter gray ribbed tubing from the Umbilical
to the office Central Vacuum within the Junction Box, and the 3/4"
diameter clear tubing from the Umbilical to the office Gravity Drain Cover
within the Junction Box. IL 1250 “Engle 300 Junction Box Installation
Diagram”.
! Using the “male” and “female” Quick Disconnects attached to the
tube ends of the Umbilical and both Air and Water Regulator
assemblies, match the tubes regarding “size” and “color” and
SECURELY connect the “male” and “female” Quick Disconnects by
twisting them together. Umbilical
Mount
EXAMPLE: 1/8“ diameter yellow tubing from the Umbilical would connect Frame
With 1/8” diameter yellow tubing from the Water Regulator assembly.
Figure 3.
! Plug the Power Cord from the Umbilical into the Electrical Box.
! Re-install the Cover to the Frame of the Engle 300 Junction Box.
10
Hidden Umbilical Junction Box Installation Diagram (34-015)
IL 1250 Page 1 of 1

11
Engle 300 Auto Control (Brake) - Description of Functions
IL 1149 Page 1 of 1

3X Automatic
Hand Piece Holder
(P/N 29-515)
2X Hand Piece Holder Automatically stops and applies
(P/N 29-516) the flow of air and water
to it’s designated hand piece.

HP3 HP2 HP1

FRONT VIEW Engle 300 A.C. Touch Pad Assembly


(P/N 96-530)
Positions dental chair
automatically and manually.
ON/OFF
Master Switch
(P/N 015-27) Water Coolant Hand Piece
Master Air and Water Flow Valve Flush Valve
Solenoids Off and On. (P/N 015-16)
When not in use, turn Controls the amount
this switch to the of Water Coolant to Flushes hand piece
Off position. each individual hand piece. in use with water.

REAR VIEW

Drive Air Air Coolant


Arm Brake Pressure Flow Valve Drive Air
Switch Adjust Controls the amount Pressure Gauge
(P/N 015-27) Controls the of Air Coolant to (P/N 015-85)
amount of air each individual Air pressure
Activates the hand piece.
pressure to reading for the
locking and
drive each hand piece in use.
un-locking of
individual
the Flex Arm’s
hand piece.
vertical
movement.

12
AUTO BLOCK
29-342/29-343 (SHOWN)/ R ON / OFF VALVE
SYRINGE LEA
29-344/OR 29-345 " -C (015-27)
(20-295) 1/8
CK
LA EXHAUST
" -B
1/8 E VALVE (29-250)
HIT
" -W
TUBING DIRECTLY ATTACHED QUICK 1/8
FROM UMBILICAL DISCONNECTS.

R
1/8 EA
E "- - CL
BLU 1/8 WH
1 /8"
"- R "- ITE
1/8 EA GR
- CL 1/8
"- EY
1/8" EY BLU
GR E
"- E CK
1/8 HIT LA
- W " -B
IL 1150

1/8" CK 1/8
BLA
BARB FITTING W "-
LO EY 1/8
(009-16) YEL GR CK
"- "- LA
1/8 1/4 " -B
1/8
CK
SEE DETAIL -A- BLA
"-
ITE 1/8
IR WH 1/8
-A "- "-
ING 1 /8 CL
UB EA
R ARM BRAKE
ET 1/8
"-
ING 1/8
VALVE (015-27)
YR E E 1/8 BLU
E
"-
IT S BLU BLU "- CL
EA
SPL "- "-
1/4
"-
BLU
E R
1/8 1/8 GR

13
EY
1/8
"-
WH
ITE

R
LEA
" -C R
R 1/8 LEA
LEA " -C
" -C R 1/8

SPLIT SYRINGE TUBING - WATER


1/8 LEA
" -C
1/8
H.P. FLUSH R
LEA
VALVE " -C
(015-16) 1/8 H.P. ACTUATOR
(29-515)

E
BLU
"-
1/8

AUTOBLOCK PORTS
PRESSURE (PER HANDPIECE)
GAUGE NOTE: WATER
(015-85) PILOT VALVE
FOR AUTOBLOCKS WITH SIGNAL COOLANT AIR
ADDITIONAL AUTOMATIC DRIVE AIR
POSITIONS, ROUTE 1/8" CLEAR PILOT VALVE
TUBES TO ADDITIONAL H.P. SUPPLY HANDPIECE TUBING
ACTUATOR THESAME AS SHOWN.
Engle 300 Auto Control (Brake) - Tubing Diagram
Page 1 of 1
Engle 310 Cuspidor/Utility Center - Description of Functions
IL 1248 Page 1 of 1

Cup Fill
Engle 310
Actuator Bowl Flush
Cuspidor
Actuator

Water Auxiliary Quick


Connect/Disconnect outlet

City Water

Water source selector for Hand tools,


Water Auxiliary, & Cup Fill - (Engle 310 Cuspidor)

Bottle Water

Utility Center

Shown with Hidden Umbilical option.

14
Engle 310 Cuspidor - Air/Water Schematic
IL 1249 Page 1 of 1

22

21

3
15 14
15
6
23 1/4 BLUE
20
18 19
4

IL249
1/8 YELLOW 1/8 YELLOW
12 16
1/4 GRAY

1 1/8 BLUE BL 1/4 BLUE

CL 1/8 CLEAR
1/8 YELLOW
11 17
1/8 CLEAR

1/8 WHITE
2 10
1/8 YELLOW
1/8 BLUE

1/8 YELLOW
14
1/8 YELLOW 9 8 7
1/8 BLUE

1/8 YELLOW

13
5
BL

1/8 BLUE

(TO ON/OFF VALVE 015-27)

(TO FOOT CONTROL)


5/16 GRAY
5/8 GRAY
TO LIGHT
GN BLK WH WATER
AIR
(REGULATED AIR TO SYRINGE) 1/8 YELLOW

(REGULATED WATER TO AUTOBLOCK) 1/8 BLUE


(UNREGULATED AIR TO ON/OFF VALVE 015-27) 1/8 WHITE
(FROM JUNCTION BOX REGULATED AIR) 1/8 YELLOW
(FROM JUNCTION BOX) 18-3 ELECTRIC

(FROM JUNCTION BOX CENTRAL VACUUM) 3/4 GRAY RIBBED


(FROM JUNCTION BOX) LOW VOLTAGE CLEAR ELECTRIC

(FROM JUNCTION BOX GRAVITY DRAIN) 3/4 CLEAR


(FROM JUNCTION BOX UNREGULATED AIR) 1/8 BLACK

(FROM JUNCTION BOX REGULATED AIR) 1/4 GRAY


(FROM JUNCTION BOX UNREGULATED AIR) 1/8 WHITE

(FROM JUNCTION BOX REGULATED WATER) 1/8 BLUE


(FROM JUNCTION BOX UNREGULATED WATER) 1/4 BLUE

NO. PART NO. DESCRIPTION QTY.


11 29-180 BOWL PILOT VALVE (2 WAY) 1
12 29-181 CUP PILOT VALVE (3 WAY) 1
13 29-640 AIR/WATER MANIFOLD 1
NO. PART NO. DESCRIPTION QTY. 14 40-102 RESTRICTOR (LARGE) 2
1 015-18 TOGGLE VALVE 1 15 69-025 TUBE MOUNT 2
2 015-35 REGULATOR (SMALL) 1 16 69-030 BOWL HOLDER 1
3 015-68 QUICK DISCONNECT 1 17 69-129 VENT TEE 1
4 018-05 O-RING (BOWL) * 1 18 69-216 FLUSH VALVE ASSEMBLY 1
5 20-295 SYRINGE - OPTIONAL 1 19 69-216R VALVE REPAIR KIT * 1
6 21-609 PREASURE CAP ASSEMBLY 1 20 69-221 CUSPIDOR MONO BLOCK ASSY. 1
7 22-299 SALIVA EJECTOR 1 21 69-224 BOWL FLUSH TUBE 1
8 22-528 C.V. MOUNT BRACKET * 1 22 69-225 CUP FILL TUBE 1
9 22-535 C.V. CAN ASSEMBLY 1 23 21-617 1 LITER BOTTLE * 1
10 22-751 H.V.E. 1 * INDICATES PART NOT SHOWN

15
Engle 320 Linear Telescoping Arm - Air/Water Schematic
IL 1340 Page 1 of 1

5
11

1 1/8 BLUE BL

CL 1/8 CLEAR
1/8 CLEAR

1/8 WHITE
2 9 9
1/8 BLUE

1/8 YELLOW 8 7 6
1/8 BLUE

1/8 YELLOW

10
4
BL

1/8 BLUE

(TO ON/OFF VALVE 015-27)

(TO FOOT CONTROL)


IL340
5/16 GRAY
5/8 GRAY
TO LIGHT
5/8 GRAY
GN BLK WH WATER
AIR
(REGULATED AIR TO SYRINGE) 1/8 YELLOW

(REGULATED WATER TO AUTOBLOCK) 1/8 BLUE


(UNREGULATED AIR TO ON/OFF VALVE 015-27) 1/8 WHITE
(FROM JUNCTION BOX REGULATED AIR) 1/8 YELLOW
(FROM JUNCTION BOX) 18-3 ELECTRIC

(FROM JUNCTION BOX CENTRAL VACUUM) 3/4 GRAY RIBBED


(FROM JUNCTION BOX) LOW VOLTAGE CLEAR ELECTRIC

(FROM JUNCTION BOX UNREGULATED AIR) 1/8 BLACK

(FROM JUNCTION BOX REGULATED AIR) 1/4 GRAY


(FROM JUNCTION BOX UNREGULATED AIR) 1/8 WHITE

(FROM JUNCTION BOX REGULATED WATER) 1/8 BLUE

NO. PART NO. DESCRIPTION QTY.


1 015-18 TOGGLE VALVE 1
2 015-35 REGULATOR (SMALL) 1
3 015-68 QUICK DISCONNECT 1
4 20-295 SYRINGE 1
5 21-609 PREASURE CAP ASSEMBLY 1
6 22-299 SALIVA EJECTOR 1
7 22-528 C.V. MOUNT BRACKET * 1
8 22-535 C.V. CAN ASSEMBLY 1
9 22-751 H.V.E. 2
10 29-640 AIR/WATER MANIFOLD 1
11 21-617 1 LITER BOTTLE * 1
* INDICATES PART NOT SHOWN

16
Pneumatic Flex Arm - Tension and Pitch Adjustment
IL 1020 Page 1 of 1

Lock Lock Tension Tension Flex Arm


Nut #1 Nut #2 Nut Spring Cover
Leveling
Bolt

Spring
DECREASE Tension INCREASE

Front Arm Knuckle


End End
Cap Cap
Section View of Pneumatic Flex Arm Assembly
End Cap End Cap
Screw Screw

Spring Tension Adjustment:


1. Remove both End Caps, and Flex Arm Cover by un-installing the “two” End Cap Screws.

2. Use a 3/16“ Allen wrench, insert it into a Slot of the Tension Nut and rotate it to increase
or decrease tension on the Tension Spring. Looking from the Front Arm Knuckle side of the
Flex Arm assembly, rotating the Tension Nut clockwise will increase tension and counter-
clockwise will decrease tension.

3. With the complete service load of the Auto Control assembly, adjust the spring tension
so that the Flex Arm will maintain all of the Auto Control’s vertical positions when left unattended.

4. Re-install the Flex Arm Cover and both End Caps with the “two” End Cap Screws.

Front Arm Knuckle Pitch Adjustment:


1. Remove both End Caps, and Flex Arm Cover by un-installing the “two” End Cap Screws.

2. Loosen and back away Lock Nut #2.

3. Adjust the Leveling Bolt until the desired pitch of the Auto Control assembly is achieved.

4. Tighten Lock Nut #2 and check to make sure Lock Nut #1 is still tight, as shown.

5. Re-install the Flex Arm Cover and both End Caps with the “two” End Cap Screws.

17
Engle Light - Focusing Instructions
IL 284 Page 1 of 1

180mm
(~7.0“)

Light
Pattern
90mm
(~3.5“) 700mm
(~27.5“)

B
1. Loosen the set screw marked “A” ½ turn, using a
1.5mm allen wrench. Over loosening this set screw
will require taking the light head apart.

2. Adjust the screw marked “B”, using a 4mm allen


wrench, until the desired light pattern is acheived.

3. Tighten the set screw marked “A” until you feel


slight pressure on the set screw.

18
General Dimensions - Engle 310™Operatory System
IL 1200 Page 1 of 1

NOTE: When locating the


Engle 310™ System allow
for it’s ability to rotate 60°.

6.8" (173mm)

41.6“
(1057mm)
R16.0"
(407mm)

14.5"
(367mm)

17.5" (443mm)

24.0" (610mm)

35.5" (903mm)

19
General Dimensions - Engle 320™Operatory System
IL 1204 Page 1 of 1

NOTE: When locating the


Engle 320™ System allow
for it’s ability to rotate 60°.

48.7“
(1236mm)

41.6“
(1057mm)

R10.3"
(262mm)

14.5"
(367mm)

17.5" (443mm)
24.0" (610mm)
35.5" (903mm)
20
Maintenance & Service Notes:
Description & Technician/Company Date

21