Meaning : The polygon did not reach its necessary speed within a certain time.
Reason _: Problem with polygon or scan control board (SCB).
Cure : Check polygon and SCB.
20377
E_SCD_SWP_TST_0_RANGE
Meaning : While testing the scan control board (SCB), the read-back
voltage of the sweep signal deviated more than 10% from
the expected value.
Reason CB defective.
Cure. : Replace SCB.
20378
E_SCD_SWP_TST_A_RANGE
Meaning : While testing the scan control board (SCB), the read-back
voltage of the sweep signal deviated more than 10% from
the expected value.
Reason _: SCB defective,
Cure : Replace SCB.
20379
E_SCD_SWP_TST_R_RANGE
Meaning : While testing the scan control board (SCB), the read-back
voltage of the sweep signal deviated more than 10% from
the expected value.
Reason CB defective,
Cure : Replace SCB.
20370
E_SCD_SWP_TST_O_RANGE
Meaning : While testing the scan control board (SCB), the read-back
voltage of the sweep signal deviated more than 10% from
the expected value.
Reason CB defective.
Cure : Replace SCB.
20380
E_SCD_SWP_CSD_TLIN_LOW
Meaning : Too short line_time, detected at calculation of sweep divider.
Reason software error or invalid parameters in FIP tables.
Cure : Check if parameters in FIP tables have been changed.
Correct parameters or delete D:\FIPXXX00,U00.
20381
E_SCD_SWP_CSD_TLIN_HIGH
Meaning : Too long line_time, detected at calculation of sweep divider,
Reason software error or invalid parameters in FIP tables.
Cure : Check if parameters in FIP tables have been changed.
Correct parameters or delete D:\FIPXXX00.U00.
20382
E_SCD_SWP_CSD_DIV_LOW
Meaning : Calculated too low sweep divider. Minimum used.
Reason oftware error or invalid parameters in FIP tables.
Cure : Check if parameters in FIP tables have been changed.
Correct parameters or delete D:\FIPXXX00.U00.
20383
E_SCD_SWP_CSD_DIV_HIGHMeaning : Calculated too high sweep divider, Minimum used.
Reason _: Software error or invalid parameters in FIP tables.
Cure : Check if parameters in FIP tables have been changed.
Correct parameters or delete D:\FIPXXX00,U00,
20384
E_SCD_SWP_OBSERVE_TIMEOUT
Meaning : The polygon speed did not stay not good for the requested time,
Reason Problem with polygon or scan control board (SCB).
Cure : Check polygon and SCB.
20390
E_SCD_RS_VER_DAQ
Meaning : Wrong version of DAQ board (ADC70 hardware).
20391
E_SCD_RS_VER_SCB
Meaning : Wrong version of scan control board (SCB) detected.
Reason Negative version code from scd_scb_version(). Software error.
Cure : Check if software
20392
E_SCD_RS_PRESENT
Meaning : Scan control board (SCB) not found.
Reason __: - SCB not or badly inserted.
~ SCB defective.
Cure : Check / insert / replace SCB.
20393
E_SCD_RS_INI_HW
1g the scan control board (SCB) failed.
20394
E_SCD_RS_GET_IVEC
Meaning : Getting interrupt vector from operating system failed.
(Used for RS position or RS optoswitch / begin of line.)
Reason _: Problems with operating system.
Cure : Check if software is installed correctly. Check CPU.
20395
E_SCD_RS_PRESENT_RMS
Meaning : No rotating shutter driver board (RMS) found.
Reason _: - Bad connection or cabelling between scan control board (SCB) and
RMS.
- RMS board defective
Check cables and connectors.
- Check / replace RMS board.
cure
20396
E_SCD_RS_VERS_RMS
Meaning : Hardware version of RMS board not correct
Reason Wrong hardware
- Connection problem
Cure : Check RMS board, cables to RMS and power supply of RMS
20397
E_SCD_RS_FCREATEMeaning : Creating event flag group with sc_fcreate() failed.
(Used e.g. for ramping rotating shutter.)
Reason roblem with operating system.
Cure : Check if software is installed correctly,
20398
E_SCD_RS_RMSCOM
Meaning : Communication with rotating shutter driver board failed
Reason Most probably hardware problem.
Cure : Check RMS board, cables to RMS and power supply of RMS
20399
E_SCD_RS_NOT_INIT
Meaning : The driver software for rotating shutter control has not been
initialized,
Reason oftware error.
Cure : Software must call scd_rs_i
0) first.
203A0
E_SCD_RS_RAMP_ENAL
Meaning : Enabling the rotating shutter motor supply failed.
Reason __: - Wrong version of rotating shutter driver board (RMS).
+ Communication problem with RMS.
Cure : Check SCB and RMS. Replace if necessary.
203A1
E_SCD_RS_RAMP_ENA2
Meaning : Enabling the rotating shutter motor supply failed.
Reason __: - Wrong version of rotating shutter driver board (RMS).
+ Communication problem with RMS.
Cure : Check SCB and RMS. Replace if necessary.
203A2
E_SCD_RS_RAMP_FCLO
Meaning : Clearing the rotating shutter position_ok event flag
with sc_fclear() failed.
Reason
Cure
2033
E_SCD_RS_RAMP_FPD1
Meaning : sc_fpend() returned with error after pending for
rotating shutter position_ok event.
Reason imeout or others.
Cure
20304
E_SCD_RS_RAMP_FCL1
Meaning : Clearing the rotating shutter position_ok event flag
with sc_fclear() failed.
Reason
Cure :
203A5
E_SCD_RS_RAMP_FPD2
Meaning : sc_fpend() returned with error after pending for
rotating shutter position_ok event.
Reason imeout or others.
Cure203A6
E_SCD_RS_RAMP_FCL2
Meaning : Clearing the rotating shutter posi
with sc_fclear() failed.
Reason
Cure
n_ok event flag
203A7
E_SCD_RS_RAMP_FPD3
Meaning : sc_fpend() returned with error after pending for
rotating shutter position_ok event.
jimeout or others.
Reason
Cure
2038
E_SCD_RS_RAMP_FCL3
Meaning : Clearing the rotating shutter posi
with sc_fclear() failed.
Reason
Cure
n_ok event flag
203A9
E_SCD_RS_RAMP_FAILED
Meaning : Ramping up the rotating shutter failed at least once,
Retry was successfull,
Reason
Cure : Only warning. Ramping was finally successfull.
If error occurs regularily, try if replacement of rotating
shutter or replacement of RMS board solves the problem.
203AA
E_SCD_RS_RAMP_FAILED_ALL
Meaning : Ramping up the rotating shutter failed completely.
Retries were not successfull.
Reason no power for rotating shutter
- cabeling problem to rotating shutter
- rotating shutter driver board defective
- rotating shutter motor defective
Cure: Check rotating shutter power supply, cabeling, driver board,
motor. Replace or repair defective part.
203AB
E_SCD_RS_RAMP
Meaning : Motor is not ramping at end of run command.
Reason
Cure
‘Check RMS board and cabeling.
= Check rotating shutter optoswitch.
- Check if replacement of scan control board (SCB) solves
the error.
203ac
E_SCD_RS_WRAMP_TMO.
1g for end of rotating shutter ramp timed out.
Reason __: - Problem with motor controller "MIG" on scan control board (SCB).
= Too short timeout time (programming error).
Cure : Check if error occurs regularily. If so,
check if replacement of scan control board solves the error.203AD
E_SCD_RS_WRAMP_FAILED
\g shutter ramp, there were no optos!
signals detected,
Retry was successfull.
Motor lost steps and is not turning.
- Rotating shutter optoswitch does not work correctly.
Cure: Only warning. Ramping was finally successfull
If error occurs regularily, try if replacement of rotating
shutter or replacement of RMS board solves the problem.
Reason
203AE
E_SCD_RS_WRAMP_FATLED_ALL
Meaning : After end of rotating shutter ramp, there were no optos\
signals detected,
Retries were not successfull
Motor lost steps and is not turning.
1g shutter optoswitch does not work correctly.
Check if rotating shutter optoswitch is working
(LED at scan control board (SCB))
- Try if replacement of rotating shutter or replacement of
RMS board solves the problem.
203AF
E_SCD_RS_WRAMP_STOP
Meaning : Stopping the rotating shutter in a ramp retry cycle failed.
Reason us errors in log file.
Cure
20380
E_SCD_RS_WRAMP_WSTOP
Meaning : Waiting for end of rotating shutter stop ramp timed out.
Error can only occur in ramp retry cycles.
Problem with motor controller "MIG" on scan control board (SCB).
= Too short timeout time (programming error).
irst check for other rotating shutter errors and solve them.
Check if replacement of scan control board solves the error.
Reason
Cure
20381
E_SCD_RS_WRAMP_RUN
Meaning : Restarting the rotating shutter motor in retry cycle failed.
Reason _: See previous error in log
Cure: See previous error in log file.
20382
E_SCD_RS_TRUN_SPEED
Meaning : Rotating shutter is not turning with the correct speed.
Reason __: Ramping problems or test while motor stopped.
Cure: Rely on automatic retries.
203c0
E_SCD_HT_VER_SWH
Meaning : Wrong version of SWH board (ADC70 hardware).
203¢1
E_SCD_HT_VER_SCB
Meaning : Photomultiplier high tension control software detected a wrong
version of scan control board (SCB).
Negative version code from scd_scb_version(). Software error.
ReasonCure. : Check if software
stalled correctly,
203¢2
E_SCD_HT_PRESENT
Meaning : No high tension control hardware found.
Reason Scan control board (SCB) not or badly inserted.
- SCB defective.
Cure : Check / insert / replace SCB.
203¢3
E_SCD_HT_INI_HW
Meaning : Initializing the scan control board (SCB) by the photomultiplier
high tension control software failed.
Reason
Cure : Check / replace SCB.
2034
E_SCD_HT_CMF_TOO_FEW
Meaning : Less than two pairs of HT and SAL for calculation of MFA / MFB.
Reason ‘oftware error or wrong behaviour of system.
Cure. : Check correct settings. Calibrate MFA / MFB again.
203¢5
E_SCD_HT_CMF_TOO_MANY
Meaning : Too many pairs of HT and SAL for calculation of MFA / MFB.
Reason oftware error or wrong behaviour of system.
Cure : Check correct settings. Calibrate MFA / MFB again.
203¢6
E_SCD_HT_CMF_MFA_LOW
Meaning : Calculated value for MFA too low. Li
Reason: MFA / MFB calibration failed.
Cure : Check correct settings. Calibrate MFA / MFB again.
19 to mi
203¢7
E_SCD_HT_CMF_MFA_HIGH
Meaning : Calculated value for MFA too high. Limiting to maximum.
Reason —_: MFA / MFB calibration failed,
Cure : Check correct settings. Calibrate MFA / MFB again.
203c8
E_SCD_HT_CMF_MFB_LOW
Meaning : Calculated value for MFB too low. Limiting to minimum,
Reason MFA / MFB calibration failed.
Cure. : Check correct settings. Calibrate MFA / MFB again,
203¢9
E_SCD_HT_CMF_MFB_HIGH
Meaning : Calculated value for MFB too high. Limiting to maximum.
Reason MFA / MFB calibration failed.
Cure : Check correct settings. Calibrate MFA / MFB again.
203cA
E_SCD_HT_OVER
Meaning : Asked photomultiplier high tension too
Limiting to maximum.
Wrong parameters to scd_ht_cale_volt().
- Wrong manually selected high tension.
Check cassette parameters, MFA, MFB and laser power,
igh.
Reason
Curethe values are correct.
- If voltage manually set: Ignore error. Scanner uses max.
203cB
E_SCD_HT_ADC
Meaning : Reading back the photomultiplier high tension failed.
Reason roblem with the A/D converter on the scan control board (SCB).
Cure : Check / replace SCB.
203cc
E_SCD_HT_READ_OVER
Meaning : The measured high tension for the photomultiplier is more than
10% above the expected value.
Cable from high tension supply to photomulti
- Photomultiplier high tension supply defective.
= SCB board defective.
Reason jer disconnected.
Cure: - Check cable from high tension supply to photomulti
- Check / replace photomultiplier high tension supply.
~ Check / replace SCB board.
203CE
E_SCD_HT_READ_UNDER
Meaning : The measured hit n for the photomultiplier is more than
10% below the expected value.
Reason _: - Cable from SCB board to high tension supply defective or
not connected properly.
- 24V supply or 24V cable to high tension supply defe
not connected properly.
- High tension supply defective
= SCB board defective.
Cure: - Check cabelling and repair.
- Check / replace High tension supply (for photomultiplier).
= Check / replace SCB board.
= Check / replace 24V supply.
203D0
E_SCD_HT_CHT_sc
Meaning : Speed class is below or equi
Corrected to 200,
Reason Wrong parameters.
Cure. : Check software configuration.
zero, which is not allowed.
203D1,
E_SCD_HT_CHT_IP_FACTOR
Meaning : IP factor is below or equal zero, which is not allowed.
Corrected to 1000,
Reason Wrong parameters.
Cure : Check software configuration.
203D2
E_SCD_HT_CHT_LAS_POW
Meaning : Laser power is below or equal zero, which is not allowed.
Corrected to 17000 [uW].
Reason roblem in laser power measurement.
Cure. : Check software configuration.
203D3
E_SCD_HT_CHT_PIXEL_SIZE
Meaning : Pixel size is below or equal zero, which is not allowed,Corrected to 1137 [um].
Reason _: Wrong parameters.
Cure : Check software configuration.
20304
E_SCD_HT_CHT_PIXEL_TIME
Meaning : Pixel time is below or equal zero, which is not allowed.
Corrected to 4000 [ns].
Reason Wrong parameters.
Cure : Check software configuration.
203D5,
E_SCD_HT_CHT_OVER
Meaning : Calculated photomult
Limiting to maximum.
Reason Wrong parameters.
Cure : Check cassette parameters, MFA, MFB and laser power,
if the values are correct.
ier high tension too high.
20306
E_SCD_HT_NOT_INIT
Meaning : The driver software for photomultiplier high tension control
has not been initialized.
Reason joftware error.
Cure : Software must call scd_ht_init() first,
20307
E_SCD_HT_OUT_RANGE
Meaning : While testing the scan control board (SCB), the read-back
voltage of the high tension D/A converter deviated more
than 10% from the expected value.
Reason —_: SCB / SMB / SSB defective.
Cure : Replace SCB / SMB / SSB.
20308
E_SCD_HT_CMF_HV0_LOW
Meaning : Calculated value for HV0 too low. Limiting to minimum.
Reason ——_: MFA / MFB calibration failed,
Cure : Check correct settings. Calibrate MFA / MFB again.
203D9
E_SCD_HT_CMF_HVO_HIGH
Meaning : Calculated value for HVO too high. Limiting to maximum.
Reason MFA / MFB calibration failed.
Cure. : Check correct settings. Calibrate MFA / MFB again,
203DA
E_SCD_HT_CMF_BAD_DATA
Meaning : Data does not allow to fit MFA, MFB and HV0,
Reason software error or wrong behaviour of system.
Cure. : Check correct settings. Calibrate MFA / MFB again.
203DB
E_SCD_HT_CMF_POSSIBLE_BAD_FIT
Meaning : Warning: possibly bad fit in MFA / MFB calibration.
Reason joftware error or wrong behaviour of system.
Cure : Check correct settings. Calibrate MFA / MFB again.
203DCE_SCD_HT_CHT_OVER_2
Meaning : Calculated photomultiplier high tension too high.
Limiting to maximum.
Reason Wrong parameters.
Cure : Check cassette parameters, MFA, MFB and laser power,
if the values are correct.
203DD
E_SCD_HT_CHT_UNDER_2
Li
Reason
Cure
Wrong parameters.
Check cassette parameters, MFA, MFB and laser power,
if the values are correct.
203E0
E_SCD_LAS_VER_SWH
Meaning : Wrong version of SWH board (ADC70 hardware),
203E1
E_SCD_LAS_VER_SCB
Meaning : Laser power control software detected a wrong version of
‘scan control board (SCB).
Reason jegative version code from scd_scb_version(). Software error.
Cure : Check if software is installed correctly.
203E2
E_SCD_LAS_PRESENT
Meaning : No laser power measuring hardware found.
Reason Scan control board (SCB) not or badly inserted.
- SCB defective.
Cure : Check / insert / replace SCB.
203E3
E_SCD_LAS_ADC
Meaning : Reading the laser power from the A/D converter failed.
Reason A/D converter timeout.
- A/D converter value mismatch,
= Wrong A/D converter number (software error).
Cure : Check / replace SCB board.
If this does not help, inform MED-CSO.
203E4
E_SCD_LAS_ADC_LOW
Meaning : The given A/D converter value is below zero,
This should never occur.
Reason oftware error.
Cure: Inform MED-CSO.
203E5
E_SCD_LAS_ADC_HIGH
Meaning : The given A/D converter value is too high.
‘Already a value with all converter bits = 1 is assumed as error.
Reason Most probably defective hardware (SCB board).
Cure : Check / replace SCB board.
20366
E_SCD_LAS_NOT_INIT
Meaning : The driver software for laser power control has not been initialized,Reason oftware error,
Cure : Software must call sed_las_init() first,
203E7
E_SCD_LAS_LOW_TESTPOW
Meaning : The measured test voltage from the PIN diode board is too low.
Reason PIN diode board disconnected or cable defective
= PIN diode board defective.
Cure : Check cable between scan control board (SCB) and PIN diode
board. If ok, replace PIN diode board.
203E8
E_SCD_LAS_HIGH_NORMPOW
Meaning : The measured voltage from the PIN diode board is too high.
Reason __: - PIN diode board disconnected or cable defective
= PIN diode board defective.
Cure : Check cable between scan control board (SCB) and PIN diode
board, If ok, replace PIN diode board,
203E9
E_SCD_LAS_NOT_PRESENT
Meaning : Laser diode control hardware not detected.
Reason Laser diode control board not connected.
- Bad cable or connector.
Cure : Check laser diode control board and cable,
203EA
E_SCD_LAS_NOT_OK
Meaning : Laser diode control hardware detected an error.
Reason jefect in laser diode control.
Cure : Replace laser diode control board.
203EB
E_SCD_LAS_SET_BOL_HOOK
Meaning : Setting the BOL interrupt hook function failed.
Reason jperating system problem.
Cure : Check if software is installed correctly.
20400
E_SCD_PRESENT_TSTREG
(No COMMENT)
20401
E_SCD_PRESENT_TIM1
(NO COMMENT)
20402
E_SCD_PRESENT_TIM2
Meaning : Timer 2 of DAQ board not present.
The software automatically checks for old ADC70 hardware
and then for COPRA hardware.
If ADC70 is not found, COPRA is checked.
Reason jo ADC70-DAQ board inserted.
Cure : Ignore this error, if no other errors present.
20403
E_SCD_PRESENT_PIT
(NO COMMENT)20404
E_SCD_PRESENT_FIFO
(NO COMMENT)
20405
E_SCD_PRESENT_HT_CTRL
(No COMMENT)
20406
E_SCD_PRESENT_SWP_OFF
(NO COMMENT)
20407
E_SCD_PRESENT_SWP_RET
(NO COMMENT)
20408
E_SCD_DAQ_RMSCOM
(No COMMENT)
20420
E_SCD_PRESENT_SWP_AMP
Meaning : Address SCB_DAC_ADD on scan control board not found.
Reason Scan control board (SCB) not or badly inserted.
= SCB defective.
Check / insert / replace SCB.
Cure
20421
E_SCD_PRESENT_INTCTRL
Meaning : Address INTERR_CTRL on scan control board not found.
Reason __: - Scan control board (SCB) not or badly inserted.
- SCB defective.
Cure : Check / insert / replace SCB.
20422
E_SCD_PRESENT_MIG
Meaning : Address MIG_STAT_CTRL on scan control board not found.
Reason __: - Scan control board (SCB) not or badly inserted,
= SCB defective.
Cure : Check / insert / replace SCB.
20423
E_SCD_ALLOC_IVEC
Meaning : Allocation of 32 contiguous interrupt vectors failed.
Reason __: Serious problems with operating system,
Cure : Check if software is installed correctly.
20424
E_SCD_FIND_IVEC
Meaning : Could not find interrupt vector with 4 least si
Reason serious problems with operating system.
Cure : Check if software is installed correctly,
nificant bits
20425
E_SCD_IVEC_BASE
Meaning : No valid interrupt vector available for scanning.
Reason nitialisation of interrupt vectors failed.
Cure : Check if software is installed correctly.