Professional Documents
Culture Documents
Table of Contents
1. Specifications, Classifications, Symbols (Aug 2014)..............3 6.17 Shower Flow, Air Flow Switch, Water Feed (Aug 2014)........49
3.1 Dimensions (Aug 2014)............................................................7 7.2 Displays and Keypad (Apr 2015)............................................51
3.3 Capacities (Aug 2014)..............................................................9 7.4 Programming Flow Chart (Apr 2015)......................................53
4.4 Connect Air Pre-Conditioner (Aug 2014)................................20 9.3 CCT-1 (Mar 2015)..................................................................62
4.5 Compressed Air (Aug 2014)...................................................22 9.4 GMW 14872, GM9540P (Mar 2015)......................................63
4.6 Water (Mar 2015)...................................................................23 9.5 Ford CETP: 00.00-L-467 (Feb 2015)......................................64
4.7 Drainage (Aug 2014)..............................................................24 9.6 SAE J2334, SAE J2721 (Mar 2015).......................................65
5.2 RH (Controlled Relative Humidity) Function (Aug 2014)........30 10.1 Chamber Temp Sensor, Wet/Dry Bulb (Aug 2014)...............69
6. Major Components..................................................................33 10.3 Shower Collection / Corrosion Rate (Aug 2014) ..................81
6.2 Fog Pump (Aug 2014)............................................................34 11.1 Connecting the Q-FOG to a Computer (Aug 2014)...............83
6.3 Shower Module (Aug 2014)....................................................35 11.2 Viewing the Data on the Computer (Aug 2014).....................84
6.5 Chamber Heaters (Aug 2014)................................................37 12.1 Removing the Blower Module (Aug 2014)............................86
6.6 Chamber Temp. Sensor & Over Temp. Switch (Aug 2014).....38 12.2 Shower Nozzle Cleaning (Mar 2015)...................................88
6.7 Diffuser (Aug 2014)................................................................39 12.3 Solution Reservoir, Shower System (Apr 2015)...................90
6.8 Bubble Tower (Aug 2014).......................................................40 12.4 Washable Air Filter Cleaning (Apr 2015)..............................96
6.12 Lid Interlock (Aug 2014).......................................................44 14. Replacement Parts (Mar 2015)............................................ 102
Page 2
Section 1: Specifications, Classifications, Symbols
Specifications, Classifications
Operating Temperature and Humidity: The recommended ambient operating temperature and relative humidity is 23
± 5oC and < 75% RH. The maximum ambient operating temperature and humidity is 40oC and 80% relative humidity.
Temperatures outside the recommended range may cause chamber temperature and/or humidity faults.
Sound Pressure Level: Sound Pressure Level does not exceed 75 dBA.
Voltage: 230V or 208V (as stated on the nameplate) ± 10% - single phase.
Frequency: 50 or 60 Hz.
Supply Connection: Permanently connected or plug/socket connection (industrial type per IEC 6309 or twist lock type
in North America).
External Over-Current Protection: Must be rated for not more than 40A (USA, Canada) or 64A (Europe).
Page 3
Section 1: Specifications, Classifications, Symbols
Symbols
Attention
Page 4
Section 2. Safety (Aug 2014)
Safety Devices
●● Chamber, Over Temperature Cut-out
Operator Safety
Hazardous Vapors
Do not use gases such as SO2 in the Q-FOG. Do not use hazardous or petroleum based organics (solvents). Purge
the chamber of airborne mist or fog before opening the chamber lid.
Product Safety
The Q-FOG meets the European Low Voltage Directive 2006/95/EC and complies with the requirements of
EN61010-1: 2010 (Third Edition), “Safety of Electrical Equipment for Measurement, Control and Laboratory Use”.
Electromagnetic Compatibility
The Q-FOG meets the European Electromagnetic Directive 2004/108/EC and complies with the requirements of
EN 55011:2007 Radiated and Conducted Emissions – class A.
Page 5
Section 3. General Description
The Q-FOG model CRH performs salt spray, Prohesion and cyclic corrosion testing with full variable humidity control
through the use of an innovative blower module which adjusts the amount of moist and dry air that circulates through
the test chamber.
An air pre-conditioner removes moisture from ambient air to supply either cool or warm dry air to expand the
chamber operating temperature and relative humidity range.
Additionally, this model includes a programmable shower feature for compatibility with most major automotive
corrosion test standards, such as GMW 14872, SAE J2334 and others from Ford, ISO, VW, Volvo, Chrysler, Renault,
etc.
Page 6
Section 3. General Description
E
A F
B G
H
Front Side
93 cm
82 cm 36.5"
32.0" 82 cm
32.0"
101 cm
39.4"
Front Side
Q-FOG Air Pre-Conditioner
Page 7
Section 3. General Description
The Q-FOG CRH test chamber and air pre-conditioner should be positioned as shown in the figure below. This will
allow sufficient room to operate the unit and allow proper ventilation.
The air pre-conditioner is attached to the Q-FOG chamber by a 96" (244 cm) flexible hose. The air pre-conditioner is
mounted on casters and may alternatively be positioned in front of or behind the Q-FOG test chamber. The air pre-
conditioner may be moved to open the Q-FOG test chamber right side access panel.
25 cm WALL
10.0"
WALL
61 cm
Air 24.0"
Q-FOG Test Chamber
Pre-Conditioner
5 cm
Minimum
2.0"
Page 8
Section 3. General Description
Chamber Volume
Volume (excluding lid) 511 liters (18.0 ft3) 857 liters (30.2 ft3)
Volume (including lid) 640 liters (22.6 ft )
3
1103 liters (38.9 ft3)
Page 9
Section 4. Set Up
4. Set Up
Do not attempt to remove the chamber lid without special instructions. Contact Q-Lab
Corporation Repair and Technical Support Department at +1-440-835-8700 for detailed
instructions.
Q-Lab China
Tel: +86-21-5879-7970
Email: RepairChina@q-lab.com
Page 10
Section 4: Set Up
Q-FOG Model
CRH600 CRH1100 Air Pre-Conditioner
Shipping Weight:(Approximate) 390 kg (860 lb) 472 kg (1040 lb) 175 kg (385 lb)
Machine Weight: 333 kg (734 lb) 378 kg (834 lb) 91 kg (200 lb)
Tools Required
Remove Remove
1. Use pry bar to remove 2 particle board sheets from the top of the Q-FOG crate.
2. Remove the 2 particle board sheets from the sides of the Q-FOG crate.
Page 11
Section 4. Set Up
Metal Tab
3. Remove the metal reinforcing tabs at each end of the 10 top support boards.
Ridge Board
Page 12
Section 4: Set Up
Remove Remove
6. Remove the 2 particle board sheets from the ends of the Q-FOG crate.
Remove Remove
7. Remove the vertical support board from both ends of the crate.
Remove Remove
8. Remove the vertical support boards from both sides of the crate.
Page 13
Section 4. Set Up
Remove Remove
9. Remove the horizontal support boards from both sides and ends of the crate.
Cut
Remove
X4
10. Remove 4 corner posts. 11. Use a utility knife to cut the straps holding the
Specimen rack package to the pallet. Remove the
specimen racks.
Q-FOG
New Image Rear
12. Locate the 4 15 mm (9/16") bolts and nuts holding the Q-FOG to the pallet.
Page 14
Section 4: Set Up
Remove
X4
Forks Must be
Placed Under
Metal Plates
13. Use a 15 mm (9/16") socket to remove the 4 nuts 14. Use a fork lift to lift Q-FOG up off the pallet.
and bolts.
Metal plate
Metal plate
Forks go here.
15. Make sure forks are under the metal plates on the bottom of the Q-FOG. See the label on the front of the Q-FOG
for correct fork position.
Forks Must be
Placed Under
Metal Plates
USE EXTREME CAUTION: Forks can easily damage the Q-FOG. Make sure forks
are under metal plates before lifting.
Page 15
Section 4. Set Up
Remove Remove
Remove Remove
Remove
Remove
16. Use a 15 mm (9/16") socket to remove the 2 nuts 17. Use a15 mm (9/16") socket to remove the 2 nuts and
and bolts in the right angle brackets under the bolts in the left angle brackets under the Q-FOG.
Q-FOG. Remove the board. Remove the board.
Adjust
Down
18. Adjust the 4 leveling feet down about 38 mm (1.5"). 19. Carefully move the Q-FOG to the installation location
See Section 5 for leveling information. and lower into position.
Remove Plastic
Wrap
20. Use a utility knife to carefully remove the plastic wrapping. Use care to avoid scratching the Q-FOG cabinet.
Page 16
Section 4: Set Up
21. Q-FOG cabinet uncrating is complete. Next uncrate the air pre-conditioner.
22. Use pry bar to remove the particle board sheets and framing from the Q-FOG Air Pre-conditioner crate
Page 17
Section 4. Set Up
The Q-FOG test chamber must be leveled correctly to assure a proper lid seal and prevent leakage of corrosive fog
into the laboratory.
2. Close the lid and close the lid latch by rotating the latch counter-clockwise until it fully engages with the center lid
handle post. A gentle push downward on the handle makes this easier.
3. Adjust the support feet so the Q-FOG test chamber is reasonably level and all four feet touch the ground (does
not rock). The support feet are adjusted using a wrench.
4. To check for leakage, put a light inside the chamber and look for light escaping. The Q-FOG test chamber has a
light built in. The power cord for an external inspection light may exit through the air exhaust pipe.
5. Make sure the lid is closed and latched. Look all along the lid seal for light escaping between the lid seal and
the chamber.
6. If light is escaping from the left side of the lid, lift the left front corner of the chamber by adjusting the left
front support foot. If light is escaping from the right side of the lid, lift the right front corner of the chamber by
adjusting the right front support foot. The support foot should be adjusted just enough to prevent light from
escaping.
7. After adjusting a support foot, the Q-FOG test chamber may rock because only three feet are touching the
ground. If it rocks, extend the foot that is not touching until it touches the ground. The foot should be extended
just enough to touch the ground. If it is extended more than this it may create another leak.
Page 18
Section 4: Set Up
Installation: A qualified electrician should connect the mains according to local codes using approved disconnect
and over-current protection. On the back of the Q-FOG tester is a Main Power Circuit Breaker and a Residual
Current Device (RCD). Below this breaker is a hole with strain relief for the mains power entry. Your electrician can
enlarge this hole to accept his chosen electrical conduit or power cord. The mains wires should be connected to the
breaker. Wire used for connection is required to be rated for 90oC or higher.
Voltage: The Q-FOG test chamber is powered by either 208V ± 10%, 1-Phase, 50/60 Hz or 230V ± 10%, 1-Phase,
50/60 Hz. The input voltage is shown on the nameplate attached to the rear of the chamber.
It is important that the cabinet be connected to the correct voltage. If the voltage is too low, the chamber will not heat
up quickly enough. If the voltage is too high, it could cause damage to the chamber. Maximum current loads follow:
Maximum
Maximum Current at
Current at
Model 208V ± 10% 230V ± 10%
Air Pre-Conditioner
The air pre-conditioner includes a cable that connects to the rear of the Q-FOG test chamber. The Q-FOG test
chamber supplies the power to the air pre-conditioner as follows: either 208V ± 10%, 1-Phase, 50/60 Hz or 230V ±
10%, 1-Phase, 18 Amps.
Page 19
Section 4. Set Up
1. Open 2 latches on Q-FOG access panel. 2. Remove the Q-FOG right side access panel.
Air Pre-conditioner
Hose from
Air Pre-conditioner
Blower
Module
3. Cut the ties on the air pre-conditioner hose and power 4. Route the air pre-conditioner hose through the hole in
cable. the Q-FOG cabinet, to the blower module as shown.
Blower
Module
5. Connect hose to the blower module. 6. Replace the access panel. Close the latches.
Page 20
Section 4: Set Up
Air Pre-conditioner
Connect
Cable Here
Page 21
Section 4. Set Up
Purity:
Purity: The compressed air supply must be clean, dry, and oil free. The presence of solid impurities, water, oil or
other contaminants in the compressed air may affect test results. The Q-FOG CRH is equipped with a moisture filter
to improve the incoming compressed air quality. As moister accumulates, the filter purges to a water drain.
Do not add a lubricator to your compressed air system. Oil in the air will reduce the corrosive effects of the electrolyte
on the test specimen.
Pollutants in the air may increase the corrosive effects. Impurities in the air supply may also clog the spray nozzle,
solenoid valves or check valves and cause maintenance problems.
Pressure:
4-10 bar (60-150 psi)
Volume:
1.7 lps (3.5 cfm) maximum
Installation:
Connect a 10 mm (3/8") I.D. compressed air hose to the inlet at the rear of the Q-FOG test chamber cabinet.
Pressure
Regulator
Supply
60 psi
Filter
Drain
Page 22
Section 4: Set Up
●● Ordinary tap water will damage this tester and void your warranty.
●● Water purity requirements are listed below.
The Q-FOG tester requires a supply of pure water for the Bubble Tower. This is used for saturating the compressed
air with water vapor during the fog function. Q-FOG CRH tester also needs pure water for producing chamber
humidity.
The water for these processes must be distilled or deionized and, except for the salt solution make-up water, must be
connected directly to the unit. Proper water quality can be achieved by one of the following methods: Distilled Water,
Reverse Osmosis (RO), De-ionizer (DI) with filter. We recommend the use of a deionizer to remove dissolved solids,
with a filter to remove suspended solid particles. Usually the least expensive way to get the proper water quality is to
use local vendors’ water purification services.
Total
Maximum Maximum
Pressure Resistivity Conductivity Dissolved pH
Flow Rate Volume
Solids
9-56 psi
0.4 liter/min. 5 liters/hour > 200k ohm•cm < 5 μS/cm < 2.5 ppm 6-8
(0.6-3.8 bar)
NOTE: Use of very high purity water (greater than 5MΩ/cm or less than 0.2Ms/cm) for extended periods is not
recommended. Very high purity water may corrode stainless steel and other corrosion resistant metals in the
Q-FOG tester.
Installation:
Connect 10mm (3/8") I.D. pressure hose to the water inlet at the rear of the cabinet and the water supply. This should
be fitted with a stop valve.
When all three services have been connected the air and water can be turned on and checked for leaks at the
connection points.
Page 23
Section 4. Set Up
Chamber Drain:
The main testing chamber is fitted with an outlet to enable excess solution to go to a sanitary sewer drain. This
system consists of a bulkhead fitting, elbow, hose barb, and threaded nipple already in place in the chamber outlet.
Run the hose to a drain. The end of the hose must be below the water level in the drain, sealed into the drain, so that
corrosive vapors cannot escape. Alternatively, a trap can be connected to the hose, so that vapors cannot escape.
Air Pre-Conditioner
The end of the drainage hose from the bottom of the air pre-conditioner should be placed into a floor drain.
Page 24
Section 4: Set Up
Be sure to comply with all local regulations regarding chemical exhaust. Usually simple salt water is not regulated.
Proper venting will prevent corrosive mist from entering the laboratory and assure correct performance of your
Q-FOG tester. Air venting from the chamber is done via a 114 mm (4.5") outside diameter vent tube. Because of the
typically corrosive nature of the exhaust, plastic is strongly recommended. The user must connect an
exhaust system with a (102 mm) 4.0" inside diameter (minimum) vent line.
The length of the venting system must be less than 30 meters (100 feet) and contain fewer than 10 elbows (i.e. 90°
turns). If it must be longer, or if it must contain more turns, there could be significant back pressure on the chamber
and this might cause corrosive mist to enter the laboratory and/or reduce the tester’s performance capabilities.
A “tee” is recommended at the end of the vent to reduce the risk of periodic back-pressure from the wind. Place a
screen over the vent to discourage birds and small animals.
Connections: Because some liquid from the salt fog may condense on the inside of the vent tube, we
recommend that all sections of the vent tube have a slope either back toward the chamber or away to a drain. Avoid
all installations that could create moisture traps. Otherwise, liquid may build up in the vent tube and cause problems
with back-pressure. Excessive amounts of moisture build up can create a weight load on the vent tubing.
The figures on the following page illustrate right and wrong ways to connect the Q-FOG chamber vent. When
variations are necessary, please consult Q-Lab Corporation for recommendations.
Page 25
Section 4. Set Up
Correct Installations
Make sure the test chamber vent is installed so that moisture cannot build up and create a trap as shown below.
X X
Incorrect Installations
Page 26
Section 4: Set Up
Utility Connections
Floor
Drain
Chamber drain hose must be inserted below Chamber Drain Hose
water level in floor drain, or a trap must be used 4.5 m (15 feet) X 32 mm (1 -1/4")
to keep solution mist from entering room supplied
Electrical
Power Entry Air Vent
114 mm (4.5") O.D.,
102 mm (4.0") I.D. tube
Air In 10 mm (3/8") must be connected to
hose required the outdoors
See Section 3.7
Water In 10
mm (3/8") hose
required
Bubble Tower
Drain
Air Filter Drain Chamber Drain Solution Reservoir Drain
Page 27
Section 5. Functions
5. Functions
The chamber can cycle between Fog, RH, and Shower functions.
The Fog and Shower functions cannot be run in the same step.
The shower module must be removed from the Q-FOG chamber to run a fog step (See Section 5.3).
Page 28
Section 5. Functions
The fog function sprays a fine mist of corrosive solution throughout the chamber. The system works as follows.
●● The fog pump sends solution from the reservoir to the spray nozzle.
●● Pump speed is selected via the keypad using the P7 program. The flow rate is displayed on the flow meter.
●● The air solenoid valve sends compressed air through the bubble tower and then to the spray nozzle where it
atomizes the solution. The air supply pressure is set via the fog pressure regulator located on the control panel.
The pressure is displayed on the pressure gage.
●● See Section 10.2 for more information on setting air pressure and pump speed.
Vent
Lid
Controller
Solution Specimens
Reservoir
Fog Pressure
Regulator and Gage
Bubble
Tower Compressed
Air In
Fog
Nozzle
Chamber
Heater
Fog Pump
Solution
To Pump
Flow
Meter
Fog Pump
Page 29
Section 5. Functions
The RH function allows the user to control the relative humidity and temperature of the test chamber. The system
works as follows:
●● The RH generation system injects atomized pure water into the air stream to control the RH level.
●● The controller senses chamber relative humidity and pulses the atomizing nozzles to inject the correct amount
of moisture into the chamber.
●● The blower module recirculates air through the chamber and adjusts the amount of outside air brought into the
chamber.
●● Outside air may be pre-conditioned to allow lower temperature and relative humidity settings than otherwise
possible.
●● Dry-off of specimens occurs by programming low relative humidity (e.g. < 30%) conditions.
●● The chamber heaters and air pre-conditioner heater maintain chamber air temperature.
●● The operator can choose one of three ramp modes when programming a controlled RH step:
oo Linear: User inputs transition time and controller automatically adjusts RH and temperature to achieve a
linear adjustment from previous conditions.
oo Auto: The controller achieves the RH and temperature as quickly as possible.
oo Less Than: User inputs the maximum transition time specified by the standard and the controller achieves the
RH and temperature as quickly as possible.
Humidification
Air Pre-Conditioner
Compressed
Air In
D.I.
Blower Water In
Module
Conditioned Air
Conditioned
Air
Blower
Module
Air in from
Air Pre-conditioner
Page 30
Section 5. Functions
90 3 Hours
80
Relative Humidity (%)
Temperature (°C) and
70 Temperature
60
Relative Humidity
50
40
30
20
900 950 1000 1050 1100 1150 1200
Cycle time (Minutes)
"Less
Auto Than" Transistion
and Less Time from
Than Transition Times
Dry to
from DryAmbient Conditions
to Ambient Conditions
80
70 40 Minutes
Relative Humidity (%)
Temperature (°C) and
60
Temperature
50
Relative Humidity
40
30
20
10
1400 1420 1440 1460 1480 1500
Cycle time (Minutes)
Page 31
Section 5. Functions
The shower function is used in GMW 14872 and other automotive test specifications. A user-adjustable volume of
solution is uniformly sprayed onto specimens through specially positioned nozzles. The spray function is used in
place of the fog function.
●● The user installs the shower module in the test chamber over the existing fog spray nozzle.
●● The shower pump sends solution to four non-atomizing spray nozzles.
●● Spray droplets are much bigger, flow rates are much higher, and shower times are much shorter than the
atomized solution mist produced by the fog function.
●● The shower nozzles spray solution in an arc pattern.
●● The shower nozzles are self-cleaning.
●● The shower flow rate is constantly monitored and an alarm is sounded if flow is not within acceptable limits.
●● An automatic flow rate detection circuit detects plugged spray nozzles or substantial leaks.
●● The shower pulse rate can be changed in the P10 controller program.
●● Mass loss rates of corrosion coupons can be adjusted by increasing or decreasing shower pulse rate.
●● See Section 10.3 for more information on setting shower pump regulator.
Shower Module
Shower Pump
Shower Regulator and
Pump Pressure Gage
To Install Shower Module,Lower Module into Connect Supply Tube to Shower Module Installed in Q-FOG
Chamber as Shown Solution Supply Chamber
Page 32
Section 6. Major Components
Fog Spray
Nozzle
Cap
X
Clean Nozzle
Page 33
Section 5. Functions6. Major Components
●● The solution reservoir filter before the pump prevents particles or contaminants from entering the pump and
spray nozzle.
Solution
Reservoir
Filter
Fog
Pump
Pump Tubing
NOTE: The Q-FOG must be running a fog step in order to set the bubble tower air supply pressure.
Page 34
Section 6. Major Components
The shower module sprays drops of liquid solution directly onto test specimens. These drops are larger than those
produced by atomizing nozzles that use compressed air. The shower pulse rate is programmable. The shower
module is installed directly over the fog nozzle. NOTE: The shower module must be removed to run a fog step.
Connect to Solution
Supply Here
Shower Module Shower Module Installed
1/4 Turn to
Remove
Shower Nozzles Turn Counter Clockwise 1/4 Turn to Remove for Cleaning
To view the value that the tester assigned to the normal flow; select: ? then UP ARROW on the control panel. The
controller displays the current shower flow value and the normal value: D25 SHOWER FLOW= XX; NORMAL= XX.
When the Q-FOG is running a test cycle, during shower steps the shower flow is monitored. If the shower flow is not
within ±10% of the normal flow set by the process outlined above, an alarm will sound. The test is not stopped. The
alarm indicates that shower flow is not normal and the operator should determine the cause of the abnormal flow.
Possible causes include clogged shower nozzles, shower module malfunction and water supply pressure fluctuations.
Page 35
Section 6. Major Components
Shower Pump
Page 36
Section 6. Major Components
Two heaters at the bottom of the chamber are used to maintain temperature during fog and relative humidity steps.
Diffuser Plates
RH Generator
Fog Spray
Nozzle
Chamber Chamber
Heater Heater
Page 37
Section 6. Major Components
6.6 Chamber Temp. Sensor & Over Temp. Switch (Aug 2014)
●● The chamber temperature sensor is mounted in a flexible probe on the right side of the chamber.
●● An over temperature switch is also mounted on the right side of the chamber to protect the chamber from
excessive heat.
Over
Temperature
Switch
Inside
Chamber
Temperature
Sensor
Page 38
Section 6. Major Components
Diffusers are perforated fiberglass sheets positioned above the chamber heaters.
●● The diffuser sheets help diffuse heat during the fog function and spread air circulated during the RH function.
●● The diffuser plates also protect people from touching the hot chamber heaters below.
●● The diffusers are not intended to hold test specimens (use panel racks or hanging rods to hold test specimens).
Diffuser Location
Diffuser
Diffuser
Diffusers in Chamber
Page 39
Section 6. Major Components
The purpose of the bubble tower is to saturate the compressed air used to atomize the salt solution.
●● The bubble tower contains a heater to keep the water 12°C hotter than the chamber temperature during
fog steps. This is adjustable in software when necessary. Contact Q-Lab Repair and Technical Support for
assistance.
●● The water level in the bubble tower is automatically maintained by the level switches and water fill solenoid
valve.
Safety Features:
Pressure Relief
Valve
Temperature
Sensor
Low & High Level Heater
Float Switches
Aerator Tube
Gasket
Compression Springs
Page 40
Section 6. Major Components
The blower module is the main control mechanism for the Q-FOG RH function.
●● The blower module controls the circulation of moist and dry air through the test chamber.
●● The blower module also contains wet and dry bulb thermometers to monitor test chamber relative humidity.
●● Additionally, the blower module can draw in ambient air to cool or reduce the relative humidity in the test
chamber.
Air to
Chamber
Air from
Pre-Conditioner
Cooling Fan
Recirc
Damper Flow Control
Air Cylinder
Duct
Seal
Page 41
Section 6. Major Components
●● Deionized water enters on the right side of the RH generator and compressed air enters on the left side. As
they mix, the compressed air atomizes the water into a fine mist, which evaporates to increase the RH in the
chamber.
RH Generator
RH Generator Nozzle
Deionized Compressed
Water In Air In
RH Generator Assembly
Page 42
Section 6. Major Components
●● The 120 liter (31.6 gallon) reservoir holds enough solution for five days of continuous salt spray.
●● Deionized or distilled water should be used when making up the solution.
●● Sensors detect when the solution level gets too low and will automatically stop the test when the reservoir is
empty.
Solution
Reservoir Lid
Solution Level
Sensors
Page 43
Section 6. Major Components
When the lid is opened the interlock switch will open and the pump, blower, and heaters will stop.
If the lid is opened in a fog step, however, the chamber will be full of salt fog which will escape into the room.
Lid
Interlock Switch
Page 44
Section 6. Major Components
●● The air pre-conditioner provides dry, temperature regulated air to the Q-FOG test chamber during the RH
function if necessary.
●● The air pre-conditioner consists of a chiller which cools and dries ambient air and a heater.
●● Depending on programmed test and laboratory ambient conditions, the air may be heated prior to being blown
into the test chamber.
Control
Control Unit
Unit Heater
Assembly Damper
Actuator
Over
Temperature Air Heater
Switch Assembly
Chiller
Assembly
Chiller
Evaporator Assembly
Air Filter
Temperature Sensor
Air Pre-conditioner Schematic Doors Open Control Unit and Chiller Components
Page 45
Section 6. Major Components
●● A temperature and relative humidity sensor is mounted under the front right corner of the Q-FOG test chamber.
●● This sensor measures both the ambient temperature and humidity in the laboratory and can be used to help
troubleshoot miscellaneous faults that might occur.
Laboratory
Temperature and Humidity
Sensor
Page 46
Section 6. Major Components
●● Solenoid valves control the flow of air and water to the fog, shower and RH components
●● Bubble tower, shower and RH control solenoid valves are located under the shelf below the blower module.
●● Two solenoid valves mounted on the blower module control air to open and close the recirculation and output
dampers.
Shelf
Water Solenoid
Valves
Solenoid Valve Location
Under Shelf Below Blower
Module
Air Solenoid
Valves
Shelf Raised to Access Solenoid Valves
Recirc
Solenoid
Output
Solenoid
From Top: Bubble Tower Water From Top: Bubble Tower Air Top: Blower Module Recirc Solenoid
Solenoid, RH Generator Water Solenoid, Shower Pump Solenoid Bottom: Blower Module Output
Solenoid and Shower Nozzle Purge and Fog/RH Generator Solenoid Solenoid
Solenoid.
Page 47
Section 6. Major Components
●● Water is filtered from the main supply line and the salt solution is filtered before the shower nozzles and the fog
spray nozzle.
●● The solution reservoir filter and the shower nozzle filter prevent particles or contaminates from entering the
pump and shower nozzles.
Solution
Reservoir Filter
Main Water
Supply Filter
Shower Nozzle
Solution Filter
Page 48
Section 6. Major Components
6.17 Shower Flow Sensor, Compressed Air Flow Switch, Water Feed Valve (Aug 2014)
●● The flow rate of the shower pump is monitored by the shower flow sensor
●● The compressed air flow switch monitors air flow to the bubble tower and then to the spray nozzle
●● The water feed valve controls water to the wet bulb temperature sensor
Shower Flow
Sensor
Page 49
Section 7. Controller
7. Controller
7.1 Description (Apr 2015)
The controller controls all functions of the Q-FOG. Interactive software allows easy programming and operation
using a keyboard and two 80 character LCD displays. The user can create custom test cycles or run one of the pre-
programmed test cycles.
The controller can store up to 10 test cycles of 25 steps each. Each step consists of a function (fog, RH, or shower),
a temperature, and a time. Step time can range from 1 minute to 99 hours 59 minutes.
The controller can be set to automatically stop a test after a predetermined amount of time. The current function,
temperature, and test time are continuously displayed. Errors messages are displayed if a failure occurs.
Page 50
Section 7. Controller
Page 51
Section 7. Controller
On the controller keypad press the PROGRAM key to enter program mode. Scroll to one of the programs menus
shown below.
●● Press the ENTER key to move one level deeper into a program.
●● Use the ESCAPE key to move up one level.
●● Use the arrow keys change parameters.
●● Use the ENTER key to move to the next level and to save the changes to any program.
Page 52
Section 7. Controller
Page 53
Section 7. Controller
Page 54
Section 8. Specimen Mounting (Mar 2015)
Test specimens can be mounted in panel racks or suspended from hanging rods. Typically, flat panels are mounted
in racks and 3D parts are suspended from hanging rods. Panel racks are available with either 6o or 15o from vertical
mounting slots.
Very large parts, can be mounted on grates. There are two different grates available. One mounts at the same
height as the panel racks, the other mounts at the same height as the diffuser panels. Use the rack level grates if
possible. Use the diffuser level grates if your parts are too tall for the rack level grates
IMPORTANT: The maximum weight of test specimens must not exceed 544 kg (1200 lb).
Page 55
Section 8: Specimen Mounting
Page 56
Section 8. Specimen Mounting (Mar 2015)
Page 57
Section 9. Running a Test
Page 58
Section 9. Running a Test
IMPORTANT: The boundaries shown in these graphs should be considered typical but we do not guarantee all
machines will operate at these limits.
Page 59
Section 9. Running a Test
ASTM B117 is the most widely used corrosion test procedure but it is not a cyclic corrosion test. Samples remain wet
at all times.
In spite of 80 years of use, there has long been general agreement that B117 test results do not correlate as well with
the corrosion seen in actual atmospheric exposures as cyclic tests do.
ASTM B117
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
1 Fog 35 24:00
2 FINAL STEP – GO TO STEP 1
Notes
Page 60
Section 9. Running a Test
The Prohesion test was developed in England for industrial maintenance coatings. The solution is much more dilute
than traditional salt fog.
ASTM G85.A5
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
1 Fog 25 1:00
2 RH 35 25* 1:00 Auto
3 FINAL STEP – GO TO STEP 1
* Indicates no value specified in the test method but a value is programmed into the tester
Notes
Page 61
Section 9. Running a Test
CCT-1 is specified by some Japanese automotive manufacturers. It is also sometimes called CCT-A.
CCT-1
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
Notes
Page 62
Section 9. Running a Test
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
3 Shower 25 0:03
4 RH 25 45 1:30 Auto
6 RH 49 95 0:30 Auto
Notes
Page 63
Section 9. Running a Test
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
1 Shower 25 0:10
2 RH 25 95 0:20 Auto
3 Shower 25 0:03
4 RH 25 95 2:27 Auto
5 Shower 25 0:03
6 RH 25 95 2:54 Auto
7 Shower 25 0:03
Notes
Shower Deposition Rate: Local – each collection vessel must be within 3-15 l/m2 after running the 19
minutes of shower time in steps 1-7, this is equivalent to 24-118 ml if using 10
cm diameter funnels to collect the solution.
Average – the average of all collection vessels must be within 5-10 l/m2 after
running the 19 minutes of shower time in steps 1-7, this is equivalent to 39-79 ml
if using 10cm diameter funnels to collect the solution.
Page 64
Section 9. Running a Test
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
3 RH 60 50 17:45 Auto
4 FINAL STEP – GO TO STEP 1
Notes
Page 65
Section 9. Running a Test
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
6 Shower 35 0:15
7 RH 35 97 1:45 Auto
Notes
Page 66
Section 9. Running a Test
Volkswagen PV 1210
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
3 RH 23 50 4:00 Auto
5 RH 23 50 48:00 Auto
6 FINAL STEP – GO TO STEP 1
Notes
Page 67
Section 9. Running a Test
Chamber Step
RH Ramp Time
STEP FUNCTION Air Temp Time Ramp
(%) (hh:mm)
(oC) (hh:mm)
1 Subcycle Repeat Steps 2-3 3x
2 Shower 35 0:15
3 RH 35 97 1:45 Auto
4 RH 45 50 4:00 Linear 2:00
5 RH 35 95 2:00 Linear 2:00
6 Subcycle Repeat Steps 7-9 7x
7 RH 35 95 4:00 Auto
8 RH 45 50 6:00 Linear 2:00
9 RH 35 95 2:00 Linear 2:00
10 Subcycle Repeat Steps 11-12 3x
11 Shower 35 0:15
12 RH 35 97 1:45 Auto
13 RH 45 50 4:00 Linear 2:00
14 RH 35 95 2:00 Linear 2:00
15 Subcycle Repeat Steps 16-18 5x
16 RH 35 95 4:00 Auto
17 RH 45 50 6:00 Linear 2:00
18 RH 35 95 2:00 Linear 2:00
19 Final Step - Go To Step 1
Notes
pH: ~ 4.2*
*As measured in the collection vessels. The concentration and acidity of
the solution are changed by heating and spraying. Adjust the mixture in the
reservoir to achieve these values in the collection vessels.
Shower Deposition Rate: 80 to160 milliliters/hour of shower time
Page 68
Section 10. Calibration
10. Calibration
10.1 Chamber Temperature Sensor, Wet and Dry Bulb Thermometers (Apr 2015)
For accurate temperature readings, the chamber air temperature sensor and chamber relative humidity wet and dry
bulb thermometers must be calibrated Calibrate the thermometers every six months in accordance with
ASTM E-220. The following items are needed to perform the calibrations:
Page 69
Section 10. Calibration
5. Open the Q-FOG right side access panel. Place the 6. At the right end of the Q-FOG chamber, loosen the
water near the tester. nut holding the chamber temperature sensor.
7. Pull the temperature sensor through the side wall of 8. Place the chamber temperature sensor in the beaker
the chamber. of water.
Remove
9. Remove the wing nut from the wet and dry bulb 10. Remove the wet and dry bulb thermometers from the
thermometer assembly on top of the blower module. Q-FOG blower module.
Page 70
Section 10. Calibration
Wet Bulb
Wick
11. Inspect wet bulb wick for any debris or deposits. 12. Place the wet and dry bulb thermometers in the
Clean or replace as necessary. beaker.
Wait
Page 71
Section 10. Calibration
? Program
Status
P1 SET TEST DURATION
Message
Escape Enter
17. Press PROGRAM on the Q-FOG control panel. P1 SET TEST DURATION is displayed.
? Program
Status
P4 CALIBRATE TEMPERATURE SENSOR
Message
Escape Enter
? Program
Status
WHICH SENSOR? CHAMBER TEMP
Message
Escape Enter
Page 72
Section 10. Calibration
? Program
Status
WHICH SENSOR? WET AND DRY BULB
Message
Escape Enter
20. Press UP ARROW to select WET AND DRY BULB WHICH SENSOR? WET AND DRY BULB is displayed.
? Program
Status
IS REFERENCE THERMOMETER IN PLACE?
Message
Escape Enter
? Program
Status
REFERENCE TEMPERATURE = XX.X
Message
Escape Enter
Page 73
Section 10. Calibration
26.1
? Program
26.1
Status
REFERENCE TEMPERATURE = 26.1
Message
Escape Enter
? Program
Status
ENTER=ACCEPT CHANGES; ESCAPE=CANCEL
Message
Escape Enter
Page 74
Section 10. Calibration
? Program
Status
SAVING CHANGES
Message
Escape Enter
? Program
Status
P1 SET TEST DURATION
Message
Escape Enter
27. Press PROGRAM on the Q-FOG control panel. P1 SET TEST DURATION is displayed.
? Program
Status
P4 CALIBRATE TEMPERATURE SENSOR
Message
Escape Enter
Page 75
Section 10. Calibration
? Program
Status
WHICH SENSOR? CHAMBER TEMP
Message
Escape Enter
? Program
Status
IS REFERENCE THERMOMETER IN PLACE?
Message
Escape Enter
? Program
Status
REFERENCE TEMPERATURE = XX.X
Message
Escape Enter
Page 76
Section 10. Calibration
26.1
? Program
26.1
Status
REFERENCE TEMPERATURE = 26.1
Message
Escape Enter
? Program
Status
ENTER=ACCEPT CHANGES; ESCAPE=CANCEL
Message
Escape Enter
Page 77
Section 10. Calibration
? Program
Status
SAVING CHANGES
Message
Escape Enter
36. Remove reference thermometer sensor from beaker. 37. Remove the chamber temperature sensor and the
wet and dry thermometers from beaker.
Wet Bulb
Tube
51mm (2")
Blower
Module
38. Reinstall the wet and dry bulb thermometers in the 39. Measure the distance from the blower module
Q-FOG blower module. Be sure the wet bulb wick opening to the top of the wet bulb tube. The top of
goes into the water supply tube. the tube should be 51mm (2.0") above the blower
module opening.
Page 78
Section 10. Calibration
51mm (2")
Strain
Relief
40. If necessary loosen the strain relief and move 41. Push the chamber temperature sensor through the side
the tube up or down so that the top of the tube is wall back into the chamber.
51mm (2.0") above the blower module opening.
Tighten
Page 79
Section 10. Calibration
The quantity and uniformity of the salt fog should be verified before each test. This is done by placing collection
vessels in the chamber. ASTM B117 calls for the use of 2 collection vessels, while several other international
standards specify the use of 4 or 6 collection vessels. Q-Lab recommends the use of 6 collection vessels, which are
provided with the Q-FOG tester. The collection vessel consists of a 10cm diameter funnel and a graduated cylinder.
The recommended placement of these vessels is 1 in each corner of the test chamber and 1 on each side of the fog
spray nozzle approximately 15 cm from the nozzle (See figure below). Replacement collection vessels are available
from Q-Lab (part no. F-9001-K).
Fog Spray
Nozzle
Collection Collection
Vessel Vessel
Test Chamber
Place the cylinders on top of the diffuser plate or through the holes in the panel holders. If placed in the panel racks,
remove the base and put an O-ring around the neck of the cylinder to hold it in the rack. Make sure that the test
samples do not shield the funnels.
100 100
90 90
80 80
70 70
60 60
50 50
40 40
30 30
20 20
10 10
Graduated Cylinder with Base Attached resting on Graduated Cylinder with Base Removed Resting in
Diffuser Plate Panel Rack (Held in Place with O Ring Positioned on
80ml Mark)
Refer to the appropriate test method for the required fog deposition, for example, ASTM B117 specifies that the
collection rate should be between 1.0 and 2.0 milliliters of solution each hour (averaged over not less than 16 hours)
per 80 cm2 of area.
If too much fog is collected in both vessels, reduce the pump speed. If there is too little fog collected, increase the
pump speed.
If there is more fog collected in the vessels closest to the nozzle, increase the air pressure to throw the spray farther
from the nozzle. If there is less fog collected in the vessels closest to the nozzle, decrease the air pressure to throw
the spray closer to the nozzle.
The spray nozzle can also be tilted to balance the amount of fog collected between the collection vessels. Tilt the
nozzle by loosening the big nut under the nozzle, tilt the nozzle, and re-tighten the nut.
Page 80
Section 10. Calibration
The quantity and uniformity of the shower should be verified before each test. This is done by placing collection
vessels in the chamber. See Section 10.2 for recommended collection vessel placement. The collection vessel
consists of a 10cm diameter funnel and a graduated cylinder. Replacement collection vessels are available from
Q-Lab (part no. F-9001-K).
Place the cylinders on top of the diffuser plate or through the holes in the panel holders. If placed in the panel racks,
remove the base and put an O-ring around the neck of the cylinder to hold it in the rack. Make sure that the test
samples do not shield the funnels.
100 100
90 90
80 80
70 70
60 60
50 50
40 40
30 30
20 20
10 10
Graduated Cylinder with Base Attached resting on Diffuser Plate Graduated Cylinder with Base Removed
Resting in Panel Rack (Held in Place with O
Ring Positioned on 80ml Mark)
Refer to the appropriate test method for the required shower solution volume.
If too much shower solution is collected in the vessels, reduce the shower pulse rate time. If too little shower solution
is collected, increase the shower pulse rate time. Shower pulse rate ON / OFF time is set in the P10 program.
Default is 15 sec. On 15 sec. OFF. Spray ON time may be increased or decreased to adjust mass loss rates of
corrosion coupons.
The shower pump regulator is used to adjust the uniformity of the shower spray in the chamber. If there is more
shower solution collected in the vessels closest to the shower module, use the shower pump regulator to increase
the pressure to project the spray farther from the shower nozzles. If there is less solution collected in the vessels
closest to the shower module, decrease the pressure to project more spray closer to the shower nozzles. Also be
sure that the nozzles are spraying with a uniform pattern and are not plugged.
Page 81
Section 11. Data Logging
The Q-FOG may be connected by an Ethernet cable to a computer or a network. Data logging software is available
to log operating information automatically.
Contact the Q-Lab Repair and Tester Support or Sales Departments to obtain the software (for Microsoft Windows
XP, Vista, Windows7 and Windows 8) and instructions.
Q-Lab Headquarters Q-Lab Europe, Ltd. Q-Lab Deutschland GmbH Q-Lab China 中国代表处
Westlake, OH USA Bolton, England Saarbrücken, Germany Shanghai, China 中国上海
Tel: +1-440-835-8700 Tel: +44-1204-861616 Tel: +49-681-857470 电话: +86-21-5879-7970
Repair@q-lab.com RepairEurope@q-lab.com RepairGermany@q-lab.com RepairChina@q-lab.com
Page 82
Section 11. Data Logging
To a Hub
Page 83
Section 11. Data Logging
Stripchart View
Table View
Page 84
Section 12. Maintenance
12. Maintenance
Air Pre-Conditioner
Air Filter Clean Salt Build-Up
Page 85
Section 12. Maintenance
Disconnect air pre-conditioner hose (1A) from the blower module (1B).
Page 86
Section 12. Maintenance
Lift (4A) and latch (4B) shelf for access to bottom components.
Tubing
Location of preventive maintenance items (5, 6, 7, 8, 9, 10). Open Peristaltic Pump to Check Tubing
Page 87
Section 12. Maintenance
Shower
Nozzles
Shower
Module
1. Open the Q-FOG chamber lid. 2. Locate the shower module and nozzles in the
chamber.
Remove 4
Nozzles
3. Turn nozzle counter clockwise 1/4 turn to remove 4. Locate the beaker in the kit (F-8064-K).
from shower module. Remove all 4 nozzles.
5. Place 4 old nozzles into the beaker. 6. Locate 4 new nozzles in the kit.
Page 88
Section 12. Maintenance
Distilled White
Vinegar
Soak 5 Minute
7. Install 4 new nozzles in the shower module. Turn 8. Fill the beaker with old nozzles 1/2 full (50 ml) of
clockwise 1/4 turn by hand to tighten. distilled white vinegar. Soak 5 minutes.
9. Under running clean tap water, fill the beaker to 12. Empty and refill breaker with clean water 4 times to
overflowing. throughly rinse nozzles.
Every 3 Months
Store Clean
Nozzles
4 Nozzles in
Q-FOG
Page 89
Section 12. Maintenance
The solution tank and shower system should be cleaned every six (6) months or after every 4000 hours of Q-FOG
operation. More frequent cleaning may be necessary when running GMW 14872.
The solution specified in GMW 14872 creates a precipitate, calcium carbonate (CaCO3). This precipitate will build up
inside the reservoir and can clog the shower system filters, hoses and nozzles. Calcium carbonate is not soluble in
water, so distilled vinegar is used to dissolve the calcium carbonate.
Tools Required:
●● 7.6 liter (2 gallon) white vinegar ●● Clean water supply
●● Safety Glasses
1. Open solution reservoir drain valve under the reservoir (See Section 4.9).
Page 90
Section 12. Maintenance
4. Use a long handled circular brush (nylon or plastic, not wire) to remove calcium carbonate deposits from sides
and bottom of reservoir
7. Use clean water to flush out any particulate material that may have accumulated in reservoir outlet
9. Pour 1/2 gallon of white vinegar into reservoir, wetting reservoir sides as vinegar is poured into reservoir
10. Use the brush to scrub sides and bottom of reservoir with vinegar
Page 91
Section 12. Maintenance
3. Remove shower nozzles from shower nozzle assembly (See Section 6.3)
Page 92
Section 12. Maintenance
Solenoid
Valves
Press Down
to Start Pump
Pump Solenoid
8. Insert screwdriver into shower pump solenoid manual override and press down to start shower pump
Page 93
Section 12. Maintenance
12. Insert screwdriver into shower pump solenoid manual override to start shower pump
15. Remove solution reservoir filter and cartridge. 16. Remove shower nozzle filter and cartridge.
17. Clean particulate material from filter bowls and cartridges with clean water
Page 94
Section 12. Maintenance
35. Clean shower nozzles (See Section 12.2 and Service Instruction F-8064-L)
Page 95
Section 12. Maintenance
Air Pre-Conditioner
Air Filter
1. Open the Air Pre-Conditioner front access door. Locate the washable air filter at the bottom of the enclosure.
Page 96
Section 12. Maintenance
Water Flow
4. Hold filter under CLEAN running water. 5. Wash all of filter. NOTE: If rinsing with water
IMPORTANT: Water flow MUST BE in direction of does not remove dirt, use an electrostatic air filter
WASH arrow. cleaner spray.
7. Reinstall the washable air filter in the Air Pre-Conditioner. IMPORTANT: Make sure arrow on the airflow label is
pointing up.
Page 97
Section 13. Trouble Shooting
Press the ? key on the Q-FOG control panel to enter the diagnostic mode for troubleshooting.
Page 98
Section 13. Trouble Shooting
Page 99
Section 13. Trouble Shooting
Page 100
Section 13. Trouble Shooting
Page 101
Section 14. Replacement Parts
Page 102
Section 14. Replacement Parts
Page 103
Section 14. Replacement Parts
Page 104
Section 14. Replacement Parts
Page 105
Section 14. Replacement Parts
Page 106
Section 14. Replacement Parts
F-9223-K Controller
Page 107
Section 14. Replacement Parts
Page 108
Section 14. Replacement Parts
Page 109
Section 14. Replacement Parts
Page 110
Section 14. Replacement Parts
Lid Seal Kit, 600 liter and 1100 liter, series 40 and
F-8266-K
up
Page 111
Section 14. Replacement Parts
Page 112
Section 14. Replacement Parts
Page 113
Section 14. Replacement Parts
Page 114
Section 14. Replacement Parts
Page 115
Section 14. Replacement Parts
X-7254 Transformer
Page 116
Section 14. Replacement Parts
Page 117
Section 15: Warranty
All Q-FOG Cyclic Corrosion Chambers and components are guaranteed against defects in workmanship or
materials for one year. Liability is limited to replacing or repairing any part or parts that are defective in materials or
workmanship.
Liability in all events is limited to the purchase price paid. Damage due to accident or abuse is not covered. Labor
cost is not covered.
Q-Lab Corporation makes no other warranties, including implied warranties of merchantability or fitness for a
particular purpose, except as may be expressly provided by the Q-Lab Corporation in writing.
Q-Lab Corporation shall not be liable for any incidental, consequential, special or contingent damages arising out of
the sale or use of any product.
Page 118
Section 16: Contact Information
Repair and Tester Support is available over the telephone Monday through Friday from 8:30 AM to 5 PM. Please
contact the nearest international branch office by phone or email for technical support. You can also visit our website
at www.q-lab.com to register your tester to access additional useful troubleshooting guides, operating manuals, and
technical information.
Q-Lab Headquarters
Tel: +1-440-835-8700
Email: Repair@q-lab.com
Q-Lab China
Tel: +86-21-5879-7970
Email: RepairChina@q-lab.com
Page 119
Page 120
Q-FOG Wiring Diagram F-8164.1
Page 121
Page 122
Q-FOG Plumbing Diagram F-8165.1
Page 123
Quick Set-up Guide LF-8165-QS
Section Page
1. Purpose............................................................................ 2
2. Unpacking........................................................................ 3
3. Dimensions...................................................................... 9
4. Position............................................................................ 10
5. Leveling............................................................................ 11
6. Electrical.......................................................................... 11
7. Compressed Air................................................................ 12
8. Water Supply.................................................................... 13
9. Water Drainage................................................................ 14
10. Venting........................................................................... 14
11. Utility Connections.......................................................... 17
12. Support.......................................................................... 19
1. Purpose
This Quick Set-up Guide is not an operating manual. It is intended to explain only the basics of setting up
a Q-FOG® Controlled Relative Humidity (CRH) cyclic corrosion tester.
The Q-FOG CRH Technical Manual LF-8165-TM is necessary for a complete understanding of how to
operate the Q-FOG CRH tester.
2
Section 2: Unpacking
2. Unpacking
Tools Required
Remove Remove
1. Use pry bar to remove 2 particle board sheets from the top of the Q-FOG crate.
2. Remove the 2 particle board sheets from the sides of the Q-FOG crate.
3
Section 2: Unpacking
Metal Tab
3. Remove the metal reinforcing tabs at each end of the 10 top support boards.
Ridge Board
4
Section 2: Unpacking
Remove Remove
6. Remove the 2 particle board sheets from the ends of the Q-FOG crate.
Remove Remove
7. Remove the vertical support board from both ends of the crate.
Remove Remove
8. Remove the vertical support boards from both sides of the crate.
5
Section 2: Unpacking
Remove Remove
9. Remove the horizontal support boards from both sides and ends of the crate.
Cut
Remove
X4
10. Remove 4 corner posts. 11. Use a utility knife to cut the straps holding
the Specimen rack package to the pallet.
Remove the specimen racks.
Q-FOG
Back
Q-FOG
Front
12. Locate the 4 15 mm (9/16") bolts and nuts holding the Q-FOG to the pallet.
6
Section 2: Unpacking
Remove
X4
13. Use a 15 mm (9/16") socket to remove the 4 14. Use a fork lift to lift Q-FOG up off the pallet.
nuts and bolts.
Remove Remove
Remove
15. Make sure forks are beneath the wood 16. Use a 15 mm (9/16") socket to remove the
runners. 2 nuts and bolts in the right angle brackets
under the Q-FOG. Remove the board.
Adjust
Remove Remove Down
Remove
17. Use a15 mm (9/16") socket to remove the 2 18. Adjust the 4 leveling feet down about 38 mm
nuts and bolts in the left angle brackets under (1.5"). See Section 5 for leveling information.
the Q-FOG. Remove the board.
7
Section 2: Unpacking
Remove Remove
19. Move the Q-FOG to the installation location 20. Use a 11 mm (7/16") socket to remove the
and lower into position. 2 nuts and bolts in each runner under the
Q-FOG. Remove the runners.
Remove Plastic
Wrap
21. Use a utility knife to carefully remove the plastic wrapping. Use care to avoid scratching the Q-FOG
cabinet.
22. Unpack the air preconditioner unit. Unpacking is complete. See Section 4 for positioning
information.
8
Section 3: Dimensions
3. Dimensions
Model CRH600-HSC Model CRH1100-HSC
A 184 cm 72.3" 220 cm 86.6"
B 188 cm 74.1" 225 cm 88.4"
C 172 cm 67.5" 183 cm 72.0"
D 122 cm 48.0" 126 cm 49.8"
E 84 cm 33.0" 100 cm 39.3"
F 103 cm 40.3" 121 cm 47.3"
G 102 cm 40.0" 119 cm 47.0"
H 114 cm 44.8" 130 cm 51.1"
E
A F
G
B H
Front Side
Q-FOG CRH Test Chamber
93 cm
82 cm 36.5"
32.0" 82 cm
32.0"
101 cm
39.4"
Front Side
Q-FOG CRH Pre-Conditioner
9
Section 4: Position
4. Position
The Q-FOG CRH test chamber and air pre-conditioner should be positioned as shown in the figure
below. This will allow sufficient room to operate the unit and allow proper ventilation. The air pre-
conditioner is mounted on casters and may alternatively be positioned in front of or behind the Q-FOG
test chamber. The air pre-conditioner may be moved to open the Q-FOG test chamber right side
access panel.
25 cm WALL
10.0"
WALL
61 cm
Air
Q-FOG Test Chamber 24.0"
Pre-Conditioner
5 cm
Minimum
2.0"
The pre-conditioner is connected to the Q-FOG chamber by a 244 cm (96") flexible hose. Connect the
pre-conditioner hose through an opening in the bottom of the Q-FOG cabinet. The hose connects to
the blower module behind the Q-FOG tester right access panel.
Blower
Module
10
Section 5: Leveling
5. Leveling
The Q-FOG tester has an all plastic chamber and, therefore, it is not perfectly rigid. Because of this,
the chamber can flex a small amount. This flexing can cause the lid seal to leak if the support feet are
not properly adjusted. Be sure to follow the procedure in Section 3.1 Leveling of the Q-FOG Technical
Manual to adjust the support feet so the lid seals properly.
6. Electrical
The Q-FOG test chamber is powered by either 208V ± 10%, 1-Phase, 50/60 Hz or 230V ± 10%,
1-Phase, 50/60 Hz. This voltage requirement was specified by the purchaser at the time the chamber
was ordered. The input voltage is shown on the nameplate attached to the rear of the chamber.
It is important that the cabinet be connected to the correct voltage. If the voltage is too low, the
chamber will not heat up quickly enough. If the voltage is too high, it could cause damage to the
chamber. Maximum current loads follow:
Maximum Maximum
Current at Current at
Model 208V ± 10% 230V ± 10%
11
Section 7: Compressed Air
Installation: A qualified electrician should connect the mains according to local codes using approved
disconnect and over-current protection. On the back of the Q-FOG tester is a Main Power Circuit
Breaker and a Residual Current Device (RCD). Below this breaker is a hole for the mains power entry.
Your electrician can enlarge this hole to accept his chosen electrical conduit or power cord. The mains
wires should be connected to the breaker. Wire used for connection is required to be rated for 90oC or
higher. For dual phase systems, such as the USA, both the brown and blue wires will be live. For single
phase systems, such as Europe, the brown is live and the blue is neutral.
NOTE: M
ake sure that power cord, plugs,
receptacles or connectors are rated Mains Power
for the specified voltage and current.
Pre-Conditioner:
The pre-conditioner includes a cable that connects to the rear of the Q-FOG test chamber. The Q-FOG
test chamber supplies the power to the pre-conditioner as follows: either 208V ± 10%, 1-Phase, 50/60
Hz or 230V ± 10%, 1-Phase, 18 Amps.
Purity: The compressed air supply must be clean, dry, and oil free. The presence of solid impurities,
water, oil or other contaminants in the compressed air may affect test results. Do not add a lubricator to
your compressed air system. Oil in the air will reduce the corrosive effects of the electrolyte. Pollutants
in the air may increase the corrosive effects. Impurities in the air supply may also clog the spray nozzle,
solenoid valves or check valves and cause maintenance problems.
12
Section 8: Water Supply
●● Ordinary tap water will damage this tester and void your warranty.
●● Water purity requirements are listed below.
The Q-FOG tester requires a supply of pure water for the Bubble Tower. This is used for saturating the
compressed air with water vapor during the fog function. Q-FOG CRH tester also needs pure water for
producing chamber humidity.
Water Consumption: Purified water consumption is 5 liters per hour maximum. Maximum consumption
rates are during the relative humidity function; typical consumption will be much lower. Additionally, the
purified water system must be sized to accommodate maximum peak demand during short duration bubble
tower refill step at 0.4 liters/min.
Water Purity: The water for these processes must be distilled or deionized and, except for the salt solution
make-up water, must be connected directly to the unit. Proper water quality can be achieved by one of the
following methods: Distilled Water, Reverse Osmosis (RO), De-ionizer (DI) with filter. We recommend the
use of a deionizer to remove dissolved solids, with a filter to remove suspended solid particles. Usually the
least expensive way to get the proper water quality is to use local vendors’ water purification services.
●● Purity: less than 8 ppm Total Dissolved Solids, greater than 50 kΩ•cm (less than 20 μS/cm)
●● Remove Solid Particles: greater than 20 μm
NOTE: Use of very high purity water (greater than 5MΩ/cm or less than 0.2Ms/cm) for extended periods is
not recommended. Very high purity water may corrode stainless steel and other corrosion resistant
metals in the Q-FOG tester.
●● The water system requires a minimum pressure of 0.2 bar (3 psi) and not more than 4.0 bar (56 psi).
●● Connect 10 mm (3/8") I.D. pressure hose to the water inlet at the rear of the cabinet and the water
supply. This should be fitted with a stop valve.
●● If line pressure is greater than 4.0 bar (56 psi), install a pressure regulator.
When all three services have been connected the air and water can be turned on and checked for leaks at
the connection points.
13
Section 9: Water Drainage
Chamber Drain: The main testing chamber is fitted with an outlet to enable excess solution to go to a
sanitary sewer drain. This system consists of a bulkhead fitting, elbow, hose barb, and threaded nipple
already in place in the chamber outlet. Run the hose to a drain. The end of the hose must be below the
water level in the drain, sealed into the drain, so that corrosive vapors cannot escape. Alternatively, a trap
can be connected to the hose, so that vapors cannot escape.
Bubble Tower Drain: This system consists of a 13 mm (1/2") hose. This hose is used only when cleaning
out the Bubble Tower and does not have to be run to a drain. However, for convenience, we recommend
that this hose be connected to a drain.
Solution Reservoir Drain: The user must attach a 19 mm (3/4") hose and ball valve shut off (all supplied).
This hose is used only when cleaning out the reservoir and does not have to be run to a drain. However,
for convenience, we recommend that this hose be connected to a drain. The user may want to supply his
own plastic pipe and plumb all the drains permanently into a sanitary drain. Make sure the ball valve is
attached and closed before filling the reservoir.
Pre-Conditioner Drain: The end of the drainage hose from the bottom of the pre-conditioner should be
placed into a floor drain.
Air Filter Regulator Drain: The end of the drainage hose from the bottom of the air filter regulator should
be placed into a floor drain.
10. Venting
Be sure to comply with all local regulations regarding chemical exhaust. Usually simple salt water is not
regulated.
Proper venting will prevent corrosive mist from entering the laboratory and assure correct performance of
your Q-FOG tester. Air venting from the chamber is done via a 114 mm (4.5") O.D. vent tube. Because of
the typically corrosive nature of the exhaust, plastic is strongly recommended. The user must connect an
exhaust system with a (102 mm) 4.0" I.D. (minimum) vent line. The length of the venting system must
be less than 30 meters (100 feet) and contain fewer than 10 elbows (i.e. 90° turns). If it must be longer,
or if it must contain more turns, there could be significant back pressure on the chamber and this might
cause corrosive mist to enter the laboratory and/or reduce the tester’s performance capabilities. A “tee” is
recommended at the end of the vent to reduce the risk of periodic back-pressure from the wind. Place a
screen over the vent to discourage birds and small animals.
Recommended Venting
14
Section 10: Venting
Connections: Because some liquid from the salt fog may condense on the inside of the vent tube, we
recommend that all sections of the vent tube have a slope either back toward the chamber or away to a
drain. Avoid all installations that could create moisture traps. Otherwise, liquid may build up in the vent
tube and cause problems with back-pressure. Excessive amounts of moisture build up can create a weight
load on the vent tubing.
The figures below illustrate right and wrong ways to connect the Q-FOG chamber vent. When variations
are necessary, please consult Q-Lab Corporation for recommendations.
Correct Vent
Installation
Correct Vent
Installation
Correct Vent
Installation
Correct Installations
15
Section 10: Venting
Important: Make sure the vent is installed so that moisture cannot build up and create a trap as shown
below.
Incorrect Vent
Installation
Incorrect Vent
Installation
Incorrect Installations
16
Section 11: Utility Connections
Electrical
Power Entry
Air In 10 mm (3/8") Air Vent
hose required) 114 mm (4.5") O.D.,
102 mm (4.0") I.D. tube
must be connected to
the outdoors
Water In 10
mm (3/8") hose
required)
Bubble Tower
Drain
17
Section 11: Utility Connections
18
Section 12: Support
12. Support
Repair and technical support is available over the telephone Monday through Friday from 8:30 AM to 5 PM. Please
contact the nearest international branch office by phone or email for technical support. You can also visit our
website at www.q-lab.com to register your tester to access additional useful troubleshooting guides, operating
manuals, and technical information.
Q-Lab Headquarters Q-Lab Europe, Ltd. Q-Lab Deutschland GmbH Q-Lab China 中国代表处
Westlake, OH USA Bolton, England Saarbrücken, Germany Shanghai, China 中国上海
Tel: +1-440-835-8700 Tel: +44-1204-861616 Tel: +49-681-857470 电话: +86-21-5879-7970
repair@q-lab.com repairEurope@q-lab.com repairGermany@q-lab.com repairChina@q-lab.com
19
Quick Set-up Guide LF-8165-QS
Section Page
1. Purpose............................................................................ 2
2. Unpacking........................................................................ 3
3. Dimensions...................................................................... 9
4. Position............................................................................ 10
5. Leveling............................................................................ 11
6. Electrical.......................................................................... 11
7. Compressed Air................................................................ 12
8. Water Supply.................................................................... 13
9. Water Drainage................................................................ 14
10. Venting........................................................................... 14
11. Utility Connections.......................................................... 17
12. Support.......................................................................... 19
1. Purpose
This Quick Set-up Guide is not an operating manual. It is intended to explain only the basics of setting up
a Q-FOG® Controlled Relative Humidity (CRH) cyclic corrosion tester.
The Q-FOG CRH Technical Manual LF-8165-TM is necessary for a complete understanding of how to
operate the Q-FOG CRH tester.
2
Section 2: Unpacking
2. Unpacking
Tools Required
Remove Remove
1. Use pry bar to remove 2 particle board sheets from the top of the Q-FOG crate.
2. Remove the 2 particle board sheets from the sides of the Q-FOG crate.
3
Section 2: Unpacking
Metal Tab
3. Remove the metal reinforcing tabs at each end of the 10 top support boards.
Ridge Board
4
Section 2: Unpacking
Remove Remove
6. Remove the 2 particle board sheets from the ends of the Q-FOG crate.
Remove Remove
7. Remove the vertical support board from both ends of the crate.
Remove Remove
8. Remove the vertical support boards from both sides of the crate.
5
Section 2: Unpacking
Remove Remove
9. Remove the horizontal support boards from both sides and ends of the crate.
Cut
Remove
X4
10. Remove 4 corner posts. 11. Use a utility knife to cut the straps holding
the Specimen rack package to the pallet.
Remove the specimen racks.
Q-FOG
Back
Q-FOG
Front
12. Locate the 4 15 mm (9/16") bolts and nuts holding the Q-FOG to the pallet.
6
Section 2: Unpacking
Remove
X4
13. Use a 15 mm (9/16") socket to remove the 4 14. Use a fork lift to lift Q-FOG up off the pallet.
nuts and bolts.
Remove Remove
Remove
15. Make sure forks are beneath the wood 16. Use a 15 mm (9/16") socket to remove the
runners. 2 nuts and bolts in the right angle brackets
under the Q-FOG. Remove the board.
Adjust
Remove Remove Down
Remove
17. Use a15 mm (9/16") socket to remove the 2 18. Adjust the 4 leveling feet down about 38 mm
nuts and bolts in the left angle brackets under (1.5"). See Section 5 for leveling information.
the Q-FOG. Remove the board.
7
Section 2: Unpacking
Remove Remove
19. Move the Q-FOG to the installation location 20. Use a 11 mm (7/16") socket to remove the
and lower into position. 2 nuts and bolts in each runner under the
Q-FOG. Remove the runners.
Remove Plastic
Wrap
21. Use a utility knife to carefully remove the plastic wrapping. Use care to avoid scratching the Q-FOG
cabinet.
22. Unpack the air preconditioner unit. Unpacking is complete. See Section 4 for positioning
information.
8
Section 3: Dimensions
3. Dimensions
Model CRH600-HSC Model CRH1100-HSC
A 184 cm 72.3" 220 cm 86.6"
B 188 cm 74.1" 225 cm 88.4"
C 172 cm 67.5" 183 cm 72.0"
D 122 cm 48.0" 126 cm 49.8"
E 84 cm 33.0" 100 cm 39.3"
F 103 cm 40.3" 121 cm 47.3"
G 102 cm 40.0" 119 cm 47.0"
H 114 cm 44.8" 130 cm 51.1"
E
A F
G
B H
Front Side
Q-FOG CRH Test Chamber
93 cm
82 cm 36.5"
32.0" 82 cm
32.0"
101 cm
39.4"
Front Side
Q-FOG CRH Pre-Conditioner
9
Section 4: Position
4. Position
The Q-FOG CRH test chamber and air pre-conditioner should be positioned as shown in the figure
below. This will allow sufficient room to operate the unit and allow proper ventilation. The air pre-
conditioner is mounted on casters and may alternatively be positioned in front of or behind the Q-FOG
test chamber. The air pre-conditioner may be moved to open the Q-FOG test chamber right side
access panel.
25 cm WALL
10.0"
WALL
61 cm
Air
Q-FOG Test Chamber 24.0"
Pre-Conditioner
5 cm
Minimum
2.0"
The pre-conditioner is connected to the Q-FOG chamber by a 244 cm (96") flexible hose. Connect the
pre-conditioner hose through an opening in the bottom of the Q-FOG cabinet. The hose connects to
the blower module behind the Q-FOG tester right access panel.
Blower
Module
10
Section 5: Leveling
5. Leveling
The Q-FOG tester has an all plastic chamber and, therefore, it is not perfectly rigid. Because of this,
the chamber can flex a small amount. This flexing can cause the lid seal to leak if the support feet are
not properly adjusted. Be sure to follow the procedure in Section 3.1 Leveling of the Q-FOG Technical
Manual to adjust the support feet so the lid seals properly.
6. Electrical
The Q-FOG test chamber is powered by either 208V ± 10%, 1-Phase, 50/60 Hz or 230V ± 10%,
1-Phase, 50/60 Hz. This voltage requirement was specified by the purchaser at the time the chamber
was ordered. The input voltage is shown on the nameplate attached to the rear of the chamber.
It is important that the cabinet be connected to the correct voltage. If the voltage is too low, the
chamber will not heat up quickly enough. If the voltage is too high, it could cause damage to the
chamber. Maximum current loads follow:
Maximum Maximum
Current at Current at
Model 208V ± 10% 230V ± 10%
11
Section 7: Compressed Air
Installation: A qualified electrician should connect the mains according to local codes using approved
disconnect and over-current protection. On the back of the Q-FOG tester is a Main Power Circuit
Breaker and a Residual Current Device (RCD). Below this breaker is a hole for the mains power entry.
Your electrician can enlarge this hole to accept his chosen electrical conduit or power cord. The mains
wires should be connected to the breaker. Wire used for connection is required to be rated for 90oC or
higher. For dual phase systems, such as the USA, both the brown and blue wires will be live. For single
phase systems, such as Europe, the brown is live and the blue is neutral.
NOTE: M
ake sure that power cord, plugs,
receptacles or connectors are rated Mains Power
for the specified voltage and current.
Pre-Conditioner:
The pre-conditioner includes a cable that connects to the rear of the Q-FOG test chamber. The Q-FOG
test chamber supplies the power to the pre-conditioner as follows: either 208V ± 10%, 1-Phase, 50/60
Hz or 230V ± 10%, 1-Phase, 18 Amps.
Purity: The compressed air supply must be clean, dry, and oil free. The presence of solid impurities,
water, oil or other contaminants in the compressed air may affect test results. Do not add a lubricator to
your compressed air system. Oil in the air will reduce the corrosive effects of the electrolyte. Pollutants
in the air may increase the corrosive effects. Impurities in the air supply may also clog the spray nozzle,
solenoid valves or check valves and cause maintenance problems.
12
Section 8: Water Supply
●● Ordinary tap water will damage this tester and void your warranty.
●● Water purity requirements are listed below.
The Q-FOG tester requires a supply of pure water for the Bubble Tower. This is used for saturating the
compressed air with water vapor during the fog function. Q-FOG CRH tester also needs pure water for
producing chamber humidity.
Water Consumption: Purified water consumption is 5 liters per hour maximum. Maximum consumption
rates are during the relative humidity function; typical consumption will be much lower. Additionally, the
purified water system must be sized to accommodate maximum peak demand during short duration bubble
tower refill step at 0.4 liters/min.
Water Purity: The water for these processes must be distilled or deionized and, except for the salt solution
make-up water, must be connected directly to the unit. Proper water quality can be achieved by one of the
following methods: Distilled Water, Reverse Osmosis (RO), De-ionizer (DI) with filter. We recommend the
use of a deionizer to remove dissolved solids, with a filter to remove suspended solid particles. Usually the
least expensive way to get the proper water quality is to use local vendors’ water purification services.
●● Purity: less than 8 ppm Total Dissolved Solids, greater than 50 kΩ•cm (less than 20 μS/cm)
●● Remove Solid Particles: greater than 20 μm
NOTE: Use of very high purity water (greater than 5MΩ/cm or less than 0.2Ms/cm) for extended periods is
not recommended. Very high purity water may corrode stainless steel and other corrosion resistant
metals in the Q-FOG tester.
●● The water system requires a minimum pressure of 0.2 bar (3 psi) and not more than 4.0 bar (56 psi).
●● Connect 10 mm (3/8") I.D. pressure hose to the water inlet at the rear of the cabinet and the water
supply. This should be fitted with a stop valve.
●● If line pressure is greater than 4.0 bar (56 psi), install a pressure regulator.
When all three services have been connected the air and water can be turned on and checked for leaks at
the connection points.
13
Section 9: Water Drainage
Chamber Drain: The main testing chamber is fitted with an outlet to enable excess solution to go to a
sanitary sewer drain. This system consists of a bulkhead fitting, elbow, hose barb, and threaded nipple
already in place in the chamber outlet. Run the hose to a drain. The end of the hose must be below the
water level in the drain, sealed into the drain, so that corrosive vapors cannot escape. Alternatively, a trap
can be connected to the hose, so that vapors cannot escape.
Bubble Tower Drain: This system consists of a 13 mm (1/2") hose. This hose is used only when cleaning
out the Bubble Tower and does not have to be run to a drain. However, for convenience, we recommend
that this hose be connected to a drain.
Solution Reservoir Drain: The user must attach a 19 mm (3/4") hose and ball valve shut off (all supplied).
This hose is used only when cleaning out the reservoir and does not have to be run to a drain. However,
for convenience, we recommend that this hose be connected to a drain. The user may want to supply his
own plastic pipe and plumb all the drains permanently into a sanitary drain. Make sure the ball valve is
attached and closed before filling the reservoir.
Pre-Conditioner Drain: The end of the drainage hose from the bottom of the pre-conditioner should be
placed into a floor drain.
Air Filter Regulator Drain: The end of the drainage hose from the bottom of the air filter regulator should
be placed into a floor drain.
10. Venting
Be sure to comply with all local regulations regarding chemical exhaust. Usually simple salt water is not
regulated.
Proper venting will prevent corrosive mist from entering the laboratory and assure correct performance of
your Q-FOG tester. Air venting from the chamber is done via a 114 mm (4.5") O.D. vent tube. Because of
the typically corrosive nature of the exhaust, plastic is strongly recommended. The user must connect an
exhaust system with a (102 mm) 4.0" I.D. (minimum) vent line. The length of the venting system must
be less than 30 meters (100 feet) and contain fewer than 10 elbows (i.e. 90° turns). If it must be longer,
or if it must contain more turns, there could be significant back pressure on the chamber and this might
cause corrosive mist to enter the laboratory and/or reduce the tester’s performance capabilities. A “tee” is
recommended at the end of the vent to reduce the risk of periodic back-pressure from the wind. Place a
screen over the vent to discourage birds and small animals.
Recommended Venting
14
Section 10: Venting
Connections: Because some liquid from the salt fog may condense on the inside of the vent tube, we
recommend that all sections of the vent tube have a slope either back toward the chamber or away to a
drain. Avoid all installations that could create moisture traps. Otherwise, liquid may build up in the vent
tube and cause problems with back-pressure. Excessive amounts of moisture build up can create a weight
load on the vent tubing.
The figures below illustrate right and wrong ways to connect the Q-FOG chamber vent. When variations
are necessary, please consult Q-Lab Corporation for recommendations.
Correct Vent
Installation
Correct Vent
Installation
Correct Vent
Installation
Correct Installations
15
Section 10: Venting
Important: Make sure the vent is installed so that moisture cannot build up and create a trap as shown
below.
Incorrect Vent
Installation
Incorrect Vent
Installation
Incorrect Installations
16
Section 11: Utility Connections
Electrical
Power Entry
Air In 10 mm (3/8") Air Vent
hose required) 114 mm (4.5") O.D.,
102 mm (4.0") I.D. tube
must be connected to
the outdoors
Water In 10
mm (3/8") hose
required)
Bubble Tower
Drain
17
Section 11: Utility Connections
18
Section 12: Support
12. Support
Repair and technical support is available over the telephone Monday through Friday from 8:30 AM to 5 PM. Please
contact the nearest international branch office by phone or email for technical support. You can also visit our
website at www.q-lab.com to register your tester to access additional useful troubleshooting guides, operating
manuals, and technical information.
Q-Lab Headquarters Q-Lab Europe, Ltd. Q-Lab Deutschland GmbH Q-Lab China 中国代表处
Westlake, OH USA Bolton, England Saarbrücken, Germany Shanghai, China 中国上海
Tel: +1-440-835-8700 Tel: +44-1204-861616 Tel: +49-681-857470 电话: +86-21-5879-7970
repair@q-lab.com repairEurope@q-lab.com repairGermany@q-lab.com repairChina@q-lab.com
19