Professional Documents
Culture Documents
2778 2777 Mobile Service Manual PDF
2778 2777 Mobile Service Manual PDF
SERVICE MANUAL
Model- 2778
Mobile Combo
Therapy System
Applies to Serial numbers 1000 and above
Model- 2777
Mobile Stim
Therapy System
Applies to Serial numbers 1000 and above
Read, understand, and follow the Safety Precautions and all other information contained in this
manual.
This manual contains the necessary safety and field service information for those field service
technicians, certified by Chattanooga Group, to perform field service on the Intelect Mobile Stim or
Combo Therapy Systems.
At the time of publication, the information contained herein was current and up-to-date. However,
due to continual technological improvements and increased clinical knowledge in the field of
electrotherapy, as well as Chattanooga Group’s policy of continual improvement, Chattanooga
Group reserves the right to make periodic changes and improvements to their equipment and
documentation without any obligation on the part of Chattanooga Group.
©2005 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without express written consent from the Chattanooga Group of Encore
Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by the Chattanooga Group of Encore Medical, L.P.
1
1 SAFETY PRECAUTIONS Intelect® Mobile Stim and Combo Therapy Systems
F. NOTE:
Throughout this manual “NOTE” may be found. These Notes
are helpful information to aid in the particular area or function being
described.
2
1 SAFETY PRECAUTIONS Intelect® Mobile Stim and Combo Therapy Systems
• These devices are restricted to sale by, or on the order of, a • Stimulation should not be applied transthoracically in that
physician or licensed practitioner. This device should be used the introduction of electrical current into the heart may cause
only under the continued supervision of a physician or licensed cardiac arrhythmia.
practitioner. • Stimulation should not be applied over swollen, infected,
• For continued protection against fire hazard, replace fuses and inflamed areas or skin eruptions, (e.g., phlebitis,
only with ones of the same type and rating.
thrombophlebitis, varicose veins, etc.).
• Make certain the unit is electrically grounded by connecting • Stimulation should not be applied over, or in proximity to,
only to a grounded electrical service receptacle conforming to
the applicable national and local electrical codes. cancerous lesions.
• Care must be taken when operating this equipment around • Output current density is related to electrode size. Improper
other equipment. Potential electromagnetic or other application may result in patient injury. If any question arises
interference could occur to this or to the other equipment. Try as to the proper electrode size, consult a licensed practitioner
to minimize this interference by not using other equipment in prior to therapy.
conjunction with it. • Unplug the unit from the power source before attempting
• The safety of TENS waveforms for use during pregnancy or removal or replacement procedures to prevent electrical
birth has not been established. shock.
• TENS is not effective for pain of central origin. This includes • Use only Degassed Water in Power Meter for testing Ultrasound
headache. Applicators. Use of other types of water will cause false test
• TENS should be used only under the continued supervision of results. Refer to page 21 for Degassed Water Recipes.
a physician or licensed practitioner. • Do not aerate water when filling Power Meter.
• TENS waveforms have no curative value. • Unit failing Dielectric Withstand Test or Leakage Test could
• TENS is a symptomatic treatment, and as such, suppresses the indicate serious internal problems! Do not place unit back into
sensation of pain which would otherwise serve as a protective service! Send unit to factory for repair! Do not attempt to repair.
mechanism.
• The user must keep the device out of the reach of children.
• Electronic monitoring equipment (such as ECG monitors and
ECG alarms) may not operate properly when TENS stimulation
is in use.
• Powered muscle stimulators should be used only with
the leads and electrodes recommended for use by the
manufacturer.
• In the event that an Error message or Warning appears
beginning with a 2 or 3, immediately stop all use of the
system and contact the dealer or Chattanooga Group for
service. Errors and Warnings in these categories indicate an
internal problem with the system that must be tested by
Chattanooga Group or a Field Service Technician certified by
Chattanooga Group before any further operation or use of the
system. Use of a system that indicates an Error or Warning in
these categories may pose a risk of injury to the patient, user
or cause extensive internal damage to the system.
• Use of controls or adjustments or performance of procedures
other than those specified herein may result in hazardous
exposure to ultrasonic energy.
• Before administering any treatment to a patient you should
become acquainted with the operating procedures for
each mode of treatment available, as well as the indications,
contraindications, warnings, and precautions. Consult other
resources for additional information regarding the application
of Electrotherapy and Ultrasound.
• To prevent electrical shock, disconnect the unit from
the power source before attempting any maintenance
procedures.
• Keep electrodes separated during treatment. Electrodes in
contact with each other could result in improper stimulation
or skin burns.
• Long term effects of chronic electrical stimulation are
unknown.
• Stimulation should not be applied over the anterior neck
or mouth. Severe spasm of the laryngeal and pharyngeal
muscles may occur and the contractions may be strong
enough to close the airway or cause difficulty in breathing.
3
1 SAFETY PRECAUTIONS Intelect® Mobile Stim and Combo Therapy Systems
4
2 THEORY OF OPERATION Intelect® Mobile Stim and Combo Therapy Systems
2.1 OVERVIEW
The Intelect Mobile Therapy Systems are comprised of several PC board assemblies housed within a common
enclosure. These assemblies each support a distinct function in the product. The basic elements are User
Interface, Control Board, Stim Boards, Ultrasound Board (Combo only), Ultrasound Applicator (Combo only),
and Power Supply Circuits.
2.2 POWER SUPPLY CIRCUIT
A universal 75 Watt input power supply provides each system with the required 24 volts DC. The supply is
connected to the mains at all times when the Mains Power Cord is attached and plugged into an outlet
supplying 100 - 240 VAC. The 24 V supply is regulated locally at each PC board as required.
2.3 CONTROL BOARD
The Control Board serves just as its name implies. It controls the operation of the Stim Boards, Ultrasound
Board, User Interface, and Accessories. The Control Board communicates to the Stim Boards and Ultrasound
Board through a proprietary bus. The Control Board drives the display. The Control Board reads the menu
buttons. The Control Board also reads the amplitude and the Contrast Control on the systems. Sound output
is generated by the Control Board and routed to an internal speaker.
2.4 STIM BOARDS
The Stim Boards create all muscle stimulation output. Communication to the Stim Boards is via a
proprietary bus. A Processor on each Stim Board acts on messages passed to it by the Control Board
to set up waveforms and adjust output amplitude. Information can likewise be passed from each
Stim Board back to the Control Board for monitoring current, etc. If a Stim Board does not respond as
expected to a command from the Control Board, output is stopped and an Error Message is generated.
2.5 ULTRASOUND BOARD AND APPLICATOR COMBINATION SYSTEMS ONLY
The Ultrasound Board generates the 1 or 3.3 MHz output to drive the Sound Head of the Applicator. The
Ultrasound Board is accessed through the proprietary bus by the Control Board. It can provide current
and voltage information about the ultrasound output of the board. The calibration data for the Sound Head
is passed through the Ultrasound Board from the Applicator to the Control Board. By storing the calibration
data in the Applicator, there is no calibration necessary for the Ultrasound Board and any
calibrated Chattanooga Group Intelect Advanced or Intelect Mobile Ultrasound Applicator can be connected
and operated to provide accurate coupling and output.
2.6 USER INTERFACE AND ACCESSORIES
The LCD display panel provides the operator visible feedback in the way of menu choices. Pressing the
User Interface buttons makes selections from the menus. The Control Board interprets these user inputs and
responds accordingly. Audible feedback is given for such events as key presses and end of treatment.
2.7 NIMH BATTERY
The NiMH Battery Module incorporates a Nickel Metal Hydride (NiMH) Battery Pack and a PC Board. The
PC Board monitors the Battery Charge Level. The Battery Pack supplies the required 24 VDC to the system
which is then distributed to the respective PCB’s through the Universal Power Supply. The Battery
Pack is interfaced with the system via a Wire Harness that facilitates communication with the Control
Board and delivery of power to an Electrotherapy or Combination Therapy System. When the Therapy
System is connected to a Mains Power Supply via the Mains Power Cord, the NiMH Battery Pack will charge.
Once the Battery Pack is fully charged, the software will stop the charging process, eliminating the possibility
of overcharging. Battery power is used only when the Therapy System is not connected to a Mains Power
Supply.
5
3 NOMENCLATURE Intelect® Mobile Stim and Combo Therapy Systems
CONTRAST
CONTROL
FAN VENT
ON/OFF
POWER
SWITCH
LCD
USER INTERFACE
ULTRASOUND
APPLICATOR
PLYNTH
ELECTROTHERAPY
LEAD WIRE
RECEPTACLES
ULTRASOUND
RECEPTACLE
FIGURE 3.1
6
3 NOMENCLATURE Intelect® Mobile Stim and Combo Therapy Systems
CONTRAST
CONTROL
FAN VENT
ON/OFF LCD
POWER
SWITCH
USER INTERFACE
PLYNTH
ELECTROTHERAPY
LEAD WIRE
RECEPTACLES
FIGURE 3.2
7
3 NOMENCLATURE Intelect® Mobile Stim and Combo Therapy Systems
DATA INTENSITY
PORT
STOP UP ARROW
TREATMENT
DOWN ARROW
PAUSE
TREATMENT
START
TREATMENT
INCREASE
DECREASE
CLINICAL
RESOURCES
BACK
CHANNEL 1
LEAD WIRES
CHANNEL 2
LEAD WIRES
ULTRASOUND
APPLICATOR
CHARGE LEVEL
BATTERY
CHARGING
8
4 SPECIFICATIONS Intelect® Mobile Stim and Combo Therapy Systems
DEPTH
HEIGHT
WIDTH
FIGURE 4.1
9
4 SPECIFICATIONS Intelect® Mobile Stim and Combo Therapy Systems
DEPTH
HEIGHT
WIDTH
FIGURE 4.2
10
4 SPECIFICATIONS Intelect® Mobile Stim and Combo Therapy Systems
The specifications found in this section provide the necessary waveform NOTE:
specifications to aid in troubleshooting. A waveform graphic from an All waveforms except High Voltage Pulsed Current (HVPC) of the Intelect
oscilloscope is also provided for clarification. Mobile Therapy System have been designed with a 200 mA current limit.
Refer to this section when performing troubleshooting, replacement, and VMS™, and all TENS waveform output intensities are measured, specified, and
repair of the Intelect Mobile Stim and Combo Systems. listed to peak, not peak to peak.
All Waveforms are available on all channels.
A. IFC (Interferential) Traditional (4 Pole)- Figure 4.3
Interferential Current is a medium frequency waveform. Current is
distributed from two channels (four electrodes). The currents cross
in the body within the area being treated. The two currents interfere
with each other at this crossing point, resulting in a modulation
of the intensity (the current intensity increases and decreases at a
regular frequency).
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA
Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2,000-10,000 Hz
Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-200 Hz
Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 seconds
Sweep Low Beat Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . .1-200 Hz
Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . .1-200 Hz
Scan Percentage . . . . . . . . . . . . . . . . . . . . . . . . .Static, 40%, and 100%
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes FIGURE 4.3
11
4 SPECIFICATIONS Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 4.10
FIGURE 4.8
FIGURE 4.11
FIGURE 4.16
FIGURE 4.17
FIGURE 4.18
FIGURE 4.20
16
4 SPECIFICATIONS Intelect® Mobile Stim and Combo Therapy Systems
A. Ultrasound
Frequency. . . . . . . . . . . . . . . . . . . . . . . . 1 MHz, ± 5%; 3.3 MHz, ±5%
Duty Cycles . . . . . . . . . . . . . . . . . . . . 10%, 20%, 50%, and Continuous
Pulse Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . .16 Hz, 48 Hz, 100 Hz
Pulse Duration . . . . . . . . . . . . . . . . . . . 1 mSec, ±20%; 2 mSec, ±20%
5 mSec, ±20%
Output Power
10 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-20 Watts at 1 MHz,
0-10 Watts at 3.3 MHz
2
5 cm Crystal . . . . . . . . . . . . . . . . . . . . . . . . 0-10 Watts, 1 and 3.3 MHz
2 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . .0-4 Watts, 1 and 3.3 MHz
1 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-2 Watts 3.3 MHz Only
Amplitude . . . . . . . . . . . . . . . . . . . 0 - 2.5 w/cm2 in Continuous mode,
0-3 w/cm2 in Pulsed modes
Output accuracy . . . . . . . . . . . . . . . . ± 20% above 10% of maximum
Temporal Peak to Average Ratios:
2:1, ± 20%, at 50% Duty Cycle
5:1, ± 20%, at 20% Duty Cycle
9:1, ± 20%, at 10% Duty Cycle
Beam Nonuniformity Ratio. . . . . . . . . . 5.0 : 1 maximum
Beam Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Collimating
Effective Radiating Areas . . . . . . . . . . .10 cm2 Crystal - 8.5 cm2, ±1.5
5 cm2 Crystal - 4.0 cm2, ±1.0
2 cm2 Crystal - 1.8 cm2, +0.2/-0.4
1 cm2 Crystal - 0.8 cm2, +0.2/-0.4
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30 Minutes
B. Head Warming Feature Specifications
The Head Warming feature of an Intelect Combination Therapy
System utilizes Ultrasound output resulting in warming of the Sound
Head to increase patient comfort.
With Head Warming enabled, ultrasound is emitted without pressing
the START button. The Applicator LED will not illuminate during the
Head Warming period. US Channel will indicate "Warming".
Output .. .. .. .. .. .. .. .. .. .. .. .. .. ..0 - 50% Cycling of maximum power
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 Mhz
Do not apply the Ultrasound Applicator to the patient during the Head
Warming period. Applicator must remain in Applicator Hook during the
Head Warming period.
17
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
18
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
5.1
5.1 INTELECT
INTELECT MOBILE
MOBILE STIM AND
SYSTEM COMBO ERROR
SOFTWARE ERRORMESSAGES
MESSAGES(continued)
(continued)
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga
Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field
Service Technician certified by Chattanooga Group before any further operation or use of the unit. Use of a unit that indicates an Error or Warning in these
categories may pose a risk of injury to the patient, user, or may cause extensive internal damage to the unit.
19
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga
Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field
Service Technician certified by Chattanooga Group before any further operation or use of the unit. Use of a unit that indicates an Error or Warning in these
categories may pose a risk of injury to the patient, user, or may cause extensive internal damage to the unit.
20
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
A. General 11. Degassed Water (<5 ppm) for Ultrasound Power Meter.
1. The following information is intended to aid in troubleshooting Recipe(s) for Degassed Water
the major components of the Intelect Stim and Combo Therapy 1) Boil Distilled Water for 30 Minutes. Place water in a
Systems to “Board Level” only. These tests are FACTORY standard non-porous container and immediately cover
testing procedures and methods used at the factory before with cellophane. Allow to cool to room temperature of
shipment of any Intelect Therapy System. approximately 70 °F (21 °C). May be refrigerated to aid
2. Due to the complex nature of the technology utilized by cooling time.
Chattanooga Group, the recommended troubleshooting techniques or
are to determine “Bad Board” and board replacement only. No
board component level troubleshooting is recommended, nor will 2) Bring Distilled Water to a boil. Place the container under
information or parts be supplied by Chattanooga Group. Any board vacuum for 5 to 10 Minutes.
component level troubleshooting performed will be at sole risk and NOTE:
liability of the Service Technician performing such troubleshooting Two liter soft drink bottles are ideal storage and transport containers for
techniques. degassed water as they are designed to keep oxygen out. Do not allow aeration
3. Once a particular PC Board has been determined as bad, refer to the of degassed water during transport or filling of the power meter.
appropriate Removal and Replacement Section of this Manual for Do not use Tap Water or Distilled water in the Ultrasound Power Meter. Use
proper replacement. only Degassed Water in order to obtain correct test results. The chart below
illustrates the oxygen content of Degassed, Tap, and Distilled Water.
B. Special Tools, Fixtures, & Materials Required
1. Certain tests require the use of special tools and fixtures. These will WATER TYPE ppm of OXYGEN
be listed at the particular test where they are required. Testing with Degassed Less than 5 ppm
any other special tool or fixture other than those stated could give (per Recipe 1 or 2)
erroneous readings or test results. Always perform the tests exactly
as stated to ensure accurate results. Tap Water Up to 35 ppm
2. Any special tools or fixtures required can be obtained through the Distilled Water Up to 20 ppm
Chattanooga Group Service Department.
3. Scope and other standard test equipment settings will be listed D. Full Functional Tests
for each test performed to aid in performing the test to FACTORY Perform the tests found in this section to verify Full Functionality of
standards and ensure proper readings. new Therapy System and accessories.
4. The troubleshooting and repair of the Intelect Therapy Systems and E. Required Handtools
Accessories should be performed only by authorized technicians
trained and certified by Chattanooga Group. 1. # 1 Phillips Screwdriver
C. Equipment Required 2. Insulated Needle Nose Pliers
1. Oscilloscope and Probes 3. 1/4 in. Wrench
2. ESTI-2 Load Test Fixture
3. Digital Multimeter
4. Microcurrent Probe (Accessory)
5. Intelect Mobile or Advanced Ultrasound Applicators (Accessories)
6. Dielectric Withstand (Hi-Pot) and ground resistance
tester
NOTE:
Adjust Dielectric Withstand tester to indicate fault with 120 k Ohm
Load across the output when at specified test voltage.
7. Carbon Electrodes
8. Milliohm Meter
9. Ohmic Instruments UPM DT 10 or DT 100 Ultrasound
Power Meter
10. Dissolved Oxygen Test Kit used to test oxygen level
of degassed water.
21
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 5.5
23
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
24
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
25
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 5.10A
26
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 5.13
27
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 5.16
28
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
29
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
8. Increase Intensity until the appropriate Watts is displayed. See 10% 0.8 - 1.2
Figure 5.20. 20% 1.6 - 2.4
5 cm2
9. Compare Power Meter reading to Figure 5.20. 50% 4.0 - 6.0
10. Press the STOP button. 100% (Continuous) 8.0 - 12.0
11. Highlight Duty Cycle and press the Enter button. Highlight the FIGURE 5.20
next level of Duty Cycle and repeat steps 6 through 10. Repeat
for remaining Duty Cycle levels.
12. Highlight Frequency. Press the Enter button until 3.3 MHz is
displayed beside Frequency. Repeat steps 4 through 11.
B. Ultrasound Duty Cycle Test Results
1. Duty Cycles fall within the specified ranges as listed in
Figure 5.20.
Unit passed test.
2. Readings fall outside specified ranges of Figure 5.20.
a) Possible bad degassed water in Power Meter.
b) Possible use of Power Meter other than Ohmic Instruments
UPM DT 10 or DT 100 Ultrasound Power Meter.
c) Possible bad or out of calibration Applicator. Retest with
known good Intelect Mobile or Advanced Applicator.
d) Possible bad internal connection at Ultrasound Board.
e) Replace possible bad Ultrasound Board.
f) Replace possible bad Control Board.
30
5 TROUBLESHOOTING Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 5.22
FIGURE 5.23
31
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 6.2
POWER SUPPLY
FAN
BATTERY
FIGURE 6.3
32
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
3. Remove the Fan Baffle from the Fan Housing. See Figure 6.5. REMOVE
BAFFLE
FIGURE 6.5
33
33
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
MOUNTING TABS
FIGURE 6.7
CAPACITOR C4
FIGURE 6.8
34
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
PLIERS NOT
C. Replacing Power Supply
SHOWN
1. Replace new Power Supply in reverse order of preceding steps. FOR CLARITY
2. Reassamble system Top to Bottom. Refer to 6.1, part C.
NOTE: FIGURE 6.9
Do not over tighten the screws. Over tightening will damage the threads of
the brass inserts.
35
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 6.11
36
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 6.14
37
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
A. Tools and
Unplug the unit from the power source before
attempting removal or replacement procedures to
prevent electrical shock.
Equipment Required
1. #1 Phillips Screwdriver
2. 1/4 in Wrench
B. Ultrasound Board Removal
1. Separate Top from Bottom. Refer to 6.1, part B.
2. Remove Channel 1 Stim Board. Refer to 6.4, part B.
3. Remove Channel 2 Stim Board. Refer to 6.5, part B. FIGURE 6.15
4. Using the 1/4 in Wrench, remove the four Stand Offs securing
Ultrasound Board in place. See Figure 6.15.
5. Carefully remove the Ultrasound Board from the 40 Pin Control CONTROL BOARD HEADER
Board Header. See Figure 6.16.
6. Remove the 40 Pin Header from the Ultrasound Board. See
Figure 6.17.
C. Replacing Ultrasound Board
1. Install 40 Pin Header to back of new Ultrasound Board. Make
certain it is completely seated against board. See Figure 6.17.
2. Position the Ultrasound Board over the Control Board aligning the
40 Pin Header with the 40 Pin Connector on Control Board. See
Figure 6.16.
3. Press the Ultrasound Board into position until the board rests
against the Control Board Stand Offs.
4. Secure the Ultrasound Board using the Stand Offs removed
in part B, step 4 above.
FIGURE 6.16
5. Install the Channel 2 Stim Board. Refer to 6.5, part C.
6. Install the Channel 1 Stim Board. Refer to 6.4, part C.
7. Re-assemble Top to Bottom. Refer to 6.1, part C.
NOTE:
Do not over tighten the Stand Offs or screws. Over tightening will damage the
threads of the brass inserts and Stand Offs.
FIGURE 6.17
38
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
A. Tools and
Unplug the unit from the power source before
attempting removal or replacement procedures to
prevent electrical shock.
Equipment Required
1. #1 Phillips Screwdriver
2. 1/4 in Wrench
FIGURE 6.18
B. Control Board Assembly Removal
REMOVE SCREWS
1. Separate Top from Bottom. Refer to 6.1, part B.
2. Remove Channel 1 Stim Board. Refer to 6.4, part B.
3. Remove Channel 2 Stim Board. Refer to 6.5, part B.
4. Remove Ultrasound Board (Combo Systems Only). Refer to 6.6,
part B.
5. Remove the Contrast Control Knob. See Figure 6.18.
6. Remove the seven screws securing the Control Board Assembly in
position. See Figure 6.19.
7. Remove the four LCD Bracket Mounting Screws. See Figure 6.19.
8. While lifting on the lower area of the Control Board Assembly
with one hand, release the Alignment Tabs with the other to
remove the Control Board from the system Top. See Figure 6.20.
C. Replacing Control Board Assembly
1. Position the new Control Board Assembly over the Alignment Tabs
of the system Top. Press the Control Board Assembly until the
Alignment Tabs lock the Control Board into position. See
Figure 6.20.
3. Install the four LCD Bracket Mounting Screws. See Figure 6.19.
4. Install the seven Control Board Mounting Screws. See Figure 6.19. REMOVE SCREWS
5. Install the Ultrasound Board (Combo Systems only). Refer to 6.6, FIGURE 6.19
part C.
6. Install the Channel 2 Stim Board. Refer to 6.5, part C. RELEASE TABS
7. Install the Channel 1 Stim Board. Refer to 6.4, part C.
LIFT
8. Re-assemble Top to Bottom. Refer to 6.1, part C. HERE
NOTE:
Do not over tighten the Stand Offs or screws. Over tightening will damage the
threads of the brass inserts and Stand Offs.
FIGURE 6.20
39
6 REMOVAL & REPLACEMENT Intelect® Mobile Stim and Combo Therapy Systems
KEYMAT
ON/OFF BUTTON
Unplug the unit from the power source before
attempting removal or replacement procedures to
prevent electrical shock.
A. Tools and Equipment Required
1. #1 Phillips Screwdriver
2. 1/4 in Wrench
B. Keymat Assembly Removal
1. Separate Top from Bottom. Refer to 6.1, part B.
2. Remove Channel 1 Stim Board. Refer to 6.4, part B.
3. Remove Channel 2 Stim Board. Refer to 6.5, part B.
FIGURE 6.21
4. Remove Ultrasound Board (Combo Systems only). Refer to 6.6,
part B.
5. Remove Control Board Assembly. Refer to 6.7, part B.
6. Lift out Keymat Assembly and On/Off Button Keymat. See
Figure 6.21.
NOTE:
Do not over tighten the Stand Offs or screws. Over tightening will damage the
threads of the brass inserts and Stand Offs.
40
7 GENERAL MAINTENANCE Intelect® Mobile Stim and Combo Therapy Systems
41
8 ULTRASOUND APPLICATOR CALIBRATION Intelect® Mobile Stim and Combo Therapy Systems
FIGURE 8.1
Use only Degassed Water in Power Meter for calibrating Ultrasound
Applicators.
Use of other types of water will cause false readings and bad test results.
See page 21 for Degassed Water Recipes.
Use of other brands or types of tools, equipment, fixtures, materials, and
supplies other than those specifically listed on page 21 will give bad test
and calibration results.
If proper equipment is not available or cannot be obtained, send the
Ultrasound Applicators to the factory for calibration.
SET METER
TO WATTS
FIGURE 8.3
42
Intelect® Mobile Stim and Combo Therapy Systems
9 PARTS
Combo Assembly
COMBO TOP
ASSEMBLY
See Pages 45
10
1
COMBO BASE
ASSEMBLY
See Page 46
2
2
43
9 PARTS Intelect® Mobile Stim and Combo Therapy Systems
COMBO TOP
ASSEMBLY
See Page 45
44
9 PARTS Intelect® Mobile Stim and Combo Therapy Systems
45
9 PARTS Intelect® Mobile Stim and Combo Therapy Systems
10
9
8
2
7
5
46
9 PARTS Intelect® Mobile Stim and Combo Therapy Systems
Stim Assembly
STIM TOP
ASSEMBLY
See Pages 49
1
9
STIM BASE
ASSEMBLY
See Page 50
2 2
47
9 PARTS Intelect® Mobile Stim and Combo Therapy Systems
Stim Boards
STIM TOP
ASSEMBLY
See Page 49
48
9 PARTS Intelect® Mobile Stim and Combo Therapy Systems
49
9 PARTS Intelect® Mobile Stim and Combo Therapy Systems
10
9
7
5
50
10- Schematics Intelect® Mobile Stim and Combo Therapy Systems
SCL_APPL
SDA_APPL
MCLR
1
1
1
C2 C1 R2
R6 R1 10nF 10nF 1K
10K 1K R4
42.2K 1%
F1
2
U1
2
VCC_APPL 1 8
1 2
VDD VSS
1
2 7
1
GP5/T1CKI/OSC1/CLKIN GP0/AN0/CIN+/ICSP
DAT
3 6
GP4/AN3/T1G/OSC2/CL
KOUT GP1/AN1/CIN-/VREF/ICSPCLK
4 5 C5
C3 C6 GP3/MCLR/VPP GP2/AN2/T0CKI/INT/
COUT RT1 Next to Pin 1
0.1uF Thermal connect
0.1uF 1.0uF/16V 100K
PIC12F675-I/SN L2
2
2
2
DGND_APPL
2
NOT Case
Connected 220uH
3
LED1
R5 R7
2
1 2 2 1 1 2 1
Q4
220 10K
1
MMBT3904
R3
2
C4 10K
10nF
1
D1
BAS16_NL
1 2
1- N/C K1
3
2,4- US+ 8
3- MCLR 1
5- +5V 5 1 50PAD
6,8- US-
1
6 PAD4
7- IC2 DATA 7 1 50PAD
1
9- IC2 CLOCK PAD3
10- Applicator Body 4 1 50PAD
C8 1
3 PAD2
2 1 50PAD
1
C7 PAD1
10nF JP1
T1 G6K-2F-Y5VDC
US+ 1 8
1 2
10nF
3 4 PIN1
C9
5 6
330pF/200V 2 7 1
7 8 1
9 10
3 6
US- 4 5
1
EFD15 100PAD 100PAD
PIN2 PIN3
1
Case Connected
Transducer
VCC
VCC
R74
C6
100K
RP1 .1uf
/CS 1 16
RS 2 15 CN2
R/W 3 14 R47 33
1
NMI 4 13 RxD2
2
/ST BY 5 12 TxD2
3
MD0 6 11
4
E 7 10
VCC
/RST 8 9 CON4
10K
VCC
+24V
U1 DSP1
R12
2 VCC
V CC
59 52 /CS 15 A 18K
VCC1 P43 /CS Anode
4 53 RS 4
VCC1 P44 RS
VCC 37 54 R/W 5 Panasonic: EVUF3AF25B14
AVCC P45 R/W
36 55 E 6 3 R35 D20 R59
AVref P46 E VO
56 /RST 16 10K 1N6263 22K
R27 P47 /RST
NMI 7
NMI
/ST BY 8 79 7 LCD 240 * 128
10K /STBY P10 DB0
MD0 6 78 8 17 R13 24V_MON
MD0 P11 DB1 VEE
C2 U4 77 9
P12 DB2 1K
.01uf
2 76 10
Vdd P13 DB3 VEE
1 1 75 11 C22 R60
RST /RES P14 DB4 R52 100
3 100 74 12 .01uf 3.01K
GND RES0 P15 DB5
73 13
P16 DB6 R53
DS1233-10(3) Y1 72 14 K 100
P17 DB7 Cathode
3 1
EXTAL VSS
67 C_INH Q6
P20 C39 C40 C41 C42 C43 C44 C45 C46
C4 20M Hz C5 66 DQ LCD Displ ay 2N7002
P21
22pf 22pf 65 AC_IN 22pf 22pf 22pf 22pf 22pf 22pf 22pf 22pf Connector: Sa mtec HLE-122-02- S-DV-BE-A BACKLIGHT
P22
64 SD_REQ
P23
63 SHUTDOWN
P24
2 62 24V_EN VCC R4
XTAL P25
MD1 5 61 Vee_EN 100K
MD1 P26
TxD1 97 60 BEEPER +24V
TxD1 P27
RxD1 98 VCC
RxD1
99 82 BATT_IN J3
SCK1 P30
83 BACKLIGHT BP1
P31 R14 R15 R16 R17 R18 R19 R20 R21 1
BUS_DI R38 100 RxD0 13 84 FAN
RxD0 P32 1K 1K 1K 1K 1K 1K 1K 1K 2
BUS_DO R39 100 TxD0 14 85 /EEPROM_CS
TxD0 P33
BUS_CLK R40 100 SCK0 12 86 /BATT_EN DIP1 FAN
SCK0 P34 SW DI P-4
87 BEEPER
P35
VCC 88 1 8
P36
89 2 7
P37 Q7
3 6 LED1 LED2 LED3 LED4 FAN
ZVN4306GV Q5
26 4 5 BEEPER
P60 2N7002
R31 27
P61
10K /CS_STIM R41 100 91 28 IND_1 IND_2 IND_3 IND_4 R54
PB0 P62
/CS_US R42 100 90 29 100K R7
PB1 P63
/CS_ AUX R43 100 81 32 100K
PB2 P64
/BUS_RST 80 33
PB3 P65
69 34
PB4 P66
Q4 68 35
PB5 P67
2N7002 58
PB6
57
PB7 49 R66 1K TxD2
TxD2
50 R67 1K RxD2
48 RxD2
PA0 51
R6 47 SCK2
PA1
100K 31
PA2
30 38 24V_MON VCC
PA3 P70
21 39
PA4 P71
20 40
PA5 P72
11 41
PA6 P73
10 42
PA7 P74
43
P75
96 44 R49 R50 R51
P83 P76
95 45 10K 10K 10K
P82 P77
94
P81 VCC
93
P80
46 25 R56 10K
AVSS P90
24
P91 J1
15 23
VSS P92 VSS MCLR
70 22 1 2 V CC
VSS P93 RxD1
71 19 3 4
VSS P94 TxD1 B6
92 18 SW2 SW3 SW4 SW5 5 6
VSS P95 MD1
17 7 8
P96 VCC B7
9 16 9 10
VCC2 P97
PUB BUS
D14 R75 R76 R77
C16 H8S2144AF 10K 10K 10K
.1uf SW6 SW7 SW8
1N6263
JP1
VSS BUS_DI
D15 1 2
VSS BUS_DO
3 4
VSS BUS_CLK
SW10 SW11 5 6
VSS /CS_US
1N6263 7 8
VSS /CS_STIM
9 10
CC
11 12
VCC VCC
D16 13 14 C47 C48
VSS
15 16
+24V VSS
SW12 SW13 SW14 17 18 22pf 22pf
VSS +24V
1N6263 19 20
+24V VSS
21 22
VSS +24V
23 24
VSS
D17 25 26
VCC
27 28
CC
29 30
VSS
1N6263 31 32
VSS VSS
33 34
VSS
35 36
VSS /CS_ AUX
37 38
VSS /BUS_RST
VCC 39 40
R22 R23 R24 R25
HEADER 20X2 C49
R57
1K 1K 1K 1K
Connector: Sa mtec TLE-120-01 -G-DV-A
22pf
10K
U8
8 1 /EEPROM_CS
Control Board
Vcc /CS
C19 7 6 SCK0
.1uF /HOLD SCLK
3 5 TxD0
/WP SI
4 2 RxD0
Vss SO
1 of 2
25AA080
52
10- Schematics Intelect® Mobile Stim and Combo Therapy Systems
VCC
V CC
+ +
C1 C12 C13 C14 C17 C18
.01uf .01uf .01uf .01uf 100uf 100uf
D1
1N6263
VCC
VCC
SD_REQ
R1 V CC
100K
+24V R73 C34
VCC .1uF
1K
BATT_IN D2
1N4148
U6 R72
/BATT_EN Q3
1 4 R46 R2
IN FB FMM T551 24V_DC 24V_B ATT Q1
2 1K 10K
OUT 680 2N7002
5 R8
ON/OFF L1
3
GND 330mH 5V_B ATT
D10 SW1 1M
C37 C8 LED6 R9
LM2575T5.0(5) C9
.01uf 330uf / 35V D7 5V OK R11
330uF
15K
1N5819 1N4148 ON/OFF
10M
+
D11
C20 R26 D12
C15
100uf 10K 1N4148
.1uF
1N4148
VSS_B ATT
U2
3 5 24V_B ATT
DG
SHUTDOWN R36 2
IN
4.7K
D19 1 4
COM
1N6263 R58 D13 D18
VCC
10K R28 1N4148 5.1V IPS521
10K
Q2
2N7002
+24V
VEE
Vee-
R3
C10 D8 TP1 C21 1M
C36 C7 L2 R33
1uf D3
4.7uF .1uF
10uH 10K
.22uF 1N5819
VCC
D9
U7 B560C-13
1N5819 R48 VCC
5 1 C27 LED7 24V_DC
Vin SW 270K +
100pf C11 Vee- OK
Vee_EN 4 10uf / 35V
/SHDN
R30
D5
2 3 10K
GND NFB 1N6263
R34 VR1 R44 VCC
10K LT1617 Vee Cal 18K
D6
8
AC_IN R5
50K U5A
3 100K
R10 R29
1 B560C-13
2
15K 1K
24V_DC
LM358D
C31
4
R70 R37
.1uF
10K 2.7K
C28 C3 LED5
.01uf .01uf CN1
R45
1
7.5K 2
24V OK
75W
R79 15
C38
R78 15
.001uf
L3
24V_DC CC
10uH L4
D21
10uH
U9 B560C-13 R61
D4 8 6 560K C23 C32 C30 C33 C26 J2
Vcc DR Q8
.1uF 10uF / 35V .1uF 10uF / 35V 100uF / 35V
C_INH 7 24V_B ATT
FA/SD IRFR4105 1
CC
2
1N6263 2 1 VSS_B ATT
COMP ISEN 3
5V_B ATT
C25 R68 4
5 C_INH 100
.022uf PGND 5
4 3 DQ R69 100
Vss FB 6
D22
1N6263
4
R32
C29 .033 OHM
.1uF
53
10- Schematics Intelect® Mobile Stim and Combo Therapy Systems
FB1
HZ1206 L3
100uH
1
1 2 GND
V CC
6
J1
HEAD_C LOCK
C11 C12 KMDG-8 S-BS
4.7uF/25V 0.1uF
7
8
6
4
2
4 R156 R151
2
5 3
5 3 0 0
1
GND
R157 R152
1
2 1
S1
S2
S3
2 1
C1 GND
HEAD_ DATA
T1 0 0
1
100pF/100V US_GND
4 5 R7 R8
S1
S2
S3
R1 R160 K1 K2 100 100
47 +40 VADJ C5 22_NL L1 C9
21
21
0.1uF/50V TX2-24V 11.5uH/2A 3300pF/250V TX2-24V
1
1
1
3 6 3 2 1 1 2 3
DRAIN
1
2 7 4 4
2
C4 C98 5 5 C13 C14
R2 33uF/50V C6 C10 10nF 10nF
2
NL
47 0.1uF/50V 10 1000pF/250V 10
2
1
1 8 9 2 1 1 2 9
2
C2 8 8
2
100pF/100V GND 1 L2 1
VIN IRF
1
ENG00787 12 2.8uH/2A 12 GND
FB9
1
V CC L4
R5 100uH_NL
1
D3 0.56 TP1 1 2
1
1
2
HZ1206 C3 BAS16 3 TESTPOINT
US Ground
1
1 2
U2 0.1uF C16 U3 V CC C20
R19
3
1 5 4.7uF/25V 5 1 0.1uF_NL
2
VDD V CC VCC VDD 1.5K_NL
C18 R112
2
2N7002
0.1uF HZ1206 0
2
Q8 R22
4
12
2 GND Q4 Q7 2 1 GND 1 2
GND GND 1_3MHZ HEAD_D ETE CT
2
IRLR110 IRLR110
2
1.5K_NL
1
GND R9 GND GND
2
3 4 1 2 1 1 1 2 4 3 GND US_GND R20
IN OUT OUT IN 300K_NL R23
1
1
R13 C19 C23
TPS2819D BV 2.2 R12 TPS2819D BV 10K 0_NL
0.1uF_NL
3
3
0.1uF C22
2.2
2 2
2
R10 R11 2 1
1
V CC
1.0M 1.0M
R17
U4
7
GND GND 0.1uF GND
2
2
14.0K
3
+
12
GND 6
VRF
2
-
R18
LMH6642 M A
1.00K C21
4
2 1
-5V
1
GND 0.1uF
GND 1 2
R21
1.00K
DU TY_C YCLE
CLK_REF U1A
C17
13
12 2 1 2
1
74HC27D
VCC R27 47pF GND
1
620
2
R25 R26
U1B
2.2K 2.2K U5 3
4 6
1 16 5
2
2
5 10 620
XIN EXTLF
6 9
XOUT LOCK/IPRG
8 14 C28
VDD VDD
4 11
VSS VSS
GND 47pF
2
FS6131-01
1
0.1uF
2
11
V CC
U1C
HZ1206 C26
74HC27D
4.7uF/25V
7
GND
2
GND
Ultrasound Board
1 of 3 (AMP)
54
10- Schematics Intelect® Mobile Stim and Combo Therapy Systems
FREQ.
D13 R115
LEDRD 270
1 12 2
1_3MHZ
V CC
TP13
I_S A MPLE
TESTPOINT
GND
V_S AMPLE
2
R158 R159
V_SUPP LY
R117
1.00K 1.00K
DU T Y_C YCLE 180
4.096V
1
V_REF
C76 3
U16 0.1uF
D14
2
C96 C97
LM4040DIM3-4.1
7 2 GND 470pF 470pF
HEAD_D ETE CT RA5/AN4/SS R A0/AN0
3
2
R A1/AN1
4
R A2/AN2/ Vref-/C Vref
8 5
Vss RA3/AN3/ V ref+ /CS_US
6 GND GND
R A4/ TOCKI/C1OUT
HEAD_ DATA
V CC GND 21
RB0/INT
FB11 22
RB1
HZ1206 23
RB2
24
RB3/PGM
1
1
20 25
Vdd RB4 HEAD_C LOCK
26
RB5 Precision PWMReference
C100 C81 27 V CC
RB6/PGC I2C_ DATA
1
3.3uF/16V 0.1uF 19 28
Vss RB7/PGD
1
2
2
11 C77 C86
RC0/T1OSO/T1CKI
3pole B esse l, -3dB @ 1K Hz, -51dB@10K Hz, -111dB @100K Hz
2
GND GND 12 3.3uF/16V C82 0.1uF
RC1/T1OSI/ CCP2
9 13 2 1
2
OSC1/CLKIN RC2/ CCP1 0.1uF
14 R125
2
RC3/SCK/SCL I2C_C LOCK TP15
Y1 R119 15 GND 33K GND
RC4/SDI/S DA
5
1 2 10 16 R123
CLK_REF OSC2/CL K OUT RC5/SDO C AL
5
17 33K R127 R128 R129 GND
RC6/ TX/CK
20MHz Clock Reference
1
1
MCLR/ Vpp RC7/ R X/ DT +
1 1 2 1 2 1 2 1
+
PWM
1
C80 C79 4 4
BUS_CLK - PWR
1
18pF 18pF U18 3
-
1
PIC18F252-20/SO D17 R121 LMV7239M5 C85
2
BUS_DO
LEDRD 270 R124 R126 C83 .033uF/5% LMV321M
2
GND GND STATUS 1 1 2 2 1 33K 10K .01uF/5% R130
2
BUS_DI
20.5K
2
GND
1
1 2 2 1 GND GND
STATUS 2 2 1
2
D18 R122
LEDGN 270 GND C84
1
.068uF/5%
D16
R120 BAS16
3
10K
GND J3
2 1
2
/BUS_RST 2 1
RX
4 3
PGC TX
6 5
3
MD1 GND
8 7
PGD
10 9 VCC
D15
2 Header 2x5
V CC B ZX84C5V1 PUB
2
VIN
1
1
Ready
1
GND
VCC
R131 R134 D21
2
LEDGN 0.1uF
11
1 2
2
R141
2 2
BAS16 R133 Q17 U17A 2.2K Shu td own Ac tive High CON2_NL
2
R137
23
LEDY W 3 2N7002 3 4 1 2 2 21 1 2 1
220K DU TY_C YCLE /BUS_RST 40 39
1
2
47 38 37
74HC04 D22 GND Digital Ground
36 35 R162
11
11
1 BAS16 V CC
2
34 33 DR AIN
74HC04 Du ty Cy cle
7
32 31 0
C87 R135 U17C
2
30 29
R132 33uF/25V 330K 5 6 1 22 1
28 27
1.00K
GND_US 26 25
GND 74HC04 R142 D23
2
24V_US 24 23
270 LEDRD
2
22 21 24V_US
V CC
20 19
GND GND GND GND GND
GND_US 18 17
16 15 GND_US
14 13
12 11
GNDD_US
10 9
V CC GNDD_US
8 7
GNDD_US
6 5
U17F GNDD_US
4 3
U17D U17E GNDD_US
2 1
9 8 11 10 13 12
1
74HC04 74HC04 HEADER 20X2 GND Digital Ground
47 47 47 47
VCC
2
BUS_DI
BUS_DO
BUS_CLK
Ultrasound Board
/CS_US
2 of 3 (Ports)
55
10- Schematics Intelect® Mobile Stim and Combo Therapy Systems
TP2
TESTPOINT
1
V_SAMPLE
R32
VOU T- FULL-WAVE RECTIFIER 499 TP3
1 2 TESTPOINT
1
FB3 VCC
VCC C31 U7D R36 R37 HZ1206
0.1uF 12 LMH6644 MA 499 301 U9B
+
1
1 2 14 1 2 1 2 5 LMV824M
+
1
13 7
-
4
R29 6 C42
-
499 U7A GND C34 C41 0.1uF
1 2 3 LMH6644 MA 2 1 33nF R38 R39 0.1uF
2
+
1 1 D4 2.00K 1.47K
2
2 BAT54S R34 1 2 1 2
-
2.00K GND
GND C32 2 1
0.1uF 2 1
11
1 2 C33 GND
2
18pF_NL C35
22nF -5V SUPP LY
2
-5V GND R33 R35 TP4
R30 C49 TESTPOINT
2.00K U8
1.00K 1.00K 3.3uF/16V
1
R31 1 6
IN NO
1 2 1 2
1
VRF
VCC 50mA max.
1
1.00K 2 5 FB4 -5V
V+ COM
HZ1206
C45
FC
OSC
CAP-
0.1uF 3 4 8 5
CAP+
GND NC V+ OUT
2
2
TP5 U12 C54
CAL
MAX4644EUT TESTPOINT C53 0.1uF
GND
GND MAX660M 0.1uF
LV
1
R52
1
IOU T- FULL-WAVE RECTIFIER 499 U7C R56 R57 U9C
1 2 10 LMH6644 MA 499 301 LMV824M C51
6
+
8 1 2 1 2 10 3.3uF/16V GND
+
1
9 8
-
9 GND GND
-
C44
1 2 33nF R58 R59
1.00K 4.02K
2
3
R54 1 2 1 2
R49 2.00K
499 U7B 2 1
1 2 5 LMH6644 MA 2 1
+
7 1 D7 C43 GND
6 BAT54S 18pF_NL C46
-
4.7nF
GND
TP6
2
1
R53 R55 C47
1.00K 2.00K 1nF/100V
1
R50
1
2
1 2 1 2 D8 D9 R62 C52
IRF
12
30BQ100 150 0.1uF/100V
T3
2
1.00K 2 1 1 2
2
R60 R61 4 5
MBRS1100
200 200 C50
C48 R63 1nF/100V TP7
0.1uF/50V 1 2 1 2 TESTPOINT
1
I_SAMPLE
1
3 6
1 1
1 1
1
1
VCC D10 2 7 150
1
C65
FEEDBK
680uF/35V C67 GND
R86 R90 R91 0.1uF/50V 1 2
MBRS1100
200 200 1 8
2
DU TY_CYCLE 47
1
2
1
1
TESTPOINT R73 ENG00789 D11
2
3
1
3.3uF/16V_NL 4.7uF/25V 0.1uF C69 VCC
R66 100_NL
1
2 1 2 1 1 2 1 0.1uF
1 2
2
1
C62 2 1
2
4.99K
C57 TP10 470pF GND GND R95
2
5
1 2 C58 R76 TESTPOINT 18.0K
1nF_NL 1.0M C60 R84 GND R161
2
1
2
4
4
2
TP8 10nF_NL 10nF GND R93 1 33K
2
2
R67 +
TESTPOINT 1 2 R80 4.99K U13 4 1 2
2.2 V_SUPP LY
0- 4.1V 4.99K Q15 Q16 3
2
29.4K -
1
1
2 1 1 8 NDT3055L TP11 NDT3055L
U9D COMP VDD
GND GND GND 2 7 1 TESTPOINT 1 U14
2
1
FB OU TA
13 LMV824M 3 6 R87 LMV321M C99
2
-
CS OUTB
1
14 4 5 2.2 0.1uF
R65 RC GND
1
1 2 12 2 1
2
+
PWR C61
1
U CC3808AN-2 R88 R92 C68
4.99K R77
2 1 1 2 1.0M 1.0M 0.1uF GND
R68 2 1 2 1 R96 GND
2.21K
1M_NL 560pF 2.00K
R71 R74
1 2 1 2 1 2 GND
2
649 6.81K
2
2
J6
11
1 3
1 3
+
1
1
2 1 2 1 LMV824M
2
V_REF
1
1
C70 R99 R109 R110
2
24V_US
4.096V
1
NL 1.0M VCC
GND R107 220K 9.09K C75
2
20 - 25.5 VDC
1
C37
2
GND_US
GND GND GND 0.1uF R108
2
1 2 GND
2
GND
1
GND 2.2K
C73
100pF
Ultrasound Board
2
GND GND
Analog G
round
Ground Links
3 of 3 (V Supply)
56
10- Schematics Intelect® Mobile Stim and Combo Therapy Systems
Chattanooga Group ("Company") warrants that the Intelect Mobile Combo and Stim Therapy Systems ("Products") are free of defects in material and workmanship.
This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If these Products fail to function during the two year
warranty period due to a defect in material or workmanship, Company or the selling dealer will repair or replace the respective Product without charge within a
period of thirty days from the date on which the Product is returned to the Company or the dealer.
All repairs to the Product must be performed by a service center authorized by the Company. Any modifications or repairs performed by unauthorized centers or
groups will void this warranty.
The warranty period for certain accessories is 180 days. Accessories consist of Lead Wires, Electrodes, and Nylatex®.
The warranty period for the Battery and Ultrasound Applicators is one year (12 Months).
This Warranty Does Not Cover:
• Replacement parts or labor furnished by anyone other than the Company, the selling dealer, or a certified Company service technician.
• Defects or damage caused by labor furnished by someone other than Company, the selling dealer, or a certified Company service technician.
• Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and required maintenance or any
use that is inconsistent with the Product User's Manual.
COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
Some locations do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
To obtain service from Company or the selling dealer under this warranty:
1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the Company should be sent to:
Chattanooga Group
4717 Adams Road
Hixson, TN 37343 USA
Phone: +1-423-870-7200
FAX: +1-423-870-2046
and
2. The Product must be returned to the Company or the selling dealer by the owner. A Return Authorization (RA) Number must be obtained before returning
any product to the Company.
This warranty gives you specific legal rights and you may also have other rights which vary from location to location.
The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the sale of the Product.
Any representative or agreement not contained in the warranty shall be void and of no effect.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
68
4717 Adams Road
P.O. Box 489
Hixson, TN 37343 U.S.A.
+1-423-870-7200 OUTSIDE U.S.A.
+1 423-870-2046 OUTSIDE U.S.A. FAX
www.chattgroup.com
0413