Professional Documents
Culture Documents
Corrosion Science
journal homepage: www.elsevier.com/locate/corsci
Short Communication
a r t i c l e i n f o a b s t r a c t
Article history: In this work, a test rig was developed to study the effect of cathodic protection (CP) on corrosion of X70
Received 25 February 2009 pipeline steel in the crevice area under disbonded coating through the measurements of local potential,
Accepted 19 May 2009 solution pH and dissolved oxygen concentration. Results demonstrated that, in the early stage of corro-
Available online 23 May 2009
sion of steel, CP cannot reach the crevice bottom to protect steel from corrosion due to the geometrical
limitation. Corrosion of steel occurs preferentially inside crevice due to a separation of anodic and catho-
Keywords: dic reaction with the depletion of dissolved oxygen in the crevice solution. The main role of CP in miti-
A. Steel
gation of sequential corrosion of steel in crevice under disbonded coating is to enhance the local solution
C. Cathodic protection
C. Crevice corrosion
alkalinity. With the increase of distance from the open holiday, a high cathodic polarization is required to
C. Oxygen reduction achieve appropriate CP level at crevice bottom. A potential difference always exists between the open
C. Polymer coating holiday area and inside crevice, reducing the CP effectiveness.
Ó 2009 Elsevier Ltd. All rights reserved.
0010-938X/$ - see front matter Ó 2009 Elsevier Ltd. All rights reserved.
doi:10.1016/j.corsci.2009.05.027
X. Chen et al. / Corrosion Science 51 (2009) 2242–2245 2243
a -600
-700
-800
E (mV/SCE)
-900
-1000
-1100
-1200 OCP
-775 mV
-1300 -1000 mV
-1200 mV
-1400
0 5 10 15 20 25
Fig. 1. Test rig for simulation of a crevice under disbonded coating. Distance (cm)
14
b OCP
2. Experimental 13 -775 mV
-1000 mV
12
Test specimen was cut from a X70 steel pipe, with the chemical -1200 mV
composition (wt.%): C 0.045, Si 0.24, Mn 1.48, S 0.0001, Cr 0.031, P 11
0.017, Ni 0.16, Mo 0.23, Cu 0.21 and Nb 0.033. A three-electrode
10
cell system was used for simulation of a crevice created by coating
disbondment, as shown in Fig. 1. The cell consisted of an X70 steel 9
pH
working electrode with 33.5 cm 7.5 cm 1 cm in dimension, 8
which was ground sequentially to #1000 grit emery paper. A satu-
rated calomel electrode (SCE) was used as reference electrode, and 7
a platinum wire (1 cm in diameter) as counter electrode. A rectan- 6
gle thermoplastic shim (0.9 mm in thickness) was used to create
the crevice, with a Lucite as a coating membrane. The plastic shim 5
was adjustable to control the crevice width. A hole of 3.5 cm 4
1 cm was opened to simulate a holiday. Three ports were then 0 5 10 15 20 25
drilled at an interval of 5 cm from the holiday along the Lucite Distance (cm)
block to position the microelectrodes for measurements of local
potential, solution pH and oxygen concentration, respectively. In c 9
OCP
particular, the local potentials of steel at individual ports were
8 -755 mV
measured by microelectrodes when various CP potentials were ap-
-1000 mV
plied at the holiday. The local solution pH was measured ex-situ, 7 -1200 mV
using a capillary to sample the local solution in a beaker, through
a Thermo Orion 9863BN pH meter. The local oxygen concentration 6
was measured in situ through a Unisense OX100 dissolved oxygen
[O2] (mg/L)
5
micro-glass electrode with a 5 micron tip. The oxygen microelec-
trodes were sealed tightly with Lucite using a 732 Multi-purpose 4
Sealant (Dow Corning) in order to avoid diffusion of oxygen
through the port openings. 3
The test solution was 0.01 M Na2SO4 solution, with a pH of 6.8.
2
The solution was made of analytical agent and distilled water.
All tests were conducted at room temperature (22 °C), and 1
open to air.
0
0 5 10 15 20 25
Distance (cm)
3. Results
Fig. 2. Potential (a), pH (b) and oxygen concentration (c) profiles within crevice
Fig. 2 shows the local potential, solution pH and the oxygen after 0.5 h of test.
concentration as a function of distance from the open holiday
after 0.5 h of immersion in the solution at corrosion potential
(680 mVSCE), 775 mVSCE, 1000 mVSCE and 1200 mVSCE, day, and there was the lowest solution pH at the crevice bottom
respectively. The concentration of dissolved oxygen in the bulk (Fig. 2b). Furthermore, the oxygen concentration dropped from
solution was 8.3 mg/L. It is seen that the applied cathodic poten- 8.3 mg/L at holiday to 1 mg/L at crevice bottom no matter if CP
tial could not penetrate into the crevice bottom. With the increase was applied.
of distance from holiday, the local potential became positive Fig. 3 shows the local potential, solution pH and oxygen concen-
(Fig. 2b). Even at 1200 mVSCE, the potential at 10 cm point was tration after 72 h of application of CP potentials. It is seen that
approximately corrosion potential. Moreover, local solution pH there was a similar local potential distribution to that shown in
decreased (more acidic) with the increase of distance from holi- Fig. 2a, i.e., local potential became less negative towards the cre-
2244 X. Chen et al. / Corrosion Science 51 (2009) 2242–2245
a -600 4. Discussion
-700
The anodic and cathodic reactions during corrosion of pipeline
steel in aerated Na2SO4 solution are oxidation of iron and the
reduction of oxygen, respectively:
-800
Fe ! Fe2þ þ 2e ð1Þ
E (mV/SCE)
Fig. 3. Potential (a), pH (b) and oxygen concentration (c) profiles within the crevice
5. Conclusions
after 72 h of test.
main role of CP in mitigation of corrosion of steel in crevice area un- [2] N. Sridhar, D.S. Dunn, M. Seth, Corrosion 57 (2001) 598.
[3] D.T. Chin, G.M. Sabde, Corrosion 5 (1999) 229.
der disbonded coating is to enhance the local solution alkalinity.
[4] M.C. Yan, J.Q. Wang, E.H. Han, Corrosion Science 50 (2008) 1331.
With the increase of distance from the open holiday, a high [5] T.R. Jack, G. van Boven, M. Wilmott, Materials Performance 33 (1994)
cathodic polarization is required to achieve appropriate CP level 17.
at crevice bottom. A potential difference always exists between [6] J.J. Perdomo, M.E. Chabica, I. Song, Corrosion Science 43 (2001) 515.
[7] Z.F. Li, F.X. Gan, X.H. Mao, Corrosion Science 44 (2002) 689.
the open holiday area and inside crevice, reducing the CP effective- [8] G.W. Walter, Corrosion Science 32 (1991) 1059.
ness. The oxygen concentration drops significantly inside crevice [9] A.Q. Fu, Y.F. Cheng, Corrosion Science 51 (2009) 914.
no matter if CP is applied. [10] W. Furbeth, M. Stratmann, Corrosion Science 43 (2001) 207.
[11] R. Brousseau, S. Qian, Corrosion 50 (1994) 907.
[12] B.W. Cherry, A.N. Gould, Material Performance 29 (1990) 22.
Acknowledgements [13] R.R. Fessler, A.J. Markworth, R.N. Parkins, Corrosion 39 (1983) 20.
[14] F.M. Song, N. Sridhar, Corrosion Science 50 (2008) 70.
[15] F.M. Song, D.W. Kirk, J.W. Graydon, D.E. Cormack, Corrosion 58 (2002)
This work was supported by Chinese National R&D Infrastructure
1015.
and Facility Development Program (Project # 2005DKA10400) and [16] F.M. Song, D.A. Jones, D.W. Kirk, Corrosion 61 (2005) 145.
Canada Research Chairs Program. [17] J.J. Perdomo, I. Song, Corrosion Science 42 (2000) 1389.
[18] C.F. Dong, A.Q. Fu, X.G. Li, Y.F. Cheng, Electrochimica Acta 54 (2008) 628.
[19] A. Leng, H. Steckel, M. Stratmann, Corrosion Science 41 (1999) 547.
References [20] A. Leng, H. Steckel, M. Stratmann, Corrosion Science 41 (1999) 579.
[21] A. Leng, H. Steckel, M. Stratmann, Corrosion Science 41 (1999) 599.
[1] National Energy Board, Report of Public Inquiry Concerning Stress Corrosion [22] Zhengfeng Li, Fuxing Gan, Xuhui Mao, Corrosion Science, 44 (2002) 689.
Cracking on Canadian Oil and Gas Pipelines, MH-2-95, 1996. [23] D.T. Chin, G.M. Sabde, Corrosion 55 (1999) 229.