TRIPLEX STACK MOUNTED

"Q" SERIES MEDICAL AIR SYSTEMS


7.5 THROUGH 20 HP
The Amico Alert-2 continuous duty Medical Air system
is a completely packaged NFPA 99 and NEC compliant
assembly featuring 100% oil-less air compressors, a U.L.
listed electrical control cabinet, an ASME receiver and the
necessary accessories required to meet and exceed the
current code requirements. All components are pre-piped
and pre-wired to single-point service connections. The
only field connections are air intake, air discharge and
power connection at the control panel. All interconnect-
ing copper/brass piping as well as wiring is completed and
operationally tested prior to shipment. Liquid tight con-
duit, fittings and junction boxes are provided for all
control and power wiring.
The entire system including the receiver is stack mounted
on a common structural steel base.

The medical air compressors are of the oil-less reciprocat-


ing air-cooled design. The air compressors feature a cast
iron crankcase, teflon-composite compression and rider
rings, lubricated and totally sealed bearings, stainless steel for compressor thermal malfunction and "Backup in use".
reed valves, non-asbestos gaskets. No oil is used in The alarms include indicating lights as well as the horn. In
operation of the compressors, so the discharge air is addition, the thermal malfunction shut-down is provided
100% oil-free. Cooling is provided by a fan type com- with a manual reset. The audible alarm can be acknowl-
pressor pulley with air shroud. edged with the "Silence" button. The visual alarm will
Compressors are V-belt driven by 3 phase, 60 cycle, 1750 remain energized until the problem has been corrected.
RPM, NEMA design B motors. Each alarm function includes a set of dry contacts for
connection to the master alarm. All control and alarm
The system includes a pressure storage tank of ASME functions will remain energized while any compressor in
construction rated for 200 PSI MWP service. The tank is the system remains electrically on-line.
equipped with a pressure gauge, safety relief valve, 3 way Field adjustable control switches are pre-set to operate the
by-pass, gauge glass and automatic tank drain with man- lead compressor between 80 PSIG and 100 PSIG, the lag
ual override. compressor between 75 PSIG and 100 PSIG. The stand-
Also included as standard equipment are compressor by compressor will automatically start at 70 PSIG if one
discharge check valves of bronze construction, safety of the other compressors fails to operate.
relief valves, bronze intake and discharge flexible con- The Medical Air system is guaranteed by the manufac-
nectors, solenoid unloaders for reliable no-load starting, turer for a period of 12 months from the date of start-up or
isolation valves, high discharge temperature shut-down 18 months from the date of shipment (whichever comes
switches, pressure control switches as well as copper first) against defects in design, materials, or construction.
tubing with shut-off cock for gauge and switches.

The system includes a UL labeled control panel in a


NEMA 12 enclosure. The panel includes the following Optional System Accessories
standard accessories for each pump: externally operable (only checked options will be supplied)
circuit breaker with a door interlock, control circuit trans- "Dura-coat" rust protection receiver lining
former with fused primary and secondary coils, H-O-A
Galvanized receiver
switch, run light, hour meter and magnetic starter with 3 leg
Air-cooled aftercoolers
overload protection and reset switch. The panel is
Dual medical air treatment system
equipped with a multiple position selector switch for selec-
with refrigerated air dryers
tion of normal operation (automatic alternation) or manual
Dual medical air treatment system
selection of lead and lag pumps if one of the pumps is taken
with desiccant air dryers
out of service due to scheduled maintenance.
Interface to AIMS (AMICO Information
Local audible and visual alarms are provided per NFPA 99
Management System)
TRIPLEX STACK MOUNTED
"Q" SERIES MEDICAL AIR SYSTEMS
7.5 THROUGH 20 HP
LAYOUT AND PERFORMANCE TABLE

DISCHARGE

INTAKE
(TYP.)

RECEIVER
CONTROL
PANEL
AIR COMPRESSOR

L W

System Horsepower Capacity SCFM (Each Pump) Disch. Tank Dimensions, In. Weight

Model Number Each Total 50 PSIG 100 PSIG Conn. (Gal.) L W H Lbs.

3TOQ7.5S200 7.5 22.5 32.5 28.5 1.5" 200 118 70 94 3645

3TOQ10S200 10 30 40.8 36.3 1.5" 200 118 70 94 3765

3TOQ15S200 15 45 58.3 51.0 2" 200 118 70 94 4440

3TOQ20S240 20 60 75.0 65.1 2" 240 118 70 102 4730

Power Requirements:
(Three) _____ HP Motors, 3 Phase 60 Hertz 208 v 230 v 460 v

Phone 1-800-962-5385, Fax 1-973-227-9223, Web Site: http://www.amico.com January-2005