You are on page 1of 3

Challenges for the design of an endoscopically implanted

electrostimulator

Lonys L1, Hiernaux M1, Cauche N1, Devière J1, Vanhoest A2, Mathys P1, Nonclercq A1

1
Université Libre de Bruxelles, Brussels, Belgium
2
University College London, London, United Kingdom

Abstract
Gastrointestinal stimulator implants have recently shown positive results in helping obese patients to achieve healthier
nutrition. However, to place the implant, the patient currently needs to go through a surgical procedure. Endoscopic
implantation has the benefit over laparoscopies and bariatric surgeries to be non-invasive. It could be used to place the
gastric stimulator in the stomach. However, the implant then needs to go through the oesophagus, which implies new
constrains on the size and the weight of the device. This paper presents a project that aims to design such an implant
and assess the electrical and mechanical challenges involved in such a design.
Keywords: gastric electrical stimulation, endoscopic implantation, obesity.

Introduction receiver or a battery providing the power and a


stimulator. The housing is placed in a
Obesity has reached epidemic proportions globally,
subcutaneous pocket. To place the implant, the
with at least 2.6 million people dying each year as
patient needs to undergo a surgical procedure (such
a result of being overweight or obese. Today, one as laparoscopy), which is expensive, requires a
billion adults are overweight (BMI above 25) and
long recovery time and presents risks for the
more than 300 million are obese (BMI above 30).
patient. In particular, to place the electrodes in the
In almost half of OECD countries, 1 in 2 people is
stomach wall, the surgeon needs to go through a
now overweight or obese. If recent trends continue,
thick layer of fat, which is risky for the patient.
projections suggest that more than 2 out of 3
Implanting the device by endoscopy would avoid
people will be overweight or obese in some OECD
such a surgical procedure.
countries within the next 10 years (Fig. 1) [1].
This paper presents a project that aims to design
such an implant and assess the electrical and
mechanical challenges involved in such a design.
Main challenges of the project
An endoscopically implanted gastric stimulator has
to fulfil various requirements. First, due to the
passage through the oesophagus, the device cannot
exceed a diameter of 18mm and a maximum
volume of 5 cm3. The large housings used for
current implants are therefore unsuitable. Second,
the device needs to have adequate anchoring in the
  stomach to avoid migration or detachment. This
Fig. 1: Past and projected future overweight rates in requirement is particularly delicate as the layer of
selected OECD countries [1] the stomach used to attach the implant may be thin
and as the implant needs to withstand stomach
contractions. Third, the device must resist highly
Gastric electrostimulation is a technique used to acidic environment (the pH in the stomach can be
fight obesity. Gastrointestinal stimulator implants as low as 1), which imposes considerable
have recently shown positive results in helping limitations on the packaging. Fourth, the device
obese patients to achieve better nutrition and are needs to be easily placed using endoscopic tools.
being clinically tested [2] [3]. Current implants For instance, the electrode should be embedded in
consist of an implanted housing composed of a the implant as in situ connections and sutures are
difficult to perform by endoscopy. Fifth, the frequency switching [4] to allow using smaller
system’s power supply should not be cumbersome blocking capacitors while insuring patient
neither require too regular endoscopies to recharge protection.
the batteries.
The design of the anchoring system
Work achieved
Natural contractions and the thin layers (a few
The whole system is currently under development. millimetres) of the stomach make the anchoring
On constate cependant que les nouvelles versions de ce stimulateur offrent des
The stimulation
paramètres protocol, the output stage design,
de plus en plus réglables. difficult. The device may drift over time because
the anchoring
La tension system
initiale de la batterie de 3,7 and the power
V, sa capacité de 2,7 Ah et supply
la longévité have
annoncée de 5 à 10 années selon le protocole de stimulation utilisé permettent
of quick cicatrization and renewal of the tissues.
been assessed
d’estimer and nécessaire
la puissance moyenne are detailed
: in this section.
)LJXUH Various issues have been previously raised such as
3, 7 * 2, 7 * 3600
= 228 µW
([HPSOHVGHFRLO migration and detachment [5], difficult removal [6]
Choice of the stimulation protocol
5 * 365 * 24 * 3600  and limited degrees of freedom in endoscopy [7].
Il convient de spécifier à ce stade que les calculs sont effectués dans l’unique but
The protocol
d’estimer la consommationwill moyennebe et de based
décider si leson the
systèmes Implanted
d’alimentation
envisagés par la suite, plus particulièrement celui inspiré de la montre, rencontrent les The pyloric sphincter has been chosen to anchor
Gastric Stimulation (IGS) from Medtronic, which
exigences énergétiques des différents types d’implants étudiés.
the device as it is quite thick (5 to 7 mm), hard
already showed good results when implanted near
2.3.1.2 Le stimulateur Endoges
Une étude préliminaire ([HIERNAUX]) a déterminé le protocole de stimulation décrit (muscle layer) and easily located. Besides,
the pylorus [2] [3] (note that the published results
ci-dessous et supposé être utilisé dans le stimulateur électrique Endoges :
stimulation in the pylorus with long pulses already
were achieved with a subcutaneous implant
proved its effectiveness to treat morbid obesity in
connected to electrodes placed next to the pylorus)
dogs [8]. Furthermore, the most common used
and requires less energy than most current
position for implantable gastric stimulation is
stimulators. Stimulation will only occur during
2s 3s

located just 2cm away from the pylorus.



mealtimes (3 toFigure 4 2-4
hours per day).
Train d'impulsions Endoges
)LJXUH
The chosen
([HPSOHVGHWXEHPLFURGpFRXSp 
 design, to fix the implant on the
TheChaqueIGS protocol
apparaît durant(Fig.
2 secondes2) is a pendant
subthreshold and
impulsion et s’arrête 3 secondes. Le
 heures de pyloric sphincter, relies on using a type of
unconditional stimulation. It consists in stimulating
phénomène se répète 3 à 4 heures par jour, vraisemblablement durant les
repas.
“piercing” (Fig. 3) to make the anchoring
8QSUHPLHUSURWRW\SHLQHUWHDpWpGpYHORSSpHWHVWSUpVHQWpjOD)LJXUH easier.
&HOXLFL

periodically with en réalité
Chaque impulsion se compose a train of pulses
d’un train d’impulsions
HVW withoutGH GHX[ GLVTXHV HQ WLWDQH SOHLQV UHOLpV SDU XQ FRLO 3OXVLHXUV GH FHV
FRQVWLWXp
que l’on peut
observer sur la figure ci-dessous qui en offre une vue détaillée : To place the implant, the pylorus is stretched and
generating Gastric Slow Waves (GSW).SURWRW\SHV VHURQW LPSODQWpV FKH] O¶DQLPDO ORUV G¶XQ HVVDL FOLQLTXH VXU DQLPDO SRXU
the device is inserted through the muscle. The
HVWURXYHUGDQVOHEUHYHWHQDQQH[H
YpULILHUODTXDOLWpGHO¶DQFUDJHDLQVLUpDOLVp
 YDULDQWHV G¶LPSODQWDWLRQ RQW pJDOHPHQW pWp

 sharp tip pierces the sphincter and allows the


Le train d’impulsions se construit de la façon suivante : les impulsions apparaissent à
abutting member to be clipped. The pylorus can
(6
une fréquence de 40 Hz et possèdent une durée de largeur 300 !s comme détaillé sur
le schéma ci-dessous. then be released. 
 
2s 3s

300 !s Figure 2-5 Train d'impulsions détaillé Endoges

4
25 ms
 
Figure 2-6 Fréquence des impulsions au sein du train
Fig. 2: Stimulation protocol from Enterra by Medtronic
(above, two seconds train of pulses provided every five
Pour évaluer la consommation de l’implant Endoges, nous allons nous baser sur les Fig. 3: Inert implant (left) and implant in the pylorus
seconds
informations and below,
fournies a
partrain
[LIU].of
Le pulses composed of 300µs

précises neurostimulateur présenté dans cette 
publication se vante de ne consommer que 200 !W. Si l’on se réfère )LJXUH 
aux données de (right) WXH WUDQVSHUFH OH S\ORUH /H FRQWUHIRUW HVW
pulses occurring every 25ms)
l’article, pour un courant de stimulation de 1 mA, la consommation durant la phase
HUDWUDQVSHUFpH/¶LPSODQWHVWHQVXLWHSRXVVp
3URWRW\SHPpFDQLTXHLQHUWH
cathodique (phase d’impulsion positive) s’élève à 9,18 mW. On retrouve en effet cette
SUpVHQWp D GHYDQWOHS\ORUHOHTXHOHVWpWLUp
valeur en se basant sur le protocole [LIU] décrit ci-dessous :  OH S\ORUH FDU XQ GHV FRQWUHIRUWV Q¶HVW SDV

A flexible body has been chosen to reduce the
RQW UHSULVHV j OD )LJXUH  /¶LPSODQW HVW
Electrodes are located near
50 ms the pylorus KWWSZZZPHGLFRLOFRPSURGXFWH[DPSOHVKWPO
to stimulate
at 40Hz, a stress endured by the device during stomach
in the opposite direction to the GSW, KWWSZZZDJRUDEYQOZHESURGXFWVVHUYLFHVODVHUDSSOLFDWLRQVODVHUFXWWLQJKWPO
FRQWUDLQWHV

movements (stronger contractions occur at the


XIILVDPPHQW IOH[LEOH TXH SRXU VXLYUH OHV
frequency superior to 10the
ms natural frequency of the
 GDQV XQH pSDLVVH FRXFKH PXVFXODLUH TX¶RQ
pyloric sphincter)
 and to ensure a good contact of
GSW (3 waves per minute). The stimulation
the electrodes. This idea has been patented by our
pGDQVOHVSKLQFWHUS\ORULTXH
induces a decrease of the peristaltism and a feeling
of1 mssatiety.
1 ms
group (Priority Date: 2009/05/08; Application 

number: PCT/EP2010/056378).
Output stage design
Figure 2-7 Protocole de stimulation Liu
Choice of the power supply strategy
On remarque ainsi une phase cathodique de 1 ms, une pause de 1 ms ainsi qu’une
The output
phase anodique stage
de décharge consists in a current source
de 10 ms.
delivering The power supply is to be integrated inside one of
En supposant une consommation de 10 !W to
10mA pulses the
durant patient.
la phase One
anodique et of dethe
les phases
the abutting members of the implant to limit the
major challenges is the miniaturization of
repos, on obtient bien, en pondérant les consommations sur une période de 50 ms, les
the
200 !W annoncés :
size of the whole device. Its weight is therefore
circuit. Usually,9,18blocking capacitors are used to
protect the patient against
+ 0, 01* 49
50
= 0,193mW
DC current during single limited to 1g and volume to 360 mm3. The power
fault failure. However, the size of the blocking consumption needed to achieve the stimulation
protocol proposed is about 200µW. Various
capacitor is considerable so a promising alternative
was chosen instead. It consists in using high options for the power supply of the implant are
5 being considered.
HIRUWV GH FRPPXQLTXHU GHV VLJQDX[ RX GH
DQW G¶pOHFWURGHV HW SRVVqGH HQ VRQ VHLQ GHV
OHWLVVXS\ORULTXH/DWLJHIOH[LEOHSUpVHQWHj
Batteries are attractive as they do not need external need outpatient operation, without risking bariatric
circuitry. However, they would require regular surgery, to benefit from this cure.
endoscopies to be replaced or reloaded and the
energy available is limited by the size of the
Acknowledgment
implant. The authors thank Prof. Nick Donaldson from the
University College London for his comments and
Transcutaneous inductive powering requires suggestions on this paper.
achieving an inductive link across a large layer of
fat (around 10 centimetres). Furthermore, an References
external powering system is required which may [1] OECD, Obesity and the economics of prevention: fit
be cumbersome and impede the comfort of the not fat, www.oecd.org, 2010.
patient. [2] V Cigaina. Long-term follow-up of gastric
stimulation for obesity: the mestre 8-year
Human energy harvesting avoids both regular experience. Obesity surgery, 2004.
endoscopy and the need for an external powering
system. An energy conversion system based on an [3] S A Shikora. “What are the Yanks doing” the U.S.
electromechanical transducer has been proposed by experience with implantable gastric stimulation fort
the treatment of obesity – Update on the ongoing
our group [9]. It showed that, considering the low
clinical trials. Obesity Surgery, 2004.
level of physical activity of an obese patient, only
about 100nW could be harvested in this way. [4] X Liu, A Demosthenous, N Donaldson. An
Further studies are ongoing to improve the energy Integrated Implantable Stimulator That Is Fail-Safe
harvesting method, for instance relying on Without Off-Chip Blocking-Capacitors. Biomedical
Circuits and Systems, IEEE Transactions on
breathing movements.
biomedical circuits and systems 2.3: 231-244, 2008.
Further work [5] O Colliou, K Nason, H TenHoff, et al. Patent US
Our first aim is to complete the development of a 7,020,531. Gastric Device and Suction Assisted
Method for Implanting a Device on a Stomach Wall.
working prototype. Three major steps are needed
Filed April 2, 2002. Issued March 28, 2006.
to achieve this goal.
[6] M A Imran. Patent WO 2005/041749 A2.
First, the anchoring of an inert implant will be Gastrointestinal Stimulation Device. Issued October,
tested in three dogs during three months. This 2003.
implant has been realized with two titanium disks
linked by a flexible body. [7] T P Herbert, W L Starkebaum. Patent US
2006/0265021 A1. Intra-Luminal Device for
Second, a prototype to stimulate the stomach at the Gastrointestinal Electrical Stimulation. Filed
pylorus will be developed. It will first be tested on November 23, 2006.
a dead animal stomach and should assess the [8] X Xu, H Zhu, J D Z Chen. Pyloric Electrical
functionality of the stimulation. Stimulation Reduces Food Intake by Inhibiting
Gastric Motility in Dogs. Gastroenterology.128.1:
Third, the prototype will be tested in an animal
43-50, 2005.
model while monitoring food intake and energy
expenditure, as well as animal weight. This will [9] L Lonys, P Mathys, A Noncercq. Human energy
assess the physiological impact of the defined harvesting used for endoscopic implant power
stimulation protocol at the pylorus. Impacts on the supply. Proceedings of the XXIIIth Congress of the
duodenum will also be observed. ISB, Brussels, Belgium, September 2011.

Further work will include clinical trials and Author’s Address


reaching regulatory approval (CE Mark, etc). If the Laurent Lonys
conclusion is favourable, the method will be BEAMS/LIST – Engineering Faculty of the
transferred to a company. Université Libre de Bruxelles - CP 165/56, Av FD
At the end of this project, we hope to propose a Roosevelt 50, B-1050
functional gastric implant that will treat obesity llonys@ulb.ac.be
noninvasively. Obese patients would then only http://list.ulb.ac.be/bistimo/

You might also like