You are on page 1of 1

Budget :- 4.5 Lac.

Technical Specifications 100mA High Frequency X-Ray Machine

Generator Type : High Frequency, DSP Controlled, Constant Potential (DC)


High Voltage Range : 40 kV – 100 kV Settable (Step size 1 kV)
Generator Frequency : Up to 200kHz
High Voltage Rise Time : < 3 mS
Tube-head current range : 10 mA – 100 mA
Maximum Exposure Time : 5 seconds
Minimum Exposure Time : 10 mS
mAs range : 0.1mAs to 250mAs
Maximum Electrical Output : 4kW
Duty Cycle : 1:30 Duty cycle for exposure ≥ 1 sec and minimum of 30sec off
/ rest time & auto limit based on temperature.
X-ray field (at collimator tip) rectangle
Collimator Rotation : +/- 90 deg
Light intensity : >160 Lux at 1 meter
Maximum X-Ray field : 480 mm X 480 mm at 1 meter from focal spot.
Leakage radiation @ 1meter : < 50 mR / Hour
Cassette Storage Box : Yes, Standard Storage box provided
Tube Head Outer Covers : Aluminium sheet metal
*Anatomical Programming : Yes

X-Ray Tube Insert Specifications


Focal Spot : 1.8 mm
Anode material : Tungsten
Anode angle : 19°
Insert Inherent filtration : 0.5 mm Al equivalent
Anode thermal capacity : 30 kJ

Mains Power Requirements


Line voltage range : 230 Vac ± 10% / 110 Vac ± 10%
Line frequency : 50 Hz ± 5% / 60 Hz ± 5%
Maximum / Momentary Input current : ≤ 16 Amps for 230 Vac / < 35 Amps for 110 Vac

Regulatory Compliance Statements


This product is designed and manufactured to meet the following standards and regulations:

Market Type Applicable Standards

General Safety / Performance IEC 60601-1, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54,
ANSI/AAMI ES 60601-1, CAN/CSA C22.2No.60601-1:2008
Europe Safety / Performance MDD 93/42/EEC: 1993, class IIB device, CE Marked

India Radiation Safety AERB/SC/MED-2/BIS

Global QMS ISO 13485:2003, ISO 9001:2008