a>
VECTOR
ihn
el.
a / MINIMUM SIZE,
wel
CAPABILITY
:
BC-10e
Hematology Analyzer
Sy
+ CBC+3-DIFF, 20parameters+3histograms
+ Throughput : 30 samples per hour
+ Intuitive operation system with 8.4 inch TFT touch screen
+ Most compact footprint and best power efficiency
+ Open vial sampling
+ 50,000 results storage with histograms
+ Reagent barcode system for better handling of reagents
+ New operation system platform
+ High WBC Linearity with accuracy
+ Support bi-directional LIS
aot ey ratte
Principles Reagent
+ Impedance method for WBC, RBC and | DILUENT
PLT counting LYSE
* Cyanide free reagent for PROBE CLEANSER
hemoglobin test
Parameters Sample Volume
+20 parameters : WBC, Lymphi, Milt, Gran, Lymph%, | Whole blood <20uL
Mid, Grane, RBC, HGB, HCT, MCV, MCH, MCHC,
RDW-CY, RDW-SD PLT, MPV, POW, PCT, P-LCR
+ Bhistograms for WBC, RBC and PLT
C€ | “Think Hematology ! Think Vecto:Performance
Parameter
‘WBC(10"/L)
RBC(10"/L)
Linearity Range
0-100
0-8.00
HGB(g/L) | 0-280
MCV(fL)
PLT(10°/L) 0-1000
Throughput
| 30 samples per hour
Display
| 8.4 inch TFT Touch Screen
Language
| English
Data Storage Capacity
Up to 50,000 results including
numeric and graphical information
Communication
LAN Port supports HL7 protocol
Support bi-directional LIS
Interface
4 USB ports (for external printer,
software upgrade, barcode reader,
‘extemal WIFI, keyboard
‘and mouse), LAN port (1)
ace Ri 13
Eft) 027
ICE
Carryover
$0.5%
$0.5%
£1.0%
<15% (70.0-110.0)
8.0% (100-500)
Precision (CV %)
£4,0% (4.0-15.0)
$2.5% (3.5-6.5)
$2.0% (100-180)
$1.5% (70.0-110.0)
8.0% (100-500)
Printout
Thermal recorder, 50 mm
wide paper, various
printouts formats
Operating Environment
Temperature : 15°C~30°C
Humidity 20%~85%
Air pressure : 70kPa~106kPa
Power Requirement
(l00v-240v)210%, <180vA
(S0Hz/60Hz) +1Hz
Earth Voltage <5V
Online UPS (230V) with isolated
transformers - 1 KVA
Dimension and Weight
Dimension
Weight
$20Kg
wuodr
Fon Pes Oo
‘Ente, sviscraseanreDenare
Depth (410 mm) x width (300 mm) x height (400 mm)
VECTOR BIOTEH PVT. LTD.
\VePLAY1910020"er.01500