ARL 3460 Metal Analyzer déf. 15.3.

2001 17:16 Page 3

Analysis of Iron and Steel Application


by Optical Emission
Summary
ARL 3460 Metals Analyzer

Introduction following elements: nickel, chromium, manganese,


silicon, cobalt, tungsten, molybdenum and
The ARL 3460 Metals Analyzer is custom designed to vanadium. This category covers steels which are
meet your specific requirements whether you are a destined for a wide variety of uses such as :
small cast iron foundry or a large steel plant. Thermo • Stainless steels (ferritic or austenitic 18/8,
ARL’s long experience in metals analysis comes from maraging and all types of special stainless steels)
an installed base of over 6000 spectrometers world- • Tool steels
wide. The ARL 3460 is the answer to your metallurgi- • High manganese steels
cal analysis needs; whether they be incoming material • High speed steels
control or metal QC and production analysis. • Special high alloys steels, such as Ni-hard,
Working 24 hours a day and 7 days a week, the ARL Ni-resist, etc..
3460 delivers dependable performance year after year.
ARL 3460 Metals Analyzer
Iron
Several kinds of iron exist, distinguished by their
composition and use. They belong to two main
categories:
• Pig irons (hot metal); basic product for the
manufacture of steel.
• Cast irons; used for the production of semi-
manufactured products.

From a metallographic point of view, a distinction can


be made between white cast iron with a cementite
structure, and grey cast iron which contains free
graphite, either in the form of laminae or nodules.

Alloy cast irons also exist where alloying elements


such as nickel, chromium, manganese, copper, etc.,
are added to improve hardness, corrosion resistance The ARL 3460 optical emission spectrometer can
or engineering properties (Ni-hard, Ni-resist, etc. …). determine up to 60 elements. The HiRep source has
the flexibility to customize the excitation parameters
Steel providing the best analytical conditions for a wide
variety of sample types.
Typically steels are divided in two main qualities:
1. Low alloy steels (≤ 94% iron) The source uses the HEPS (High Energy-Spark)
2. High alloy steels (with notable quantities of Cr, technique to minimize the metallurgical effects and
Ni, Mn, Mo, Co, W, etc. …). produce consistently accurate results.

1. Low alloy steels The printout of average concentration for the


required elements from a repeat analysis is obtained
The total element alloying is generally less than in typically one minute.
5-7 %. The main elements are: Mn (≤ 2%),
Cr (≤ 3%), Mo (≤ 1.5%), Cu (≤ 2%), Ni (≤ 5%), Sample preparation
and V (≤ 1%). The so called “free-cutting steels”
with a concentration of 0.2 - 0.3% sulfur or lead, The sample is generally prepared by using a grinding
need a special calibration. machine. Milling machines can also be recommended.

2. High alloy steels Sample analysis time


High alloy steels contain, in addition to iron and Low Alloy Low Alloy Cast Iron Free
Steels with Steels with Cutting
carbon, notable quantities of one or more of the Nitrogen & Nitrogen Steels
sol./insol. Al
27s 25s 31s 68s
ARL 3460 Metal Analyzer déf. 15.3.2001 17:16 Page 4

Calibration summary Factory calibration (CARL)


Twelve “qualities” are available for specific alloy types: Thermo ARL's optical emission spectrometers can
be factory calibrated for iron and steel utilizing
• Low Alloy Steel
CARL (Calibration Applied Research Laboratories).
• Free Cutting Steel (S and Pb ≤0.3 %)
CARL is a very sophisticated multi-variable regression
• Chrome Steel (ferritic stainless steel)
tool that corrects for matrix effects as well as spectral
• Chrome-Nickel Steel (austenitic stainless steel)
interferences. CARL can provide an immediate
• Manganese Steel (Mn ≤20 %)
"turnkey" system which gives the user the highest
• High Speed Steel (Co ≤10 %)
accuracy possible. The calibrations are available for
• Valve Steel (Ni, Mn and Cr ≤20 %)
the different qualities given left. For each quality
• Low Alloy Steel NIST (USA calibration with
Thermo ARL uses certified material as standard
NBS samples)
samples and setting-up samples are delivered with
• Cast Iron (including Ni-hard - Ni ≤7 %)
the instrument to maintain the accuracy of the
• High Alloy Cast Iron (Cr ≤32 % & Ni ≤16 %)
calibration.
• Nickel Resist (Ni ≤35 %)
• Global (including all qualities except for free
cutting steel)

Example of factory calibration

CAL-34Fe-01 CAL-34Fe-02 CAL-34Fe-03 CAL-34Fe-04 CAL-34Fe-05 CAL-34Fe-06 CAL-34Fe-09 CAL-34Fe-12


Element Low Alloy Steel Free Cutting Steel Cr Steel Cr-Ni Steel Mn Steel High Speed Steel Cast/Nodular Global
Iron Ni-Hard
Fe 99.9%-80% 99.9%-90% 90%-65% 89%-45% 91%-70% 95%-48% 97%-70% 99.8%-42%
Al 0.001-0.3 0.001-0.10 0.002-0.3 0.001-0.2 0.003-0.1 0.001-0.1 0.003-1.5
As 0.001-0.1 0.01-0.07 0.002-0.1 0.003-0.1
B 0.0002-0.01 0.0002-0.01 0.0005-0.1 0.001-0.1
Bi 0.002-0.01 0.003-0.03
C 0.003-1.3 0.003-1.3 0.03-2. 0.01-1. 0.5-1.5 0.3-1.2 1.7-4.2 0.01-4.2
Ca 0.0002-0.01 0.0005-0.015
Ce 0.002-0.08 0.002-0.015
Co 0.001-0.3 0.008-0.15 0.01-0.2 0.02-0.4 0.1-0.4 0.05-10. 0.001-0.2 0.002-12.
Cr 0.001-5. 0.001-3. 10.-25. 10.-26. 0.1-4. 2.-5. 0.008-9. 0.005-32.
Cu 0.001-0.7 0.001-0.7 0.04-1.5 0.03-4. 0.003-0.5 0.01-0.1 0.003-2.5 0.005-6.5
La 0.003-0.025 0.002-0.04
Mg 0.001-0.1 0.001-0.15
Nodular Cast only
Mn 0.001-2. 0.05-1.7 0.2-1.5 0.02-10. 8.-20. 0.08-0.5 0.002-2. 0.005-20.
Mo 0.001-1.5 0.001-1. 0.03-2.5 0.01-6. 0.02-2. 0.1-9. 0.001-1.3 0.005-9.
Nb 0.001-0.6 0.005-0.05 0.005-1.5 0.005-2.5 0.002-0.06 0.002-2.7
Ni 0.002-5. 0.002-3. 0.05-4. 3.-35. 0.03-1.5 0.03-0.45 0.003-7. 0.005-35.
P 0.001-0.1 0.001-0.1 0.003-0.03 0.002-0.04 0.01-0.06 0.01-0.05 0.002-1. 0.005-1.4
Pb 0.001-0.02 0.001-0.3 0.002-0.03 0.002-0.03
S 0.0005-0.08 0.001-0.3 0.001-0.03 0.001-0.04 0.005-0.06 0.001-0.04 0.001-0.2 0.002-0.2
Sb 0.002-0.07 0.002-0.15 0.005-0.15
Si 0.001-3. 0.002-1.4 0.2-1.2 0.05-4. 0.2-1.4 0.1-0.4 0.2-3.5 0.005-5.5
Sn 0.001-0.1 0.001-0.1 0.002-0.03 0.005-0.03 0.001-0.3 0.002-0.3
Ta 0.008-0.15 0.008-0.04 0.01-0.1
Te 0.002-0.03 0.002-0.03
Ti 0.001-0.4 0.001-0.1 0.02-0.2 0.002-2. 0.001-0.3 0.002-2.
V 0.001-1. 0.001-0.5 0.03-0.9 0.005-0.25 0.02-0.3 0.4-3. 0.001-0.4 0.002-9.5
W 0.004-0.6 0.05-0.3 0.1-0.5 0.02-4. 1.8-22. 0.003-0.1 0.01-22.
Zn 0.001-0.02 0.002-0.025 0.002-0.025
Zr 0.001-0.1 0.005-0.05 0.002-0.1

CAL-34Fe-01 CAL-34Fe-34 CAL-34Fe-312


N 0.001-0.02 0.003-0.65 0.005-0.65
ARL 3460 - TYPICAL DETECTION LIMITS (3 sigma) AND PRECISION VALUES (1 sigma) FOR IRON BASE

ELEMENT Al As B Bi C Ca Ce Co Cr Cu Mg Mn Mo N Nb
Typical
1.1 2.4 0.3 4.1 6.9 0.4 6.0 1.2 1.6 0.6 2.8 1.2 1.7 3.5 2.6
DL [ppm]
ARL 3460 Metal Analyzer déf.

Guaranteed
≤ 2.2 ≤ 6.2 ≤ 0.9 ≤ 8.3 ≤ 12 ≤ 1.5 ≤ 11 ≤ 2.0 ≤ 4.0 ≤ 1.8 ≤ 5.6 ≤ 2.0 ≤ 3.9 ≤ 6.9 ≤ 5.0
DL [ppm]
Level [%] SD SD SD SD SD SD SD SD SD SD SD SD SD SD SD
0.001 0.00006 0.00015 0.00002 0.00018 0.00023 0.00007 0.00025 0.00005 0.00015 0.00006 0.0001 0.00005 0.0002 0.0003 0.00013
15.3.2001

0.002 0.00009 0.00015 0.00004 0.00026 0.00023 0.00014 0.00025 0.00005 0.00015 0.00007 0.0001 0.00005 0.0002 0.0003 0.00016
0.005 0.00016 0.00015 0.00012 0.00042 0.00023 0.00033 0.00025 0.00005 0.00015 0.00009 0.00025 0.00005 0.0002 0.00045 0.00022
17:16

0.01 0.00025 0.00020 0.00038 0.00062 0.00023 0.00065 0.00030 0.00010 0.0002 0.00012 0.0005 0.00008 0.0003 0.0006 0.00035
0.02 0.00038 0.00025 0.0010 0.00085 0.0003 0.0013 0.00050 0.00015 0.0003 0.00018 0.0009 0.00015 0.0006 0.0008 0.0005
0.05 0.0007 0.00035 0.0006 0.0013 0.0004 0.0005 0.0003 0.0022 0.00032 0.0010 0.0013 0.0010
Page 5

0.1 0.0012 0.00050 0.0010 0.0007 0.0008 0.0006 0.0043 0.0006 0.0015 0.0018 0.0015
0.2 0.0020 0.0020 0.0015 0.0012 0.0012 0.0080 0.0011 0.0020 0.0025 0.0025
0.3 0.0030 0.0025 0.0020 0.0015 0.0020 0.0026 0.0030 0.0031 0.0048
0.5 0.0050 0.0037 0.0030 0.0020 0.0035 0.0048 0.0050 0.009
1. 0.008 0.0062 0.0060 0.0035 0.0065 0.0091 0.008 0.017
2. 0.010 0.010 0.0050 0.010 0.0135 0.010 0.030
3. 0.015 0.015 0.007 0.015 0.017 0.012
4. 0.020 0.020 0.008 0.020 0.021 0.015
5. 0.025 0.010 0.025 0.039 0.025
10. 0.05 0.015 0.065 0.040
20. 0.025
30. 0.035
40. 0.045

Remarks: This data applies when homogeneous samples are prepared by recommended sample preparation methods.
The precision given is typical performance. Guaranteed values will be 1.5 times higher.
For multibase instruments, some analytical performance may vary based on the analytical line selected.
Values in brackets for detection limits can be obtained by using the most sensitive lines, or a longer analysis time.
Guaranteed DLs are calculated at 95% confidence limit.
ARL 3460 - TYPICAL DETECTION LIMITS (3 sigma) AND PRECISION VALUES (1 sigma) FOR IRON BASE

ELEMENT Ni P Pb S Sb Si Sn Ta Te Ti V W Zn Zr
Typical
1.8 2.0 3.3 1.5 8.5 3.6 1.8 16 4.3 0.7 1.4 9.0 3.0 1.7
DL [ppm]
ARL 3460 Metal Analyzer déf.

Guaranteed
≤ 5.1 ≤ 4.2 ≤ 5.7 ≤ 3.0 ≤ 23 ≤ 8.4 ≤ 3.5 ≤ 32 ≤ 22 ≤ 1.0 ≤ 2.0 ≤ 20 ≤ 5.0 ≤ 3.3
DL [ppm]
Level [%] SD SD SD SD SD SD SD SD SD SD SD SD SD SD
0.001 0.00015 0.00008 0.00017 0.00009 0.00025 0.00015 0.0001 0.0006 0.00015 0.00003 0.00007 0.00045 0.0002 0.00012
15.3.2001

0.002 0.00015 0.00009 0.00020 0.00014 0.00025 0.00015 0.0001 0.0006 0.00025 0.00005 0.00007 0.00045 0.00025 0.00012
0.005 0.00015 0.00010 0.00035 0.00028 0.00027 0.00015 0.0001 0.00075 0.0004 0.00012 0.00007 0.00050 0.0003 0.00013
17:16

0.01 0.00015 0.00013 0.0007 0.00050 0.00033 0.00020 0.00015 0.0010 0.0007 0.00022 0.0001 0.00050 0.00035 0.0003
0.02 0.00016 0.00026 0.0014 0.0009 0.00046 0.00025 0.00025 0.0015 0.0005 0.0002 0.00055 0.00038 0.0006
0.05 0.0003 0.0006 0.0035 0.0020 0.0010 0.00030 0.00040 0.0025 0.0010 0.0005 0.00065 0.00042 0.0015
Page 6

0.1 0.0005 0.0016 0.007 0.0040 0.0022 0.0007 0.00065 0.0035 0.0020 0.0008 0.00085 0.0052
0.2 0.0009 0.0033 0.014 0.0075 0.0012 0.0030 0.0015 0.0015 0.021
0.3 0.0013 0.005 0.021 0.012 0.0018 0.0050 0.0020 0.002 0.046
0.5 0.0020 0.009 0.0028 0.010 0.0030 0.003 0.10
1. 0.004 0.019 0.0053 0.020 0.006 0.005
2. 0.007 0.010 0.011 0.010
3. 0.011 0.015 0.015 0.015
4. 0.013 0.02 0.020 0.020
5. 0.016 0.025 0.025
10. 0.035 0.045
20. 0.065 0.09
30. 0.10 0.15
40. 0.13

Remarks: This data applies when homogeneous samples are prepared by recommended sample preparation methods.
The precision given is typical performance. Guaranteed values will be 1.5 times higher.
For multibase instruments, some analytical performance may vary based on the analytical line selected.
Values in brackets for detection limits can be obtained by using the most sensitive lines, or a longer analysis time.
Guaranteed DLs are calculated at 95% confidence limit.
ARL 3460 Metal Analyzer déf. 15.3.2001 17:16 Page 1

Accuracy Stability
Precision is only a small part of providing accurate Stability of the instrument is of the utmost importance
analyses. The most important factor is the accuracy when doing routine analysis. Typical mid-term stability
and quality of the calibration standards. Next is the measured over 24 hours shows that the standard
development of the calibration curve relative to a deviation achieved is below two times the precision
specific analytical task. Matrix matching and high value, which is excellent.
energy pre-burn reduce or eliminate matrix effects,
and spectral interferences are significantly reduced by The example below shows the long term stability of
applying appropriate corrections. two elements recorded over a period of more than 15
days without any intermediate drift correction. The
The example shown below illustrates the significant values almost never went outside the control limits.
improvement with the CARL calibration.
Moving Average Chart: C%
0.0200

0.0180

0.0160

0.0140

0.0120

11-12-9 15-12-9 20-12-9 24-12-9 28-12-9 04-01-9


18:30:0 19:55:0 23:21:3 23:13:5 22:46:0 19:00:0
50 100 150 200 250 300

Moving Average Chart: Cr%


0.0310

0.0300

BEFORE CARL 0.0290

0.0280

0.0270

0.0260

0.0250

11-12-9 15-12-9 20-12-9 24-12-9 28-12-9 04-01-9


18:30:0 19:55:0 23:21:3 23:13:5 22:46:0 19:00:0
50 100 150 200 250 300

Performance guarantee
Thermo ARL guarantees the precision shown in the
attached tables using homogeneous samples and
recommended sample preparation

This list will be updated as improvements are announced.


AFTER CARL Please contact your nearest Thermo ARL office for
the most recent values.

Options The precision is calculated from the formula:

1. Global calibration for iron i=n


where:
Σ(Xi-Xi)2
SD(1σ)=± i=l Xi the individual readings
This calibration offers more than a sorting pro- n–1 Xi the arithmetic mean of the
gram in terms of accuracy and number of analyzed individual values
elements. The global calibration for iron can be RSD = SD n the number of determinations
considered as the basic low cost calibration of the Xi
instrument, allowing the analysis of unknown samples.
The DL (Detection Limit)) is defined as three times
2. PIM 2 the standard deviation of the background expressed
in concentration units.
This option permits the determination of
soluble/insoluble aluminum and boron in low
alloy steel. It does not need special skills from the
operator and replaces a costly wet chemistry procedure.

For both options, separate application notes are


available on request.
ARL 3460 Metal Analyzer déf. 15.3.2001 17:16 Page 2

Thermo ARL
ARL Applied Research Laboratories S.A.
En Vallaire Ouest C, case postale, CH-1024 Ecublens, Switzerland
Tel. ++41-(0)21-694 71 11, Fax ++41-(0)21-694 71 12
e-mail: marketing@thermoarl.com - www.thermoarl.com

AUSTRALIA, Rydalmere (Sydney)


Conclusion Tel.: ++61-(0)2-9898-1244, Fax : ++61-(0)2-9684-4244
e-mail: sales@thermo.com.au
The ARL 3460 has all the total system features which AUSTRIA, Vienna
meet the critical needs of the metals analysis markets: Tel.: ++43-(0)1-333 50 34-0, Fax : ++43-(0)1-333 50 34-26
e-mail: info.optek@thermo.at
• Unmatched hardware for stability and reliability BELGIUM, Bornem
• Excellent performance in detection limits, precision, Tel. : ++32-(0)3-890 47 70, Fax : ++32-(0)3-890 47 74
accuracy, stability and analysis time. e-mail: info@thermo-optek.be
• Most advanced software technology CANADA, Mississauga
• Operation by unskilled worker or research chemist Tel.: ++1-(905)-890-1034, Fax: ++1-(905)-890-5775
• Widest range of metals analysis e-mail: thermoserv@aol.com
• Adaptable to the ARL automatic Sample FRANCE, Montigny le Bretonneux
Manipulation System (SMS-1000) Tel.: ++33-(0)1 39 30 53 00, Fax: ++33-(0)1 39 30 53 01
• Advanced technical/service support e-mail: arl@thermo-optek.fr
• Laboratory accreditation guidance GERMANY, Offenbach am Main
• Immediate access to parts inventory Tel.: ++49-(0)69- 98 408-170, Fax : ++49-(0)69-98 408-157
e-mail: arl_offenbach@compuserve.com
All these features allow you to optimize your ITALY, Cinisello Balsamo (MI)
productivity and to achieve the shortest payback times: Tel. ++39-(0)2-6601.1808, Fax: ++39-(0)2-6601.1689
e-mail: arl@thermo-optek.it

• Your production costs are reduced by more accurate NETHERLANDS, Breda


and reproducible analyses that are available quickly. Tel.: ++31-(0)76-572 48 40, Fax: ++31-(0)76-581 09 61
• Your production costs are reduced by the increased e-mail: info@thermo-optek.nl
instrument availability thanks to its high stability SOUTH AFRICA, Kempton Park
and drift corrections being less frequently required. Tel.: ++27-(0)11-394-1410, Fax : ++27-(0)11-394-1204
• Your operating and maintenance costs are reduced e-mail: arlabs@arlabs.co.za
through reduced consumption of drift correction SPAIN, Alcobendas (Madrid)
samples, and simple maintenance. Tel.: ++34-(91)-657 49 30, Fax: ++34-(91)-657 49 37
• Your overall cost management is reduced by e-mail: comercial@thermo.es
optimum utilization of materials and extremely low SWEDEN, Bromma
running costs compared to other methods. Tel.: ++46-(0)8-445 27 70, Fax : ++46-(0)8-627 52 20
e-mail: info@thinab.se

Thermo ARL's 65 years of experience and history of UNITED KINGDOM, Crawley


innovative technology makes it the world leader in Tel.: ++44-(0)1293-561 222, Fax ++44-(0)1293-561 980
e-mail: enquiries@thermooptek.co.uk
OE metals analysis. Thermo ARL works with its
customers to improve the efficiency of their analytical UNITED STATES, Dearborn, MI
Tel.: ++1-(313)-271-5760, Fax : ++1-(313)-271-5542
tasks and thereby increase productivity. e-mail: arl@thermo-optek.com

UNITED STATES, Franklin, MA


Tel.: ++1-(508)-553-1600, Fax : ++1-(508)-553-1679

Other offices, representatives and service centres


are located throughout the world

The descriptions contained herein were correct at the time of printing and
are subject to alteration without notice. All orders are accepted subject to
our current conditions of sale, copies of wich are available on request.
© Thermo ARL-CH publication OE-05-941-1100-CH-e

A Thermo Electron business


incorporating ARL, Baird, Hilger and Scintag