Professional Documents
Culture Documents
Manual de Aseguramiento de La Calidad para Sistemas de Medición de Contaminación Atmosférica Volumen II, Metodos Especificos para Aireambiente PDF
Manual de Aseguramiento de La Calidad para Sistemas de Medición de Contaminación Atmosférica Volumen II, Metodos Especificos para Aireambiente PDF
25 h. The actual sampling
period must be known to within +15,
min (see Section 2.1 2)
4.3.4. Sample Handling - The
procedure for sample handling is a5
follows:
1. First remove the accordion
tubing (color-coded) from the absorber
impinger tube.
2. For 24-n samples, estimate the
volume of absorber reagent remaining
in the absorber flask after sampling,
mark the TCM level n the absorber
with tape oF some other easily
removable indicator, and record the
estimate on the sample record form.
To assist in this estimation, mark the
tubes at the 35- and 50-ml levels with
a marking pen. If <35 ml of sampling
reagent remains after sampling, void
the sample, and indicate this on the
form. If 50 ml of sampling reagent
mains after sampling, state (under
Remarks” on the sample record
form) that no evaporation occurred
during the sampling.
3. If there 1s evidence of
malfunction (¢ g, absorbing reagent
forced out of absorber into the
vacuum system), record this on the
sample record form Normally, the
sample 1s invalid under this
circumstance
4. Place the tube caps first on the
impinger tube (color coded) of the
absorber and then on the suction
tube Press caps firmly in place. and
place the absorber in the shipping
block, Note: For 30-min and 1-h
samples collected in glass impingers,
shake the absorber thoroughly.
quantitatively transtar all of the
‘exposed reagent to a test tube or
small bottle with a Teflon-tined
threaded cap, and place the sample in
the shipping block
5 Record on the sample record
form any pertinent observations
relative to sources, weatherSection 2.1.4 6 Jan, 1983
Gas Bubbler Data Record
pen Clamp
248
240 248
‘Station Location Vacuum Reading Start
Cay @ State Vacuum Resting End
Ste & Adress se TS
mo fay 51 mn
Project SMELTER site no SZ ‘Stop Semping 6 6 BI 23 67.
mmo dey wr =
Pollutant Elapsed Time Start
SO,
Sompler 10 No, — 22
Type of Flow Measuring Device
Identification Number of Device
Inival Flow 204 rain
a
veo 205 enl/onin—
semoie no, 471 __
—_ROTAMETER
‘Nonunol Flow Rate
Final Flow 2OB ml/min» aiterence LO
Average
Average sample temperature ZO °c Retawve humeity 5090
Ambvent temperature: Start —
Barometric pressure. Start —_mm Hg. Stop
REMARKS: :
°C, Stop —
"C, Average 25. 2c
_ mm Ha, avera90 AVL Ome Hg
‘Meteorological conditions for use when anomaly occurs
Elapsed Tune Stop L437
mn
Sample collected within
‘guidelines gwen below
WIND calm taht —— gusty
wisiiiry clear hary
sky Z. clear scatered overcast
HUMIDITY dry moderate humid rainy
TEMP °F <20___ 20-40 41-60 “61.80 >90,
Gudennes
Proper How rate —1/2:h samples - 900-1100 em?/mn
1h samples
24h samples
"450-550 em'/min
180.220 emn/min
24: sampling ~ 23 h < samphng time < 25h
|= 335 mi of absorbing reagent
— sampler timer accuracy #15
‘Sample record form.
Figure 4.10.
conditions, etc., that might affect the
S02 level
6. Complete the sample record form
in duplicate, return the original copy
with the absorber in the shipping
‘block, and file the duplicate in the site
log book.
7. If samples must be stored before
shipment for analysis, store them in a
refrigerator or cooler at 6°C (41°F). A
shipping container that can maintain,
2 temperature of 5° +5°C is used to
transport the sample from the
collection site to the analyti
laboratory. Ice coolers or refrigerated
shipping containers have been found
‘remain atter sampling
‘min’26
to be satisfactory. The use of eutectic
cold packs instead of ice will provide a
better temperature control. Such
equipment is commercially available
from Cole-Parmer Company, 7425
North Oak Park Avena, Chicago,
Minois 60848,
4.4 Determination of the
Flow Rate of SO2 Sampling
Trains
‘The SO2 sampling train that is used
{for short-term periods of 30 minutes.
and 1 hour normally has one of two
flow control systems. One system has
2 needle valve and indicates the flow
with a rotameter. The other uses
various sizes of stainless steo!
hypodermic needles to control the
flow. The 24-h sampling trains also
use staini | hypodermic
needles to the flow. For short-
mpl indard flow rate
is determined at the sampling site at
the initiation and completion of
sample collection by use of a
calibrated flow-measuring device
connected to the inlet of the absorber.
For 24-hour samples, the standard
flow rate is determined at the time
the absorber is placed in the sampling
train and again when the absorber isJan. 1983
Section 2.1.4
moved from the train for shipment
to the analytical laboratory. A
librated flow-measuring device
connected to the inlet of the sampling
train is used for this purpose. The
flow rate determination must be made
with all components of the sampling
system in operation (e.g., th
absorber temperature controller and
any sample box heaters must be
operating).
4.4.1 Flow Rate Measurement -
‘Three step-by-step procedures for
‘measuring the sampling flow rates
are presented here, They involve the
use of a mass-flow meter, a
rotameter, or a soap-bubble meter.
The reference method requires that
the flow rate of each SOz sampling
train be measured before and after
each sampling period. The initial and
final flow rates must agree within +5
Percent or the sample is voided.
Therefore, itis very important that the
flow-measuring device selected have
‘the appropriate range, accuracy, and
readability to meet this specification.
The typical flow rates for 30-min, 1
hour, and 24-hour samples are 1000,
500, ‘and 200 ml/min, respectively
The range of a mass-flow meter
sed as a flow measuring device
hould be such that the measured
flow is between 50 and 90 percent of
full scale Regardless of the range,
the mass-flow meter must have @
readability and reproducibility of 1
percent of the measured flow to
ensure that only samples from trains
whose flow rate did change by more
than 5 percent are invalidated. Mass-
flow meters should be calibrated
according to procedures given in
Subsection 2.1.3,
Rotameters used as flow measuring
devices should have a minimum scale
of 6-in. (150 mm) with no less than
100 divisions The measured flow
should be between 50 and 90 percent
Of full scale, If these conditions are
met, only samples from the trains
whose flow rate did change by more
than 5 percent should be invalidated
Rotameters should be calibrated
according to procedures presented in
‘Subsection 2.1.4
Bubble meters should be of a size
that the volume is displaced in 10 to
30 seconds during the flow
‘measurement. Bubble meters are
calibrated according to procedures
resented in Subsection 2 1.2
When flow measurements are made
y inserting @ mass-flow meter or
rotameter Between the probe and the
absorber, the tume required for the
initial and final flow measurements
should be included in the total
sampling time,
Bubble meters are not as easily
ingerted between the probe and
absorber inlet, and the solution in the
bubble meter will absorb at least
some of the SO.. Therefore, the time
required to do the initial and final flow
measurements should not be included
in the sampling time, and a charcoal
scrubber should be Inserted between
the bubble meter and the absorber to.
ensure the removal of SOs. Care must
be exercised in the selection of the
charcoal tube so that the pressure
drop across the sampling system will
not be increased enough to alter the
sampling rate.
4.4.2. With a Mass Flow Meter - The
following procedure should be used to
measure the flow before and after
sampling.
1, Assemble the equipment for
sampling, as shown in Figure 4.11
2. Be Sure all connections are tight.
3, Be sure the absorber inlet line is
connected to the mass-flow meter
transducer outlet,
4, Zero the mass-flow meter,
5. Turn the vacuum pump on and
adjust the needle valve to obtain
approximately the desired flow (prior
to sampling only). Record the vacuum
reading on the sheet
6. Check for bubbles in the
sampling tube. Plug the inlet of the
mass-flow meter, and check for zero
flow on the rotameters. If a zer0 flow
Js not obtained, check for leaks.
Unplug the inlet
7. Read the flow rate on the mass-
flow meter after the meter indicator
has stabilized (about 30 s) and shut
off the pump.
&. Turn the pump on, and repeat
Step 7 two more times.
9. Average the three flow rates.
4.6.3, With a Rotameter - The
following procedure should be used to
measure the flow before and after
sampling
1. Assemble the equipment,
‘shown in Figure 4.12.
2. Place a new rubber septum on
the tube (prior 10 sampling only).
3. Be sure all connections are
tight
'4. Be sure the line from the surge
tank outlet is connected to the inlet of
the sampling tube absorber.
5 Turn the vacuum pump on, and
ccheck the vacuum by placing a finger
‘over the end of the vacuum line. (The
‘vacuum gauge should read at least
530 mm (21 in.) Hg at a barometric
pressure of 760 mm (29.92 in), Turn
the vacuum pump off.
6. Insert a needle through the
center of the septum, perpendicular to
the plane of the septum face.
7. Turn the vacuum pump on. Slide
the needle holder snugly into the
recess at the base of the needle, and
ccheck for bubbles in the sampling
tube. Plug the inlet to the rotameter
land check to see that bubbling ceases
in the absorber. if bubbling persists,
check for leaks. Unplug the inlet.
8. Read the ball position on the
rotameter; record the flow rate
corrected to STP, 760 mm {29.92 in.)
Hg and 25°C (77°F), from the
rotameter calibration curve; and read
the vacuum gauge to make sure the
vacuum is sufficient during the flow
measurement
9. Repeat Step 8 two more times.
Turn pump off and then back on after
each run.
10. Record the ambient temperature
and pressure oF obtain the average for
the sampling period from the local
‘weather bureau.
11, Average the flow rates of the
three runs,
12. If the rotameter is used in a
field location where the barometric
pressure and/or temperature is
which the rotameter was calibrated,
apply the following corrections to
convert the flow rate to standard
conditions. This correction factor is
not advisable for conditions that differ
‘greatly from those ich th
Fotameter was calibrated, Greater
accuracy can be obtained by
developing calibration curves for
expected conditions.
=a, ee
760 Ty
Equation 4-1
where
,= flow rate corrected to standard
conditions from field conditions,
m/min
i= flow rate at field conditions
‘from calibration curve, ml/min
P,= barometric pressure at field
conditions, mm Hg
T= temperature at field conditions,
°K
4.4.4 With a Bubble Meter - The
following procedure should be used to
measure the flow before and after the
sampling,
1. Assemble the equipment for
sampling, as shown in Figure 4.13,
2. Place a new rubber septum on
the sampling tube only prior 10
sampling,
3, Be sure all connect
2 10 tightSection 2.1.4 a
Jan. 1983
ro Probe =
‘Mass Flow meter
i0 ous
30 mi
20 mi
10m
‘Note: Rotameter and needle valve can
bbe replaced with appropriately sized
hypodermic needle (Section 2.1.11]
Needle
Rotameter
Glass woo! or
Silica Get
—
diosy "Ora, amma
—
Note A midget mpinger 18
sed for T-hour sampling
Figur
12
4, Be sure the charcoal tube is
connected between the bubble meter
and the inlet of the sampling tube
impinger.
5. Turn the vacuum pump on and
ccheck the vacuum by placing a finger
‘over the end of the vacuum line, (The
vacuum gauge should read at least
530 mm (21 in) Hg at a barometric
pressure of 760 mm (29.92 in) Hg.
Turn the pump off
6. Insert a needle through the
center of the septum, perpendicular to
the plane of the septum face.
7. Turn the vacuurn pump on,
8. Slide the needle holder snugly
into the recess at the base of the
needle, and check for bubbles in the
Absorber for
24n Sampling
Determination of Hw rate of long-term sompling trein using @ rotometer.
sampling tube. Plug the inlet to the
bubble meter and check to see that,
bubbling ceases in the absorber. If
bubbling persists, check for leaks.
Unplug the intet.
9, Touch the surface of the soay
solution with the open end of the
bubble meter so that a soap bubble
will start to travel up the bubble meter
tube. Repeat several times or until the
bubble travels the full length of tube
without breaking
10. Touch the surface of the soap
solution with the open end of bubble
meter so that a single bubble starts to
travel up the bubble tube. When the
bubble passes the first line of the
‘graduated scale, start the stopwatch
‘and record the time and volume
displaced, Read the vacuum gauge to
make sure the vacuum is sufficient
during flow measurement,
11. Repeat Step 10 two more
times. Record the time and the
volume each time,
12, Record the ambient
temperature and pressure, or obtain
the average for the sampling period
from the local weather bureau.
13, Average the times and volume
displacements for the three runs.
14, Divide the average volume
displacement by the average time to
determine the flow rate.Jan. 1983 9 Section 2.1.4
un. Tube
e 6-10 Mesh
Polypropylene
2-Port Tube Closure
Membrane Filter
Bubble
Meter
Absorber for
24h Sampling
Figure 4.13
18. Correct the average flow rate 10
standard conditions of 760 mm (29.92
in.) Hg and 25°C (77°F), using
Equation 4-2.
Qs = Quer x Pe {1-BHIPi0
Pasa
298.16
(Tate: * 273.16)
Equation 4-2"
wher
‘Qay= flow rate at standard
‘conditions, std liters/min
(25°C and’760 mm Hg)
Qac:= flow rate at monitoring sit
‘conditions, liters/min
jarometric pressure at
monitoring site conditions ,
mm or in. Hg
fractional relative humility
of the air being measured;
determined with a sling
psychrometer or similar
evice, oF obtained from the
local weather station.
apor pressure of water at
the temperature of the air in
the flow or volume standard,
in the same units as Ps (for
‘wet volume standards only,
‘.e., bubble flowmeter of wet
test meter, see Section 2.1.2:
for dry standards.
barometric
pressure, in the same units as
P, (760 mm Hg or 29.92 in,
Ha)
Pro:
Tram= temperature of the air in the
‘flow or volume standard, °C
{e.g., bubble flowmeter)
Ha barometer is not available, the
following equation may be used to
determine an average barometric
pressure:
Px = 760-.076(H) mm Hg or Pe
29,92-.0030(H) in. Hg
Equation 4-3
where
H = sampling site elevation above sea
level in meters.
4.5 Flow Rate Validation
.6.1 Initial Flow Rate I the flow
meter or the hypodermic needle in the
sampling train has been calibrated,
the initial flow rate (determined by
‘one of the flow-measuring devices
listed in Subsection 4.4) should be
compared with the rate of the flow-
controlling device to determine if the
train is leak-free.
If the initial flow rate (Q,) differs
from the flow rate of the critical
orifice or the flow rate indicated by
the flowmeter in the sampling train
(Qc) by more than 5 percent (as
determined by Equation 4-4), checks
for leaks should be made and the O,
should be determined
% Diff -Q-Qe x10
Qe.
Equation 4-4
Hypodermc Needle
Determination of flow rate of long-term sampling train using soap-bubble meter
4.8.2 Final Flow Rate The initial
flow rate and the final flow rate must
agree within +5 percent for a valid
‘sample to be collected. This check
should be performed in the field when
the final flow measurement is made
0 that any discrepancies can be
determined at that time. Immediate
‘onsite determination of the reasons
{for variable flow rates can prevent
further data loss.
The sample should be invalidated if
the difference between the initial (Q,)
and final (Qi) flow rates is more than
5 percent as determined by Equation
45,
Q1x100
a
96 Diff =O.
Equation 4-5Section 2.1.4 10 Jan. 1983
Table4.1 Activity Matrix for Sarmpling Procedure
Action it
Frequency and method uicements
Activity Acceptance limits ‘of measurement are not met
Absorber selection 1) Short-term, <3 h; Visually check each ‘Select proper absorber
30-mi all-glass midget sample.
impinger
2) 24-h sample: poly
propylene tube and
closure with glass
impinger tube
3) Orttice 1D >#79 drill
‘and <178 drill |
Cleaning absorber Clean absorber Water and acid bath prior | Clean it
(0 each use
Dispensing TOM 1) Short-term period, 10 | Visually check each Correct the discrepancy
miTCM sample.
2) 24-h period, 60 mi TCM
Assembly of absorber As diagrammed Visually check Reassemble.
Determination of flow rates | Percent difference <5% | Calibrated flow measuring | Void sample.
before and after sampling device (Sec. 2.1.2)
‘Sample 10 All required information | Visually check Complete the stick-on
Tabel/
Packing for shipment Shipped in insulated coo!- | As above Repack as appropriate.
pack container
‘Sampler assembly, 24-h 'NASN specifications (Figs | As above Reassemble if necessary.
bubbler 46,47, 48)
Installation of absorber and | See Subsec. 4.2.6 and As above Reassemble and/or select
critical oritic 427. proper orifice
Collection of sample Air flow rate +5% (Subsec. | Initial and Final(Subsection | Void sample.
4.3); absorbing solution at | 4.4). Check thermometer
18 +10°C during sampling | for each sample, Subsec. | Void sample.
and prior to recovery 43.2
‘Storage of sample before Store at 8° +5°C As above Void sample.
analysis. _Jan. 1983
The quality control functions for
analysis of samples are summarized
in Table 5 2 at the end of this section
5.1. Verification of Field
Record Form
Procedures for verifying the field
record are as follows:
1. Examine the sample record form,
and invalidate the sample if vital
information 1s missing and cannot be
obtained from field personnel Remove
the samples from the shipping
container. If the shipping period
exceeded 12 hours, verify that the
sample temperature 1s below 10°C. If
the temperature 1s above 10°C,
invalidate the sample, Check t0 see
that the initial and final flow rates
meet the +5 percent criterion (see
Subsection 4.6.2). invalidate the
sample if the flow rates differ by more
than 5 percent. If the flow rates differ
by less than 5 percent, the average
value will be used to determine the
SOz concentration. Calculate the
sample volume, using Equation 5-1
Vor @
‘Ves sampling volume, std liters
Q.= standard flow rate determined
at the initiation of sampling, std
liters/emin,
Qy= standard flow rate determined
‘at the completion of sampling
Std liters/min
1= total sampling time, min.
2. Log the sample into the sample-
receiving log book
3. Verify the volume marked on the
absorber and recorded on the sample
record form against the sample
volume received. If the volume
received is more than 10 ml for long-
term sampling (2 ml for short-term
‘sampling) lower than the volume
marked on the absorber, significant
leakage has occurred, Void the
sample. If valid samples have <50 mi
of absorbing reagent, add distilled
water to bring the volume to 60 mi
and continue with the analysis.
‘samples have @ volume of >50 mi,
measure the volume with a graduated
cylinder (Class A) and record the
volume. Continue with the analysis.
Valid short-term samples are ready for
analysis. f samples are to be stored
Equation 5-1
5.0 Analysis of Samples
before analysis, place them ina
refrigerator at 5°C
5.2 Colorimetric Analyses
The following step-by-step
procedure should be used to obtain
the calibration curve for the
spectrophotometer and to show the
relationship of the yg of SOs to the
‘absorbance. Calibration Option 1 of,
the reference method is discussed.
The procedure described is intended
to be used with matched
spectrophotometer cells. if an
unmatched pair of cells is used, the
correction listed in the reference
method (Subsection 2.1.11) must be
applied.
1 Prepare a dilute working sult
TCM solution by diluting 10 mi of the
sulfite-TCM solution to 100 mil with
TCM absorbing reagent.
2 Accurately pipette the indicated
volumes (Table 5.1) of the sulfite-TCM
solutions into six 25-ml volumetric
flasks.
3. Add the indicated volumes of
TCM absorbing reagent to bring the
volume in each flask to 10 ml
‘4, Add 1 ml of 0.6% sulfamic acid
to each, and allow to stand 10 min,
5. Accurately pipet 2 mi of 0.2%
formaldehyde to each.
6. Add 8 ml of PRA solution to
‘each, and start the timer.
7. ‘Add recently boiled and cooled
distilled water to bring each flask to
the 25-ml mark, Mix thoroughly,
8. Place all six flasks in a water
bath for at least 30 min, but no longer
than 60 min, at 20°+-0.2°C.
(68°+0.4°F),
9. Allow at least § min for the
sspectrophotometer to warm up; adjust
the zero control to bring the meter
needle to infinity absorbance on the
scale; standardize the light control by
Section 2.1.8
inserting a 1-cm cuvette filled with
distilled water into the sample holder
land adjusting the light control until
the meter reads zero absorbance.
Note: The wavelength scale should be
checked with a standard wavelength
filter traceable to the National Bureau
‘of Standards initially and after each
160 hours of normal use or every 6
months, whichever occurs first. Use
‘commerically available, optically
matched 1-cm cuvettes.
10. Remove the six samples from
‘the water bath, pour a portion of each
fone at a time) into a cuvette,
immediately read the absorbance at
548 nm, and record on the SOs
Calibration Data Sheet (Figure 5.1),
Compute the ug SO2/lask, using the
following equation:
19 SOs = Vroms0 X Crowiso, x D
Equation 5-2
where
‘View so volume of sulfite-TCM
solution used, ml
Crems0,= concentration of sulfur
‘dioxide in the working
sulfite-TCM, ug SO2/ml
from Equation 3-3)
D= dilution factor (D = 1 for
the working sulfite-TCM
solution; 0 = 0.1 for the
diluted working sulfite-TCM
solution).
each 25-m flask (Figure 5.2).
12. Use regression analysis to
determine the slope, intercept, and
correlation coefficient of the
Calibration curve, (Refer to Appendix J
(of Volume | of the Handbook’ for
discussion of linear regression
analysis.) See Figure 6.3 for
calculation of the linear regression
parameters.
Table 5.1
Volume of Approximate
Sulfite-ToM Suifite-Tom Volume of totar
‘solution solution, ml TCM, ral ug SO
00 10.0 0.0
Ditute working 50 50 36
Ditute working 100 00 72
Working 20 80 144
Working 30 70 216
Working. 40 60 288
jased on working sullite-TCM solution concentration of 7.20 ug SOz/ml. the
‘actual total ug S02 must be calculated by use of Equation 8:2 (shown in Step 10).Section 2.1.5 2
oe
Instrument 1D
Color Development time BS
Wavelength SFE nm yg SOs/mi (Working sultte-TCM solution) LL Ong.
502 Cahbration Data Showt
Temperature —
Jan, 1983
SOzfme.
roe, ° La “Fisorbance
os B 1 . 00 0.160.
° o 3.6 [0.226
- = es 22 0.378
= ZI ! 144 0.590
3 1 ea 0.8il
s | 1 28.8 1.024
Veen so
Creu sop® Concentration of $02 n the working sulte-TCM solution. ng SOx/ml and
D> dilution factor
‘Regression analysis results.
Mossured
Slope 0.05 OOS
O/595
Intercept
coneaioneettoon 027998 3058
33.28 1 s0v’absorbance unit
$02 spectrophotameter calibration dat
4
slope
Calibration factor By
Figure 5.1
13. Check each new calibration
curve for conformance with these
eniteria
a. Slope is 0.03+0 002 absorbance
units/ug $02. If not, repeat the
calibration. If still outside the limits,
restandardize the sulfite solution and
check the dye.
b. The intercept 1s <0.17
absorbance units at 22°C (72°F) with
2 t-cm optical path length, and the
blank 1s + 0.030 absorbance units of
the regression intercept. If not, check
the temperature of the water bath. If
the new intercept still 1s outside the
limits, check the chemicals from
which the reagents were prepared
and prepare new reagents
¢. The correlation coefficient 1s
greater than 0.998
14. File the calculations and the
calibration curve in the calibration log
book.
15 Determine the calibration
factor, B, (ug SO2/absorbance unt),
whiett 1s the reciprocal of the slope.
jolume of sullite-TCM solution used, ml
Critena
0.0300.002 absorbance units/ug SO
5.3 Sample Analysis (24-h
Sampling)
The following step-by-step
procedure 1s to be used for analysis of
field samples.
1. If no leakage occurred during
‘sample shipment (as indicated by
comparing the volume marked on the
data form after sampling with the
volume received), dilute the sample
with distilled water to 50 ml If the
volume is greater than 50 ml
‘measure the volume with a Class A
graduated cylinder and record the
volume.
2. I leakage occurred, invalidate
the sample.
3. Pipette 10.0 mi of the sample
into a 25-ml volumetric flask. Note:
For shorter sampling periods (e g.. 1
or 3 h), quantitatively transfer the
entire 10 mi sample to increase the
detection limit
4 Delay the analysis for 20 min to
allow any ozone to decompose.
<0 170 at 22°C (add 0.015 per °C above 22°C)
5. Prepare a reagent blank by
pipetting 10.0 ml of absorbing reagent
into a 25-ml volumetric flask for use
throughout the analytical procedure.
Prepare two internal control standards
containing approximately 5 and 15 ug
‘of SO2 and add absorbing reagent to
bring the volume to 10 ml in 25-ml
volumetric flasks,
6. Analyze the reagent blank and
internal control standards before
the first sample, after every
subsequent 10th sample, and after
the last sample. Each reagent blank
absorbance should be within +0.03
absorbance units of zero, and the
measured value of each internal
control standard must be within + 1
ug of the true value
7. To each 25-ml flask add the
following
a. 1.0 ml of 0.6 percent sulfamic
‘acid (allow to react for 10 min to
destroy the nitrate from oxides of
nitrogen).
aniys: —f—_shematche=__ eJan. 1983,
Section 2.1.5
50; calibration curve
EPA pararosapiine releronce method
Analysis
Calibration Temperature 20.
DatewR2-BL
Calbration actor (Bs) 3.3.2
ope nS 53-3 yj 80,/abeobance unt
Sipe! = nn a8 309
—
10
oe
£ os
g
2
i
Boe ;
o2
0 ° 1016S
19580,
Figure 6.2. Example of $0, cohrevon cure
b. Pipette 2.0 ml of 0.2 percent
formaldehyde solution.
°c. Add 5 0 ml of PRA solution
8. Start the laboratory timer
previously set for 30 min
9. Dilute all flasks to 26 mi with
recently bailed and cooled distilled
water, stopper them, and place them
tna water bath at 20°4.0.2°C
(68:0.3°F)
10. Set the spectrophotometer
wavelength to 548 nm, and allow at
least 5 min for the spectrophotometer
to warm up. If necessary, adjust the
zero control to bring the meter needle
to infinity on the absorbance scale.
‘Standardize the light contol by
inserting a cuvette filled with distilled
water into the sample holder and by
adjusting the light control until the
meter reads zero absorbance
11. Determine as quickly and
accurately as possible, the absorbance
of each sample of the internal control
standard and of the reagent blank
Use distilled water, not the reagent
blank, as the reference.
12 Record all absorbance units
from the samples and the internat
‘control standard on the laboratory
data form. Figures §.4a and 5 4b)
13 If the absorbance of the sample
solution ranges between 1.0 and 20,
the sample can be diluted 1:1 with a
portion of the reagent blank and the
absorbance redetermined within 5
minutes Solutions with higher
absorbances can be diluted up to
sixfold with the reagent blank to
‘obtain scale readings of less than 1.0,
absorbance unit It is recommended,
however, that a smaller portion (<10
ml) of the onginal sample be
reanalyzed (if possible) if the sample
requires a dilution greater than 1:1
14, Dispose of all reagents
containing mercury by using one of
the procedures described in
Subsection 5.6. Until disposal, the
discarded solutions can be stored in
closed glass containers, and they
should be left im a fume hood
15. Immediately after their use,
clean the cuvettes with isopropanol to
avoid dye deposition on the cuvette
walls
16 Determine the pH of the
samples in step 11. Maximum pH
Sensitivity 1s 16 0.18
17, Calculate the SO
concentration as follows,
9 S02/m? = 105A - Ad) BaVexD
Va Va
Equation 6-3
10°= conversion of liters tom?
‘A= sample absorbance
‘Ac= reagent blank absorbance
By = calibration factor, ng
'$02/ absorbance unit
\Vn= the sample air volume
corrected to 25°C (77°F) and
760 mm Hg (29.92 1n Ha),
Iners,
volume of absorber solution
analyzed, ml
Ve= total volume of solution in
absorber, ml
D= dilution factor (f any) required
to reduce sample absorbance
below 1
And calculate the SO3 concentration
(pm:
bm S02 = 4g SO2/m? x 3.82 x 10-4
Equation 5-4
18 If the absorbance of the blank 1s
within +0.03 absorbance unis of the
intercept and if the measured values
of the internal control samples are
within £0 07 ig/ml SO2 of the actual
value, the analytical values
determined for the field samples can
bbe accepted as valid, If the reagent
blank and/or the control sample falls
outside the above limits, the reagent
blank, the control sample, and the
calibration curve should be checked
by replication to venfy that the above
limits have been met before
reanalyaing the field samples.
5.4 Quality Control Check
of Analysis Procedure
ts recommended that each
laboratory participate in the EPA
rational performance audit program
as an independent quality control
check on the analytical method In
this audit program, each participant
receives five sampies each 6 mo from
the US. EPA Environmental
Monitoring Systems Laboratory. If the
results of these five analyses all fall
within the acceptable range as
published in the EPA survey report,
the analysis procedure 1s considered
to be in control. Otherwise, the
laboratory should take corrective
action and analyze a second set of
samples to determine if the action
taken has corrected the bias or
imprecision of analytical results
VeSecton 2.8 ‘ Jan. 1903
Ara ronan
a
—~Talban marpgans airboee
i : ae
_——_humber. a) ty. ew ay ¥
1 0.0 0.100
2 0.266 12.96 0.9576 0.070156.
a 0378 5,84 —272/6 _0.142884
—1_ 07.36 8.4960 0.348100 _
——5 0.6!) 466.56 75176 O65 T 72) _
i 1.024 629.44
75.64). 3.229 2 1S0BI6. ty 59/84 ry 2243637
—4 _tmumber of calibration ports}
L294NZ 48576
*“Calibration Slope, Intercept, and Correlation Coefficient. The method of least squares is used to calculate a calibration equation mn the form
ot
where
Y= corrected absorbance,
m= slope, absorbance units/ug SOs
1X5 micrograms of SOs
b= y intercept absorbance units)
yemxeb
The slope (mi. intercept (bl. and correlation coeticrent (rare calculated as follows
monkey ex) Oy = 6
nb Y (6) (1568,
546)(3:224). - 0.030050
6)
b -Eysmbx s (3.229) 70.930050) (75:6 = ONS9S
° €
p= gf tay zety/n = {=
By yn
= 0.99998
Figure5.3.
The performance audit program is
coordinated through the 10 EPA.
regional offices by the Quality
Assurance Division of the
Environmental Monitoring Systems
Laboratory. To request further
information about the program or to
participate in the program, write to:
‘Ambient Air Audit Coordinator
Quality Assurance Division
Environmental Monitoring Systems
Laboratory
Research Triangle Park, North
Carolina 27708
A description of the audit material
for S02 analysis is given in Section
2.0.10 of this Handbook.
p= (75,
229) G)
Cteutation of near regression parameters
5.5 Disposal of Mercury-
Containing Solutions
‘The TCM absorbing solution and
any reagents containing mercury
‘compounds must be treated and
disposed of by one of the two
methods discussed below. Both
methods remove greater than 99.99
percent of the mercury.
6.8.1 Method for Forming an
Amalgam 1. Put the waste solution in
‘an uncapped vessel and place it in @
hood.
2. For each liter of waste solution,
add approximately 10 g of sodium
Equation 1
6)(3.229)/()I
carbonate until neutralization has
‘occurred (NaOH may have to be used).
3. Following neutralization, add 10
@ of granular zinc or magnesium.
4. Stir the solution in a hood for 24
hours. Caution must be exercised, as
hydrogen gas is evolved by this.
treatment process.
5, After the 24 hours of stirring, let
the solution stand without further
stirring to allow the mercury amalgam
(solid black material) to settle to the
bottom of the waste receptacle
6. After the settling has occurred,
decant and discard the supernatant
Fiquid.
"7. Quantitatively transfer the solidSection 2.1.6
Jon. 1983
(hv0y) 80) ep Arosenoge) ve'sesnBig
ssyewoy
fy
aumeubis 4
2 uidy “7 40 opis asvenay
luo seupaping yum aoueprasay ut paypuey
EE
HH
tH
fry
uedesd 01 pasn suoHinos 21475 J)
Wepre)
05
dorouning uN
cos - s9ang 500)
EEE CBT |B EOE | OPRT
2
soue | awnjon | owe
qwosay | aw ay Buyduses.
(u014) 607
Be a.
vequinny wis [40a _qucw
odes Fie Bunauwes
peasouy oun 19loug
"SBHAZ AIOMNZOSSection 2.1.8. 6 Jan. 1983
Preparation and Calibration Sample Analysis | a
reagent water double-distiled | Colorimetric analysis Bank Frequency
{and stored n'a borosiicate within 40.03 absorbance ‘of
‘lass vessel with vert guard | calibration intercept. control
Reagent Container Preparation,
sunimieaot, 0.6% Glass soopered Before analyts
Formaldehyde. 0.2% Glass stoppered Before anslysis
Sample absorbance within
£1 yg SO, of actual value
Absorbing reagent prepared Working iodine, "Glass stoppered, Before analysis
every 6 mo, clear with 30 < pH O1n ‘dark
<50 ‘Stock iodine. Gass stoppered, Yearly
‘O.01N ‘ark
24-h sampler with polypropylene | Calirauon curve for ab- Sodium thiosulfate, Glass stoppered Yearly
lubes and closure with glass | sorbance vs. ug SOx/m ‘oon
‘mpinger, % and I-h sample with | slope Between 0.03 20.002 | Stock sodium Glass stoppered Yearly
laih glass maget impinger ‘absorbance units/ug SOs thiosulfate,
| imercept <0.17, allponts | OIN
‘against specihicauon. S04) of best-fit curve Working sullite-TCM solution accurately prepared from standard
Solution prior to analysis, stable or 30 days at 8°C (41°F) if stored in
All data points within +2% of the dark glass-stoppered bottle
best-ta calibration curve for How. |
‘measuring rotameter ‘Standard sulfite-ICM solution 320-400 pg SO/ml
Flow rate measured before and |
Pararosanuline (PRA) reagent 100% assay: stable if stored in dark
‘sher each sampling period
‘g/ass-stoppered bottle
Rotameter cleaned and calibrated
every 30 days of operation or
every 6 mo
Assayed PRA stock solution stored in dark glass-stoppered bottle:
Solution prepared it slope of calibration curve deviates
from 0.03 0.002
Sample umer within 18 min/ Analytical balance performance within 20.1 mg of manufacterer's
speciheauons
|
|
|
|
|
Check of flow-measuring device | within 0.04 wg SO2/ml (1 ug |
Figure5.4B8 Laboratory data log fbackside)
material to @ container and allow it to
ary.
8, The solid material can be sent to
a mercury reclaiming plant. It must
not be discarded
5.5.2. Method Using Aluminum Foil
Sirpps 1. Put the waste solution in an
luncapped vessel and place it im a
hood,
2. For each liter of waste solution,
add approximately 10 g of aluminum
{oil strips. fall the aluminum appears
10 be consumed and no gas is
evolved, add an additional 10 g of foil
Repeat until the foil is no longer
consumed and allow the gas to evolve
for 24 hours.
3. Decant the supernatant liquid
and discard,
4, Transfer to a storage container
the elemental mercury that has
settled to the bottom of the vessel
5. The mercury can be sent to a
mercury reclaiming plant. It must not
be discarded.Jan, 1983 7
Table5.2 Activity Matrix for Analysis of Samples
@ Activity
Section 2.1.6
Action if
Frequency and method requirements
Acceptance limits of measurement are not met
Verification of field
record form
Flow rate check of system
Calibration of
spectrophotometer
Reagent blank and internal
control standard
All vital information
obtained and verified
Ditterence in initial and
final flow rates after
‘sampling <5%
Each calibration curve:
Slope 0.03 +0.002 ab-
‘sorbance unit/ug SOx
intercept <0.17 absorbance
units: each correlation
coefficient >0.998
Analysis in accordance
with Subsec. § 3; absorb-
ance of blank within £0.03
absorbance units of zero:
‘measured value of each
internal control standerd
within 1 ug SO2 actual
value
Visually inspect after each
sampling period.
Repeat flow rate measure
‘ment after sampling.
Check the calibration
curve atter sampling
Analyze a reagent blank
‘and internal control
‘standard before the first
samples, atter every sub:
Sequent 10th samples and
after the last sample
Obtain missing data or
invalidate the sample.
Invalidate the sample.
Repeat calibration and/or
standardize the sulfite
solution, and check dye:
‘check water bath temp.
and/or check chemicals
used to prepare reagents:
repeat reading for that
point; average new and
original values; replat
Take corrective action
(recalibrate it necessary)
and reanalyze the last 10
samples.Jan. 1983 1 Section 2.1.6
6.0 Data Reduction, Validation, and Reporting
‘The quality control activities are
summarized in Table 6.1 at the end of
this section
6.1 Flow Rate Data Review
Samples should be invalidated if the
difference between the initial and
final flow rate is greater than 5
percent
6.2 S02 Data Review
The analytical data should be
invalidated if any of the following
Conditions exist regarding SO,
calibration.
1. Fresh reagents are not used.
2. The internal control samples
were analyzed, and the results
deviated by more than +1 1g SOz
from the known value, but a new
calibration curve was not constructed.
The slope of the final calibration
curve was nat within 0.03 £0,002
absorbance units/ug SO2.
4. The intercept of the final
calibration curve was not <0.17 at
22°C (72°F) with a 1-cm optical path
length, and/or the blank was not
within +0.03 of the regression
intercept.
5. The correlation coefficient of the
least square regression was <0.998
6.3 Standardize Format
Reporting
The standardized procedure for
reporting is as follows
1. Record the 24-h sample
analytical results on the SAROAD
daily data form (Figure 6.1),
2. Record the 30-min or the 1-h
‘sample analytical results on the
‘SAROAD hourly data form (Figure
62)
3. Refer to EPA-450/2-76-029,
Dec. 1976, OAQPS Guidelines,
AEROS Manual Serves, Volume I
‘AEROS User's Manual, U.S, EPA
Office of Air and Waste Management,
Research Triangle Park, North
Carolina.Section 2.1.6 2 Jan. 1983
24-Hour or Greater Sampling Interval
21 Smelter own Pollution Control r )
1 gener
Site Area Ste
—pmelter Town __________ (arg oralolo lols
£00 Worth Street ger Frye © re aon
Bae Aadress
802 Network _2# ovr (P] {olz| [7] [7[4} [o[z)
Fare Tine interval 7 oR 116 17 Te
SO — _
ome Tame Tame | ware
pananieren panamerer pananrer panainerer
Code Code cove ove
° |
4izwiol7] |_LLLLL) LTT)
232475 26-27 3 38 3940 7 Penne |e wee
7A (17) fo] (LLILOO }
2 wa 2 |\fe www @ wn RT
19 20.21 22. "3934 38 96. 4748.49.60 “8 16 77-78
oT 1o}o) felatz{s E |
2lolol felalzié | | |
fofet T | | “TI |
+t |
Pots{ | I { | {
BAS ood | EES
ofsl | [ | + +— |
oe | | 4 | 1
TotW] i CI Cen
tft] fi | —— r+ |
bo2zt4 - J | i
Crest] lea al
14 } | t |
i \ { 1 i
i{s i peal I 4 |
tte | | | oe
rz 1 iL
Cre 1 | CPT
rps | el]
rete | A C
ait t | —|
Be I
ts i | I
io ee
2.5 y Ly yyy |
216 a
lett Tf 7 [EC
zfe|_| |
am | Z| 7 C
7 { i
o- 204892 1 O aoe sret 43270 o3210
Figure6 1 SAROAD doy data ormSection 2.1.6
Jan, 1983,
Ze |Aola\e loll olo|o[olalololo|ale/olo\o
fe [ele[o ol jojolojojolols|o ojala] oe °
zitoy | on
4
mp Ninoy avOWWS Fg einbLy
E|O Oo) 4|0|AAO) a|A|A| O[0/0] 3] 10/0) @
7[o[919[ 7[O[a|0] 7fo|o|o} 7/9] efofolajojojalojojojolo °
ZZ
pu |
01 Spy 6 py 8 Spy 9 Spy g Spy Spy |g bpy 2 Spy 1 Opy aH as! keg
ze 1 of 6k 8 28 SE be Ee
e a0, sw __s90 1p ours ons
Zz] 412] BI) Vielzlz|4 Wed Apaoy Yao aH TOD
RoR Gar lemme rower __pauseye sntureg ssompy 015
archon shes —_ 2 _
ia FUG OF Fees GH Oe
V/\ [slZ] (4 [Z] : : fe owen h9
aceite meng oor) cet ” ee
O16 o£ 9 yueg e10g 2naWi0.y /euoney Asueby
7je|efolo AINIDY NOUD310Ud TYINIMNOIANT BAG Tey] au
ais
ensewuy Bunctues non-ye veuL S507Section 2.1.6 4 Jan. 1983
6.4 Filing the Calculat
1. Identify all of the sample
calculations with the sample numbers
and dates.
2. File the calculations in the
laboratory data log book.
Table6.1 Activity Matrix for Data Reduction, Validation, and Reporting
Action if
Frequency and method requirements
Activity _ Acceptance limits ____ of measurement _ are not met
Flow dota Deviation in flow rates Visually observe reported Void the sample
less than 5% data
S02 data Five conditions of Sub As above As above
sec 6.2 met
Documentation and report All needed data available As above As above
of results.Jan, 1983
7.1. The 24-h Bubbler Train
Every 3 months or every 15
sampling periods, whichever occurs
first, the following components should
be replaced
1. Membrane filter - Replace the
membrane filter by removing the
Tygon tubing from the inlet and outlet
of the plastic fiter holder (The fiter
and filter holder are disposable ) Seal
the new filter holder to prevent
leakage by pulling the filter holder
apart and coating the junction of
the two halves with a thin layer of
cyclohexanone and by rejoining the
two halves and pressing them
together until they are dry Connect
the new filter in place
2 Mist trap ~ Change the trap by
removing the Tygon tubing and
replacing it with a clean dry trap
Disassemble the trap that was,
removed, wash it, and allow it to dry
before reassembling and using it
again
3. Rubber septa - Remove the old
septa from the glass adapters, and
replace them with new septa
4. Vacuum gauge - Check the
vacuum gauge against a calibrated
vacuum guage or a mercury
manometer. Replace the guage if its
defective, (not within +25 mm (1
in) Hg).
5 Vacuum tubing - Replace the
black vacuum tubing with @ new
section to minimize the possibility of
leakage.
7.0 Maintenance
6 Vacuum pump - If the vacuum
pump cannot provide >530 mm (>21
in) Hg vacuum at 760 mm (29.92 in)
atmospheric pressure, replace ts
carbon vanes or its diaphragm. At an
atmospheric pressure of 760 mm, the
530 mm (21 inj Hg vacuum provides
a safety margin of 100 mm (3.9 in)
Hg vacuum
7.2. The Short-Term
Sampling Train
Every 3 months or every 30,
sampling periods, whichever occurs
Section 2.1.7,
7.3 Calibration Equipment
and Related Apparatus
Periodic maintenance requirements
{for equipment such as the mass -flow
meters, soap-bubble meters, and
spectrophotometers vary with the
specific manufacturer. Thus, the
manufacturer's recommendations
should be followed as well as one’s
own experience.
first, the following components should
be serviced
1 Rotameters - Disassemble and
clean the rotameters with detergent,
and rinse with distilled water and
alcohol Recalibrate before use
2 Mist trap - Clean the trap, and
replace the glass wool
3 Needle valve - Inspect the valve
{for proper seating of the stem, check
the vaive-siem packing nut for proper
tension, and clean the valve with
detergent Rinse with distilled water
and alcoho!
4. Vacuum pump - Perform the
same maintenance as that described
tn Subsection 7 1
5. Vacuum tubing - Perform the
same maintenance as that described
in Subsection 71
Table 7.1 Activity Matrix for Regular and Preventive Maintenance
Action if
Frequency and method requirements
Equipment Acceptance limits ‘of measurement re not met _
24-h bubbler train ‘Membrane filter. mist rap. Replace every 3 moor after Implement maintenance
rubber septa, and vacuum — 18 sampling periods. actions.
tubing replaced regularly
Vacuum gauge
Vacuum pump
Readings correct within
426-4 mm (41 in.) Hg
Capability to provide at
Check every 3 mo against
calibrated vacuum gauge
Replace it defective.
or Hg manometer.
Teast 530 mm (21 in)
Hg vacuum at STP.
Short-term sampling train
Rotameter, mist trap
needle valve, vacuum
‘pump, and vacuum ser-
viced regularly.
Calibration equipment and
related apparatus
Maintained periodically
according t0 manufac:
turer's recommendations.
Test the vacuum.
Clean every 3 mo or after
30 sampling periods.
Follow manufacturer's
recommendations.
Replace carbon vanes or
diaphragm: retest.
Implement maintenance
‘actions.
Implement maintenance
‘actions,Jan. 1983,
1 Section 21.8
‘An audit is an independent
assessment of the accuracy of data,
Independence is achieved by having
the audit made by an operator other
than the one conducting the routine
‘measurements and by using audit
‘standards and equipment different
trom those routinely used in
‘monitoring,
The audit should be a true
assessment of the measurement
process under normal operations—
that is, without any special
preparation or adjustment of the
system. Routine quality assurance
checks conducted by the operator are
necessary for obtaining and reporting
{good quality data, but they are not
Considered part of the auditing
procedure.
Two types of audits are
recommended herein—performance
audits and system audits. Three
performance audits and one system
audit are detailed in Subsections 8.1
and 8.2 and summarized in Table 8.2
(at the end of this section}. In addition
to these audits, the precision of the
entire measurement process is
assessed by comparing the
‘measurements made by collocated
samplers, as described in Sections
2.0.9 and 2.1.8.
Proper implementation of an
auditing program serves a twofold
purpose: to ensure the integrity of the
data and to assess the accuracy of the
data. The technique for estimating the
accuracy of the data is given in
Section 2.08 of this volume.
8.1 Performance Audits
Performance audits conducted by
another operator/analyst are
quantitative evaluations of the quality
of data produced by the total
‘measurement system (sample
collection, sample analysis, and data
processing). The following three
performance audits of individual
variables are recommended:
1. Audit of flow rate for sample
collection.
2. Audit of analysis process by use
of reference samples (mandatory).
3. Audit of data processing. (Refer
to Appendix A for auditing frequency.)
Auditing of 7 of 100 sampling
periods for each site is suggested as a
Starting frequency for the first and
third audits; frequency for the second
audit is given in Subsection 8.1.2
8.0 Auditing Procedure
Where one sample is collected every
6th day, one audit per month at each
site is recommended: if the data are
reported quarterly, this would mean
auditing 3 of each 18 sampling
periods. if the number of sampling
periods 1s >16 but <0, four audits
are recommended. These are the
suggested starting frequencies: they
should be altered if either experience
fr data quality indicate a need for
change. For example, the frequency
should be reduced if experience
indicates that data are of good quality;
the opposite would apply if the data
were of poor quality
In a determination of the number of
audits needed, itis more important to
be sure that the audit sample is
representative of the various
conditions that may influence the data
{quality than to adhere to a fixed
frequency. The supervisor will specify
the frequencies according to
‘mointoring requirements,
8.1.1 Flow Rate Audit - For 30-mit
1-h, and 24-h samples, a flow rate
audit is recommended to assess the
sampling collection phase of the
‘measurement process. The audit
should be performed as follows.
1. Have the regular operator
the sampler for sample
‘usual; this must include
filling in the sample record form
Figure 4.10)
2. Have the operator compute the
average sampled volume, Viv aS
measured by the regular flow
measurement device and corrected to
reference conditions (if necessary)
3. Insert the audit device in the
‘sample inlet line and measure the
flow in the usual manner at the
beginning and at the end of the
sampling period.
4. Calculate the average audited
volume, Vat reference conditions for
liters/minute and t is the sampling
time in minutes).
5. Compute the percent difference,
(a measure of inaccuracy):
d=(Vn=Va) 100.
ve
Equation 8-1
6. Record Vin. Vs, and d. tis
recommended that the d's be plotted
fon the X-and-R chart as illustrated in
Figure 8.1. (Only the X chart is used
for quality control of accuracy, the
standard X-and-R chart 1s used for
convenience.) If d is greater than 7%
oF if Vn/ Va does not fall between 0.93
and 1.07, begin the troubleshooting
and take corrective action before
resuming the sampling, All flow rate
data for the audit period is invalidated
if an unexplained malfunction has
occurred and resulted in errors in the
flow rates,
8.1.2 Audit of Analysis Process
Using Reference Samples - Reference
(audit) samples should be included at
ramdom at the recommended
frequency among the samples
awaiting analysis. If possible, they
should not be recognizable to the
analyst as reference samples. The
‘audit procedure for the SOz method is
2s follows
1. Prepare the audit solutions from
‘2 working sulfite-TCM solution, as
described previously in Subsection
3.12 of Section 2.1.3. Prepare these
‘audit samples independently from the
standardized sulfite-TCM solution
Used in the routine analysis
procedure. Prepare new sulfite-TCM
‘audit samples every 30 days, and
‘store them between 0° and °C.
2. Prepare the audit samples in
‘each of the concentration ranges of
0.2 100.3, 0.5 100.6, and 0.8 0.0.9
19 SO2/mi.* Analyze an audit sample
in each of the three ranges at least
‘once each day that samples are
analyzed and at least twice per
calendar quarter. The differences
between the measured concentrations
(Vg $O2/mi) and the audit
concentrations are used to calculate
the percent difference (a measure of
inaccuracy! as described in Section
2.0.8. The calculation for percent
difference for individual audit samples
is also shown here at Item 3.
3. The agency/organization
determines the percent difference (d)
between the measured SOs
concentration and the audit or known
Sis same acjurrmant eNovié be applied al
Corterponding ara in ths sectonJan. 1983
Section 2.1.8
(ae
‘uonay
3201109)
sou:
ro
{|
|
|
uy sebu0y
Wie Ie ar za ua .
_ bexlzcpa-bsspozlaox|zor/levoez|nelevr|s zee léect 7 ue
= a | | X Wesay
an! 4 — — — —- i cat
_ - _ |_| ie | : ‘J
soe] czpeE roe oe weleerrerleerigerver\Zér| 7e|7#|2|2 | F
LC —_ 7 a sep cr ip sels lade rarer Por| of ros! x bus|ser|z0c| | * @
C ase J [| [ { ee
| | CI ale
t+ Po ez jaeperha [i
_ { Bees [Ee re 17e ke th Peele Peep
Gonusosnseay,
ea y pe x
| % 7
27 Sn SBW ATO BL PIPE POITELIW i BDU STEHT peu —————opny area
sutra y mi suns CUPL V ET ar gt eta oor ee men ‘peteJan. 1983
values of concentration. The dis @
measure of both the bias and the
random error of the analytical phase
of the S02 reference method.
Calculate d by using Equation 8-2,
See Cree A
a)
Csoe
Equation 8-2
where
Cao, (M)= concentration measured
by the lab analyst ug/ml
C20, (A)= audit or known
concentration of the audit
sample, ug/ml
The recommended contro! limits for
the three audit sample ranges (0.2 10
0.3, 0.5 10.06, and 0.8 100.9 ug
'$02/mi) are the 90th percentile value
for d based on the results of five
audits (11/76, 4/77, 10/77, 4/78,
and 10/78) performed by the
Environmental Monitoring Systems
Laboratory, USEPA, Research Triangle
Park, North Carolina.” "° By definition,
90 percent of the laboratory
participants in the audit obtained
values of d less than the values
tabulated below. The control limits
axpected to be exceeded by 10
percent of the laboratories to be
audited, based on these five audits.
The 90th percentile values and the
known audit concentrations are given
below for each audit concentration,
range.
3 Section 2.1.8
Based on the results or these five
audits, the recommended 90th
Percentile control limits for audit
‘samples are +33 percent for the 0.2
t0 0.3 ug SO2/ml concentration range
and +18 percent for both the 0.5 to
06 and the 0.8 10 0.9 ug SOz/m!
concentration ranges.
itis recommended that each
laboratory participate in the EPA
National Performance Audit program
as described in Section 2.1.5,
Subsection 5.4. This will serve as an
independent audit of the analysis
technique.
8.1.3 Data Processing Audit - The
data processing audit is most
conveniently performed soon after the
original calculations have been
completed, so that corrections can be
made immediately and additional
explanatory data can be retrieved from
ield personnel when necessary. The
procedural stops are as follows.
1. The audit must be performed by
an individual other than the one who
originally reduced the data. The check
should start with the raw date and
conti ‘ough the recording of the
concentration (ug SOz/m) on the
‘SAROAD form.
2. if the mass concentration of SOz
computed by an audit does not agree
with the original or indicated value
within round-off error, the
calculations for all samples collected
since the previous audit should be
checked and corrected. The audit
value is always reported as the
0.2 10.0.3 yg SOz/mi _
‘Known audit
concentration 90th percentile ford.
Audit date ug SO_/ml %
11/76 0.259 33
477 0.289 32
10/77 0.338 35
0.5 100.6 yg SO/mi
‘Known audit
concentration 90th percentile ford.
Audit date SOo/ml %
11/76 0.617 24
“7 0.688 17
10/77 0.562 18
10/77 0.562 duplicates 17
10/78 0.533 18
0.8 100.9 yg SO/mt
‘Known audit
concentration 90thpercentile ord.
Audit date. 'S0;/ml %
11/76 0.895 18
4/77 0.930 23
10/77 1.07 16
4/78 0.755 7
10/78 0.757 7
correct value, under the assumption
that the audit should be checked if a
discrepancy exists and corrected if
necessary.
3. The audit and the original values
and the percentage difference should
bbe recorded in the laboratory log book
and reported to the supervisor.
8.2 System Audit
A system audit is an onsite
inspection and review of the quality
assurance system used for the total
‘measurement system (sample
collection, sample analysis, data
processing, etc.) A system audit is a
‘qualitative appraisal of system quality.
‘A system audit should be conducted
at the startup of a new monitoring
system and thereafter as appropriate
to ascertain significant changes in
system operations. A checklist for use
in a system audit is provided in Figure
8.2. These questions should be
reviewed for their applicability to the
particular local, State, or Federal
agency being auditedSection 2.1.8 4 Jan, 1983
1. What type of manual sampler 1s used for SOs collection?
(a) BAC @DSport (c)2-port (4 other — =
2 Ie sraight tube mpingr used nthe pohpronyione sampling tube? eg b=
3 15 sample probe made of accepted materia? Teflon, Glass Pe
4 le the roe ond mano) cated pig mag urggardeneyon whe vambling n hun ongtone age
5 Woot imaiod of salen sis lnc eget Ina copy avatatesggaane (Mei should be om 40
CER 50, July 1, 1982. as amended 47 FR 54896, Decerhber 6, 1982 and 48 FR 17955, Apri, 1963
6
i
Wihet quality of reagents is sed 'o makeup the ehenveos forthe Sr aelyan? Legale
{they Shut be reagent grade ber
2. Is the sampling train routinely checked for leaks? — :
tow otra) Once a weet, (6) Gnce roth, (Mone a cvarter (aot sana,
8. What calration check procedure is used for flow measuring device used i the Held? 72.1.2 ID
Isa writen copy avarlable?
9. Are sampling low rates detefmnined before and after each sampling period? Is thera an established procedys for
isang needs that are not wth flow ae nts ical? aglaw at eats? LAB 2 (any
needle outs of 180 t0 200 e”m/min should be discarded for 24% samples) Is the absorbing solution mamntained af 15° 1
Bee Are samples vere? gta Do nal and tinal How rte agree within = agama iors sample they 70
me)
10, Have calibration curves been made, ad are they avalable? You might also like