• Double click SmartSEM icon on the desktop to start the software
• Enter your Username (first initial, last name) and Password (7 digit student ID), click Ok.
Load or Exchange Sample (See the SOP for sample changing for details)
• Click the “Beam Stage” drop down box and select “Beam Off”
• Load samples.
• Move the sample into position using the Z stage navigation joystick
• Center the sample under the objective lens using either the XY stage navigation knob or
alternatively double clicking the sample number on the carousel thematic
• Click the “Sigal A” drop down box and choose “SE1” to enable the Secondary Electron
Detector
• Click the “Beam State” drop down box and select “Beam On”
• Set Magnification to lower value using knob on Keyboard Console (tip: click the “MagWD”
button in the top tool bar to show the Mag and WD values in the status bar)
• Click the “Levels” button in the tool bar to show the Brightness and Contrast values in the
status bar.
• Use Contrast as “gain” control to adjust signal for the best image.
• Adjust focus using Focus knob on Keyboard Console to obtain coarse focus.
• Adjust stigmatism
• Change WD, HV or I-Probe if necessary and repeat above steps to improve resolution
• Wait for Red Dot in lower right corner of image to appear indicating a Frozen Image
—Method 2
• Default setting with scan speed 8, Line Int., and N=4 will be used for imaging
• Speed and N can be manually changed and the scanning will restart
• Wait for Red Dot in lower right corner of image to appear indicating a Frozen Image
—Method 3
• Default setting with scan speed 6, Line Int., and N=4 will be used for imaging
• Speed and N can be manually changed and the scanning will restart
• Wait for Red Dot in lower right corner of image to appear indicating a Frozen Image
—Method 4
• From Photo drop down menu, select Frame Average
• Default setting with scan speed 4, Frame Int., and N=30 will be used for imaging
• Speed and N can be manually changed and the scanning will restart
• Wait for Red Dot in lower right corner of image to appear indicating a Frozen Image
• File>Save Image
• Images are stored in the User’s “Image’ folder on the SEM PC.
To Finish
• Turn “Beam Off”
• Turn on camera
• Vent chamber
• Remove sample
• Pump chamber