SATELEC Scaler System Components:

Scaler Tubing:
(153765-001, 005) Procenter Delivery Unit Components ...... 1 1/8 " Clear ........ AR
Installation Instructions Pressure Swt. .....
Check Valve ........
1
1
1/4" Clear ......... AR
Fittings:
Washer ............... 1 1/16" barbed T ... 1
Intended Use: Wire Connector ... 1 Uni-Clamp 1/8" ... 5
Velcro, Loop ........ 1 1/16" to 1/8"
To provide a means for a dental hygentist to remove
Velcro, Hook ...... 1 barbed ............... 2
plaque and build-up on patient’s teeth.
Bracket ............... 1 1/16" barbed
Cover ................. 1 plug .................... 2
Requirements / Restrictions: Screws ............... 4
None Special Tools:
none

STANDARD HANDPIECE TUBING REMOVAL


Note
Determine which current handpiece position
will be replaced with the SATELEC Scaler
System. Ask unit owner to indicate preferred
placement as Scaler will be connected to the
valve at that location.

Step 1: Position delivery unit cover Step 3: Disconnect 1/4" handpiece


in fully opened position. tubing from 1/4" clear tubing.

Step 4: Disconnect 1/4" handpiece


tubing from connector; then
remove handpiece tubing
from delivery unit.

Step 2: Disconnect 1/8"


handpiece tubing.

Kink
Step 5: Disconnect 1/8" Valve
green tubing.

KA975501i

Style A 003-1672-00 Rev. CA1 (3/13/18)


© Midmark Corporation 2006 1
SCALER COMPONENTS INSTALLATION Step 8: Position Pressure Switch in
delivery unit housing so that one
of the leads can be connected to
the terminal board & one to the
Grey lead.

Step 7: Insert scaler handpiece tubing up


thru hole in bottomof delivery unit
housing. Tie off the restraining
string to a hole in the hinge.

Step 3: Feed black and grey wires from


Step 4: Assure indicator on potentiometer Potentiometer connector, up thru
wire hole on bottom of delivery
knob is set so it points straight up unit housing.
when turned all the way counterclockwise;
then install washer, and nut on shaft.

Indicator
Step 2: Install scaler holder bracket
10
on bottom side of delivery
9 1
2 unit housing, secure with
8
7
6 5 4
3 two screws.

Step 1: Install controller on mounting


Step 6: Position decal, that indicates inten- 9
8
10
1
2 bracket using velcro strips.
sity settings, so indicator on knob 7
6 5 4
3

is in the space between #1 and #10


settings then secure decal, and
knob. Step 5: Install cover on scaler holder
bracket, two screws.

© Midmark Corporation 2006 2


SCALER TUBING CONNECTIONS
Note
Install 1/4" sleeve clamps and 1/8" uni-clamps onto
comparable tubing before making connections.
Cut a 2" length of 1/8" clear tubing from tubing
supplied in kit for use in connections. Scaler
Exhaust Handpiece
Air Tubing

Step 4: Connect Pressure Switch to T-fitting 1/4"


with 1/4" Clear tubing, 1/4" - 1/8" Clear
Reducer, 2" length of 1/8" Clear tubing,
uni-clamps and sleeve clamps.
Reducer
1/4" to 1/8"
Reducer
1/8" 1/4" 1/4" to 1/8"
Clear Clear

1/8"
Pressure Clear
Check
Switch
Valve Step 1: Plug off end of Exhaust Air
1/8" tubing using a 1/4" - 1/8"
Clear Plug Reducer, 2" length of 1/8"
Clear tubing, uni-clamps,
To sleeve clamps, and a plug
Pressure
Gauge 1/4" 1/8"
Clear Clear

A
Step 2: Plug end of Coolant Air (1/8" Step 3: Connect 1/8" Clear tubing from
Green) tubing using a 1/8" plug
and uni-clamp. Scaler Handpiece to Kink Valve
Kink (A). Use uni-clamp to secure.
Valve

Coolant Air Plug


1/8" Grn.
KA975701i

© Midmark Corporation 2006 3


SCALER ELECTRICAL CONNECTIONS
PROCENTER

Scaler
Handpiece

1/8" Clear
Terminal Tubing to
Board Kink Valve
Note
Loop each wire thru
EMC Filter once.
EMC Black
24V Filter
Wire
From Grey Connector
0V
Potentiometer
Connector on
Controller
White
White Pressure
Switch

Black

Grey

Midmark Corporation
Potentiometer
60 Vista Drive From Scaler
Versailles, OH 45380-0286
Controller
Handpiece
Phone: 937-526-3662
Fax: 937-526-5542 KA975802i
midmark.com

© Midmark Corporation 2006 4