Professional Documents
Culture Documents
Human Humastar 180 - Service Manual PDF
Human Humastar 180 - Service Manual PDF
| Service Manual
SYSTEM VERSION
COPYRIGHT
No part of this documentation may be reproduced in any form, nor processed, copied or distribut-
ed by means of electronic systems, without prior permission of Human GmbH in writing. Since all
precautionary measures were taken into account in producing these operating instructions, the
manufacturer accepts no responsibility for any errors or omissions. This includes any liability for
damage that could arise from possible incorrect operation based on this information. Subject to
changes without notice as result of technical development.
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS 5
1.1 INTRODUCTION 5
1.2 USER WARRANTY 5
1.3 INTENDED USE OF THE INSTRUMENT 5
1.4 GENERAL SAFETY WARNINGS 6
1.5 DISPOSAL MANAGEMENT CONCEPT 6
1.6 BIOHAZARD WARNING 7
1.7 INSTRUMENT DISINFECTION 7
2 SYSTEM INTRODUCTION 11
2.1 INTRODUCTION 11
2.2 CHARACTERISTICS 11
2.3 SPECIFICATIONS 12
2.4 FUNCTIONAL DESCRIPTION 13
2.4.1 Process Management 13
2.4.2 Liquid Handling 14
2.4.3 Carousels, Temperature Control and Cooling 17
2.4.4 Measurement System 20
3 ELECTRONICS 23
3.1 INTRODUCTION 23
3.2 SCHEMATIC DIAGRAM OF ELECTRONICS 23
5 MAINTENANCE 33
5.1 INTRODUCTION 33
5.2 SYRINGES 33
5.3 SYRINGE VALVES 33
5.4 TUBINGS 33
5.5 PHOTOMETER 34
5.6 X-DRIVE 34
5.7 Z-DRIVES 34
5.8 LIQUID DETECTION 35
5.9 MAGNETIC VALVES 35
5.10 PROBES 35
5.11 MEMBRANE PUMPS 35
6 TESTS/TROUBLESHOOTING 37
6.1 INTRODUCTION 37
6.2 FUNCTIONAL TESTS 37
6.3 TROUBLESHOOTING 39
6.3.1 Reagent Rotor 39
6.3.2 Cuvette Rotor 40
6.3.3 Belt 42
6.3.4 Pipettor 44
6.3.5 Liquid Detection 45
6.3.6 Liquid Flood 47
6.3.7 Communication 48
7 REPLACEMENT 49
7.1 INTRODUCTION 49
7.2 CAROUSEL UNIT WITH NEW REAGENTROTORSENSOR 49
7.3 CAROUSEL UNIT WITH OLD REAGENTROTORSENSOR
(POTENTIOMETER) 54
7.4 CENTERING MAGNET 59
7.5 COVER 64
7.6 CUVETTE DRIVE MOTOR ASSEMBLY 67
7.7 CUVETTE ROTOR BOARD 71
7.8 MASTER BOARD 72
7.9 PHOTOMETER 74
7.10 PIPETTOR ARM 76
CONTENTS
7.11 PROBES 80
7.12 REAGENT ROTOR CONTROL BOARD 82
7.13 REAGENT ROTOR POTENTIOMETER 83
7.14 SYRINGE PUMPS 89
7.15 WASHING PUMPS 92
8 ADDITIONAL INFORMATIONS 95
8.1 INTRODUCTION 95
8.2 PIPETTOR ARM 95
SAFETY INSTRUCTIONS 5
1 SAFETY INSTRUCTIONS
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the ”GENERAL SAFETY WARNINGS”, describing the appropriate operating pro-
cedures. Please contact your HUMAN authorised local Technical Service in the
event of instrument failure or other difficulties with the instrument.
Batteries, power packs and similar power sources must be removed from
electric/electronic parts and disposed of in accordance with applicable local
regulations.
FIGURE 1
Biological Hazard
Symbol
Notes:
10
2 SYSTEM INTRODUCTION
- Characteristics, Specifications & Performance
- Functional Description
2.1 Introduction
HumaStar 180 is a fully automated, random access Clinical Chemistry Analyser
for wet chemistry applications. It has an open structure, so that reagents from
most manufacturers can be used on the system.
HumaStar 180 is computer controlled and uses the Software Package „DETect“
for data reduction and result evaluation.
HumaStar 180 has been developed in a way as to maximise throughput at
maximum comfort, and at the same time to minimise maintenance efforts to be
taken by the user.
HumaStar 180 is fitted with a patented, maintenance free measuring street,
featuring 20 photometer channels for individual wavelength selection and
maximum flexibility.
2.2 Characteristics
2.3 Specifications
TABLE 1
Throughput: - around 150 Tests/hr under speed – optimised
conditions
Carousel: - 25 reagent positions (with SW update up too
50 positions)
- 25 sample positions (random access)
- 25 serum cup positions (minimum 300μl
volume)
- 300 test positions (75x4 cuvettes)
- Customized carousels available !
Liquid Handling: - 2 independent Syringes:
Sample: 1 - 200μl in 1μl steps
Reagent: 1 - 900μl in 1μl steps
- CV less 2,0% for Volumes > 10μl
- High pressure washing
- Autocleaning
- Backflush function
- Easy maintenance
Measurement System: - Maintenance Free !
- Light sources: LED
- Detectors: Photodiodes
- Temperature controlled
- 5 inbuilt photometers, 4 channels each
- 20 independent channels
- up to 20 different wavelegths at a time
- Wavelength range: 340 – 900 nm
- Standard filters:
- 340, 405, 505, 546 & 620nm (4 each)
- Resolution: 0,0001 ABS
- Accuracy: <1% CV at 1ABS / 405nm
- Linearity: 0.0000 – 3.0000 OD better than +/- 1%
- Drift: less than 0.001 OD / hr
Temperature Control: - Reagent cooling (ambient minus 10°C)
- (water circulation w/ external cooler)
- Cuvette heating (electrical) in carousel and
- reading path: 37°C
- Liquid heating (electrical) in probes
- Process Management
- Liquid Handling
- Carousels
- Temperature control
- Cooling (optional)
- Measurement System
- DETect Software package
The following explains the functions of all these units, as well as their
interference within eachother.
Each module is fitted with it’s own “intelligence” and functions as a slave to the Process
Manager. The process manager is connected to a PC via RS 232 and receives his orders from DETect
Software package running on the PC. When a process is started on DETect (e.g.: testing a sample
for certain paramters), DETect cuts the complete process into the required steps:
These process steps are sent to the process manager in the instrument, one at a time, exactly
whenappropriate. The process manager himself distributes the necessary commands to the
modules. The modules “fulfill the order” and report back to the process manager, which himself
provides DETect with the actual status accordingly.
FIGURE 2
Liquid Path
Z Movement Valve Positions:
VP a
X Movement
3
VP b
6 4
VP c
13 11 5 2
10 7
9 8
12
Waste System
14 Liquid 1
Explanation:
Both liquid paths are built up identical, with the only difference of the size of the
syringes and the probes:
Syringe: 1ml for Reagent path Probe: 0.5 mm inner diameter for Reagent path
250 μl for Sample path 0.4 mm inner diameter for Sample path
X- and Z – movement are realised by the pipetting arm, which holds the
probes and the probes drives, together with all necessary motors, sensors and
electronics.
Wash probe
- move probe in Z direction into wash station
- flush with System liquid for a pre-defined time span
- lift probe in Z direction to top position
REAGENT Pipetting:
Pipette liquid into target container
- move probe in X direction to target container (Cuvette)
- move probe in Z direction into target container
- pipet required test volume into target container
- lift probe in Z direction to top position
SAMPLE Pipetting:
Pipette liquid into target container
- move probe in X direction to target container (Cuvette)
- move probe in Z direction into target container until liquid is detected
- pipet required test volume into target container
- mix by aspirating and re-pipetting
- lift probe in Z direction to top position
FIGURE 3
Carousels, loaded, inside
Instrument
Inner
max 10,5 mm
FIGURE 8
Cuvette Block
FIGURE 7
Outer Carousel
The Outer Carousel features Each Cuvette Block contains 4 Cuvettes, for
75 Positions for Cuvette Blocks one test each.
Total loading volume: Cuvettes for 300
Tests. These Cuvette Blocks are unique
and can only be ordered from Human:
Cat number 16900/1
Insulation
In order to prevent temperature – interferences between the (optionally) cooled
inner and the temp. controlled outer carousel, an insulation ring is dividing the
two carousels.
Rotation
Both Carousels are controlled and rotated independently, within a co-axial
mechanical system.
Besides keeping samples, reagents and test containers in stable temperature
environment, actually the main function of the carousels is to provide the
Y – movement of the source and target containers within the pipetting and
measuring system of HumaStar 180.
The carousels are rotated while other processes within the complete system are
being performed. By doing so, the time required for rotating and positioning the
carousels is not a critical factor in terms of process time and throughput.
These 20 channels are organised in 5 groups of 4 channels each, whereas each group of 4 channels
is controlled by two electronic boards: One light emitting board and one detection board. These
are directly connected to the central Process Manager
Each of the 5 light emitting boards is fitted with 4 light sources (LED, light emitting diodes) and
the respective control electronics.
Each of the 5 detection boards is fitted with 4 detectors (Silicon Diodes) and the respective signal conver-
ter electronics, as well as with a RS 232 port which connects it with the Process Manager.
system
Cuvette
Transport
Motor
Detection board
Detector
Optics (lens)
Light beam
Optics (lens)
For reading, the cuvette block is transported from the outer carousel through the
linear photometer by a transport belt.
Depending on the wavelength required for the measurement, the cuvette block
will be positioned into the 4 light beams of the respective photometer block.
The light emitted by the LED’s is directed through a focusing lens. This lens
ensures a light beam, which is as parallel as required for having homogenous
measurement conditions.
While passing the cuvettes, respectively the liquid in the cuvettes, a part of the
light energy is being absorbed by the liquid in the cuvette. The remaining light
energy is now passing another lens and the respective interference filter and is
then being measured by the detectors.
22
3 ELECTRONICS
- Schematic Diagrams
3.1 Introduction
This chapter gives information on the electrical and electronical components of
the Human system.
FIGURE 10
3.2 Schematic Diagram of Electronics
24
FIGURE 11
4.1 Introduction
This chapter gives the instructions on how to test functionality of all
independent modules of HumaStar 180 by using the Service Software
“EulyService”.
4.4 Script
A script consists of a string of commands, which are explained in this section of
the Service manual.
26
FIGURE 12
EuLy Service Script
Screen
Script features a command set that allows service personnel to check and
confirm functionality of modules of HumaStar 180. Every single module and it’s
sub-functions can be checked independently.
When processing commands, Script requires command strings that determine the module, the subfunction
Euly Service does not check and the actual command for the respective function.
for sensors ! It is the responsibili-
ty of the Service engineer to make A command – string consists of:
sure, that only command strings - Module Code
are being entered, that can not - Command
damage mechanical parts. - Parameter
The Module codes, Commands and Parameters are explained on the following
pages (all letters are case sensitive !)
IMPORTANT NOTES:
Before starting X-movement of the pipetting arm, both probes must be in top
position !
Commands: “ Zi “ (initialises left Probe) and “ Zj “ (initialises right Probe)
Before starting Cuvette Rotor movements, make sure the Belt drive is
initialised! Command: “ Mi ”
3 0w3 MV off WP ON AP ON
5 0w5 MV ON WP off AP ON
6 0w6 MV ON WP ON AP off
7 0w7 MV ON WP ON AP ON
FIGURE 14
30
FIGURE 15
FIGURE 16
Commandset for reading
with Photometer 1.
After the command set has been entered and maybe stored as a script, select the
“Start”, button and one command after the other will be sent to the instrument.
During processing all steps, the feedback from the instrument will be displayed
in the Result Window of Euly Service.
4.5 ReagentRotor
Menucommand for adjusting and testing the Reagent Rotor.
4.6 CuvetteRotor
Menucommand for adjusting and testing the Cuvette Rotor.
4.7 Belt
Menucommand for adjusting and testing the Belt.
32
4.8 Pipettor
Menucommand for adjusting and testing the Pipettor.
4.9 EulyTest
Menucommand for testing the liquid detection.
4.12 FW Update
Menucommand for updating the firmware.
4.13 Photo
Menucommand for testing the Photometer.
4.14 Pumps-Valves
Menucommand for testing the Pumps and Valves.
4.15 Info
Menucommand for displaying program information.
5 MAINTENANCE
5.1 Introduction
This chapter gives information on maintenance procedures for the Eurolyser sy-
stem. The required maintenance varies with usage of the Eurolyser system. The-
refore the following recommendations are only general guidelines and should
be adapted according to the needs.
5.2 Syringes
Reasons to change the part:
- The Syringe Seal looks worn.
- Leakage of the Syringe.
- At least once a year.
Remarks:
You can either change the Syringe Seal or the complete Syringe. If you change
only the Syringe Seal change the whole Syringe every forth time.
5.4 Tubings
Reasons to change the part:
- The Tubings look worn.
- Leakage of the Tubings.
34
5.5 Photometer
Reasons for maintenance of the part:
- The Photometer is dirty.
- At least once every 6 months.
Instructions:
1. Clean the wall of the photometer with a small amount of glass cleaner on a
cotton wool stick.
2. Perform a Belt Adjust (either with the Service Software or with the
Application Software).
5.6 X-Drive
Reasons for maintenance of the part:
- The X-Positioning is not accurate (e.g. small (unwanted) movements on a
certain position).
- At least once every 6 months.
Instructions:
1. Clean the spindle and the two axis (use only alcohol) with a cotton wool stick.
2. Lubricate spindle and axis with a small amount of the Special Lubricant
(BE1072).
5.7 Z-Drives
Reasons for maintenance of the part:
- The Z-Positioning is not accurate (e.g. small (unwanted) movements on a
certain position).
- At least once every 6 months.
Instructions:
1. Clean the spindle (use only alcohol) with a cotton wool stick.
2. Lubricate spindle and axis with a small amount of the Special Lubricant
(BE1072).
5.10 Probes
Reasons for maintenance of the part:
- Probe is blocked.
Instructions:
1. Use the Cleaning Set Reagent & Sample Probe.
6 TESTS/TROUBLESHOOTING
6.1 Introduction
This chapter gives information on testing components of the Human system.
Sensors:
Start EulyService-Script “SensorTest” with 1000 iterations. Check the state of the
doorsensor, cuevettesensor, button and liquidsensors.
Reagentrotor:
Insert empty tray on the reagentrotor.
Select menucommand „EulyService“ -> „ReagentRotor“ and click „Test Digital“.
Start EulyService-Script „ReagRotorTest1“.
Start EulyService-Script „ReagRotorTest2“.
Insert full tray on the reagentrotor.
Start EulyService-Script „ReagRotorTest1“.
Start EulyService-Script „ReagRotorTest2“.
Cuevtterotor:
Select menucommand „EulyService“ -> „CuevetteRotor“ and click “Test Digital”.
Start EulyService-Script „CuvRotorTest1“. All positions are positioned in both
directions.
Select menucommand „EulyService“ -> „CuevetteRotor“ and click „Test Sensor“.
Cuevettesensor is tested with and without cuevettes.
Insert a cuvette on position 1 (marked with 2 dots) and remove all other cuvettes.
Start EulyService-Script „CuvRotorTest5“.
38
Pipettor:
Start EulyService-Script „PipettorSensorTest“ with 1000 Iterations. Open and
close the pipettor and check the results.
Start EulyService-Script „PipettorXTest“. Positions the X-drive.
Fill traypositions R2 and S2 with water (traypositions R1 and S1 have to be
empty). Select menucommand „EulyService“ -> „EulyTest”.
Enable “LiquidDetection”.
Click „Start“.
Check results on sheet „EulyTest“.
Dispenser:
Start EulyService-Script „DispenserTest“. Initalizes the dispenser.
Photometerstreet:
Start EulyService-Script „PhotoLEDsOn“. Switches on the photometer-LEDs.
Insert a cuvette on position 1 (marked with 2 dots).
Start EulyService-Script „BeltTest1“. The cuevette is positioned to all
photometer-modules.
Endurance Test:
Fill traypositions R2 and S2 with water. Empty all other sample-positions and R1.
Insert a cuvette filled with water to position 1 (marked with 2 dots) of the
cuvette rotor.
Select menucommand „EulyService“ -> „EnduranceTest“.
Click „Start“.
Check results on sheet „EnduranceTest“.
6.3 Troubleshooting
6.3.3 BELT
No Replace photometer.
Yes Start EulyService-Script
BeltFastCCW.
Does the belt move in counter-clockwise No Remove possible blocking.
direction?
No Replace photometer.
Yes Start EulyService-Script
BeltSlowCCW.
Does the belt move in counter-clockwise No Remove possible blocking.
direction?
No Replace photometer.
No Proceed with step 5.
5 Select menucommand
EulyService->Belt and click Adjust
and check results on sheets Belt1
and Belt2.
Does the adjustment fail? Yes Replace photometer.
No Proceed with step 6.
6 Insert cuvette in outer carussel
position 1. Start EulyService-
Script BeltTest1.
Result: „M: State: READY M-Timeout“ Yes Replace photometer.
No If the error occurs from time to
time replace photometer.
44
6.3.4 PIPETTOR
6.3.7 COMMUNICATION
7 REPLACEMENT
7.1 Introduction
This chapter gives information on replacing components of the Human system.
FIGURE 17
Liquid connection
block
FIGURE 18
Reagent-code-flat-cable
Reagent-position-cable
FIGURE 19 Cuvette-
temperature-cable
6. Disassemble the cuvette rotor board (see “Disassembling the Cuvette Rotor
Board”).
7. Remove the 5 screws on the bottom of the instrument fixing the carousel
unit.
FIGURE 20
Screws for
carousel unit
6. Adjust the cuvette rotor board until the cuvette rotor position is aligned with
the photometer.
FIGURE 21
Centering-magnet-
cable
FIGURE 22
Liquid connection
block
FIGURE 24
Reagent-code-flat-cable
Reagent-position-
cable
Cuvette-tempera-
ture-cable FIGURE 25
6. Disassemble the cuvette rotor board (see “Disassembling the Cuvette Rotor
Board”).
7. Remove the 5 screws on the bottom of the instrument fixing the carousel
unit.
FIGURE 26
FIGURE 27
Centering-magnet-
cable
6. Adjust the cuvette rotor board until the cuvette rotor position is aligned with
the photometer.
FIGURE 28
12. Insert tray and align border between position 1 and 25 with photometer.
FIGURE 29
FIGURE 30
FIGURE 31
> 0.3mm
2. Adjust the height of the cuvette ring to 0.3mm above the photometer. To
change the height open all 6 screws, rotate the spacing bolts 1/6-rotation
each and tighten the screws. Repeat this until the height is as desired.
60
FIGURE 32
0.3mm
Top of photometer
FIGURE 33
FIGURE 34
3. Adjust the height of the centering magnet to the centre of the fixing area on
the cuvette rotor.
FIGURE 35
FIGURE 36
64
Adjust the position of the centering magnet so that the cuvette rotor position is
aligned with the photometer.
FIGURE 37
Alignment
7.5 Cover
4. Remove the 6 screws on the bottom of the instrument fixing the main cover.
FIGURE 39
Doorlock-cable
FIGURE 41
Syringe-pump-
cable
Doorsensor-LED-key-
flat-cable
FIGURE 42
Cuvette-motor-
cable
68
FIGURE 43
Slug
4. Disassemble the cuvette drive motor assembly (1 screw on the bottom of the instrument).
FIGURE 44
FIGURE 45
Pressure = 8N
FIGURE 46
Cuvette-flat-cable
Centering-magnet-
Cuvette-motor-cable cable
X-15V-cable
Cuvette-tempera- Photometer-tem- Z-24V-cable
ture-cable perature-cable Cuvette-
sensor-cable
Photometer-
flat-cable
Reagent-flat-cable Cuvette-flat-cable
FIGURE 48
15. Remove the master board (4 screws on the bottom of the instrument).
7.9 Photometer
Disassembling the Photometer
1. Disassemble the cover (see “Disassembling the Cover”).
FIGURE 49 Photometer-tempe-
rature-cable
Photometer-flat-cable
FIGURE 50
5. Remove the 8 screws on the bottom of the instrument fixing the photometer.
6. Remove the photometer.
Assembling the Photometer
1. Assemble the photometer.
2. Fix the photometer with 8 screws.
3. Assemble the centering magnet.
4. Adjust the centering magnet (see “Adjusting the Centering Magnet”).
5. Connect the photometer-temperature-cable (red and black) to the master
board.
6. Connect the photometer-flat-cable to the master board.
7. Assemble the cover (see “Assembling the Cover”).
8. Execute functional test for photometer.
76
FIGURE 51
Saftey ring
5. Disconnect the Z-24V-cable (red and black) from the Z–drive board.
6. Open the cable driller fixing the Z-24V-cable behind the X-drive board.
7. Disconnect the X-15V-cable (yellow and black) from the X-drive board.
8. Disconnect the X-flat-cable from the X-drive board.
FIGURE 54
Z-24V-cable
X-flat-cable
X-15V-cable
FIGURE 55
Needles
Holes in washing-
device
3. Press „Adjust Z“ (adjust the needles just above the cuvette rotor).
FIGURE 56
Needles
Top of cuvette
rotor
7.11 Probes
Disassembling the Probe
1. Remove the top cover (5 screws below the top cover).
FIGURE 57
2. Disconnect the Z-24V-cable (red and black) from the Z–drive board.
FIGURE 58
Reagent-probe-cable Sample-probe-cable
FIGURE 59
Shield (middle)
Liquidsensor-sample-
probe-potentiometer
Liquidsensor-reagent-
probe-potentiometer
Temperature-sample-
probe-potentiometer
Temperature-reagent-
probe-potentiometer
FIGURE 60
Reagent-flat-cable Reagent-code-flat-
cable
Reagent-position-
cable (not present on
all instruments)
FIGURE 61
FIGURE 62
Screws
Centerring
FIGURE 63
Coupler
Slug
FIGURE 65
Reagent rotor poten-
tiometer
86
Wires
1. Disconnect the wires from the old reagent rotor potentiometer and connect
to the new reagent rotor potentiometer one by one.
2. Assemble the reagent rotor potentiometer and fix it with silicon.
3. Assemble the reagent rotor potentiometer ring (as low as possible).
4. Assemble the coupler.
5. Assemble the centerring.
Adjusting the Reagent Rotor
1. Start EulyService-Script “ReagRotorPosition“ with 1000 Iterations.
2. Insert tray and align border between position 1 and 25 with photometer.
FIGURE 67
3. Remove tray.
4. Adjust the centerring until the position shown by EulyService-Script
“ReagRotorPosition” is on the threshold between 1 and 25.
FIGURE 68
Tubings FIGURE 69
FIGURE 70
Syringe-pumps-
cable
Jumper
1. Install the jumpers on the syringe pump according the figure.Connect the
syringe-pumps-cable to the syringe pump.
2. Assemble the syringe pump (2 screws).
3. Assemble the tubings to the syringe pump.
92
Washing-pumps-
FIGURE 72 cable
FIGURE 73
Tubings
FIGURE 74
Screws for
washing pumps
8 ADDITIONAL INFORMATIONS
8.1 Introduction
This chapter gives additional information on the components of the Human
system.
2. X-Motor:
There are two types of X-motors, which are from different manufacturers (escap
and maxon)
- After either changing the X-drive-board or the X-motor start one of the
following EulyService-Script:
- “X_Enc16_EEPROM” for motor from manufacturer escap
- “X_Enc32_EEPROM” for motor from manufacturer maxon
There are two types of sensor systems for the reagent rotor:
- No hardware-change of the Reagent Rotor Control Board