Professional Documents
Culture Documents
and often overlooked components of the GC system contamination problems are suspected (e.g., ghost Clean Blank Run (no injection) Take steps to prevent sample
backflush (reduce injection volume,
Mixed sample solvent Change sample solvent to
a single solvent
Worse for solvents with large
differences in polarity or boiling
or analysis. Many of these items are transparent in peaks or erratic baseline). lower inlet temperature, use larger
volume liner)
points
2. Run a blank analysis (i.e., start the GC, but with no sample cleanup, handling,
transfer, and storage Poor sample focusing Use a retention gap For splitless and on-column
–– Gases: pressures, carrier gas average linear velocity, injection
injection) using the normal temperature conditions Semivolatile contamination Bake-out column. Solvent Limit bake-out to 1 to 2 hours.
and flow rates (detector, split vent, septum purge) (peak widths will be broader rinse the column. Check for Only for bonded and cross-linked
and instrument settings. than sample peaks with similar contamination in the inlet, phases
retention) carrier gas, or carrier gas lines
Retention Time Shift Possible Cause Solution Comments
–– Temperatures: column, injector, detector, and transfer Change in carrier gas velocity Check the carrier gas velocity All peaks will shift in the same
3. Collect the chromatogram for this blank run. direction by approximately the
lines Excessive Possible Cause Solution Comments 2.00 2.75 same amount
4. Immediately repeat the blank run when the first one Baseline Noise
Injector contamination Clean the injector; replace liner, Try a condensation test; gas lines
4.75 5.50
Change in column temperature Check the column temperature Not all peaks will shift by the same
–– System parameters: purge activation times, detector gold seal may also need cleaning amount
is completed. Do not allow more than 5 minutes to Column contamination Bake out the column Limit the bake-out to 1 to 2 hours 4.00 Change in column dimension Verify column identity Measure the carrier gas velocity
attenuation and range, mass ranges, etc.
3.25
with an unretained compound
elapse before starting the second blank run. Solvent rinse the column Only for bonded and cross-linked
phases Large change in compound Try a different sample May also affect adjacent peaks.
–– Gas lines and traps: cleanliness, leaks, and expiration concentration concentration Sample overloading is corrected
5. Collect the chromatogram for the second blank run Check for inlet contamination with an increase in split ratio or
sample dilution
Detector contamination Clean the detector Usually the noise increases over
–– Injector consumables: septa, liners, O-rings, and and compare it to the first chromatogram. time and not suddenly Leak in the injector Leak check the injector A change in peak size usually also
occurs
ferrules Contaminated or low-quality Use better grade gases; also Usually occurs after changing a
6. If the first chromatogram contains a larger amount gases check for expired Gas Clean
filters
gas cylinder Blockage in a gas line Clean or replace the plugged line More common for the split line;
also check flow controllers and
–– Sample integrity: concentration, degradation, solvent, of peaks and baseline instability, the incoming carrier Column inserted too far into
the detector
Reinstall the column Consult GC manual for proper
insertion distance Septum leak Replace septum
solenoids
Check for needle barb
and storage gas line or the carrier gas is contaminated. Incorrect detector gas flow Adjust the flow rates to the Consult GC manual for proper flow Sample solvent incompatibility Change sample solvent For splitless injection
rates recommended values rates Use a retention gap
–– Syringes: handling technique, leaks, needle sharpness, 7. If both chromatograms contain few peaks or little Leak when using an MS, ECD,
or TCD
Create leak-free column unions
with an UltiMetal Plus Flexible
Usually at the column fittings or
injector
and cleanliness baseline drift, the carrier gas and incoming carrier gas Metal ferrule or a Self Tightening
column nut Change in Peak Size Possible Cause Solution Comments
lines are relatively clean. Old detector filament, lamp, or Replace appropriate part Change in detector response Check gas flows, temperatures, All peaks may not be equally
–– Data system: settings and connections electron multiplier and settings affected
Septum degradation Replace septum For high temperature applications, Check background level or noise May be caused by system
use an appropriate septum contamination and not the detector
View the Agilent GC troubleshooting videos: Change in the split ratio Check split ratio All peaks may not be equally
affected
agilent.com/chem/gctroubleshooting Baseline Instability Possible Cause Solution Comments All Peaks Some Peaks Change in the purge activation Check the purge activation line For splitless injection
time
Injector contamination Clean the injector Try a condensation test; gas lines
For Agilent Technical Support, please visit or Disturbances may also need cleaning Change in injection volume Check the injection technique Injection volumes are not linear
Column contamination Bake out the column Limit a bake-out to 1 to 2 hours
agilent.com/chem/techsupport Unequilibrated detector Allow the detector to stabilize Some detectors may require up to
Change in sample concentration Check and verify sample
concentration
Changes may also be caused
by degradation, evaporation, or
24 hours to fully stabilize variances in sample temperature
Locate supplies and parts with ease: Incompletely conditioned Fully condition the column More critical for trace-level
Leak in the syringe Use a different syringe
or pH
Sample leaks passed the plunger
column analyses
agilent.com/chem/partsfinder Change in carrier gas flow rate Often normal MS, TCD, and ECD respond to
or around the needle; leaks are not
often readily visible
during the temperature program changes in carrier gas flow rate
Find the correct GC column for your application: Column contamination Trim the column Remove 0.5 to 1 meter from the
front of the column
reliability your lab needs at Loss of Resolution Possible Cause Solution Comments
Decrease in separation
www.agilent.com/chem/gcproductivity Tailing Peaks Possible Cause Solution Comments
Different column temperature Check the column temperature Differences in other peaks will be
Column contamination Trim the column Remove 0.5-1 meter from the 10.59 10.83 10.59 10.77 10.59 10.77
front of the column visible
Separation
Solvent rinse the column Only for bonded and cross-linked Different column dimensions Verify column identity, measure Differences in other peaks will be
phases or phase the carrier gas velocity visible
Peak
Check for inlet contamination width Coelution with another peak Change column temperature Decrease column temperature
and check for the appearance of
Column activity Irreversible. Replace the column Only affects active compounds Wh = 0.059 Wh = 0.059 Wh = 0.105
a peak shoulder or tail
R = 2.4 R = 1.5 R = 0.84
Decrease in Increase in Increase in peak width
Solvent-phase polarity mismatch Change sample solvent to a More tailing for the early eluting Separation Peak width
single solvent peaks or those closest to the Change in carrier gas velocity Check the carrier gas velocity A change in the retention time
solvent front also occurs
Use a retention gap 3 to 5 meter gap is sufficient Column contamination Trim the column Remove 0.5 to 1 meter from the
front of the column
Solvent effect violation for Decrease the initial column Peak tailing decreases with
splitless or on-column injections temperature retention Solvent rinse the column Only for bonded and cross-linked
phases
Too low of a split ratio Increase the split ratio Flow from split vent should be
20 mL/min or higher Change in the injector Check the injector settings Typical areas: split ratio, liner,
temperature, injection volume
Poor column installation Reinstall the column More tailing for the early eluting
peaks Change in sample concentration Try a different sample Peak widths increase at higher
concentration concentrations
Some active compounds Utilize inert flow path Most common for amines and
always tail consumable components carboxylic acids Improper solvent effect, Lower oven temperature, better For splitless injection
This information is subject to change without notice.
(agilent.com/chem/inert) lack of focusing solvent, sample phase polarity
© Agilent Technologies, Inc. 2019
match, use a retention gap
Published in the USA, April 15, 2019
5994-0451EN