Tisch Environmental, Inc.
Flow Look-Up Table for PM10 VFC
High Volume Air Sampler
Serial # P10311 X
Calibrated with Rootsmeter scrial # 0438320
Date Calibrated: 01/16/17
G Factor = 0,01570695
Village of Cleves, Oh 4502
Torts aD -ruo
an
FAX! (S13) 467-5009G-Factor Excel Spreadsheet Instructions
‘The TE-6070V (use for PMI2.5 VFC) calibration worksheet can be found on our
Website at www.tisch-env.com. If you have the G Factor that accompanies each
VEC, go to "TE-6070V High Vol. PM10 VFC with G-Factor’, if you do not have a G
Factor then go to “TE-6070V High Vol. PM-10” excel spreadsheet.
Note: Calibration orifices should be sent back to Tisch Environmental for
calibration on an annual basis per US EPA Compendium Method I0-2.1 Part 7.3.2
‘Sampling of Ambient Air For Total Suspened Particulate Matter (SPM) and PMzo
Using High Volume (HV) Sampler
1. Enter the following information in the corresponding cells in the
worksheet:
Site Information
Location ‘The location ofthe instrument
ate Current Oate
Tech ‘eehnitian performing the calbration
seal Saril number ofthe instrument, Px
VEC Factor The factor ofthe UFC you are callrating. Tis can be found onthe lokup
table documentation (fst gage ofthis doc) ar the sicker located onthe VEC
Callbcation onfie Information
Make ‘Tae make of the oie, typeallyTsch Environmental
Mode! ‘The model number ofthe eric, typlalyTE-50264
Serial ‘The Serial number ofthe eairation orcs you are using
‘GaSiope (m} Tae Ga slope ofthe elation onfiee you are using. This fund onthe
calbration documentation provided with the ealibration orifice
a3 ne) ‘The Qa intercept ofthe cllbaton orice you are using. This found on the
calibration documentation provided with the calloration ace
Calration ue ‘The date thatthe ealbaton ofthe office is de. Ores should be calbrated
Date fon an annual bass. Call Tisch Environmental at 1-S®-AND-PMIO to sehedle 2
calibration
Ambient Conditions
‘Tem (Dee 1 Enter the current ambient temperature at caliration, Tain Oagrees Kanda a
grees C willbe calculated by the spreadsheet
Barometric Pressure _ Enter the ambient barometrie pressre (Pa) inches of Hg calibration the Pa in
‘mn wl be seated bythe spreadsheet2, Enter the calibration information by performing each calibration point
and entering the following information into each corresponding row for
each poi
albration information
orice "0 ‘The pressure measured at the ore port using 2 manometer, The fst point
shouldbe performed with the ores knob turned counter clockwise wide
‘open, then four consecutive points turning the orfiee kno clockwise ret
ose)
(Good ides to take afew entr points here.
Sampler"H0 The pressure measured at the sampler side port using # manometer {cea tubing
thats connect to bulkhead ting thats cannectd to side of fite elder)
The calrator flow is eateuated (0a) using the formula
‘The alulated flow in m/min wil be aleulated using the gator formula, ths flow will correspond to
the low found nthe lookup able supplied with the VEC
The percent ference wil be clelated using the forme
Calculated Flow ~ (Qa) Calibrator Flow
tilt (a) Calibrator Flow
x 100
As per statedin the method 0-21, % Difference calculations shouldbe less than + 4%
3. To calculate the total air volume during the sample enter the following
information:
aleulate Total Ale Volume Using G-Fator
average Enter the average temperature of te sample throughout the sample period in
Temperatre Deg. The temperature wil then be calculated in ee K
Barometric Pressure Enter the average barometric pressure ofthe sample throughout the sample
period in Inches of Hg. The barometric pressure in mig will then be akuated
Clean fiter“Hi0 Enter the clean fiter pressure in Inches of water prior to samoling,
Dirty Fter "Hid Enter the ery fter pressure in inches of water after sampling. The ae
sample pressure wil then be calelated in mig
Ponting Enter the total untime in hours ba ofthe sample
Using the g-octor formula, Po/Pa willbe caleulted and an average fw cate ofthe sample wil be
Calelated in m/min. Using his information te total sample volume willbe eakulstedar Gee eIpenne accelTe EY an Oe En Gn Se aN an Ge OS Ge Oe Ee