Professional Documents
Culture Documents
MEDITOM
DT-400S/300S
User Manual
DAIWHA
2
5.1 Twin Button Handle & Return Electrode ................................................................................ 28
5.2 Unipolar Handle & Return Electrode ..................................................................................... 28
5.3 Bipolar Forceps ..................................................................................................................... 29
5.4 Foot switches......................................................................................................................... 29
6. Operating and Sequence ........................................................................................... 30
6.1 Monopolar.............................................................................................................................. 30
6.2 Bipolar ................................................................................................................................... 34
6.3 Summary ............................................................................................................................... 34
6.3.1 Preparations ............................................................................................................... 34
6.3.2 Operation Sequence .................................................................................................. 34
6.3.3 Guidelines after use ................................................................................................... 35
6.3.4 Precautions during use .............................................................................................. 36
7. Output Setting Guide.................................................................................................. 37
8. Sterilization ................................................................................................................. 38
9. Load Regulation & Vpeak Graphs ............................................................................. 39
9.1 Pure Cutting Mode ................................................................................................................ 39
9.1.1 Load Regulatio ........................................................................................................... 39
9.1.2 Open Circuit Peak Voltage ......................................................................................... 40
9.2 Blend1 Cutting Mode ............................................................................................................. 41
9.2.1 Load Regulation ......................................................................................................... 41
9.2.2 Open Circuit Peak Voltage ......................................................................................... 42
9.3 Blend2 Cutting Mode ............................................................................................................. 43
9.3.1 Load Regulation ......................................................................................................... 43
9.3.2 Open Circuit Peak Voltage ......................................................................................... 44
9.4 Blend3 Cutting Mode ............................................................................................................. 45
9.4.1 Load Regulation ......................................................................................................... 45
9.4.2 Open Circuit Peak Voltage ......................................................................................... 46
9.5 Soft Coagulation Mode .......................................................................................................... 47
9.5.1 Load Regulation ......................................................................................................... 47
9.5.2 Open Circuit Peak Voltage ......................................................................................... 48
9.6 Spray Coagulation Mode ....................................................................................................... 49
9.6.1 Load Regulation ......................................................................................................... 49
9.6.2 Open Circuit Peak Voltage ......................................................................................... 49
9.7 Cutting Biploar Mode ............................................................................................................. 50
9.7.1 Load Regulation ......................................................................................................... 50
9.7.2 Open Circuit Peak Voltage ......................................................................................... 50
9.8 Soft Bipolar Coagulation Mode.............................................................................................. 51
9.8.1 Load Regulatio ........................................................................................................... 51
3
9.8.2 Open Circuit Peak Voltage ......................................................................................... 51
10. User Maintenance ....................................................................................................... 53
10.1 General Maintenance Information ..................................................................................... 53
10.2 Responsibility of the manufacturer .................................................................................... 53
10.3 Cleaning and Disinfecting .................................................................................................. 53
10.4 Periodic Inspection ............................................................................................................ 53
10.5 Periodic Performance Testing ............................................................................................ 54
10.6 Product Life Span .............................................................................................................. 54
10.7 How to Scrap the Unit and Accessories ............................................................................ 54
10.8 Warranty ............................................................................................................................ 54
11. Guidance and manufacture’s declaration on electromagnetic emissions ............. 56
12. Document History....................................................................................................... 61
4
1. Precautions
1.1 Overview
◎ This manual describes how to use the electrosurgical unit (Model: DT-400S/300S) manufactured
by DAIWHA Corporation, and its technical specifications.
◎ Repair or circuit alteration of this equipment can only be carried out by personnel authorized by
DAIWHA. The warranty shall not apply to defects or damage caused by repairs or disassembly by
anyone other than personnel authorized by DAIWHA, and safety shall not be assured.
◎ Parts or circuits which are used for this equipment can be replaced to improve the safety and
performance of this equipment.
◎ Read this manual carefully before using the electrosurgical unit, and understand its functions and
how to use it for safe use.
◎ Contact DAIWHA if you have any questions regarding the equipment or the User Manual.
◎ As the medical equipment, this product can be used only trained doctors.
◎ When the patient contacts grounded electric conductors such as the metal part of the grounded
operation table during the output, it may burn or shock the contacted body part.
◎ Since it operates on high frequency and high voltage, electromagnetic waves are generated
naturally. This might affect other equipment such as distorting an ECG monitor screen in a
negative way. When HF SURGICAL EQUIPMENT and physiological monitoring equipment (for
example ECG monitor) are used simultaneously on the same PATIENT. The grounding of ECG
should be placed through neutral plate in case that the units are using the same ground earth.Any
monitoring electrodes should be placed as far as possible from the surgical electrodes (Minimum
distance 150mm). Needle monitoring electrodes are not recommended.
The lead of Radio-Frequency electrode is the shorter is the better not to make loops between other
leads. For this reason Using device that apply to EMC(Electro-magnetic Compatibility) is strongly
recommended.
◎ Set the output to the lowest level which can achieve the desired operation effect, and then
gradually increase the output volume to the desired level.
◎ Prior to use, check the equipment-whether it has a problem.
◎ Power voltage and frequency must be supplied in a stable manner within the designated range on
product rear side. However, the output may vary depending on power. If the wall outlet voltage in
the hospital exceed the designated range by ± 5%, the equipment should be calibrated again by
manufacturer or supplier.
◎ If the power frequency in the hospital different from the designated range, the equipment should be
calibrated again, too. Please refer to product label on the back for exact power usage.
◎ Considering its cooling feature, continuous output can result in abnormal operation of the product.
Be sure duty cycle.
◎ A short circuit that might occur between electrodes during activating output will damage the
product.
◎ The Failure of the HF SURGICAL EQUIPMENT could result in unintended increase of output
power.
◎ Do not reuse disposable accessories.
◎ All accessories has to be always keep disinfected when you use, especially metal parts which is
contacting with the patient's body
◎ This manual is only applicable to pumps with Program Version v1.xx or above.
1. Precautions5
Safety precautions are classified as below in accordance with the expected danger, seriousness of
damage or urgency.
A precaution which can result in critical personnel injury or immediate loss of life if
Danger instructions are not followed.
A precaution which could result in critical personnel injury or loss of life if instructions are
Warning not followed.
A precaution which could result in minor personnel injury or damage to the product if
Caution instructions are not followed.
1.2 Danger
1) The use of anesthetics or oxidizing gases such as nitrous oxide (N2O) and oxygen should be
avoided if a surgical procedure is carried out in the region of the thorax or the head, unless these
agents are sucked away.
Non-flammable agents should be used for cleaning and disinfection wherever possible.
Flammable agents use for cleaning or disinfecting, or solvents of adhesives, should be allowed to
evaporate before the application of H.F. surgery. There is a risk of pooling of flammable solutions
under the PATIENT or in body depressions such as the umbilicus, and in body cavities such as the
vagina. Any fluid pooled in these areas should be mopped up before HF SURGICAL EQUIPMENT is
used.
Attention should be called to the danger of ignition of endogenous gases. Some materials, for
example cotton, wool and gauze, when saturated with oxygen may be ignited by sparks produced in
NORMAL USE of the HF SURGICAL EQUIPMENT.
1.3 Warnings
1) The entire surface area of the Neutral Plate(Patient Plate, Return Electrode) must be reliably
attached to the patient's body. Be sure to check the contact status frequently during the operation.
When the contact is poor, parts of body touching the Neutral Plate(Patient Plate, Return Electrode)
will be burned. Avoid areas such as near heart, scars, joints, and parts with bones projected, parts
with a small area, and palaces where contact is likely to be unstabilized when the patient moves.
2) If the patient complains about electric shock, stop using the product immediately and have it
examined throughly.
3) Equipotential Grounding
The Electro Surgical Unit must be grounded to Earth. The Grounding Resistance must be Maximum
10Ω. Unexpectable Electric shock may be happened when you use the unit with Non-grounding.
Expecially, Equipotential Grounding is required between the devices in the same operating room, it
prevent micro shock. The Micro Shock means electric shock to heart. Sometimes it can be caused a
ventricular and so on. The resistance between the Grounding of equipments in same operating room
must be Maximum 0.1Ω.
4) Cardiac pacemaker or Internal Cardiac Defibrillator-(ICD)
HF SURGICAL EQUIPMENT poses a potential risk to patients with implants ( e.g. cardiac
pacemakers) or something. For PATIENT with cardiac or active implants, a possible hazard exists
because interference with the action of the pacemaker may occur, or the pacemaker may be
damaged. In case of doubt, approved qualified advice should be obtained. Before using this product
to patients who are currently using Pacemaker or ICD, consult with the manufacturer or cardiac
experts at the hospital.
1.4 Caution
1) As the medical equipment, this product can be used only trained doctors.
2) When the patient contacts grounded electric conductors such as the metal part of the grounded
operation table during the output, it may burn or shock the contacted body part.
3) Since it operates on high frequency and high voltage, electromagnetic waves are generated
1. Precautions6
naturally. This might affect other equipment such as distorting an ECG monitor screen in a negative
way.
When HF SURGICAL EQUIPMENT and physiological monitoring equipment (for example ECG
monitor) are used simultaneously on the same PATIENT. The grounding of ECG should be placed
through neutral plate in case that the units are using the same ground earth.
Any monitoring electrodes should be placed as far as possible from the surgical electrodes (Minimum
distance 150mm). Needle monitoring electrodes are not recommended.
The lead of Radio-Frequency electrode is the shorter is the better not to make loops between other
leads. For this reason Using device that apply to EMC(Electro-magnetic Compatibility) is strongly
recommended.
4) Set the output to the lowest level which can achieve the desired operation effect, and then
gradually increase the output volume to the desired level.
5) Prior to use, check the equipment-whether it has a problem.
6) Power voltage and frequency must be supplied in a stable manner within the designated range on
product rear side. However, the output may vary depending on power. If the wall outlet voltage in the
hospital exceed the designated range by ± 5%, the equipment should be calibrated again by
manufacturer or supplier.
If the power frequency in the hospital different from the designated range, the equipment should be
calibrated again, too. Please refer to product label on the back for exact power usage.
7) Considering its cooling feature, continuous output can result in abnormal operation of the product.
Be sure duty cycle.
8) A short circuit that might occur between electrodes during activating output will damage the product.
9) The Failure of the HF SURGICAL EQUIPMENT could result in unintended increase of output power.
10) Do not reuse disposable accessories.
11) All accessories has to be always keep disinfected when you use, especially metal parts which is
contacting with the patient's body
1) Please do operate the unit by the instructed order.
The following rules must be observed and always pay keen attention.
2) Do not open or disassemble the unit.
- The repairing and circuit change of this equipment must only be handled by those who have
been authorized by DAIWHA Corporation, LTD.
3) The unit is designed to be used at intervals.
Duty cycle ; 10 sec On, 30 sec Idle
4) When installing the unit,
Do not locate the unit the place where easily affected by shock, impact & etc.(including the times of
movement & transportation). Do not locate the unit the place where easily affected by
temperature, humidity, barometric pressure also. Keep clean always. At least one grounded
duplex wall receptacle and proper line voltage should be provided.
5) In using of electricity,
Connect the system AC line voltage after correct line voltage has been verified. Always check the
grounding. After using the unit, please be sure to main power off and get rid of the power cord.
6) How to prevent skin burn
① The entire area of the NEUTRAL ELECTRODE should be reliably attached to the PATIENTS
an appreciable capacitance to earth(for example operating table supports, etc.). The use of
1. Precautions7
antistatic sheeting recommended for this purpose.
- It is good to spread ECG gel or cream to increase the effectiveness of firm contact on the
neutral electrode.
③ Areas subject to extreme perspiration-extremities lying against the body, or skin-to-skin
contact(arm/trunk, leg/leg, breast.) must be kept dry by placing sheets between them. It is
easily to get burnt when these parts are applied with other metal parts. Skin-to-skin contact (for
example the arms and body of the PATIENT) should be avoided, for example by insertion of
dry gauze.
The use of wet towel as conductive agents should not be used since they dry out, resulting in
high-frequency skin burn to the patient. Urine has to be discharged through catheter or
something.
④ Initial self-test
- First off, test the unit at very low RF power and then increase the RF power smoothly to
prevent any problem caused from improper touch or disconnection of neutral electrode at
normal RF power.
⑤ Do not use children's patient plate (neutral plate) for adults. You may increase the risk of
⑥Recommendation to position patient leads in such a way that contact with the patient or
unwanted tissue damage for surgical procedures where hf current could flow through relatively
⑧Instructions indicating the output power should be selected as low as possible for
intended purpose
7) Do not use cords as a handles ; damage to the insulation and increase of burnsor other injury
may result.
8) Use accessories, active and neutral electrodes provided by DAIWHA Corporation, LTD. in order to
avoid incompatibility and unsafe operation. If you intend to purchase accessories other than
those that DAIWHA Corporation, LTD., please check that the accessories are CE certified for Electro-
surgical unit to make sure safety. Use CE certified accessories for patient contacting parts except
supplied by DAIWHA.
9) The cables to the surgical electrodes should be positioned in such a way that do not contact with
1. Precautions8
the PATIENT or other leads is avoided. Temporary unused ACTIVE ELECTRODES should be
stored so that they are isolated from the PATIENT.
10) For surgical procedures where the h.f. current flow through parts of the body having a relatively
small cross-sectional area, the use of bi-polar techniques may be desirable in order to avoid
unwanted coagulation.
11) Apparent low output or failure of the HF SURGICAL EQUIPMENT to function correctly at the
normal operating settings may indicate faulty application of the NEUTRAL ELECTRODE or poor
contact in its connections. In this case the application of the NEUTRAL ELECTRODE and its
connections should be checked before selecting higher power.
12) The supplier will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions, or other information which will assist the USER's appropriately qualified
technical personnel to repair those parts of EQUIPMENT which are designed by the
manufacturer as repairable.
13) It may cause a malfunction when the unit is used in an environment that has strong magnetic
fields i.e. MRI room.
14) Do not use the unit other than your intended purpose. It may cause unexpected side effects.
1. Precautions9
This is medical equipment.
DAIWHA Corporation
135, Donghwagongdan-ro, Munmak-eup, Wonju-si,
Gangwon-do, Korea
Product Name Electrosurgical unit
1. Precautions10
1.5 Checklist before use
1) Check and clear away flammable gas or solvent.
2) Check the cable connection on the back panel of the unit.
3) Check that all accessories that have direct contact with patients are sterilized.
4) Check all accessories in order to prevent damage or malfunction of the unit.
5) Check the DRM function in order to prevent patients from burning due to wrong measurement
of contact size that happens at the time of unit troubles.
6) Check contact status between patient plate and patient.
1. Precautions11
2. Terms and Symbols
2.1 Terms
- The terms used in this User’s Manual is explained below.
- The terminology described below is limited to this User’s Manual and note that it may be
different from those generally used.
Non-ionizing
ionizing Radiation Equipotential Grounding
Off On
(Power : Disconnection form the (Power : Disconnection to the
mains) mains)
Replace Fuse only with type and Replace Fuse only with type and
rating as shown. rating as shown.
Authorized representative in
Manufacturer
European Community
3. Single Pad or Dual Pad Selection Switch 4. Bi-polar Mode Activation Indicator
Pressing this switch will select the Single Pad This lamp lights on when the unit is activated in
return electrode or the Dual Pad return Bi-polar modes
electrode.
5. Mono-polar Cut Mode Activation Indicator 6. Mono-polar Coagulation Mode Activation
This lamp lights on when the unit is activated in Indicator
Mono-polar cut modes This lamp lights on when the unit is activated in
Mono-polar coagulation modes
7. Bi-polar Power Increase Switch 8. Bi-polar Power Decrease Switch
Increases the output power level of the Bi-polar Decreases the output power level of the Bi-polar
mode when depressed. mode when depressed.
40. Mono-polar Foot Switch Inlet 41. Main Inlet with Main Fuse case
Is a 3-pin threaded connector designed to Supplies ac main power to the unit. It only
accept plugs attached to any two treadle should be connected to a source of power
mono-polar foot switch available from corresponding to that listed on the Patcher
DAIWHA. attached right by. A detachable power cord with
a standard 3-prong are supplied with unit. The
fuses in main fuse box provide Dual-pole power
line overload protection.
44. Single Pad Mode Indicator 45. Dual Pad Mode Indicator
Illuminates when the Single pad mode is Illuminates when the Dual pad mode is
selected. selected.
Installation
5cm clearance required on each side unit for cooling
Conditions
Cooling 1 fan mounted
Recommended Duty
10 sec ON, 30 sec, OFF(Output for 10 seconds, pause for 30 seconds )
Cycle
3. Specifications19
3.2 Operating Modes and Output Parameters
(DT-400S) Tolerance:±20%
Carrier Frequency Repeat Output RF Power
Mode Load Resistance(Ω)
(㎑) Frequency(㎑) (W)
Pure mode 400 - 400 500
(DT-300S) Tolerance:±20%
Carrier Frequency Repeat Output RF Power
Mode Load Resistance(Ω)
(㎑) Frequency(㎑) (W)
Pure mode 400 - 300 500
3.3 Functions
1) Good linearity and stabilization of output power by microprocessor.
2) EMC design that is in accord with EN60601-1-2 [2001] and EN60601-2-2 [2001].
3) Remote Control using two Hand pieces (Selection between cutting and coagulation)
4) Selection between cutting and coagulation by Foot switch (Can be also used as an
exclusive switch for laparoscope.
3. Specifications20
5) A safety circuit by D.R.M.(Daiwha Return electrode Monitoring)
- To prevent burns, if the contact is lower than 80% of the adhesive area, the output power
is stopped and the alarm will be operated.
6) When the cutting and coagulation and Bi-polar is operated, it can be differentiated by the
sound and the lamp.
7) The output power can be controled by pressing the Up / Down switch on front panel and
be also shown in numbers.
8) Bi-polar is available even if Return Electrode has not been mounted.
then press (Single pad or Dual pad Selection Switch) to switch on 'Single
pad Lamp'.
② Mount Return Electrode to Return Electrode jack by using single pad, metal plate
with cable and rubber plate with cable.
➂ Patch the pad on the patient's body.
3. Specifications21
4) Alarm
: If Return Electrode is dismounted from the device or its cable is broken, Return Fault
Indicator flickers and alarm tone sounds, cutting off the output.
single pad mode. Press (Single pad or Dual pad Selection Switch) in order
to switch on ‘Dual pad Mode Lamp’.
② Mount Return Electrode to Return Electrode jack.
③ Patch the pad on the patient's body as closely as possible so that more Assist
Resistance Indicators could light for the safety, if possible. (Refer to the ‘5) Relation
between the light counts of Assist Resistance Indicator and pad resistance’ in the next
page.)
④ See ‘Assist Resistance Indicator’ change. If more than three lights are turned on, it is
possible to use them.
4) Alarm
(1) Error
a
n
error occurs, check out the error number displayed in equipment, and contact DRE Isales
agent or DRE
(2) Alarm
Alarm Alarm
Alarm Situation How to cancel the alarm
No. Display
When connecting ‘Dual Pad’ while using
the single pad mode
Connect ‘Single Pad’ to ‘Neutral
1 “ALr 1"
When the ‘Single Pad’ is generated from plate’ jack.
‘Neutral plate’ jack while using the single
pad mode
3. Specifications22
When Neutral plate cable is single lined Replace the single lined
while using the single pad mode cable.
When the contact area is small while using Apply ‘Dual Pad’ to secure
the dual pad mode larger contact area.
2 “ALr 2”
When the cable is disconnected or short Replace the disconnected or
while using the dual pad mode short line.
During the use of dual pad mode, more Re-attach the peeled-off Dual
than 20% of the originally attached sheet Pad, or Replace with a new
is fallen off over 20. Dual Pad.
- ‘DRM Setting Lamp’ and ‘Return Fault Indicator’ start to flicker and alarm tone sounds,
cutting off the output.
※ Alarm condition
- Eight or more lights are on and alarm tone sounds.
: Short circuit is generated between cables.
- 3~8 lights are on and alarm tone sounds.
: Pad contact area decreased 80% smaller than the initial area.
: HF current from surgical unit has deteriorated pad adhesives during long time
operation.
: In this case, Patch the pad to the patient and press 'DRM Setting Switch' to
continue to operate.
- One or two lights are on and alarm tone sounds.
: Return Electrode is dismounted from the device.
: Cable of Return Electrode is broken.
: Pad is detached from the patient.
: Pad is taken out from clamp.
5) Relation between the light count of Assist Resistance Indicator and pad resistance
- The following table shows the resistance at the both ends of pad foil when dual pad is
sticked to the patient.
- In general, 6~8 lights are on in case of new dual pad. If four or less lights are on even
though the pad is completely sticked to the patient, it is necessary to replace the pad
with a new one. The more the lights are on, the higher the conductivity degree of pad
becomes. The high conductivity of pad secures safety of operations by reducing
output loss.
- The wider the pad area sticked to the patient becomes, the smaller the resistance
becomes.
3. Specifications23
Tolerance : ±10%
31-50 8
51-80 7
81-100 6
Dual PAD
101-110 5
111-120 4
121-130 3
131-141 2
Above 150 0
Open PAD ∞ 0
3. Specifications24
4. Effect using the unit
4.1 Intended Purpose
This product including its associated ACCESSORIES intended for the performance of surgical
operation, such as the CUTTING or COAGULATION of biological tissue by means of high
frequency (H.F) current.
2) Blend 1
3) Blend 2
4) Blend 3
2) Soft Coagulation
- It is very important to minimize the spark between electrode and tissue for the quality coagulation.
The following procedures are to ensure the quality coagulation.
- Let the output out with the electrode adhered to the tissue.
- The output should be low enough to make coagulation. High output produces high voltage to the
tissue, resulting in spark increase.
- It is desirable to use a ball type of electrode because it generates less spark compared with other
types of electrodes. If possible, an electrode with small diameter is recommended if possible as you
need higher output in proportion to the diameter. Low output can make coagulation possible. Don't
use electrode of lancet or needle type as possible as it is subject to make spark at the edge.
- If you hold tissue with Forceps, HF current circulated between the both handles makes the
tissues in the midst of handles coagulated by heating them.
1) BipolarCut
2) BipolarSoft Coagulation
Return Electrode is made of wide area metal, conductive rubber or aluminum foil in order to minimize the
burn risk. It is sometimes called Dispersive Electrode, Neutral Plate or Patient Plate.
Plug Uni-polar Handle, mono-polar forceps, laparoscopic instrument, endoscopic instrument, TUR
Caution
- If they are plugged in the wrong hole, be warned of such malfunction that output may continue
even when the foot switch is off.
1) After checking the main power voltage and frequency, the power cord should be connected to
the main power inlet located at the rear panel of the equipment. If there is no ground circuit in
the main power outlet, the Equipotential grounding cable must be connected to the quipotential
Grounding. Contact Daiwha sales department when you need the Equipotential Grounding
cable.
2) The Power Switch is turned on - If the Pad is not connected, The Return Fault Indicator will be
twinkle and the alarm will ring to warn the fault of the Return Electrode.
① Connect the Single Pad to the Return Electrode Jack. The Alarm will stop and it will be
ready to be used.
② Attach the Pad firmly to the patient. (Please read carefully the precautions)
② Press (Single Pad or Dual Pad Selection Switch) to set up the Dual Pad Mode.
The Dual Pad Mode Indicator will be turned on and the DRM will be ready.
③ Attach the Pad firmly to the patient. (Please read carefully the precautions)
④ The changes in the Assist Resistance Indicatorwill be seen. You can set DRM by pressing
※ The Assist Resistance Indicator has 2 steps or 0 step when the Return Electrode cable
may be cut off or the Pad is not attached properly to the patient. The Indicator has 8 steps
when the Return Electrode cable may be short-circuited.
5) Select the Coagulation mode (Contact, Spray) and choose the appropriate output power [W]
1) This slide switch alters the path of RF Output by Mono-polar Foot Switch. RF Output will go out
to Monopolar Outlet1 When the slide switch is selected Mono Outlet 1.
2) This slide switch alter the path of RF Output by Mono-polar Foot Switch. RF Output will go out
to Monopolar Outlet 2 When the slide switch is selected Mono Outlet 2.
1) Connect the foot switch to the Bi-polar Foot Switch Inlet located at rear panel.
2) Select the appropriate RF Power [W] by Bi-polar Power Increase or Decrease Switch.
3) Connect the Bi-polar Forceps with Cable.
6.3 Summary
6.3.1 Preparations
1) Check if the electrode for the ESU is sterilized.
2) After checking the main power voltage, connect the power cord to the ‘Main Inlet’ located at the
rear panel of the equipment.
If there is no ground circuit in the power cord, the equipotential grounding cable must be
connected to the ‘Equipotential Grounding’.
3) Patch Neutral plate as closely and firmly to the surgery area as possible and secure as the large
contact area as possible.
- Check if the current output setting is appropriate to the existing surgical environment including
- Make sure that patients do not touch grounded metals. It may cause partial burn.
- Do not give long time output to tissue areas or higher output beyond the need. Manipulate
output when electrode is always contacted with tissue to prevent tissue burn.
- If the patient complains about electric shock, stop using the unit immediately and have it
examined thoroughly.
- The sound volume should be at least the medium of level to prevent sound-recognition-fail
caused dangers.
2) Storage condition or storage method
- Storage Condition: -10℃~60℃, 30%~ 95% noncondensing, 700mbar~1060mbar
- Put in the product package box, and place in the cool area.
Part Cut Mode Cut Power Coag. Mode Coag. Power BI-Coag.
Pure 10~120
Skin Incision
or Blend1 8~100
Muscle Pure Above 15
Dissection or Blend1 Above 15
Tumor
Blend3 Above 10
Excision
Stomach,
Blend2 20 and up
Intestine
or Blend3 20 and up
Resection
Hemostasis Contact 10~65 3~22
Massive
Spray 10~30
Coagulation
Loop
BALL
Neuro Surgery Blend2 20 ~ 80 Contact 1~8
10~25
or Blend3 20 ~ 70
Prostatic Pure 65 and up
Spray 15 and up
Resection or Blend1 55 and up
Bladder
Spray 12~30
Fulguration
Blend2 15 ~ 80
Tumor Excision Spray 10~15
Blend3 15 ~ 70
Cervical
Blend2 10 ~ 80 Contact 25~70
Conization
Bartholin
Blend3 15 ~ 30 Spray 12~22
and Skeneis
1) For cleaning and sterilizing, the unit must be disconnected from the mains. Accessories
have to be always keep sterile when you use in operating room, especially metal parts which is
contacting with the patient's body.
2) Be sure to clean and sterile the accessories before keeping them.
3) Before sterilization, inspect the accessories and their cord for breaks or cracks in the insulation,
corrosion, or loose conditions. Do not use the accessories if any of these conditions exist.
4) Sterilization
Sterilization
Accessory Steam EtO
(Autoclave) (EO Gas)
Caution
8. Sterilization38
9. Load Regulation & Vpeak Graphs
9.1 Pure Cutting Mode
9.1.1 Load Regulation
(DT-400S)
(DT-300S)
(DT-400S)
(DT-300S)
(DT-400S)
(DT-300S)
(DT-400S)
(DT-300S)
(DT-400S)
(DT-300S)
(DT-400S/300S)
(DT-400S/300S)
(DT-400S/300S)
(DT-400S/300S)
Warning
- Electric shock hazard Always turn off and unplug the unit before cleaning.
- When cleaning with damp cloth, exert cautions to prevent moist from leaking into the product. It
may cause the product to malfunction.
Caution
- Do not clean the unit with abrasive cleaning or disinfectant compounds, solvents thinner, solvent,
benzene, ammonia or acetone or other materials that could scratch the panels or damage the unit
- Turn off the unit and unplug the power cord from the wall outlet.
- Thoroughly wipe all surfaces of the unit and the power cord witch a mild cleaning solution or
disinfectant and a damp (not dripping) cloth.
- Do not allow impurities to enter the chassis.
- Do not sterilize with autoclave or EO gas.
10.8 Warranty
The manufacturer warrants that the ESU is free from defects in material and workmanship
under normal use and service for a period of months after the purchase data.
The manufacturer its authorized representative takes obligation to carry out the warranty
repair of the ESU or to replace the ESU with an operational one in case the Manufacturer
or its authorized representative determines that the cause of the ESU's failure was related
If the Buyer finds the defect in the ESU during the warranty period, he must report it and
authorized representative in its original or equivalent packaging. The ESU is sent for
If no defect is found during testing, the Manufacturer or its authorized representative reserves
the right to submit the invoice to the Buyer for the work carried out.
This warranty is not applicable to ESUs with damaged seal or when failure caused by
packaging is damaged.
Serial No :
Delivery date :
Quality inspector :
The DT-400S/300Sis intended for use in the electromagnetic environment specified below.
The customer or the user of the DT-400S/300Sshould assure that it is used in such an environment.
The DT-400S/300S is intended for use in the electromagnetic environment specified below.
The customer or the user of the DT-400S/300S should assure that it is used in such an
environment.
Immunity test IEC 60601-1-2 Compliance level Electromagnetic environment -
Test level guidance
Electrostatic ±6kV Contact ±6kV Contact Floors should be wood, concrete
discharge (ESD) ±8kV air ±8kV air or ceramic tile. If floors are
IEC 61000-4-2 covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast ±2kV for power ±2kV for power Mains power quality should be
transient/burst supply lines supply lines that of a typical commercial or
IEC 61000-4-4 ± 1kV for ± 1kV for hospital environment.
input/output lines input/output lines
NOTE Uт is the a.c. mains voltage prior to application of the test level.
survey, a
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the DT
DT-400S/300SS is used exceeds the applicable RF compliance
level above, the DT-400S/300S S should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re re-orienting
orienting or relocating
the DT-400S/300S.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1]V/m.
There is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the DT-400S/300S can help Prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the DT-400S/300S as recommended below, according to the maximum
output power of the communications equipment.
Equation
NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the DT-400S/300S is used exceeds the applicable RF compliance
level above, the DT-400S/300S should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating
the DT-400S/300S.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
The DT-400S/300S S is intended for use in the electromagnetic environment specified below.
The customer or the user of the DT-400S/300S S should assure that it is used in such an
electromagnetic environment.
61000-4-3
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
NOTE 2) It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to assure that they meet the minimum specification.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telepho
telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be consider
considered.
ed. If the measured field strength
outside the shielded location in which the DTDT-400S/300S S is used exceeds 3V/m, the DT DT-400S/300S
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, suc such as relocating the
DT-400S/300S S or using a shielded location with a higher RF shielding effectiveness and filter
attenuation.