—G20901— “gable 1. Thickness of test material and type of applicable probe ‘Thickness of test material of te ‘Type of applicable probe Double crystal normal probe 19 or over up to and incl. 60°] HOUR ory © over 60 Normal probe japparatus for flaw detection jaw detector performance Gy Automalic Naw delecloe ‘The suloialic Maw detector whites uses the jie crystal normal prove shall be checked at least once every three years, pill have the following performance: itude linearity For the amplitude linearity of a digital type automat- (edetoctor, the first bottom echo or the electric false signal of the test block “for setting sensitivity of Annex 2, is set at an appropriate level, the ampli- io Hinearity is measured hy respective lines ~6 dR, — 12 dR and —18.dR jrom the set level, and the sum of the absolute value of the positive’ and negative maximum deviation between the ideal value and the measured