You are on page 1of 2

Vertical Laminar Flow Cabinet

For Product Protection

● Reliable industry standard switches operate


the blower, fluorescent lights and the optional UV
lamp. To prevent accidental personnel exposure
to UV radiation, the UV and normal light controls
are integrated onto a 3-way switch to prevent both
from being activated at the same time. The light
and the blower can be operated independently for
example during cleaning the cabinet.

Control Panel with a 3-way switch for UV and fluorescent light,


electronic blower monitor and a blower speed controller

● Electronic blower monitor with LED display


mounted on the front panel acts as an indicator to
the blower performance and fan speed.

● Easily accessible blower speed controller


mounted on the control panel on the cabinet front
allows the operator to monitor blower performance
and take necessary steps to ensure optimum
Streamline Laboratory Products’ Vertical Laminar Flow Cabinets offer a speed.
proven level of ISO Class 4 product protection at an economical cost
without compromising performance. Streamline Laboratory Products vertical ● Permanently lubricated direct drive centrifugal
laminar flow models feature many key innovations which ensure its superior blower; unique blower mounting design allows
performance : mini-pleat separatorless HEPA filter technology, the best product for 360 degree airflow distribution, thus increasing
protection in the world, superior filter mechanical construction aspects and air uniformity.
mounting methods.
● Each unit manufactured is tested at factory, and
individual test report is sent with each cabinet.
● Streamline Lab’s Vertical Laminar Flow Cabinet is more compact than any other
vertical flow cabinet. This is due to the new, unique blower design. ● Like all Streamline Lab cabinets, all potentially
hazardous parts of the cabinet are colour-coded
● Mini-pleat HEPA filter with a typical efficiency of 99.999% at 0.3 and 0.12 red to advise the operator for caution.
microns. Single piece moulded-in-place gasket prevents leaks typical with conventional
multi-piece gasket designs. Aluminium frame is more durable and sheds no particles ● Universal electrical socket outlets (2 no’s)
compared to the conventional wooden frames. Integrated metal guards on both sides are provided as standard features on back walls.
of the filter protect it from mechanical damage.
● Lifetime warranty for the cabinet with the
● Washable polyurethane pre-filter extends life of main HEPA filter. Easily cleanable exception of: Blower and electrical systems (12-
with water, compressed air, or vacuum - needs no replacement over the operating life months) and consumables such as light fixtures,
of cabinet, therefore lowering maintanence costs. pre-filters and HEPA filters (6 months)

● Generously sized stainless steel work surface is scratch-proof, chemically


resistant, and easily cleanable with a variety of sterilizing agents. Rounded edge of the
Optional Accessories*
work surface in front ensures maximum comfort for the operator’s arms during extended
hours of work inside cabinet workzone. All joints in the work area are permanently
• 253.7nm germicidal ultraviolet lamp:
sealed at the factory to prevent ingress of contaminants.
mounts in cabinet internal work zone ceil-
ing. NB: the UV lamp is not a substitute for
● Transparent UV-tempered glass side walls and white non-reflective powder-
surface decontamination. Protect eyes and
coated back wall. Glass sides provide maximum visibility into the work zone, optimizing
skin when using.
operator comfort. White powder-coated back wall is non-reflective and will not cause
glare. They also contribute to increased and uniform overall lighting levels in the work
• Service fixtures: European and American
zone. Main body of the cabinet has a textured powder-coated finish that has higher
style fixtures are available.
resistance to routine wear than glossy finishes.
• Support stand: available with castor wheels
● Staggered service fixture penetrations on side walls allow for better access for
in standard 28” / 711mm (seated operator)
the operator than standard level provisions (see engineering diagram on page 2 for
height. Steel metal back panel.
illustration)
*Streamline Lab’s optional accessories can be fitted on
● Quick-start T8 industry standard fluorescent twin lighting is employed. The warm the cabinets onsite after purchase.
white 5000k fluorescent lamps (industry standard length) are comfortable to the eyes.
Streamline Lab Catalogue Revision No: SCV.v1.10.2004
Specifications subject to change without notice.
Streamline Vertical Laminar Flow Cabinet Specifications ©2004 Streamline Laboratory Products.

1340 / 1950 mm 584 mm / 23.0”


52.8” / 76.8”

2 280 mm / 11.0”
10
ESCO C. Vertical Laminar Flow Cabinet

100

ESCO

11 3

4
950 mm / 37.4”
5 13
550
6
380 mm / 15.0”
14
ACCESS OPENING 7
12

50 mm /2.0”
478 mm / 18.8”
1240 / 1850 mm
48.8” / 72.8”
Engineering Details
1. Pre-filter 2. Blower 3. HEPA Filter 4. Fluorescent Light
5. Front Air Guide 6. Tempered Glass Side Panel
7. Staggered Service Fixture Provisions (2 on Each Side)
8
8. Electrical Panel 9. Operating switches
10. Electronic Blower Monitor 11. Blower speed controller
1 12. Standard Universal Electrical Socket Outlet (2 no’s)
13. Optional UV Light 14. Stainless Steel Work Top

General Specifications SCV-4AX SCV-6AX


Air Volume
750 cmh / 415 cfm 1114 cmh / 648 cfm
(At Initial Velocity)
Laminar Airflow Velocity Average of 0.35 m/s or 70 fpm measured 150mm / 6” from filter face for 38 air changes / minute in work zone
ISO 14644.1 Class 4, US Federal Standard 209E Class 10 / 2.5, AS 1386 Class 2.5, JIS B9920 Class 4, BS5295
Air Cleanliness Within
Class D, Class M100,000 as per KS 27030.1 and other equivalent cleanliness classifications of the VDI 2083 and
Working Area
AFNOR X44101
H13 HEPA filter with a typical efficiency of 99.999% at 0.3 microns; integral metal guards and filter frame gaskets on
Main Filter Type
both sides; fully compliant with EN 1822 and IEST-RP-CC001.3 requirements
Pre-Filter Washable non-woven polyurethane fibers with an efficiency of 20%; framed in aluminium with an integral metal guard
Noise Level < 67 dBA at initial blower speed setting measured at typical operator work position
Light Intensity >800 lux / >75 foot candles at work surface level
Main Body Construction Electro-galvanised steel with white oven-baked epoxy powder-coated finish
Choose from the following options and specify option number (e.g. SCV-4A1) when ordering
Power Supply Options 1. 220-240VAC 50HZ, 1 phase 2. 110-130VAC 60HZ, 1 phase 3. 220-240VAC 60HZ, 1 phase
4. 110-130VAC 50HZ, 1 phase 5. 100-110VAC 50HZ / 60HZ
Power Rating 142W (230VAC 50HZ units) 380W (230VAC 50HZ units)
Net Weight 100 kgs / 220.5 lbs 160 kgs / 353 lbs
Max Shipping Weight 127 kgs / 280 lbs 200 kgs / 441 lbs
Shipping Dimensions 1450 x 750 x 1100 mm / 57.0” x 29.5” x 43.3” 2050 x 750 x 1100 mm / 80.7” x 29.5” x 43.3”
Shipping Volume 1.20 cbm / 42.4 cbf 1.69 cbm / 59.7 cbf

NOTE: Laminar flow cabinets are to be used for product (i.e.


specimen) protection only and do not offer protection for the operator Your local distributor:
or the environment from aerosol contamination generated by the
specimens placed within the cabinet. Thus no toxins or chemicals
hazardous to humans should be used inside the cabinet. For this
purpose, a biohazard safety cabinet is recommended.

Streamline Laboratory Products


32 Maxwell Road ∙ #03-13 Whitehouse ∙ Singapore 069115
Tel: +65 6785 1020 ∙ Fax: +65 6785 1027
mail@streamline-lab.com ∙ www.streamline-lab.com