You are on page 1of 21

Versana PremierTM Ultrasound System

Specification sheet Rev4

2018-Oct-30
GE Healthcare

8.10 Advance 3D............................................................. 7


Contents 9. General System Parameters ................................... 7
Contents ................................................................. 1
9.01 System Setup .......................................................... 7
1. General specifications ............................................ 3
9.02 System scanning parameters .................................. 8
1.01 Dimensions and weight .......................................... 3
9.03 B-Mode ................................................................... 8
1.02 Electrical power...................................................... 3
9.04 Coded Harmonic Imaging ....................................... 8
1.03 Console design ....................................................... 3
9.05 SRI-HD..................................................................... 8
1.04 User interface ......................................................... 3
9.06 CrossXBeam ............................................................ 8
1.05 Monitor .................................................................. 3
9.07 Color Flow Mode .................................................... 9
2. System overview .................................................... 3
9.08 Power Doppler Imaging .......................................... 9
2.01 Applications ............................................................ 3
9.09 Auto Optimization .................................................. 9
2.02 Scanning methods .................................................. 4
9.10 M-Mode.................................................................. 9
2.03 Transducer types..................................................... 4
9.11 Pulse Wave Doppler Mode ................................... 10
2.04 Operating modes .................................................... 4
9.12 Continuous Wave Doppler (Option) ..................... 10
3. System standard features....................................... 4
9.13 CINE Memory/Image Memory ............................. 10
4. System Options ...................................................... 4
9.14 Image Storage ....................................................... 10
5. Peripheral Options ................................................. 5
9.15 Connectivity & DICOM.......................................... 11
6. Display modes ........................................................ 5
9.16 Virtual Convex ...................................................... 11
7. Selectable alternating Modes ................................ 5
9.17 LOGIQ View (Option) ............................................ 11
8. Display annotation ................................................. 6
9.18 Easy 3D (Option) ................................................... 11
8.01 General user interface ............................................ 6
9.19 Advanced 3D (Option) .......................................... 11
8.02 B-Mode ................................................................... 6
9.20 Static 3D/ Realtime 4D (option)............................ 11
8.03 Color Flow Mode .................................................... 6
9.21 B-flow (Option) ..................................................... 12
8.04 PDI mode ................................................................ 6
9.22 TVI (Option) .......................................................... 12
8.05 B-flow/ B-flow color Mode ..................................... 6
9.23 Stress Echo (Option) ............................................. 12
8.06 M-Mode .................................................................. 7
9.24 Follow-up tool....................................................... 12
8.07 Doppler Mode ........................................................ 7
9.25 Needle recognition (Option)................................. 12
8.08 3D/4D ..................................................................... 7
9.26 Scan Coach............................................................ 12
8.09 Easy 3D.................................................................... 7
9.27 My trainer ............................................................. 13
1
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

9.28 Battery (Option) .................................................... 13 13.11 6S-RS ................................................................... 20


10. Controls Available While “Live”............................ 13 13.12 12S-RS ................................................................. 20
11. Controls Available on “Freeze” or Recall .............. 13 13.13 RAB2-6-RS ........................................................... 21
12. Measurements/Calculations ................................ 14
12.01 General B-Mode ................................................. 14
12.02 General M-Mode ................................................ 14
12.03 General Doppler Measurements/Calculations ... 14
12.04 Real-time Doppler Auto
Measurements/Calculations......................................... 14
12.05 OB Measurements/Calculations ......................... 14
12.06 GYN Measurements/Calculations ....................... 15
12.07 Vascular Measurements/Calculations ................ 15
12.08 Urological Calculations ....................................... 15
12.09 Cardiac Measurements/Calculations .................. 16
12.09-1 B-Mode Measurements .................................. 16
12.09-2 M-Mode Measurements ................................. 16
12.09-3 Doppler Mode Measurements ........................ 17
12.09-4 Color Flow Mode Measurements .................... 18
12.09-5 Combination Mode Measurements ................ 18
12.10 Cardiac Worksheet.............................................. 19
13. Probes ..................................................................... 19
13.01 4C-RS................................................................... 19
13.02 L6-12-RS .............................................................. 19
13.03 L8-18i-RS ............................................................. 19
13.05 LK760-RS ............................................................. 19
13.06 E8C-RS................................................................. 19
13.07 E8Cs-RS ............................................................... 20
13.08 BE9CS-RS............................................................. 20
13.09 8C-RS................................................................... 20
13.10 3Sc-RS ................................................................. 20
2
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

Versana Premier™ ultrasound is designed to help – Integrated speakers


you apply your clinical and imaging skills to deliver – Probe holders, removable for clean and washing
high-quality care to the wide range of patients you – Gel holder, removable for clean and washing
see daily. This high-quality – Gel warmer
system is well suited for – Front and rear handles
general practice clinics, – Probe Cable management slots
physical check-up centers, – Easily removable air filters
community healthcare – Wheels - Locking mechanism that provides rolling lock
centers, scanning centers, and and caster swivel lock
other facilities worldwide that 1.04 User interface
offer basic medical care. It is Operator keyboard
versatile enough to cover a – Height adjustable
broad range of exams: – Full alphanumeric keypad covered with washable
abdominal, OB/GYN, cardiac, protection film
small parts, urology, vascular, – 10 TGC pods
pediatrics and MSK. – 9.9” (213.4 mm) touch screen

1. General specifications 1.05 Monitor


– 21.5” (482.6 mm) high-resolution
1.01 Dimensions and weight – Articulating monitor arm
Height with 21.5” monitor – Tilt/Rotate/Translate:
• Max 1780mm (70.1 in) • Tilt angle: +25˚/–90˚
• Min 1405mm (55.3 in) • Rotate angle: -130˚,50 ˚
Width • Translate horizontal 650 mm
• Keyboard: 593 mm (23.3 in) • Translate vertical 150 mm
• Caster: 717 mm (28.2 in) – Fold-down and lock mechanism for transportation
Depth – Brightness and contrast adjustment
• Maximum: 839 mm (33.0 in)
• Caster: 803 mm (31.6 in) 2. System overview
Weight (no Peripherals): less than 70 kg
2.01 Applications
1.02 Electrical power – Abdominal
– Voltage 100-240 VAC
– Obstetrical
– Frequency 50/60 Hz
– Gynecological
– Power consumption maximum of 450 VA with
– Small Parts
peripherals
– Musculoskeletal
1.03 Console design – Vascular/ Peripheral Vascular
– Max 4 active probe ports – Urological
– Integrated HDD (500GB) – Pediatric

3
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

– Transcranial – SRI-HD (High Definition Speckle Reduction Imaging)


– Cardiac – B-Steer
– Coded Phase Inversion Harmonic Imaging
2.02 Scanning methods
– Virtual Convex
– Electronic Convex
– Patient information Database
– Electronic Linear
– Image Archive on integrated HDD
– Electronic Micro Convex
– Raw Data Analysis
– Electronic Sector
– Voice comments
– Mechanic Volume Sweep
– Real-time automatic Doppler calculations
2.03 Transducer types – OB Calculations
– Convex Array – Sono Biometry (BPD/ HC/ AC/ HL/ FL)
– Linear Array – Fetal Trending
– Microconvex Array – Multi-gestational Calculations
– Sector Phased Array – Hip Dysplasia Calculations
– Volume Probes (4D) – Gynecological Calculations
– Bi-plane convex array – Vascular Calculations
2.04 Operating modes – Breast Productivity
– B-Mode – Urological Calculations
– Coded Phase Inversion Harmonic Imaging – Auto Bladder
– M-Mode Auto detect bladder contour in 2 orthogonal planes and calculate

– Anatomical M-Mode (Option) bladder volume

– Curve AMM (Option) – Renal Calculations


– Color M Mode – Cardiac Calculations
– Color Flow Mode (CFM) – On-board reporting package
– Power Doppler Imaging (PDI) – MPEGVue
– Directional PDI – Network Storage
– B-flow (B-flow color) (Option) – Remote capability: InSite ExC
– PW Doppler with High PRF – My trainer
– CW Doppler Mode (Option) – Scan Coach
– TVI Mode (Option) – Scan Assistant
– 3D/4D Volume Modes (Option) – Follow-up tool
– DICOM© 3.0 Connectivity
3. System standard features – Tricefy™ Uplink

– Installation wizard
4. System Options
– Whizz
Dynamical image optimization, Auto detection, Auto measurement and Auto – CW Doppler
annotation) Auto Bladder – Anatomical M-Mode
– CrossXBeam – Curve AMM

4
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

– LOGIQ View • B + CFM/M


– Advanced 3D (includes Easy 3D) • Real Time Triplex Mode (B + CFM or PDI+PW or CW)
– B-flow (includes B-flow color) • B-flow or B-flow color
– Tissue Velocity Imaging (TVI) with Q-Analysis – Zoom
– Stress Echo • Write (HD)/Read 20X
– Auto EF – Colorized Image
– Static 3D/ Real-time 4D • Colorized B
• TUI • Colorized B-flow
• VOCAL • Colorized M
– Auto IMT • Colorized PW
– Needle recognition • Colorized CW
• Colorized 3D
5. Peripheral Options • Colorized 4D
– Time line display
– Sony UP-D898MD B/W thermal printer
• Independent Dual B/PW or CW Display
– Sony UP-D898MD B/W thermal printer with fixture kit
• Display Formats
– Sony UP-898DC Printer Sony UP-D25MD Color thermal
– Top/Bottom selectable format (Size: 1/2:1/2;
printer
1/3:2/3; 2/3:1/3)
– 1-Pedal type footswitch ‘Whanam FSU-1000’
– Side/Side selectable format (Size: 1/2:1/2;
– Footswitch MKF 2-MED USB GP26
1/4:3/4; TL only)
– Keeber USB Stick 8G
– Virtual convex
– 1TB mobile USB HDD
– LOGIQ View
– Transcend TS8XDVDS-K DVDRW kit
– TUI
– USB ECG Kits (AHA/IEC)
– ECG Module(option)
– USB Wireless adaptor
7. Selectable alternating Modes
– Blue tooth adapter – B/M
– B/PW
6. Display modes – B + CFM/M
– B + CFM (PDI)/PW (CW)
– Live and Stored Display Format
– Freehand 3D-Mode
– Wide Screen
– Freehand 3D-Mode Color
• Full size and Split screen
– B/CW
• Both with thumbnails for still and Cine
– B + CFM (PDI)/CW
– Review Image Format:
– Multi-image split screen (quad screen)
• 4x4 and “thumbnails” for still and Cine
– Live and/or frozen
– Simultaneous Capability
– B + B/CFM or PDI
• Dual B (B/B)
– PW/M
• B/CFM or PDI
– Independent CINE playback
• B/PW or B/CWB/M

5
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

– Frame Average
8. Display annotation – Frame Rate
– Gray Map
8.01 General user interface
– ATO On /Off
– Patient Name: First, Last (Max 28 characters displayed
– SRI-HD
per each, Up to 64 total characters per each)
– CrossXBeam
– Patient ID (Max 64 characters)
– Other ID (Max 64 characters) 8.03 Color Flow Mode
– Age, Sex and Date of Birth – Line Density
– Hospital Name (Max 23 characters displayed) – Frame Average
– Date format: 3 Types selectable – Packet Size
• MM/DD/YYYY – Color Scale: 2 types
• DD/MM/YYYY • Power
• YYYY/MM/DD • Directional PDI
– Time format: 2 types selectable – Color Velocity Range and Baseline
• 24 hours – Color Threshold Marker
• 12 hours – Color Gain
– Gestational Age from – PDI
• LMP • GA – Inversion
• EDD • BBT – Doppler Frequency
– Displayed Acoustic Output
8.04 PDI mode
• TIs: Thermal Index Soft Tissue
– Line Density
• TIc: Thermal Index Cranial (Bone)
– Frame Averaging
• TIb: Thermal Index Bone
– Packet Size
• MI: Mechanical Index
– Directional PDI
– % of Maximum Power output
– Color Velocity Range and Baseline
– Probe Name
– Color Threshold Marker
– Map Names
– PDI Gain
– Probe Orientation
– Inversion
– Depth Scale Marker
– Lateral Scale Marker 8.05 B-flow/ B-flow color Mode
– Focal Zone Markers – Gain
– Image Depth – Background
– Zoom Depth – SRI HD
– Accumulation
8.02 B-Mode – Flow Module
– Gain
– Rejection
– Dynamic Range
– Colorize
– Imaging Frequency
– Edge Enhance
– Edge Enhance
– Gray Map
6
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

– Frame Average – Colorize


– Power OutPut – Direction
– Sensitivity/PRI – Rend 1 Gray
– Dynamic Range – Rend 2 Gray
– Frequency – Adv. Rend
– Suppression – Transparency
– FlashSuppression – Ref Image
– Enhance – Cut Mode
– Threshold – Cut Depth
– Map Compress – Depth
– Default Map – Rotational Angle
– Wall Filter – Step Angle
– Transparency – Rotation Axis
– Start Angle
8.06 M-Mode
– End Angle
– Gain
– Loop Speed
– Dynamic Range (Use the Dynamic Range of B mode)
– Time Scale 8.09 Easy 3D
– AMM – Utilities
– Curve AMM – Texture
– GreySurface
8.07 Doppler Mode
– Render
– Gain
– Threshold1
– Angle
– Threshold2
– Sample Volume Depth and Length
– ScanDistance
– Wall Filter
– Colorize
– Velocity and/or Frequency Scale
– Spectrum Inversion 8.10 Advance 3D
– Time Scale – DefineAxis
– Scale – Group Planes
– Doppler Frequency – Reslice
– Tile
8.08 3D/4D
– Tile/Mix
– Active Curve
9. General System Parameters
– Reset Curve
9.01 System Setup
– Lower Threshold
– 9 Pre-programmable Categories
– Volume Angle
– User Programmable Preset Capability
– B Quality
– Factory Default Preset Data
– 3D Orient
– Languages: English, Latin American Spanish, French,
– Render Mode
7
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

German, Italian, Brazilian Portuguese, Chinese – Line Density: 5/6 steps (depending on probe)
(Simplified), Swedish, Russian, Norwegian, Danish, – Line Density Zoom: 5 steps
Dutch, Finnish, – Thermal Index: TIc, TIs, TIb
– OB Report Formats including Tokyo Univ., Osaka Univ., – Image Reverse: On/Off
USA, Europe, and ASUM – Focus Number: 8 steps
– User Defined Annotations – Focus Width: 3 types
– Body Patterns – Suppression: 6 steps
– Customized Comment Home Position – Edge Enhance: 7 steps
– Rejection: 6 steps
9.02 System scanning parameters
– Steered Linear: ±12°±15˚ (probe dependent)
– Digital Agile Beamformer Architecture
– Scanning Size (FOV or Angle – depending on the probe)
– 225,792 system processing channels
– SRI-HD: Up to 6 Levels selectable
– Max. Frame Rate: 1447 F/s (depends on probes &
– CrossXBeam: Up to 7 Angles selectable
modes)
– Depth: 1– 33 cm, 0.5 or 1 cm step, Probe dependent
– Displayed Imaging Depth: 1 – 33 cm
– Minimum Depth of Field: 0 – 2 cm (Zoom) (probe 9.04 Coded Harmonic Imaging
dependent) – Coded Phase Inversion Harmonic Imaging
– Maximum Depth of Field: 0 – 33 cm (probe dependent) – Available on all Probes
– Transmission Focus: 1 – 8 Focal Points selectable (probe – Line Density: 5/6 steps (depending on probe)
and application dependent) – Line Density Zoom: 5 steps
– Quad Beamforming – Suppression: 6 steps
– Continuous Dynamic Receive Focus/Aperture – Edge Enhance: 7 steps
– Multi-Frequency/Wideband Technology – Gray Scale Map: 7 types
– Frequency Range: 1.7 to 18 MHz – Tint Map: 9 types
– 256 Shades of Gray – Gain: 0 – 90 dB, 1 dB step
– 224dB systematic Dynamic Range – Dynamic Range:
– Adjustable Dynamic Range (36 - 96dB) – 51 to 78 dB 3dB per step; 36~48/78~96 6dB per step;
– Adjustable Field of View (FOV): up to 168degree – Rejection: 6 steps
(depending on probe) – Frequency: Up to 4 steps, probe depended
– Image Reverse: Right/Left
9.05 SRI-HD
– Image Rotation of 0° 90°180° 270°
– High Definition Speckle Reduction Imaging Provides
9.03 B-Mode multiple levels of speckle reduction
– Acoustic Power Output: 0 – 100%, 2, 5, and 10 steps – Compatible with Side by Side DualView Display
– Gain: from 0 – 90 dB, 1 dB steps – Compatible with all linear, convex and sector transducers
– Adjustable Dynamic Range: 36 – 96 dB, 3 or 6 dB steps – Compatible with B-Mode, 3D/4D imaging
– Frame Averaging: 8 steps
9.06 CrossXBeam
– Gray Scale Map: 6 or 8 types probe dependent
– Provides 3, 5, 7 of spatial compounding
– B Colorization: 9 types
– Live Side by Side DualView Display
– Frequency: Up to 11 selectable (depending on probe)
– Compatible with:
8
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

• Color Mode – CF/PDI Angle Steer: 0, ±10˚, ±15˚, ±20˚


• PW – Packet Size: 8 – 24, dependent on probe and application
• SRI-HD – Spatial Filter: 6 steps
• Coded Harmonic Imaging – Frame Average: 7 steps
• Virtual Convex – PRF: 0.1 – 25 KHz/19 steps
– Available on 4C-RS, L6-12-RS, E8C-RS, 8C-RS, RAB2-6-RS, – Power Threshold: 0 – 100%, 10% steps
L8-18i-RS, BE9CS-RS, E8Cs-RS, 12L-RS, LK760-RS, RIC5- – Gain: 0 – 40 dB, 0.5 dB steps
9A-RS – Wall Filter: 4 steps depending on probe and application
– CF/PDI Frequency: Up to 15 steps, depending on probe
9.07 Color Flow Mode
– Transparent Map: 5 steps
– Baseline: 0 – 100%, 10% steps
– Invert: On/Off
– Invert: On/Off
– Accumulation: 8 steps
– CF/PDI Focus Depth: default pre-settable for 10 – 100%
of ROI in depth, 15% or 20% step 9.09 Auto Optimization
– CF/PDI Flash Suppression: 5 steps – Optimize B-Mode image to improve contrast resolution
– CF/PDI Angle Steer: 0, ±10˚, ±15˚, ±20˚ – Selectable amount of contrast resolution improvement
– Packet Size: 8 – 24, dependent on probe and Application (low, medium, high)
– Line Density: 5 steps – Auto-Spectral Optimize adjusts
– Line Density Zoom: 5 steps • Baseline
– Frame Average: 7 steps • Invert
– PRF: 0.1 – 25 KHz/19 steps • PRF (on live image)
– Spatial Filter: 6 steps
9.10 M-Mode
– Gain: 0 – 40 dB, 0.5 dB steps
– Gain: –20 -20 dB, 1 dB step
– Wall Filter: 4 steps, dependent on probe and Application
– Dynamic Range: 36 to 96dB, 36~48/78~96 6dB step;
– Scanning Size (FOV or Angle): Probe dependent
48~78 3dB step. (Use Dynamic Range of B mode)
– CF/PDI Vertical Size (mm) of ROI: default pre-settable
– Gray Scale Map: 6 or 8 types (Probe dependent)
– CF/PDI Center Depth (mm) of ROI: default pre-settable
– Colorization: 9 types
– CF/PDI Frequency: Up to 4 steps, depending on probe
– Scanning Size (FOV or Angle – depending on probe, see
– Color Maps, including velocity-variance maps: 14 types
probe specifications)
depending on Application
– Rejection: 6 steps
– Transparent Map: 5 steps
– M/PW Display Format: V-1/3B, V-1/2B, V-2/3B, H-1/2B,
– Color Threshold: 0 – 100%, 10% steps
H-1/4B, Timeline only (Vert 1/3B, Vert 1/2B, Vert 2/3B,
– Accumulation: 8 steps
Horiz 1/2B, Horiz 1/4B, TL Only)
– Flash Suppression
Anatomical M-Mode (Option)
9.08 Power Doppler Imaging – M-Mode cursor adjustable at any plane
– PDI Map: 14 types – Can be activated from a Cine loop from a live or stored
– CF/PDI Focus Depth: default pre-settable for 10 – 100% image
of ROI in Depth, 15% or 20% step – Measure & Analysis capability
– CF/PDI Acoustic Output: 0 – 100%, 2%, 5% or 10% step – Available with Color Flow Mode

9
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

Curve AMM • CW Colorization: Tint Map A/B/C/D/E/F


• Velocity Scale Range: 0.1~6105cm/s
9.11 Pulse Wave Doppler Mode
• Spectrum Inversion
– Acoustic Power: 0 – 100%, 2, 5, and 10 steps
• Trace Method: Max, Mean
– Gain: 0 -85 dB, 1 dB step
• Doppler Auto Trace: Frozen, Live, Off
– Gray Scale Map: Up to 8 types
• Trace Direction: Above, Below, Both
– PRF: 0.3 – 27.9 KHz
Trace Sensitivity: 0~40, step: 2
– Transmit Frequency: 1.7~10MHzMHz, probe depend
– Wall Filter: 5.5 – 5000Hz, 27 steps, dependent on probe 9.13 CINE Memory/Image Memory
– PW Colorization: Up to 6 types – 384 MB of Cine Memory
– Velocity Scale Range: 0.1-6670cm/s – Selectable Cine Sequence for Cine Review
– Sweep Speed: 0~7, 8 steps – Prospective Cine Mark
– Sample Volume Depth: 0.1~33 cm, probe depend – Measurements/Calculations and Annotations on Cine
– SV Gate: 1, 2, 3, 4, 5, 6, 7, 8, 10, 12, 14, 16 mm Playback
– Angle Correction: -90~90, 1 steps – Scrolling timeline memory
– M/PW Display Format: V-1/3B, V-1/2B, V-2/3B, H-1/2B, – Dual Image Cine Display
H-1/4B, Timeline only – Quad Image Cine Display
– Spectrum Inversion – Cine Gauge and Cine Image Number Display
– Duplex: Simultaneous: On/Off (PW only) – Cine Review Loop
– PW Angle Steer: 0, ±10, 15, 20˚ (use Angle Steer of B – Cine Review Speed: 11 steps (11, 13, 14, 17, 22, 25, 31,
mode) 48, 100, 200, 400%)
– Trace Method: Off, Max, Mean
9.14 Image Storage
– Baseline Shift: 11 steps
– On-board database of patient information
– Doppler Auto Trace
– Storage Formats:
– Compression: 0.5~2.4(0.5,0.7,0.9,1,1.1,1.4,1.6,2,2.4)
• DICOM – compressed/uncompressed, single/multi-
– Trace Direction: Above, Below, Both
frame, with/without Raw Data
– Trace Sensitivity: 0~40, 2 steps
• Export JPEG, WMV (MPEG 4) and AVI formats
9.12 Continuous Wave Doppler (Option) • DICOM Still Image Storage Size: ~2.1 MB
– Gray Scale Map: 8 types • Display Format: Full Size, 4x4 and “thumbnails”
– Baseline: 11 steps • Storage Devices:
– Angle Correct: ±90˚, 1˚ step o Internal Hard Drive Partition of 356 GB for
– Spectral Color: 6 types Image Storage
– Invert: On/Off o External USB HDD and USB Memory Stick
– Spectral Averaging: 5 steps Support for Import, Export, DICOM Read,
– Gain: 0 – 85 dB, 1 dB steps SaveAs, and MPEGVue
– Wall Filter: 5.5 – 5000Hz, 27 steps, dependent on probe o CD-RW storage: 700 MB
and application o DVD storage: -R (4.7 GB)
– CW-Mode includes: – Conversion to Formats: JPEG, AVI, WMV
• Transmit Frequency: 1.9,4.2,5,6.2 MHz – Live Image and stored image side-by-side Display

10
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

– Reload of archived data sets Medium/Average strong


– Network Storage support for Import, Export, DICOM – GreySurface 0-100%
Read, SaveAs, MPEGVue – Threshold1 0-255
– Threshold2 0-255
9.15 Connectivity & DICOM
– ScanDistance 1.0-15.0
– Ethernet network connection
– Colorize 0-360
• DICOM 3.0
• Verify 9.19 Advanced 3D (Option)
• Print – DefineAxis: select two points as start and end point of
• Store long Axis.
• Modality Worklist – Group Planes: OFF/Main/Parallel/Angular
• Storage Commitment – Reslice: Cube/Virtual Rescan/Cubic Plane
• Modality Performed Procedure Step (MPPS) – Tile:1/2/4/6
• Query/Retrieve
9.20 Static 3D/ Realtime 4D (option)
• Structured Reporting Template – which can be
– Available on RAB2-6-RS, RIC5-9A-RS
compared to vascular and OB standard
– Acquisition Modes:
• Remote capability InSite ExC
• Static 3D
9.16 Virtual Convex • Realtime 4D
– Provides a convex field of view Static 3D
– Compatible with CrossXBeam Static 3D mode
– Available on linear and Sector transducers Visualization Modes:
• 3D Rendering (Diverse surface and intensity
9.17 LOGIQ View (Option)
projection modes)
– Extended Field of View Imaging
• Sectional Planes (3 Section planes perpendicular to
– Available on 4C-RS, L6-12-RS, 8C-RS, 3Sc-RS, E8C-RS,
each other)
E8Cs-RS, RAB2-6-RS, L8-18i-RS, 6S-RS, BE9CS-RS, RIC5-
Render Mode
9A-RS, LK760-RS, 12L-RS, 12S-RS probes
– Surface texture, Surface Smooth, max-, min-, X-ray, Mix
– For use in B-Mode
Mode of two render Modes
– CrossXBeam is available on linear probes
– Display Format:
– Auto detection of scan direction
• Quad: A-/B-/C-Plane/3D
– Post-process zoom
• Dual: A-Plane/3D
– Rotation
• Single: 3D
– Auto fit on monitor
– Curved 3point Render start
– Measurements in B-Mode
3D Movie
– Up to 60cm scan length
• LoopSpeed:6-400
9.18 Easy 3D (Option) • Run/Stop
– Allows unlimited rotation and planar translations • Scalpel: 3D Cut tool
– 3D reconstruction from Cine sweep • Cut Mode: Inside Contour/Outside Contour/Inside
– Utilities: Average Off/Average Light/Average Box/Outside Box/Small Eraser/Big Erase

11
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

• Cut Depth: Full/UserDefine image


• Depth: 0-1000 – Available on the sector probes
• 3D Rotation Cine – Tissue color overlay can be removed to show just the 2D
• Rotational Angle:30-360 image, still retaining the tissue velocity information
• Step Angle1-15 – Q-Analysis: Multiple Time Motion trace display from
• Rotation Axis: X/Y selected points in the myocardium
• 3D Volume Review
9.23 Stress Echo (Option)
• Start Frame: define start Frame
– Advanced and flexible stress-echo examination
• End Frame: define end Frame.
capabilities
• Run/Stop
• Provides exercise and pharmacological protocol
Niche
templates
TUI: Tomographic ultrasound imaging
• 8 default templates
• Display Format: 1*1/1*2/2*2/3*3
• Template editor for user configuration of existing
• Slices: 3-19
templates or creation of new templates
• Slices Distance: 0.5-40mm
• Reference scan display during acquisition for stress
VOCAL (Option)
level
Virtual Organ Computer-aided Analysis
– Comparison (dual screen)
Realtime 4D
• Baseline level/Previous level selectable
4D Volume Cine
• Raw data continuous capture
• LoopSpeed:6-400
• Over 87.5 seconds available
• Run/Stop
• Wall motion scoring (bulls-eye and segmental)
9.21 B-flow (Option) • Smart stress: Automatically set up various scanning
– Available on 4C-RS, L6-12RS, 12L-RS probes parameters (for instance geometry, frequency, gain, etc.)
– Background: On/Off according to same projection on previous level
– Sensitivity/PRI: 1 – 50, 17 steps
9.24 Follow-up tool
– Line Density: 5 steps
– The Follow-up tool is intended to more accurately
– Edge Enhance: 7 steps
perform serial scans on a patient and compare the
– Frame Average: 0 – 7, 8 steps
images of a previous ultrasound exam to the current
– Gray Scale Map: 8 maps
exam.
– Tint Map: 5 maps
– Dynamic Range: 36 – 96 dB, 16 steps 9.25 Needle recognition (Option)
– Rejection: 6 steps – Needle recognition allows you to obtain precise needle
– Gain: 0 – 90 dB range, 1 dB steps imaging in the dashed box. it is available with linear
– B-Flow Color probes on L6-12L-RS, L8-18i-RS,12L-RS.
– Accumulation: 8 levels 9.26 Scan Coach
B-flow color – Scan Coach is a contextual reference tool. It is with
9.22 TVI (Option) clinical guidance for scan plane acquisition and
– Myocardial Doppler Imaging with color overlay on tissue references for anatomical structures.it can be displayed

12
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

on-demand by the user. Clinical reference images & • Length


animations to depict information related to each step. It • Depth
covers 5 applications: – Velocity Scale
• Abdomen – Color Flow Mode
• Obstetrics – CFM Gain
• Gynecology – CFM Velocity Range
• Cardiology – Acoustic Output
• Vascular – Wall Echo Filter
– Packet Size
9.27 My trainer
– Frame Rate Control
– Abstracted from basic user manual, it lists out FAQs from
– CFM Spatial Filter
customers and instructs customer how to solve problems
– CFM Frame Averaging
by themselves timely.
– Frequency/Velocity Base Line Shift
9.28 Battery (Option)
– The lithium ion battery provides power when an AC 11.Controls Available on “Freeze” or Recall
power source is not available. About 15 minutes of
– Automatic Optimization
battery life can be expected with fully charged battery in
– SRI-HD
use to supply power to the system.
– CrossXBeam – Display non-compounded and
compounded
10.Controls Available While “Live”
– Image simultaneously in split screen
– Write Zoom – 3D reconstruction from a stored Cine loop
– B/M/CrossXBeam-Mode – B/M/CrossXBeam Mode (CrossXBeam is disable on
– Gain “Freeze” or recall)
– TGC – Gray Map Optimization
– Dynamic Range – TGC
– Acoustic Output – Colorized B and M
– Transmission Focus Position – Frame Average (loops only)
– Transmission Focus Number – Dynamic Range
– Line Density Control – Anatomical M-Mode (Freeze only)
– Sweep Speed for M-Mode – Sweep Speed
– Number of Angles for CrossXBeam – Gray Map
– PW-Mode – Post Gain
– Gain – Baseline shift (PW,CW)
– Dynamic Range – Sweep Speed
– Acoustic Output – Invert Spectral wave form
– Transmission Frequency – Compression
– -Scale – Rejection
– Wall Filter – Colorized Spectrum
– Sample Volume Gate
13
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

– Display Format – AT (Acceleration Time)


– Doppler Audio – ACCEL (Acceleration)
– Angle Correct – TAMAX (Time Averaged Maximum Velocity)
– Quick Angle Correct – Volume Flow (TAMEAN and Vessel Area)
– Auto Angle Correct (can’t adjust in freeze) – Heart Rate
– Overall Gain (loops and stills) – PI (Pulsatility Index)
– Color Map – RI (Resistivity Index)
– Transparency Map
12.04 Real-time Doppler Auto
– Frame Averaging (loops only)
Measurements/Calculations
– Flash Suppression (disable in freeze)
– PS (Peak Systole)
– CFM Display Threshold
– ED (End Diastole)
– Spectral Invert for Color/Doppler
– MD (Minimum Diastole)
– Anatomical M-Mode on Cine loop
– PI (Pulsatility Index)
– RI (Resistivity Index)
12.Measurements/Calculations – AT (Acceleration Time)
– ACC (Acceleration)
12.01 General B-Mode
– PS/ED (PS/ED Ratio)
– Depth and Distance
– ED/PS (ED/PS Ratio)
– Circumference (Ellipse/Trace)
– HR (Heart Rate)
– Area (Ellipse/Trace)
– TAMAX (Time Averaged Maximum Velocity)
– Volume (Ellipsoid)
– PVAL (Peak Velocity Value)
– % Stenosis (Area or Diameter)
– Volume Flow (TAMEAN and Vessel Area)
– Angle between two lines
12.05 OB Measurements/Calculations
12.02 General M-Mode
– Gestational Age by:
– M-Depth
• GS (Gestational Sac)
– Distance
• CRL (Crown Rump Length)
– Time
• FL (Femur Length)
– Slope
• BPD (Biparietal Diameter)
– Heart Rate
• AC (Abdominal Circumference)
12.03 General Doppler • HC (Head Circumference)
Measurements/Calculations • APTD x TTD (Anterior/Posterior Trunk Diameter by
– Velocity Transverse Trunk Diameter)
– Time • FTA (Fetal Trunk Cross-sectional Area)
– A/B Ratio (Velocities/Frequency Ratio) • HL (Humerus Length)
– PS (Peak Systole) • BD (Binocular Distance)
– ED (End Diastole) • FT (Foot Length)
– PS/ED (PS/ED Ratio) • OFD (Occipital Frontal Diameter)
– ED/PS (ED/PS Ratio) • TAD (Transverse Abdominal Diameter)

14
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

• TCD (Transverse Cerebellum Diameter) 12.06 GYN Measurements/Calculations


• THD (Thorax Transverse Diameter) – Right Ovary Length, Width, Height
• TIB (Tibia Length) – Left Ovary Length, Width, Height
• ULNA (Ulna Length) – Uterus Length, Width, Height
– Estimated Fetal Weight (EFW) by: – Cervix Length, Trace
• AC, BPD – Ovarian Volume
• AC, BPD, FL, HC – ENDO (Endometrial thickness)
• AC, FL, HC – Ovarian RI
• BPD, APTD, TTD, FL – Uterine RI
– Calculations and Ratios – Follicular measurements
• FL/BPD
12.07 Vascular Measurements/Calculations
• FL/HC
– SYS DCCA (Systolic Distal Common Carotid Artery)
• CI (Cephalic Index)
– DIAS DCCA (Diastolic Distal Common Carotid Artery)
• CTAR (Cardio-Thoracic Area Ratio)
– SYS MCCA (Systolic Mid Common Carotid Artery)
– SonoBiometery
– DIAS MCCA (Diastolic Mid Common Carotid Artery)
• BPD
– SYS PCCA (Systolic Proximal Common Carotid Artery)
• HC
– DIAS PCCA (Diastolic Proximal Common Carotid Artery)
• AC
– SYS DICA (Systolic Distal Internal Carotid Artery)
• HL
– DIAS DICA (Systolic Distal Internal Carotid Artery)
• FL
– SYS MICA (Systolic Mid Internal Carotid Artery)
– Measurements/Calculations by: ASUM, ASUM 2001,
– DIAS MICA (Diastolic Mid Internal Carotid Artery)
Berkowitz,
– SYS PICA (Systolic Proximal Internal Carotid Artery)
– Bertagnoli, Brenner, Campbell, CFEF, Chitty, Eik-Nes,
– DIAS PICA (Diastolic Proximal Internal Carotid Artery)
Ericksen,
– SYS DECA (Systolic Distal External Carotid Artery)
– Goldstein, Hadlock, Hansmann, Hellman, Hill, Hohler,
– DIAS DECA (Diastolic Distal External Carotid Artery)
Jeanty,
– SYS PECA (Systolic Proximal External Carotid Artery)
– JSUM, Kurtz, Mayden, Mercer, Merz, Moore, Nelson,
– DIAS PECA (Diastolic Proximal External Carotid Artery)
Osaka
– VERT (Systolic Vertebral Velocity)
– University, Paris, Rempen, Robinson, Shepard,
– SUBCLAV (Systolic Subclavian Velocity)
Shepard/Warsoff,
– Automatic IMT
– Tokyo University, Tokyo/Shinozuka, Yarkoni
– Fetal Graphical Trending 12.08 Urological Calculations
– Growth Percentiles – Bladder Volume
– Multi-Gestational Calculations (4) – Prostate Volume
– Fetal Qualitative Description (Anatomical survey) – Lt/Rt Renal Volume
– Fetal Environmental Description (Biophysical profile) – Generic Volume
– Programmable OB Tables – Post-Void Bladder Volume
– Over 20 selectable OB Calculations
– Expanded Worksheets
15
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

12.09 Cardiac Measurements/Calculations • E-Point-to-Septum Separation (EPSS)


• Mitral Valve Area Planimetry (MVA Planimetry)
12.09-1 B-Mode Measurements
Pulmonic Valve
Aorta
• Pulmonic Diameter (Pulmonic Diam)
• Aortic Root Diameter (Ao Root Diam)
Right Ventricle
• Aortic Arch Diameter (Ao Arch Diam)
• Right Ventricle Internal Diameter (RVIDd, RVIDs)
• Ascending Aortic Diameter (Ao Asc)
• Right Ventricle Outflow Tract Diameter (RVOT Diam)
• Descending Aortic Diameter (Ao Desc Diam)
System Inferior Vena Cava
• Aorta Isthmus (Ao Isthmus)
• Systemic Vein Diameter (Systemic Diam)
• Aorta (Ao st junct)
Aortic Valve 12.09-2 M-Mode Measurements
• Aortic Valve Cusp Separation (AV Cusp) Aorta
• Aortic Valve Area Planimetry (AVA Planimetry) • Aortic Root Diameter (Ao Root Diam)
• (Trans AVA) • Aortic Valve Diameter (AV Diam)
Left Atrium • Aortic Valve Cusp Separation (AV Cusp)
• Left Atrium Diameter (LA Diam) • Aortic Valve Ejection Time (LVET)
• LA Length (LA Major) Left Atrium
• LA W idth (LA Minor) • Left Atrium Diameter to AoRoot Diameter Ratio
• Left Atrium Area (LAA(d), LAA(s)) (LA/Ao Ratio)
• Left Atrium Volume, Single Plane, Method of Disk • Left Atrium Diameter (LA Diam)
(LAEDV A2C, LAESV A2C) (LAEDV A4C, LAESV A4C) • Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI
Left Ventricle Ds)
• Left Ventricle Mass (LVPWd, LVPWs) • Left Ventricle Posterior Wall Thickness (LVPWd,
• Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI LVPWs)
Ds) • Left Ventricle Ejection Time (LVET)
• Left Ventricle Internal Diameter (LVIDd, LVI Ds) • Left Ventricle Pre-Ejection Period (LVPEP)
• Left Ventricle Length (LVLd, LVLs) • Interventricular Septum (IVS)
• Left Ventricle Outflow Tract Diameter (LVOT Diam) Mitral Valve
• Left Ventricle Posterior Wall Thickness (LVPWd, • E-Point-to-Septum Separation (EPSS)
LVPWs) • Mitral Valve Anterior Leaflet Excursion (D-E
• Left Ventricle Length (LV Major) Excursion)
• Left Ventricle Width (LV Minor) • Mitral Valve D-E Slope (D-E Slope)
• Left Ventricle Outflow Tract Area (LVOT) • Mitral Valve E-F Slope (E-F Slope)
• Left Ventricle Mass Index (LVPWd, LVPWs) Pulmonic Valve
• Ejection Fraction, Teichholz/Cube (LVIDd, LVIDs) • QRS complex to end of envelope (Q-to-PV close)
• Left Ventricle Posterior Wall Fractional Shortening • Right Ventricle Internal Diameter (RVIDd, RVIDs)
(LVPWd, LVPWs) • Right Ventricle Outflow Tract Diameter (RVOT Diam)
• Mitral Valve • Right Ventricle Ejection Time (RVET)
• Mitral Valve Annulus Diameter (MV Ann Diam) • Right Ventricle Pre-Ejection Period (RVPEP)

16
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

Tricuspid Valve • Mitral Valve Velocity Peak E (MV E Velocity)


• QRS complex to end of envelope (Q-to-TV close) • Mitral Valve Area according to PHT (MV PHT)
• Mitral Valve E-Peak to A-Peak Ratio (A-C and D-E)
12.09-3 Doppler Mode Measurements
(MV E/ARatio)
Aortic Valve
• Mitral Valve Acceleration Time
• Aortic Valve Mean Velocity (AV Trace)
• (MV Acc Time)
• Aortic Valve Velocity Time Integral (AV Trace)
• Mitral Valve Deceleration Time (MV Dec Time)
• Aortic Valve Mean Pressure Gradient (AV Trace)
• Mitral Valve Acceleration Time/Deceleration Time
• Aortic Valve Peak Pressure Gradient (AR Vmax)
Ratio (MVAcc/Dec Time)
• Aortic Insufficiency Peak Velocity (AR Vmax)
Pulmonic Valve
• Aortic Insufficiency End-Diastolic Velocity (AR Trace)
• Pulmonic Insufficiency Peak Pressure Gradient (PR
• Aortic Valve Peak Velocity (AV Vmax)
Vmax)
• Aortic Valve Deceleration TIme (AV Trace)
• Pulmonic Insufficiency End-Diastolic Pressure
• Aortic Valve Ejection Time (AVET)
Gradient (PRTrace)
• Aortic Valve Area according to PHT
• Pulmonic Valve Peak Pressure Gradient (PV Vmax)
Left Ventricle
• Pulmonic Insufficiency Peak Velocity (PR Vmax)
• Left Ventricle Outflow Tract Peak Pressure Gradient
• Pulmonic Insufficiency End-Diastolic Velocity (Prend
(VLOT Vmax)
Vmax)
• Left Ventricle Outflow Tract Peak Velocity (LVOT
• Pulmonic Valve Peak Velocity (PV Vmax)
Vmax)
• Pulmonary Artery Diastolic Pressure (PV Trace)
• Left Ventricle Outflow Tract Mean Pressure Gradient
• Pulmonic Insufficiency Mean Pressure Gradient (PR
(LVOT Trace)
Trace)
• Left Ventricle Outflow Tract Velocity Time Integral
• Pulmonic Valve Mean Pressure Gradient (PV Trace)
(LVOT Trace)
• Pulmonic Insufficiency Mean Square Root Velocity
• Left Ventricle Ejection Time (LVET)
(PR Trace)
Mitral Valve
• Pulmonic Insufficiency Velocity Time Integral (PR
• Mitral Valve Regurgitant Mean Velocity (MR Trace)
Trace)
• Mitral Regurgitant Mean Pressure Gradient (MR
• Pulmonic Valve Mean Velocity (PV Trace)
Trace)
• Pulmonic Valve Velocity Time Integral (PV Trace)
• Mitral Regurgitant Velocity Time Integral (MR Trace)
• Pulmonic Insufficiency Pressure Half Time (PR PHT)
• Mitral Valve Mean Velocity (MR Trace)
• Pulmonic Valve Flow Acceleration (PV Acc Time)
• Mitral Valve Velocity Time Integral (MR Trace)
• Pulmonic Valve Acceleration Time (PV Acc Time)
• Mitral Valve Mean Pressure Gradient (MR Trace)
• Pulmonic Valve Ejection Time (PVET)
• Mitral Regurgitant Peak Pressure Gradient (MR
• QRS complex to end of envelope (Q-to-PV close)
Vmax)
• Pulmonic Valve Acceleration to Ejection TIme Ratio
• Mitral Valve Peak Pressure Gradient (MR Vmax)
(PV Acc Time, PVET)
• Mitral Regurgitant Peak Velocity (MR Vmax)
Right Ventricle
• Mitral Valve Peak Velocity (MR Vmax)
• Right Ventricle Outflow Tract Peak Pressure
• Mitral Valve Velocity Peak A (MV A Velocity)
Gradient (RVOT Vmax)
17
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

• Right Ventricle Outflow Tract Peak Velocity (RVOT • Tricuspid Valve Velocity Time Integral (TV Trace)
Vmax) • Tricuspid Valve Time to Peak (TV Acc/Dec Time)
• Right Ventricle Outflow Tract Velocity Time Integral • Tricuspid Valve Ejection Time (TV Acc/Dec Time)
(RVOTTrace) • Tricuspid Valve A-Wave Duration (TV A Dur)
• Right Ventricle Ejection Time (RV Trace) • QRS complex to end of envelope (Q-to-TV close)
• Stroke Volume by Pulmonic Flow (RVOT Planimetry, • Tricuspid Valve Pressure Half Time (TV PHT)
RVOTTrace) • Tricuspid Valve E-Peak to A-Peak Ratio (TV E/A
• Right Ventricle Stroke Volume Index by Pulmonic Velocity)
Flow (RVOT Planimetry, RVOT Trace)
12.09-4 Color Flow Mode Measurements
System
Aortic Valve
• Pulmonary Artery Peak Velocity (PV Vmax)
• Proximal Isovelocity Surface Area: Regurgitant Flow
• Pulmonary Vein Velocity Peak A (reverse) (P Vein A)
(AR Trace)
• Pulmonary Vein Peak Velocity (P Vein D, P Vein S)
• Proximal Isovelocity Surface Area: Regurgitant
• Systemic Vein Peak Velocity (PDA Diastolic, PDA
Volume Flow (AR Trace)
Systolic)
• Proximal Isovelocity Surface Area: Aliased Velocity
• Ventricular Septal Defect Peak Velocity (VSD Vmax)
(AR Vmax)
• Atrial Septal Defect (ASD Diastolic, ASD Systolic)
Mitral Valve
• Pulmonary Vein A-Wave Duration (P Vein A Dur)
• Proximal Isovelocity Surface Area: Regurgitant Flow
• IsoVolumetric Relaxation Time (IVRT)
(MR Trace)
• IsoVolumetric Contraction Time (IVCT)
• Proximal Isovelocity Surface Area: Regurgitant
• Pulmonary Vein S/D Ratio (P Vein D, P Vein S)
Volume Flow (MR Trace)
• Ventricular Septal Defect Peak Pressure Gradient
• Proximal Isovelocity Surface Area: Aliased Velocity
(VSD Vmax)
(MR Vmax)
• Pulmonic-to-Systemic Flow Ratio (Qp/Qs)
Tricuspid Valve 12.09-5 Combination Mode Measurements
• Tricuspid Regurgitant Peak Pressure Gradient (TR Aortic Valve
Vmax) • Aortic Valve Area (Ao Diam, LVOT Vmax, AV Vmax)
• Tricuspid Valve Peak Pressure Gradient (TV Vmax) • Aortic Valve Area by Continuity Equation by Peak
• Tricuspid Regurgitant Peak Velocity (TR Vmax) Velocity (Ao Diam, LVOT Vmax, AV Vmax)
• Tricuspid Valve Peak Velocity (TV Vmax) • Stroke Volume by Aortic Flow (AVA Pl ani met ry, AV
• Tricuspid Valve Velocity Peak A (TV A Velocity) Trace)
• Tricuspid Valve Velocity Peak E (TV E Velocity) • Cardiac Output by Aortic Flow (AVA Planimetry, AV
• Tricuspid Regurgitant Mean Pressure Gradient (TR Trace, HR)
Trace) • Aortic Valve Area by Continuity Equation VTI (Ao
• Tricuspid Valve Mean Pressure Gradient (TV Trace) Diam, LVOT Vmax, AV Trace)
• Tricuspid Regurgitant Mean Velocity (TR Trace) Left Ventricle
• Tricuspid Regurgitant Velocity Time Integral (TR • Cardiac Output, Teichholz/Cubic (LVIDd, LVI Ds, HR)
Trace) Mitral Valve
• Tricuspid Valve Mean Velocity (TV Trace) • Stroke Volume by Mitral Flow (MVA Planimetry, MV

18
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

Trace) • Applications MSK Superficial


• Cardiac Output by Mitral Flow (MVA Planimetry, MV • Number of Element: 168
Trace, HR) • Footprint: 34.8 X 11.1 mm
• B-Mode Imaging Frequency:
12.10 Cardiac Worksheet
8.0,12.0,14.0,16.0 MHz
• Parameter: lists the mode, the measurement
• Harmonic Imaging Frequency:
folder and the specific measurement
9.0,15,18.0 MHz
• Measured Value: Up to six measurement values for
• CFM/PDI/PWD Frequency: 6.7,8.3,10 MHz
each item. Average, maximum, minimum, or last
• Steered Angle: +/-20°

13. Probes • 13.04 12L-RS


Linear Probe
13.01 4C-RS • Applications Vascular, Small Parts, Pediatrics, MSK
Convex Probe • Number of Element: 192
• Applications: Abdomen, OB/Gyn, Vascular, Urology • Footprint: 47.1 x 12.7mm
• Number of Element: 128 • B-Mode Imaging Frequency:
• Convex Radius: 60 mmR 6.0, 8.0, 10.0, 12.0 MHz
• FOV: 58° • Harmonic Imaging Frequency:
• Footprint: 66.2 x 18.3 mm 8.0, 10.0, 12.0, 13.0 MHz
• B-Mode Imaging Frequency • CFM/PDI/PWD Frequency: 4.2, 6.3, 7.7 MHz
2.0, 3.0, 4.0, 5.0 MHz • Steered Angle: +/-20°
• Harmonic Imaging Frequency: 13.05 LK760-RS
3.0, 4.0, 5.0 MHz Linear Probe
• CFM/PDI/PWD Frequency: 2.0(CFM/PDI), 2.5, 2.8, • Applications MSK
3.6 MHz • Number of Element: 128
13.02 L6-12-RS • Footprint: 67.0 x 13.0 mm
Linear Probe • B-Mode Imaging Frequency:
• Applications Vascular, Small Parts, Pediatrics 5.0, 7.0, 9.0MHz
• Number of Element: 128 • Harmonic Imaging Frequency:
• Footprint: 47 x 11.4 mm 6.0, 8.0, 10.0MHz
• B-Mode Imaging Frequency: • CFM/PDI/PWD Frequency: 3.5, 4.2, 5.0 MHz
6.0, 8.0, 10.0, 11.0 MHz • Steered Angle: +/-10°
• Harmonic Imaging Frequency: 13.06 E8C-RS
8.0, 10.0, 12.0, 13.0 MHz Endo Micro Convex Probe
• CFM/PDI Frequency: 4.0, 5.0, 6.0 MHz • Applications: OB/Gyn, Urology, Transvaginal,
• PWD Frequency: 4.0, 4.5, 5.0MHz Transrectal
• Steered Angle: +/-20° • Number of Element: 128
13.03 L8-18i-RS • Convex Radius: 10.73 mmR
Linear Probe • FOV: 128°

19
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

• Footprint: 16.9 x 21.2 mm • Convex Radius: 10.73 mmR


• B-Mode Imaging Frequency: 6.0, 8.0, 10.0 MHz • FOV: 131°
• Harmonic Imaging Frequency: • Footprint: 22.0 x 12.0mm
7.0, 8.0, 10.0 MHz • B-Mode Imaging Frequency: 6.0, 8.0, 10.0 MHz
• CFM/PDI/PWD Frequency: 4.2, 5.0, 6.3 MHz • Harmonic Imaging Frequency:
• Biopsy Guide: Fixed Angle Reusable Bracket 6.0, 7.0, 8.0, 10.0 MHz
• CFM/PDI/PWD Frequency: 4.2, 5.0, 6.3 MHz
13.07 E8Cs-RS
Endo Micro Convex Probe 13.10 3Sc-RS
• Applications: OB/Gyn, Urology, Transvaginal, Phased Array Sector Probe
Transrectal • Applications: Cardiac, Transcranial
• Number of Element: 128 • Number of Element: 64
• Convex Radius: 8.73 mm • FOV: 120°
• FOV: 168° • Footprint: 23.7 x 18.4 mm
• Footprint: 18.6 X 13.9 mm • B-Mode Imaging Frequency: 2.0, 3.0, 4.0 MHz
• B-Mode Imaging Frequency: • Harmonic Imaging Frequency:
6.0,8.0,10.0 MHz 3.0, 3.2, 3.5, 4.0 MHz
• Harmonic Imaging Frequency: • CFM/PDI/PWD Frequency: 1.7, 2.0, 2.5, 3.3 MHz
7.0,8.0,10.0 MHz • CWD Frequency: 1.9 MHz
• CFM/PDI/PWD Frequency: 4.0,5.0,6.0 MHz
13.11 6S-RS
• Biopsy Guide: Fixed Angle Reusable Bracket
Phased Array Sector Probe
13.08 BE9CS-RS • Applications: Cardiac, Transcranial
Simultaneous Bi-plane, Micro convex probe • Number of Element: 64
• Applications: Urology, Transrectal • FOV: 120°
• Number of Element: 96 • Footprint: 23.5 X 16.8 mm
• Convex Radius: 9 mm • B-Mode Imaging Frequency: 4.0,5.0,6.0 MHz
• FOV: 133° • Harmonic Imaging Frequency: 4.0,5.0,6.0, 7.0 MHz
• Footprint: 19 x 18.7 mm • CFM/PDI/PWD Frequency: 3.0,4.0,4.5 MHz
• B-Mode Imaging Frequency:
13.12 12S-RS
6.0,8.0,10.0 MHz
Phased Array Sector Probe
• Harmonic Imaging Frequency:
• Applications: Neonatal, Transcranial
8.0,9.0,10.0 MHz
• Number of Element: 96
• CFM/PDI Frequency: 4.0,5.0,6.3 MHz
• FOV: 120°
• PW Frequency: 4.2,5.0,6.3 MHz
• Footprint: 17.6 X 13.2 mm
• Biopsy Guide: Fixed Angle Reusable Bracket
• B-Mode Imaging Frequency: 6.0,8.0,10.0,11.0 MHz
13.09 8C-RS • Harmonic Imaging Frequency: 9.0,10.0,12.0 MHz
Micro Convex Probe • CFM/PDI/PWD Frequency: 4.5,5.0,5.6,6.7 MHz
• Applications: Pediatrics • CW Frequency: 4.2,5.0,6.2 MHz
• Number of Element: 128
20
Versana PremierTM Ultrasound System
Specification sheet Rev4

2018-Oct-30
GE Healthcare

13.13 RAB2-6-RS devices—Part 1 Evaluation and testing


Convex Volume Probe • EN 62366 Medical devices —Application of
• Applications: Abdomen, OB/GYN, Urology usability engineering to medical devices
• Number of Element: 128
• Convex Radius: 47.1 mmR About GE Healthcare
• Footprint: 62.2 x 34.0 mm GE Healthcare provides transformational medical
• Volume Sweep Radius: 24.11 mm technologies and services to meet the demand for increased
• FOV: 70 (B), 84° x 70° (Volume scan) access, enhanced quality and more affordable healthcare
• B-Mode Imaging Frequency: 3.0, 4.0, 5.0 MHz around the world. GE (NYSE: GE) works on things that matter
• Harmonic Imaging Frequency: 4.0, 5.0, 6.0 MHz - great people and technologies taking on tough challenges.
• CFM/PDI/PWD Frequency: 2.0, 3.0, 4.0 MHz From medical imaging, software & IT, patient monitoring and
diagnostics to drug discovery, biopharmaceutical
Inputs and Outputs manufacturing technologies and performance improvement
– CVBS output (RCA) solutions, GE Healthcare helps medical professionals deliver
– S-Video output great healthcare to their patients.
– VGA output (1920*1080 resolution)
– HDMI output (1920*1080 resolution) GE Healthcare World Headquarters
– Audio line-out (RCA) 9900 Innovation Drive
– Ethernet (RJ45) Wauwatosa, WI, 53226
– USB (3x in rear, 1 beside touch panel) USA
www.gehealthcare.com
Safety Conformance
– The Versana PremierTM is CE Marked to Council Directive © 2018 General Electric Company.

93/42/EEC on Medical Devices Dec 2018/DOC2129260


– Conforms to the following standards for safety:
USCAN Version: JB63215US
• IEC 60601-1 Medical electrical equipment—Part 1:
GE, the GE Monogram, and Versana are trademarks of General Electric Company.
General requirements for safety (basic safety and
Trademark of General Electric Company. DICOM is the registered trademark of the
essential performance)
National Electrical Manufacturers Association for its standard publications relating to
• IEC 60601-1-2 Medial electrical equipment—Part
digital communications of medical information. GE Medical Systems Ultrasound &
1-2 General requirements for safety (basic safety
Primary Care Diagnostics, LLC, a General Electric Company, doing business as GE
and essential performance)—Collateral Standard:
Healthcare.
Electromagnetic disturbances—requirements and
tests EMC Emissions Group 1 Class A device Reproduction in any form is forbidden without prior written permission from GE.

requirements as per CISPR 11 Nothing in this material should be used to diagnose or treat any disease or condition.

• IEC 60601-2-37 Medical electrical equipment—Part Readers must consult a healthcare professional.

2-37: Particular requirements for the safety (basic


safety and essential performance) of ultrasonic
medical diagnostic and monitoring equipment
• ISO 10993-1 Biological evaluation of medical
21