You are on page 1of 2
. «@ 2 @ @2 @ @ @2 @ ®@ Gas Chromatographs Detectors The SRI GC does not have a standard detector configuration. You choose the detector(s) you want installed. Usually you can satisfy the majority of your analytical needs with a single detector, but either SRI GC chassis can accommodate as many as four detectors to give you maximum flexibility. Table 1 includes a listing of the detectors that can be installed on your SRI GC. The table shows the typical detection limits for each detector along with their common applications. If you are not sure of which detector is needed for your specific analysis, do not hesitate to contact our technical support for assistance, Table 1 Dersrox Dexecton Lint Arrucaron Tc 100ppm _Detects any compound, Thermal Conductivity used mostly for fixed gas analysis. FD Ippm Detects organic Flame Ionization compounds having a carbon-hydrogen bond. NPD ippm Detects compounds Nitrogen Phosphorous containing nitrogen and phosphorus atoms. ECD ppb Detects halogenated Flectron Capture compounds. PID 10ppb _ Detects aromatics and Photo Ionization any compound with a carbon-carbon double bond. FPD S0ppb P/ Simultaneously detects Flame Photometric 500ppb S_— compounds containing sulfur or phosphorus atoms. DFPD Soph P/ Simultaneously detects Dual FPD 500ppb $ compounds containing sulfur or phosphorus atoms, DELCD 1Oppb Detects halogenated Dry Electrotytic compounds. Conductivity HD ‘ippm Detects any compound, Helium Ionization used mostly for fixed gas analysis. ™ TOppb Detects compounds Thermionic Ionization containing nitro functional groups or chlorinated phenols. cep 100ppm Detects hydrogen and Catalytic Combustion hydrocarbons. CHROMATOGRAPHY TTD oe