Professional Documents
Culture Documents
How To Obtain Sample Potential Data For KPFM Measurement 18 PDF
How To Obtain Sample Potential Data For KPFM Measurement 18 PDF
www.parksystems.com
INTRODUCTION
Fundamentals of KPFM
Kelvin Probe Force Microscopy, or KPFM, was introduced as a tool
to measure the local contact potential difference between a
conducting atomic force microscopy (AFM) tip and the sample,
thereby mapping the work function or surface potential of the
sample with high spatial resolution. Since its first introduction by
Nonnenmacher [1], KPFM has been used extensively as a unique
method to characterize the nanoscale electrical properties of metal
or semiconductor surfaces and semiconductor devices. Recently, + ++ - --
KPFM has also been used to study the electrical properties of
organic materials, devices [2–4], and biological materials. To The KPFM measures Contact Potential Difference (CPD) between a
eliminate any confusion, let us look into KPFM’s synonyms for this conducting AFM tip and a sample. The CPD (VCPD) between the tip and
technique: sample is defined as:
tip ϕsample
• KPFM: Kelvin Probe Force Microscopy VCPD (2.1)
• SKPM: Scanning Kelvin Probe Microscopy e
• SSPM: Scanning Surface Potential Microscopy
• SKFM: Scanning Kelvin Force Microscopy where ϕsample and ϕtip are the work functions of the sample and tip,
• SPM: Surface Potential Microscopy and e is the electronic charge. The different Fermi energy levels
• SP-AFM: Surface Potential Atomic Force Microscopy between the AFM tip and sample surface causes an electrical force as
the AFM tip is brought close to the sample surface. Figure 1 shows the
KPFM will be used in this document, as it is the most widely used energy level diagram of the tip and sample surface where ϕsample and
descriptor for this technique. The term ’Kelvin force’ refers to ϕtip are different. Figure 1(a) depicts the energy levels of the tip and
similarities between this microscopic technique and the macroscopic sample surface when separated tip and sample surface are close
technique, which is the Kelvin probe method. However, the enough for electron tunneling, equilibrium of the states require Fermi
methodology is somewhat different, but the measured value is levels to line-up at steady state. Upon electrical contact, the Fermi levels
equivalent for both techniques. For clarity, this note will refer only will align through electron current flow, and the system will reach to an
to the microscopic technique KPFM. equilibrium state as shown in Figure 1(b).
C ( z ) 1 2
F2 V AC cos(2t ) 1 (2.7)
Z 4
The sample consisted of three different materials: Au, Si, and Al. For
the calibration surface sample, Au was selected to become the base
material. The work function of Si and Al were determined as