Professional Documents
Culture Documents
PRESETS
BODY MARK SEL (ABD) (FETAL) (OB/GYN) (S ORG) (P.V.) (CARD) (ORTHOP) Default anatomical region to be displayed when Body Mark is
(RGST) turned ON
CHARACTER EXP (OFF) (ON) Whether the annotation is displayed in double height and
width size as the default when Annotation is turned ON
COM PORT SEL (EXTERNAL) (INTERNAL) Input/output of RS-232C
C PALETTE ( 1) ( 2) ( 3) ( 4) ( 5) ( 6) ( 7) ( 8) Default color pallet to be set when Color Palette is turned ON
HIGH DENSITY (OFF) (ON) Whether high density mode is set as the default
HIGH FRAME (OFF) (ON) Whether high frame mode is set as the default
M SPEED (x1) (x2) (x4) (x8) Scroll speed for the M-Time mark
No. 2D730-154E
M TIME MARK (OFF) (ON) Whether the M-Time mark is displayed or not
FU CAPA SEL ( 0) ( 1) ( 2) Image compression ratio for image storage
This is displayed only when Filing Unit is selected for
PROTOCOL.
MULTI FREQ (PENE) (NORMAL) (RESO) Default type of multi-frequency
Item Options Description
NEEDL GUIDE A (OFF) (ON1) (ON2) (ON3) (OFF1) (OFF2) (OFF3) Whether Type A biopsy mark is displayed as the default type
For Famio5, only one biopsy mark (Type 1) is supported.
NEEDL RANGE (OFF) (ON) Whether the range of the Type A biopsy mark is displayed or
not
NEEDL GUIDE B (OFF) (ON) Whether the Type B biopsy mark is displayed as the default
or not
PROBE SELECT (A) (B) Default transducer selection
If no transducer is connected to the default connector and a
transducer is connected to the other connector, the
connected one is selected.
5-2
No. 2D730-154E
5.2 MEASURE Page
No. 2D730-154E
measurement is performed multiple times for an item
Item Options Description
OB DATE INPUT (LMP) (EDD) (U LMP) (PRV) (CLN) (U PRV) Default gestational age input
One of (LMP), (EDD), and (U LMP), and one of (PRV), (CLN),
and (U PRV) can be selected using the Next switch.
For (U LMP) and (U PRV), the method registered for Birth is
selected.
OB GA DISP (OFF) (ON) Whether GA is displayed in the measurement window
OB GRAPH (OFF) (ON) Whether the trend graph is displayed during OB
measurement
OB IDLIST KEY (ID NO) (NAME) (EDD) Default sorting item for the ID List on the ID registration
screen
OB ITEM SEL — Startup of the OB Item Select registration function
OB MEAS UNIT (mm) (cm) Unit for the measurement values displayed on the Report and
Worksheet screens
5-4
OB MIN DISP (OFF) (ON) Whether the measurement results are displayed in a single
line only
When this is set to ON, All Display is disabled.
OB REPRT FORM (BASIC) (BASIC+ANAT) (ADV) (ADV+ANAT) OB report display format
Basic:
Only the basic items (maximum 8 items) are displayed.
Basic + Anatomy:
The basic items and Anatomy page are displayed.
Advanced:
The Advanced items (maximum 16 items) are displayed.
Advanced + Anatomy:
No. 2D730-154E
The Advanced items and Anatomy page are displayed.
OB SD DISP (OFF) (ON) Whether SDs are displayed in the measurement window
Item Options Description
OB TREND DISP (CURRENT) (PREVIOUS) (CUR+PREV) Default data to be displayed in the trend graph
Current:
Only the data obtained for the current examination is
plotted.
Previous:
Only the data for the previous examination is plotted.
Current + Previous: Both sets of data are plotted together.
OB TREND FORM (SINGLE) (QUAD) Default format of the trend graph
OB TREND GRID (OFF) (ON) Whether the grid is displayed or not for the trend graph
OB U/S DISP (OFF) (ON) Whether U/S GA is displayed or not in the measurement
window
OB WEEK DISP ( w d) (w t ) (w d ) ( . w) GA display format
OB WEEK SEL (OFF) (CURRENT) (AVERAGE) (U/S) GA calculation method to be used for the Weeks of
5-5
Pregnancy function
OFF: The Week switch is not displayed.
Current: The last measured GA value is used.
Average:
The mean of the GA values obtained in the most recent
measurement is used.
U/S:
The mean of the values obtained for the item that is
specified for U/S GA calculation is used.
OB WT. UNIT ( g ) (lbs) ( g +lbs) (g + GA) Unit for EFW
SIMP TRACE (TRACE) (SPLINE) Trace method to be used for LV MOD measurement
No. 2D730-154E
5.3 SERVICE Page
When the Preset menu is displayed, set to ON and then press < .
,
DEPTH>6.0MHz ( 4 cm) ( 6 cm) ( 8 cm) (10 cm) (12 cm) Default depth when a transducer with a frequency higher than
6.0 MHz is connected
ECHO BASE UP (OFF) ( A) ( B) Whether the color of the background of the ultrasound image
is set to gray for easier viewing
END OF TEXT (ETX) (CR) (NONE) (CR+LF) Packet end character string
ID DISP LINE (SECOND LINE) (THIRD LINE) The line to display the patient ID
INTERPOLATION (OFF) (ON) Whether image raster interpolation is performed or not
Normally, this item should be preset to ON.
LANGUAGE (ENGLISH-UK) (FRENCH) (GERMAN) (ENGLISH-US) Language to be displayed on the system
NAK (0X0F) (0X15) Whether the NAK message is handled as 0x0F or 0x15.
No. 2D730-154E
(Usually, NAK is handled as 0x0F.)
PARAMETER (8, STP1, NON) (7, STP1, EVN) Communication parameters
(Data bit, stop bit, parity)
Item Options Description
PRINT A (PRINTER) (FU) (SEND) (CAPTURE) The function to be set when the Print A switch is pressed
(Changed)
PRINT B (PRINTER) (FU) (SEND) (CAPTURE) The function to be set when the Print B switch is pressed
(Changed)
PROTOCOL (STANDARD) (FU) (RSVD) RS-232C communication protocol
Standard: α protocol
Filing Unit: Filing unit control
(Changed)
START OF TEXT (STX) (NONE) Packet start character string
TERM CHAR (ETX) (CR) (NULL) (ACK) (NAK) Character string for end of communication
TIME FORM (12 H) (24 H) Time display format
TRNS SEQ (FU) Startup of measurement code setting function
5-7
VERSION DISP (OFF) (ON) Whether the software version, copyright, and copyright year
are displayed at power ON
No. 2D730-154E
*