TECHNICAL DATA
Classification in accordance with the Council Directive 93/42) Class IIb
EEC concerning medical devices, appendix IX
Protection type in accordance with IEC 60529 IPX1 (Protection against
vertically dripping water)
|Classification in accordance with IEC 60601-1 Type B applied part
Protection class in accordance with IEC 60601-1 Protection class | device (with
protective ground connection)
UMDNS code 17-648
Dimensions (W x H x D) approx. 200 x 290 x 330 mm
Weight (full) approx. 9 kg
Tank capacity (minimum/maximum markings) approx. 0.7- 1.41
Noise level approx. 35 dB(A) (1m)
Alarm level > 65 dB(A) (3 m)
eed sae ices
Sensor measuring range approx. 7 - 50°C.
Measuring tolerance: Temperature display — water temperature < +0.1°C.
Scanned with CamScannerCLUE Temcuy cer ic) CCT Te
Correction value: Water temperature - temperature display +0.5°C
Setpoint range 3 33 - 39°C
Autonomous safety shutdown 40°C
[SVC TOEL yd) ed] Se on REA PEL
Rated voltage (factory setting) 110 - 115 VAC, 50/60 Hz 230 V AC, 50/60 Hz
Power consumption 700 W. 530 W
Current consumption 23A 2.3A
Pump capacity 10 W (up to 0.34 bar), 16 min 10 W (up to 0.34 bar), 16 min
Heating capacity 500 W 500 W
approx. 5 - 10 min
Heating-up time approx. 5- 10 min
(from 19°C to 37°C)
Fuse rating 2xT5L250V 2*T3.15L 250V
PUT sede leeds
Ambient temperature 10 - 30°C
Storage temperature 10 - 40°C
30-70%
Relative humidity