Professional Documents
Culture Documents
Five GC Methods For Biodiesel QC • Measure C17:0 (ISTD) retention time over ten runs
7.
8.
Transfer vial to heater and react for 30 minutes at room temperature.
Transfer vial to front tower.
7.
8.
Transfer vial to front tower.
Add 800 uL n-heptane to vial using front tower.
– C17:0 retention time; 4.082 min., +/- 0.001 min. 9. Add 800 uL n-heptane to vial using front tower. 9. Transfer vial to mixer and mix for 1 minute.
10. Transfer vial to mixer and mix for 1 minute. 10. Transfer vial to rear tower (10 uL syringe).
Method Scope GC Column Column Temperature Inlet/Detector C16:0 11. Transfer vial to rear tower (10 uL syringe). 11. Inject 1 uL on-column using rear tower.
C17:0 C18:1
ASTM D6584 Analysis of Free and Total High Temp 5% DB-5 Temperature programmed 50 oC to Cool-on-column/FID 12. Inject 1 µL on-column using rear tower. 12. Repeat for other B100 samples
C18:0 C18:2 Palm B100
Glycerin 15 m x 0.32 mm x 0.1 um 380 oC
EN14105 Analysis of Free and Total High Temp 5% DB-5 Temperature programmed 50 oC to Cool-on-column/FID
Palm B100 ASTM D6584 – GC Analysis of Free and Total Glycerin After Silylation
Glycerin 10 m x 0.32 mm x 0.1 um 380 oC Palm B100 Comparison of Manual and Automated Calibration Standard Preparation
EN14103 Ester and Linoleic Acid Methyl HP-Innowax Isothermal 200 oC or 210 oC Split-splitless/FID
30 m x 0.25 mm x 0.25 um Palm B100
Ester Content Tricaprin
pA
Rape B100 800
0.6 Glycerol Monoolein
EN14110 Residual Methanol Content HP-Innowax Isothermal 50 oC Split-splitless/FID 600
Std 1 0.5
2.0
response ratio
400
1,2,4-Butanetriol
response ratio
by Headspace 30 m x 0.25 mm x 0.25 um 200
0.4 1.5
Rape B100 0
pA
5 7.5 10 12.5 15 17.5 20 22.5 25 min
0.3
800 1.0
0.2
EN14106 Determination of Free HP-Innowax Isothermal 200 oC Split-splitless/FID Rape B100
600
400
Std 2 Diolein 0.1
y = 1.3366x - 0.0036
R 2 = 1.000
0.5 y = 1.7325x - 0.1156
R 2 = 0.9936
Glycerol 30 m x 0.25 mm x 0.25 um 200
0 0 0
5 7.5 10 12.5 15 17.5 20 22.5 25 min
pA
Rape B100 800
600 Monoolein 0 0.1 0.2 0.3 0.4 0.5 0.6 0 0.5 1 1.5
amount ratio amount ratio
400
Std 3
There are five industry standard methods designed to measure some aspect of biodiesel quality. Most labs run 2 or 3 of these methods. Soy B100
200
0
25 min
pA
800
5 7.5 10 12.5 15 17.5 20 22.5
0.8
Diolein 0.4 Triolein
This table shows the methods grouped according to the column and upper temperature used by the method. As you can see, two
response ratio
600
response ratio
Soy B100 400
Glycerol Std 4 0.6 0.3
methods use a high temperature column ramped to 380 oC. Three other methods use a an HP-Innowax column run isothermally at 60 o 200
0
5 7.5 10 12.5 15 17.5 20 22.5 25 min
0.4 0.2
3 4.082
4 min. 5 6 7
pA
800
y = 1.3859x - 0.0174 y = 0.6531x - 0.0337
C or 210 oC. +/- 0.001 min.
600 0.2 2
R = 1.000
0.1
R2 = 0.9961
400
200
Std 5 Triolein 0 0
Bioidiesel labs have two choices if they wish to run the high temperature column and the wax column methods. They could buy two 0
5 7.5 10 12.5 15 17.5 20 22.5 25 min 0 0.2 0.4 0.6 0.8 0 0.2 0.4 0.6 0.8
amount ratio amount ratio
instruments to accommodate the temperature limitations of the columns. While this offers the greatest flexibility, many labs, especially in
emerging economies cannot afford this option. Additionally, some production labs do not have the space or resources to support two
Low Temperature (50 oC) Isothermal Precision
GCs. The other alternative is to buy one GC and change the columns when changing methods. While economical, this is time
consuming and requires technical skills not always found in many QC labs. Most labs prefer to keep a GC configured for a set of
EN14110 – Determination of Methanol Content (50 oC) ASTM D6584 – GC Analysis of Free and Total Glycerin After Silylation
methods with minimal changes. Measured oven temperature precision: 50 oC, +/- 0.1 oC Comparison of Manual and Automated Sample Preparation
Observed retention time precision:
Soybean B100 Sample
Tricaprin
7890A External Isothermal Capillary Column Oven methanol A. Manual Prep
(ISTD 2)
ISTD 1
Monoolein,
Monolinolein,
1.606 +/- 0.001 min Monopalmitin Monolinolenin
Manual Monostearin
A solution for biodiesel analysis is to have an two column ovens on a single GC. The high temperature column methods require precise Prep
Butanetriol
oven programming performance, therefore that column should reside in the high performance main column oven. Since the methods
n-propanol (ISTD) (ISTD 1) Glycerol 7693A
using the HP-Innowax column are isothermal, the performance criteria for a second oven are less stringent. The 7890A GC has a Auto Prep
2.646 +/- 0.002 min
separate heated region on the top of the GC between the inlets and detectors. Normally, this heated zone is used for valves. 50 oC Isothermal
5 5.5 6 6.5 16 16.5 17 17.5
Fortunately it is large enough to accommodate a properly designed external column oven provided the column can be wound on a
smaller cage. ISTD 2
B. 7693A
Auto Prep Manual Diglycerides Triglycerides
HP-Innowax Column, 30m x 0.32mm ID x 0.25um 0.5 1 1.5 2 2.5 3 3.5 Min. Prep
Column Oven Heater Cartridge and on 4 inch Column Basket
Insulation Thermocouple 7693A
Box Auto Prep
5 10 15 20 25 20 21 22 23 24 25
System Inertness for Isothermal Column Comparison of a soybean B100 sample prepared (A) manually according to Details of the glycerol, mono-, di- and tri-glycerides contained in a soybean B100
EN14106 – Determination of Free Glycerol Content (210 oC) the ASTM D6584 protocol and (B) the 10% scaled automatic protocol using sample. The automated sample preparation yields the same chromatographic result
the 7693A ALS. as the manual sample preparation.
Butanetriol (ISTD)
Isothermal Relative Response Summary
Column Oven Factor Standard Deactivated stainless steel tubing and
Thermal Isolation Glycerol Capillary Flow Technology unions are
From Main Oven Purged CFT Union for
used to connect the external column to • New external capillary column oven for isothermal chromatography
Column Backflush
the system’s split/splitless inlet and • New Liquid Sampler Automates Complex Standard and Sample Preparation
This photo shows the interior of the isothermal column FID. This provides an inert flow path
This photo shows the isothermal oven mounted on the 7890A.
oven. A 4-inch cage contains the 30 m x 0.25 mm ID x • A single GC for complete analysis of biodiesel using 5 different methods:
for the analysis of polar compounds.
The oven insulation box is removed and shown on the left. The Palm B100
0.25 um capillary column. A Capillary Flow Technology Liquid-Liquid This is demonstrated by the analysis of • ASTM D6584: Free and Total Glycerin - Automated derivatization of samples and standards
external oven is thermally isolated from the main column oven Glycerol
(CFT) union connects the column to the split/splitless 0.03 wt % Extract free glycerol extracted from biodiesel.
using specially designed insulation. A heater cartridge and • EN14105: Free and Total Glycerin - Automated derivatization of samples and standards
inlet via deactivated stainless steel tubing. The column
thermocouple automatically controls the external oven
outlet is connected to the FID using a purged CFT inlet. • EN14103: FAME Content - Automated addition of internal standard to samples
temperature.
The purged inlet enables the use of backflushing to 1 2 3 4 5 6 Min. • EN14110: Residual Methanol Content
improve column performance and extend column life.
• EN14106: Free Glycerol Content