Professional Documents
Culture Documents
Troubleshooting Guide
Elecsys® 2010
cobas e 411
Version 5.3
Troubleshooting Elecsys® Systems Page 2 of 142
Table of Contents
1. TROUBLE SHOOTING GUIDE ELECSYS® SYSTEMS: QUICK INFO 4
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 3 of 142
All lists for Elecsys® 2010 are also valid for cobas e 411, when there is no further comment that it
is not valid.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 4 of 142
“reagent handling”
instrument
When did the problem start e.g. after reagent lot change?
Within Grips under Document Type “Assay Summary” there is the HETIA complaint
check list available.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 5 of 142
reagent kit
calibrator
control
is probably the cause of the trouble.
We recommend sending the used "reagents" together with as much
information as possible to GCS Penzberg. A CAPA (corrective and preventive
action) is initiated in case of a confirmed complaint.
Please note that within the Troubleshooting Guide Elecsys® 2010 there is much more
detailed information concerning troubleshooting as well as support lists, technical service
instructions etc. We have summarized experiences that from both internal and external
sources - from different countries - to share them with all those concerned.
We would very much appreciate it if we could add your experiences to the Troubleshooting
Guides for Elecsys® Systems.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 6 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 7 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 8 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 9 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 10 of 142
Others
• Hardware failure
* Microbead mixing speed out of specification.
* Microbead mixer shape bent.
* S/R Probe adjustment not correct.
* LLD adjustment not correct.
* Magnet adjustment not correct.
* Sipper adjustment not correct.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 11 of 142
Additional information Sometimes calibration can be released but problems occur with:
• Recovery of controls.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 12 of 142
Calibrator handling
• Recommended calibration frequency not followed (see point
6 assay calibration).
• Calibration not carefully carried out.
Others
• Hardware failure.
* Measuring Cell (MC), Photomultiplier (PMT), incubator or
system reagent temperature not within specifications.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 13 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 14 of 142
Inter assay precision out of expected • Foam on assay or system reagents. • Have you handled the reagents
range • Reagent kit and/or sample not at proper temperature and calibrators according to the
(drift!). pack. insert?
• Reagent kit stressed (storage or transport conditions not as • Careful refilling of the water
recommended e.g. temperature, upright position), when container (not to cold degassed
bottle change over. water).
excluding explained individual false • Calibration not carefully carried out. • Have you performed
measurements • Recommended calibration frequency not followed (see recommended maintenance?
(see top 2.1 this chapter) point 6 assay calibration). • Call service (check adjustments,
• Air bubbles in the distilled water container when refilling the check bead mixer, carry out APC
container with distilled or deionized water. tests, carry out service
• Hardware failure. maintenance?).
* Microbead mixer speed not within specifications.
* Microbead mixer shape bent.
* Measuring Cell (MC) defective (lifetime 1 year or 50000
cycles warranty).
* S/R Probe adjustment not ok.
* Tubings or seals not tight.
* Clogged distilled water filter (MD6).
* Clogged heat pipe.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 15 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 16 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 17 of 142
3. SUPPORT LISTS
3.1 Expected counts for calibration of all Elecsys® 2010 assays
Assay calibration (mean observed signals)
Please find below a table of the mean assay calibration signals observed on Elecsys® 2010 and cobas e 411.
This signal count lists should only be a hint in which magnitude the assay signals are expected. Please note that it is not necessary to match the listed
counts. These signals should be used as approximate values in order to simplify trouble shooting in case of complaints. We recommend that customer
compare current versus previous measured calibration signals available in instrument / calibration trace screen or calibration print outs.
This data are excerpts of the standardization and Quality Control (QC) release documents of the respective Elecsys® reagents measured in the Roche
standardization / QC department.
This list is updated approx. once a year (in the end of the year) and inbetween for new / improved tests when they have been introduced.
If there are any questions or if lot specific calibration signals are requested, please do not hesitate to contact us.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 18 of 142
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 19 of 142
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 20 of 142
- no information
The counts can also differ between Reagent lots and CalSet lots due to calibrator target values might change.
The above mentioned count differences should not have any effect on the concentrations of analyte in the sample!!!!!!
This table summarizes the mean of data, collected by the QC department.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 21 of 142
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 22 of 142
3.2 Assay Summary Elecsys® 2010: LDL, measuring range, stability, CV, etc.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 23 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 24 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 25 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 26 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 27 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 28 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 29 of 142
Insulin 0.20-1000 2.6-24.9 0.20 µU/ml Expiry date 12 weeks 4 weeks ∅ WHO-reference standard
12017547 µU/ml µU/ml 1.39 pmol/l
(18)[s] 1.39-6945 17.8-173
pmol/l pmol/l *
PlGF 3-10000 pg/ml * ≤ 3 pg/ml Expiry date 12 weeks 6 weeks ∅ Commercially available
05144671 PlGF assay
(18)[s]
sFlt-1 10-85000 * ≤ 10 pg/ml Expiry date 12 weeks 6 weeks ∅* Commercially available
05109523 pg/ml sFlt-1 assay
(18)[s]
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 30 of 142
Assay Measuring Reference LDL Reagent Reagen Reagent Dilution Reference material
Cat.no. range range (functional stability t stability (conc. of or Reference
(total duration in sensitivity, diluted
min.)
unopened stability on method /Standar-
[competitive/ if available) 2-8°C opened board sample disation
sandwich assay]
2-8°C must be)
AFP 0.500-1000 IU/ml ≤ 5.8 IU/ml 0.50 IU/ml Expiry date 12 weeks 8 weeks 1:50 WHO-reference
04481798 0.605-1210 ng/ml ≤ 7.0 ng/ml 0.61 ng/ml (> 20 IU/ml standard
(18)[s] > 24 ng/ml)
*
CA 125 II 0.600-5000 U/ml < 35 U/ml 0.60 U/ml Expiry date 12 weeks 6 weeks 1:5 Enzymun-Test® CA 125
11776223 * (> 1000 U/ml) II (this in turn was
(18)[s] calibrated against CA
125 II RIA from
Fujirebio Diagnostics)
CA 15-3 II # 1.00-300 U/ml ≤ 25 U/ml < 1.00 U/ml Expiry date 12 weeks 5 weeks 1:10 Elecsys® CA 15-3
11776169 * (> 30 U/ml)
(18)[s]
CA 19-9 0.600-1000 U/ml ≤ 27 U/ml < 0.60 U/ml Expiry date 8 weeks 8 weeks 1:10 Enzymun-Test® CA 19-
11776193 * (> 50 U/ml) 9
(18)[s]
CA 72-4 0.200-300 U/ml < 6.9 U/ml < 0.20 U/ml Expiry date 12 weeks 8 weeks 1:2 Enzymun-Test® CA 72-
11776258 * (1.0 U/ml) (> 150 U/ml) 4
(18)[s]
CEA 0.200-1000 ng/ml * 0.20 ng/ml Expiry date 12 weeks 6 weeks 1:50 WHO-reference
11731629 (> 20 ng/ml) standard
(18)[s]
* Please check in package insert. # Elecsys® Universal Diluent needed for pre-dilution.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 31 of 142
Assay Measuring Reference LDL Reagent Reagen Reagent Dilution Reference material
Cat.no. range range (functional stability t stability (conc. of or Reference
(total duration in sensitivity, diluted
min.)
unopened stability on method /Standar-
[competitive/ if available) 2-8°C opened board sample must disation
sandwich assay]
2-8°C be)
Cyfra 21-1 0.100-500 ng/ml < 3.3 ng/ml ≤ 0.10 ng/ml Expiry date 12 weeks 8 weeks 1:2 Enzymun-Test® Cyfra
11820966 * (> 250 ng/ml) 21-1
(18)[s]
HE4 15.0-1500 pmol/l * 15.0 pmol/l expiry date 12 weeks 28 days Diluent MA HE4 EIA method from
05950929 (20.0 pmol/l) 1:20 Fujirebio Diagnostics, Inc.
(18)[s] (> 75 pmol/l)
NSE 0.05-370 ng/ml < 16.3 ng/ml <0.05 ng/ml Expiry date 12 weeks 8 weeks Elecsys® Enzymun-Test® NSE
12133113 * (0.25 ng/ml) Diluent NSE
(18)[s] 1:2
(> 50 ng/ml)
free PSA 0.010-50.00 ng/ml * ≤ 0.01 ng/ml Expiry date 12 weeks 6 weeks ∅ WHO standard
03289788 (0.02 ng/ml) (100% free PSA)
(18)[s]
total PSA 0.002-100 ng/ml age dependend 0.002 ng/ml Expiry date 12 weeks 8 weeks 1:50 Stanford reference
04641655 * (0.03 ng/ml) (> 2 ng/ml) standard
(18)[s] (90% PSA-ACT + 10%
free PSA)
S100 0.005-39 μg/l * < 0.005 μg/l Expiry date 12 weeks 8 weeks 1:5 with S 100 Weighed-out S100 β/β
03175243 (< 0.02 μg/l) negative protein
(18)[s] human sera or
CalSet 1
(> 1 μg/l)
Please check in package insert.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 32 of 142
Assay Measuring Reference LDL Reagent Reagent Reagent Dilution Reference material
Cat.no. range range (functional stability stability stability (conc. of or Reference
(total duration in sensitivity, unopened opened on board diluted method /Standar-
min.)
[competitive/ if available) 2-8°C 2-8°C sample disation
sandwich assay] must be)
Free ßhCG 0.1-190 IU/l < 0.13 IU/l < 0.1 IU/l Expiry date 4 weeks 4 weeks 1:10 NIBSC
04854071 * (0.5 IU/l) (> 19 IU/l)
(18)[s]
PAPP-A 4-10000 mIU/ll < 7.15 mIU/l < 4 mIU/l Expiry date 4 weeks 3 weeks 1:10 Commercially
04854098 * (20 mIU/l) (> 500 mIU/l) available PAPP-A
(18)[s] Test *
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 33 of 142
Assay Measuring Reference LDL Reagent Reagent Reagent Dilution Reference material
Cat.no. range range (functional stability stability stability (conc. of or Reference
(total duration in sensitivity, diluted
min.)
unopened opened on board method /Standar-
[competitive/ if available) 2-8°C 2-8°C sample disation
sandwich assay] must be)
Vitamin B12 30-2000 pg/ml 243-894 30 pg/ml Expiry date 12 weeks 5 weeks 1:2 Commercially
04745736 22-1476 pmol/l pg/ml 22 pmol/l (> 1000 available radiobinding
(27)[c] 179-660 pg/ml B12 assay
pmol/l > 738
* pmol/l)
Ferritin 0.500-2000 ng/ml * 0.5 ng/ml Expiry date 12 weeks 6 weeks 1:50 Elecsys® Ferritin
03737551 (> 40 ng/ml) cat.no. 11820982*
04491785
(18)[s]
Folate III 0.64-20 ng/ml * ≤ 0.64 Expiry date 8 weeks 2 weeks 1:2 Elecsys® Folate II
03253678 1.45-45.4 nmol/l ng/ml manually cat.no. 03253678
(27)[c] ≤ 1.45 (> 10 ng/ml
nmol/l > 22.7
nmol/l)
RBC Folate * * * Expiry date 14 days - * *
Hemolyzing
Reagent
12017741
* Please check in package insert.
Pretreatment assay
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 34 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 35 of 142
Assay Measuring range Reference LDL Reagent Reagent Reagent Dilution Reference material
Cat.no. range (functional stability stability stability (conc. of or Reference
(total duration in sensitivity, unopened opened on board diluted method /Standar-
min.)
[competitive/ if available) 2-8°C 2-8°C sample must disation
sandwich assay] be)
IgE II 0.100-2500 IU/ml * 0.10 IU/ml Expiry date 12 8 weeks 1:20 WHO-reference
04827031 0.240-6000 ng/ml 0.24 ng/ml weeks (> 125 IU/ml standard
(18)[s] (0.50 IU/ml > 300 ng/ml)
1.20 ng/ml)
Anti-CCP 7-500 U/ml * 7 U/ml Expiry date - 1 week diluent: Commercially
05031656 (8 U/ml) * negative available second-
(18)[s] human serum generation anti-CCP
pool assay
1:2 to 1:5
BRAHMS PCT 0.02-100 ng/ml 0.046 ≤ 0.02 Expiry date 12 4 weeks diluent: BRAHMS PCT LIA
05056888 ng/ml* ng/ml weeks negative
(18)[s] (≤ 0.06 human serum
ng/ml) 1:4
(> 1.0
ng/ml)
IL-6 1.5-5000 pg/ml 7 pg/ml* Approx. 1.5 Expiry date 12 weeks 4 weeks 1:10 Dil MA NIBSC 1st 89/548
05109442 pg/ml (> 50 pg/ml) standard
(18)[s]
* Please check in package insert.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 36 of 142
Anti-HBc - Cut off index > < 0.8 PEI 100% 99.6% Expiry date 8 weeks 8 weeks possible ”HBc-
11820559 1 are non- U/ml resp. * Reference
(27)[c] reactive * 99.7% material 82
* * (IgG anti-
HBc)” of the
Paul Ehrlich
Institute
Anti-HBc - Cut off approx. < 3.0 100% 100% Expiry date 8 weeks 4 weeks possible "HBc
IgM # 100 PEI U/ml, PEI-E/ml * * Reference
11820567 equival cut off serum IgM 84
(18) [µ-Capture] index is 0.9- (IgM anti-
HBc)" of the
1.1 *
Paul Ehrlich
Institute
* Please check in package insert. - no information # Elecsys® Universal Diluent needed for pre-dilution.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 37 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 38 of 142
Assay Measuring Reference LDL/LoD Clinical Clinical Reagent Reagent Reagent Dilution Reference
Cat.no. Range range sensitivit specificity stability stability stability (conc. of material
(total duration in diluted
min.)
y unopene opened on board or
[competitive/ d 2-8°C sample Reference
sandwich assay]
2-8°C must be) method
/Standar-
disation
HBsAg II quant 0.05-130 0.05 IU/ml - - Expiry 8 weeks 8 weeks 1:100 with NISBC
05957435 IU/ml * date Diluent standard
(18)[s] (undiluted Universal
samples)
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 39 of 142
Assay Measuring Reference LDL Clinical Clinical Reagent Reagent Reagent Dilution Reference
Cat.no. Range range sensitivity specificit stability stability stability (conc. of material or
(total duration diluted
in min.)
y unopened opened on board Reference
[competitive/ 2-8°C 2-8°C sample method
sandwich must be) /Standar-
assay]
disation
HIV Ag - Cut off index ≤ 4.0 * * Expiry date 4 weeks 4 weeks - NIBSC
11971611 < 0.9 are not pg/ml code 90/636
(18)[s] reactive
Cut off index
0.9-<1.0 are
deemded
borderline
HIV Ag - x < 50% = * * * Expiry date 4 weeks - - *
Confir- neg. or false-
matory Test reactive
12001101 * *
HIV combi - Cut off index < 6 U/ml 100% (95% 99.76% expiry date 4 weeks* 2 weeks; 70 - Antigen: 1st
04860446 < 0.9 are not (antigen- conf. limit = (blood h in case of Int. Ref.
(18)[s] reactive sensitivity) 99.8%) donors); alternate Reagent
Cut off index 99.42% storage at 1992, 90/636
0.9-<1.0 are (others) 2-8°C and (NIBSC)
deemed 20-25°C
borderline during 4
weeks
* Please check in package insert.
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 40 of 142
Assay Measuring Reference LDL / Clinical Clinical Reagent Reagent Reagent Dilution Reference
Cat.no. Range range Detection sensitivity specificit stability stability stability (conc. of material or
(total duration in limit unopened opened on board diluted Reference
min.) y
[competitive/ 2-8°C 2-8°C sample method
sandwich must be) /Standar-
assay] disation
HIV combi - Cut off index < < 2 IU/ml 100% (95% 99.88% expiry date 12 weeks 4 weeks - NIBSC, code
PT 0.9 are non- conf. limit = (blood 90/636
05390095 reactive 99.76%) donors)*
(27)[s] Cut off index
0.9-<1.0 are
borderline
Cut off index ≥
1 are reactive
CMV IgG 0.25-500 * 0.25 U/ml - - expiry date 12 weeks 3 weeks 1:20 Internal Roche
04618815 U/ml (> 15 standard for
(18)(s) U/ml) CMV IgG
CMV IgG 0.25-500 * 0.25 U/ml - - expiry date 12 weeks 3 weeks* 1:20 Internal Roche
Avidity U/ml manually standard for
05909708 with CMV IgG
(18)(s) Diluent.
Universal
(> 15
U/ml)
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 41 of 142
Assay Measuring Reference LDL Clinical Clinical Reagent Reagent Reagent Dilution Reference
Cat.no. Range range sensitivity specificit stability stability stability (conc. of material or
(total duration
in min.)
y unopened opened on board diluted Reference
[competitive/ 2-8°C 2-8°C sample method
sandwich must be) /Standar-
assay]
disation
CMV IgM - Cut off index < - - - expiry date 12 weeks 2 weeks; - Roche
04618858 0.8 are not 84 h in standard
(18)[µ- reactive case of
capture) Cut off index ≥ alternate
0.8 - < 1.0 COI storage at
are 2-8°C and
indeterminate 20-25°C
Cut off index ≥ during 12
1 are reactive weeks
HSV-1 IgG - Non-reactive < - - - expiry date 8 weeks 28 days - Roche
05572185 0.6 COI standard
(18)[s] Gray-zone: ≥
0.6-<1.0 COI
Reactive ≥ 1.0
COI
HSV-2 IgG - Non-reactive < - - - expiry date 8 weeks 28 days - Roche
05572193 0.51 COI standard
(18)[s Gray-zone: ≥
0.51-<1.0 COI
Reactive ≥ 1.0
COI
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 42 of 142
Assay Measuring Reference LDL Clinical Clinical Reagent Reagent Reagent Dilution Reference
Cat.no. Range range sensitivity specificit stability stability stability (conc. of material or
(total duration
in min.)
y unopened opened on board diluted Reference
[competitive/ 2-8°C 2-8°C sample method
sandwich must be) /Standar-
assay]
disation
Rubella IgG 0.17-500 * 0.17 IU/ml - - expiry date 12 weeks* 2 weeks, 1:20 1st
04618793 IU/ml 84 h in (> 10 International
(18)[s] case of IU/ml) Standard for
alternate Anti-Rubella
storage at Immun-
2-8°C and globulin
20-25°C
during 12
weeks
Rubella - Cut off index - * * expiry date 12 weeks* 2 weeks; - Roche
IgM < 0.8 are not 84 h in standard
04618831 reactive case of
(18)[µ-capture) Cut off index alternate
≥ 0.8 - < 1.0 storage at
COI are 2-8°C and
indeterminate 20-25°C
Cut off index during 12
≥ 1 are weeks
reactive
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 43 of 142
Assay Measuring Reference LDL Clinical Clinical Reagent Reagent Reagent Dilution Reference
Cat.no. Range range sensitivity specificit stability stability stability (conc. of material or
(total duration
in min.)
y unopened opened on board diluted Reference
[competitive/ 2-8°C 2-8°C sample method
sandwich must be) /Standar-
assay]
disation
Toxo IgG 0.13-650 * 0.125 - - expiry date 12 weeks* 2 weeks, 1:20 3rd
04618815 IU/ml IU/ml 84 h in (> 3 International
(18)(s) case of IU/ml) Standard for
alternate anti-
storage at Toxoplasma
2-8°C and serum
20-25°C (TOXM) from
during 12 NIBSC •
weeks
Toxo IgM - Cut off index - - - expiry date 12 weeks* 2 weeks; - Roche
04618858 < 0.8 are not 84 h in standard
(18)[µ- reactive case of
capture) Cut off index alternate
≥ 0.8 - < 1.0 storage at
COI are 2-8°C and
indeterminate 20-25°C
Cut off index during 12
≥ 1 are weeks
reactive
* Please check in package insert.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 44 of 142
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 45 of 142
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 46 of 142
Assay Specified intraassay internal evaluation Specified interassay precision internal evaluation
precision intraassay precision interassay precision
CK-MB 0.3-5 ng/ml 0.15 ng/ml SD 0.02, 0.06 ng/ml 0.3-5 ng/ml 0.25 ng/ml SD 0.03, 0.1 ng/ml
>5-15 ng/ml ≤ 3% CV 1.3% CV >5-15 ng/ml ≤ 5% CV 2% CV
>15 ng/ml ≤ 5% CV 1.4, 1.2% CV >15 ng/ml ≤ 10% CV 2.6, 2.0% CV
Myoglobin 15-50 ng/ml < 8% CV 2.1% CV 15-50 ng/ml < 10% CV 2.6% CV
STAT 50-500 ng/ml < 5% CV 2.9, 1.3% CV 50-500 ng/ml < 6% CV 3.6, 1.6% CV
> 500 ng/ml < 8% CV 2.9, 3.4, 5.3, 1.9% CV > 500 ng/ml < 10% CV 3.8, 4.0, 6.7, 2.3% CV
proBNP II 50-100 pg/ml < 10% CV 2.1% CV 50-100 pg/ml < 13% CV 2.8% CV
>100-250 pg/ml < 5% CV 2.4% CV >100-250 pg/ml < 8% CV 2.6% CV
> 250 pg/ml < 5% CV 1.2, 1.4, 2.7% CV > 250 pg/ml < 5% CV 1.6, 2.1, 3.8% CV
< 50 pg/ml (not specified) 4.1% CV < 50 pg/ml (not specified) 4.6% CV
Troponin T 3 - < 25 ng/l < 1 ng/l SD 0.6 ng/l SD 3 - < 25 ng/l < 2 ng/l SD 1 ng/l SD
STAT hs 25-< 100 ng/l ≤ 5% CV 2.1, 1.1, 1.1% CV 25-< 100 ng/l ≤ 10% CV 3.6%, 4.4, 2.1% CV
> 100 ng/l ≤ 3% CV 1.5, 0.9, 0.8% CV > 100 ng/l ≤ 6% CV 2.2, 1.4, 1.3% CV
Troponin I 0.1-0.2 ng/ml ≤ 25% CV 0.1-0.2 ng/ml ≤ 40% CV
STAT 0.2-0.4 ng/ml ≤ 10% CV 9.1, 4.8% CV 0.2-0.4 ng/ml ≤ 20% CV 16, 8.9% CV
> 0.4 ng/ml ≤ 8% CV 3.3, 2.2, 4.2, 2.9% CV > 0.4 ng/ml ≤ 8% CV 5.9, 5.3, 3.2, 6.7, 3.6% CV
Digoxin < 2.2 ng/ml < 0.075 ng/ml SD 0.048, 0.045 ng/ml SD < 2.2 ng/ml < 0.15 ng/ml SD 0.07, 0.066 ng/ml SD
2.2-4 ng/ml < 3.5% CV 3.3% CV 2.2-4 ng/ml < 5.5% CV 4.1% CV
> 4 ng/ml < 5% CV >4 ng/ml < 7% CV
Digitoxin < 10 ng/ml < 8% CV 5.7% CV < 10 ng/ml < 10% CV 6.5% CV
10-40 ng/ml < 4% CV 2.6, 2.3% CV 10-40 ng/ml < 6% CV 3, 2.4% CV
> 40 ng/ml < 6% CV 1.8, 1.7% CV > 40 ng/ml < 8% CV 3, 2.6% CV
HCG 10-500 mIU/ml < 5% CV 2.2, 1.5, 1.3% CV 10-500 mIU/ml < 7% CV 3.1, 1.7, 1.3% CV
STAT 500-1000 mIU/ml < 3% CV 1.2% CV 500-1000 mIU/ml < 5% CV 1.4% CV
>1000 mIU/ml < 3% CV 1.2% CV >1000 mIU/ml < 5% CV 1.4% CV
HCG+β 2-5 mIU/ml ≤ 9% CV 4.9% CV 2-5 mIU/ml ≤ 10% CV 5.9% CV
5-10 mIU/ml ≤ 5% CV 1.9, 1.7% CV 5-10 mIU/ml ≤ 5% CV 2.9, 3.6, 2.9% CV
> 10 mIU/ml ≤ 4 % CV 3.3, 2.3, 1.3% CV > 10 mIU/ml ≤ 5 % CV 2.1, 2.7, 1.9% CV
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 47 of 142
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 48 of 142
SHBG 5-200 nmol/l < 4% CV 2.2, 2.1, 2.4, 2.8% CV 5-200 nmol/l < 6% CV 2.6, 2.7, 2.8, 5.6% CV
Testosterone 0.15 – 0.5 ng/ml ≤ 10% CV 2.8% CV 0.15 – 0.5 ng/ml ≤ 20% CV 3.8% CV
II 0.5-1 ng/ml ≤ 6% CV 2.1% CV 0.5-1 ng/ml ≤ 8% CV 3.2% CV
1-3 ng/ml ≤ 5% CV 1.9, 1.8% CV 1-3 ng/ml ≤ 6% CV 2.8, 3.7% CV
3-15 ng/ml ≤ 5% CV 2.6, 1.2% CV 3-15 ng/ml ≤ 6% CV 2.8, 2.9% CV
< 0.15 ng/ml (not specified) 4.7% CV < 0.15 ng/ml (not specified) 8.4% CV
C-Peptide 0.5-20 ng/ml ≤ 5% CV 4.2, 2.4, 4.7% CV 0.5-20 ng/ml ≤ 6% CV 3.0, 5.0, 1.8% CV
serum
C-Peptide 0.5-20 ng/ml ≤ 5% CV 1.6, 4.2, 2.9% CV 0.5-20 ng/ml ≤ 6% CV 4.2, 2.4, 1.8% CV
urine
Insulin 5-10 µU/ml < 5% CV 1.9% CV 5-10 µU/ml < 6% CV 2.6% CV
10-300 µU/ml < 4% CV 1.5, 1.7, 1.9, 2.2, 1.9% 10-300 µU/ml < 5% CV 2.5, 2.1, 2.8, 2.6, 2.3% CV
300-1000 µU/ml < 5% CV 2% CV 300-1000 µU/ml < 6% CV 2.5% CV
PlGF ≤ 50 pg/ml ≤ 4 pg/ml SD 0.28 pg/ml SD ≤ 50 pg/ml ≤ 5 pg/ml SD 0.45 pg/ml SD
≥ 50 pg/ml ≤ 8% CV 1, 0.8, 1% CV ≥ 50 pg/ml ≤ 10% CV 1.3, 1.4, 1.3% CV
≥ 200 pg/ml ≤ 6% CV 0.8% CV ≥ 200 pg/ml ≤ 8% CV 1.3% CV
≥ 1000 pg/ml ≤ 6% CV 0.8, 0.9% CV ≥ 1000 pg/ml ≤ 8% CV 1, 2% CV
sFlt-1 ≤ 25 pg/ml ≤ 2 pg/ml SD ≤ 25 pg/ml ≤ 3 pg/ml SD
≥ 25 pg/ml ≤ 8% CV 1.6% CV ≥ 25 pg/ml ≤ 12% CV 4.3% CV
≥ 200 pg/ml ≤ 7% CV 0.8, 1% CV ≥ 200 pg/ml ≤ 8% CV 2.3, 2.3% CV
≥ 10000 pg/ml ≤ 6% CV 1, 1.2, 1% CV ≥ 10000 pg/ml ≤ 7% CV 2.6, 2.9, 3.7% CV
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 49 of 142
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 50 of 142
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 51 of 142
Ferritin 10-40 ng/ml < 7% CV 3.5, 1.8, 2.1% CV 10-40 ng/ml < 7% CV 4.4, 2.9, 3.2% CV
50-400 ng/ml < 4% CV 2.7, 1.8, 1.9% CV 50-400 ng/ml < 7% CV 3, 3.2, 2.2% CV
> 400 ng/ml (not specified) 2% CV > 400 ng/ml (not specified) 2.6% CV
Folate III ≤ 5 ng/ml <0.35 ng/ml SD 0.23, 0.24 ng/ml SD ≤ 5 ng/ml < 1 ng/ml SD 0.39, 0.45 ng/ml SD
> 5 ng/ml < 7% CV 5.7, 4.7 3.0, 4.3% CV > 5 ng/ml < 12% CV 6.4, 6.3, 7, 5% CV
RBC Folate na * na *
* Please check in package insert.
na not available at the moment
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 52 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 53 of 142
Assay Specified intraassay precision internal evaluation Specified interassay precision internal evaluation
intraassay precision interassay precision
IgE II 1-5 IU/ml < 9% CV - 1-5 IU/ml < 10% CV -
6-250 IU/ml < 6% CV 2, 4.1% CV 6-250 IU/ml < 7% CV 3.7, 5.1% CV
> 250 IU/ml < 5% CV 2.3, 2.4, 2.6, 2.4% CV > 250 IU/ml < 8% CV 4, 3.8, 3.9, 4% CV
Anti-CCP 10-25 U/ml ≤ 2.5 SD 0.6, 1% CV 10-25 U/ml ≤ 3.8 SD 3, 3.1% CV
25-200 U/ml ≤ 6% CV 1.4% CV 25-200 U/ml ≤ 8% CV 2.5% CV
200-1000 U/ml ≤ 8% CV 2.3% CV 200-1000 U/ml ≤ 10% CV 4.5% CV
BRAHMS 0.075-1 ng/ml ≤ 15% CV 1.3, 2.1% CV 0.075-1 ng/ml ≤ 20% CV 3.7, 4.2% CV
PCT > 0.1 ng/ml ≤ 10% CV 0.9, 2.1% CV > 0.1 ng/ml ≤ 15% CV 4, 4.9% CV
(< 0.075 ng/ml, not specified) 8.8% CV (< 0.075 ng/ml, not specified) 16.3% CV
IL-6 5-25 pg/ml rec. +/-5 pg/ml 5-25 pg/ml rec. +/-5 pg/ml
25-300 pg/ml ≤ 5% CV 2.3, 1.7, 2.5% CV 25-300 pg/ml ≤ 7% CV 5.4, 3.1, 3.2% CV
> 300 pg/ml ≤ 5% CV 2.6% CV > 300 pg/ml ≤ 7% CV 2.9% CV
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 54 of 142
Assay Specified intraassay precision internal evaluation Specified interassay precision internal evaluation
intraassay precision interassay precision
Anti-HAV HS, neg < 3 IU/ml SD - HS, neg < 3 IU/ml SD -
HS, low pos < 7% CV 1.2% CV HS, low pos < 10% CV 4% CV
HS, positive < 5% CV 1.2% CV HS, positive < 10% CV 2.9% CV
Anti-HAV HS, neg ≤ 20% CV 2% CV HS, neg ≤ 20% CV 3.8% CV
IgM HS, low pos ≤ 7% CV 3.4% CV HS, low pos ≤ 10% CV 2.8% CV
HS, positive ≤ 5% CV 3.1% CV HS, positive ≤ 10% CV 5.4% CV
Anti-HBc HS, neg ≤ 7% CV 1.2% CV HS, neg ≤ 10% CV 1.4 % CV
HS, low pos ≤ 7% CV 3.9% CV HS, low pos ≤ 10% CV 1.1 % CV
HS, positive ≤ 7% CV 3.4% CV HS, positive ≤ 10% CV 1.3 % CV
PC a-HBc 1 ≤ 7% CV 1.9 % CV PC a-HBc 1 ≤ 10% CV 1.0 % CV
PC a-HBc 2 ≤ 7% CV 2.3% CV PC a-HBc 2 ≤ 10% CV 1.6 % CV
Anti-HBc HS, neg ≤ 5% CV 2.2% CV HS, neg ≤ 13% CV 2.8% CV
IgM HS, low pos ≤ 5% CV 3.5% CV HS, low pos ≤ 8% CV 1.8% CV
HS, positive ≤ 7% CV 3.7% CV HS, positive ≤ 10% CV 3.4% CV
PC a-HBc IgM 1 ≤ 5% CV 1.7% CV PC a-HBc IgM 1 ≤ 10% CV 3.3% CV
PC a-HBc IgM 2 ≤ 7% CV 4.5% CV PC a-HBc IgM 2 ≤ 13% CV 3.6% CV
Anti-HBe HS, neg ≤ 5% CV 2.4% CV HS, neg ≤ 10% CV 2.6% CV
HS, low pos ≤ 3% CV 2.0% CV HS, low pos ≤ 5% CV 2.2% CV
HS, positive ≤ 3% CV 3.0% CV HS, positive ≤ 5% CV 3.5% CV
PC a-HBe 1 ≤ 5% CV 2.1% CV PC a-HBe 1 ≤ 10% CV 2.0% CV
PC a-HBe 2 ≤ 3% CV 2.1 % CV PC a-HBe 2 ≤ 5% CV 2.5% CV
HS = human sera
- no information
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 55 of 142
Assay Specified intraassay precision internal evaluation Specified interassay precision internal evaluation
intraassay precision interassay precision
HBeAg HS, neg ≤ 10% CV 4.0% CV HS, neg ≤ 10% CV 4.0, 1.7% CV
HS, low pos ≤ 5% CV HS, low pos -
HS, positive ≤ 5% CV 4.0, 2.8% CV HS, positive ≤ 5% CV 4.9% CV
PC HBeAg 1 ≤ 10% CV 6.6% CV PC HBeAg 1 ≤ 10% CV 4.5% CV
PC HBeAg 2 ≤ 5% CV 1.6% CV PC HBeAg 2 ≤ 5% CV 1.7% CV
Anti-HBs ≤ 5 IU/l ≤ 20% CV 3 IU/l 13.1% CV ≤ 5 IU/l ≤ 25% CV 3 IU/l 9.9% CV
5-10 IU/l ≤ 15% CV 5-10 IU/l ≤ 20% CV
10 IU/l ≤ 8% CV 10.7 IU/l 6.3% CV 10 IU/l ≤ 10% CV 10.7 IU/l 16.2% CV
10-1000 IU/l ≤ 5% CV 15.4, 498, 3.4-1.5% CV 10-1000 IU/l ≤ 8% CV 15.4, 498, 11-3.4% CV
603 IU/l 603 IU/l
Anti-HCV HS, neg ≤ 0.5 COI SD 0.01 COI SD HS, neg ≤ 0.5 COI SD 0.01 COI SD
HS, low pos ≤ 7 % CV 3.3, 1.2 % CV HS, low pos ≤ 10 % CV 3, 4.2% CV
HS, positive ≤ 8 % CV 1.1% CV HS, positive ≤ 10 % CV 3.5% CV
HBsAg II HS, neg ≤ 0.07 COI SD 0.03, 0.04 COI SD HS, neg≤ 0.7 COI SD 0.04, 0.04 COI SD
HS, low pos ≤ 7 % CV 2.2, 1.5% CV HS, low pos ≤ 10 % CV 5.7, 3.3% CV
HS, positive ≤ 7 % CV 3.9% CV HS, positive ≤ 10 % CV 1.1% CV
HBsAg II 0.05-5 IU/ml SD ≤ 0.5 IU/ml 0.13 IU/ml SD 0.05-5 IU/ml SD ≤ 0.6 IU/ml 0.17 IU/ml SD
quant. undiluted 0.6% CV undiluted 5.6% CV
5-130 IU/ml ≤ 10% CV undiluted 2.6% CV 5-130 IU/ml ≤ 12% CV undiluted 5.6& CV
5-13000 IU/ml ≤ 15% CV diluted 5-13000 IU/ml ≤ 17% CV diluted
HIV Ag HS, neg - 5.2% CV HS, neg - 7.1% CV
HS, low pos ≤ 5 % CV 2.7% CV HS, low pos - 3.7% CV
HS, positive - 2.6% CV HS, positive - 4.1% CV
PC HIV Ag 1 ≤ 10 % CV 6.6 % CV PC HIV Ag 1 ≤ 20 % CV 5.9% CV
PC HIV Ag 2 ≤ 5 % CV 2.0 % CV PC HIV Ag 2 ≤ 10 % CV 5.0% CV
HS = human sera - no information
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 56 of 142
Assay Specified intraassay precision internal evaluation Specified interassay precision internal evaluation
intraassay precision interassay precision
HIV combi HS neg: ≤ 10 % CV 3.4% CV HS neg: ≤ 15% CV 10.1% CV
HS, pos. # ≤ 6 % CV 1.2, 1, 0.9% CV HS, pos. # ≤ 10 % CV 3.6, 3.9, 3.6% CV
PC HIV com. 1 ≤ 10 % CV na 5.8% CV PC HIV com. 1 ≤ 20% CV 7.9% CV
PC HIV com. 2 ≤ 5% CV 1% CV PC HIV com. 2 ≤ 10% CV 1.5% CV
PC HIV com. 3 ≤ 5% CV 1.5% CV PC HIV com. 3 - 2.8% CV
HIV combi HS neg (COI ≤ 0.6): SD ≤ 0.06 0.026 COI SD HS neg (COI ≤ 0.6): SD ≤ 0.09 0.031 COI SD
PT Anti-HIV-1 pos. ≤ 6 % CV 0.9% CV Anti-HIV-1 pos. ≤ 10 % CV 1.4% CV
Anti-HIV-2 pos. ≤ 6 % CV 0.9% CV Anti-HIV-2 pos. ≤ 10 % CV 1.5% CV
HIV Ag pos. 3 ≤ 6% CV 0.9% CV HIV Ag pos. 3 ≤ 10% CV 1.8% CV
CMV IgG 0.5-2.0 U/ml ≤ 10 % CV 0.9, 1.2, 1.5% CV 0.5-2.0 U/ml ≤ 12 % CV 3.2, 2.6, 3.3% CV
> 2.0 U/ml ≤ 7 % CV 0.7, 1.5, 1.4% CV > 2.0 U/ml ≤ 10 % CV 3.9, 3.3, 3.6% CV
CMV IgG < 10 Avi% ± 2 Avi% SD 0.04 Avi% SD < 10 Avi% ± 3 Avi% SD 0.1 AVI% SD
Avidity 10-100% Avi ≤ 8% CV 1, 2.6, 1.5, 3.1% Avi 10-100% Avi ≤ 8% CV 2.0, 3.4, 2.6% AVI%
CMV IGM < 0.6 COI ≤ 0.06 COI SD 0.003, 0.002 COI SD < 0.6 COI ≤ 0.06 COI SD 0.005, 0.004 COI SD
> 0.6 COI ≤ 7% CV 1.5, 1.8% CV > 0.6 COI ≤ 7% CV 5.3, 3.4, 3.1% CV
HSV-1 IgG ≤ 0.8 COI ≤ 0.1 SD 0.003, 0.001, 0.006 COI SD ≤ 0.8 COI ≤ 0.15 SD 0.006, 0.001, 0.012 COI SD
> 0.8-2.0 COI ≤ 6% CV 1, 1.3% CV > 0.8-2.0 COI ≤ 8% CV 2.9, 2.8% CV
> 2 COI ≤ 8% CV 1, 1.3% CV > 2 COI ≤ 10% CV 2.5, 2.8% CV
HSV-2 IgG ≤ 0.8 COI ≤ 0.1 SD 0.005, 0.004 COI SD ≤ 0.8 COI ≤ 0.15 SD 0.01, 0.007 COI SD
> 0.8-2.0 COI ≤ 6% CV 1.2, 1.2% CV > 0.8-2.0 COI ≤ 8% CV 2.8% CV
> 2 COI ≤ 8% CV 1.6, 1.6% CV > 2 COI ≤ 10% CV 3.3, 3.6% CV
Rubella IgG < 5 IU/ml < 1 IU/ml SD 0.075 IU/ml SD < 5 IU/ml < 1.1 IU/ml SD 0.074 IU/ml SD
> 5 IU/ml < 10% CV 1.9, 2.5, 1.6% CV > 5 IU/ml < 12% CV 3.8, 3.4, 3.6% CV
Rubella IgM HS neg: ≤ 7 % CV 2.4% CV HS neg: ≤ 10 % CV 3% CV
HS, pos. ≤ 7 % CV 1.2, 1% CV HS, pos. ≤ 10 % CV 1.9, 4.1% CV
PC Rub. IgM 1 ≤ 7 % CV 1.4% CV PC Rub. IgM 1 ≤ 10 % CV 4.1% CV
PC Rub. IgM 2 ≤ 7 % CV 1.1% CV PC Rub. IgM 2 ≤ 10 % CV 4.1% CV
HS = human sera - no information
# HS low pos/pos.: for anti-HIV 1, anti-HIV 2, HIV Ag determined (separate available in package insert)
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 57 of 142
Assay Specified intraassay precision internal evaluation Specified interassay precision internal evaluation
intraassay precision interassay precision
Toxo IgG HS neg: ≤ 7 % CV HS neg: ≤ 10 % CV
HS, pos. ≤ 7 % CV 1.5, 1.3 % CV HS, pos. ≤ 10 % CV 4.2, 5.3% CV
PC Toxo IgG 1 ≤ 7 % CV 1.6 % CV PC Toxo IgG 1 ≤ 10 % CV 5.7% CV
PC Toxo IgG 2 ≤ 7 % CV 2.1% CV PC Toxo IgG 2 ≤ 10 % CV 3.0% CV
Toxo IgM HS neg: ≤ 7 % CV HS neg: ≤ 10 % CV
HS, pos. ≤ 7 % CV 1.9, 1.6% CV HS, pos. ≤ 10 % CV 4, 3.6 % CV
PC Toxo IgM 1 ≤ 7 % CV 2.5% CV PC Toxo IgM 1 ≤ 10 % CV 2.7% CV
PC Toxo IgM 2 ≤ 7 % CV 1.6% CV PC Toxo IgM 2 ≤ 10 % CV 3% CV
HS = human sera - no information
# HS low pos/pos.: for anti-HIV 1, anti-HIV 2, HIV Ag determined (separate available in package insert)
The system to system precision is defined as interassay precision x 1.5 e.g. 6% interassay precision = 9% system to system precision, when not specified for the test.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 58 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 59 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 60 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Cat.no. frequence target stability stability stability possible material volume
Calibrator Cal 1/Cal unopened opended on board per bottle/
2 2-8°C 2-8°C 20-25°C lyophilized Pipetting
ready- volume
to-
use
CK-MB ° as required, ≈ 1.5/25 Expiry date 2 weeks 5 hours Y 1x lyo. human 1 ml
05957664 e.g. control ng/ml (-20°C / 3 serum with 15 μl
(9 min)/ not within month) added CK-
05957656 range MB
(18 min)
° every 1
Myoglobin month (same ≈ 35/1200 Expiry date 12 weeks 5 hours Y - buffer/ 1 ml
11820893 lot), ng/ml protein 15 µl
(9 min)/ ° every week matrix with
12178222 (same RP) added
(18 min) myoglobin
proBNP II ≈ 140/2700 Expiry date 2 weeks 5 hours Y 1x equine 1 ml
04842472 pg/ml (-20°C / 3 serum 15 µl
≈ 16.6/320 month) matrix with
pmol/l added
proBNP
Troponin T ≈ 18/4200 Expiry date 2 weeks 5 hours Y 1x lyo. rec. 1 ml
hs ng/l (-20°C / 3 human 50 μl
05092736 month) serum with
(9 min)/ added
05092752 Troponin T
(18 min)
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 61 of 142
Assay Calibratio Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Cat.no. n target stability stability stability possible material volume
Calibrator frequence Cal 1/Cal unopened opended on board per bottle/
2 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
to- use volume
Troponin I ° as ≈ 0.4/30 Expiry date 4 days 5 hours Y 3 months human 1 ml
STAT required, ng/ml, (freeze serum 30 µl
05094801 e.g. control only once) matrix with
not within added TNI
range
Digoxin ° every 1 ≈ 0.61/3.61 Expiry date 12 weeks 5 hours Y - bovine 1.5 ml
11820907 month ng/ml, serum 10 µl
(same lot), ≈ 0.78/4.62 matrix with
° every nmol/l added
week Digoxin
(same RP)
Digitoxin ≈ 5.00/60 Expiry date 12 weeks 5 hours Y - lyo. human 1 ml
03002667 ng/ml, serum with 15 µl
≈ 6.55/79 added
nmol/l Digitoxin
HCG STAT ≈ 10/5000 Expiry date 12 weeks 5 hours Y - lyo. human 1 ml
03303071 mIU/ml serum with 10 μl
added HCG
HCG+β ≈ 1.5/2000 Expiry date 12 weeks 5 hours Y - lyo. human 1 ml
03302652 mIU/ml serum with 10 µl
added HCG
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 62 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 63 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 64 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 65 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Cat.no. frequence target stability stability stability possible material volume per
Calibrator Cal 1/Cal unopened opended on board bottle/
2 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
to- use volume
AFP II ° as required, ≈ 5/50 Expiry date 6 weeks 5 hours Y 1x lyo. human 1 ml
04487761 e.g. control IU/ml, (-20°C / serum with 10 μl
not within ≈ 6/60 6 weeks) added human
range ng/ml AFP
CA 125 II ° every 1 ≈ 35/500 Expiry date 12 weeks 5 hours Y - human serum 1 ml
11776240 month U/ml with added 40 µl
(same lot), human CA 125
° every week
(same RP)
CA 15-3 II ≈ 15/100 Expiry date 12 weeks 5 hours Y - human serum 1 ml
03045846 U/ml with added 20 µl
human CA 15-
3
CA 19-9 ≈ 20/250 Expiry date 4 weeks 5 hours Y 1x lyo. human 1 ml
11776215 U/ml (-20°C / serum with 10 µl
2 month) added human
CA 19-9
CA 72-4 ≈ 1/70 U/ml Expiry date 6 weeks 5 hours Y - lyo. human 1 ml
11776274 serum with 30 µl
added human
CA 72-4
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 66 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrato
Cat.no. frequence target stability stability stability possible material r volume
Calibrator Cal 1/Cal unopened opended on board per
2 2-8°C 2-8°C 20-25°C lyophilized ready- bottle/
to- use Pipetting
volume
CEA ° as required, ≈ 5/50 Expiry date 12 weeks 5 hours Y - buffer/protein 1 ml
11731645 e.g. control ng/ml matrix with 10 μl
not within added CEA
range
Cyfra ° every 1 ≈ 0/50 Expiry date 8 weeks 5 hours Y - lyo. human 1 ml
21-1 month ng/ml serum with 20 µl
11820974 (same lot), added
° every week cytokeratin
(same RP)
NSE ≈ 0.5/50 Expiry date 6 weeks 3 hours Y - buffer/protein 1 ml
12133121 ng/ml matrix with 20 µl
added NSE
free PSA ≈ 0.1/20 Expiry date 12 weeks 5 hours Y - buffer/protein 1 ml
03289796 ng/ml matrix with 20 µl
added
human PSA
total PSA II ≈ 0/60 Expiry date 6 weeks 5 hours Y - - human 1 ml
04485220 ng/ml serum matrix 20 μl
with added
human PSA
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 67 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Cat.no. frequence target stability stability stability possible material volume
Calibra- Cal 1/Cal 2 unopened opended on board per
tor 2-8°C 2-8°C 20-25°C lyophilized ready- bottle/
to- use Pipetting
volume
S100 ° as required, ≈ 0.2/2.0 Expiry date 4 weeks 5 hours Y 3 month lyo. human 1 ml
03289834 e.g. control µg/l (freeze serum with 20 µl
not within only once) added
range S100B(ββ)
antigen
Free ° every 1 ≈ 1/64 IU/L Expiry date 6 weeks 5 hours Y 3 month lyo. human 1 ml
ß-hCG month (freeze serum with 10 µl
04854080 (same lot), only once) added free
° every week ßhCG
(same RP)
PAPP-A ≈ 75/2500 Expiry date - 5 hours Y 3 month lyo. hzman 1 ml
04854191 mIU/L (freeze serum matrix 15 µl
only once) added PAPP-
A
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 68 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Cat.no. frequence target stability stability stability possible material volume
Calibra- Cal 1/Cal 2 unopened opended on board per bottle/
tor 2-8°C 2-8°C 20-25°C lyophilized Pipetting
ready- volume
to- use
Vitamin ° as required, ≈ 100/1500 Expiry date 3 days use only Y 3 months human 1 ml
B12 II e.g. control pg/ml, once (freeze serum with 15 µl
04572459 not within ≈ 74/1107 only added B12
range pmol/l once)
Ferritin ° every 1 ≈ 10/300 Expiry date 12 weeks 5 hours Y - human 1 ml
03737586 month ng/ml serum with 10 µl
(same lot), added
° every week Ferritin
Folate III# (same RP) ≈ 4/17 ng/ml Expiry date 3 days use only Y 3 months lyo. human 1 ml
04476441 ≈ 9.1/38.5 once (freeze serum 25 µl
nmol/l only albumin
once) matrix with
added
Folate
- no information
# reconstitution time 30 minutes
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 69 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Cat.no. frequence target stability stability stability possible material volume
Calibrator Cal 1/Cal unopened opended on board per bottle/
2 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
to- use volume
β- ° as required, ≈ 0.05/2.0 Expiry date 12 weeks 5 hours Y - β-CTx (synthetic 1 ml
CrossLaps/ e.g. control ng/ml peptide) in an 50 µl
serum not within ≈ 50/2000 equine matrix
11972316 range pg/ml
every 1
N-MID ° month (same ≈ 0/280 Expiry date 1 week 5 hours Y 1x human serum 1 ml
Osteocalcin lot), ng/ml (-20°C / 3 with added 20 µl
11972111 ° every week month) Osteocalcin
(same RP) (synthetic
° after seven peptide, human
days (20- sequence)
25°C)
PTH ≈ 0.05/4500 Expiry date 2 weeks 5 hours Y 1x PTH (synthetic 1 ml
04894138 (9 pg/ml (-20°C / 3 peptide, human 50 µl
min) ≈ 0.005/477 month) sequence) in
11972219 pmol/l human serum
(18 min) matrix
PTH (1-84) ≈ 0.1/420 Expiry date 5 days 5 hours Y 1x PTH (synthetic 1 ml
05608554 pg/ml (-20°C / 90 peptide, human 50 µl
≈ 0.01/42 days) sequence) in
pmol/l human serum
matrix
Total P1NP ≈ 25/850 µg/l Expiry date 1 week up to 24 Y 5x P1NP (human) 1 ml
03141080 hours (-20°C / 3 in human serum 20 µl
month matrix
Vitamin D ≈ 2/45 ng/ml expiry date 120 hours up to 5 hours Y 1x 25-OH Vitamin 1 ml
total ≈ 5/113 (-20°C / 28 D3 in human 15 µl
05894921 nmol/l days) serum matrix
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 70 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Cat.no. frequence target stability stability stability possible material volume per
Calibrator Cal 1/Cal 2 unopened opended on board bottle/
2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
to- use volume
IgE ° as ≈ 1/100 IU/ml Expiry date 12 weeks 5 hours Y - equine 1 ml
11930427 required, ≈ 2.4/240 serum 10 µl
e.g. control ng/ml matrix with
not within added
range human IgE
Anti-CCP ° every 1 ≈ 20 / 200 Expiry date - 2 hours Y 4 weeks Anti-CCP 1 ml
month U/ml (freeze antibodies 15 µl
(same lot), only in human
° every once) serum
week matrix
BRAHMS (same RP) ≈ 0.1 / 54 Expiry date - 2 hours Y 3 months PCT in 4 ml
PCT ng/ml (freeze human 30 µl
only serum
once) matrix
IL-6 ≈ 18 / 700 Expiry date - 5 hours Y 3 months IL-6 (human, 2 ml
pg/ml (freeze rec.) 30 µl
only once)
calibrators are part of the test kit
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 71 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 72 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Calibrator frequence target stability stability stability possible material volume
are within range unopened opended on board per bottle/
the testkit Qual.: 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
counts to- use volume
Quant.:
units
Anti-HBe ° as required, Cal1: Expiry date 8 weeks 5 hours Y - Cal1: neg. 1 ml
e.g. control 300000- calibrator, 35µl
not within 1500000 human serum
range Cal2: 1000- Cal2: pos.
° every 1 6000 calibrator,Anti-
month HBe (human) >
(same lot), 3 PEI U/ml in
° every week human serum
HBeAg (same RP) Cal1: 400- Expiry date 8 weeks 5 hours Y - Cal1: neg. cal., 1 ml
2000 human serum 35 µl
Cal2: Cal2: pos.
20000- calibratorHBeAg
100000 (E.coli, rDNA)
approx. 5 U/ml
[PEI units] in
HEPES
Anti-HBs Cal1: 4-15 Expiry date 8 weeks 5 hours Y - Cal1: Anti-HBs 1.3 ml
IU/l (human) in 40 µl
Cal2: 350- human serum
600 IU/l Cal2: Anti-HBs
(human) in
human serum
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 73 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Calibra- frequence target stability stability stability possible material volume
tor are range unopened opended on board per bottle/
within Qual.: 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
the counts to- use volume
testkit Quant.:
units
HBsAg II ° as required, Cal1: 600- Expiry date 8 weeks 5 hours Y - Cal1: neg. 1.3 ml
e.g. control not 1700 calibrator, human
within range Cal2: 3000- serum 50µl
° every 1 month 11000 Cal2: pos.
(same lot) calibrator, HBsAg
° every week 0.5 IU/ml in human
(same RP) serum
HBsAg II ≈ 0 / 0.5 IU/ml Expiry date 8 weeks 5 hours Y - Cal1: neg. 1.3 ml
quant calibrator, human
serum 50µl
Cal2: pos.
calibrator, HBsAg
0.5 IU/ml in human
serum
Anti-HCV Cal1: 350- Expiry date 8 weeks 5 hours Y - Cal1: neg. cal. , 1.3 ml
1200 human serum 40 µl
Cal2: 10000- Cal2: pos.
38000 calibrator, anti HCV
- no information Calibration must be performed with every new reagent kit
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 74 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Calibra- frequence target stability stability stability possible material volume
tor are range unopened opended on board per bottle/
within Qual.: 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
the counts to- use volume
testkit Quant.:
units
HIV Ag ° as required, Cal1: 600- Expiry date 4 weeks 5 hours Y - Cal1: neg. cal. , 1 ml
e.g. control 1400 human serum 50 µl
not within Cal2: 18000- Cal2: pos.
range 55000 calibrator,HIV
° every 1 p24-antigen
month (same (E.coli, rDNA) 240
lot) pg/ml in acetate
° every week buffer
HIV combi (same RP) Cal1: 1600- Expiry date 8 weeks 5 hours Y - Cal1: neg. cal., 1 ml
5600 human serum 30 µl
Cal2: 24 000 Cal2: pos.
– 110 000 calibrator, human
serum, anti-HIV
positive
HIV combi Cal1:1200- Expiry date 12 weeks 5 hours Y Do not Cal1: neg. cal., 1 ml
PT 5000 freeze. human serum 40 µl
Cal2: 17000- Cal2: pos.
75000 calibrator, human
serum Anti-HIV-1
positive
- no information
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 75 of 142
For all Torch assays: Calibrators are part of the test kit; lot number of RackPack and calibrators are identical
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Calibra- frequence target stability stability stability possible material volume per
tor are range unopened opended on board bottle/
within the Qual.: 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
testkit counts to- use volume
Quant.:
units
CMV IgG ° as required, e.g. Cal1: neg. expiry date 8 weeks 5 hours Y - Cal1: neg. 1 ml
control not within Cal2: ~ 40 calibrator, human 20 µl
range U/ml serum (hs)
° every 1 month Cal2: pos.
(same calibrator, hs
CMV IgG lot) Cal1: neg. expiry date 8 weeks 5 hours Y - Cal1: neg. 1 ml
Avidity ° every week Cal2: ~ 40 calibrator, human 20 µl
(same RP) U/ml serum (hs)
Cal2: pos.
calibrator, hs
CMV IgM Cal1: neg. expiry date 4 weeks 5 hours Y - Cal1: neg. 1 ml
Cal2: pos. calibrator, hs 10 µl
Cal2: pos.
calibrator, hs
HSV-1 IgG Cal1: 600- expiry date 28 days 5 hours Y 1 freeze / Cal1: neg. 1 ml
20000 thaw cycle calibrator, hs 20 µl
Cal2: 35000- possible Cal2: pos.
500000 calibrator, hs
HSV-2 IgG Cal1: 600- expiry date 28 days 5 hours Y 1 freeze / Cal1: neg. 1 ml
6500 thaw cycle calibrator, hs 20 µl
Cal2: 28000 - possible Cal2: pos.
300000 calibrator, hs
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 76 of 142
Assay Calibration Calibrator Calibrator Calibrator Calibrator Calibrator Freezing Calibrator Calibrator
Calibrator frequence target stability stability stability possible material volume
are within range unopened opended on board per bottle/
the testkit Qual.: 2-8°C 2-8°C 20-25°C lyophilized ready- Pipetting
counts to- use volume
Quant.:
units
Rubella IgG ° as required, Cal1: neg. expiry date 8 weeks 5 hours Y - Cal1: neg. 1 ml
e.g. control Cal2: ~ 400 calibrator, 10 µl
not within IU/ml human serum
range (hs)
° every 1 Cal2: pos.
Month (same calibrator, hs
Rubella IgM lot) Cal1: neg. expiry date 4 weeks 5 hours Y - Cal1: neg. 1 ml
° every week Cal2: ~ 700 calibrator, hs 10 µl
(same RP) U/ml Cal2: pos.
calibrator, hs
Toxo IgG Cal1: neg. expiry date 8 weeks 5 hours Y - Cal1: neg. 1 ml
Cal2: ~ 100 calibrator, hs 10 µl
IU/ml Cal2: pos.
calibrator, hs
Toxo IgM Cal1: neg. expiry date 4 weeks 5 hours Y - Cal1: neg. 0.67 ml
Cal2: ~ 130 calibrator, hs 10 µl
U/ml Cal2: pos.
calibrator, hs
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 77 of 142
Control ready- lyo- Package Concen- Volume Stability Stability Stability Stability Control
Cat.no. to-use philized content tration per unopened opened or freezend 20-25°C/ materia
range bottle 2-8°C reconstitu- on board l
ted
2-8°C
PreciControl Y 2 levels, * 3 ml Expiry date 3 days one month -/ human
Universal each level (only freeze 5 hours serum
11731416 two bottles once) # matrix
PreciControl Y 2 levels, * 2 ml Expiry date 3 days three month -/ human
Cardiac II each level (only freeze 3 hours serum
04917049 two bottles once) matrix
after thawing
use only once
PreciControl Y 2 levels, * 2 ml Expiry date 4 days three month -/ rec
Troponin each level (freeze only 5 hours human
05095107 two bottles once) serum
after thawing matrix
use only once
PreciControl Y 1 level, ≈ 0.2 2 ml Expiry date 12 weeks - -/ equine
TSH four µIU/ml 5 hours serum
11776479 bottles matrix
- no information
* Please check in package insert.
# Stability of insulin in the reconstituted control serum: on the analyzers at 20-25°C and at 20-25°C: up to 5 hours/at –20°C 1 month (only freeze
once)
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 78 of 142
Control ready- lyo- Package Concen- Volume Stability Stability Stability Stability Control
Cat.no. to-use phili- content tration per unopened opened or freezend 20-25°C/ material
zed range bottle 2-8°C reconsti- on board
tuted
2-8°C
PreciControl Y 2 levels, * 3 ml Expiry date 2 weeks one month 24 hours/ human
Tumormarker each level (only freeze 5 hours serum
11776452 two bottles once) matrix
PreciControl Y 2 levels, * 3 ml Expiry date 72 hours 31 days 5 hours/ human
Varia$ each level (only freeze 5 hours serum
05618860 two bottles once) matrix
- no information
* Please check in package insert.
$ controls not barcoded for cobas e 602
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 79 of 142
Control ready- lyo- Package Concen- Volum Stability Stability Stability Stability Control
Cat.no. to-use philize content tration e per unopened opened or freezend 20-25°C/ material
d range bottle 2-8°C reconsti- on
tuted board
2-8°C
PreciControl Y 2 levels, * 2 ml Expiry date 72 hours 31 days -/ equine serum
Multimarker each level (freeze only up to 5 matrix
05341787 three bottles once) hours
PreciControl Y 2 levels, ≈ 4/16 IU/l 2ml Expiry date - three months 3 hours / 3 human serum
ThyroAB each level Anti-TSHR (only freeze hours matrix
05042666 one bottle once)
PreciControl Y 3 levels, * 2 ml Expiry date - three months up to 5 human serum
Maternal Care each level 2 (only freeze hours /- matrix
04899881 bottles once)
PreciControl Y 2 levels, ≈ 30/100 2 ml Expiry date - one month 5 hours / 5 human serum
Anti-CCP$ each level U/ml (freeze only hours matrix
05031664 two bottles once)
PreciControl Y 2 levels, ≈ 0.5 / 10 4 ml Expiry date - three months -/ human serum
BRAHMS each level ng/ml (only freeze 2 hours matrix
PCT# two bottles once)
* Please check in package insert. # Calibrators and Controls are part of the test kit. $ controls not barcoded
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 80 of 142
Control ready- Package Nominal Volume Stability Stabilit Stabilit Control material
Cat.no. to-use content range for per unopened y y on
cutoff index bottle 2-8 oC opened board
(qual.) 2-8 oC
Concentration
range (quant.)
PreciControl No 2 levels PC1: approx. 20 4.0 ml Expiry date 8 weeks 7 hours Anti-HAV antibodies (human)
Anti-HAV each level IU/l PC 1: ≈ 20 IU/l
11876350 two bottles PC2: approx. 35 PC 2: ≈ 36 IU/l
IU/l in human serum
PreciControl Y 2 levels PC1: negative 0.67 ml Expiry date 8 weeks up to 14 PC 1: human serum
Anti-HAV each level PC2: approx. 3 hours PC 2: Anti-HAV IgM antibodies
IgM eight bottles U/ml (human) ≈ 3 U/ml (PEI
11876368 units) in human serum
PreciControl Y 2 levels PC1: 1.05-3 1.3 ml Expiry date 8 weeks 5 hours PC 1: human serum
Anti-HBc each level PC2: 0.14-0.87 PC 2: Anti-HBc antibodies
11876325 eight bottles (human) ≈ 0.75 U/ml
(PEI units) in human
serum
PreciControl Y 2 levels PC1: 0-0.3 1.0 ml Expiry date 8 weeks 6 hours PC 1: human serum
Anti-HBc IgM each level PC2: 1.1-2.5 PC 2: Anti-HBc IgM antibodies
11876333 eight bottles (human) > 130 U/ml (PEI
units) in human serum
PreciControl Y 2 levels PC1: 1.2-2.1 1.3 ml Expiry date 8 weeks 6 hours PC 1: human serum
Anti-HBe each level PC2: 0.30-0.90 PC 2: Anti-HBe antibodies
11876384 eight bottles (human) ≈ 0.25 U/ml
(PEI units) in human
serum
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 81 of 142
PreciControl Y 2 levels PC1: 0.0-0.80 1.3 ml Expiry date 8 weeks 5 hours PC 2: HBsAg (human) ≈ 0.2 IU/ml in
HBsAg II each level PC2: 2.6-5.0 human serum
04687876 eight bottles
PreciControl Y 2 levels PC1: 0-0.3 1.3 ml Expiry date 8 weeks 5 hours PC 2: Anti-HCV in human serum
Anti-HCV each level PC2: ~ 8
03290379 eight bottles
* see package insert
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 82 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 83 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 84 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 85 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 86 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 87 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 88 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 89 of 142
HIV Ag *
HIV combi
HIV combi PT *
Torch CMV IgG 1:20 with Diluent ≥ 15 U/ml
Universal
CMV IgG Avidity 1:20 with Diluent Universal*
CMV IgM∞ *
HSV-1 IgG *
HSV-2 IgG *
Rubella IgG 1:20 with Diluent > 10 IU/l
Universal
Rubella IgM∞ #*
Toxo IgG 1:20 with Diluent > 3 IU/l
Universal
Toxo IgM∞ #*
Autodilution possible for bold mentioned dilution ratios
∅ Dilution not necessary due to the broad measuring range, or cannot be diluted as this changes the
concentration of the binding proteins and therefore alters the equilibrium.
# Diluent Universal can be used to dilute the samples.
♣ Auto-dilution is not possible as assay uses a three-step method.
∞ Use Elecsys® Diluent Universal for automatic sample predilution.
∇ Autodilution not possible.
* Please check package insert.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 90 of 142
Used Documents:
Assay Package Application report
insert
FT3 1109/12 2ft2050b.xls (NCCLS precision)
FT4 0810/17 2ft4050a.xls (NCCLS precision)
T3 0409/18 2_t3050a.xls (NCCLS precision)
T4 0510/15 2_t4050a.xls (NCCLS precision)
TSH 0510/19 2tsh050a.xls (NCCLS precision)
T-uptake 0409/16 2tbk050a.xls (NCCLS precision)
TG 0310/05 2tg_050a.xls (NCCLS precision)
TG Confirmatory Test 0710/03
Anti-TG 0710/02 2atg050a.xls (NCCLS precision)
Anti-TPO 1010/03 2tpo050c.xls (NCCLS precision)
Anti-TSHR 0610/08 2ATR050a.xls (NCCLS precision)
CK-MB 0511/01 2_3AR01_Applicaiton Report Summary_CK-MB.doc
Myoglobin STAT 0810/16 2myo050a.xls (NCCLS precision)
proBNP II 0510/05 2bn2050a.xls (NCCLS precision)
Troponin T STAT hs 0110/03 2TS9050a.xls (CLSI precision)
Troponin I STAT 0310/02 2_050a_EP5Precision_TNIstat.xls (CLSI precision)
Digoxin 0610/15 2dig050a.xls (NCCLS precision)
Digitoxin 0310/09 2dit050a.xls (NCCLS precision)
HCG STAT 1109/13 2HC2050c.xls (NCCLS precision)
HCG+β 0310/14 2hb2050a.xls (NCCLS precision)
ACTH 0810/05 2acth050a.xls (NCCLS precision)
Cortisol 0710/17 2cor050a.xls (NCCLS precision)
DHEA-S 0509/14 2dhe050a.xls (NCCLS precision)
Estradiol II 0509/15 2e22050a.xls (NCCLS precision)
FSH 0509/16 2fsh050b.xls (NCCLS precision)
LH 0509/16 2_lh050a.xls (NCCLS precision)
Progesterone II 0810/13 2pg2050b.xls (NCCLS precision)
Prolactin II 0509/03 2pro050a.xls (NCCLS precision)
SHBG 0509/08 2shg050a.xls (NCCLS precision)
Testosterone II 0710/04 2_050a_EP5Precision_TestoII.xls (CLSI precision)
C-Peptide 0810/05 2cpe050a.xls (NCCLS precision)
Insulin 0810/10 2in1050a.xls (NCCLS precision)
PlGF 0810/03 2PLGF050a.xls (CLSI precision)
sFlt-1 0710/03 2SFL050b.xls (CLSI precision)
AFP 0410/09 2af2050a.xls (NCCLS precision)
CA 125 II 1110/12 2125050a.xls (NCCLS precision)
CA 15-3 II 0210/13 2215050a.xls (NCCLS precision)
CA 19-9 0410/18 2192050a.xla (NCCLS precision)
CA 72-4 0410/08 2724050a.xls (NCCLS precision)
CEA 0410/18 2ce2050c.xls (NCCLS precision)
Cyfra 21-1 0310/14 2cyf050a.xls (NCCLS precision)
NSE 0310/03 2nse050b.xls (NCCLS precision)
free PSA 0610/09 2pf050b.xls (NCCLS precision)
total PSA 0510/07 2ps2050c.xls (NCCLS precision)
S100 0510/07 2sc_050a.xls (NCCLS precision)
Free ßhCG 0610/04 2PAP050a.xls (NCCLS precision)
PAPP-A 0907/01 2fHC050a.xls (NCCLS precision)
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 91 of 142
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 92 of 142
Used Documents:
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 93 of 142
Used Documents:
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 94 of 142
Used Documents:
For this document are the contents of the current Elecsys® reagent kit, CalSet,
PreciControl and System reagent package inserts and Application Reports used. The
used package insert versions and Application Reports are listed above.Original datas
are available in the corresponding package inserts and Application Reports.
b-TroubleshootingElecsys2010v53.doc
Troubleshooting Elecsys® Systems Page 95 of 142
This is a list of questions you should ask operators with unresolved problems.
Reagent Handling
• Did you adjust temperature of the reagent kit before starting the instrument?
• Were the reagent kits within the allowed stability after opening or were the expiry
dates exceeded?
• Did you observe any foam?
• Did you store the reagent kit in an upright position and at the right temperature (2-
8°C*)?
* The temperature of the fridge should be ≥ 2°C because the microbeads must not
freeze.
Calibrator Handling
• Did you bring the calibrators to room temperature before starting the instrument?
• Were the lyophilized calibrators carefully dissolved? Did you bring the
reconstituted calibrators to room temperature and allow to stand closed for 15
minutes after 1 ml distilled water was added?
• Did you carefully mix the calibrators avoiding the formation of foam?
• Did you transfer the calibrators into the correct bar-coded vial?
• Did you close the calibrator bottles (no single use calibrators) after calibration as
soon as possible? Because of possible evaporation effects, the on board stability
of the calibrators must not be exceeded.
• Did you aliquot and freeze the calibrators? This is recommended for some
CalSets, e.g. CK-MB, Troponin T.
• Did you check that the calibrators were within the allowed stability after opening?
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 96 of 142
Control Handling
• Did you carefully reconstitute the lyophilized controls (see calibrators) e.g. with 3
ml for PreciControl Universal (stand closed for 30 minutes) according to the
information in the package insert?
• Did you transfer the controls into the correct bar-coded vial?
• Did you check that the controls were within the allowed stability after opening and
that the expiry date was not exceeded?
• Did you observe foam on the controls?
• Did you aliquot and freeze controls? This is recommended e.g. for PreciControl
Universal.
Sample Handling
• Did you use the recommended sample containers (outside diameter < 13 mm)?
• Did you use the recommended cup adapter for Elecsys® 2010 rack system when
13 mm sample tubes are used?
• Did you notice air bubbles or foam on the surface of the sample?
• Did you check that the sample stability were not exceeded? Check the package
insert.
• Was dead volume within the specified range? Please check attached list of dead
volume.
• Did you prepare the sample material carefully? Did you allow 30 minutes waiting
time for clotting? Did you centrifuge the samples before putting them on the
analyzer?
• Did the sample contain precipitates? Sample containing precipitates must be
centrifuged before testing at a minimum of * 2500 rpm for 10 min (primary and
normal cups).
• Did you check if there was enough sample volume (plasma)?
• Did sample cups fit correctly in sample disk or rack?
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 97 of 142
Diluent
Maintenance
Lab
• Did you check, whether the temperature and humidity of the lab is within the
defined specification mentioned in the operator’s manual?
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 98 of 142
A reduced dead volume may only be used with Hitachi standard cups (not with primary or secondary cups).
Cup adapter is recommend when primary tubes with 13 mm are used on sample racks.
August 2011
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems
Note: The list in Section 3.6 is based on results and experience from internal studies and
from customer complaints. Please note that this compilation cannot be complete,
since in most cases complaints and troubleshooting actions are not based on an
assessment with all assays.
- mixer speed not within FT3, E2, AFP False results, outliers Check mixer speed
specifications (SN 32/98)
Transport system
- gripper dirt falls into vessels HCG False results, outliers Clean gripper finger
Measuring unit
- magnet disalignment Anti-HBc, CEA Diff. in signal levels Check magnet
Imprecision adjustment
- Tighten torque / alignment TNT STAT False positive results Loosen connection
problem of tubings attached between Möller
to measuring cell tubings & mc. Re-
align tubings & mx
inlet using nylon cord,
SB 2/2000.
- MC/Ref. Electrode Anti-HB>, Diff. in signal levels LFC (SN 22/98)
(RE lifted, deposits, ?) Testosterone, FSH, Imprecision, drifts Change MC or RE
E2 (SN 08/98; 21/98)
Fluidic system
- tubings (leaky, dirty) FT3, T3, FT4, T4 False results Change tubings and
(preheat pipe MC check
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 103 of 142
RackPack
all cal criteria o.k. one cal criteria not o.k. auto system release
auto system release release by operator
Operator released RP calibration only on Elecsys® 2010 and not available on cobas e 411.
Thomas Hartke, Global Product Management, Claudia Schäfer, Global Systems Support December 06
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 104 of 142
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 105 of 142
• single calibrator
• multi use
• Calibrators are measured in duplicate
• CBR Concept for non-infectious diseases
• primary vial concept (plastic vial, snap cap)
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 106 of 142
5.5 Calibration procedure when > one Reagent pack for one assay on board
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 107 of 142
2 Reagent packs (RPs)(same lot): Automatic calibration request of the Automatic calibration request of the RP
1 Reagent pack (new) RP with the lower position number. with the lower position number.
1 Reagent pack (new) Control(s) and samples are also Control(s) and samples are also
pipetted with this RP. pipetted with this RP, when no
additional manual calibration request
of the second RP.
2 Reagent packs (RPs)(different lot): Automatic calibration request of Automatic calibration request of the RP
1 Reagent pack (old, previous used both Rps. Control(s) and samples with the new lot. A manual request of
RP) are pipetted with the old RP. the RP with the old lot is possible.
1 Reagent pack (new) Depending on the control position(s)
and the calibration request is the
control(s) pipetted with the new and/or
old lot (please check first case).
Samples are pipetted with the old lot.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 108 of 142
2 Reagent packs (RPs)(same lot): Take off the old RP. After finishing Automatic calibration request of the
1 Reagent pack (old, previous used the calibration incl. measurement of new RP. Control(s) and samples are
RP) control(s) take off all RP of the test pipetted with the old RP, when the
1 Reagent pack (new), but new and perform a reagent scan. If this control vial(s) is not positioned directly
Reagent pack should be controlled scan has finished replace both RPs behind CalSet vials.
after calibration but sample should and perform an additional reagent In the case that the control vial is
measured with the old Reagent pack scan. If the additional scan is not directly positioned behind the CalSet
performed the new Reagent pack the control is pipetted with the new RP
will be used for pipetting because it and the samples are pipetted with the
is the current Reagent pack (used old RP.
for calibration).
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 109 of 142
QC mode
Please note, that it is not possible to transfer a newly generated lot calibration to a reagent pack which was already used.
To avoid two recommendations “money” and “time” for each example, only the “money” recommendation was described in most of the cases.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 110 of 142
Time out triggered calibration not available on Elecsys 2010 and cobas e 411!
Within the inventory there is the information when the lot calibration and reagent pack calibration has to be carried out. But if the calibration is not
carried out, there is no hint.
Workaround for Anti-TSH Receptor and Anti-TPO (each reagent pack has to be calibrated and in addition each day)
usage of daily calibration feature calibration request for each reagent kit and every day (lot calibration not possible, calibration is
deleted every night)
Usage of daily calibration feature not possible for Anti-HCV (each reagent pack has to be calibrated and in addition each 7 days), because otherwise
each reagent pack would receive a calibration request, although the calibration is only necessary every 7 days.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 111 of 142
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 112 of 142
Calibration reference
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 113 of 142
5.6 Examples of control of calibration in Elecsys® 2010 software version 03-08/12 (not avail. in 06-x or cobas e 411)
b) TSH CalSets, T4 CalSets, empty space, PCU(s): Control as specified in the control definition.
c) TSH CalSets, PCU(s), T4 CalSets, PCU(s): Control of Calibration for RP TSH and RP T4.
d) TSH CalSets, empty space, PCU(s), T4 CalSets, PCU(s): Controls after TSH CalSets as specified in the control definition and control of
calibration with chosen RP T4.
e) HCG CalSets (RP1), HCG CalSets (RP2), PCU(s): Control of calibration with both RPs HCG.
Note: The Control of calibration feature is not any more available in software version 06-x, as each reagent kit can be selected to be
controlled. Please note that in case that a reagent pack has been selected in the Inventory screen to be controlled this has
priorisation. The additional assays selected for e.g. PCU in the control definition will then be ignored.
If no reagent kit will be in the Inventory screen selected to be controlled the assays (current reagent kits) will be controlled as defined
in the control definition screen.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 114 of 142
5.7 Main differences concerning assay calibration between different Elecsys® 2010 software versions
Differences concerning Software V01-39/V 02-13/14 Software V03-08/12 and 06-0x (final software)
Expired Reagent packs Can not be used for calibration. Can be used for calibration, but it will be then
”only” a reagent pack calibration.
Calibration factor The first Lot calibration utilizes a calibration factor Each Lot calibration utilizes a calibration factor 1.
1. The following calibrations (independend of Lot The following Reagent pack calibrations are
or Reagent pack calibration) are compared with compared with the last measured Lot calibration.
this first Lot calibration.
Software 03-x
Software 01-39
Calibration successful 0.6 - 1.4
Calibration successful 0.8 - 1.2
Calibration questionable 0.4 - 0.59 or 1.41 - 1.6
Calibration questionable 0.6 - 0.79 or 1.21 - 1.4
Calibration failed < 0.4 or > 1.6
Calibration failed < 0.6 or > 1.4
Software 06-x (final software)
Advantage: Five instead of four calibration
validation criteria for each calibration. Calibration successful 0.8 - 1.2
Disadvantage: Is the first Lot calibration e.g. Calibration questionable 0.6 - 0.79 or 1.21 - 1.4
because of a ”not recommended reagent handling” Calibration failed < 0.6 or > 1.4
not ok the subsequent calibrations can sometimes
not be released. This is the reason why the limits Advantage: No change in workdisks is necessary
of the factor are widen in software 02-x. Change of when the first Lot calibration e.g. because of a ”not
the workdisk is necessary. recommended reagent handling” is not ok.
Software 02-x Disadvantage: Only four calibration validiation
criterea when a Lot calibration is carried out.
Calibration successful 0.6 - 1.4
Calibration questionable 0.4 - 0.59 or 1.41 - 1.6
Calibration failed < 0.4 or > 1.6
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 115 of 142
Differences concerning Software V01-39 and V02-13/14 Software V03-08/12, 06-x (final software)
Calibration validation criteria Value below minimum signal will be used for Value below minimum signal or above maximum
calibration calculation signal will not be used for calibration calculation.
Differences concerning Software V01-39 and V02-13/14 and V03- 06-x (final software version)
08/12
Calibration validation criteria Monotony is ok, when there is only 1 count Monotony is ok, when there is only 1 count difference
difference between CalSet 1 and CalSet 2. between CalSet 1 and CalSet 2 (as in previous
CalSet 1 counts < CalSet 2 counts for sandwich software versions). But one test independend
assays and additional calibration criteria (minimum acceptable
CalSet 1 counts > CalSet 2 counts for difference) will be added for quantitative assays to
Competitive assays. avoid that the calibration will be released with this
small signal difference between CalSet 1 and CalSet
2.
Calibration reference Reagent packs (unused or partly used) which Reagent packs (used) which have not been
have not been calibrated due to a valid Lot calibrated due to a valid Lot calibration is available,
calibration is available, receive ”new calibration do not change the calibration parameter when a new
parameter” when a new valid Lot calibration is Lot calibration is measured.
measured.
System Pipetting Priorisation 1st Priorisation: Number of determinations in 1st Priorisation: Liquid level detection
the reagent pack 2nd Priorisation: Number of determinations in the
2nd Priorisation: Liquid level detection reagent pack
Manual entered control value 1st priority is the manual entered control value. 1st priority is the manual entered control value, but it
Therefore only when a new control lot is used, is possible to enter this control value reagent lot
the control values are used from the control specific. And therefore when new reagent lot is
barcode card or reagent barcode. scanned in the correct control value is used from the
control barcode card or reagent barcode.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 116 of 142
Each lot calibration uzilizes a calibration factor 1. The following rackpack calibrations are compared with the last measured lot calibration. The
calibration factor is the ratio between the calibrator signals (difference of CalSet 1 and CalSet 2) of the lot and rackpack calibration.
The calibration factor is only used as a calibration validation criteria and not used for sample calculation.
t1
Calibration factor for each Lot calibration = ____ = 1
t1
This “easy” formula is only valid when the same calibrator concentrations are used for the reagent pack and lot calibration. If the concentrations are differnet
the calibration signals of the standardization have to be considered.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 117 of 142
after a service action when there have been changes made or adjustments to the
detection unit.
Exception: When the current RackPack has been on board < 24 hours and the
calibration results in a lot calibration.
To ensure that all needed assay calibrations will be carried out the SCLT Tool
(Software Calibration Liquidation Tool) has been developed to delete all available
assay calibrations after service action.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 118 of 142
An actual reagent lot is measured with a new manufactured Roche control. The
control value is written on the control bar code card independent of the reagent lot.
This control value is valid until a specific target value for a reagent lot exists.
Note
If the difference between the medians of the target values for Elecsys® 2010 <
approx. 1 SD (corresponding depending on the assay approx. 5-10%) the mean of
these medians are used as target value.
A reagent will be standardized against the master lot (master reagent and master
calibrator) with all valid calibrators available.
All valid Roche controls are measured and checked for deviations.
Case 1: All controls are within this target range < +/- 1 SD
The target values of the existing control bar code card are used for this reagent lot,
i.e., the control values are not changed when using this reagent lot.
Disadvantage: The target value of the control is identical, but the recovered value
can have another level within the given range, e.g., 105% for reagent
lot 1 and 94% for reagent lot 2.
Case 2: The controls are out of this target range > +/- 1 SD for this reagent lot
The target values of the controls are on the RackPack bar code, i.e., specific target
values for reagent lots exist and the control values on the control bar code card are
not valid. As a consequence an extra information sheet is put inside the reagent kit
indicating the re-assigned values and the new values are stated on the reagent pack
bar code.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 119 of 142
Notes
Due to the priority rules, the target value of a control will not be taken from the control
barcode card or the reagent barcode for a PreciControl lot –reagent lot combination,
if the target value has been entered manually.for this special PreciControl lot –
reagent lot combination.
Priority 1 Target values of controls put in manually for a special reagent lot.
Priority 2 Target values of controls on the reagent bar code.
Priority 3 Target values of controls on the control bar code card.
That means once a target value of a control is entered manually for a special reagent
lot – PreciControl lot combination, this value is valid for this reagent lot as long as the
customer uses a new reagent lot or a new control lot.
The main point of each standardization action is to receive the same human
serum recovery independent of the reagent lot.
The multi analyte controls are spiked, stripped and preserved and do
unfortunately not always react in the same way.
Biorad Control Values will be assigned with the actual available reagent lot (3
systems, 1 run each, threefold determination) when the Biorad Control is produced,
therefore it will be measured only once per Biorad lot.
Because of this "handling" we do not have experience with Biorad control recovery
deviation when changing reagent lots.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 120 of 142
Connection via test no. and RP or CalSet lot no. Connection via first two digits of test no. and lot identifier
RackPack Lots which will be produced after the CalSet lot, will have the CalSet lot number printed on the RackPack barcode and so the
connection will be vial CalSet lot number.
CalSet lots which will be produced after the RackPack lot, will have the RackPack lot number printed on the CalSet barcode card and so the
connection will be via RackPack lot number.
Connection via test no. and control lot no. (when there is a reassignment) Conn. via control no. and lot
identifier
RackPack lots which will be produced after the Control lot, will have the Control lot number printed on the RackPack barcode when there is a
target values reassignment and so the connection will be via Control lot number.
The test number of each assay is stated on the data disk in one file called refer.dat.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 121 of 142
* only for assays where calibrators are not in the reagent package
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 122 of 142
8. SOFTWARE INFORMATION
Within GRIPS there is within the document type “Software information” where for the defined system all software issues and hints are
described. It consists of an EXCEL spreadsheet representing the reocrd indwex. Via the index a WORD document contains all record details
and optional attachment files can be opened. All these files are connected via hyperlinks.
As soon as an update of a “software information” is published an e-mail notification is send out to all registered users, when the topic “software
and connectivity is defined“
January 05
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 123 of 142
Dear colleagues,
The attached information is a summary of our current knowledge about the throughput of
Elecsys 2010 when different combinations of assays are performed.
This information becomes important with the upcoming official launch of Vitamin B12 and
Folate and for further evaluation activities with infectious disease tests.
The information is a summary of information from different evaluations. Please feel free to
copy this information in any format you judge desirable for distribution to your customers.
Frank Willemse
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 124 of 142
The workflow on the Elecsys® 2010 system is completely sample orientated. Due to the
availability of a new disposable tips for each test, there is no risk of contamination. It
becomes possible to perform assays in any sequence, thus allowing samples to be
completed one after the other.
When all assays on the system are ”18 minute assays” the optimal throughput of 88 results
per hour can be reached, producing a result every 42 seconds. In combination with 9 or 27
minute assays, or in combination with 2-step dilution assays, the instrument will slow down,
dependent on the percentage and sequence of tests with other incubation times.
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 125 of 142
1. 18 minute tests
Typical throughput Elecsys 2010 disk Typical throughput Elecsys 2010 rack
300 300
250 250
200 200
150 150
100 100
50 Disk system
50
Rack system
Theoretical Theoretical
0 0
0:00:00 1:00:00 2:00:00 3:00:00 4:00:00 0:00:00 1:00:00 2:00:00 3:00:00 4:00:00
All 18 minute tests follow the same time-protocol. Therefore, there will be no timing conflicts.
The Elecsys® 2010 will perform in one 42 second time cycle simultaneously
S1 (1st pipetting), S2 (2nd pipetting) and D (detection).
D (detection)
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 126 of 142
The throughput of Elecsys® 2010 depends on the percentage and distribution of the 9 minute
tests. As a limiting factor it is not possible during one 42 second cycle to plan the detection of
2 tests. When scheduling the first pipetting of a 9 minute assay, the system has to be sure to
have an open cycle for detection 9 minutes later. Dependent on the percentage and
distribution of the 9 minute assays, throughput can be influenced or not.
D (detection)
Throughput reduction !
If the number of requested 9 minute assays is very small, bigger ”THROUGHPUT GAPS” will
exist (second part of below graph).
200
180
160
Number of results produced
140
120
100
80
60
THEORETICALOUTPUT
40
Mix of 9and 18 min tests
0
0 0.5 1 1.5 2 2.5 3 3.5 4
Operationtime (inhours)
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 127 of 142
3. 27 minute tests
If only 27 minute assays are performed, the throughput of Elecsys® 2010 is reduced to 44
results per hour. Every 13 cycles, the Elecsys® 2010 comes into a timing problem. It is not
possible to perform within one 42 second cycle a S0 (pretreatment pipetting) together with a
S1 (first reagent pipetting). The instrument will stand for 13 cycles (9 minutes) until it is
possible again to pipette without conflict.
S0 (pretreatment pipetting)
D (detection)
200
180
160
140
results produced
120
100
80
60
0
00:00
00:15
00:30
00:45
01:00
01:15
01:30
01:45
02:00
02:15
02:30
02:45
03:00
03:15
03:30
03:45
04:00
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 128 of 142
When mixing 18 and 27 minute assays, it depends on the assay-mix and on the exact test
sequence, how many gaps are created. The ”gaps” can vary from 1 to 13 idle cycles (42
seconds to 9 minutes). Limiting factors are that only 1 detection can take place during one 42
second cycle and that pretreatment step (S0) can not be combined with first reagent pipetting
(S1). In below example of 20% random distribution of 27 minute assays, the throughput is
reduced with 16.6% (5 cycles on 30 results).
S0 (pretreatment pipetting)
Following data from the internal evaluation show the decrease in throughput, dependent on
the distribution of 18 and 27 minute tests.
40 27 min Assays
20 Only 27 minute
assays
0 00: 00: 00: 00: 01: 01: 01: 01: 02: 02: 02: 02: 03: 03: 03: 03: 04: 04: 04: 04: 05:
00 15 30 45 00 15 30 45 00 15 30 45 00 15 30 45 00 15 30 45 00
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 129 of 142
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 130 of 142
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 131 of 142
Start
No
First order ?
Yes
Were reagents
exchanged ?
No
Yes
Counting tips
Reagent barcode scan and vessels
PC/CC
volume
90 min or more Clearing the check
passed since last incubator
Inventory mixing ? No and the T/V
sending to buffer
the host Yes
computer Beads mixing
Alarm:
Inventory OK? Inventory
Short
Short(something)
Enough 45-xx-01
First pipetting
First sipping
Pipetting is continuing
Resume Cycle
Sipping is continuing
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 132 of 142
Preparation 1
(Flowing PC/CC without
voltage)
Preparation 3-1
(Flowing PC/CC with
voltage)
Preparation 3-2
(Flowing PC/CC with
voltage)
Preparation 3-3
(Flowing PC/CC with voltage
and AB current/count check)
End
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 133 of 142
Count/Current
Check Start
Attach Flag 41 or 43
to all measurements.
Alarm; PC/CC LLD It depends on the
PC LLD ? Fail 37-05-01 timing the alarm
occured that which
OK
flag will be attached.
OK
OK
End
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 134 of 142
11.2 Resume
Yes
Yes
13 cycles or more ?
No
Yes
Next determination
(starts direct after resume)
End
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 135 of 142
11.3 Finalization
Last sipping
10 cycles waiting for the new oder* *10 cycles for Disk system
10 cycles for RS
60 cycles for RC
40 cycles for Disk since 7/E version
Finalization 40 cycles for RS since 7/E version
Filling the MC
Yes with the PC
Stand by
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 136 of 142
11.4 LFC
Fill in ISE-Solution
Y = X+10 (N-1)
Sipper move down Y mm
N = repeat times
X = fixed Aspirate 2.5 ml, 150 μl/s ISE-Solution
Incubate 6 min.
Sipper Prime
Finalization
Standby
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 137 of 142
Liquid type ?
Reagent
Sample or Calibrator
Sucking, Z motor
Sucking starts
down and CD start
NG
Pressure check ? when Alarm ; Clot detected, 49-01-01
OK
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 138 of 142
No Bright
Alarm
Abnormal descnet senor ? when Z motor stops
Sipper Z 20-02-05
Dark
Z-motor stops
OK
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 139 of 142
Valve Block
MC
System Water Circulation
17
Liquid Flow System 15
SV3 SV4
Waste Tubing/Modification 0
18
T Parts Name Parts No Material
Line Heater
No 9
8
0 Tube 0 741-0996 19
1 Tube 1 741-0807
16
2 Tube 2 741-0808
8 Tube 8 741-0543
7
5
SV1/2/7/5/6
14
Waste
12/13 Modification
System Water
6 Liquid Waste
10
Pump
b-TroubleshootingElecsys2010v53.doc-SC 11 Filter
Troubleshooting Elecsys® Systems Page 140 of 142
Why could it be Due to Elecsys® Universal Diluent is one bottle only, it is Therefore we would recommend to inform
possible that an possible that the diluent is not correct placed on the reagent your customer that the Elecsys® Universal
auto dilution is not rotor that means instead of the Elecsys® 2010 barcode will Diluent RackPack should be placed very
possible any more, be used the E 170 module for MODULAR ANALYTICS carefully that the Elecsys® 2010 barcode
although a bottle barcode. And then it is not possible to use this diluent for and not the E 170 module for MODULAR
Elecsys® Universal auto dilution on Elecsys® 2010, due to the needed Elecsys® ANALYTICS barcode will be used. The
Diluent is on test number is not available but the Application Code barcode reader of Elecsys® 2010 is outside
board? number for E 170 module for MODULAR ANALYTICS. In the reagent rotor, whereas the barcode
case that the first RackPack of the Elecsys® Universal reader of E 170 for MODULAR
Diluent Lot is not placed correctly, it is not possible to use ANALYITCS is inside the reagent rotor.
another RackPack of the same Elecsys® Universal Diluent
again, due to if the same lot number will scanned, the
information will be used from the data base.
What to do when a Please repeat within the next 24 hours with the same If the calibration is repeated > 24 hours, it
„bad“ calibration reagent pack a „good“ calibration e.g. with new CalSets on is possible that the RackPack calibration is
was released as lot Elecsys® 2010. Then the „bad“ calibration is overwritten and not released because of the calibration
calibration on not available any more. factor. It is also possible to carry out a new
Elecsys® 2010? lot calibration with another reagent pack.
Then this reagent pack with the „bad“
calibration can be calibrated again. The
RackPack calibration refers with the
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 141 of 142
What could be the We would like to inform you that starting with software
reason when flag version 06-x for each determination an effective signal check
58 signal level is carried out to improve the system reliability (flag 58 signal
below limit appears level below limit). In the Operators Manual is the remedy
for competitive "rerun the sample". But please note that if an extremely high
assays? sample concentration is measured for a competitive assay,
this result will also be flagged. This can easily be checked
by the dilution of the sample. Therefore please rerun the
diluted sample. When a new version of Operators Manual
will be introduced, the remedy "rerun diluted sample" will be
added.
What has to be Please check the unit, because for the instrument is this a new assay. That means when the customer
considered when an does not use the default unit, the unit has to be adapted. In addition possibly a new CalSet, Control and
improved test with Diluent has to be used, where the new test number is corresponding.
new test number is
used?
What has to be Please use the default unit when the intercept is used as instrument factor, even when the customer
considered when an does not use the intercept.
instrument factor is
used?
b-TroubleshootingElecsys2010v53.doc-SC
Troubleshooting Elecsys® Systems Page 142 of 142
What is the dead The dead volume in the CalSet and Control vials is 250 µl for software version 03-x, but 150 µl for the
volume in the software version 06-x.
CalSet vials for
software version
03-x and 06-01
(final software)?
b-TroubleshootingElecsys2010v53.doc-SC