Manufactured by:
NuAire, Inc.
2100 Fernbrook Lane
Plymouth, Minnesota USA 55447
Toll Free 1-800-328-3352
In MN 612-553-1270
OM0080
Fax 612-553-0459
OM0080 2
Congratulations!
You have just purchased one of the finest Laminar Flow Biological Safety Cabinets
available. With proper care, maintenance (certification), and laboratory procedure, this cabinet
will give you years of product and personnel protection from particulate contaminants as
prescribed in National Sanitation Foundation (NSF) Standard No. 49. Please read this manual
carefully to familiarize yourself with proper installation, maintenance and operation of the cabinet.
Appendix A lists Training Aid Materials, Films, Pamphlets and Books that may assist you in the
proper application/procedure of the cabinet.
Acknowledgment
NuAire, Inc. acknowledges that some material in this manual reflects information supplied by the National Institutes of
Health personnel both in verbal and written specifications. In particular, NuAire acknowledges that information in Section 8
was obtained from the following sources:
1. Technical Report No. FPS 56500000001. Prepared by Dow Chemical Co., for the National Cancer Institute, May 1,
1972.
2. Stolar MH, Power LA, Vielo CS: Recommendations for handling cytotoxic drugs in hospitals. Am J Hosp Pharm
1983;40:
1163-1171.
OM0080 3
3. Anderson R.W., Director of Pharmacy, University of Texas, M.D. Anderson Hospital and Tumar Institute at Houston.
OM0080 4
ABOUT THIS OPERATION & MAINTENANCE MANUAL
The information contained in this manual is intended to reflect our current production standard
configuration model along with the more frequently purchased options. Any unique
additions/modifications/shop drawings are appended in the back flap of this manual, along with any
modifications and/or additions to procedures as outlined in this manual. A copy of the original
factory test report is also appended to this manual. In case this manual and/or test report is lost or
misplaced, NuAire retains a copy in our files. A replacement copy can be obtained by calling or
writing NuAire, Inc. stating the model number and serial number and a brief description of the
information desired.
OM0080 5
Labgard Class II, Type A/B3
Laminar Flow Biological Safety Cabinet
Models NU-425-300/400/600
Operation & Maintenance Manual
TABLE OF CONTENTS
Section No. 1---------------------------------------------------General Description
Section No. 2---------------------------------------------------Construction Features
Section No. 3---------------------------------------------------Models & Options
Section No. 4---------------------------------------------------Test Performance & Procedures
Section No. 5---------------------------------------------------Warranty
Section No. 6---------------------------------------------------Shipments
Section No. 7---------------------------------------------------Installation Instructions
7.1------------------------------------------------------------Location
7.2------------------------------------------------------------Set-up Instructions
7.3------------------------------------------------------------Certification
Section No. 8---------------------------------------------------Operating the NU-425
8.1------------------------------------------------------------Operator Controls & Indicators
8.2------------------------------------------------------------Operating Guidelines
8.3------------------------------------------------------------Operating Sequence
8.4------------------------------------------------------------Antineoplastic Decontamination Procedure
Section No. 9---------------------------------------------------General Maintenance
9.1------------------------------------------------------------Fluorescent Lamp Replacement
9.2------------------------------------------------------------HEPA Filter/Motor Replacement
9.3------------------------------------------------------------Sliding Window Replacement & Adjustment
9.4------------------------------------------------------------Airflow Calibration
9.5------------------------------------------------------------Filter Integrity Check
9.6------------------------------------------------------------Cabinet Leak Tightness Test
Section No. 10--------------------------------------------------Error Indicators & Troubleshooting
Section No. 11--------------------------------------------------Remote Contacts
Section No. 12--------------------------------------------------Optional Equipment
12.1-----------------------------------------------------------Ultraviolet Lamp
Section No. 13--------------------------------------------------Electrical/Environmental Requirements
ASSEMBLY DRAWINGS
BCD-05147----------------------------------------------------Base Stand Assembly
BCD-05146----------------------------------------------------Base Stand Storage Cabinet Assembly
BCD-05145----------------------------------------------------Control Center & Front Decorative Panel Assembly
BCD-05144----------------------------------------------------Sliding Window Assembly & Adjustment
BCD-05153----------------------------------------------------Base Cabinet Assembly
ELECTRICAL SCHEMATICS
BCD-05033---------------------------------------------------NU-425-300/400/600 Electrical Schematic
OM0080 6
Labgard Class II, Type A/B3
Laminar Flow Biological Safety Cabinet
Models
NU-425-300/400/600
MANUFACTURED BY:
NuAire, Inc. - Plymouth, Minnesota, U.S.A.
The Laminar Flow Biological Safety Cabinet, (LFBSC) is a product resulting from the
development of the "laminar flow" principle and the application of environmental controls as
required in the field of biological research or chemical containment. The LFBSC, when used with
proper technique, is an effective laboratory aid in obtaining the optimum control over product
quality while reducing the potential for exposure of both product and personnel to airborne
biological or particulate chemical agents in low to moderate risk-hazard research and drug
preparation or product operations, as prescribed by the Center for Disease Control (CDC)
Atlanta, Georgia.
The NU-425 bench LFBSC is known as a "convertible cabinet", meeting the requirements
of Class II, Type A and, when vented to the outside, meeting the requirements of Class II, Type
B3, hence the A/B3 designation. This is possible since NuAire's cabinet conforms to the following
requirements:
1. Maintains inflow velocity of 100 LFPM (.51 M/S) through the work access
opening.
2. Has HEPA filtered downflow air that is mixed with the inflow air from a
common exhaust plenum (A or B3 requirements are identical).
3. Discharges all air to the outside atmosphere after HEPA filtration. (Type A may
exhaust back to the laboratory).
4. Has all biologically contaminated ducts and plenums under negative pressure or
surrounded by negative pressure. (Type A permits positive pressure
contaminated ducts and plenums).
The LABGARD Bench/Console Model NU-425 has six distinct characteristics that
enables it to perform the foregoing objectives:
HEPA Filtration
Controlled Airflow, Internally Balanced
Sealed Construction
Hepex Pressure Plenum
Minimum Air Turbulence
Front Maintenance
OM0080 7
1.1 High Efficiency Particulate Air (HEPA) filtration is utilized for both supply
(recirculated) and exhaust air systems. The HEPA filters are rated at 99.99% efficiency
on removal of all particulate matter 0.3 micron with greater efficiency on larger and/or
smaller particles.
1.2 Controlled Air Flow (see ACD-05179 Airflow Schematic) with an internal balancing
damper establishes an optimum air balance between the cabinet's recirculated air system
and the exhaust air system while:
c) Ducting the work zone airflows such that the quantity of air leaving
through the rear perforated area is half of the recirculated air quantity.
The remainder, that is the other one-half (50%) of the recirculated air
and the air entering through the work access opening enters the front
perforated area.
1.3 Sealed construction requires that all recirculated and/or exhaust air plenums (from
immediately downstream of the work surface to the discharge side of the supply and
exhaust HEPA filters) be leak-tight. Plenums and penetrations as necessary for utilities,
access panel doors, etc., are sealed in such a manner as to prevent escape of potentially
contaminated air into the room.
1.4 HEPEX Pressure Plenum provided by the HEPEX Absolute Filter System results in
greater reliability for the continuing Class 10, Federal Standard 209d (1), Bio-clean
environment within the working zone of the cabinet. The air-space between the HEPEX
and the cabinet structure is always less than room ambient pressure which prevents any
air outflow from the LFBSC other than the filtered outflow from the exhaust filter. The
absolute filter is nested in non-flammable elastomeric foam in a manner which ensures
zero leak and preserves the Class 10 environment free from the rigors of shipping and
moving.
OM0080 8
1.5 Minimum air turbulence is maintained by having all work zone enclosure walls
(including the viewing window) provide a fixed flow area perpendicular to the air flow
streamlines. This condition of minimum turbulence with no upward or refluxing
component is optimized by having each enclosure wall essentially become an extension of
the true flow area of the HEPA filter - thus permitting the clean filtered air to flow
downward like a huge piston forcing clean air onto the work tray before the air
recirculates through the cabinet airflow/blower system.
Minimizing turbulence at the front opening is effected by the airfoil turning the air stream
entering the cabinet to a direction parallel to the flow of the recirculated air within the unit
thus preventing the recirculated air from flowing out through the front opening into the
room and the room air entering the cabinet from overflowing onto the work surface.
Truncating the inlet grill against the side wall of the cabinet produces air velocities 25%
greater around the side wall airfoil, protecting this critical area from migrating particles.
In addition, the volume of air provided by the 10 inch (254mm) access opening produces a
comprehensive "sphere of influence" throughout the work access opening for particulate
control, since particles must have more energy to escape. Escape velocities are directly
proportional to the volume, and inversely proportional to the area plus the square of
distance from the inlet grill.
1.6 Vibration is minimized (less than 100 micro inches (2.54 micrometers) on the work
surface) through the use of vibration isolators; first around the motor mounts fastening
the motor to the blower and second when attaching the motor/blower combination to the
frame. All blower scrolls are dynamically balanced.
OM0080 9
(ACD-05179)
(REV A)
OM0080 10
2.0 Construction Features
Standard Features
2.1 The NU-425 cabinet shell is pressure tight, consisting of an all welded, 16 gauge, type 304
stainless steel wrap. All welds are meticulously hand-finished and polished to be free of
burrs. The wrap is structurally reinforced to minimize vibration and maintain surface
flatness. The stainless steel is finished in easily cleanable No. 4 satin. The cold-rolled
steel 14 gauge access panel and light/control center is finished in textured baked urethane.
Colors available upon request.
2.2 All LABGARD controls are contained within an easily removable Control Center, placed
at eye level for convenience. All connections from the Control Center to the cabinet are
hermetically sealed at the cabinet shell, with quick disconnect electrical connections at the
Control Center. The Control Center is hinged for easy access to electrical components
and the fluorescent light.
2.3 All cabinet surfaces, interior and exterior, are constructed of, or furnished with, materials
which are corrosion, flame and moisture resistant and which will not deteriorate under
exposure to liquid or vapor phase decontaminates such as hydrogen peroxide,
formaldehyde, alcohol, iodophers, peracetic acid. USE OF CHLORINATED OR
HALOGEN MATERIALS MAY DAMAGE STAINLESS STEEL.
2.4 The work area consists of the following items going from front to back: the airfoil, the
work access opening, the removable perforated stainless steel inlet grill, the removable
solid stainless steel work tray and the perforated stainless steel air exhaust area located in
the rear wall of the work area. The top lip of the airfoil is one inch above the removable
solid stainless steel work tray which extends the full length and width of the work space.
2.5 Air leaves the work space via the perforated area in front of the work tray, and via a
similar perforated opening extending the full length of the work space in the rear wall just
above the work tray. The air removed via the front and rear perforated areas is mixed by
the blower, exhausting approximately 30% through a HEPA filter located on top of the
LFBSC and recirculating approximately 70% through a HEPA filter supplying the interior
downflow.
2.6 The exhaust filter is protected with a perforated metal grill, to prevent blockage and/or
damage of the exhaust HEPA. However, items should not be stored on the top of the
cabinet. When exhausting the cabinet to the atmosphere, the grill is not used.
2.7 The supply filter is of full work zone depth and width and is protected with an easily
removable, perforated metal diffuser, in order to check filter integrity.
2.8 Both the supply and exhaust HEPA filters are 99.99% efficient on particulate 0.3 microns
in diameter with even greater efficiency on larger and/or smaller particles. When
replacing the HEPA filters use only filters of the same size and flow rating as originally
installed (see Replacement Parts List).
2.9 A drain-spillage plenum is located below the work surface, so arranged as to catch any
spillage in the work area, and is easily cleaned after the removal of the solid work tray. A
3/8 inch (10mm) ball-type drain valve is located on the right front of the plenum to handle
large spills. The drain spillage trough accommodates a minimum of five (5) liters. This
area also forms the return air path to the blower.
OM0080 11
2.10 The viewing window is a full counter balanced sliding 1/4 inch (6mm) tempered glass
window that provides for smooth opening/closing. The window provides a 19 1/2 inch
(495mm) maximum open area to fully closed. The window frame is constructed from
type 304 stainless steel with all exposed edges deburred for easy cleaning, finished in No.
4 satin grain. Both audible and visual alarms are provided when the window is raised
above the standard work opening height (10 inches (254mm) for the 4 & 6 foot cabinet, 8
inches (203mm) for the 3-foot cabinet, and optional for the 4 & 6 foot cabinet). The
blower is shut off when the window is closed to prevent unnecessary motor
overheating. The blower does not shut off for hinged window cabinets.
2.11 The two (T8) fluorescent lamps are cool white, rapid start type, located outside the work
space, obscured from direct view providing an average uniform light intensity of 100 foot
candles (1076 LUX) at the work surface. The electronic ballast contains thermal
protection with automatic reset. PLEASE NOTE, ELECTRONIC BALLASTS
OPERATE WITH HIGH IN RUSH CURRENT. IT IS NOT RECOMMENDED
TO USE THIS PRODUCT WITH GROUND FAULT CIRCUIT
INTERRUPTERS (GFCI'S) BECAUSE THE BALLASTS MAY CAUSE THE
GFCI TO TRIP.
2.12 Electrical power is required from one 115 Vac, 60 Hz, 20 Amp power source. The power
is distributed internally for the blower, lights and duplex outlets via solid state relays and
triacs. Cabinet shell and all electrical components are grounded back to electrical source,
to meet grounding continuity requirements for electrical safety.
2.13 The motor/blower requires 115 Volt, single phase, 60 Hz power and will not exceed a
temperature of 105°C in a maximum ambient temperature of 48°C (120°F) under any
maximum load condition (class "B" insulation). The thermal protector will not trip at 115%
of the rated voltage under maximum load and ambient temperature conditions. The
motor/blower is rated for 24 hour continuous operation and is lubricated for life.
2.14 The LABGARD has only one motor/blower fan system controlled by the motor voltage
regulator for both the recirculated and exhaust air. The motor/blower and blades are
fabricated of or protected with corrosion resistant materials to withstand normal
laboratory or chemical fumes. The motor/blower will automatically compensate for
airflow as the filters load with particulate to achieve a fan delivery fall off of no more than
10% as a result of a 80% increase in pressure drop across the filters. The Electronic
Airflow Control System will provide a greater than 200% increase in pressure drop across
the filters is achievable.
2.15 The LABGARD is designed for clean ability: surface decontamination is made simple by
use of coved corners and an easily removable work tray over the spillage trough.
Complete biological decontamination can be accomplished following the recommended
NIH procedure using formaldehyde. The NIH procedure can be obtained from the
National Audiovisual Center (GS) Washington, D.C., 20409 in the form of a tape cassette
and slides.
2.16 Two outlets with drip-proof covers are located on either side of the rear divider panel
above the work tray as standard (300 unit has only one centered on the rear divider
panel). Collectively the outlets are rated at 115 Vac at 3 amperes maximum.
2.17 One ground-key cock for gas service is located in the right side wall as standard. One
plugged penetration for future service is also provided. The valve body and handle is
OM0080 12
made from chrome plated brass forgings with the valve ground and lapped and tested at
100 psi (7.0 Bar). Cocks have integral ten serration outlet and individually coded by a
removable button on the handle. Cocks are rated 30 psi (2.0 Bar) working pressure.
OM0080 13
2.18 A normally plugged D.O.P. challenge port is provided on the top of the cabinet. The port
provides access to the pressure plenum for checking the concentration of D.O.P.
challenging the HEPA filter, if the measuring instrument requires such access.
2.19 The textured baked urethane finish is a high-solids chemical agent resistant coating
(CARC), and meets new United States Environmental Protection Agency (EPA) solvent
emission regulations with less than 3.5 lb./gal. (.419Kg/Lit.) volatile organic content. The
CARC coating contains no chlorinated hydrocarbons and withstand chemical and
biological reagents absorption into the paint. They survive corrosive, humid and salt
environments.
Optional Features
2.20 An ultraviolet light provides a minimum of 40 micro-watts per square centimeter on the
work tray, controlled from the Control Center with a switch which is controlled so that
the UV light cannot be used when the sliding window is opened or the normal fluorescent
lamps are on (see Section 12.1).
2.21 Ground-key cocks (up to three) can be installed in either or both side walls. The type of
service is indicated by a removable button on the handle. Needle valves for high pressure
gases are optionally availa ble.
2.22 The base support stand is constructed from 16 gauge, all welded 2-inch (51mm) square
steel tubing, finished in textured baked urethane. The base stand heights are
26 1/2-inches or 32 1/2-inches (673mm or 825mm) which provide work surface heights of
30-inches or 36-inches (762mm or 914mm) respectively. The base support stand is
available as a base storage cabinet, that is fabricated from 16-gauge cold-rolled steel. It
consists of rear, end and bottom panels fastened to the base stand via screws. Sliding
doors provide complete access to the 16-inch (406mm) cabinets interior height (depth is
22-inches) (559mm). NSF approved leg levelers permit ± 3/4-inch (20mm) of height
adjustment. The base support stand is shipped knocked-down for on-site assembly.
2.23 NuAire offers an EXHAUST TRANSITION for use with LABGARD Laminar Flow
Biological Safety Cabinet to provide a transition between the LABGARD and a plant
exhaust system which preserves the air balance within the LABGARD. The hood
provides access for replacement of the exhaust HEPA filter and means to conduct
fumigation of the cabinet.
2.24 A lay-in Sorbent filter is available for the exhaust efflux. The Sorbent filter is a 2-inch
(51mm) pack, bolted and gasketed to the exhaust HEPA filter via a CRS frame and held
in place with spring clips
2.25 The cabinet can be configured with plastic storage trays, which are attached to the
underside of the cabinet via stainless steel rails. The trays are high-impact polyethylene
plastic available in red. The base stand without storage cabinets is a prerequisite for the
trays.
2.26 Decorative side panels can be added to cover plumbing lines to the services within the
interior. The panels are constructed from 16 gauge, cold rolled steel finished in textured
baked urethane. Colors optional.
OM0080 14
3.0 Models & Features
The model NU-425, Class II, Type A Laminar Flow Biological Safety Cabinet is
manufactured in four sizes: 2 ft., 3 ft., 4 ft., and 6 ft. This manual only covers the 3, 4
and 6 foot cabinets.
3.1 Dimensions
Interior Dimensions:
Width (W): 36 3/8 (873) 46 3/8 (1178) 70 3/8 (1788)
Depth (D): 23 1/2 (597) 23 1/2 (597) 23 1/2 (597)
Height (H): 28 1/2 (724) 28 1/2 (724) 28 1/2 (724)
OM0080 15
3.3 Optional Features
OM0080 16
(BCD-04902)
(REV C)
OM0080 17
(BCD-04903)
(REV D)
OM0080 18
(BCD-04904)
(REV D)
OM0080 19
4.0 Test Performance & Procedures
All equipment is thoroughly inspected at the NuAire Factory at the time of shipment.
Quality control is maintained by constant surveillance over the product, beginning at the receipt of
purchased material and concluding with a final inspection which certifies cabinet performance to
the National Sanitation Foundation Standard No. 49, as well as any unique customer requirements.
In all instances where product quality cannot be easily assessed on the end item, product
inspection is performed during sub-assembly fabrication. The following test procedures are
conducted on each cabinet and a copy of the test report is included with each unit.
OM0080 20
4.3 HEPA Filter Integrity/Housing & Frame Leak Test
Both supply and exhaust HEPA filters are scanned for leaks around the filter seal
periphery, filter housing and filter media. Passes are made overlapping to ensure
complete coverage. D.O.P. smoke is introduced upstream using one Laskin Nozzle per
each 135 CFM (752 m3/H). Downstream measurements using an Aerosol Linear
Photometer must be less than 0.01% of upstream concentration.
4.10 Vibration
Vibration amplitude will not exceed 200 micro inches RMS is measured with a Bruel and
Kjaer, type 2511 in the center of the work tray in three principle plane (no dimensional
requirements) at 20 to 20,000 hertz.
Note: NuAire has conducted the test on a representative production cabinet in order to validate
the design and containment properties of the design. NuAire will periodically conduct
these tests as a Quality Assurance measure. Cabinets so tested are labeled and
decontaminated prior to shipment. These tests may also be conducted at customer
request and expense. All of the other above tests are conducted on each cabinet prior to
shipment and a copy of the test report accompanies this Operation and Maintenance
Manual.
OM0080 22
5.0 Warranty
NuAire, Inc. warrants that it will repair F.O.B. its factory or furnish without charge
F.O.B. its factory a similar part to replace any material in its equipment within 36 months after the
date of sale if proved to the satisfaction of the company to have been defective at the time it was
sold provided that all parts claimed defective shall be returned, properly identified to the company
at its factory, charges prepaid. Factory installed equipment or accessories are warranted only to
the extent guaranteed by the original manufacturer, and this warranty shall not apply to any portion
of the equipment modified by the user. Claims under this warranty should be directed to NuAire,
Inc. setting forth in detail the nature of the defect, the date of the initial installation and the serial
and model number of the equipment.
This warranty shall not apply to any NuAire product or part thereof which has been
subject to misuse, abuse, accident, shipping damage, improper installation or service, or damage by
fire, flood or acts of God. If the serial number of this product is altered, removed or defaced as to
be illegible, the Warranty shall be null and void in its entirety.
The warranty is for the sole benefit of the original purchaser and is not assignable or
transferable. Prior to returning any item, for any reason, contact NuAire for a Return
Authorization Number. This number must accompany all returns. Any product shipped to NuAire
without this number will be returned refused shipment or collect freight.
6.0 Shipments
NuAire takes every reasonable precaution to assure that your LABGARD cabinet arrives
without damage. Motor carriers are carefully selected and shipping cartons have been specially
designed to insure your purchase. However, damage can occur in any shipment and the following
outlines the steps you should take on receipt of a NuAire LABGARD cabinet to be sure that if
damage has occurred, the proper claims and actions are taken immediately.
6.1.1 Terms are factory, unless stated otherwise. Therefore, it is important to check
each shipment before acceptance.
6.1.2 If there is visible damage, the material can be accepted after the driver makes a
notation on the consignee's copy of the freight bill. Then an inspection must be
made to verify the claim against the carrier. This inspection is the basis of your
filing the claim against the carrier.
OM0080 23
7.0 Installation Instructions
7.1 Location
Within the laboratory, pharmacy, etc., the ideal location of the biological safety cabinet is
away from personnel traffic lanes, air vents (in or out), doors and/or any other source of
disruptive air currents.
If drafts or other disruptive air currents exceed the inflow velocity of the cabinet through
the access opening, the potential exists for contaminated air to exit or enter the work
zone area of the cabinet. It depends on the severity of the air current. REMEMBER:
A BIOLOGICAL SAFETY CABINET IS NO SUBSTITUTE FOR GOOD
LABORATORY TECHNIQUE.
Where space permits, a clear 12" (305mm) area should be permitted on each side of the
cabinet for maintenance purposes. The electrical outlet into which the cabinet is
connected should be readily accessible for maintenance purposes. A MINIMUM
CLEARANCE OF 6" (152MM) IS REQUIRED FROM THE TOP OF THE CABINET
TO THE CEILING FOR PROPER VENTILATION OF THE EXHAUST EFFLUX.
HOWEVER, FOR CERTIFICATION OR COMMISSIONING, 18 INCHES (451MM)
IS REQUIRED TO OBTAIN VALID EXHAUST MEASUREMENTS.
If it is desired to exhaust the cabinet to the outside, NuAire offers two general categories
of exhaust transitions which will capture the exhaust efflux from the cabinet. These are:
Thimble or Air Gap Exhaust Transitions (with and without integral fan)
Gas-Tight Exhaust Transitions
Both types of transitions have some common attributes, in addition to some that are
unique. The exhaust transition selected is dependent on desired performance of the
cabinet, capabilities of the plant exhaust system and expenditure desired. See separate
instruction sheets for a discussion of exhaust transitions and installation requirements.
OM0080 24
The base stand storage cabinets will usually be shipped according to
customer requirements. If it is shipped unassembled, it can be assembled per
drawing BCD-05146. It is recommended that the upper and lower base stand
braces be installed first, then the rear and bottom panels (the end panels are
always prefastened). Once assembled, fasten the cabinet per the above
instructions.
Remove the plastic cap protecting the drain valve threads and install the drain
valve, on the bottom right front of the cabinet using Loctite 242 furnished to the
threads and rotate the valve body until it is secure (see ACD-01855).
7.2.2 Leveling
Using a level placed on the work tray, adjust the leg levelers, first, end to end,
then, front to back. The NSF approved leg levelers provide a ± 3/4" (20mm)
adjustment.
If a drain valve is desired, remove the handle from the valve stem to gain
clearance for valve body rotation. Add Loctite 242 (furnished) to the threads
and rotate valve body until secure, with the valve stem (for handle) on the left
side. Re-install handle to valve stem. Adjust the cabinet on bench to provide a
1-1/2" (38mm) overhang and seal the interface of the bench and cabinet, using
RTV caulk as above.
External connection is to 3/8 inch NPT coupling in the inner side walls.
Connection to plant utilities should be made with proper materials for the
individual service and according to national and/or local codes. It is not
recommended that flammable gases be used in the cabinet. However, if
flammable gas is used, emergency shut-off valves should be located in an
accessible area external to the cabinet. Observe all labels pertaining to the type
of service and operating pressure. THIS UNIT HAS NOT BEEN
EVALUATED FOR USE WITH FLAMMABLE, TOXIC OR
EXPLOSIVE SUBSTANCES. USE AT YOUR OWN RISK.
OM0080 25
(BCD-05154)
(REV B)
OM0080 26
7.2.5 Electrical Services
The NU-425 series Biological Safety Cabinets may be "hardwired" (optional) or
connected via an electrical power cord which is standard. The unit requires 115
VAC at 20 Amps, 60 Hz, single phase. It is recommended that power to the
unit be on its own branch circuit, protected with a 20 Amp circuit breaker
respectively or fuse at the distribution panel.
Do not attempt to clean the HEPA filter media. Cabinet interior walls or work
surface are easily cleaned with any mild household detergent cleaner using a soft
cloth. Turn the cabinet on and let it operate for 60 minutes before using it as a
LFBSC.
7.3 Certification
After installation and prior to use, NuAire recommends that the cabinet be certified or
commissioned to factory standards. At a minimum, the following tests should be
performed (see Section 4.0).
1. HEPA filter media
2. Filter frame leak test
3. Airflow velocities
4. Airflow smoke patterns
Of these tests, in order to insure that no disruptive air currents are penetrating the air
inflow barrier (see Section 7.1) smoke flow tests must be performed at a minimum (only
biological testing can confirm containment). These tests must result in the containment of
smoke passed around the perimeter of the work access opening, as well as, no refluxing
or drift of smoke within the interior of the cabinet.
Note that the LABGARD cabinets, filters and seals provide premium performance;
Quality Control in both design and manufacturing assure superior reliability. However,
protection to both product and operator is so vital that certification to the performance
OM0080 27
requirements should be accomplished as stated to ensure biologic al safety established by
the factory standards.
OM0080 28
Labgard Class II, Type A/B3
Laminar Flow Biological Safety Cabinet
Models NU-425-300, 400, 600
Catalog Number
Catalog Number NU-425-300 NU-425-400 NU-425-600
Nominal 3 foot (0.9m) Nominal 4 foot (1.2m) Nominal 6 foot (1.8m)
Performance Specifications NIH-03-112c NIH-03-112c NIH-03-112c
1. Personal Protection NSF Std. No. 49 NSF Std. No. 49 NSF Std. No. 49
2. Product Protection NSF Std. No. 49 NSF Std. No. 49 NSF Std. No. 49
3. Air Cleanliness/Laminarity Fed. Std. 209d, Class 10 Fed. Std. 209d, Class 10 Fed. Std. 209d, Class 10
NSF Std. No. 49 Class Class 11, Type A/B3 Class 11, Type A/B3 Class 11, Type A/B3
Style of Cabinet Bench top/console w/base Bench top/console w/base Bench top/console w/base
stand/storage cabinet stand/storage cabinet stand/storage cabinet
Cabinet Construction All welded stainless steel All welded stainless steel All welded stainless steel
16GA, Type 304 pressure 16GA, Type 304 pressure 16GA, Type 304 pressure
tight design tight design tight design
Diffuser for Air Supply (Metal) Non-flammable Non-flammable Non-flammable
HEPA Filter Seal Type:
Supply Filter-99.99% Eff. on 0.3 HEPEX Seal HEPEX Seal HEPEX Seal
microns Neoprene, Springloaded Neoprene, Springloaded Neoprene, Springloaded
Exhaust Filter-99.99% Eff. on 0.3
microns
Fumigation per NIH/NSF Procedure Yes Yes Yes
Standard Services:
Service Coupling (3/8 inch NPT) One, Right Sidewall One, Right Sidewall One, Right Sidewall
Gas Valve/Service Coupling (3/8inch One, Right Sidewall One, Right Sidewall One, Right Sidewall
NA) One, Backwall Center Two, Backwall Two, Backwall
Duplex Outlet
Optional Services:
Gas Cocks 3/8" NPT Up to 3 ea. Sidewall Up to 3 ea. Sidewall Up to 3 ea. Sidewall
Ultraviolet Light One, Backwall One, Backwall One, Backwall
Standard/Cup Sinks Left or Right Work Surface Left or Right Work Surface Left or Right Work Surface
Cabinet Size Inches (mm):
Height (Fully Assembled) 63 (1600) 63 (1600) 63 (1600)
Height (Minimum for Transport) 60 (1524) 60 (1524) 60 (1524)
Width 41 5/8 (1057) 53 5/8 (1362) 77 5/8 (1972)
Depth (with Control Center) 32 7/8 (835) 32 7/8 (835) 32 7/8 (835)
Work Access Opening Inches (mm):
Standard Opening Height/Optional 8 (203) 10 (254)/8 (203) 10 (254)/8 (203)
Standard Inflow Velocity 105 FPM (.53 m/s) 105 FPM (.53 m/s) 105 FPM (.53 m/s)
Work Zone Inches (mm):
Height 28 1/2 (724) 28 1/2 (724) 28 1/2 (724)
Width 34 3/8 (873) 46 3/8 (1178) 70 3/8 (1788)
Depth 23 1/2 (597) 23 1/2 (597) 23 1/2 (597)
Viewing Window Inches (mm): Fully closed to Fully closed to Fully closed to
Standard is tempered sliding glass 19 1/2 (495) open 19 1/2 (495) open 19 1/2 (495) open
Hinged Tempered Glass (optional) 8 (203) opening 8 (203) & 10 (254) 8 (203) & 10 (254)
openings openings
Required Exhaust CFM/CMH Standard/
Optional:
Gas-Tight (NU-916/919) 200 (340) 338 (575)/270 (459) 513 (872)/410 (697)
Thimble (NU-918/917) 282 (479) 438 (739)/370 (624) 647 (1100)/545 (925)
Thimble (NU-916) 326 (554) 496 (842)/428 (727) 709 (1205)/606 (1030)
Plant Duct Static Pressure Eng/Metric 0.05-0.1"/1.27-2.54mm 0.05-0.1"/1.27-2.54mm 0.05-0.1"/1.27-2.54mm
Heat Rejected,BTU,Per Hour(non- 1442 1937 2435
vented) 865 1162 1460
(vented)
OM0080 29
Electrical: UL/UL-C Listed UL/UL-C Listed UL/UL-C Listed
Volts, AC 60 Hz 115 115 115
Amps: Blower/Lights 7 9 11/*9
Amps: Duplex 3 3 3
Amps: Total 10 12 14/*12
12 ft. Power Cord (one) 14 GA - 3 Wire, 15A 14 GA - 3 Wire, 15A *12/14 GA-3 Wire, 20/15A
Crated Shipping Weight: 340 lbs./154 kg. 490 lbs./222 kg. 680 lbs./308 kg.
Net Weight 305 lbs./138 kg. 441 lbs./200 kg. 598 lbs./271 kg.
*15A Configuration does not include accessory outlet
OM0080 30
8.0 Operating the NU-425
8.1 Operator Controls & Indicators
The following is a description of the controls and indicators found on both the front panel
(see Drawing BCD-05322) and cabinet.
The purpose of the outlet is to provide "switched" power for the following (see
also 8.1.11):
OM0080 32
(BCD-05322)
(REV A)
OM0080 33
8.1.8 Airflow Control
The operating airflows within the cabinet (i.e. 70 LFPM (.35 m/s) downflow and
105 LFPM (.53 m/s) air inflow barrier) are controlled by a potentiometer and an
exhaust damper (see Section 9.4). The potentiometer controls the operating
voltage applied to the motor/blower. The potentiometer is adjustable over 270
degrees with a slotted screwdriver, which varies the applied voltage from 70 to
115 VAC. THIS ADJUSTMENT SHOULD ONLY BE MADE BY A
QUALIFIED TECHNICIAN EMPLOYING THE PROPER
INSTRUMENTS IN ORDER TO INSURE AIRFLOWS PER NSF
STD. NO. 49.
8.2.1 Procedure protocols defined in terms of the barrier or control concepts unique to
LFBSC must be developed in order to obtain a maximum potential for safety and
protection. The pre-planning necessary to develop these protocols is based on
OM0080 34
several fundamental considerations, each of which will contribute to optimum
benefits from the equipment:
Unnecessary raising of the hands inside the cabinet above the level of the work
opening should be avoided. This presents an inclined plane from hands to elbows
along which the downflow of air may run to, and possibly out, the open face.
Note: When working with agents of lower risk, it is not as important for all materials to
be placed in the cabinet before starting, or for the procedure to be completely
finished before materials are removed. Also, the time period for a unit may be
continued over a more extended period during which entries and withdrawals
from the cabinet may be made.
A solid object placed in a laminar air stream will disrupt the parallel flow and
consequently, the capability of controlling lateral movement of airborne
particulates. A cone of turbulence extends below the object and laminarity of the
air stream is not regained until a point is reached downstream, approximately
equal to three to six times the diameter of the object. Within the parameters of
this cone, particles may be carried laterally by multidirectional eddy currents.
OM0080 35
8.2.5 Employ Aseptic Technique
The operator must not assume an attitude of "let the cabinet do it" when
performing procedures within a LFBSC. Properly balanced and properly used
cabinets will do an excellent job of controlling airborne contamination and
containing viable agents, but the cabinet will not eliminate contact transmission
of contamination. Normal laboratory contamination control procedures and
basic aseptic techniques are necessary to obtain maximum benefit from the
cabinet. For example, open bottle, tube or flask mounts should be kept as
parallel as possible to the downflow to minimize capture of chance particulates.
This precaution is merely an extension of good aseptic technique as practiced on
open bench tops. The good laboratory practices designed to minimize creation
and/or release of aerosols to the environment should not be discontinued.
Items of equipment in direct contact with the etiologic agent must remain in the
cabinet until enclosed or until surface-decontaminated. Trays of discard pipettes
must be covered before removal from the cabinet (aluminum foil may substitute
for fabricated covers).
Applying a burner flame to flask and tube necks when mating surfaces of sterile
assemblies is a conventional method of minimizing chance contamination.
However, the efficiency of this operation is usually related to the removal of
airborne contamination occurring while the item is uncovered. If the manipulation
is carried out in an environment free of airborne particulates, then the need for the
flaming operation is essentially removed. This is one of the additional advantages
of the LFBSC - use of the gas burner is seldom necessary.
The gas burner flame in one of these units not only contributes significantly to the
heat build-up, it also disrupts the laminar air streams which must be maintained for
maximum efficiency. If the procedure demands use of a flame, a burner with a
pilot light should be used so that the flame can be ignited only when actually
required and immediately turned off again after use. DO NOT USE
CONSTANT FLAME GAS BURNERS. It should also be placed to the rear
of the work space where resulting air turbulence will have a minimal effect. If
cabinet air is inadvertently turned off, the flame could damage the HEPA filters.
OM0080 36
8.3 Operating Sequence
8.3.1 Start Up
Turn on cabinet blower and lights, check air intake and exhaust portals of the
cabinet to make sure they are unobstructed. The electronic airflow control
system will automatically control airflows to specified setpoints. However,
upon filter loading, the cabinet may be required to be rebalanced or filters
replaced. Only a qualified maintenance technician should perform cabinet
balancing and filter replacement.
Note: Some cabinets are equipped with ultraviolet (UV) lights. Good procedure
includes the decontamination or wipedown of cabinet surfaces with chemical
disinfectant before work commences. This practice eliminates the need for UV
lights, whose primary utility in this application is inactivation of surface
contamination since the filters effectively remove all airborne contaminants.
UV lights, therefore, are not recommended in the LFBSC.
8.3.2 Wipedown
The interior surfaces of the work space should next be disinfected by wiping
them thoroughly with 70% alcohol or similar non-corrosive anti microbial
agents. USE OF CHLORINATED OR HALOGEN MATERIALS IN
THE CABINET MAY DAMAGE STAINLESS STEEL.
OM0080 37
8.3.5 Perform Work
The work can now be performed. The technician performing the work is
encouraged to wear a long-sleeved gown with knit cuffs and rubber gloves. This
will minimize the shedding of skin flora into the work area and concurrently
protect the hands and arms from viable agent contamination. At a minimum, the
hands and arms should be washed well with germicidal soap before and after
work in the cabinet. For the preparation of antineoplastic drugs, the following
procedures summarize those contained in OSHA Instruction PUB 8-1.1, "Work
Practice Guidelines for Personnel Dealing with Cytotoxic (Antineoplastic)
Drugs." The above document should be thoroughly studied/reviewed prior to
drug preparation in the cabinet.
OM0080 38
8.3.8 Shut Down
Turn off blowers and lights. Do not use cabinet as a depository for excess lab
equipment during periods of non-operation. If antineoplastic agents are being
prepared in the cabinet, it is recommended to let the cabinet run 24 hours per
day. This lessens the possibility that contaminants may escape.
8.4.1 Preparation
Prior to beginning decontamination activity, personnel should put on a Tyvek1
isolation gown, 2 pair of vinyl gloves and a full faced HEPA filtered respirator.
All protective garments should be contained in 4 mil plastic bags and labeled for
disposal as chemotherapy waste after completion of the procedure (see also
Section 9.0 of the Operation and Maintenance Manual). For the purpose of this
procedure, the term CLEANING is defined as the operation of wiping down
with a cloth wetted with a clean hot (above 60°C) detergent solution, followed
by wiping down repeatedly with sterile water to rinse. All cloths shall be
contained in 4 mil plastic bags and la beled for disposal as chemotherapy waste.
8.4.2 Procedure
a. Make sure that the cabinet remains in operational mode with internal
blower on.
b. Open the hinged or sliding view screen and secure in the full open
position.
CAUTION: With the view screen in the full open position, personnel
protection is compromised and a full faced HEPA filtered
respirator must be worn.
Note: Depending on the model, the diffuser screen is secured to the cabinet by
#8-32 screws or 1/4" - 20 acorn nuts, 3 places. It is purposely a tight fit
and is secured to the back wall with projecting threadless studs.
e. Clean both sides of the perforated metal diffuser screen and remove it
from the cabinet.
f. Lift the cabinet work tray, clean both sides and remove it from the
cabinet.
g. Remove the front perforated grill, place on the cabinet floor and clean
both sides. Remove from cabinet.
OM0080 39
1 Available from Lab Safety Supply, Janesville, WI 53547-1368, or other laboratory, industrial, or hospital supply
distributors.
OM0080 40
h. Clean work tray supports.
i. Working from top to bottom, clean all inside surfaces of the cabinet.
Take care not to wet the HEPA filter. If liquid has collected in the
plenum drain, aspirate it using an IV tubing into an evacuated container.
Label the evacuated container for disposal as chemotherapy waste.
8.4.3 Assembly
b. Replace the work tray and carefully tighten the thumb screws.
d. Wipe down all exposed surfaces of the work area with 70% isopropyl
alcohol.
OM0080 41
9.0 General Maintenance
CAUTION: All maintenance actions on this equipment must be performed by a
qualified technician who is familiar with the proper maintenance
procedures required for this equipment. This includes both
certification as well as repair.
If the cabinet has been used to prepare antineoplastic drugs, (chemotherapy), or other
toxic chemicals, decontamination of the cabinet cannot be accomplished by the above procedure.
It is recommended that the following protective measures be taken:
1. Gloves
Gloves must be worn. Care must be taken not to cut, puncture, or tear the
gloves. No one glove material is impervious to all CYTAs; disposable surgical
or polyvinyl chloride (PVC) gloves provide substantial but not complete
protection. PVC gloves probably are more protective than surgical gloves, but
they are stiffer and le ss tactile. Gloves should be discarded after each use.
Gloves should be tucked into the cuffs of the gown. Double gloving should be
considered.
3. Gowns
A protective garment must be worn. The garment should be made of lint-free,
low-permeability fabric and must have a closed front, leg sleeves, and elastic or
knit closed cuffs. Tyvek1 isolation gowns are one example of an acceptable
garment. The garment must be worn outside the work area. Disposable gowns
are preferred over reusable. Front-buttoned coats are not recommended.
5. Motion
Slow and deliberate motions are necessary when working in the interior of the
cabinet, in order to minimize the generation of particulates.
Please consult with NuAire, Inc. about any unique contamination problems.
Normally, no preventive maintenance is required on the interior of the cabinet (i.e., the
area behind the access panel containing the HEPA filters and motor (blower assembly). All
required adjustments in order to maintain proper cabinet airflows are external to the cabinet
interior. The motor is lubricated for life and is thermally protected with automatic reset.
OM0080 42
1 Available from Lab Safety Supply, Janesville, WI 53547-1368, or other laboratory, industrial, or hospital supply
distributor
OM0080 43
NuAire does, however, recommend that the LABGARD have the integrity of the HEPA
filters verified by a qualified technician per paragraph 4.3 after the unit has been initially installed.
Therefore, certification per Section 4.0 should be performed on an annual basis for whenever the
operator has reason to believe it necessary, especially if the cabinet has been moved to a new
location, or the minihelic gauge has increased by 0.1" w.g.
2. Second, remove the screws at each upper side of the Control Center
and allow the Control Center to rotate down, resting on the safety straps.
4. The lamp bulb is removed by displacing the lamp to one side against the
compressible lamp holder and lifting out the lamp.
5. Reverse the procedure to reinstall the lamp assembly being careful not to pinch
the safety straps, cable or tubing during closure of the control center.
The HEPA Filters should not be replaced until the entire cabinet has been decontaminated
or known to be biologically "clean". See Section 9.0 General Maintenance for additional
precautions for chemical contamination.
Step 1: Remove screws at each upper side of the control center and allow the
control center to rotate down, resting on the safety straps. Second,
remove the front decorative panel which is held into position by (3)
knurled nuts on the top edge and (6) knurled screws on the front.
Step 2: Place sliding window into lowest position and remove front filter panel,
which is held into position by Phillip pan head screws. Once the screws
are removed, the panel is held into position by smooth weldstuds located
on the top corner of the front filter panel. Use the window stop brackets
as handles to remove the panel.
OM0080 44
CAUTION: Screws are used in lieu of acorn nuts, and lockwashers. The
screws have O-rings and should be replaced if damaged or
badly deformed.
The interior of the cabinet is now fully exposed for replacement of the filters and/or
motor/blower.
Note: Double sticky back gasket is used to hold the plenum to the
choke tray and should be replaced when reassembling.
2. Pull the exhaust choke tray free and remove the filter. It is not
necessary to remove the tray, although it is free to move forward
several inches, if necessary, to free the HEPA filter.
OM0080 45
Step 4: Filter Installation
When installing new filters, use only filters of the same rated flow and size as
originally installed. It is recommended that a new HEPEX/Supply filter be
installed since the HEPEX is factory installed to the filter. However, field
installation kits are available separately from the filter (see Section 13.0).
a. To install the supply filter, simply reverse the procedure outlines in Step
3a, above.
Note: Be sure to open the choke plate fully before inserting the filter into the
tray. This will assist in adjusting the airflow.
b. To install the exhaust filter, apply a thin layer of silicone grease to the top
and bottom gaskets of the filter and carefully insert into the exhaust
choke tray.
Position the filter frame within the outside walls of the exhaust opening on
the top of the hood. Tighten the spring loaded bolts, 4 places, depressing
the gasket material by 1/8 inch (3mm).
OM0080 46
9.4 Airflow Calibration
The NU-425 Airflow Calibration Consists of adjustments to balance the airflow within
the cabinet. THIS WORK SHOULD BE DONE ONLY BY A QUALIFIED
TECHNICIAN WHO CAN MEASURE THE AIRFLOW FROM THE FILTERS
WITH A SUITABLE VELOMETER. NuAire provides two adjustments to balance
the airflow within the cabinet. These are:
The blower speed control system adjusts the cabinets total volume of airflow while the
choke adjusts or balances the exhaust airflow as well as makes up for filter resistance
tolerances. Since it has been NuAire's experience that the filters may not "load" evenly,
both adjustments are necessary for proper cabinet performance.
b. Inflow average: 105 LFPM ± 5 LFPM (.53 m/s ± .025 m/s) using the direct
inflow measurement method or related value using the calculated
inflow velocity measurement method. Both values are published
in the NSF or NuAire listing.
OM0080 47
9.4.2 Inflow Calibration
The diffuser plate placed below the HEPA to protect the filter during normal usage may
be removed as follows: The diffuser is secured to the cabinet shell by #1/4-20 acorn nuts
located immediately behind the front viewing window. After removing the fasteners, drop
the front of the diffuser plate several inches and pull forward gently. Note, that the
diffuser is purposely a tight fit - it is secured to the back wall of the cabinet interior by a
light push - fit with projecting studs.
The exhaust filter is typically more difficult to check, since charcoal filters, or exhaust
transitions could cover the filter. Access panels are usually provided and should be
removed. If an air gap exhaust transition is provided, the air gap must be sealed with duct
tape or other suitable means to prevent contaminated air from migrating into the exhaust
efflux. All exhaust blowers/fans should be turned off during the check.
OM0080 48
Table 9.0
Recommended Measurement Methods for Cabinet Downflow & Inflow.
A. Downflow Measurement
a. Instruments: Alnor 8500 or TSI 8355 Thermoanemometer
b. Procedure: Supply filter efflux is measured on a grid, in a horizontal plane
defined by the bottom edge of the window. No readings should be taken closer
than 6 inches (152mm) from the inside perimeter.
c. Test Data - Inches (mm):
300 6 11.594 17.188 22.782 28.375
(152) (295) (437) (597) (721)
400 6 11.729 17.458 23.187 28.916 34.645 40.375
(152) (298) (443) (589) (735) (880) (1026)
600 6 11.838 17.676 23.514 29.352 35.190 41.028 46.866 52.704 58.542 64.375
(152) (301) (449) (597) (746) (894) (1042) (1190) (1339) (1487) (1635)
6
(152)
11.750
(298)
17.5
(445)
B. Inflow Measurement
a. Instrument: Shortridge Flowhood ADM-870 or Alnor 8500 or TSI 8355
Thermoanemometer.
b. Procedure:
The inflow velocity is measured by using a Direct Inflow Measurement (DIM)
Instrument (i.e. shortridge flowhood). The DIM Instrument can be used directly
on the cabinet with NO CORRECTION FACTORS REQUIRED since NSF has
tested the cabinet using the same technique. The DIM Instrument should be
equipped with a flowhood that is as close as possible to the width of the cabinet
(i.e. NU-425-400 should use 1 x 4 foot flowhood). The DIM Instrument should
also be duct taped to the cabinet to prevent any sneak air paths from occurring.
The DIM Instrument will read inflow volume (i.e. CFM). Use the window
access opening area to calculate inflow velocity.
The calculated inflow velocity measurement method may also be used, but the
setpoint values must be obtained in the NSF Standard 49 cabinet listing. Exhaust
filter air velocity is measured on a four by four (102mm by 102mm) grid scale
four inches (102mm) above the filter, starting from the inner edge of the filter
frame. Use the area table to calculate the inflow velocity based upon the exhaust
air volume.
1. Dim Measurement
Inflow Volume ft.3/min.(m3/s) Access Opening ft.2(m2) Inflow Velocity
OM0080 49
ft./min(m/s)
OM0080 50
2. Calculated Inflow Velocity Measurement - Inches (mm):
d. Acceptance Criteria:
1. Access Opening inflow velocity = 100 to 110 fpm (.51 to .56 m/s)
OM0080 51
10.0 Error Indicators & Troubleshooting
Audible alarms and error indicators occur for a variety of reasons. Whenever an alarm
condition is present, the audible alarm and error indicator will be presented and stay on until the
error is cleared. When presented with an error indicator, please perform the following:
Step 1: NOTE ALL ERROR INDICATORS. When the cabinet is running, any and all
red indicators display an error.
OM0080 52
Cabinet outlets won't turn on. Check outlet circuit
breaker on top of control
center. Check voltage to
outlets.
OM0080 53
11.0 Remote Contacts
12.1.1 Overview
The germicidal ultraviolet is primarily intended for the destruction of bacteria and
other microorganisms in the air or on directly exposed surfaces. Approximately
95% of the ultraviolet radiation's from germicidal tubes are in the 253.7
manometer region. This is a region in the ultraviolet spectrum which is near the
peak of germicidal effectiveness. The exposure necessary to kill bacteria is the
product of time and intensity. High intensities for a short period of time, or low
intensities for a longer period are fundamentally equal in lethal dosage on bacteria
(disregarding the life cycle of bacteria). The intensity of light falling on a given
area is governed by the inverse law; that is the killing intensity decreases as the
distance increases from the tube.
Since ultraviolet rays will not penetrate ordinary glass, it is recommended that the
sliding window be closed while the ultraviolet light is on within the cabinet; or that
personnel leave the cabinet face area.
12.1.2 Operation
The operation of the ultraviolet lamp is accomplished by closing the sliding
window and pressing the UV switch located on the front panel.
12.1.3 Precaution
The rays from germicidal tubes may cause a painful but temporary irritation of
the eyes and reddening of the skin, if of sufficiently high intensity, or if exposure
covers a prolonged period of time. For this reason, one should avoid direct eye
and skin exposure to ultraviolet light. If exposure cannot be avoided, it is
necessary for personnel to wear eye goggles or face shields, and long sleeve
gowns with rubber gloves.
Since ultraviolet rays will not penetrate ordinary glass, it is recommended that the
sliding window be closed while the ultraviolet light is on within the cabinet; or that
personnel leave the cabinet face area.
OM0080 54
12.1.4 Maintenance
The output of an ultraviolet lamp deteriorates with burning age. The useful life
of the lamp is approximately 3750 hours under specific test conditions. If the
tube is turned on every day for 12 hours, the tube will last approximately one
year.
OM0080 55
13.0 Electrical/Environmental Requirements
13.1 Electrical
NU-425-300 115V, 60Hz, 1 Phase, 10 Amps
NU-425-400 115V, 60Hz, 1 Phase, 12 Amps
NU-425-600 115V, 60Hz, 1 Phase, 14 Amps
(Note: The EMC performance requirements are generated within the product enclosure.
The enclosure will be all metal grounded to earth. In addition, the membrane front panel
will also include a ground plane for maximum protection and an electrostatic shie ld.
OM0080 56
Energies Required to Destroy Some Microorganisms By Ultraviolet Radiations(e)
Microwatt Microwatt
Mold Spores seconds Protozoa seconds
per cm/2 per cm/2
Penicillum roqueforti 26,400 Paramecium 200,000(a)
Penicillium expansum 22,000
Penicillium digitiatum 88,000 Nematode Eggs 40,000(b)
Aspergillus glaucus 88,000
Aspergillus flavus 99,000 Algae 22,000(c)
Aspergillus niger 330,000
Rhizopus nigricans 220,000 Virus
Mucor racemosus A 35,200 Baceriophage (E. Coli) 6,600
Mucor racemosus B 35,200 Tobacco Masaic 440,000
Oospora lactis 11,000 Influenze 3,400(d)
Yeasts
Saccharomyces 13,200
ellipsoideus 17,600
Saccharomyces cerevisiae 13,200
Brewers' yeast 6,600
Baker's yeast 8,800
Common yeast cake 13,200
Bacteria
Streptococcus lactis 8,800
Strep. hermolyticus (alpha type) 5,500
Staphylococcus aureus 6,600
Staphylococcus albus 5,720
Micrococcus sphaeroides 15,400
Sarcina lutea 26,400
Pseudomonas fluorescens 7,040
Escherichia coli 7,040
Proteus vulgaris 7,480
Serratia marcescens 6,160
Bacillus subtilis 11,000
Bacillus subtilis spores 22,000
Spirillum rubrum 6,160
References:
(a) Luckiesh, Matthew (1946) Application of Germicidal, Ethyemal and Infrared Energy, D.
Van Nostrand o., New York, New York, pp 253
(b) Hollaender (1942) Aerobiology, A.A.A.S. (for 90% inactivation), pp 162
(c) Ellis, C. and Wells, O.O. (1941) The Chemical Action of Ultraviolet Rays, Reinhold
Publishing Corp., pp. 713-714
(d) Hollaender, A., Oliphant, J.W. (1944) The inactivation effect of monochromatic
ultraviolet. Radiation on Influenze Virus (for 90% inactivation) Jour. of Bact. 48, pp. 447-
454
(e) This table, "Energies Required to Destroy Some Microorganisms by Ultraviolet
Radiations" comes from Westinghouse brochure entitled - "Westinghouse Sterilamp
Germicidal Ultraviolet Tubes"
OM0080 57
APPENDIX A
These slide-type cassettes are available for purchase from the National Audiovisual Center.
Write to:
Sales Branch
National Audiovisual Center (GSA)
Washington, D.C. 20409
Phone: 1-800-788-6282
Films
OM0080 59
Books, Pamphlets & Magazines
NIH-03-112c, Specification for a Class I, Type A: National Institutes of Health; Public Health
Service; Bethesda, Maryland 20014
NSF Standard No. 49: National Sanitation Foundation; NSF Building; Ann Arbor, Michigan
48105
General Purpose Biological Safety Cabinet Class I, Type B; National Institutes of Health; Public
Health Service; Bethesda, Maryland 20014
National Cancer Institute Safety Standards for Research Involving Chemical Carcinogens DHEW
Publication No. (NIH) 77-900: National Institutes of Heath; Public Service; Bethesda, Maryland
20014
Guide for the Care and Use of Laboratory Animals DHEW Publication No. (NIH) 74-23:
National Institutes of Health; Public Health Service; Bethesda, Maryland 20014
Biohazard Control and Containment in Oncogenic Virus Research: DHEW Publication No.
(NIH) 73-459: National Institutes of Health; Public Health Service; Bethesda, Maryland 20014
Assessment of Risk of Human Infection in the Microbiological Lab: 2nd Ed., A.G. Wedum and
R.H. Kruse, Mis Pub-30; Department of the Army; Fort Detrick; Frederick, Maryland 21701
Design Criteria for Viral Oncology Research Facilities: DHEW Publication No. (NIH) 78-891
National Institutes of Health; Public Health Service; Bethesda, Maryland 20014
Classification for Etiologic Agents on the Basis for Hazard: Department of Health, Education and
Welfare; Public Health Service; Center for Disease Control; Atlanta, Georgia 30333
Safety and Operation Manual High Containment Research Facility: National Cancer Institute;
Public Health Service; Bethesda, Maryland 20014
Contaminant Dispersion and Dilution in a Ventilated Space: National Cancer Institute; National
Institutes of Health; Bethesda, Maryland 20014
Federal Register; Recombinant DNA Research; Vol. 41, No. 131: National Institutes of Health;
Public Health Service; Bethesda, Maryland 20014
OM0080 60
(BCD-05147)
(REV G)
OM0080 61
(BCD-05146)
(REV E)
OM0080 62
(BCD-05145)
(REV B)
OM0080 63
(BCD-05144)
(REV A)
OM0080 64
(BCD-05153)
(REV A)
OM0080 65
(BCD-05033)
(REV M)
OM0080 66