You are on page 1of 8

NOTICE: This standard has either been superseded and replaced by a new version or discontinued.

Contact ASTM International (www.astm.org) for the latest information.


Designation: E 1387 – 95
AMERICAN SOCIETY FOR TESTING AND MATERIALS
100 Barr Harbor Dr., West Conshohocken, PA 19428
Reprinted from the Annual Book of ASTM Standards. Copyright ASTM

Standard Test Method for


Ignitable Liquid Residues in Extracts from Fire Debris
Samples by Gas Chromatography1
This standard is issued under the fixed designation E 1387; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope 3. Summary of Test Method


1.1 This test method covers the identification of residues of 3.1 The sample extract or preparation is introduced into the
ignitable liquids in extracts from fire debris samples. Extrac- gas chromatographic column containing a liquid phase suitable
tion procedures are described in the referenced documents. for the separation of common ignitable liquid components. The
1.2 The values stated in SI units are to be regarded as the resulting chromatogram is interpreted by techniques of pattern
standard. The values given in parentheses are for information recognition and pattern comparison described in this test
only. method. Ignitable liquids may fall into one of five major
1.3 This standard does not purport to address all of the classifications, or into a “miscellaneous” category described
safety concerns, if any, associated with its use. It is the herein.
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica- 4. Significance and Use
bility of regulatory limitations prior to use. For a specific 4.1 The identification of an ignitable liquid residue in a
precautionary statement, see 6.3. sample of fire debris can support a field investigator’s opinion
regarding the incendiary nature of a fire.
2. Referenced Documents 4.1.1 The identification of an ignitable liquid residue in a
2.1 ASTM Standards: fire scene does not necessarily lead to the conclusion that a fire
E 260 Practice for Packed Column Gas Chromatography2 was incendiary in nature. Further investigation may reveal a
E 752 Practice for Safety and Health Requirements Relating legitimate reason for the presence of ignitable liquids.
to Occupational Exposure to Carbon Disulfide3 4.2 Due to the volatility of ignitable liquids and to variations
E 1385 Practice for Separation and Concentration of Ignit- in sampling technique, the absence of detectable quantities of
able Liquid Residues from Fire Debris Samples by Steam ignitable liquid residues does not necessarily lead to the
Distillation2 conclusion that ignitable liquids were not present at the fire
E 1386 Practice for Separation and Concentration of Ignit- scene.
able Liquid Residues from Fire Debris Samples by Solvent 4.3 When the gas chromatographic pattern is not sufficiently
Extraction2 complex, as described in 10.3, additional analytical techniques
E 1389 Practice for Cleanup of Fire Debris Sample Extracts are required.
by Acid Stripping2
E 1412 Practice for Separation and Concentration of Ignit- 5. Apparatus
able Liquid Residues from Fire Debris Samples by Passive 5.1 Gas Chromatograph—A chromatograph equipped with
Headspace Concentration2 a flame ionization or mass spectral detector may be used. Other
E 1413 Practice for Separation and Concentration of Ignit- detectors may be used if it can be shown that they have
able Liquid Residues from Fire Debris Samples by Dy- sensitivity and selectivity equal to the above detectors.
namic Headspace Concentration2 5.1.1 Sensitivity—The system shall be capable of detecting
E 1618 Guide for Ignitable Liquid Residues in Extracts 0.05 volume % in carbon disulfide (or any appropriate solvent)
from Fire Debris Samples by Gas Chromatography-Mass of any common ignitable mixtures.
Spectrometry2 5.1.2 Sample Inlet System—A sample inlet system that
allows a reproducible volume of liquid to be injected.
5.1.3 Column—A non-polar capillary column is recom-
1
This test method is under the jurisdiction of ASTM Committee E-30 on mended, but any column may be used provided that, under the
Forensic Sciences and is the direct responsibility of Subcommittee E30.01 on
Criminalistics.
conditions of use, the test mixture can be resolved into its
Current edition approved Oct. 10, 1995. Published December 1995. Originally component peaks.
published as E 1387 – 90. Last previous edition E 1387 – 90. 5.1.3.1 The test mixture shall consist of equal parts by
2
Annual Book of ASTM Standards, Vol 14.02. weight of the even-numbered normal alkanes ranging from
3
Annual Book of ASTM Standards, Vol 11.03.

1
E 1387
n-hexane through n-eicosane, plus the following aromatic change composition. Extracts in carbon disulfide may be
components: toluene, p-xylene, o-ethyltoluene, covered with water prior to removing the extracts from the
m-ethyltoluene, and 1,2,4-trimethylbenzene. sample preparation hood. Alternatively, septum seal vials may
5.1.3.2 If total resolution of the test mixture cannot be be used for storing any solvents or extracts.
obtained on a single column or program, changing columns or 7.2.1 If water is used as a sealant, exercise care to avoid the
programs is permitted as long as using both columns or introduction of water onto DMCS treated columns.
programs results in the complete resolution of the test mixture. 7.2.2 Avoid the use of water as a sealant if the presence of
5.2 Column Oven—A column oven capable of reproducible water soluble compounds is suspected.
temperature and temperature program settings in the range of
50 to 300°C should be used. 8. Calibration
5.3 Strip Chart Recorder—A recording potentiometer with 8.1 Calibrate the chromatographic instrument frequently
a full scale deflection of 10 mV or less should be used. The full using standards of known concentrations of known ignitable
scale response of the recorder should not exceed 1 s. An liquids as well as blanks. Optimize gas flows periodically.
integrator, or computerized data station and printer which Refer to Practice E 260 for detailed instructions on operation of
meets or exceeds these requirements, is acceptable. the gas chromatograph.
5.4 Syringes: 8.1.1 Run appropriate blanks and controls periodically.
5.4.1 For liquid samples—a microsyringe, capable of repro- 8.1.1.1 Clean syringes thoroughly between injections.
ducibly introducing sample sizes in the range of 0.1 to 10.0 8.1.2 Run known standards as necessary.
mL. 8.1.2.1 Standard chromatograms must be run under the
5.4.2 For gas samples—a gas-tight syringe capable of same chromatographic conditions as those used to produce the
reproducibly introducing sample sizes in the range of 0.5 to 5 sample chromatogram.
mL. 8.1.2.2 Every case file that includes a positive identification
of an ignitable liquid or residue must include the standard
6. Reagents and Materials chromatogram used to confirm the identification.
6.1 Purity of Reagents—Reagent grade chemicals shall be 8.2 Chromatogram Evaluation—A good chromatogram for
used in all tests. Unless otherwise indicated, it is intended that comparison work is one in which the peaks of interest are 50
all reagents conform to the specifications of the Committee on to 100 % of full scale. Rerun samples, or replot chromatogram,
Analytical Reagents of the American Chemical Society where using different parameters (attenuation or sample size) to
such specifications are available.4 Other grades may be used, achieve a good chromatogram.
provided it is first ascertained that the reagent is of sufficiently 8.2.1 In addition to the chromatogram described above, it is
high purity to permit its use without lessening the accuracy of sometimes necessary to produce other, off-scale plots, in order
the determination. to bring some features into view for comparison. Such off-scale
6.2 Solvents—The solvent in which the extract is dissolved plots may be required when there are one or more components
will depend upon the extraction technique employed. Fre- present at a significantly higher concentration than the other
quently check solvents for purity by running appropriate components in the residue.
blanks, both neat and evaporated, to at least twice the extent
used in the analysis. 9. Petroleum Distillate Classification System
6.2.1 If Carbon disulfide is used, read and follow the safety 9.1 Six major classes of complex liquid products are recog-
precautions described in Practice E 752. nized as usually identifiable by GC patterning alone when
6.3 Carrier Gas—Caution: hydrogen, helium, and nitrogen recovered from fire debris.
are compressed under high pressure, and hydrogen is an 9.1.1 This test method is only intended to allow isolated
extremely flammable gas. residues to be characterized as to one of the types of products
6.4 Combustion Gases—Air and hydrogen (if a flame ion- listed in Table 1. Other characterizations of samples may be
ization detector is used). possible, but are not within the scope of this test method.
6.5 Activated Charcoal.
10. Procedure
7. Sample Handling 10.1 Obtain a chromatogram of the fire debris sample
7.1 Methods of obtaining the extracts or preparations for extract.
analysis are described in Practices E 1385, E 1386, E 1389, 10.2 Obtain a chromatogram of a matching, or nearly
E 1412, and E 1413. matching, standard, and compare the pattern of peaks visually.
7.2 Due to the volatility of the solvents and the analytes, 10.2.1 The essential requirement for making a classification
take care to ensure that samples do not evaporate or otherwise using this procedure is the matching of the sample chromato-
gram with a known standard chromatogram obtained under
similar conditions, noting sufficient significant points of corre-
4
Reagent Chemicals, American Chemical Society Specifications, American lation or similarities. Make all comparisons using only good
Chemical Society, Washington, DC. For suggestions on the testing of reagents not chromatograms, as described in 8.2.
listed by the American Chemical Society, see Analar Standards for Laboratory
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
10.2.2 Pattern matching requires that the entire pattern used
and National Formulary, U.S. Pharmaceutical Convention, Inc. (USPC), Rockville, for comparison be displayed at the same sensitivity.
MD. 10.2.2.1 To provide sufficient detail for some comparisons,

2
E 1387
TABLE 1 Ignitable Liquid Classification System
“Peak Spread” Based on
Class Number (Class Name) N-Alkane Carbon Numbers Examples
(Unevaporated Liquid)
1 C4–C11 Petroleum ethers, Pocket lighter fuels, Some rubber cement solvents, Skelly solvents, V M &
Light Petroleum Distillates (LPD) P Naphtha, Some camping fuels
2 C4–C12 All brands and grades of automotive gasoline, including gasohol.
Gasoline
3 C8–C12 Mineral spirits, Some paint thinners, Some charcoal starters, “Dry-cleaning” solvents, Some
Medium Petroleum Distillates (MPD) torch fuels, Some solvents for insecticides and polishes, some lamp oils
4 C9–C17 Number 1 Fuel Oil, Jet-A (aviation) fuel, Insect sprays, Some charcoal starters, Some torch
Kerosene fuels, Some paint thinners, Some solvents for insecticides and polishes, some lamp oils
5 C9–C23 Number 2 fuel oil, Diesel fuel
Heavy Petroleum Distillates (HPD)
0
Miscellaneous Variable Single compounds, Turpentines, Specialty mixtures that cannot be further classified into one
of the categories below
0.1
Oxygenated solvents Variable Alcohols, Esters, Ketones
0.2
Isoparaffins Variable Isoparaffin products, Some charcoal starters, Some copier fluids, Some aviation gasolines,
Some lamp oils, Some solvents for insecticides and polishes, Some camping fuels
0.3
Normal alkanes Variable Specialty products formulated from normal alkanes, Some lamp oils, Some solvents for
insecticides and polishes
0.4
Aromatic solvents Variable Light, medium and heavy “aromatic naphtha” used as solvents for paints and plastics
0.5
Naphthenic/paraffinic solvents Variable Specialty solvent/fuel products made from Class 3 or Class 4 distillates-treated to remove
normal alkanes and aromatics, with higher cycloalkane content than isoparaffin products

different amplitudes or presentations of the data may be 10.3 Use the following criteria to determine whether suffi-
necessary. cient similarities exist between the sample and the standard to
10.2.2.2 The carbon number range is determined by com- classify the residue:
paring the chromatogram to a standard containing known 10.3.1 . Class 1—At least four major peaks matching a
normal alkanes, as shown in Fig. 1. known Class 1 standard are in the C4 to C11 range. No major
10.2.3 Store the standard chromatogram(s) in the case file, peak beyond C11 is associated with ignitable liquids. Examples
along with the sample chromatogram(s). of Class 1 liquids are shown in Fig. 2 and Fig. 3.

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 1 C6–C20 Normal Alkanes

3
E 1387

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 2 Example of a Class 1 Pattern; Brand A Cigarette Lighter Fluid

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 3 Example of a Class 1 Pattern; Brand B Cigarette Lighter Fluid, Evaporated 90 %

10.3.2 Class 2—The C3 alkylbenzene four-peak group, evaporation or combustion. Other peak groupings characteris-
which represents m- and p-ethyltoluene, 1,3,5- tic of gasoline must also be present. Examples of Class 2
trimethylbenzene, o-ethyltoluene, and 1,2,4-trimethylbenzene liquids are shown in and Fig. 4. A time-expanded view of the
(pseudocumene), must be present; this group occupies the C3 alkylbenzene four-peak group is shown in Fig. 5.
range between C9 and C10 and is still present in gasolines that 10.3.2.1 The mere presence of these alkylbenzenes does not
have lost as must as 90 % of their initial (“fresh”) weight by justify an identification of gasoline. These compounds must be

4
E 1387

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 4 Example of a Class 2 Pattern; Gasoline, Evaporated 90 %: (3) C3 Alkylbenzenes, (4) C4 Alkylbenzenes, (5) Naphthalene, and (6) 2-
and 1-Methylnaphthalene

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 5 Time Expanded View of C3 Alkylbenzene Region of Gasoline: (1) m-Ethyltoluene, (2) 1,3,5-rimethylbenzene, (3) o-Ethyltoluene, (4)
1,2,4-Trimethylbenzene, (*) Unresolved p-Ethyltoluene, and ( + ) Branched C10 Alkane

present at approximately the same relative concentrations as 10.3.2.2 Benzene, toluene, ethylbenzene, xylenes, cumenes,
are observed in samples of known gasoline. Many carpet and naphthalenes, which are present in gasoline, are also
samples that have been exposed to fire conditions contain these sometimes found in fire debris samples containing no foreign
compounds in some concentration. ignitable liquid residues.

5
E 1387
10.3.2.3 Class 1 and Class 2 patterns may change signifi- industrial products which are ignitable but which fall into more
cantly as the residue becomes more evaporated. than one category or do not fall into any of the above
10.3.3 Class 3—The pattern starts near C8 and ends near categories, other than “miscellaneous.” Many of these are
C12 and contains at least two significant N-alkane peaks synthetic mixtures consisting of only a few compounds, rather
between C8 and C12 . An example of a Class 3 pattern is shown than distillation fractions, and require multiple column analysis
in Fig. 6. in order to achieve an identification. Other techniques such as
10.3.4 Class 4—The pattern starts near C8. At least five GC/MS, Infrared Spectroscopy, or Ultraviolet Spectroscopy
consecutive N-alkane peaks between C12 and C17 must be may be indicated when the gas chromatographic pattern is not
present. An example of a Class 4 pattern is shown in Fig. 7. sufficiently complex to identify a mixture of components.
10.3.5 Class 5—The pattern starts near C9. At least five 10.5 Extraneous Components—It is recognized that the
consecutive N-alkane peaks between C14 and C20 must be matrix from which the sample is extracted may contribute
present. An example of a Class 5 pattern is shown in Fig. 8. extraneous components to an extract. These components may
10.3.5.1 In Class 3, 4, and 5 patterns, N-alkanes are typi- cause some difficulty in pattern interpretation, but the presence
cally the tallest peaks and those peaks frequently describe a of extraneous components does not invalidate the proper
roughly bell-shaped curve. classification of a sample.
10.3.5.2 Class 3, 4, and 5 patterns tend to shift upward in 10.5.1 Extracts that meet the criteria for Class 4 or Class 5
carbon number range with evaporation, but the overall appear- should be reviewed carefully for “extraneous components” that
ance of the pattern does not usually change as a result of elute near N-alkanes and are the result of polyolefin decompo-
evaporation. sition. Peaks representing the corresponding 1-alkene or 1,
10.3.5.3 It is recognized that a distinction between Class 4 (N-1) diene, and having a concentration near the concentration
and Class 5 may be difficult or impossible due to fire exposure (within 1⁄2 an order of magnitude) of the alkane, should be
and fractionation during sample preparation. A category that considered as indicating the presence of polyolefin products
includes both classes may be referred to as “fuel oil” as long as rather than fuel oil products. Polyolefin decomposition prod-
five consecutive N-alkane peaks can be identified. ucts typically do not exhibit the same pattern of branched
10.3.5.4 When N-heptadecane (C17 H36) and N-octadecane alkanes as fuel oils.
(C18 H38) are identified, observe the pattern to determine 10.6 Missing Components—The experience of the ignitable
whether pristane and phytane are also present. Fuel oils that liquid during a fire may result in the loss of lighter components,
contain C17 and C18 will usually also contain identifiable and the experience of the residue during the separation process
quantities of these two branched alkanes. Fuel oils also contain may result in the loss of heavier components. The selective loss
significant quantities of other branched alkanes that exhibit a of normal alkanes and the lower aromatics from soil samples
characteristic pattern. by microbial degradation has been documented, but it has not
10.4 No classification system is likely to describe all pos- been reported in other types of substrate materials. Once a
sible ignitable liquids. There are numerous commercial and pattern starts, it should match a standard until the pattern ends.

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 6 Example of a Class 3 Pattern; 50 % Evaporated Mineral Spirits

6
E 1387

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 7 Example of a Class 4 Pattern; 50 % Evaporated Kerosene: (1) Pristane, and (2) Phytane

Chromatographic Conditions
Instrument: HP 5890 Gas Chromatograph Initial Temperature: 60°C, 6 min
Detector: HP 5971 Mass Selective Detector Ramp: 20°C/min
Column: 25 m, HP-1, 0.2 mm, ID, 0.5 µm film thickness Final Temperature: 280°C, 4 min
All injections are 1 µL of a 1 % solution in diethyl ether.

NOTE 1—These are typical chromatograms, provided for illustration only. Each laboratory is likely to obtain slightly different retention times and
relative abundances of individual components, depending on sample history, separation procedure, and chromatographic conditions.
FIG. 8 Example of a Class 5 Pattern; 50 % Evaporated Diesel Fuel

Peaks missing from the middle of a pattern are usually those cases, the use of GC/MS as described in Guide E 1618 is
sufficient grounds for concluding that there is not a match recommended.
between the sample and the standard.
10.7 Diffıcult Patterns—Patterns will occasionally be en- 11. Report
countered that make a distinction between an ignitable liquid 11.1 Report the following information:
and a background material such as carpet very difficult. In 11.1.1 Laboratory case number,

7
E 1387
11.1.2 Submitting agency’s name, address, and case number techniques were used and which analytical techniques were
or other identifier, used.
11.1.3 Date(s) of sample delivery, 11.4 The conclusion should give the scientist’s opinion as to
11.1.4 Name of the person(s) making the delivery, whether an ignitable liquid was identified in the sample. If a
11.1.5 Name of the person(s) receiving the sample, negative result was obtained, a disclaimer to the effect that
11.1.6 Type of examination requested, negative results do not preclude the possibility that ignitable
11.1.7 Name of the analyst, liquids were present at the fire scene may help to avoid
11.1.8 Itemized list describing the submitted samples, and misunderstanding by readers of the report.
11.1.9 Results of the laboratory examination. 11.5 Certain words should not appear without explanation
11.2 A conclusion, summarizing the results in terms under- within the report. All extracts from organic materials are likely
standable to a lay person, may be added to the report. to contain“ hydrocarbons.” The word “hydrocarbon” should
11.2.1 The description of the evidence would seem to be not appear in a report unless those hydrocarbons can be
merely a clerical matter, but it is important that the analyst be specifically identified and classified. The phrase “hydrocarbons
sure that the evidence is described accurately, not simply as it from an unknown source” is expressly prohibited. Similarly,
was identified by the submitting agent. Fire debris samples, words such as “consistent with,”“ in the boiling range of,”
especially, tend to appear similar from the outside. It is possible “similar to,” or “characteristic of” a particular ignitable liquid
for samples coming from different locations within a fire scene, should not be used unless that liquid has been positively
or even from different fire scenes, to be confused with each identified using the methods described in Section 10.
other.
11.2.1.1 While it may not be possible for the analyst to 12. Precision and Bias
distinguish by visual inspection the difference between carpet-
12.1 Precision—It is not practicable to specify the precision
ing from the living room and carpeting from the hallway, it is
of this test method because of the unavailability of interlabo-
possible to determine by visual inspection the difference
ratory test data.
between bedding from the master bedroom and carpeting from
the hallway or concrete from the basement. The information 12.2 Bias—Because the test method yields a qualitative
which the analyst puts into the report should be verified by the identification, there is no bias.
analyst to the extent possible. An analyst’s first hand observa-
tions, and information supplied by a submitter when a sample 13. Keywords
is delivered, should be easily distinguishable. 13.1 fire debris samples; forensic sciences; gas chromatog-
11.3 The results section should state which preparation raphy; ignitable liquid residues

ANNEX

(Mandatory Information)

A1. SAMPLE STORAGE

A1.1 Store the original sample after extraction using Long term stability can be obtained by adding activated
appropriate procedures for handling and documentation. charcoal to the solvent containing the extract and allowing the
A1.1.1 Extract Storage-Short Term solvent to slowly evaporate. The sample can later be reconsti-
Extracts may be stored in a refrigerator in a stoppered tube tuted by addition of the solvent. A charcoal adsorption package,
to prevent evaporation. such as a C-strip, may serve the same purpose as added
A1.1.2 Extract Storage-Long Term charcoal.

The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.

You might also like