You are on page 1of 112

 

 
W‐PD60 
 
 
USER’S MANUAL 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
Version 6.0 
 
 
 
PREDICTIVE DIAGNOSTICS DIVISION 
 
 

 
W‐PD60  user’s  manual  Ver6.0 

Table of Contents 
 
FOREWORD ........................................................................................................................................................... 6
1 DESCRIPTION...................................................................................................................................................... 7
1.1 GENERAL ........................................................................................................................................................ 7
1.2 PRINCIPLES OF OPERATION ................................................................................................................................ 8
1.3 SPECIFICATION ............................................................................................................................................... 10
1.4 IMPORTANT FEATURES AND OPTIONS ................................................................................................................ 11
1.4.1 Data Structure and storage. ................................................................................................................11
1.4.2 Trend .................................................................................................................................................. 12
1.4.3 Alarms ................................................................................................................................................ 12
1.4.4 Continuous Watch Feature ................................................................................................................. 13
1.4.5 Phase Resolved Data and Phase Reference ........................................................................................ 13
1.4.6 Schedule ............................................................................................................................................. 14
1.4.7 Self‐test and Self‐calibration .............................................................................................................. 14
1.4.8 PDI (PD Intensity) and PPS (PD pulse repetition rate) Calculation ..................................................... 14
1.4.9 Auxiliary Inputs ................................................................................................................................... 15
1.4.10 Measurement Suspend ..................................................................................................................... 15
1.4.11 Communication Options and Device Address ................................................................................... 15
1.4.12 Software ........................................................................................................................................... 16
1.4.13 W‐PD60 configuration (settings) ...................................................................................................... 16
1.5 W‐PD60 MOUNTING OPTIONS AND SENSOR CONNECTION .................................................................................... 17
1.5.1 Panel mount W‐PD60 ......................................................................................................................... 17
1.5.2 Door mount W‐PD60 .......................................................................................................................... 22
1.6 SENSORS FOR W‐PD60 .................................................................................................................................. 24
2 APPLICATION ................................................................................................................................................... 25
2.1 GENERAL ...................................................................................................................................................... 25
2.1.1 Type and specifics of equipment to monitor ...................................................................................... 25
2.2 CONFIGURATION OF PD SENSORS ..................................................................................................................... 25
2.2.1 Configuration of Alarm Thresholds..................................................................................................... 26
2.2.2 Noise Aspects...................................................................................................................................... 26
2.2.3 Device Location and Environmental Issues......................................................................................... 26
2.2.4 Additional Parameters to Measure .................................................................................................... 27
2.2.5 Networking and Communicating with W‐PD60 ................................................................................. 27
2.2.6 Motor and Small Generator Applications ........................................................................................... 28
2.2.7 Large Generator Applications ............................................................................................................. 29
2.2.8 MV Switchgear Applications............................................................................................................... 29
3 ON‐SITE INSTALLATION ................................................................................................................................... 30
3.1 W‐PD60 KIT CONTENTS ................................................................................................................................. 30
3.2 IMPORTANT NOTE ON SENSOR INSTALLATION...................................................................................................... 31
3.3 STEP 1 ‐ HARDWARE INSTALLATION ................................................................................................................... 32
3.4 STEP 2 ‐ SENSOR AND POWER CONNECTIONS....................................................................................................... 33
3.5 STEP 3 ‐ CONNECTIONS TO THE DEVICE............................................................................................................... 33
3.6 STEP 4 ‐ DEVICE INSTALLATION AND FINAL CONNECTIONS ..................................................................................... 33
4 STARTUP........................................................................................................................................................... 35

Page 3 of 112 
W-PD60 user’s manual Ver6.0

4.1 GENERAL SETTING WITH KEYPAD....................................................................................................................... 35
4.2 QUICK START UP ............................................................................................................................................ 35
4.3 COMMUNICATIONS BOARD SETUP..................................................................................................................... 37
4.3.1 USB ..................................................................................................................................................... 38
4.3.2 Ethernet .............................................................................................................................................. 38
5 TROUBLESHOOTING ........................................................................................................................................ 42
5.1 TROUBLESHOOTING TIPS ................................................................................................................................. 42
5.2 ERROR MESSAGES .......................................................................................................................................... 44
5.3 FUSE REPLACEMENT ....................................................................................................................................... 45
6 APPENDIX I W‐PD60 FRONT PANEL KEYPAD OPERATION .............................................................................. 47
6.1 KEYPAD ........................................................................................................................................................ 47
6.2 SETUP VIA KEYPAD.......................................................................................................................................... 47
6.2.1 Setting Date ........................................................................................................................................ 47
6.2.2 Setting Time........................................................................................................................................ 48
6.2.3 Set Active Channels ............................................................................................................................ 49
6.2.4 Set Display Parameters....................................................................................................................... 49
6.2.5 Set Sensor Sensitivity .......................................................................................................................... 53
6.2.6 Set Measurement Schedule ................................................................................................................ 53
6.2.7 Set Data Saving Mode ........................................................................................................................ 55
6.2.8 Set Pulse Width and Interval .............................................................................................................. 56
6.2.9 Set Alarm and Trend Thresholds......................................................................................................... 57
6.2.10 Set Rated Voltage ............................................................................................................................. 59
6.2.11 Set Rated Current ............................................................................................................................. 59
6.2.12 Set Number of Power Cycles for Data Acquisition ............................................................................ 60
6.2.13 Set Alarm Relay Mode ...................................................................................................................... 60
6.2.14 Set Minimum Noise Filter Level ........................................................................................................ 61
6.2.15 Set Display Scroll Time ...................................................................................................................... 62
6.2.16 Set ModBus Address ......................................................................................................................... 62
6.2.17 Set Continuous Watch Options ......................................................................................................... 63
6.2.18 Set Baud Rate ................................................................................................................................... 63
6.2.19 Select PDI or Qmax ........................................................................................................................... 64
6.2.20 Start/Stop W‐PD60 ........................................................................................................................... 64
6.2.21 Set Phase Shift .................................................................................................................................. 65
6.2.22 Set Channel Phase ............................................................................................................................ 66
6.2.23 Set Frequency Range ........................................................................................................................ 66
6.2.24 Set Suspend for Temperature ........................................................................................................... 67
6.2.25 Set Suspend for Voltage ................................................................................................................... 67
6.2.26 Set Suspend for Current .................................................................................................................... 68
6.2.27 Set Trend Window Length ................................................................................................................ 69
6.2.28 Calibrate Temperature ..................................................................................................................... 69
6.2.29 Calibrate Voltage .............................................................................................................................. 71
6.2.30 Calibrate Current .............................................................................................................................. 72
6.2.31 Calibrate 4‐20mA Loop ..................................................................................................................... 73
6.2.32 Calibrate Humidity Sensor ................................................................................................................ 75
6.2.33 Set Noise Filter Offsets ..................................................................................................................... 76
6.2.34 Set Channel Low Magnitude Limit .................................................................................................... 77
6.2.35 Set Software Filters........................................................................................................................... 79
6.2.36 Set ModBus Protocol ........................................................................................................................ 79

Page 4 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.37 Start a Single Measurement ............................................................................................................. 80
6.3 ACCESSING MEMORY VIA KEYPAD ..................................................................................................................... 80
6.3.1 Clear All Memory ................................................................................................................................ 80
6.3.2 Check Data Saved in Memory ............................................................................................................. 81
7 APPENDIX II SYNCHRONIZATION AND PHASE REFERENCE SHIFT .................................................................. 83
7.1 SYNCHRONIZATION ......................................................................................................................................... 83
7.2 ENTERING TYPICAL PHASE REFERENCE SHIFT FOR EXTERNAL SYNCHRONIZATION ........................................................ 84
7.3 DETERMINING AND ENTERING PHASE REFERENCE SHIFT ........................................................................................ 85
8 APPENDIX III COMMUNICATIONS BOARD INTERFACE ................................................................................... 88
8.1 GENERAL ...................................................................................................................................................... 88
8.2 NETWORK CONFIGURATION ............................................................................................................................. 88
9 APPENDIX IV USB DEVICE DRIVER INSTALLATION .......................................................................................... 91
10 APPENDIX V ETHERNET VIRTUAL COMMUNICATIONS PORT DRIVER INSTALLATION................................. 98
10.1 INSTALLING COM0COM NULL MODEM EMULATOR.............................................................................................. 98
10.2 CONFIGURING COM0COM NULL MODEM EMULATOR ....................................................................................... 104
10.3 CONFIGURING COM0COM (USING COMMAND LINE UTILITY) ............................................................................... 106
10.4 COM PORT REDIRECTOR INSTALLATION ........................................................................................................... 107
10.5 SETTING ENVIRONMENT VARIABLES ................................................................................................................111
10.6 RUNNING COM PORT REDIRECTOR APPLICATION .............................................................................................. 112
10.7 TROUBLESHOOTING .................................................................................................................................... 112

Page 5 of 112 
W-PD60 user’s manual Ver6.0

Foreword 
The purpose of this manual is to describe the functionality of the new generation of W‐PD60. New 
features include a graphic dot matrix display and more communication options embedded, such as USB 
port, Ethernet port over network, webpage and FTP server. It serves as primary guideline for qualified 
personnel involved in the onsite installation, start‐up, commissioning and continuous monitoring. 

The whole document is organized in the following way: 

Section 1 Description introduces the basic functions and operations of W‐PD60 

Section 2 Application presents the general considerations for W‐PD60 applications, especially 
configuration of different sensors for different applications. 

Section 3 On‐Site Installation shows the procedures for onsite installation of W‐PD60 systems. 

Section 4 Startup presents the procedures for initial setup using the front panel keypad without a 
computer upon completion of W‐PD60 installation. 

Section 5 Troubleshooting provides troubleshooting suggestions in different scenarios. 

Section 6 Appendix I W‐PD60 Front Panel Keypad Operation give details about how to operate using the 
front panel keypad and display on the W‐PD60. 

Section 7 Appendix II Synchronization and Phase Reference Shift shows the procedures to make proper 
synchronization and measure phase shift between W‐PD60 and monitored object. 

The following three sections should be ignored if W‐PD60 does not include the optional communication 
board. 

Section 8 Appendix III Communications Board Interface presents the features available through the 
optional communications board. 

Section 9 Appendix IV USB Device Driver Installation gives exact steps to install the driver for 
communications between a computer and W‐PD60 with a USB cable through the USB port on the 
communication board. 

Section 10 Appendix V Ethernet Virtual Communications Port Driver Installation gives exact steps to 
install the driver and configure the interface to use the Ethernet port on the communications board as a 
serial port to communicate between a computer and W‐PD60 online. 

 
 
 
 
 
 

Page 6 of 112 
W‐PD60  user’s  manual  Ver6.0 

1 Description 

1.1 General 
W‐PD60 is a standalone microprocessor‐controlled continuous partial discharges monitoring device for 
use with a wide range of medium voltage power equipment. It works with 50/60 Hz frequency powered 
equipment as well as with variable frequency applications. It can even work at 5Hz or 400Hz application. 

Partial discharges (PD) are the small sparks that commonly occur within or on surface of insulation in 
various types of medium and high voltage equipment. Over time, PD destroys the insulation, eventually 
causing insulation breakdown (full discharge). In general, it takes months or even years before PD will 
cause a failure of insulation. Therefore, PD is an important indicator of insulation degradation for prompt 
maintenance or repair. 

W‐PD60 is designed to provide an alarm based on PD characteristics at an early stage of insulation 
degradation. It measures PD from up to 15 high frequency PD sensors, stores the information in internal 
memory and provides various alarm interfaces to the users if any set points are exceeded. Depending on 
the application various PD sensors can be used. 

W‐PD60 also has three auxiliary inputs for measuring auxiliary parameters of the monitoring equipment: 
temperature, load and humidity or voltage, depending on the application. Those parameters will be 
correlated with PD signals for predictive diagnostics.   

W‐PD60 includes a main unit and a Sensor Interface Board connected by flat ribbon cable (Figure 1‐1). All 
sensors have to be connected to the Sensor Interface Board. 

Five LEDs on the front panel of the main unit indicate Alarm status (green LED for Normal, yellow LED for 
Warning (Alarm 1) and red LED for Alarm (Alarm 2)), Setup and Memory operation modes. 

A graphic dot matix display shows measurement data, setup or memory content and error messages. 
During normal operation the display continuously scrolls down the last measured data, auxiliary 
parameters and other information. 

With the front panel keypad and display, users can operate W‐PD60 directly without a computer. The 
membrane keypad has four navigation ARROW keys (←→↑↓), two function keys (Esc and Enter) and 
two specific function keys (Set and Memory). 

W‐PD60 has several interfaces (Figure 1‐1) to be easily implemented into any alarm and SCADA system. 
These options include three C‐form relays for Warning (Alarm 1), Alarm (Alarm 2) and Device Status 
indication, 4‐20mA output, RS485 and two USB interfaces and an Ethernet interface which enable 
convenient online communication, webpage display and FTP server for data downloading. 

Page 7 of 112 
W-PD60 user’s manual Ver6.0

 
 

 
 
Figure 1‐1 a ‐ Top: Sensor Interface Board. Bottom: b – Side view of the W‐PD60. 
 

1.2 Principles of Operation 
W‐PD60 has 15 signal input channels, channels1 ~15, for partial discharge measurement and a noise 
input channel dedicated for noise suppression (channel 16). All sixteen input channels have identical 
conditioning circuits (CC) that provide signal isolation, transient suppression, and high‐pass filtering of 
the input signals. The frequency band of the PD signals for W‐PD60 is from 1 MHz to 20 MHz. 

For each channel W‐PD60 acquires PD data in the form of three dimensional Phase‐Resolved Pulse 
Height Distribution (PRPHD) ‐ PD pulse count as a function of pulse magnitude relative to phase of 

Page 8 of 112 
W‐PD60  user’s  manual  Ver6.0 

monitored device voltage (60/50 Hz or voltage phase of VFD ‐ Variable Frequency Drive). The data is 
sorted into 24 phase windows (150 each) and 21 magnitude windows in the range of about 70dB. 

Using this data W‐PD60 calculates integral PD parameters: Partial Discharge Intensity (PDI) or Maximum 
Pulse magnitude (Qmax), Pulse Repetition rate (pps), and Trend of the PD activity. 

The data is stored in the internal device memory in the form of integral parameters for both brief and 
full mode measurement as well as in the form of three‐dimensional PRPHD matrixes for full mode 
measurement. Each record is accompanied by three auxiliary operating parameters, i.e. temperature, 
humidity, load current or voltage. 

Before each measurement, W‐PD60 performs a self‐calibration and self‐test. If a problem is detected, its 
Status relay dry contacts open and an appropriate message appear on the W‐PD60 display. Loss of power 
is indicated in the same manner, via the opening of status relay contacts. 

W‐PD60 acquires signals from the measurement channels sequentially multiplexing them to a single 
metering channel. Each pulse from each sensor is validated by the allowed pulse width. Noncompliant 
pulses will be discarded. 

After each measurement, PDI, Qmax and Trend levels will be compared to the preset Warning and Alarm 
thresholds for each active channel. If the Warning threshold limits are exceeded, the Warning yellow LED 
will be turned on and the Warning (Alarm 1) relay dry contacts will close. If the Alarm level is exceeded, 
W‐PD60 will trigger an additional measurement and, if confirmed, a red LED Alarm will be turned on and 
the Alarm (Alarm 2) relay dry contacts will close. If Alarm is detected, full mode measurement data will 
be stored in the memory. Alarm events also will be stored in log file in device memory. 

PD measurements can be performed on a time schedule (up to 50 different times per day) or in specified 
time interval (from 1 minute to 23 hours 59 minutes) as long as the time interval is longer than the time 
it takes W‐PD60 to make one measurement. Four measurements per day are recommended. 

Between scheduled measurements, the “Continuous Watch” feature is enabled. All signal sensors are 
connected to a summation unit and then to a separate “Continuous Watch” channel. W‐PD60 
continuously searches for an appearance of high magnitude pulses and pulse series. Magnitude 
threshold and repetition rate in series are configurable. If five events of pulse series are detected 
between the scheduled measurements, W‐PD60 will trigger a full PD measurement and displays an alarm, 
if confirmed. 

To provide correct measurements W‐PD60 must be synchronized with the voltage frequency of the 
monitored equipment (see Section 7 Appendix II Synchronization and Phase Reference Shift) and must 
have correct settings for the type of equipment, rated voltage, and PD sensors used. All settings can be 
made using Setup mode from the front panel keypad and display (see Section 6 Appendix I W‐PD60 Front 
Panel Keypad Operation ) or more easily using W‐PD60 PC software supplied with W‐PD60 (see W‐PD60 
Software Manual). Normally W‐PD60 already configured based on information received from a customer 
at its order. 

Page 9 of 112 
W-PD60 user’s manual Ver6.0

1.3 Specification 
General:   
Applications  HV and MV equipment (motors including VFD, 
switchgear, generators, bus ducts, cable terminations, 
transformers, etc.) 
Mounting options  (1) In NEMA 4X enclosure.
(2) Panel Mount. 
(3) Door (Flush) mount.
Installation category  II
Pollution Degree  2
Temperature Range  ‐400C ‐ +850C (+700C for enclosure mount). 
UL Certified for 00C ‐ +700C ambient 
Relative humidity  0%RH – 90%RH
Maximum Altitude (m)  2000
Power Source  115V / 230V AC +/‐10%
60 / 50 Hz
Power Consumption of device (VA max)  15 VA
Display  Graphic Dot Matrix Display, 2 lines 
Keypad  4 arrows and
4 function keys
LEDs  W‐PD60 Status (Normal)
Warning (Alarm 1) 
Alarm (Alarm 2) 
Memory mode 
Setting mode
Approximate Size (Length x Width x Height) and   
Weight: 
Main Unit  9.2”x 7” x 2.5” (23.4cm x 17.8 cm x 6.4 cm) 
4.2 lb (1.9 kg)
Door Mounted option (Main Unit with Sensor  9.2”x 7” x 4” (23.4cm x 17.8 cm x 10.2 cm) 
InterfaceBoard)  4.6 lb (2.1 kg)
Panel Mounted option  14.8” x 12.9” x 4” (37.6 cm x 32.8 cm x 10.2 cm) 
10.3 lb (4.7 kg)
Enclosure (NEMA 4X) Mounted option  17.2” x 15.4” x 8.8”, (43.7 cm x 39.0 cm x 22.3 cm)
21.6 lb ( 9.8 kg)
PD Measurement:   
Number of PD Channels  15
PD Channel Dynamic Range 68dB
Frequency Bandwidth  1‐20MHz
Phase‐Resolved Pulse Height Distribution (PRPHD):
Number of Magnitude Windows (3.23 dB each)   
Number of Phase Windows (150 each)  21 
24
Power Frequency at a Monitored Equipment  3‐20Hz, 20‐400Hz
(including VFD) 
Synchronization type  Internal and External
Maximum Measured Pulse Repetition Rate  367,300 Pulses/second
Integral PD Parameters  Partial Discharge Intensity (PDI) 
Maximum Pulse Magnitude (Qmax) 
Trend of PDI / Qmax

Page 10 of 112 
W‐PD60  user’s  manual  Ver6.0 

Warning and Alarm  by PDI/Qmax
Number of Noise Filter Channels  1
Continuous Watch (High PD activity) Function Active between measurements 
Data Record Types  Full/Brief
Internal Data Memory  2 MB. Provides up to 1000 days of data storage at 4 
measurements per day 
Self‐test and Self‐calibration At powering up and before every measurement 
Setup  Configurable from keypad and PC 
Allowed RG‐58 Coaxial Cable Length to PD Sensors Up to 150ft (50m)
Auxiliary Inputs:   
Input specified for temperature measurement 1 
calibrated for 100 Ω Platinum RTD sensor. 
Analog Inputs specified for Current, Voltage or 2 
Humidity measurement 
Interfaces:   
C‐form Dry Type Relays for Device Status, Warning  3
(Alarm 1) and Alarm (Alarm 2) PD levels (fully  120VAC/ 5 Amps 
configurable).  28VDC/ 5 Amps 
For other ratings refer to the relay specification 
4–20 mA isolated interface represents Highest PDI or  Slope is 1mA per 10%
Max. Magnitude as % of the Alarm (Alarm 2) Threshold 
RS‐485 interface optically‐isolated (231 addresses). Communication Protocol: ModBus RTU, Binary, 
Additional Communication Options  USB for communication with PC 
Ethernet, Webpage and FTP
Regulatory and Standards Compliance  UL 61010‐1: Second Edition Standard for Electrical 
Equipment for Laboratory Use, with evisions dated 
2008‐10‐28 
  FCC Part 15, Subpart B, Class A 
  CAN/CSA C22.2 No. 61010‐1‐04, Second Edition, 
with revisions through 2008‐10‐01 
  IEC61010‐1:2001 and EN61010‐1:2001 
  EN61326‐1, EN61000‐3‐2, EN61000‐3‐3 
  IEC61000‐4‐2, IEC61000‐4‐3, IEC61000‐4‐4, 
IEC61000‐4‐5, IEC61000‐4‐6, IEC61000‐4‐8, 
IEC61000‐4‐11 
 

1.4 Important Features and Options 
1.4.1 Data Structure and storage. 

A full PD measurement by W‐PD60 involves a measurement of statistical Phase Resolved Partial 
Discharge Distribution (PRPDD) for every active channel. After each measurement for every active 
channel, W‐PD60 calculates integral PD parameters: Partial Discharge Intensity (PDI), maximum PD 
magnitude Qmax, Pulse Repetition rate (pps) and Trend of the PDI activity (rate of PDI change). The 
calculated parameters are compared to preset levels for Warning and Alarm. 

All calculated and alarm status parameters are stored in the internal memory for each measurement. 
Additionally, three auxiliary parameters (temperature, humidity, percent (%) of full load current or 
operating voltage) are stored and related to corresponding measurement data. 

Page 11 of 112 
W-PD60 user’s manual Ver6.0

Depending on what data will be stored in the internal memory, there are two modes of the W‐PD60 
operation: Test mode and Normal mode. 

Test Mode – during this mode “Full" data (PRPDD with the calculated integral parameters for each 
channel mentioned above) is stored in the memory for every measurement. Please note this is very 
memory‐consuming and stored data should be downloaded more often. Usually this mode is reserved 
for specialized testing. 

Normal Mode – in order to save memory, PRPDD can be stored in the memory only several times a 
month. An operator should set the interval in days and the measurement number in the current day for 
storing PRPDD. During the remaining measurements W‐PD60 will store the “Brief” version of PD data 
(Only integral parameters without PRPDD). This mode should typically be used for continuous 
monitoring. 

W‐PD60 has 2 MB of internal flash memory for data storage. When the memory is filled, the device starts 
replacing the oldest data with the latest data. The rate of the memory consumption depends upon the 
number of active channels, frequency of measurements, and frequency of PRPDD storage. As an 
example, if all 15 channels are active for measurements four times a day and PRPDD are stored twice a 
month, the device holds 17 months of the latest PD data in its internal memory. All stored data and 
settings can be accessed from the keypad or remotely from a PC. 

1.4.2 Trend 

W‐PD60 calculates trend of the parameter – PDI or Qmax whichever is enabled for alarm. Trend is 
calculated as a linear approximation of data over a specified time interval (default is 90 days). This 
specified time interval window is sliding over time while the device is monitoring PD. Trend is normalized 
to the value of the parameter change in times per year. Like PDI or Qmax levels, trend also has two 
alarming thresholds, Warning (Alarm 1) and Alarm (Alarm 2). When the Warning or Alarm level is 
exceeded, the corresponding LEDs on the front panel and relays will respond in the same manner as 
when PDI or Qmax level is exceeded. Details are described in the following section. 
 

1.4.3 Alarms 

W‐PD60 has two configurable alarm levels, Warning (Alarm 1) and Alarm (Alarm 2), each of which is 
connected to a C‐form relay. 

There are two groups of parameters that can generate an alarm. One is the preset level of PDI or Qmax 
for Warning and Alarm. The other is the preset trend level for PDI or Qmax. Only one of two, PDI or PD 
magnitude Qmax, can be configured for alarming at any time. 

The same parameter, PDI or Qmax is automatically configured for the 4‐20 mA interface output. The 4‐ 
20 mA output increases or decreases by 1 mA for every 10% increase or decrease of the ratio of the 
maximum measured PDI or Qmax among all active channels versus the preset alarm level. An output of 
4mA corresponds to 0% and 14mA corresponds to 100%. 

The alarm level for PDI is represented in terms of mW while the magnitude Qmax is represented in terms 

Page 12 of 112 
W‐PD60  user’s  manual  Ver6.0 

of mV. 

Trend is set in terms of times per year for both PDI and Magnitude. Alarm on trend is enabled after a 
training period of 1/3 of the trend‐sliding window. 

Warning and Alarm operate differently. A Warning‐level condition activates the corresponding relay and 
turns on the yellow Warning LED at the time the condition is detected. When an Alarm‐lcondition is first 
detected, W‐PD60 will initiate an additional measurement, and only if confirmed will it indicate the 
Alarm via relay and red Alarm LED.. If the Alarm conditions are not confirmed, the LED and Relay status 
will be set per the last measurement. If at any measurement the alarming status will be reduced, 
W‐PD60 will indicate the reduced alarming status with both LED and relay. 

Alarm (or Alarm 2) and Warning (or Alarm 1) relays can operate in two modes (configurable). Relays lock 
in the status received at the last measurements, until the next measurement. Or a relay can operate for 
a limited configurable time and then open the contacts. At the next measurement, if an alarm/warning 
status is detected, the relay will hold the contacts closed for the same time. 

1.4.4 Continuous Watch Feature 

Between the scheduled measurements the Continuous Watch feature (“High Alarm”) is initiated. At this 
time all signal sensors are summarized and connected to the separate High Alarm channel. W‐PD60 is 
continuously watching for the events of high magnitude pulses (See software manual for how to 
configure the settings) and their series. If five series of such events are detected, W‐PD60 starts PD 
measurement and, in the case of Alarm (Alarm 2) confirmed, W‐PD60 indicates it and stores full PD data. 

1.4.5 Phase Resolved Data and Phase Reference 

In order to provide phase‐resolved data, W‐PD60 must be synchronized with a signal that is synchronous 
with the voltage on the object being monitored and the phase shift between the voltage on monitored 
object and W‐PD60 power input voltage should be detected and input as a parameter. 

There are two options for synchronization: 

In most of cases the power source (115/230 VAC) powering up W‐PD60 is used as a synchronization 
signal. The device has an internal connection of the power circuit voltage to its synchronization circuitry. 
This option (default) always works on monitored objects powered from a regular power system (60/50Hz) 
in the area. Phase shift between the power voltage and the object high voltage should be determined 
during W‐PD60 startup and can be used at all times, while the same object is monitored and the same 
power circuit is used. W‐PD60 allows for automated detection of a phase shift between its power circuit 
and an external phase reference signal. (For more detailed information on synchronization and phase 
reference shift detection refer to Section 7 Appendix II Synchronization and Phase Reference Shift.) 

Be cautious that an W‐PD60 can be powered from an UPS only when the UPS has constant phase shift to 
the voltage in the monitored object in Standby mode. If not, W‐PD60 will be in an “asynchronous” state 
and can’t provide accurate phase‐resolved measurements.. 

Page 13 of 112 
W-PD60 user’s manual Ver6.0

External synchronization should be used if the W‐PD60 power supply voltage is not synchronous with the 
voltage on the test object, although it can also be used when W‐PD60 power supply is in synchronization 
with the monitored object voltage. An example of an asynchronous test object is a variable frequency 
drive. An external synchronization signal should be continuously provided to the special synchronization 
input and appropriate jumpers in the device should be configured. Low frequency component of a signal 
applied to the channel #1 (coupling capacitor option) may also be used for external synchronization. For 
more information on synchronization and device configuration refer to Section 7 Appendix II 
Synchronization and Phase Reference Shift 

1.4.6 Schedule 

PD measurements can be performed at specified times during a day or at specified time interval. The 
device is shipped with “time basis” schedule enabled and set to make measurements four times per day. 
This is sufficient for all common applications. If necessary, W‐PD60 can be set to measure up to 50 times 
per day at scheduled times or in specific time intervals varying from 1 minute to 23 hours 59 minutes as 
long as the time interval is shorter than the time of the measurement. 

The time T required for a measurement is T_ 21*NCh*NC / F, where NCh – number of active channels, NC – 
number of cycles per acquisition (usually 60), F – synchronization frequency (usually 50/60 Hz). Thus an 
W‐PD60 with all 15 channels active set for 60 cycles per acquisition and synchronized to 60 Hz would 
require 21*15*60/60=315 seconds, or 5 minutes, 15 seconds to complete a measurement. 

1.4.7 Self‐test and Self‐calibration 

W‐PD60 performs a self‐test and calibration of PD and noise filter channels when powering up as well as 
before each PD measurement. In the case of a fault, an error message will be displayed and the status 
relay contacts will open. Error status will be shown in software as well as on the webpage through the 
optional communications board. 

1.4.8 PDI (PD Intensity) and PPS (PD pulse repetition rate) Calculation 

PDI and pulse count (PPS ‐ pulse per second) are calculated for each channel at every measurement 
starting from some preset magnitude level (zone) and up. This feature allows for rejecting background 
noise from affecting PDI and pulse count calculations. The low magnitude threshold to calculate 
parameters can be set for each channel independently, based on its own background noise level, if any. 
There are options to set these values manually or automatically. Automatic low magnitude threshold 
detection can only be launched from the keypad (usually during device start up or during the initial 
period of operation). 

Page 14 of 112 
W‐PD60  user’s  manual  Ver6.0 

1.4.9 Auxiliary Inputs 

It is well known that operating conditions of monitored equipment can significantly affect PD behavior. 
Correlation between PD activity and operating conditions can be used as valuable diagnostic information 
for further diagnosis. In general, insulation temperature, mechanical vibration reflected by load current, 
voltage on the object, and humidity of the air are the most important parameters to correlate with PD 
characteristics. 

W‐PD60 has three auxiliary inputs. One of the inputs is strictly designated for temperature measurement. 
A 100 _ platinum RTD can be used as a sensor. See the temperature sensor manual for details. 

Two other inputs can accept any AC or DC signals within a specified voltage range. As factory default, 
they are configured for auxiliary sensors for load current and humidity measurement supplied with the 
device. 

1.4.10 Measurement Suspend 

Frequently measurements need to be suspended for some time. Measurements can be suspended in 
three different ways: 

Monitoring can be stopped. The “STOPPED” message will be blinking on W‐PD60 display and Red Alarm 
LED will be ON. The only way to resume the monitoring is through keypad operation or software 
communication by an operator. 

Monitoring can be paused. In this case, monitoring will resume automatically in three hours. 

Auto‐suspend mode can be controlled by any of three auxiliary parameters. The device allows the user 
to set low and high thresholds for these three auxiliary parameters and enable the auto‐suspend mode 
via the front panel display and keypad or the PC software (see software manual for details). The W‐PD60 
will stop monitoring when any enabled auxiliary parameter is outside the range of the low and high 
boundaries. 

The auto‐suspend mode is widely used to suspend measurements when necessary. For example, 
assuming the low threshold for load current is 5% of the nominal current, the W‐PD60 will stop making 
any measurement when the monitored equipment is out of service (load current =0). This keeps W‐PD60 
from collecting meaningless PD data which can lead to incorrect trend calculations. 

1.4.11 Communication Options and Device Address 

W‐PD60 has various interfaces that allow its easy implementation into any alarm or SCADA system as 
well as remote communications: 

  ‐Three C‐form relays dry contacts for Warning (Alarm 1), Alarm (Alarm 2) and Device Status 
indication. 

Page 15 of 112 
W-PD60 user’s manual Ver6.0

  ‐4‐20 mA isolated interface can be configured to represent PD Intensity (PDI) or Maximum PD 
magnitude (Qmax) to any SCADA system. 

  ‐RS‐485 optically isolated interface based on ModBus RTU protocol allows remote device 
configuration and data downloading. IG can be networked with an existing ModBus network, allowing 
for up to 231 addressable devices. 

W‐PD60 is shipped with the device address #1, unless otherwise specified. The address can be changed 
from the keypad and set to the value from 1 to 231. In any case, the address must be unique at each 
particular local device network. 

W‐PD60 is supplied with database software that allows automated communication with a device for data 
acquisition and analysis. The software allows either direct networking using a USB/Ethernet to RS485 
converter or dial up connection to a device by a regular telephone landline or a cellular connection using 
appropriate modems with RS485 interface. 

USB ports (only available with optional communications board): one USB (Device) for a direct link to a 
computer for communications with the W‐PD60 software or for communication board firmware 
upgrading, the other USB (Host) is reserved for future use. 

Ethernet port (only available with optional communications board): It enables direct communication 
between computer and W‐PD60, convenient online webpage and FTP server for data downloading or 
FTP client for data uploading. 

Telephone landline or cellular modems with RS485 interface are optional. 

1.4.12 Software 

W‐PD60 software allows for communicating with W‐PD60 through a PC on a local RS‐485 network or a 
modem. It also provides automated or manual data downloading and storage for several devices at 
different locations, device setting, data analysis, and presentation. (See W‐PD60 Software Manual) 

1.4.13 W‐PD60 configuration (settings) 

To provide proper PD monitoring W‐PD60 must be configured in accordance with the type of equipment 
to be monitored, its voltage, sensors used and preferable mode of monitoring. This configuration 
requires setting of many W‐PD60 parameters. 

W‐PD60 is typically supplied preconfigured for particular equipment in accordance with information 
received from the customer. 

There are two ways to make changes of the W‐PD60 configuration if it is necessary: 

  ‐From W‐PD60 keypad (see Section 6 Appendix I W‐PD60 Front Panel Keypad Operation ), 

Page 16 of 112 
W‐PD60  user’s  manual  Ver6.0 

  ‐From Software supplied with W‐PD60. The Software has some typical templates to help with the 
configuration (see W‐PD60 Software Manual). 

There are some settings which can only be configured through the keypad. Please refer to Section 4.1 
General Setting with Keypad for details. 

1.5 W‐PD60 mounting options and sensor connection 
W‐PD60 is normally shipped in one of the following optional configurations: 

  ‐In NEMA‐4X enclosure with or without transparent windows. 

  ‐On the panel for installation in a customer cabinet. This option is identical to the first one, except 
that the panel is not mounted into an enclosure. 

  ‐Door mount option. This option is used for indoor installation usually on a switchgear cabinet door. 
For Door mount W‐PD60 Sensor Interface Board is mounted on the W‐PD60 back cover. Also 
temperature and humidity sensors can be installed inside Sensor Interface Board. 

1.5.1 Panel mount W‐PD60 

An example of a panel mounted W‐PD60 and its wiring diagram for 115 VAC power voltage with optional 
landline modem is shown on Figure 1‐3 a, b and Figure 1‐4. For panel mounted W‐PD60 powered at 
230VAC, there is no power outlet. 

The W‐PD60 panel must be grounded at the place of its installation. 

 
a ‐ Left view. 
 

Page 17 of 112 
W-PD60 user’s manual Ver6.0

 
b ‐ Right view 
 
Figure 1‐3 Panel mount ed W‐PD60 

Page 18 of 112 
W‐PD60  user’s  manual  Ver6.0 

 
Figure 1‐4 A typical wiring diagram for panel mounted W‐PD60 with optional modem for 
230VAC power voltage. 

Page 19 of 112 
W-PD60 user’s manual Ver6.0

Note: The wiring diagram is just an example. For the W‐PD60 installation use the wiring 
diagram supplied with each W‐PD60.) 
 
All connections (excluding coaxial cables from PD sensors and connection to USB and Ethernet ports) 
must be done to the terminal strip on the right side of the panel in accordance with panel wiring diagram 
supplied with each W‐PD60. The table below: shows connection for typical wiring diagram of Figure 1‐4. 

Terminal #  Connections Note 


1  RTD sensor termination  White wire (common) 
2  RTD sensor termination  Black wire (signal) 
3  RTD sensor termination  Red wire (signal) 
4  CT – Load sensor termination Red wire (signal) 
5  CT – Load sensor termination Black wire (common) 
6  Humidity sensor termination White wire (common) 
7  Humidity sensor termination Red wire (+5V output) 
8  Humidity sensor termination Black wire (signal) 
9  4‐20mA current loop (+ ) output
10  4‐20mA current loop ( ‐ ) output
11  RS485 interface (A)   
12  RS485 interface (B)   
13  RS485 interface (insulated GND) Not connected
14  Alarm (Alarm 2) relay NC contact  
15  Alarm (Alarm 2) relay COM contact  
16  Alarm (Alarm 2) relay NO contact  
17  Warning (Alarm 1) relay NC contact  
18  Warning (Alarm 1) relay COM contact  
19  Warning (Alarm 1) relay NO contact  
20  Status relay NC contact   
21  Status relay COM contact  
22  Status relay NO contact   
23  Spare terminal  Not connected
24  Spare terminal  Not connected
25  AC Power Line protected by Fuse  
26  AC Power Line protected by Fuse  
27  Fuse input for External AC Power Line   
115/230VAC, 60/50Hz 
28  External AC Power Neutral  
29  Input for External AC Power Neutral  
30  External AC power Ground  
31  Input for External AC Power Ground
 
Coaxial cables from PD sensors (and from the noise sensor if used) must be connected directly to the 
Sensor Interface Board installed on a panel under the main unit (Figure 1‐3 b). Tie all cables together by 
tie‐wrap at the top of the hub inside the device enclosure. Connect the coaxial cables to their respective 
terminals at the Sensor Interface Board. Each channel has two input contacts: first (upper) for signal and 
second (lower) for coaxial shield connection. 

Page 20 of 112 
W‐PD60  user’s  manual  Ver6.0 

To make these connections a customer should remove the W‐PD60 from the panel, make all connections 
and then install the W‐PD60 back on the panel. 

 
Figure 1‐5 a, b ‐ Connection of the coaxial cable from PD sensors to the Sensor Interface Board, c – 
cable preparation, d ‐ stripper for RG58 cable, e – Switch. 
 
Using the coax cable stripper for RG58 cable (IDEAL part#45264 STRIP MASTER for RG58 coax cable, 
Newark catalog#05H8227 or similar) strip the outer PVC jacket to the distance of 1‐inch, unbraid the 
shield and remove the inner conductor insulation to the distance of about 0.3‐inch, twist shield wires to 
avoid “hairs” and crimp the supplied blade terminal to the shield (Figure 1‐5 c). Cut any loose “hairs” to 
avoid short‐circuiting. Do not use a blade terminal to connect the signal (central) conductor, mount it 
directly into the terminal block. Distribute cables evenly and secure them in place. 

To measure humidity (any signal applied to contact #12) the switch (Figure 1‐5 e) must be in the H% 
position. The switch position Ref is used only to measure a phase shift between external reference signal 
and inner IG synchronization (Section 7 Appendix II Synchronization and Phase Reference Shift). After 
this measurement the switch must be switched back to the H% position.. 

Page 21 of 112 
W-PD60 user’s manual Ver6.0

1.5.2 Door mount W‐PD60 

For door mounted W‐PD60 the Sensor Interface Board is mounted on the W‐PD60 back cover (Figure 
1‐6). 

 
 
Figure 1‐6 Door (Flush) mount W‐PD60 (back view) 
 
The W‐PD60 must be grounded at the place of its installation. 

Power line and output signals (Status, Warning (Alarm1) and Alarm (Alarm2) relays, RS485 and 4‐ 20mA 
interfaces) should be connected to W‐PD60 directly through molded plug‐in connectors supplied with 
the W‐PD60 in accordance with the label on the right side of the W‐PD60 (Figure 1‐3 c). 

All sensors should be connected directly to the Sensor Interface Board in accordance with its label. 
Coaxial cables from PD sensors should be connected to Ch1‐Ch15 inputs as it was described above for 
the panel mount W‐PD60. Auxiliary sensors should be connected to the 12 contact connectors on the left 
side of the Sensor Interface Board in accordance with Figure 1‐7 and table below. 

Page 22 of 112 
W‐PD60  user’s  manual  Ver6.0 

 
 
Figure 1‐7 Auxiliary sensors connection to sensor interface board. 
 
Auxiliary sensors connection: 
 
Contact  Sensor  Sensors wire Note 
#  connected 
1  Temperature  RTD common wire (white wire) Board common
2  sensor (100Ω  RTD signal wire (black wire) Output dc current ≈ 2mA 
3  Platinum RTD)  RTD signal wire (red wire) Output dc current ≈ 2mA 
4  Cable shield  Board common
5  Load (CT) sensor  Signal wire (red wire) Input impedance 100 _, 
6  Common wire (black wire) Board common
7  External  Usually not used Input impedance > 27k Ω. Input voltage should 
Reference  not exceed 1VAC. Disconnect jumper 1 if 
external reference signal is connected. 
8  n/a    test point
9  n/a    Board common
10  Humidity sensor Sensor common (white wire) Board common
11  Output + 5V for sensor (red wire) Output current is limited by 100Ωresistor.
12  Sensor output signal (black wire) Input impedance > 27kΩ. Input voltage should 
not exceed +5V. 
 
Notes: 

  ‐To measure humidity (or any signal applied to contact #12) the switch (Figure 1‐5 e) must be in the 
H% position. The switch position Ref is used only to measure a phase shift between external reference 
signal and inner IG synchronization during the start up procedure. After this measurement the switch 
must be returned to the H% position. 

  ‐The Jumper 1 (Figure 1‐5 b and Figure 1‐6) must be removed if external reference signal (e.g. from 
potential transformer) is applied to contact #7 on Sensor Interface Board (Section 7 Appendix II 
Synchronization and Phase Reference Shift). 

Page 23 of 112 
W-PD60 user’s manual Ver6.0

  ‐For W‐PD60 mounted on switchgear door the temperature and humidity sensors are usually 
installed inside the Sensor Interface Board because its temperature is close to the switchgear 
temperature. The switch must be in H% position. 

  Additional communication options, such as a modem, are separate items. 

1.6 Sensors for W‐PD60 
There are various PD sensors. Some basic sensors are: 

        IPDS ‐ Integrated Partial Discharge Sensor with voltage rating up to 38kV. It is an 80pF coupling 
capacitor with special protection circuitry built in. 
  RFCT ‐ Radio Frequency Current Transformer. 
  RTD‐6 ‐ PD sensor. An RTD installed in a rotating machine stator works as a local RF antenna for PD 
pulses. This sensor uses up to 6 RTDs to capture PD originated in motor stator. It is designed as a single 
board replacing the existing RTD terminal board in the motor frame and does not affect temperature 
measurements. 
The auxiliary sensors include: 

  TS ‐ Temperature sensor. It is a 3‐wire 100Ω platinum RTD. If the motor has a spare 100Ωplatinum 
RTD installed in its stator it can be used as an alternative. 
  HS ‐ Humidity sensor. 
  CT ‐ Load (CT) sensor (5 Amps primary, 1000:1) for rotating machines. It should be installed on a 
wire from secondary winding of the existent current transformer. 
  CTF ‐ Split‐core Current Transformer (FlexCore) with preinstalled CT sensor described above. It can 
be used if it is difficult to bring a load current signal to the W‐PD60 from a distant existing current 
transformer (frequently in switchgear cubicle). In this case consider installing CTF around a feeder cable 
of the phase A in the motor terminal enclosure. It should be installed over the shielded part of a cable 
and, therefore, is at ground potential and does not endanger insulation integrity. The cable shield should 
be grounded through the CT. 
All above sensors have default cable length of 65’ (20m). Specify other lengths if necessary. Every sensor 
is described in separate manuals. 

 
 
 
 
 
 
 

Page 24 of 112 
W‐PD60  user’s  manual  Ver6.0 

2 Application 

2.1 General 
W‐PD60 is a versatile device that can be applied to most medium and high voltage equipment. Specifics 
of a particular application (optional components and planning installation) should be taken into account 
when W‐PD60 is ordered. 

Issues that should be considered are: 

  Type and specifics of equipment to monitor 
  Configuration of PD sensors 
  Noise aspects 
  Device location and environmental issues 
  Additional parameters to measure 
  Networking and Communicating to the device 
 
You may wish to use separate documents like “Install Considerations” for different type of equipment. 
For each sensor installation use corresponding EE PD Sensor Installation Manual. 

2.1.1 Type and specifics of equipment to monitor 

In general, W‐PD60 was designed to monitor partial discharges in rotating machines, medium voltage 
switchgears, cast transformers and associated connections. The connections may include cable 
terminations and portions of a cable or a bus feeding a device. W‐PD60 can also be used on other types 
of medium voltage and high voltage equipment. The application of the technology may require special 
pre‐engineering and adaptation. Consideration should be given while attempting to use W‐PD60 on high 
voltage equipment above 120kV. External noise in the form of corona discharges should be addressed in 
such cases. 

The nature of PD indicates it will not occur in the equipment at voltages below 2300 Volts. Marginally, 
2300V equipment can be considered, but the technology is inconclusive. It may or may not detect a 
significant problem at 2300V. Equipment with voltage ratings of 4160V and higher are considered 
reliable for PD diagnostic applications. In general, higher voltage provides greater destructive power for 
partial discharges and problems can be more readily detected and predicted. 

2.2 Configuration of PD Sensors 
The configuration of PD sensors is specifically application oriented, although there are some general 
guidelines below. We recommend discussing any sensor installations with an Electrical application 
engineer. 

Page 25 of 112 
W-PD60 user’s manual Ver6.0

2.2.1 Configuration of Alarm Thresholds 

The Warning and Alarm thresholds for each channel of W‐PD60 have to be configured before any 
measurement. They depend on the type of sensor used (sensor sensitivity), type of equipment and its 
voltage as well as other conditions, e.g. sensor installation location, equipment environment et al. 

There are no universal standard values for Warning or Alarm thresholds. The following table just shows 
some typical values of the thresholds in some applications. 

Application  Sensor Type PDI PDI Qmax  Qmax 


Warning  Alarm  Warning  Alarm 
(mw) (mw) (mV)  (mV) 
5kV Motor / Small  Coupling Capacitor 3  10  50  100 
Generator 
6.9kV Motor / Small  Coupling Capacitor 10  30  100  500 
Generator 
6.9kV Motor / Small  RTD  10  30  100  500 
Generator 
15kV Motor /  Coupling Capacitor 20  100  200  1000 
Generator 
15kV Motor /  RTD  20  100  200  1000 
Generator 
5 kV Switchgear  Coupling Capacitor 3  5  50  100 
15 kV Switchgear  RFVS  3  10  50  200 
15 kV Switchgear  Coupling Capacitor 3 10 50  200 
38 kV Switchgear  RFCT  5  20  50  200 
 

2.2.2 Noise Aspects 

Noise is a major issue in any type of partial discharge measurement. Noise is always site‐related. There is 
no universal solution for noise suppression in the modern world, however most noise issues can be 
overcome. W‐PD60 has been designed with several noise suppressing algorithms and, along with 
recommended sensors, is able to suppress most industrial noise. While planning W‐PD60 installation, 
keep in mind these most common noise sources: 

  Power thyristor firing 
  Switching power supplies in modern metering and protective electronics 
  Arcing and sparking somewhere at the facility 
  Aerial corona on overhead lines on high voltage substation 
  Partial discharges in the adjacent equipment 
  Pulse signals coming from incoming lines and feeders 
 

2.2.3 Device Location and Environmental Issues 

W‐PD60 can be shipped with or without an environmental enclosure. Without the enclosure, it can be 

Page 26 of 112 
W‐PD60  user’s  manual  Ver6.0 

used indoors on switchgear lineups given a relatively clean environment. 

In most applications, indoor or outdoor, W‐PD60 is contained in a NEMA‐4 enclosure protecting the 
device from direct rain, sun, and dust. If the application environment is a chemically aggressive medium, 
or a company policy requires a stainless steel enclosure (NEMA‐4X), please specify it on the order for 
W‐PD60. Also you may use a specified NEMA4X plastic enclosure with a clear window. 

In a switchgear lineup W‐PD60 can be installed without an environmental enclosure. 

When planning to install W‐PD60, consider a non‐vibrating location, conveniently located in a centralized 
position between all sensors and possible communicating means (phone jack, tray to control room for 
alarms, etc.). The length of PD signal coaxial cables (RG58 or similar, 50Ω) can be up to 150ft. It is 
preferred to keep the PD signal cables short, but 150’ is not a solid limitation. 

2.2.4 Additional Parameters to Measure 

Several factors can affect PD level: temperature, humidity, voltage, mechanical vibration, etc. In North 
America system voltage commonly is not an issue. It is stable and does not affect PD significantly. If more 
than +/‐5% voltage variation is expected, the voltage should be measured and correlated to PD activity. 
Temperature is very important for rotating machine monitoring. Temperature variations can] change size 
of voids and surface conductivity which subsequently cause large changes on PD. Humidity is also a 
factor for line terminals in rotating machines, especially in an outdoor installation. 
Temperature and humidity are important factors in switchgear; often affecting contaminated surface 
conductivity or condensing moisture on some surfaces and pooling. Therefore, a varying range of PD is 
expected. If insulation can vibrate against support or core rotating machines, load current is a good 
parameter to correlate. Also zero load currents could be used to suspend monitoring during equipment 
outage (Refer to Section 1.4.10 Measurement Suspend W‐PD60 Software Manual). 
 

2.2.5 Networking and Communicating with W‐PD60 

W‐PD60 provides several interfaces for alarms, SCADA, and data retransmission. Always wire at least 
Alarm (Alarm 2) and Status relays to an annunciation system. If necessary, all relays can be wired in 
series or in parallel to the same annunciator. 

4‐20 mA output is for SCADA systems. A SCADA system can trend the most important parameters. 

If the optional communications board is installed, the USB Device port can be connected to a local PC 
running W‐PD60 software to provide device setting, initiating measurements, and downloading data. 
Communication options also include RS485‐USB converter, RS485‐Ethernet converter and landline 
modem. If an expert opinion will occasionally be needed, designate a permanent or temporary phone 
line with a modem. 

Through the Ethernet port of the optional communications board W‐PD60 status and information can be 
displayed online. Data can be downloaded through the FTP server. It also can be configured as a virtual 
serial port to be accessed by the W‐PD60 software. 

Page 27 of 112 
W-PD60 user’s manual Ver6.0

2.2.6 Motor and Small Generator Applications 

PD sensors for small generators and medium voltage motors typically include three sensors located at 
the line terminal and six to twelve RTD PD channels. Due to the large attenuation of high frequency 
signals in motor winding, sensors located at the line terminals will not produce valuable information for a 
winding depth; likewise, a sensor embedded into a winding will not sense discharges at the line terminal. 

Two types of sensors can be used at line terminals. 
  ‐Coupling Capacitors (IPDS) commonly with an 80pF capacitance. They have sufficient insulation and 
connect directly to motor line terminals. These sensors exhibit the most noise‐immune characteristics 
due to their low capacitance and, therefore, can suppress most low frequency noise by themselves. 
Using IPDS is a highly recommended option for line terminals. Note: Installation of line terminal sensors 
always requires an outage. 
  ‐Radio Frequency Current Transformers (RFCT) placed on incoming feeders’ shield grounding 
conductor (one per phase) or on single‐phase surge capacitor grounding conductor (one per phase). 
Using a single three‐phase surge capacitor is not recommended because phase‐to‐phase 

PD signals are shorted inside the capacitor and will not be detected. An advantage of RFCT is that they 
are non‐invasive (no connection to a live part is required). A disadvantage of using RFCT on a cable shield 
ground is, in some cases, the presence of ground noise. This means that RFCT measures all radio 
frequency signals coming through the cable shield. Using RFCT on a singlephase surge capacitor ground is 
more noise immune. 

As stated above, additional sensors embedded into winding sensors are required. RTD outputs are 
commonly used for this purpose. An RTD supplies a DC temperature signal to a metering device. An RTD 
with its wires also operate as a local RF antenna for PD acquisition. High frequency signals can be filtered 
out and used to for PD representation. Common noise signals from an RTD come from temperature 
metering system wires. A possible source of noise can be the operation (switching pulses) of the 
temperature metering system itself and/or pulses cross‐coupled to wires from another wire at the same 
tray. 

The RTD‐6 sensor board with embedded filters to suppress external noise. This sensor is designed as a 
single board to replace the existing RTD terminal board at the motor frame. It does not affect 
temperature measurements, and produces PD signals for W‐PD60. Each board can handle up to six RTDs. 
If in a particular application the number of RTDs is greater than 6, an additional board will be needed. An 
RTD‐6 board can be installed online which requires temporarily disabling the temperature alarm or relay 
system. 

Winding temperature, load current, and humidity in a terminal enclosure are the three most important 
parameters to correlate. Use one of the winding RTDs for temperature measurement. It can be difficult 
to bring a load current signal to the W‐PD60 enclosure from some distant CT (frequently in switchgear 
cubicle). In this case you can consider installing a FlexCore CT around a feeder cable of the phase A in the 
motor terminal enclosure. The FlexCore CT is installed over the shielded part of a cable and, therefore, is 
at ground potential and does not endanger insulation integrity. The cable shield should be grounded 
through the CT. 

Note: One W‐PD60 device is required for a motor. 

Page 28 of 112 
W‐PD60  user’s  manual  Ver6.0 

2.2.7 Large Generator Applications 

The large generator application is, in general, similar to one on a motor or small generator. Some 
specifics include: 
  ‐A coupling capacitor (IPDS) is commonly used as a sensor at line terminals. A generator equipped 
with iso‐phase bus duct might produce sparking in bus supporting insulators. A software filter has been 
built into the firmware. Another option is using two sets of IPDS for noise suppression. 
  ‐Certain generators do not have RTDs embedded into a stator winding. For example, large 
Westinghouse hydrogen cooled machines have RTDs in cooling gas pipelines. These RTDs cannot be used 
as PD sensors. Verify the RTDs are located in a winding. 
  ‐Large generators may have very long signal wires from the RTD to the first RTD terminal board. 
Frequently, these wires are also placed into a spiral armature. These factors can significantly affect high 
frequency signal propagation, attenuating the PD pulse. In order to minimize signal attenuation; one can 
install the RTD‐6 board at the generator gland, where the RTD wires come out of generator. 
 

2.2.8 MV Switchgear Applications 

Two types of PD sensors can be used for switchgear lineup: coupling capacitor (IPDS – preferable option) 
and RFCTs. 
  Three IPDS sensors (one per phase) are installed, commonly in every third cubicle. Therefore, each 
set of three IPDSs assess the cubicle where it is installed, and the cubicles on each sides. Increasing a 
distance between IPDS sets might cause some distant PD activity to be missed. Keep in mind that PD 
signals attenuate approximately twice from cubicle to cubicle due to energy splitting in every 
T‐connection. IPDS can be installed on a bus side or load side of a breaker. In the case of a load side, the 
sensor will not measure PD while the breaker is open. In general, one W‐PD60 can handle up to 15 
cubicles. 
  An RFCT sensor is installed on the incoming and outgoing feeder shield grounding conductors; 
commonly, one per cubicle. Phase to phase signals can be missed in such a configuration. Installing three 
RFCT sensors in every cubicle is very difficult. RFCT sensors on cable shields are susceptible to ground 
noise. 
   Two main factors, temperature and humidity, should be considered to correlate with PD. Plan on 
installing temperature and humidity sensors with W‐PD60. You may also consider voltage signal, if 
expected voltage variation can exceed +/‐5%. 

In indoor installations W‐PD60 can be mounted without NEMA‐4 enclosure in one of the cubicles on a 
back panel or door mounted. Temperature and Humidity sensors for that application can be 
incorporated inside W‐PD60. Check for available space before ordering a device. In some cases, the bus 
and breaker structure is very complicated. Consult with an Electrical application engineer before 
ordering sensors. 

Page 29 of 112 
W-PD60 user’s manual Ver6.0

3 On‐Site Installation 
USE START UP MANUAL step‐by‐step procedures as guidelines in actual installations and checking. 

3.1 W‐PD60 Kit Contents 
W‐PD60T is normally shipped in one of the following options: 
  ‐In a NEMA‐4X enclosure. 
  ‐On the mounting panel. This option is identical to the first one, except that the mounting panel is 
not mounted into an enclosure. 
  ‐Door mount option. This option is mainly used for W‐PD60 installation on a switchgear 
compartment door in switchgear line up. W‐PD60 comes with Sensor Interface Board installed on its 
back cover. Communication options e.g. modems and their power supplies are separate items. End‐user 
should install W‐PD60 with communication at a convenient location(s). 
See Section 1.5 W‐PD60 mounting options and sensor connection which explains W‐PD60 mounting 
options and sensor connection to its terminals. 

W‐PD60 Kit Contents: 

Item Description  Included/Optional Q‐ty  Comments 


W‐PD60 Device in enclosure, or  Included  1  Specify power voltage 115 
panel mount, or door–flush  VAC or 230 VAC at order 
mount (specify at order) 
115VAC, 10 A receptacle  Included for 115VAC power 1  Not included for door 
mount option and 230VAC 
power 
Vinyl insulated Straight Blade  Included  20 
Terminals TV14‐10BL for 
termination of coaxial 
cable sheaths 
W‐PD60 Communications  Optional 1  If included, it will be 
Board  assembled inside W‐PD60 
TS ‐Temperature sensor, 3‐wire  Optional 1  As default for switchgear 
100Ω RTD with 65’ cable  applications, installed on 
the sensor interface 
board. 
CT ‐ Load current Sensor CT  Optional 1  Needed for rotating 
(5 Amps primary, 1000:1)  equipment 
with 65’ cable 
CTF ‐ Split‐core Current  Optional 1  Convenient to install in 
Transformer (FlexCore) with 65’  motor termination box. 
cable  Specify current ratio while 
ordering. 
HS ‐ Humidity Sensor with  Optional 1  As default for switchgear 
65’ cable  applications, installed on 
the sensor interface 

Page 30 of 112 
W‐PD60  user’s  manual  Ver6.0 

board. 
Land Line Modem  Optional 1  Installed on the mounting 
panel, if order 
RS485/USB or RS485/Ethernet  Optional 1  Will need one for a 
converter with power supply  location or for every 
module  device in Ethernet 
network. May not be 
needed if optional 
communications board is 
ordered. 
IPDS‐XX ‐ PD Sensor (supplied  Optional 3  Specify it when ordering
with 65’ coaxial cables for 
connection to the device) 
RTD‐6 ‐ PD Sensor (supplied  Optional 1  Specify it when ordering
with 65’ coaxial cables for 
connection to the device) 
PD Sensor – RFCT with 65’  Optional 3  Specify it when ordering
coaxial cable 
 
Additional not included cables can be required: 
  ‐AC power cable 3‐wire (11 Amps). 
  ‐Twisted pair cables (AWG16‐20) for communication 
We recommend for these applications the following tray cables from Allied Electronics: 

Twisted pair: 
Stock # 216‐2722 (22AWG) by BELDEN Mfr’s #9322 
Stock # 216‐2724 (20AWG), by BELDEN Mfr’s #9320 (Preferable for communications over 1000ft) 
Stock # 216‐2660 (20AWG), by BELDEN Mfr’s #1033A 
Shielded 3‐wire cable (if T<800C), 
Stock # 216‐2738 (22AWG), by BELDEN Mfr’s #9363 
Stock # 216‐2740 (20AWG), by BELDEN Mfr’s #9364 
Stock # 216‐2666 (18AWG), by BELDEN Mfr’s #1036A 

You may order these cables through EE PD. Also additional RG‐58 coaxial cables can be ordered through 
EE PD in 500’ or 1000’ spools. 

3.2 Important Note on Sensor Installation 
W‐PD60 software has several useful configuration templates allowing simple configuration for a majority 
of standard applications. 

These templates for rotating machines assume that all coupling capacitors are connected sequentially to 
first PD channels in sequence of A‐B‐C. RTDs come after coupling capacitors again in sequences of A‐B‐C. 
Following this rule while connecting sensors to IG will simplify configuration later on! 

For switchgear application first connect all coupling capacitors to the first PD channels in sequence of 

Page 31 of 112 
W-PD60 user’s manual Ver6.0

A‐B‐C and then connect the RFCT sensors 

3.3 Step 1 ‐ Hardware Installation 
This section provides general guidelines. Specific options ordered with W‐PD60 must be taken into 
account during each installation. 

  ‐Select a convenient location for the W‐PD60 enclosure. It should be as close as possible to the 
equipment being monitored. Avoid locations with excessive vibrations, heat and direct sun light. Cable 
length from PD sensors should not exceed 150ft (50m). It is usually convenient to install the device in the 
center of the sensor locations, keeping PD signal cable lengths approximately the same. It is not 
necessary to have cable lengths exactly the same. 

  ‐Run conduits (if necessary) for the cables to all PD sensors, communication, analog signal cable, 
and for the power supply. Follow the corresponding installation manual for wiring to the PD sensors. 

  ‐Install the current sensor on the secondary circuit of a current transformer measuring the 
equipment load current. Phase A is preferred, but is not necessary. 

Note: It is often less expensive to install an additional current transformer inside a motor termination 
enclosure, rather than run conduits and wires from a distant breaker.   

  ‐Install the temperature sensor (if ordered separately) inside the switchgear at a suitable location 
closely representing the inside temperature. A rear compartment at the middle of a lineup usually works 
well for this. In the case of a motor or generator, use one of the spare winding RTDs. Otherwise locate 
the temperature sensor at a suitable location closely representing the stator winding temperature. 

  ‐Install the humidity sensor (if ordered separately) inside the switchgear at a suitable location 
closely representing the inside humidity. A rear compartment at the middle of a lineup usually works 
well for this also). In the case of a motor or generator, the humidity sensor should be installed inside the 
motor terminal enclosure to represent humidity at the line terminals. Insulation near the line terminals is 
most influenced by humidity variations. 

  ‐Connect the modem (if ordered separately) to a dedicated analog telephone line using a 2‐wire 
telephone cable with RJ11 connectors. The modem can also be installed at separate location. Several 
W‐PD60s can be networked together via RS‐485 network, using one modem for all of them. 

  ‐When networking several W‐PD60s together via RS‐485, use a daisy‐chain configuration. The chain 
must be terminated with a (typically) 120 _ resistor at the end. 

  ‐Run all cables into the W‐PD60 enclosure. Leave at least 24” of excess cabling in the enclosure to 
ease terminating. 

Note: Label all cables!!!. 

Page 32 of 112 
W‐PD60  user’s  manual  Ver6.0 

3.4 Step 2 ‐ Sensor and power connections 
The unit is to be provided with a disconnect device near the unit during installation. Note: Do not apply 
115/230 VAC power at this time. 

  ‐Connect coax cables to all PD sensors according to PD sensor installation manual. 

  ‐Connect a 3‐conductor shielded cable to spare platinum 100_ RTD terminals in a motor (or 
generator) according to drawings and RTD type. If an external temperature sensor was installed, connect 
its leads to W‐PD60 according to the drawing shipped with this W‐PD60. Label all conductors on both 
ends. (Check documentation! The temperature sensor may have already been installed on the sensor 
interface board for a door mount device.) 

  ‐Connect the current sensor according to the drawing shipped with this W‐PD60 and sensor manual. 

  ‐Connect the humidity sensor according to its manuals. A humidity sensor has three wires: +5 Volt, 
Ground, and Signal. Refer to the W‐PD60 drawing and sensor manual. (Check documentation! The 
humidity sensor may have already been installed on the sensor interface board for a door mount device.) 

  ‐Connect the voltage sensor if applicable. Label the positive polarity terminal with a (+) or dot. 

Note: Do not connect to the PT at this time. 

The sensor cable connections are not to be user accessible after installation. 

3.5 Step 3 ‐ Connections to the device. 
Connect all cables from the sensors to the device in accordance with Section 1.5. 

Use the actual wiring diagram supplied with W‐PD60. 

3.6 Step 4 ‐ Device Installation and Final Connections 
  ‐Connect the voltage sensor to the PT secondary at this time (if applicable). 

  ‐Check proper sensors connections with a multi‐meter. Use the detailed check‐list from the Start Up 
manuals. 

  ‐Install W‐PD60 and fasten it to the standoffs with the supplied screws. 

  ‐Connect the flat ribbon cable from W‐PD60 to the Signal Interface Board. Make sure that the two 
connector latches are locked. 

  ‐Check that the 3‐pin output relay molded connectors are inserted in their respective terminals on 

Page 33 of 112 
W-PD60 user’s manual Ver6.0

the right side of W‐PD60. The relay outputs are, from top to bottom: 

• Alarm (or Alarm 2) Relay 

• Warning (or Alarm 1) Relay 

•Device Status Relay 

  ‐Switch Power On to the W‐PD60. “Loading” and firmware version should appear on the device 
display. Device is now ready for setup and operation. 

  ‐Verify the signal wires and power are connected to the modem or/and converter. They typically 
have LEDs indicating power ON. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Page 34 of 112 
W‐PD60  user’s  manual  Ver6.0 

4 Startup 
REFER TO START UP MANUALS for short verification and start up procedures. 
 

4.1 General Setting with Keypad 
W‐PD60 setting can be done from the front panel keypad or from a PC using W‐PD60 software via the 
RS485 communication port or USB port on the optional communications board. (For software settings 
please refer to the Software manual.) 

Setting the device from a PC is easier, but there are some settings which can be changed ONLY from the 
keypad. These are parameters or procedures that must be set at device commissioning time or at factory 
calibration: 
  Modbus device address on a network 
  Modbus transmission Baud Rate 
  Phase Shift detection (Learn more in Section 7 Appendix II Synchronization and Phase Reference 
Shift before starting this). 
  Temperature channel calibration (commonly calibrated in the factory) 
  Current / Voltage channel calibration 
  Set ModBus Protocol 
  Set Frequency Range 
  Set minimum level for Noise filter 
 

4.2 Quick Start Up 
Normally, W‐PD60 is pre‐configured at the factory for a particular application in accordance with the 
questionnaire filled out by the customer during the ordering process. The setup procedures below are 
minimum requirements for setting up for monitoring. Specific applications might require some additional 
procedures. This sequence of steps is optimized for most typical cases, so following this sequence should 
minimize setup problems. 

Once W‐PD60 is installed, all sensors are connected and the equipment is powered up, a number of 
critical parameters must be configured at commissioning. The following description presents an ordered 
procedure for device settings. For a full description of setup options refer to the sections below and the 
software manual. Some of the setting procedures may be skipped if the default factory settings are used. 
The following procedure assumes the most difficult manual setup. If set from a PC, many settings can be 
combined into one setup command. Refer to the W‐PD60 Software Manual for details. 

 
Quick Startup Procedures through Keypad 
 
Parameter or  Function/Note See section #
Procedure 
Clear All  Clears all device data memory. 6.3.1

Page 35 of 112 
W-PD60 user’s manual Ver6.0

Memory  This is necessary at the beginning of PD monitoring of NEW equipment. 
Enter Setup  Enter the appropriate password to enter setup (Set). Your password is 5421  6
Date & Time  Note that setting the date earlier than the date of the last recorded  6.2.1, 6.2.2
measurement data causes an Alarm condition. 
Device Address  Device is shipped with the ModBus address 01. Make sure that the device  6.2.16
address is unique in RS‐485 network. The same addresses assigned to 
multiple devices will cause problems and contradictions in communications. 
Rated Voltage  Should be set before voltage channel calibration. 6.2.10
Voltage is used in PDI calculations.
Rated Current  Should be set before current channel calibration. 6.2.11
Activate  Activate PD channel that will be used for monitoring. 6.2.3
Channels  It is preferable to activate only the necessary channels and leave all the rest 
inactive. 
If none of channels is active, the device will not start a measurement. The 
Noise filter normally should be ON 
Channel  Ideally, sensor sensitivity should be determined by off line calibration test  6.2.5
Sensitivity  during the sensors’ installation. This does not always happen. Use suggested 
options in such case. Keep in mind that you may not be using an accurate 
number. Make sure that sensors of the same type and similar purpose have 
the same sensitivity. 
Channel Phase  Sensors connected to a particular channel are in most cases associated with  6.2.22
some phase. For example, a coupling capacitor connected to the phase B 
should be associated with phase B. Set associated phase for all active 
channels. 
In some cases a sensor senses all three phases (example, RFVS sensor in 
switchgear). There is not a phase association in such a case ‐ leave the phase 
A. 
Parameters to  Set parameters to scroll on the display you need in monitoring mode.  6.2.4
Display 
Determine Low  This procedure is not necessary in most applications. You can omit it.  6.2.34
Magnitude Limit    This utility, available only from the front panel keypad, will detect the Low 
for Channels  Limits for each channel independently. Select the Auomatic mode and wait 
until it completes. Scroll through the results and compare detected values for 
all channels with the same sensitivity. If they differ by 3 zones or less, allow 
the device to align them. Otherwise, leave them as is. Remember: 
sensitivities should be set for each channel before this operation. 
Calibration for  The goal of these procedures is to convert current and voltage inputs into  6.2.29
Current and  meaningful numbers on the display and in the database.  6.2.30 
Voltage  You may enter coefficients manually or automatically determine them while 
Channels  setting the device. 
The automatic procedure is available only from the front panel keypad. It 
converts the voltage and current inputs into % of rated value. 
Activate  These options use Current, Voltage, and/or Temperature (not recommended)  6.2.25
Auxiliary  readings to suspend measurements when any of them go beyond pre‐set  6.2.26 
Parameters to  boundaries. 
Suspend  Current is the best parameter to determine if equipment is out of service and 
Measurements  suspend measurements. Activate current for continuous check and set its 
boundaries. It is sufficient to set the current % lower boundary at 10–20%. 
Measurements will be suspended if the current goes below this value. 
Number of  Determines the length of time for a single measurement for one magnitude  6.2.12
Power Cycles  window. 
for Each  60 cycles is optimal but you may set it higher if better statistics is needed. 

Page 36 of 112 
W‐PD60  user’s  manual  Ver6.0 

Measurement  Increasing the number of cycles will proportionally increase the 
(per acquisition)  measurement time. 
Estimate the measurement time in seconds is as follows: number of active 
channels * 21 magnitude windows * selected number of cycles / power 
frequency. As an example with 15 channels and 60 cycles and 60Hz = 
15*21*60/60=315sec _ 5 minutes. 
Select PDI or  Choose PDI or Qmax as the parameter to represent the overall PD 6.2.19
Qmax  activity. This parameter is used for Alarm, Display and 4‐20mA output. 
Detection and  The phase reference shift is the phase angle between the InsuGard  6.2.21
setting of Phase  synchronization signal and a reference signal from the phase A voltage in the 
Reference Shift  monitored object. W‐PD60 uses this value for correct phase‐resolved data 
representation in W‐PD60 software and for analysis associated with this data. 
If this phase shift is known or determined by any tool (UPDA, oscilloscope or 
any phase meter) it should be manually entered in W‐PD60 settings. If it is 
unknown it can be detected by W‐PD60 using a reference signal from phase A 
of the monitored object or any other reference signal which has a known 
phase shift relative to phase A. For this measurement W‐PD60 works as 
simple phase meter. 
(See Section 7 Appendix II Synchronization and Phase Reference Shift 
for complete details.) 
Measurement  W‐PD60 is shipped with a schedule to take four measurements a day. This is  6.2.6
Schedule  the factory default setting. If needed, a single measurement can be done at 
any time with a command from a PC or the front panel keypad.. 
There is also an option to schedule measurements by time interval. 
Remember: the interval must be greater than the time of a single 
measurement! 
Set Data Save  The Alarm and Warning Relays can be set in one of three modes: 6.2.13
Mode  1. Disabled (OFF) – The relay will not close contacts in the case of an 
alarm/warning condition. 
2. Enabled (On) – The relay will keep alarm/warning contacts closed 
continuously while alarm/warning conditions are present 
3. Enabled for Time (t) – The alarm/warning relay contacts will be closed for 
the specified time (in sec.) when an alarm/warning condition is detected 
Exit Setup  Press ESC.   
Start a Single  This immediately starts a full measurement. As soon as the measurement  6.2.37
Measurement  completes check that the results are reasonable! 
 

4.3 Communications Board Setup 
If the optional communications board is installed some configuration and driver installation must be 
completed to gain complete access to all of its features. There are three communication ports available. 

 
Figure 4‐1 Communication Board Ports for W‐PD60 

Page 37 of 112 
W-PD60 user’s manual Ver6.0

4.3.1 USB 

1. Device USB port. This port is used for local communications between the W‐PD60 and a computer 
through a USB A/B cable. A software driver must be installed on the computer prior to use. The complete 
driver installation instructions can be found at Section 9 ‐ Appendix IV USB Device Driver Installation. 

2. Host USB port. This port is reserved for future use. 

4.3.2 Ethernet 

There are several functions available through the Ethernet port including webpage display, FTP upload or 
download and operation as a virtual serial port. 

Before any settings can be changed, you may need to change your computer’s network setting and 
access the W‐PD60 with an Ethernet cable connected directly between your computer and the 
communications board’s Ethernet port. The following are typical steps to initialize the communications. 

The default IP address for the communications board is 192.168.1.1. Note, if you plan to access W‐PD60 
on a LAN, or you need to remote monitor W‐PD60 online, you will need to change the IP address to an 
address compatible with the network where W‐PD60 is installed.. 

 
4.3.2.1 Change on a Computer’s Network Settings 

1. Go to “Network Connections”. If a Wireless Network is used, disable it, then right Click on “Local Area 
Connection”, and click on “Properties 

2. Click on “Internet Protocol (TCP/IP)”, then click on “Properties” 

Page 38 of 112 
W‐PD60  user’s  manual  Ver6.0 

3. Before making any changes, please note the current settings so that you can restore the settings upon 
completion. Change the settings to what is shown below. 

4. Connect your computer to the Ethernet port on W‐PD60 with an Ethernet cable. Open Internet Explorer, type in 

Page 39 of 112 
W-PD60 user’s manual Ver6.0

“192.168.1.1” on the address bar and the following screen will appear. Type in “admin” for the User name and 
Password and you are ready to go. 

 
4.3.2.2 Webpage 

W‐PD60 information can be accessed via webpages through the Ethernet port,. Some parameters must 
be configured before connecting to a network. Several other parameters should be customized to match 
the particular location. Details can be found at Section 8 Appendix III Communications Board Interface. 
The following table provides steps for configuring a typical installation. 

 
Parameter or  Function/Note See section #
Procedure 
Date and Time  Set the current date and time, synchronization to Network Time Protocol  8.5
servers, date format, time zone, and temperature units. 
Network  Set Ethernet port configuration if connecting to an Ethernet network.  8.2
Modbus RTU  Set Modbus Address and baud rate if connecting to a Modbus RS‐485  8.4
network. 
Email  Configure email server, recipient, and event trigger information for periodic  8.7
and alarm event email generation. 
FTP Client  Configure FTP server information and upload interval for automatic upload  8.6
of the measurement database. 
Access Control  Change passwords for web page access 8.3
Web Page  Customize location, contact information, and display names for web pages.  8.8, 8.9, 8.10
Customization 
 
4.3.2.3 Remote Communications As A Serial Port 

The Ethernet port can be configured as a virtual serial port which acts exactly the same way as the USB 
port does. When connected to a Local Area Network (LAN), W‐PD60 software can communicate with 
W‐PD60 remotely with the use of the serial port number created during the setup procedures. 

Page 40 of 112 
W‐PD60  user’s  manual  Ver6.0 

When users log into the company’s network through a VPN from an outside network, communications 
with W‐PD60 may also be possible depending on the company’s firewall and network settings. 

Please follow the steps in Section 10 Appendix V Ethernet Virtual Communications Port Driver 
Installation to set up the virtual serial port on the computer. 

 
4.3.3 Tips for Communications Setup 

The following tips only apply to W‐PD60 with the optional communications board. 

1. There is always a waiting time ranging from 20 seconds to 5 minutes before W‐PD60 software can 
communicate with an W‐PD60 unit after the following operations: W‐PD60 is powered up (1 minute), 
after keypad Memory and Set operations, during and for a short time after a measurement cycle. 

2. If communications through the USB port is not successful, unplug the USB cable and then plug it back 
into the computer. Note, please do not connect a USB cable between W‐PD60 and a computer while the 
computer is starting up or shutting down. 

3. During a measurement cycle, W‐PD60 cannot be accessed using W‐PD60 software through the virtual 
serial port or USB port, however, the webpage will still display the information on W‐PD60 stored before 
the measurement. 

4. If you are communicating through a virtual serial port via Ethernet, you may need to check if the IP 
address has been changed on the communications board. It is preferable to have a static IP address for 
the W‐PD60. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Page 41 of 112 
W-PD60 user’s manual Ver6.0

5 Troubleshooting 

5.1 Troubleshooting Tips 
Possible problems and probably causes and remedies are listed in the table below: 

##  Problem  Probable Cause Recommended remedy 


1  The front panel  Lost power  ‐Check incoming power;
display is blank.    ‐Remove the power leads and check the resistance 
  between neutral and line on the W‐PD60 input. It should 
  be about 47 _ for 115 VAC and about 157_ for 230VAC A 
  high resistance reading may indicate a blown fuse. If 
  necessary, replace the fuse and reconnect the power. If 
  significantly lower resistance is measured return the 
  unit for repair. 
Firmware programming  Try to reprogram in accordance with Firmware Upgrade 
has not competed  manual. If this fails, return the unit for repair. 
successfully 
2  The front panel  No data stored in W‐PD60 Scroll manually using the arrow keys. Start a single 
display is not  memory  measurement to place data in memory. 
scrolling.  Internal clock failure  Scroll until Time appears on the display Observe that the 
seconds are changing. If not, try reprogramming the 
firmware. If this action fails, return the unit for repair. 
3  When  W‐PD60 was sitting for a  Let the clock battery charge and update the date/time 
communicating to  long time without power.  from a PC or the front panel keypad. 
the device, there is  If the problem persists, return the unit for repair. 
a message that 
Date/Time differs 
from computer 
time significantly. 
4  The front panel    Upgrade the firmware. If this action fails return the unit 
display is blinking,  for repair. 
can’t communicate 
to the device from 
software. 
5  Flash Error    Cycle input power to the unit off and back on. 
message appears  If this doesn’t help upgrade the firmware. If this doesn’t 
on the front panel  help download data to the PC database and clear the 
display.  W‐PD60 memory. 
If the problem persists return the unit for repair. 
6  Error I Channel or  X Channel calibration  Upgrade the firmware. In rare instances this may correct 
Error N Channel  failed  the problem. If this does not, return the unit for repair. 
appears on the 
front panel display. 
7  The front panel  Lost contact between  Remove W‐PD60 back cover, unlock the connector push 
keypad is not  front panel flat cable and  cable in gently, lock connector. Carefully replace the 
working.  the connector on the  back cover. 
processor board  Don’t pinch the cable. If this does not correct the 
problem, return the unit for repair 
8  Communications  The RS485 network  Check the address setting in the software and in 
from the PC to  address is incorrect.  W‐PD60. 

Page 42 of 112 
W‐PD60  user’s  manual  Ver6.0 

W‐PD60 failed.  The communications baud  Set the same baud rate in the PC and W‐PD60 – the 


rate is incorrect.  default baud rate is 9600. 
The serial COM port # in  Check the computer port setting and type the proper 
software is incorrect  COM port # into the software. 
The modem or converter  Restore the settings to the as‐shipped values or 
setting is incorrect.  according to the settings in the modem/converter 
  manuals. 
The Comm. Board has not  Wait for about 1 minute before trying any 
fully started up  communication operation. 
USB from the Comm.  Disconnect and reconnect the USB cable on the Comm. 
Board does not work  Board 
9  W‐PD60 is not  The Load Threshold  Uncheck the box if W‐PD60 has no load sensor option.
making  Control check box is   
measurements.  software is checked, while   
the load current is below   
the threshold or zero   
The load sensor not  Check the current sensor connections. 
properly connected   
The load current  Set the rated current first, and then perform calibration 
calibration has not been  in accordance with this manual. 
performed or is incorrect  Note that after the rated current has been set from the 
  keypad, exit the set mode and go to the set mode again 
  for calibration. 
The flat ribbon cable  Check the flat rbbon cable connections. 
between W‐PD60 and the 
Sensor Interface board is 
not properly connected on 
either end, 
10  The load current  See above  See above
readings are 
incorrect. 
11  The readings for  The flat ribbon cable  Check the flat ribbon cable connections. 
temperature,  between W‐PD60 and the 
humidity and load  Sensor Interface board is 
current are  not properly connected on 
incorrect.  either end. 
12  The temperature  The temperature sensor  Check the sensors connection with an Ohmmeter. Check 
readings are  not connected properly.  that the voltage on the connected sensor signal 
incorrect.    terminals is about 0.25 VDC 
The flat ribbon cable  Check the flat ribbon cable connections. 
between W‐PD60 and the   
Sensor Interface board is   
not properly connected on   
either end.   
The temperature   
calibration coefficients  Recall the BackUp settings in the W‐PD60 software and 
have been corrupted.  “Send Settings to W‐PD60” with “Edit Analog Input 
Calibration Coef.” window opened. 
13  The humidity  The switch on the Sensor  Set the switch to the H% position 
readings are  Interface board is in the   
incorrect.  “Reference” position.   
The humidity sensor is not   

Page 43 of 112 
W-PD60 user’s manual Ver6.0

connected properly. Check that the voltage between contacts #10 and #12 
  on the Sensor Interface board is between 1.5 and 4 Vdc 
The humidity calibration  Recall the BackUp settings in the W‐PD60 software and 
coefficients have been  “Send Settings to W‐PD60” with “Edit Analog Input 
corrupted.  Calibration Coef.” window opened. 
14  W‐PD60 doesn’t  PC or W‐PD60 rebooted  Disconnect and reconnect USB cable 
communicate  with USB cable connected.
through USB port 
 
W‐PD60 replacement procedure: 
1. Have all configuration and data ready in the PC database file for the OLD W‐PD60. 
2. Run W‐PD60 Software. DO NOT open “Edit Analog Input Calibration Coef.” Window in software 
Configuration 
3. Change the “Magnitude Threshold, V” of the Continuous Watch function in the Configuration Tab with 
the new value which should be labeled on the front panel of the NEW W‐PD60. 
4. Send the Settings to the NEW W‐PD60. Get the Settings from the NEW W‐PD60 and click “Update the 
Database with the Current Device Configuration”. Backup the Settings. The database software will have 
analog coefficient for the NEW W‐PD60 with the remainder of the settings from the OLD W‐PD60. 
 

5.2 Error Messages 
W‐PD60 performs various tests when it powers up, before every measurement, and under some other 
conditions. If a fault is detected, an error message is displayed, and the Red LED is turned on. W‐PD60 
may display the following error messages: 

  No Synch. Signal ‐‐ ‐ Indicates a loss of the synchronization signal. Check that the external 
synchronization signal is above 0.4VAC and also refer to Section 7 Appendix II Synchronization and Phase 
Reference Shift for correct jumper configuration and more details. 

  Clock Error – This message indicates that some data in the device’s internal memory has a date 
stamp that is later than the current date. This condition degrades data consistency. To correct this 
condition download all data to the PC software, then delete all data in W‐PD60 and set the correct date 
and time. (Refer to Section 6.2.1 Setting Date and Section 6.2.2 Setting Time). 

  Error | Channel – Indicates that the PD measuring channel did not pass calibration. If this message 
appears consistently, the device must be returned for repair. 

  Error N Channel – Indicates that the PD noise filter channel did not pass calibration. If this appears 
consistently the device must be returned for repair. 

  FLASH Error – Indicates an error reading from internal FLASH memory. This may be fixed by deleting 
data from device memory. If this does not help, the device must be returned for repair. 

  STOPPED – Indicates measurements are stopped and the device will not resume measurements 
automatically. Please refer to Section 6.2.20 Start/Stop W‐PD60 for details on resuming operation from 
the front panel keypad or refer to the Software manual to resume operation from a PC. 

Page 44 of 112 
W‐PD60  user’s  manual  Ver6.0 

5.3 Fuse Replacement 
If 115/230VAC power is present, but the device display and LEDs do not turn on, the fuse may have 
blown. If no permanent damage has occurred fuse replacement may correct the problem. The display 
and LEDs may also be blank during reprogramming of the firmware or after an unsuccessful attept at 
programming. 

Remove input power and check the resistance between Line and Neutral inputs. For 115VAC applications 
it should be about 47_, for 230VAC applications about 157 _. Change the fuse if the resistance is very 
high. If the resistance is significantly less, the power transformer has probably failed and the unit should 
be returned for repair. 

Attention!!! Fuse replacement requires partial disassembly of the device and must be performed by a 
qualified person.   

W‐PD60 uses 5x20 mm, 250VAC rated Time Lag fuses by BEL Inc.: for 115VAC – 200mA fuse 5ST200‐R, 
for 230VAC – 100mA fuse 5ST100‐R. 

The fuse location is shown in Figure 5‐1 with the device back cover removed. 
 

 
Figure 5‐1 Fuse inside W‐PD60 
 
For fuse replacement: 
  Disconnect the device from 115/230 VAC power; 
  Disconnect the flat ribbon cable from the Sensor Interface Board; 
  Disconnect all connectors located on the right side of the device; 
  Unscrew the four screws on the front panel, holding the device to the panel; 
  Unscrew the two screws on the small back cover and disconnect flat ribbon cable from the device; 
  Unscrew the four screws holding the device back cover and carefully remove the cover. Pay 
particular attention to the flat cable connecting device’s keypad; 
  Check the old fuse, and if it blown replace it with new one. 
  Carefully reassemble the device and make all connections; 
  Check for correct connections and then reapply power. 
 

Page 45 of 112 
W-PD60 user’s manual Ver6.0

Attention!!! If after fuse replacement the second fuse opens while powering the device up, do not 
attempt replacing the fuse for a second time.   

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Page 46 of 112 
W‐PD60  user’s  manual  Ver6.0 

6 Appendix I W‐PD60 Front Panel Keypad Operation 

6.1 Keypad 
W‐PD60 has eight pushbuttons on the keypad. The functionality of each pushbutton is: 
  «ESC» ‐used for canceling a command and returning to the previous or higher‐level submenu; 
  «←»«→»«↑»«↓» – used to change values or navigate through menu options;   
  «Enter» ‐ used to select a menu option, start a procedure, or to confirm a value change; 
  «Memory» ‐ used to read measurement data in device memory; 
  «Set» ‐ used to enter the setup mode. 

The «Set» key should be pressed to enter the setup mode. The password prompt should appear on the 
display as shown: 

 
Enter the password 5421 by using «←» and «→» to move to different digits and «↑» and «↓» to 
change the values, then press «Enter». The password is coded into device memory and cannot be 
changed. 

Wait about 20 seconds for the display to change from dashed lines to “Set date” before pressing any 
other key. If the wrong password is entered the dashed lines remain on the display until another key is 
pressed. 

When W‐PD60 is in «Set» mode, the LED beside «Set» will be on. 

Note that if W‐PD60 is in «Set» mode for about 20 minutes, the setup menu will exit to the upper menu 
until it fully exits the «Set» mode to normal operating mode. 

Scroll through the setup options by pressing «→» or «←» keys and select an option by pressing «Enter». 
Exit the setup mode by pressing «ESC». All changes will be saved in setup memory upon exit. 

The following settings are listed in the order of pressing «→» after entering «Set» mode. If «←» is 
used, the settings order listed below should be reversed. 

6.2 Setup via Keypad 
6.2.1 Setting Date 

Page 47 of 112 
W-PD60 user’s manual Ver6.0

This setting is used to set the date in W‐PD60 with the format of MM/DD/YY. Press «Enter». 

Use «←» and «→»to move to the different digits to change. Use «↑» and «↓» to change the 
corresponding values. When finished, press «Enter» to confirm and exit. To discard the changes, press 
«ESC» to exit. 

In case the date is set earlier than the measurement data record with the latest preexisting date/time 
stamp, then the following screen appears: 

You can exit without erasing records by pressing «ESC». This will cause the “Alarm” LED to be turned on 
and “Clock Error” will be displayed upon exiting the Set mode. If pressing «Enter», all the measurement 
records later than the set data will be cleared. 

If the date is changed to an earlier date by the W‐PD60 software, the memory will not be cleared, but 
the “Alarm” LED will be turned on and “Clock Error” will be displayed. 

6.2.2 Setting Time 

 
This setting is used to set the time in W‐PD60 with the format of HH:MM:SS. Press «Enter». 

Use «←» and «→»to move to the different digits to change. Use «↑» and «↓» to change the 
corresponding values. When finished, press «Enter» to confirm and exit. To discard the changes, press 
«ESC» to exit. 

Page 48 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.3 Set Active Channels 

W‐PD60 has 15 signal channels. Depending upon the application, some channels are used while the rest 
remain inactive. To conserve memory and minimize measurement time do not activate more channels 
than necessary. This setting is used to set the active and inactive channels. 

Press «Enter». 

Use «←» or «→» to select the Channel number, use «↑» or «↓» to switch it “On” or “Off”.   

When all of the channels are properly set, press «Enter» to confirm the changes or «ESC» to discard the 
changes. 

Then the following screen appears 

Use «←» or «→» to switch the Noise Filter to “On” or “Off “, then press «Enter». 

6.2.4 Set Display Parameters 

This setting is used to determine which parameters will be automatically scrolled on the display when 
W‐PD60 is not making a measurement. 

Press «Enter» 

Page 49 of 112 
W-PD60 user’s manual Ver6.0

Use «←» or «→» to select “On” or “Off”. 

Use «↑» to continue to scroll through the items as listed below, or press «↓» to scroll back through the 
items. The following is a list of all the options. 

Turn on the date display: 

An example of the displayed information: 

Turn on the time display: 

An example of the displayed information: 

Turn on the display for next scheduled measurement time: 

An example of the displayed information: 

Page 50 of 112 
W‐PD60  user’s  manual  Ver6.0 

Turn on the PD data display: 

The PD data acquired in last measurement in each active channel will be displayed. Inactive channel(s) 
will not be displayed. An example of the displayed information if PDI is the selected parameter: 

The “Chan” means the Channel number, “%PD” means the measured PD as a percentage of the preset 
alarm level. For “Status”, Low means below the ”Normal” Threshold, Norm means elevated PD activity 
but still normal (above the “Normal” Threshold, but below the “Warning” threshold), Warn means above 
the “Warning” threshold but below the “Alarm” threshold and Alarm means above the ”Alarm” 
threshold. The displayed example means for Channel 2, the measured PDI is 30% of preset alarm level. 
The status is normal. The thresholds that control this status are set in Section 6.2.9: Set Alarm and Trend 
Thresholds. 

An example of the displayed information if Qmax is the selected parameter (For details on how to switch 
to PDI or Qmax, please refer to Section 6.2.19: Select PDI or Qmax): 

“Qmax” shows the measured data. “Chan” and “Status” show the same information as for PDI described 
above. 

Turn on the frequency display. The frequency of the signal used for synchronization is displayed. The 
default is the power frequency for W‐PD60. 

An example of the displayed information: 

 
Turn on the temperature display: 

Page 51 of 112 
W-PD60 user’s manual Ver6.0

An example of the displayed information: 

Turn on the voltage display: Note that the input channel used to measure voltage is typically used to 
measure humidity. In this case the voltage display should be turned off. 

An example of the displayed information.    The voltage is displayed as a percent of rated voltage. 

Turn on the load current display: 

An example of the displayed information: The current is displayed as a percent of rated current. 

Turn on the relative humidity display: Note that if a humidity sensor is installed and this display is turned 
on then either the voltage display (typical) or current display should be turned off, depending on which 
input channel the humidity sensor is connected to. 

An example of the displayed information: 

Page 52 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.5 Set Sensor Sensitivity 

 
W‐PD60 can use almost any PD sensor that is capable of operating in the radio‐frequency band. Sensors 
differ by their response to the same discharge event. In other words, sensors have different sensitivity. In 
order to take into account the specifics of a particular sensor, proper sensitivity for each sensor should 
be set. Generally sensor sensitivity can be determined through off line sensor calibration. 

It is recommended that you use the same sensitivity for all sensors of the same type and function. 

If off‐line sensor calibration cannot be easily performed, use the suggested values below (the value in 
parenthesis is the preferred one for each type of sensor): 
1. Coupling Capacitor 3 ‐ 5 nC/Volt (3) 
2. RFCT on cable drain shield 5 ‐ 10 nC/Volt (10) 
3. RFCT on surge capacitor ground 3 ‐ 5 nC/Volt (5) 
4. RTD sensor in motor/small generator (no spiral shielding on RTD wires) 5 – 15 nC/Volt (10) 
5. RTD sensor in large generator in spiral shields 50 – 70 nC/Volt (50) 
6. RFVS sensor in switchgear 10 – 20 nC/Volt. (15) 

Press «Enter» 

 
This shows the sensitivity for channel 1 is currently set at 10.00 nC/Volt. Use «↑» or «↓» to change the 
Channel number. Press «Enter», then use «↑» or «↓» to change the values. Use «←» or «→» to move 
to the different digits. For each channel, press «Enter» again to confirm the change. After all change are 
completed, press «ESC» to exit. 

6.2.6 Set Measurement Schedule 

 
W‐PD60 can be set with a fixed time or time interval to make PD measurement automatically. This 
setting is used to set the time or time interval. Press «Enter». 

Page 53 of 112 
W-PD60 user’s manual Ver6.0

1. To configure W‐PD60 to make measurement according to the time of a day, use «←» or «→» until 
“Time of Day” is displayed as shown above . Press «Enter». Use «↑» or «↓» to scroll through the 
different measurement time settings. 

When the time shown is 00:00, it means that no time has been set for “# XX”. To change the time, press 
«Enter». Then use «↑» or «↓» to change the value of the hour or minute. Use «←» or «→» to switch 
between hour and minute. After each change is made, press «Enter» to confirm it. Press «ESC» when all 
changes are made. W‐PD60 allows up to 50 measurement times to be set up. 

2. To configure W‐PD60 to make measurement by time interval, use «←» or «→» until “Time Interval” 
is displayed as shown above. Press «Enter». Use «_» or «_» to switch between hour and minute. Use 
«↑» or «↓» to change the value of the hour or minute. 

The example shown above means W‐PD60 will make a measurement every 6 hours. Press «Enter» to 
save the time interval and exit. Press «ESC» to discard changes to the time interval and exit. 

Please note that the Time Interval set must be less than the time for a single measurement. Otherwise 
you may create an infinite cycle and will not be able to get out without rebooting. To estimate the time 
of a measurement use the formula t= Ncycle *Nch*21/ f +1 minute, Ncycle – number of power cycles, Nch – 
number of active channels, f is the power frequency, usually 60 or 50. 

Note: The W‐PD60 will make measurements according to “Time of Day” or “Time Interval”, whichever is 
set last. 

Page 54 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.7 Set Data Saving Mode 

 
Two data saving modes are available: Full (saving all the data including full three dimensional 
Phase‐Resolved Pulse Height Distribution (PRPHD) matrixes for every measurement) and Brief (saving 
header only including PDI or Qmax for most measurements, and saving full PRPHD matrixes at a 
programmable interval). 

Press «Enter» 

1. To save data in the Full mode, use «←» or «→» until “Full” is displayed as shown above, then press 
«Enter». 

2. To save data in the Brief mode use «←» or «→» until “Brief” is diplayed as shown above, then press 
«Enter». 

The next setting is used to set how often measurement data will be saved in the Full mode. At all other 
times measurement data is saved in the Brief mode. 

Use «←» or «→» to switch between the two digits. Use «↑» or «↓» to change the digits. 

Then press «Enter» 

The last setting for this mode is used to determine which measurement on the specified day will be 
saved in full mode. 

Use «←» or «→» to switch between the two digits. Use «↑» or «↓» to change the digits, then press 

Page 55 of 112 
W-PD60 user’s manual Ver6.0

«Enter» to save the changes.. 

Note: We suggest saving full data twice per month. This is sufficient for later diagnostics. Full data can be 
saved at any time by the operator with a single measurement. For details, please refer to Section 6.2.37 ‐ 
Start a Single Measurement. 

6.2.8 Set Pulse Width and Interval 

 
This procedure sets the allowed PD pulse width and the allowed time interval (or dead time) between 
two consecutive pulses. The pulse width setting helps to reject industrial noise since this noise often 
produces a much longer pulse than PD. The time between pulses sets the criteria to determine when a 
new pulse appears vs. when a signal is a continuation (tail) of the previous pulse. 

Press «Enter» 

Use «↑» or «↓» to change the channel number. Then press «Enter» again 

Use «↑» or «↓» to switch between “PD Width” and time “Interval”. Use «←» or «→» to change the 
values. 

For the PD width, the selections are: 
_ 640nS (recommended for RTD, IPDS, RFVS and RFCT sensors) 
_ 1280nS (may be used for RFCT sensors on cables) 
_ 2560nS 

For the time interval, the selections are 640nS, 1280nS or 2560nS (recommended). 

After all changes for a channel are done, press «Enter» to confirm and move to the next channel . After 
changes for all channels are confirmed, press «ESC» to exit. 

Page 56 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.9 Set Alarm and Trend Thresholds 

 
W‐PD60 operates the alarm contacts based on the PD parameter level and trend (rate of parameter 
change). Two PD parameters are available for alarming: PDI – partial discharge intensity and Qmax – 
maximum PD pulse magnitude @ 0.2 PPC repetition rate. Only one parameter can be enabled for 
alarming at a time. Refer to Section 6.2.19 ‐ Select PDI or Qmax for how to switch between the two 
parameters. The same parameter applies to the 4‐20mAmp output. 

  There are three level thresholds: Alarm, Warning, and Normal. The Alarm and Warning thresholds 
determine when the alarm and warning relays will be operated.. The Normal threshold suppresses trend 
alarm calculations while the PD parameter is at or below this level. In general, we recommend having the 
following proportion between threshold limits: Warning 25‐50% of Alarm, and Normal 15‐25% of 
Warning. 

Press «Enter» 

Use «↑» or «↓» to change the channel number. Then press «Enter» again: 

The screen above is for the PDI Normal level setting. 

Use «↑» or «↓» to scroll through the different threshold level settings as listed in the following screens, 
then press «Enter» to modify the individual thresholds: 

The screen above is for the PDI Warning Level setting. 

The screen above is for the PDI Alarm level setting. 

Page 57 of 112 
W-PD60 user’s manual Ver6.0

The screen above is for the Qmax Normal level setting. 

The screen above is for the Qmax Warning level setting. 

The screen above is for the Qmax Alarm level setting. 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. Press «Enter» 
to confirm each change and «Esc» to exit to the next higher level menu.. 

To change the trend level, keep pressing «↑» at the channel selection screen   

After Channel 15, the following screen appears, 

Press «Enter» 

The screen above is for the Trend Warning level setting. 

Use «↑» or «↓» to scroll between the alarm and warning threshold level settings, then press «Enter» 
to modify the individual thresholds: 

Page 58 of 112 
W‐PD60  user’s  manual  Ver6.0 

The screen above is for the Trend Alarm level setting. 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. Press «Enter» 
to confirm each change and press «ESC» to exit to the next higher level menu. 

6.2.10 Set Rated Voltage 

 
 
This setting is used to set the rated voltage rating for the monitored device. Set the rated line‐to‐line 
voltage in terms of kV. This must be set before the voltage channel is calibrated. This parameter is also 
used for the PDI (Partial Discharge Intensity) calculation. 

Press «Enter» 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. After all 
changes are made, press «Enter» to confirm the changes and exit, or «ESC» to discard any changes and 
exit. 

6.2.11 Set Rated Current 

 
This setting is used to set the rated current rating for the monitored device in terms of Amps. It must be 
set before the current channel is calibrated. Press «Enter» 

Page 59 of 112 
W-PD60 user’s manual Ver6.0

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. After all 
changes are made, press «Enter» to confirm the changes and exit, or «ESC» to discard any changes and 
exit. 

6.2.12 Set Number of Power Cycles for Data Acquisition 

 
This setting determines the length of time that data will be aquired for each scheduled or single 
measurement to get sufficient statistical PD representation. The parameter is measured in terms of the 
number of cycles of power frequency. 60 cycles is recommended for rotating machines and 120 cycles 
for switchgear. More cycles can be set for higher statistics precision. 

Measurement time can be estimated as the following: number of active channels * 21 magnitude 
windows * selected number of cycles / power frequency. For example, for 15 channels and 60 cycles, 
Measurement Time = 15*21*60(cycles)/60(Hz)=315sec~5 minutes. 

Press «Enter» 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. After all 
changes are made, press «Enter» to confirm the changes and exit, or «ESC» to discard any changes and 
exit. 

6.2.13 Set Alarm Relay Mode 

 
This setting is used to set up how the alarm and warning relays operate. 
Disabled – The relays will not close their contacts in case of alarm/warning conditions. 
Enabled – Each relay will keep its contacts closed for the complete duration that the respecitve alarm or 
warning condition is present 
Timed Enable – The alarm/warning relay contacts will be closed for the specified time (in sec.) when an 
alarm/warning condition is detected 

Press «Enter» 

Page 60 of 112 
W‐PD60  user’s  manual  Ver6.0 

Use «←» or «→» to change to either of the following settings, 

Or 

To select “Enabled” or “Disabled” option, just press «Enter» at corresponding screen. To select “Timed 
Enabled”, press «Enter» at “Timed Enable” screen, 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. After all 
changes are made, press «Enter» to confirm the changes and exit, or «ESC» to discard any changes and 
exit. 

6.2.14 Set Minimum Noise Filter Level 

 
This setting is used to set the minimum level for the noise filter. The noise filter level slides down as the 
magnitude window goes down during PD measurements. The noise filter will stop at this minimum 
magnitude setting for the noise filter level while the PD signal channel will continue to slide until it 
reaches the lowest magnitude window for PD measurements. 

Press «Enter» 

Page 61 of 112 
W-PD60 user’s manual Ver6.0

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. After all 
changes are made, press «Enter» to confirm the changes and exit, or «ESC» to discard any changes and 
exit. 

6.2.15 Set Display Scroll Time 

 
This setting sets the time that each item of information set in Section 6.2.4 ‐ Set Display Parameters stays 
on the display before scrolling to the next item. Press «Enter» 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. The default 
time is 3 seconds. After all changes are made, press «Enter» to confirm the changes and exit, or «ESC» to 
discard any changes and exit. 

6.2.16 Set ModBus Address 

 
This is a critical setting for communications. Make sure the address is unique in the RS485 network that 
the W‐PD60 is attached to. Press «Enter» 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. After all 
changes are made, press «Enter» to confirm the changes and exit, or «ESC» to discard any changes and 
exit. 

Page 62 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.17 Set Continuous Watch Options 

 
W‐PD60 has a Continuous Watch feature. This option is enabled between measurements and utilizes a 
separate counter. All PD signals are brought to this counter to catch the high level pulses and their series. 
If the device counts 5 series of high level pulses, it starts a regular measurement.,If the alarm threshold is 
exceeded,it stores the full measurement data and turns on the alarm. 

Press «Enter» 

This sets the pulse magnitude level. Use «←» or «→» to move between the digits, and use «↑» or «↓» 
to change the value. After all changes are made, press «Enter» to confirm the changes and move to the 
next screen, or «ESC» to discard the changes and move to the next screen. 

 
This sets the pulse repetition rate PPS (Pulses Per Second). Use «←» or «→» to move between the digits, 
and use «↑» or «↓» to change the values. After all changes are made, press «Enter» to confirm the 
changes and exit, or «ESC» to discard any changes and exit. 

6.2.18 Set Baud Rate 

 
This sets the baud rate for serial communications through the RS‐485 port. Press «Enter» 

Use «←» or «→» to change between 9600BPS and 38400BPS. 9600 BPS is default for most RS‐485 
network applications. 38400 BPS can be applied on good transmission lines or when the optional 
communications board is used without an RS‐485 network. Press «Enter» to confirm the selection and 
exit. 

Page 63 of 112 
W-PD60 user’s manual Ver6.0

6.2.19 Select PDI or Qmax 

 
This setting is used to select either PDI (Partial Discharge Intensity) or Qmax (Partial Discharge 
Magnitude) as the measurement parameters for all PD channels as well as the 4‐20 mA output. 

Press «Enter» 

To select PDI press «Enter». Otherwise, press «←» or «→» 

This is the option to select PD magnitude Qmax as the output. Press «Enter» to confirm it and exit. 

6.2.20 Start/Stop W‐PD60 

 
This function is used to stop the monitoring mode for an undefined length of time (prolonged outage 
and so on) and to restart monitoring later. Note: It is the responsibility of the operator to resume 
monitoring. 

Press «Enter» 

Use «←» or «→» to switch between Start or Stop monitoring. Press «Enter» to confirm the selection and 
exit. 

Page 64 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.21 Set Phase Shift 

 
By default, W‐PD60 uses the internal phase reference from its 115/230 VAC power source. (External 
phase reference is also available for specific applications such as variable frequency drives‐VFD). 
Obtaining a correct relative phase shift between the internal phase reference and the voltage on the 
monitored device is very important for correct interpretation of phase‐resolved PD distribution (PRPDD). 

This value needs to be established by an operator during setup time. It is the value of the phase shift in 
degrees between the Line (Phase A)‐to‐ground voltage of the monitored device and the Line‐to‐Neutral 
(Ground) voltage of the W‐PD60 power supply. The phase shift is counted starting from the power 
supply positive zero crossing. 

The required Phase Shift can be entered into the device manually if determined by any independent tool 
(e.g. oscilloscope), or it can be determined automatically with the precision of 5 degrees. 

For automatic detection, a phase reference signal from the A‐phase should be temporarily connected to 
the W‐PD60. Please refer to section 7 : Appendix II Synchronization and Phase Reference Shift.. 

Note: in all cases, the reference signal should not exceed 5V AC. 

Press «Enter» 

Use «←», «→», «↑» or «↓» to switch to auto detect mode as shown below 

To manually input the phase shift, press «Enter» at the “Manually Input” menu, 

Use «←» or «→» to move between the digits, and use «↑» or «↓» to change the values. After all 
changes are made, press «Enter» to to confirm the changes and exit, or «ESC» to discard any changes 
and exit. 

Page 65 of 112 
W-PD60 user’s manual Ver6.0

6.2.22 Set Channel Phase 

 
As the W‐PD60 uses a variety of sensors that can be associated directly or indirectly with any phase (or 
combination of phases) of the electrical system, it is important that each sensor be assigned to a correct 
phase. This is important for the accurate interpretation of PD data. For example, a coupling capacitor 
connected to B‐phase should be assigned to phase B. In some cases a sensor senses all three phases, for 
example, RFVS sensor in SWG. There is no phase association in such a case ‐ leave the default setting of 
Phase A for such a sensor. 

Press «Enter» 

Use «←» or «→» to change the Phase. Use «↑» or «↓» to change the channel number. After all 
changes are made, press «Enter» to confirm all changes and exit, or «ESC» to discard any changes and 
exit. 

6.2.23 Set Frequency Range 

 
This setting is used only for PD measurement on very low frequency devices. For regular frequency 
devices in the range of 20Hz~400Hz, the selection should be “> 20Hz”. For devices operating at 
frequencies between 3Hz and 20Hz, “< 20Hz” should be selected. 

Press «Enter» 

Use «←», «→», «↑» or «↓» to change between “>” and “<”. Press «Enter» to confirm the selection and 
exit. 

Page 66 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.24 Set Suspend for Temperature 

 
This setting is used to suspend W‐PD60 at some extreme temperature in terms of ℃. Press «Enter» 

Use «←», «→», «↑» or «↓» to switch between “Off” and “On”. When “Off” is selected, pressing 
«Enter» will confirm the selection and exit. When “On” is selected, press «Enter» 

Use «←» or «→» to move between the digits. Use    «↑» or «↓» to change the values. After all changes 
are made, press «Enter» to confirm or «ESC» to discard any changes and move to the next setting. 

The example above means when the temperature is below ‐50℃, W‐PD60 will suspend monitoring until 
the temperature returns to over ‐50℃. 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values. After all changes 
are made, press «Enter» to confirm all changes and exit, or «ESC» to discard any changes and exit. 

The example above means when the temperature is above 150℃, W‐PD60 will suspend monitoring until 
the temperature returns below 150℃. 

6.2.25 Set Suspend for Voltage 

 
This setting is used to suspend W‐PD60 at some extreme voltage in terms of percentage of the rated 
voltage. For example, when a monitored device is shut down, W‐PD60 will suspend monitoring 
automatically if the Suspend for Voltage Low Limit is set close to zero. 

Press «Enter» 

Page 67 of 112 
W-PD60 user’s manual Ver6.0

Use «←», «→», «↑» or «↓» to switch between “Off” and “On”. When “Off” is selected, pressing 
«Enter» will confirm the selection and exit. When “On” is selected, press «Enter» 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values. After all changes 
are made, press «Enter» to confirm or «ESC» to discard any changes and move to the next setting. 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values. After all changes 
are made, press «Enter» to confirm all changes and exit, or «ESC» to discard any changes and exit. 

Activating voltage suspend mode means when the monitored voltage is out of range, i.e. lower than the 
low limit or higher than the high limit as set above, W‐PD60 will stop monitoring. 

6.2.26 Set Suspend for Current 

This setting is used to suspend W‐PD60 at some extreme current in terms of percentage of the 
ratedcurrent. For example, when a monitored device is shut down, W‐PD60 will suspend monitoring 
automatically if the Suspend For Current Low Limit is set close to zero. 

Press «Enter» 

Use «←», «→», «↑» or «↓» to switch between “Off” and “On”. When “Off” is selected, pressing 
«Enter» will confirm the selection and exit. When “On” is selected, press «Enter» 

Page 68 of 112 
W‐PD60  user’s  manual  Ver6.0 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values. After all changes 
are made, press «Enter» to confirm or «ESC» to discard any changes and move to the next setting. 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values. After all changes 
are made, press «Enter» to confirm all changes and exit, or «ESC» to discard any changes and exit. 

Activating current suspend mode means when the monitored current is out of range, i.e. lower than the 
low limit or higher than the high limit as set above, W‐PD60 will stop monitoring. 

6.2.27 Set Trend Window Length 

 
W‐PD60 provides alarms and warnings based on levels of PDI or Magnitude as well as the trend of either 
of these levels as set in section 6.2.9: Set Alarm and Trend Thresholds. This setting sets the sliding time 
window represented in weeks for trend calculations. The trend calculation is performed on the stored 
data between the current time and this set number of weeks in the past. 

Press «Enter» 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values. After all changes 
are made, press «Enter» to confirm all changes and exit, or «ESC» to discard any changes and exit. 

6.2.28 Calibrate Temperature 

 
This setting is used to calibrate W‐PD60 temperature channel. It is calibrated at the factory for a 100Ω 

Page 69 of 112 
W-PD60 user’s manual Ver6.0

Platinum RTD. Field calibration should not be performed unless absolutely necessary, such as when using 
a different type of RTD.. 

The calibration requires two precision resistors: For a Platinum RTD 100.0Ωrepresenting a temperature 
of 0℃ and 138.6Ωrepresenting a temperatures of 100℃. They should be connected as shown in Figure 
6‐1. Terminals 1 and 2 must be shortened and the resistor must be connected between the terminals 2 
and 3 of the sensor interface board. 

Figure 6‐1 Temperature channel calibration 

Press «Enter» to start the calibration. 

Connect a resistor whose resistance is equal to the resistance of the RTD type being used at 0℃. The 
displayed ADC parameter can be ignored. Press «Enter» 

Remove the first resistor and connect a resistor whose resistance is equal to the resistance of the RTD 
type being used at 100℃. Press «Enter», the first calibration coefficient will be shown 

Press «Enter», the second calibration coefficient will be shown 

Press «Enter» to confirm and exit. 

Page 70 of 112 
W‐PD60  user’s  manual  Ver6.0 

6.2.29 Calibrate Voltage 

 
This setting is used to calibrate a voltage sensor. Voltage calibration is normally done on site. Sensors 
must be connected and the monitored equipment must be energized and carrying a load. Voltage 
calibration must be done after the rated value is set (Section 6.2.10 Set Rated Voltage). There are two 
ways to calibrate voltage – manual or automatic. 

1. Manual Calculation and Entry 

The voltage coefficient should be calculated in the following way. Get the reading of the operating 
voltage of the equipment (Vrun) using whatever method is available e.g. calling control room. Get the 
voltage reading (P%) from the front panel display of W‐PD60. Assuming the rated voltage is Vrate, the 
current voltage coefficient is _read, the new coefficient should be calculated by _read *Vrun/(Vrate*P%). 

Press «Enter», use «←», «→», «↑» or «↓» to switch to the screen below 

Press «Enter» again. 

Input the coefficient as calculated by the equation above. Use «←» or «→» to switch between the digits. 
Use «↑» or «↓» to change the values. After all changes are made, press «Enter». 

2. Automatic 

Press «Enter», use «←», «→», «↑» or «↓» to switch to the screen below 

Press «Enter» again. 

Page 71 of 112 
W-PD60 user’s manual Ver6.0

Obtain the present operating voltage and enter as the percentage of rated voltage (Voper/Vrated *100) Use 
«←» or «→» to move between the digits. Use «↑» or «↓» to change the values. After all changes are 
made, press «Enter». 

The automatic calculated voltage coefficient is shown. Press «Enter» again to exit. 

6.2.30 Calibrate Current 

This setting is used to calibrate a CT sensor for load current. Current calibration is normally done on site. 
Sensors must be connected and the monitored equipment must be energized carrying a load. 

Current calibration must be done after the rated value is set (Section 6.2.11 Set Rated Current). There 
are two ways to calibrate it – manual or automatic. 

1. Manual Calculation and Entry 

The current coefficient should be calculated this way. Get the reading of the running current (Irun) in any 
way, e.g. calling control room, get the current reading (P%) on W‐PD60. Assuming the rated current is 
Irate, the current coefficient is Kread, the new coefficient should be calculated by Kread *Irun/(Irate*P%). 

Press «Enter», use «←», «→», «↑» or «↓» to switch to the screen below 

Press «Enter» again. 

Input the calculated coefficient. Use «←» or «→» to switch between the digits. Use «↑» or «↓» to 
change the values. After all changes are made, press «Enter». 

2. Automatic 

Page 72 of 112 
W‐PD60  user’s  manual  Ver6.0 

Press «Enter», use «←», «→», «↑» or «↓» to switch to the screen below 

Press «Enter» again. 

Obtain the present operating curren and input as the percentage of rated current (Ioper/Irated *100). Use 
«←» or «→» to move between the digits. Use    «↑» or «↓» to change the values. After all changes are 
made, press «Enter». 

The automatic calculated current coefficient is shown. Press «Enter» again to exit. 

6.2.31 Calibrate 4‐20mA Loop 

 
The interface is isolated and powered from W‐PD60. It is normally pre‐calibrated at the factory. 

This output can be calibrated in order to adjust for specific circuitry. Using this procedure the device will 
set 5, 7, 9, 11, 13, 15, 17 and 19 mAmps on its output. If the receiving interface reads differently, its 
values can be entered into W‐PD60, which will calculate a linear approximation with the least square 
method. 

Press «Enter» 

To change the “Actual” value as read on the receiving interface, use «←» or «→» to move between the 
digits, and use «↑» or «↓» to change the values. The same operation rules apply to all of the “Actual” 
values below. 

Page 73 of 112 
W-PD60 user’s manual Ver6.0

Press «Enter» 

Press «Enter» 

Press «Enter» 

Press «Enter» 

Press «Enter» 

Press «Enter» 

Press «Enter» 

Press «Enter», the new calibration coefficients will be shown: 

Page 74 of 112 
W‐PD60  user’s  manual  Ver6.0 

Press «Enter» 

Press «Enter» again to exit. 

6.2.32 Calibrate Humidity Sensor 

 
Do not calibrate the humidity channel unless absolutely necessary. It is pre‐calibrated at the factory. 

To calibrate the humidity sensor, press «Enter», then use «←», «→», «↑» or «↓» to switch between 
three options: 

1. Turn off the humidity monitoring. Press «Enter» when the screen below appears. 

2. There are two channels to monitor two of three parameters: humidity, voltage and current. To use 
the voltage channel to monitor the humidity sensor, switch to the following screen, 

Press «Enter» 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values, then press 
«Enter» to confirm the changes or «ESC» to discard any changes and move to the next setting. 

Page 75 of 112 
W-PD60 user’s manual Ver6.0

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values, then press 
«Enter» to confirm all changes and exit, or «ESC» to discard any changes and exit. 

3. To use the current channel to monitor the humidity sensor, switch to the following screen, 

Press «Enter» 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values, then press 
«Enter» to confirm the changes or «ESC» to discard any changes and move to the next setting. 

Use «←» or «→» to move between the digits. Use «↑» or «↓» to change the values, then press 
«Enter» to confirm all changes and exit, or «ESC» to discard any changes and exit. 

6.2.33 Set Noise Filter Offsets 

 
The Noise Filter offset allows setting the noise filter level from –5 to +5 magnitude zones relative to the 
signal channel independently for each signal channel. The default level for all channels is set to +1 zone 
of noise filter offset. Do not change this value unless necessary. 

Press «Enter» 

Page 76 of 112 
W‐PD60  user’s  manual  Ver6.0 

Use «↑» or «↓» to scroll through the channels. Use «←» or «→» to change the noise filter offset value 
for each channel. After all changes are made, press «Enter» to confirm all changes and exit, or «ESC» to 
discard any changes and exit. 

6.2.34 Set Channel Low Magnitude Limit 

 
The PDI and pulse count are calculated for each channel starting from a low magnitude zone and moving 
up through the remaining zones. This feature prevents background noise from affecting PDI and pulse 
count calculations. The low magnitude limit has to be set for each channel independently based on every 
channel’s own background noise level. Sensitivities should be set for each channel before this 
operation. Refer to Section 6.2.5: Set Sensor Sensitivity 

There are two options to set these limits – manual and automatic. 

1. Manual: Press «Enter», use «←», «→», «↑» or «↓» to choose the manual mode as shown below: 

Press «Enter», 

Use «↑» or «↓» to scroll through the channels. Use «←» or «→» to scroll through the available values 
in dB as shown in the following table. After all changes are made, press «Enter» to confirm all changes 
and exit, or «ESC» to discard any changes and exit. 

Magnitude windows (zone) threshold for W‐PD60 

Magnitude window  Threshold (V)  dB 


  10.000 0 
1  6.894 ‐3.23
2  4.753 ‐6.46
3  3.277 ‐9.69
4  2.259 ‐12.92
5  1.558 ‐16.15
6  1.074 ‐19.38
7  0.740 ‐22.61
8  0.511 ‐25.84

Page 77 of 112 
W-PD60 user’s manual Ver6.0

9  0.352 ‐29.07
10  0.243 ‐32.3
11  0.167 ‐35.53
12  0.115 ‐38.76
13  0.080 ‐41.99
14  0.055 ‐45.22
15  0.038 ‐48.45
16  0.026 ‐51.68
17  0.018 ‐54.91
18  0.012 ‐58.14
19  0.009 ‐61.37
20  0.006 ‐64.6
21  0.004 ‐67.83
 
2. Automatic: In automatic mode the device will measure actual signals from each channel with the 
noise filter set off. It will count all pulses received by the channel and find a magnitude that the total 
number of pulses above this magnitude exceeds 50,000 per second. The device will suggest the next 
magnitude window above this detected level as the low boundary for calculations of PDI and pulse 
count. Press «Enter», use «←», «→», «↑» or «↓» to choose the auto detect mode as shown below: 

Press «Enter» to start a measurement of “background” noise, 

The background noise on all channels summed together is displayed. 

Press «Enter» to measure noise on the individual channels, 

Page 78 of 112 
W‐PD60  user’s  manual  Ver6.0 

Use «↑» or «↓» to scroll through the channels number to see the results. Then press «Enter» to accept 
these values and move to the next setting or «ESC» to discard any changes and exit. 

If “Yes” is selected, an additional sub‐function will align the boundaries for all channels with the same 
sensitivity to the maximum of all detected values. 

Before aligning channels, please compare the results between channels with the same sensitivity. If they 
differ by 3 zones or less, select “Yes” to allow the device to align the values for all channels with the 
same sensitivity. Otherwise, select “No” to leave the values at the detected levels. 

Use «←», «→», «↑» or «↓» switch between “Yes” and “No”. Then press «Enter». 

6.2.35 Set Software Filters 

 
 
These filters have been specifically developed for coupling capacitor sensors on large turbine generators 
equipped with long iso‐phase bus ducts. We do not recommend using these filters for other equipment 
or sensors without consulting an expert. 

Use «↑» or «↓» to scroll through the channels. Use «←» or «→» to switch between filter options: Off, 
1, 2, or 3. After all changes are made, press «Enter» to confirm all changes and exit, or «ESC» to discard 
any changes and exit. 

6.2.36 Set ModBus Protocol 

 
W‐PD60 supports ModBus RTU and ModBus TCP protocol. Press «Enter». 

Page 79 of 112 
W-PD60 user’s manual Ver6.0

Use «←», «→», «↑» or «↓» switch between “RTU” and “TCP”. Then press «Enter». 

NOTE: The ModBus Protocol must be set to “RTU” if the optional communications board is installed. 
The communications board handles the ModBus TCP protocol separately. 

6.2.37 Start a Single Measurement 

 
 
This procedure forces a measurement to start immediately. The data will be saved in Full mode. 

Once «Enter» is pressed, the single measurement starts with the following screen displayed during the 
measurement: 

When the measurement completes the display returns to the set menu. 

6.3 Accessing Memory via Keypad 
6.3.1 Clear All Memory 

This operation will erase all data saved in memory. This is typically only necessary when W‐PD60 is 
placed in service on a new monitored object. 

Press three keys simultaneously ‐ «←», «→» and «↑»: 

Use «←», «→», «↑» or «↓» switch between “Yes” and “No”, then press «Enter». If “No” is selected, 
the screen returns to normal status. If “Yes” is selected, then 

Page 80 of 112 
W‐PD60  user’s  manual  Ver6.0 

After all records are cleared, 

Press «Enter», the screen returns to normal status. 

6.3.2 Check Data Saved in Memory 

Press «Memory» key on the keypad, then the number of records is shown, 

Press «Enter», 

Use «←» or «→» to move between the digits for Year, Day or Month respectively, use «↑» or «↓» to 
change the date, then press «Enter». If there is no data on that day, the data recorded at the time 
closest to that date will be shown. 

Use «↑» or «↓» to scroll through data recorded at different times. When the date and time of the 
desired data is shown press «Enter» 

Use «↑» or «↓» to scroll through the data for different channels, and, use «←» or «→» to switch 
between Qmax and PDI (%PD). Please refer to Section 6.2.4 Set Display Parameters for details on the 

Page 81 of 112 
W-PD60 user’s manual Ver6.0

displayed message meaning. Press «Enter» again to return to the previous menu and choose data from a 
different date/time. Press “ESC” to exit Memory mode and return to normal status. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Page 82 of 112 
W‐PD60  user’s  manual  Ver6.0 

7 Appendix II Synchronization and Phase Reference Shift 
One should clearly distinguish between W‐PD60 synchronization and phase reference shift. Phase 
reference shift is a phase angle between the W‐PD60 synchronization signal and the reference signal 
from the phase A voltage of the monitored object. 

While the first is a vital part of taking PD reading, the second is used for correct phase‐resolved data 
representation in W‐PD60 software and for analysis associated with this phase‐resolved data. 

Loss of synchronization stops W‐PD60 operation and causes the display “No Synch. Signal”. 

Incorrect phase reference shift doesn’t affect data acquisition by W‐PD60, integral quantities calculations 
and alarm operation. Even more, establishing the correct phase reference shift later will even provide 
correct phase‐resolved data analysis of the data acquired earlier. 

The correct phase reference shift should be set regardless of the selected synchronization option. It does 
not matter how this number is obtained. One may use W‐PD60, a phase meter, oscilloscope or other 
instrumentation. 

7.1 Synchronization 
W‐PD60 allows external synchronization or internal synchronization from its 115/230VAC power supply. 
Internal synchronization is the most reliable option since it stops operation only if the power is off. This is 
the default option for W‐PD60. 

External synchronization is used only when it is really necessary, when the power supply voltage is not 
synchronous with the voltage on the monitored object, e.g. for Variable Frequency Drive (VFD) 
applications. 

Be cautious when powering W‐PD60 from a UPS. It is acceptable only when the UPS has a constant 
phase shift to the voltage on the monitored object in both normal and standby modes. If not, W‐PD60 
will be “asynchronous” to the monitored object and cannot make reliable measurements. If this type 
of UPS power is absolutely necessary external synchronization must be used. 

Two jumpers JU1 & JU2 inside W‐PD60 enclosure (on the right side from the power supply transformer) 
are used to change the synchronization type. By default they are positioned for internal synchronization 
as shown in Figure 7‐1a. They should be moved to the upper position for external synchronization as 
shown in Figure 7‐1 b. 

Page 83 of 112 
W-PD60 user’s manual Ver6.0

 
Figure 7‐1 Jumpers JU1 & JU2 position for a) Internal and b) External synchronization. 

The external synchronization signal should be connected to the W‐PD60 Sensor Interface Board. 

There are two external synchronization options in the Sensor Interface Board: 

  Using the signal from a coupling capacitor installed on phase A and connected to PD channel #1. The 
coupling capacitor signal contains both high frequency PD components and power frequency 
components. The power frequency component may be used for synchronization. For this 
synchronization option jumper J1 has to be installed as shown in Figure 7‐2 b, the default position. 

  Using an external signal source, e.g. from a potential transformer (PT) on phase A. This signal should 
be connected to the “Ref” # 7 terminal and the ground wire should be connected to terminal #9. The 
signal should be in the range of 0.1‐1VAC. If necessary, it should be externally conditioned, e.g. resistive 
divider as shown in Fig 7a. Note that “Ref” input #7 has a 27k_ impedance. For this synchronization 
option jumper J1 (Figure 7‐2 b) has to be removed or be in a disconnected position. 

Figure 7‐2 a‐ Divider for reference voltage from PT connected to the Sensor Interface Board, b – 
Jumper J1, c – Switch position for phase reference shift measurement. 

7.2 Entering Typical Phase Reference Shift for External Synchronization 
 

Page 84 of 112 
W‐PD60  user’s  manual  Ver6.0 

The table below presents phase angles for the most common external synchronization options. These 
numbers should be manually entered into W‐PD60 from the front panel keypad as described in section 
6.2.21: Set Phase Shift, and synchronized with software at W‐PD60 startup and commissioning. 

 
External Synchronization Signal from    Phase Shift
Channel 1 (Coupling Capacitor on Phase A)  0 
PT Phase A  0 
PT Phase B  +240
PT Phase C  +120 
PT Phase A‐B (Open Delta)  +30 
PT Phase C‐B (Open Delta)  +90
 
VFD applications uses a modified sensor interface board with additional filtering. The phase shift slightly 
depends on frequency and varyies from 277 to 273 in 40‐60 Hz frequency range. For typical applications 
50Hz +‐10Hz enter 275 degrees. 

7.3 Determining and Entering Phase Reference Shift 
W‐PD60 uses internal synchronization from its 115/230 VAC power source as the default. If the phase 
shift between this voltage and phase A voltage of the monitored object is known it can be manually 
entered using software or the front panel keypad (section 6.2.21 Set Phase Shift). If it is unknown 
W‐PD60 can be used to determine this phase shift in the setup mode from its keypad. The phase 
reference shift can be measured from either of two sources: 

  using the reference signal from PD channel # 1 (if a coupling capacitor is connected on phase A), 

  using an external reference signal, which should be temporarily connected for this measurement 

to the terminal # 7 and terminal # 9 (Figure 7‐2 a) as described above for external synchronization (don’t 
forget to temporarily remove J1 and plug it back in upon completion of this measurement). 

For the phase shift measurement temporarily set the switch on the sensor interface board to the “Ref” 
position (Figure 7‐2 c). 

Don’t forget to return it to the H% position (Humidity measurement mode) upon completion! 

From the keypad enter the setup mode as described in 6.1: Keypad. Scroll through the setup options 
until “Set Phase Shift” is display as shown below: 

Press «Enter» 

Page 85 of 112 
W-PD60 user’s manual Ver6.0

Use «←», «→», «↑» or «↓»to switch to auto detect mode as shown below 

Press «Enter» to start a phase angle measurement between W‐PD60 power source voltage and the AC 
signal brought to the W‐PD60. The display will be updated with a measured value once every 15 seconds. 

Normally measured value deviates within the range of +/‐ 50. Usually you need to observe W‐PD60 
display for 1‐2 minutes and derive average of displayed angles. 

To manually input the phase shift, press «Enter» at manually input menu, 

Use «←» or «→» to switch to different digits, and use «↑» or «↓» to change the value. After all 
changes are made, press «Enter» to confirm and exit. 

 
 
The table below shows correction angle for the most common phase reference signal options. 

Phase Reference Option  Additional Phase Shift
Channel 1 (Coupling Capacitor on Phase A)  0 
PT Phase A  0 
PT Phase B  ‐240
PT Phase C  ‐120
PT Phase A‐B    +30 
PT Phase C‐B    +90
 
A number that represents the difference must be added to the calculated value (pay attention to signs). 

New number should be adjusted to 00‐3600 range by adding or subtracting 3600 and resultant number 
manually entered into W‐PD60 (see 6.2.21 Set Phase Shift). 

Don’t forget to return switch into H% position (Humidity measurement mode) upon completion! 

Note: Normally measured value deviates within the range of +/‐ 50. Below are some hints that may 
identify incorrect phase angle measurement: 
  Phase angle varies on more than +/‐100 (low reference signal magnitude or high noise). 

Page 86 of 112 
W‐PD60  user’s  manual  Ver6.0 

  W‐PD60 displays dashes (reference signal is likely close to zero). 
  IG display shows continuously 357 (reference signal magnitude is too high). 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Page 87 of 112 
W-PD60 user’s manual Ver6.0

8 Appendix III Communications Board Interface 
The communications board is an optional part when W‐PD60 is being ordered. Sections 8 9 and 10 can 
be ignored if W‐PD60 is not equipped with the communications board. 

8.1 General 
The W‐PD60 Communications Board provides access through USB and Ethernet interfaces. The USB 
Device interface provides a standard serial communications port to connect to a local PC. This interface 
is typically used for local access of W‐PD60 data using the W‐PD60 PC software as well as for upgrading 
firmware in the communications board. 

The USB Device Driver must be installed on the PC accessing the W‐PD60 prior to using the USB Device 
interface. (See Section 9 Appendix IV USB Device Driver Installation for complete installation 
instructions). Once driver is installed, a serial communications port such as COM5 should be assigned to 
this interface (the exact port number will vary depending on what other devices are installed on the PC). 
When accessing the W‐PD60 using the W‐PD60 PC software simply configure this serial port number on 
the Communication tab of the software. 

NOTE: If the W‐PD60 or the PC accessing the W‐PD60 is rebooted or turned off and back on, the USB 
cable from the PC to W‐PD60 must be disconnected and reconnected to reestablish the link. 

The USB Host is reserved for future use. 

The Ethernet interface supports the following network protocols: 
∙ Modbus TCP/IP on Port 502 for interface to 3rd party building management systems. 
∙ Web Server/HTTP on Port 80 providing on‐board web pages viewable with standard browsers such as 
Microsoft Internet Explorer (version 7 recommended) or Mozilla Firefox (version 2 or later 
recommended). 
∙ SMTP on Port 25 for sending Email across the Internet. 
∙ FTP Server on Port 21 to download W‐PD60 database file on demand. 
∙ FTP Client to upload W‐PD60 database file to a server at a specified time/interval. 
∙ SSH Secure Shell on Port 22 for certain administrative tasks. 
∙ Virtual Communication Port for simplified access by the W‐PD60 PC software. See Section 10 Appendix 
V Ethernet Virtual Communications Port Driver Installation for driver installation. 
 

8.2 Network Configuration 
The Ethernet Interface requires general configuration to provide access to the Internet for all protocols 
as well as specific configurations for several protocols. Prior to configuration provide your local network 
administrator with the MAC address of the Ethernet port located on the label on the side of the unit and 
request the following: 
∙ MAC Address (copy from label) _________________________ 
∙ IP Address (static address*) _________________________ 
∙ Subnet Mask _________________________ 

Page 88 of 112 
W‐PD60  user’s  manual  Ver6.0 

∙ Gateway Address _________________________ 
∙ DNS #1 (name server) IP Address** _________________________ 
∙ DNS #2 (name server) IP Address** _________________________ 
∙ DNS #3 (name server) IP Address** _________________________ 
∙ Domain ____________________ 
*If you are setting up the W‐PD60 in a dynamic host configuration protocol (DHCP) environment the 
local network administrator must set up the DHCP server to provide a static address each time the card 
makes a DHCP request. 

** Obtain the Domain Name Server (DNS) IP address and Domain if you plan to use host names. The 
number of available nameservers will vary depending on your network. In a DHCP environment these 
may be populated automatically when connecting to the network. 

If you need e‐mail functionality, obtain the SMTP IP address or host name and determine if you need 
login/password information to access the server. If using the SMTP host name, verify that you also have 
the DNS IP address above. 

SMTP (mail server) IP Address or Host Name _________________________ 

Once you have the required information configure your PCs LAN connection TCP/IP properties with IP 
address 192.168.1.2, Subnet Mask 255.255.255.0 and connect the W‐PD60 to your PC with a CAT5 
crossover cable. Note that newer computers may have “autosensing” Ethernet adapters that will work 
with a standard CAT5 cable and don’t require a special crossover cable. 

Open your web browser and enter W‐PD60’s default IP address 192.168.1.1 on the address bar select Go. 
A login screen will appear similar to Figure 8‐1 Typical login screen. The default User name and Password 
for read‐only access is user, and for configuration (read/write) access is admin. Log in using admin and 
select OK. 

 
 
Figure 8‐1 Typical login screen 

Page 89 of 112 
W-PD60 user’s manual Ver6.0

To ensure the security of your system, immediately change the user and admin passwords after the 
initial login. 

Page 90 of 112 
W‐PD60  user’s  manual  Ver6.0 

9 Appendix IV USB Device Driver Installation 
The following instructions describe the installation procedure for the USB Device driver. 

1. Copy the ‘gserial.inf’ file from the Gadget_Serial_Driver directory of the W‐PD60 installation CD into 
“C:\WINDOWS\ system32 \drivers” directory. 

2. Make sure usbser.sys file exists in the “C:\WINDOWS\ system32 \drivers” directory. If file does not 
exists, then copy this file from one of the following location (Locations are specific to Windows XP, 
Service pack2) 

∙ C:\WINDOWS\Driver Cache\i386\SP2.cab or 

∙ The Windows installation disk (Typical location of the file on Installation CD is “/I386/SP2.cab” for 
service pack2.) 

3. Connect the communications board’s USB device interface to PC’s USB host connector. 

4. Gadget serial device should automatically be detected and the following window will be displayed. 

1. Select Install from a list or specific location (Advanced), click Next, and the following window will be 
displayed. 

Page 91 of 112 
W-PD60 user’s manual Ver6.0

2. Select Don’t search. I will choose the driver to install., click Next, and the following window will be 
displayed. 

Page 92 of 112 
W‐PD60  user’s  manual  Ver6.0 

3. Select Ports (COM & LPT), click Next, and the following window will be displayed. 

4. Click Have Disk… and the following window will be displayed. 

5. Click Browse… to select the driver from the location we copied in step 1. The following window will 
be displayed. 

Page 93 of 112 
W-PD60 user’s manual Ver6.0

6. Navigate to “C:\WINDOWS\ system32 \drivers” directory, select gserial and click Open. The 
following window will be displayed. 

7. Click OK and the following window will be displayed. 

Page 94 of 112 
W‐PD60  user’s  manual  Ver6.0 

8. Click Next and the following window will be displayed. 

9. Click Continue Anyway and the following windows will be displayed. 

Page 95 of 112 
W-PD60 user’s manual Ver6.0

10. Click Finish to complete the installation 

11. To determine the comm port assigned to this USB interface open Windows Control Panel and choose 

Page 96 of 112 
W‐PD60  user’s  manual  Ver6.0 

System/Hardware/Device Manager. Click on the + beside Ports (COM & LPT) to see the available 
ports and note the port number assigned to Gadget Serial. The following window is an example 
showing Gadget Serial assigned to COM5. Make a note of this number as you will need it to configure 
the W‐PD60 PC software to communicate through this USB interface. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Page 97 of 112 
W-PD60 user’s manual Ver6.0

10 Appendix V Ethernet Virtual Communications Port Driver 
Installation 
The following instructions describe the installation procedure for the Virtual Serial Port driver. 

NOTE (Windows Vista/Windows Server 2008): Before installing/uninstalling the com0com driver or 
adding/removing/changing ports the User Account Control (UAC) should be turned off (requires reboot). 

NOTE (x64‐based Windows Vista/Windows Server 2008): The com0com.sys is a test‐signed kernelmode 
driver that will not load by default. To enable test signing, enter the following command: bcdedit.exe ‐set 
TESTSIGNING ON and reboot the computer. 

NOTE Turning off UAC or enabling test signing will impair computer security. 

10.1 Installing com0com Null modem emulator 
Run setup.exe from the com0com directory of the W‐PD60 installation CD. 

At the "Welcome to the Null‐modem emulator (Com0Com) Setup Wizard" screen, click Next to start the 
software installation. 

The "License Agreement" screen will be displayed. Click I Agree to continue the installation of the 

Page 98 of 112 
W‐PD60  user’s  manual  Ver6.0 

software. 

The "Choose Components" screen will be displayed. Do not change the default settings. The bottom 
option, CNCA0<‐>CNCB0 represents the definition of a default Virtual Port Pair and should be left 
checked so that one new com port pair is created during installation. Notice that the Virtual Port Pair 
comport names do not have the standard naming convention such as COMx, where x is the comport 
number. Do not worry about this. We will be changing the Virtual Port Pair’s comports name. Click Next 
to continue the software installation. 

Page 99 of 112 
W-PD60 user’s manual Ver6.0

The "Choose Install Location" screen will be displayed. It is recommended that you use the default 
location. Click Install to continue the software installation. 

Page 100 of 112 
W‐PD60  user’s  manual  Ver6.0 

The software installation process will begin. You can monitor the installation progress. Once it is 
complete, the "Installation Complete" screen is displayed. Click Next to continue the software 
installation. 

Since you installed the default Virtual Port Pair as part of the software installation, Windows will 
recognize that new hardware has been discovered. 

Note: Since there are two comports defined in each Virtual Port Pair, the following hardware installation 
process will occur twice for every Virtual Port Pair you define. 

When the "Welcome to the Found New Hardware Wizard" screen is displayed, if it asks for the “Can 
windows connect to windows update to search for the software” select the "No, not this time" and click 
Next. 

Select "Install the software automatically (recommended)" option and click Next. 

Page 101 of 112 
W-PD60 user’s manual Ver6.0

The "Please wait while the wizard searches" progress screen will be displayed while the driver is 
installed. 

Page 102 of 112 
W‐PD60  user’s  manual  Ver6.0 

Once the driver installation is complete, you will receive the "Completing the Found New Hardware 
Wizard" confirmation screen. Click Finish to complete the installation of the comport driver. 

You should see an indication in the bottom right hand portion of the screen that the new hardware has 
been found and is ready to use indicating that the comport driver successfully installed. Since this is just 
the first comport of the Virtual Port Pair that has been installed, immediately following the installation of 
the first comport, the second comport will start its installation. Follow the instructions above for the 
second comport installation. 

After both comports of the Virtual Port Pair have been installed, you must click on the Finish button to 
complete the Com0Com software installation 

Page 103 of 112 
W-PD60 user’s manual Ver6.0

 
 
 
 
 

10.2 Configuring com0com Null Modem Emulator 
After the comport hardware drivers have been successfully installed, it is time to change the com port 
name of the two comports in the pair to names that W‐PD60 PC and other software packages recognize 
so they can be used. 

The reason Com0Com uses this unique comport naming convention is to prevent newly created 
comports from conflicting with existing virtual and physical com ports already defined on your computer. 

NOTE: Microsoft .NET Framework 2.0 is required to run the Com0Com Setup program. If your PC does 
not have “Microsoft .NET Framework 2.0”, you can download it from the Microsoft site. Or use the 
command line utility described in next section to configure the com0com. 

To start the Com0Com Setup application, Click on the Windows Start button and run the setup program 
from "Programs‐>com0com‐>setup" 

Below is the "Setup for Com0Com" application screen. Note that there is already one Virtual Port Pair 
defined using comports CNCA0 and CNCB0. 

Page 104 of 112 
W‐PD60  user’s  manual  Ver6.0 

Now we need to change the names of the individual comports in the Virtual Port Pair to names that are 
recognizable by other software programs. 

NOTE: Enter the comport names of the Virtual Port Pair in capital letters to ensure the greatest 
compatibility with other programs using the Virtual Port Pairs. 

Highlight the Virtual Port Pair where you want to change the two comport names. Once highlighted, 
change the com port names in the two text boxes in the top right hand side of the Setup screen. In the 
example below, Virtual Port Pair 0 has been selected and CNCA0 has been changed to COM9 and CNCB0 
is left as it is, because we do not want this port be used by any other software than our com port 
redirector (com2tcp). The new port names will be in blue until they have been activated. Although COM9 
is shown in this example, you may use any available COM port. If you have just installed the USB Device 
Driver you can probably use the next sequential COM port. 

Activate the “Enable plug in mode” for COM9 and “Enable exclusive mode” for CNCB0 by clicking on the 
associated check box as shown below. Ensure all other check boxes are cleared. 

Page 105 of 112 
W-PD60 user’s manual Ver6.0

Apply these settings by clicking Apply. It will take several seconds to reconfigure both comports in the 
Virtual Port Pair so please be patient. Once the change is completed, the com port names will be 
displayed in black. 

Click the X button in the upper right corner button to close this application. 

10.3 Configuring com0com (using command line utility) 
If you have completed the com0com setup as described in above section, skip this section. This section is 
provided only if you could not run the com0com setup program described above, because 
“Microsoft .NET Framework 2.0” is not installed on your PC. 

To start the Com0Com command line setup application, Click on the Windows Start button and run the 
setup program from "Programs‐>com0com‐>Setup Command Prompt" 

Then type the following commands on the “command>” prompt one by one. 

change CNCA0 PortName=COM9 

change CNCA0 PlugInMode=yes 

change CNCB0 ExclusiveMode=yes 

quit 

Page 106 of 112 
W‐PD60  user’s  manual  Ver6.0 

Note: COM9 is selected assuming it is unused com port; user can select any unused port name. 

10.4 Com port redirector installation 
Run setup.exe from the com2tcp directory of the W‐PD60 installation CD. 

Click Next, which will take you to the license page, then click I Agree button on this page. 

Page 107 of 112 
W-PD60 user’s manual Ver6.0

Do not deselect any components, click Next 

Default installation directory is “C:\Program Files\com2tcp”, do not change the installation path, as we 

Page 108 of 112 
W‐PD60  user’s  manual  Ver6.0 

will be using this path in a batch file. Click Install. 

The software installation process will begin. You can monitor the installation progress. Once it is 
complete, the "Installation Complete" screen is displayed. Click Next to complete the installation, then 
Finish to close the final window. 

Page 109 of 112 
W-PD60 user’s manual Ver6.0

Page 110 of 112 
W‐PD60  user’s  manual  Ver6.0 

10.5 Setting Environment variables 
Two Environmental Variables must be created prior to running the com2tcp application. 

COM2TCP_SERVER_IP: Defines the IP address of the W‐PD60 that you will be communicating with. 
COM2TCP_SERVER_PORT: Defines the port used to communicate with the W‐PD60 with this virtual serial 
port application. 

NOTE: The following instructions apply to Windows XP Professional Edition. 

From the start menu select Control Panel and then System to open the System Properties window. 

Click on the Advanced tab of this window, and then the Environmental Variables button. 

Under the User variables section click New and type the following: 

Variable Name: COM2TCP_SERVER_IP 

Variable Value: 192.168.1.1 

Replace the above IP address with your actual W‐PD60 device IP address. 

Click OK, and then New again under the User variables section and type the following: 

Variable Name: COM2TCP_SERVER_PORT 

Variable Value: 49152 

Click on the OK button of the Environmental Variables window, and then the OK button of the System 
Properties window. 

NOTE: Windows Vista has different procedure to set the environment variable You cannot modify user 
environment variables in the System Properties dialog box if you log on by using a standard user account 
in Windows Vista 

For example, if you try to access the System Properties dialog box by clicking Advanced system settings 
in the System item in Control Panel, you are prompted for administrator account credentials. If you type 
the credentials for an administrator account, the user environment variables that you can access are for 
that administrator account only. 

To resolve this issue, modify the user environment variables by using the User Accounts item in Control 
Panel. To do this, follow these steps: 

1. Click Start, type Accounts in the Start search box, and then click User Accounts under Programs. If 
you are prompted for an administrator password or for a confirmation, type the password, or click 
Allow. 

Page 111 of 112 
W-PD60 user’s manual Ver6.0

2. In the User Accounts dialog box, click Change my environment variables under Tasks. 

3. Make the changes that you want to the user environment variables for your user account, and then 
click OK. 

10.6 Running Com port redirector application 
Once installation and configuration described in above sections is complete com_port_redirector batch 
file will be available under your windows start menu. 
“Start‐>Programs‐>com2tcp‐ >com_port_redirector(com2tcp)”. 

You must run this batch file before the W‐PD60 PC software is run. When your run this batch file a 
command prompt window is opened. This window must remain open while using the virtual comm. port, 
but may be minimized if desired. 

NOTE: If com_port_redirector encounters a problem a Notepad window will open titled com2tcp_log.log. 
This information in the window will help troubleshoot the problem (see section “10.7 Troubleshooting”). 

After the com_port_redirector(com2tcp) has successfully started, you can start the W‐PD60 PC software. 
Once you choose the desired W‐PD60 go to the Communication section and program in the Serial Port # 
(9 in this example) that we configured in section “10.2 Configuring com0com Null Modem Emulator”. 

Note that virtual com port (COM9) will be available only if the com_port_redirector has successfully 
started. 

When finished communicating with the W‐PD60, you may close the com_port_redirector window if 
desired. 

10.7 Troubleshooting 
∙ I can not configure the COM9 as a com port as it is already in use. 

COM9 port is selected in this document assuming it is unused port. If your PC is already using COM9 for other 
application, select any unused port. 

∙ When I start com_port_redirector batch file, it runs for few seconds and then displays the “connect() 
error” in a com2tcp_log.log file. 

com_port_redirector application is not able to connect to the “Modbus RTU over Ethernet” data server 
that is running on W‐PD60. Check the following things 

1. W‐PD60 device is up and running and is connected over Ethernet to your PC (directly or through local 
network). Software version running on the W‐PD60 is greater than or equal to 1.0.5 (you can find this 
from web page “W‐PD60” link) 

Page 112 of 112 
W‐PD60  user’s  manual  Ver6.0 

2. IP address that you configured in environment variable COM2TCP_SERVER_IP is the correct IP address 
of W‐PD60 device. Also check the Name (spelling) of the environment variable itself. 

3. Batch file window displays the command that it runs, which displays the IP address that it is trying to 
connect. Check this IP address is correct. 

4. If all the above checks are fine, make sure you can ping the W‐PD60 from you PC. You can test this by 
opening a command window and typing ping 192.168.1.1 

(Replace above IP address with actual W‐PD60 device IP address). If ping is not working, then there is a 
network issue, which needs to be resolved first. 

∙ When I start com_port_redirector batch file, it immediately opens up the com2tcp_log.log file which 
says Connect OK and then Disconnected OK. 

“Modbus RTU over Ethernet” data server running on W‐PD60 can handle only one connection from PC. 
Above condition will occur if there is already one connection active. Make sure there is no 
com_port_redirector application running from another PC. 

Page 113 of 112 

You might also like